Professional Documents
Culture Documents
Measurement Instrument
1
2
Physiology of respiration
Oxygen/Carbon dioxide interaction: Metabolism
breath
CO2 produced by cellular metabolism diffuses across the
CO2 cell membrane into the circulating blood. muscles + organs
lungs
5-10% carried in solution Oxygen
CO2 20-30% bound to haemoglobin
60-70% carried as bicarbonate in the red blood cell
cells
blood energy
Oxygen
+
Glucose
CO2
3
4
Respiratory Monitoring
• Rapid progress with greater safety in Anesthesia field and better ICU outcome.
5
Blood ➔ RBC ➔ Hemoglobin
6
Inadequate Oxygen Transport
• Anemia
• Reduces red blood cells reduce oxygen carrying capacity
• Inadequate hemoglobin (Hemoglobin is the protein inside red blood cells. It
carries oxygen. Red blood cells also remove carbon dioxide from your body,
bringing it to the lungs) results in the loss of oxygen saturation
• Poisoning
• Carbon monoxide on-loads on the hemoglobin more readily preventing
oxygen saturation and oxygen carrying capacity
• Shock
• Low blood pressures result in inadequate oxygen carrying capacity
7
Pathophysiology
• Oxygen is exchanged by diffusion from higher concentrations to lower
concentrations
• Most of the oxygen in the arterial blood is carried bound to hemoglobin
• 97% of total oxygen is normally bound to hemoglobin
• 3% of total oxygen is dissolved in the plasma
8
Terminology
• SaO2 : Arterial (invasive)Oxygen Saturation (oxygen bound to the hemoglobin
molecules)
• SpO2 indicates the oxygen bound to hemoglobin
• Non invasive oxygen saturation. SpO2 indicates the saturation was obtained with non-
invasive oximetry
• Closely corresponds to SaO2 measured in laboratory tests
• The key difference between SAO2 and SPO2 is in the type of measurement of O2
levels in the blood. SAO2 or Saturation of Oxygen is the direct measurement of O2
bound to heme protein of hemoglobin in the blood. SPO2 measurement or
oximetric measurement of O2 bound to hemoglobin is an indirect measurement
of saturation of haemoglobin with O2.
• SAO2 and SPO2 are measured during different disease conditions to monitor the
level of available O2 in hemoglobin. The saturation levels of O2 bound to
hemoglobin will provide details regarding the efficiency of the alveolar lung
system.
• PaO2 : Arterial Partial Pressure, oxygen dissolved in the plasma (only about 3% of
total content) or PO2
• CaO2: Total amount of oxygen in the blood or the (SaO2 + PaO2).
9
Oxygen Saturation
• Percentage of hemoglobin saturated with oxygen
• Normal SpO2 is 95-98%
• 97-100% sat :Good gas exchange .
• 90-95% sat: Mild hypoxia
• <90% sat : Severe hypoxia
10
Saturation O2
11
PULSE OXIMETRY: WHAT DOES IT DO?
• MEASURES/DISPLAYS
• O2 SAT OF HbG
• PULSE RATE
• INDICATES PERFUSION
• PULSATE FLOW
12
Various forms of pulse ox’s
13
Pros of Pulse Oximeters
PROS
• Non-invasive
• Allows continuous measurement in real time
• Easy to use
14
Cons of Pulse Oximetry
CONS
• Measures Hb saturation rather than the actual level of Hb. Only
measures oxygenation status.
• Does not detect carbon dioxide levels in the blood. CO2 determines
the ventilation status.
• Measurements are not always accurate. Inaccuracy may occur due
to nail polish, light interference, poor peripheral perfusion,
intravenous dyes, the presence of carboxyhemoglobin and
hemoglobinopathies.
15
Pros/Cons of an arterial blood gas
PROS CONS
• Accurate • Invasive
• The gold standard for • Not easy to perform on a
measuring respiratory status patient
• Does not reflect measurements
in real time status
16
Photospectrometry
17
18
Pulse Oximetry
• The pulse oximeter has Light-emitting diodes (LEDs) that produce red and
infrared light
• LEDs and the detector are on opposite sides of the sensor
• Sensor must be place so light passes through a capillary bed
• Requires physiological pulsatile waves to measure saturation
• Requires a pulse or a pulse wave (Adequate CPR)
Beer s law:
The concentration of a liquid is
exponentially related to the intensity of
light that will pass through it.
