Interpretation

of GC MS Results
Dr. Dhia Falih Al-Fekaiki
Expert of GC MS
University of Basrah
alfekaiki@yahoo.com
Iraq- Basrah
2017
[AOC-20i+s]
1

2 [GC-2010]

[GC Program]
3

[MS-2010]
4
AOC-20i+s 1
GC and GCMS Autosampler System
GC -2010 2
MS-2010 3
GC Program

300C
hold 3Min
1 0C
a se
re
Inc
i n
m
ch1
Ea
180C
hold 1Min
1 5C
ase
e
cr
In
in
1m
40C ch
Ea

hold 3Min
Target

Line#:16

Hit 1
Formlibrary
NIST28

Hit 2
Formlibrary
NIST28

Hit 3
Formlibrary
NIST28
Mean of peak
number
Target
The compute 0f
our samples

Hit 1
Formlibrary
NIST28

The computes
which suggest Hit 2
from NIST Formlibrary
NIST28
Library

Hit 3
Formlibrary
NIST28
Synonyms:
1.2-Furancarboxaldehyde
2.2-Furaldehyde
Hit#:1 Entry:6349 Library:NIST08.LIB 3.α-Furole
4.Artificial ant oil
SI:96 Formula:C6H6O3 CAS:67-47-0 MolWeight:126 5.Fural
RetIndex:1163 6.Furaldehyde
7.Furale
8.Furancarbonal
9.Furfuraldehyde
10.Furfurole
11.Furfurylaldehyde
12.Furole
13.Pyromucic aldehyde
14.2-Formylfuran
15.2-Furanaldehyde
16.2-Furancarbonal
17.2-Furfural
18.2-Furfuraldehyde
19.2-Furylaldehyde
20.Furol
21.2-Furylmethanal
22.Artificial oil of ants
23.Furfurale
24.Furfurol
25.NCI-C56177
26.2-Furil-metanale
27.2-Furankarbaldehyd
28.Rcra waste number U125
29.2-Furylcarboxaldehyde
30.Qo furfural
31.Furan-2-aldehyde
32.Furan-2-carbaldehyde
33.NSC 8841
5.3.2.1 Injection – Using the autosampler with a 10 μL syringe injection, perform three solvent washes from each wash bottle. Inject a 0.2
μL sample with a fast plunger speed, no sample washes, and 3 sample pumps.

5.3.2.2 Inlet – Run in split mode at 200 °C with a 100:1 split ratio.

5.3.2.3 Carrier gas – Helium.

5.3.2.4 Oven – Run a temperature program starting at 50.0 °C for 2.50 minutes. Then ramp at 10.0 °C per minute until 300 °C is obtained.
Hold at 300 °C for 5.00 minutes.

5.3.2.5 Column – Use a (5 %-Phenyl)-methylpolysiloxane (such as a HP-5MS or DB-5MS) column which is 0.25 mm in diameter that is
approximately 30 m long with a 0.25 μm film thickness. The column shall be kept at a constant flow of 0.6 mL/min.

5.3.2.6 Mass spectrometer – The transfer line temperature shall be set at 300 °C. The MS Quad temperature is set at 150 °C and the MS
source is set at 230 °C. The MS shall be run in scan mode using the settings from the atune.u file with an electron multiplier offset of 200
eV. The mass scan range is 20.0 amu to 400.0 amu with a threshold abundance of 150. The mass spectrometer shall have a 2.50 minute
solvent de
Thank you
Dr . Dhia alfekaiki