Professional Documents
Culture Documents
Associate Consultant
Department of Intensive Care
PYNEH
28 Sep 2007
Gas exchange
Pulse oximetry
Capnometry
Continuous blood gas analysis
Transcutaneous monitoring (PTCO2
and PTCCO2)
Question
Which patient is more hypoxemic, and why?
A 7.48 34 85 95 7
B 7.32 74 55 85 15
Answer: only 5
• 1 affects only content, not oxygen saturation or PO2.
• 2 through 4 affect only oxygen saturation and content,
not PO2
Important
• For example
SpO2 95% can represent true SaO2 of
For example:
• True high SaO2 (>85%) condition: MetHb results in a falsely low
SpO2
• True low SaO2 (<85%) condition: MetHb results in a falsely high
SpO2
Etiology
Fe within Hb is oxidized from
the ferrous (Fe2+) state to
the ferric (Fe3+) state
Effects:
• formation of methemoglobin
• inability to transport oxygen
and carbon dioxide
• brownish discoloration of the
blood
Causes of MetHb
Acquired:
• Exposure to Nitrites, Aniline dyes, Silver nitrate,
Nitroprusside, Antimalarials;
• Local anesthetics (Benzocaine, prilocaine, and lidocaine,
particularly when applied to mucosa, such as during
bronchoscopy, or after repeated cutaneous exposure to
eutectic mixture of lidocaine-prilocaine (EMLA(R) cream)
over a short period of time;
• Nitric and nitrous oxides;
• Vegetables (eg, spinach, beets, carrots) inadequately
cooked or contaminated with bacteria
Hereditary: deficiency of NADH cytochrome b5
reductase or NADPH-flavin reductase or the
presence of hemoglobin M.
The patient had been wrongly given sodium nitrite instead of the herbal NatriiSulfas
(chiefly NaPO4) as a laxative for clearing heat and decreasing edema.
Low perfusion-resistant pulse oximeter
Perfusion Index (PI) with trending capability indicates arterial pulse signal
strength and may be used as a diagnostic tool during low perfusion.
Pleth Variability Index (PVI): captures vital thoracic pressure changes that may
compromise normal cardiac function affecting systemic circulation
Accurate on cyanotic patients.
Signal IQ® waveform for signal identification and quality indication during
excessive motion and low signal to noise situations.
Transcutaneous monitoring
Measures O2 and CO2 diffusing
through the skin
Mainly used in infants (NICU,
PICU)
Relis on the oxygen content of
capillary blood
agrees well with arterial blood
pO2 when tissue perfusion is
adequate, but not in states of
hypoperfusion
measured by heating skin locally
to dilate capillaries
The heat emitted by the electrode
may cause areas of redness on
the skin. Hence, the site of
placement of the sensor needs to
be changed regularly.
Continuous blood gas
monitoring
A fiber optic sensor with three
sensing elements for monitoring pH,
PCO2, and PO2 plus a thermocouple
for measuring temperature.
Length: 30 cm; diameter <0.5 mm,
dead space 6/10,000 of a milliliter
A Y connector on the sensor allows
simultaneous continuous blood
pressure monitoring and enables
intermittent withdrawal of blood
samples and infusions.
Performance in the clinical setting
was not as satisfactory, especially
for PO2 values. (Ganter M and
Zollinger A. Br J Anaesth 2003)
Mainstream analyzer
• sampling window inside the ventilator
circuit for CO2 measurement
Sidestream analyzer
• aspirates gas from the ventilator circuit,
and the analysis occurs away from the
ventilator circuit
Mainstream CO2 sensor
During exhalation, exhaled gas
passes directly over the sensor
Designed primarily for intubated
patients
Difficult to use in nonintubated
patients because
1. it requires a mouthpiece or
mask that patients in
respiratory distress find
uncomfortable
2. In addition, when a mask is
used, any supplemental
oxygen must be delivered at
a flow rate of greater than 6
Lpm to ensure the patient
does not rebreathe CO2 that
can accumulate in the mask
at lower oxygen flow rates.
3. extra weight of the sensor
dragging on the
endotracheal tube
Sidestream
In sidestream capnography, a sample of exhaled
gas is aspirated from the patient’s airway
interface into the monitor, which houses the
sensor.
Designed for use in both intubated and
nonintubated patients.
Prone to obstruction with moisture.
Microstream technology is a unique low-flow (50
cc/min) system that permits precise
measurement of CO2 levels without problems of
dilution or moisture accumulation.
Normal capnogram
Expiration:
1. A or phase 1: Carbon dioxide cleared
from the anatomic dead space
2. B or phase 2: dead space and
alveolar carbon dioxide
3. C or phase 3: alveolar plateau
4. D or phase 4: end-tidal carbon
dioxide tension (PETCO2): normal
range of ETCO2 is 35 – 45 mm Hg .
Inspiration
Inhaled PCO2 is zero (the amount in
inhaled air is negligible)
DISLODGED ETT: Loss of
waveform, Loss of ETCO2
reading
Esophageal intubation:
Small CO2 spikes
(MIP)
Maximal Expiratory Pressure (MEP)
Transdiaphragmatic pressure
Airway Occlusion Pressure
= Negative pressure generated 0.1 sec following onset of
inspiration against an briefly and surreptitiously occluded
airway
Measured by inserting a shutter in the ventilator's inspiratory
line as near as possible to the patient, and recording airway
pressure tracing at the Y piece by maintaining occlusion only
for the first 200 to 300ms of inspiration
Why first 100 ms?
• chosen because a normal subject requires at least 150 ms to
sense the occlusion and react against it
P0.1 measurement
Properties
Unaffected by properties of the respiratory system
Does not require scaling to patient size
Normal values: 1 cmH2O
A measure of neural respiratory drive (central drive)
• P0.1 is maintained following curare-induced muscular
weakness
• However, may be reduced in the presence of severe
inspiratory muscle weakness when capability to produce
pressure is profoundly impaired (consider the P0.1/MIP
ratio in such case)
Uses
Predicts weaning failure
• “High” levels of P0.1 (i.e. more negative
than – 4 to -6 cmH2O) are associated
with increased respiratory effort and
indicate an inability to breath
independently with success
• lower values of P0.1 are associated with
effective weaning.
closely correlated with the work of
breathing
Eight patients recovering
from acute respiratory
failure of various causes
Respiratory Mechanics
Primary measurements
• Pao, Pes, Flow, auto-PEEP)
Derived measurements
• Static compliance, Dynamic compliance,
Pressure-volume curves, Resistance