* Lambert s Low:
distance
The of light travelled through the
liquid is exponentially related to the
intensity of light that will pass through it.
Oxygenated hemoglobin absorbs a
different wavelength of light than
does deoxygenated blood
20
Pulse Oximetry
Beer-Lambert Law
• The combination of both Beer’s Law and Lambert’s Law
• Beer’s Law – the absorption of light is proportional to the concentration
of a sample
• Lambert’s Law – absorption is proportional to the thickness of a sample
Extinction Coefficient
21
Characteristics of Common Hb Species
Spectrophotometric
Name Symbol Normal (%) Peak (nM)
Oxy O2Hb 97 530
Reduced RHb <1 585.2
Adult HbA 97 530
Fetal HbF 85 NA
Carboxy COHb 2-5% 594.5
Sulf SulfHb <0.5 618
Meth Methb 1.5% 620
22
ABSORPTION SPECTRA
23
Pulse Oximetry
HEMOGLOBIN ABSORBS LIGHT.
- THE ABSORBED LIGHT VARIES WITH:
* OXYGEN SATURATION
* TYPE OF HEMOGLOBIN
* LENGTH OF THE OPTICAL PATH.
ABSORBENCE CAN BE CALCULATED
* EXTINCTION CO-EFFICIENTS
* OPTICAL DENSITY EQUATIONS
* BEERS-LAMBERT EQUATION
24
Pulse Oximetry
First Principle of operation – 1
Infrared absorption by oxygenated and de-oxygenated
haemoglobin at 2 different wavelengths
25
Pulse Oximetry
26
Second principle for pulse oximetry
27
Pulse Oximetry
The variable absorption due to pulse added volume of arterial blood is used to calculate the
saturation of arterial blood
28
Second principle for Spo2
29
Pulse Oximetry
30
Pulse Oximetry
The red and infrared LEDs within the probe are driven in
different ways, depending on the manufacturer. Most probes
have a single photodetector (PIN-diode), so the light sources
are generally sequenced on and off. A typical pulsing scheme of
the LEDs is shown in Fig. 10.7. To compensate for ambient light
during the time when both LEDs are off, the light level is
measured and then subtracted from each light channel
between cycles. This minimizes the effects due to
ambient conditions which may vary during monitoring.
Depending on the make and model of pulse oximeters, the
drive currents of LEDs, pulse widths, off and on cycles between
pulses and cycle times can all vary (Ackerman and Weith,
1995).
31
Pulse Oximetry
• Main Limitations of SPo2 :
- Ambient light
- Patient movement or shivering.
- Hypothermia.
- Peripheral shut down.
- Hypovlemia and shock.
- Carbon monoxide poisoning(carboxy HB).
- Other dysfunctional Hemoglobins(met HB).
- Skin pigmentation.
- Dye injection(methylene blue). Elaborated in
subsequent slides…
32
Patient Environments
• Ambient Light
• Excessive Motion
33
Ambient Lighting
• Any external light exposure to capillary bed where sampling is occurring may
result in an erroneous reading
• Most sensors are designed to prevent light from passing through the shell
• Shielding the sensor by covering the extremity is acceptable
34
SOURCES OF ERROR
• Sensitive to motion
• Standard deviation is certified to 4% down to 70% saturation
• Sats below 85% increase the importance of error in the reading
• Calibration is performed by company on normal patients breathing
various gas mixtures, so calibration is certain only down to 80%
35
Hypothermia
36
SOURCES OF ERROR
• Skin Pigmentation
• Darker color may make the reading more variable due to optical shunting.
• Dark nail polish has same effect: blue, black, and green polishes
underestimate saturations, while red and purple have no effect
• Hyperbilirubinemia has no effect
• Low perfusion state(hypotension-shock).
• Ambient Light
• Delay in reading of about 10 seconds
37
SOURCES OF ERROR
• Methylene blue and indigo carmine underestimate the saturation
• Dysfunctional hemoglobin
• Carboxyhgb leads to overestimation of sats because it absorbs at 660nm with an
absorption coefficient nearly identical to oxyhgb
• Methgb can mask the true saturation by absorbing too much light at both 660nm
and 940nm. Saturations are overestimated, but drop no further than 85%, which
occurs when methgb reaches 35%.
38
• Suspect the presence of carboxyhemoglobin in patient with:
• - Smoke inhalation
• - Intentional and accidental CO poisoning
• - Heavy cigarette smoking
Treat carboxyhemoglobin with high flow oxygen irregardless of the
pulse oximetry reading!
39
SOURCES OF ERROR
• Affect of anemia is debated
• Oxygen-Hemoglobin Dissociation Curve
• Shifts in the curve can affect the reading
• Oximetry reading could correspond to a PaO2 of 60mmHg (90% saturation) or 160mmHg
(99% saturation)
40
• Normal PaO2 is 80-100 mmHg
• Normally
• 80-100 mm Hg corresponds to 95-100% SpO2
• 60 mm Hg corresponds to 90% SpO2
• 40 mm Hg corresponds to 75% SpO2
41
Clarck Electrode
42
Blood Flow Meter
43
Measurement of Blood Flow & Volume
44
Measuring Cardiac Output-Several Methods
• Fick method: the indicator is O2, consumption is measured by a spirometer. The arterial-venous
concentration difference is measure by drawing samples through catheters placed in an artery and in
the pulmonary artery.
• Dye-dilution method: dye is injected into the pulmonary artery and samples are taken from an artery.
• Thermo-dilution method: cold saline is injected into the right atrium and temperature is measured in
the pulmonary artery.
45
Indicator Dilution
• Indicator Dilution that uses continuous infusion (Indicator – Oxygen)- samples from artery &
Pulmonary artery
• Indicator Dilution method that uses rapid injection
• Dye dilution –indocyanine green – cardio green- dye injected to pulmonary artery – samples from
artery
• Thermo Dilution – cold saline- injected to RA- temp measured in pulmonary artery
46
Indicator Dilution
• When a given quantity of m0 of an indicator is added to a volume V, the resulting concentration C of the
indicator is given by
• C = m0/V
• When an additional quantity m of indicator is then added, the incremental increase in concentration is
• ΔC = m/V
47
Indicator Dilution with Continuous Injection
• Measures flow / cardiac output averaged over several heart beats
Change in [ΔC] due to continuously added indicator m to volume V
dm dt dV dm dt
C = F = =
dV dt dt C
Clinician must add a fixed quantity of indicator in order to maintain a fixed change in concentration
• Fick’s technique: the amount of a substance (O2) taken up by an organ / whole body per unit time is
equal to the arterial level of O2 minus the venous level of O2 times the blood flow ➔
dV dm dt
Blood flow, liters/min F = =
dt C Consumption of O2 (mL/min)
(cardiac output)
dm dt Arterial and venous
=
Ca − Cv concentration of O2
(mL/L of blood) 48
Indicator – Dilution Curve
• where t1 is the time at which all effects of the first pass
of the bolus have died out (point E).
49
Indicator Dilution with Rapid Injection
• A known amount of a substance, such as a dye or radioactive isotope, is injected into the
venous blood and the arterial concentration of the indicator is measured through a series
of measurements until the indicator has completely passed through given volume.
• The cardiac output (blood flow) is amount of indicator injected, divided by average
concentration in arterial blood.
m
F = t
C (t )dt
0
50
Fick’s technique
O 2 consumption(mL/ min)
Cardiac Output =
[O 2 ]a − [O 2 ]v
250 mL/ min
• How is dm/dt (O2 consumption) measured?
= = 5 L/ min
190mL / L − 140mL/L
• Where and how would we measure Ca and Cv?
• The blood returning to the heart from the upper half of the body has a different concentration of
O2 from the blood returning from the lower half.
• after it has been mixed by the pumping action of the right ventricle.
• Cv can be measure it in the pulmonary artery
51
Fick’s Technique to measure blood flow from the heart
After the bolus is injected at time A, there is a transportation delay before the concentration begins rising at time
B. After the peak is passed, the curve enters an exponential decay region between C and D, which would continue
decaying alone the dotted curve to t1 if there were no recirculation. However, recirculation causes a second peak
at E before the indicator becomes thoroughly mixed in the blood at F. The dashed curve indicates the rapid
recirculation that occurs when there is a hole between the left and right sides of the heart.
53
Indicator – Dilution Curve
• Bolus is injected at time A
• There is a transportation delay before the concentration begins rising at time B.
• After the peak is passed, the curve enters an exponential decay region between C and D,
which would continue decaying along the dotted curve to t1 if there were no recirculation.
• Recirculation causes a second peak at E before the indicator becomes thoroughly mixed in
the blood at F.
• The dashed curve indicates the rapid recirculation that occurs when there is a hole
between the left and right sides of the heart.
• An increment of blood of volume dV passes the sampling site in time dt.
• quantity of indicator dm contained in dV is the concentration C(t) times incremental
volume.
• Hence dm =C(t) dV . Dividing by dt, we obtain (dm/dt)= C(t) (dV/dt)
• dm= Fi C(t) dt
54
An Example
55
Dye Dilution
• A common method of clinically measuring cardiac output is to use a
• colored dye, indocyanine green (cardiogreen).
• It meets the necessary requirements for an indicator
• The dye is available as a liquid that is diluted in isotonic saline and injected directly through a
catheter, usually into the pulmonary artery.
• About 50% of the dye is excreted by the kidneys in the first 10 min, so repeat determinations are
possible.
• The plot of the curve for concentration versus time is obtained from a constant-flow pump, which
draws blood from a catheter placed in the femoral or brachial artery.
• Blood is drawn through a colorimeter cuvette which continuously measures the concentration of
dye, using the principle of absorption photometry.
56
Thermo-dilution
• The indicator is cold – saline, injected into the right atrium using a catheter
• Temperature change in the blood is measured in the pulmonary artery using a thermistor
• The temperature change is inversely proportional to the amount of blood flowing through the
pulmonary artery
57
Electromagnetic Flowmeters
• Based on Faraday’s law of induction that a conductor that moves through a uniform magnetic field,
or a stationary conductor placed in a varying magnetic field generates emf on the conductor:
L
e= u B dL
0
59
Principle of Electromagnetic Blood Flow Meters
60
Electromagnetic Flowmeter Probes
61
Electromagnetic Flow Meters
62
Electro-magnetic Flowmeter: Issues
63
Types of Electromagnetic Blood Flow Meters
• DC Flow meters
• Use DC Magnetic field.
• Cause electrode polarization and amplifier drift.
• o/p same as ECG
• Poor SNR
• AC Flow meters
• Electromagnets are driven by alternating currents.
• The transducer acts like a Transformer and induces error voltages that often exceed the signal
levels by several orders of magnitude.
64
Electromagnetic AC flow meters
• Error recovery is achieved by using several different waveforms for magnet current
• Suitable balancing circuits are used to balance out the error voltage.
65
Thermal Convection
66
Thermal Convection
67
Thermal Convection
68
Ultrasonic Flowmeters
• Based on the principle of measuring the time it takes for an acoustic wave launched from a
transducer to bounce off red blood cells and reflect back to the receiver.
• All UT transducers, whether used for flowmeter or other applications, invariably consists of
a piezoelectric material, which generates an acoustic (mechanical) wave when excited by an
electrical force (the converse is also true)
• UT transducers are typically used with a gel that fills the air gaps between the transducer
and the object examined
69
Ultrasonic Flowmeters
70
Ultrasonic Flowmeters
71
Ultrasonic Flowmeters
72
Near / Far Fields
• Due to finite diameters, UT transducers produce diffraction patterns, just like an aperture does in optics.
• This creates near and far fields of the UT transducer, in which the acoustic wave exhibit different
properties
• The near field extends about dnf=D2/4λ, where D is the transducer diameter and λ is the wavelength.
During this region, the beam is mostly cylindrical (with little spreading), however with non-uniform
intensity.
• In the far field, the beam diverges with an angle sinθ=1.2 λ/D, but the intensity uniformly
attenuates proportional to the square of the distance from the transducer
73
UT Flowmeters
Zero-crossing
detector / LPF
Determine
direction
High acoustic
impedance material
74
Transit time flowmeters
Effective velocity of sound in blood: velocity of sound (c) + velocity of flow of blood averaged along the
path of the ultrasound (û)
û=1.33ū for laminar flow, û=1.07ū for turbulent flow ū: velocity of blood averaged over the cross
sectional area, this is different than û because the UT path is along a single line not over an averaged of
cross sectional area
76
Doppler Flowmeters
• The Doppler effect describes the change in the frequency of a received signal , with respect
to that of the transmitted signal, when it is bounced off of a moving object.
• Doppler frequency shift
u
F = f s (cos + cos )
c
78
Doppler Flowmeters
79
Problems Associated with Doppler Flowmeters
• Also unlike what the equation may suggest, the Doppler shift is not a single frequency, but
rather a band of frequencies because
• Not all cells are moving at the same velocity (velocity profile is not uniform)
• A cell remains within the UT beam for a very short period of time; the obtained signal needs
to be gated, creating side lobes in the frequency shift
• Acoustic energy traveling within the beam, but at an angle from the bam axis create an
effective ∆θ, causing variations in Doppler shift
• Tumbling and collision of cells cause various Doppler shifts
80
Directional Doppler
• Directional Doppler borrows the quadrature phase detector technique from radar in determining
the speed and direction of an aircraft.
• Two carrier signals at 90º phase shift are used instead of a single carrier. The +/- phase difference
between these carriers after the signal is bounced off of the blood cells indicate the direction,
whereas the change in frequency indicate the flowrate
81
Blood Pressure and Heart Sound Measurement
Instrument
82
• What is blood pressure and how is it measured?
• The heart supplies the organs and tissues of the body with blood. With every beat, it pumps blood into the
large blood vessels of the circulatory system. As the blood moves around the body, it puts pressure on the
walls of the vessels. Blood pressure readings are made up of two values:
• Systolic blood pressure is the pressure when the heart beats – while the heart muscle is contracting
(squeezing) and pumping oxygen-rich blood into the blood vessels.
• Diastolic blood pressure is the pressure on the blood vessels when the heart muscle relaxes. The diastolic
pressure is always lower than the systolic pressure.
• Blood pressure is measured in units of millimeters of mercury (mmHg). The readings are always given in pairs,
with the upper (systolic) value first, followed by the lower (diastolic) value.
• So someone who has a reading of 132/88 mmHg (often spoken “132 over 88”) has a
• systolic blood pressure of 132 mmHg, and a
• diastolic blood pressure of 88 mmHg.
83
• High blood pressure itself usually goes unnoticed. Only if it is extremely high
can it sometimes result in symptoms like dizziness or trouble seeing. Over the
long term, high blood pressure increases your risk of cardiovascular problems
like heart attacks, strokes, and heart and kidney failure. So if you or your
doctor think you have high blood pressure, it's important to have your blood
pressure checked regularly. If the readings are repeatedly too high, there are
several different ways of lowering your blood pressure and decreasing the risk
of long-term health consequences.
84
• Blood circulation: to transport oxygen
and other nutrients to the tissues and
carry metabolic waste products away
from cells
• Blood pressure measurement: to
determine the functional integrity of
the cardiovascular system
• Sound: fluctuations in pressure
recorded over audio frequency,
vibration due to acceleration and
deceleration of blood
85
Pressure at different regions around heart
86
Measurement of blood pressure
• Invasive
• Pressure catheter and transducer
• Non invasive
• Sphygmomanometry
• Auscultation (by ear or automatically by microphone)
• Oscillometry
• Volume clamp: Continuous non-invasive BP monitoring. at the finger with an inflatable cuff combined
with a photodiode. The diameter of the artery in the finger is measured by the photodiode; the pressure
in the cuff is adjusted to keep the diameter of the artery constant. From the pressure changes in the cuff,
a BP curve can be calculated and transferred to correspond to brachial artery BP
• Tonometry (arterial applanation tonometry) : Continuous non-invasive BP monitoring. A transducer
strapped to an artery with a bone underneath, can obtain the arterial pulse wave. The technique has
been refined and now is able to assess mean arterial pressure in the radial artery and allows the
calculation of diastolic and systolic arterial pressure
87
Measurement of blood pressure
Advantages/ drawbacks
• Invasive
• Accurate reproduction of central pressure waveforms
• Risk of thrombosis and arrhythmias
• Non-invasive
• Quick, cheap, widely used
• Lack of central pressure measurement
• Requires skilled and experienced operators
88
BP measurement
• According to the location of the sensor element there are two general
categories:
• Extravascular sensor:
• The most common clinical method for directly measuring pressure is to couple the
vascular pressure to an external sensor element via a liquid filled catheter.
• Intravascular sensor:
• In the second category the liquid coupling is eliminated by incorporating the sensor into
the tip of a catheter that is placed in the vascular system.
89
Different sensors for BP measurement
• Strain gauge
• Linear variable differential transformer
• Variable inductance
• Variable capacitance
• Optoelectronic
• Piezoelectric
• Semiconductor devices
90
Measurement of blood pressure: Direct Method
• Extravascular Sensors: Vascular pressure is connected to the sensor via a liquid filled
catheter.
92
Extravascular sensors
• Catheter-sensor system:
liquid-filled catheter (saline-
heparin), three-way
stopcock, pressure sensor
(usually Si diaphragm with
piezoresistive sensor), flush
solution
93
Extravascular sensors
• The physician inserts the
catheter either by means of a
surgical cut down which
exposes the artery or vein or
by means of percutaneous
insertion which involves the
use of a special needle or
guide wire techniques.
• Blood pressure is transmitted
via the catheter liquid column
to the sensor and finally to
the diaphragm which is
deflected.
94
Extravascular sensors
• Limitation of Catheter-sensor system :
• The frequency response of the catheter sensor system is limited by the hydraulic
property of the system.
• Catheter insertion: surgical cut-down or percutaneous insertion
95
Intravascular sensors
• Catheter tip sensors
• Sensors used at the Catheter tip:
• Strain-gage systems bonded onto a flexible diaphragm at the catheter tip.
• Catheter-tip micropressure sensor (micro silicone pressure sensor chips): Si
diaphragm and piezoelectric strain gage
• Disadvantages:
• More expensive than other
• may break after only a few uses, further increase its cost
97
Auscultatory method
• What is the Auscultatory method of measuring blood pressure?
• When measuring blood pressure using the auscultation method, turbulent
blood flow will occur when the cuff pressure is greater than the diastolic
pressure and less than the systolic pressure. The "tapping" sounds associated
with the turbulent flow are known as Korotkoff sounds.
• Auscultatory method: Keep the bell of stethoscope over the brachial artery
and inflate blood pressure cuff to a level higher than the systolic pressure
determined by the palpatory method. Steadily deflate. Record systolic and
diastolic pressures based on the Korotkoff sounds.
98
Auscultatory method
• Select an appropriate size cuff. The bladder in the cuff should encircle at least half of the arm. You have to use a
larger cuff in obese patients. Identify systolic blood pressure by palpatory method.
• Palpatory method:
• Empty air from the cuff and apply the cuff firmly around the patient's arm.
• Feel the radial pulse.
• Inflate the cuff until the radial pulse disappears.
• Inflate 30-40 mm over and release slowly until the pulse returns. That denotes systolic pressure.
• Diastolic blood pressure cannot be obtained by this method.
• Identification of systolic blood pressure by palpatory method helps one to avoid a lower systolic reading by
auscultatory method if there is an auscultatory gap.
• It also minimizes the discomfort of over inflating the bladder of the cuff.
• Auscultatory method:
• Keep the bell of stethoscope over the brachial artery and inflate blood pressure cuff to a level higher than
the systolic pressure determined by the palpatory method. Steadily deflate.
• Record systolic and diastolic pressures based on the Korotkoff sounds.
99
Sphygmomanometer Manometer
(mercury or capsule type)
d
Pulse detector
d + 20% (stethoscope or microphone)
100
Sphygmomanometer
• Occlusive cuff is inflated above the systolic pressure, then slowly bleed off.
• If cuff pressure is higher, no blood flow, no sound.
• When cuff pressure is slightly lower, blood spurts under the cuff.
• A palpable pulse in the wrist, a stethoscope is placed over brachial artery.
• Audible sound (Korotkoff sound) in stethoscope due to blood flow and vibrations.
• First detection pressure is systolic pressure.
• As the pressure of cuff continues to drop, at a certain pressure below which Korotkoff sounds
disappear….this is diastolic pressure.
• Automated Sphygmomanometer
– Microphone replaces stethoscope
– Korotkoff sound is detected by microphone
– Cycle start with rapid inflation of the cuff, 30mmHg higher than the expected systolic pressure.
– Systolic pressure noted when first Korotkoff sound heard, diastolic pressure noted when no sound.
– Automatically repeats the cycle.
– Block diagram available in Cromwell….(important)
101
Oscillometric Method
102
Oscillometric Method
103
Components of Oscillometric instrument
104
• http://www.meddean.luc.edu/lumen/meded/medicine/pulmonar/pd/pstep6.
htm
• https://www.ncbi.nlm.nih.gov/books/NBK279251/
• VVIMP: https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3639494/
105