Professional Documents
Culture Documents
Coughing
Accumulation of secretions
Biting the endotracheal tube
Kinking in the circuit or artificial airway
Impingement of the endotracheal tube against
the carina
Herniation of the endotracheal tube cuff
TROUBLESHOOTIN
G
•Perform endotracheal suctioning if indicated.
•Insert a bite block or pharyngeal airway to prevent the patient
from biting the tube.
•Check for and rectify any kinks in the circuit or artificial
airway.
•Administer a bronchodilator to relieve bronchospasm.
•Reposition the endotracheal tube if it is not correctly placed.
•Check for and remove any water in the circuit.
•Evaluate and address patient-ventilator asynchrony.
•Inspect for a malfunction of the inspiratory or expiratory
valves.
LOW PRESSURE ALARM
A low pressure alarm in mechanical ventilation is
triggered whenever the peak inspiratory pressure
(PIP) decreases below a preset level. This most
commonly occurs due to a leak or disconnection in
the system.
The preset low pressure level for this alarm is
typically set around 5-10 cmH2O below the PIP.
TROUBLESHOOTING
•Check for a leak or disconnection in the circuit.
•Inspect the exhalation valve for leaks.
•Ensure all connections are secure and tight.
•Examine the pilot balloon for leaks.
•Verify that the endotracheal tube is correctly
positioned.
•Ensure the endotracheal tube cuff is adequately
inflated.
•Adjust the ventilator settings as necessary.
Note: If this alarm is triggered, the nurse
/respiratory therapist must promptly ensure that the
patient is being ventilated adequately. If the cause
of the alarm is unknown, the patient should be
manually ventilated until the source of the leak is
identified.
LOW volume ALARM
A low volume alarm in mechanical ventilation is
triggered whenever the expiratory volume falls
below a preset low volume threshold. This alarm
ensures that the patient is receiving and exhaling a
minimum tidal volume.
TROUBLESHOOTING
•Checking for leaks or disconnections in the circuit.
•Inspecting the exhalation valve for leaks.
•Ensuring all connections are secure.
•Examining the pilot balloon for leaks.
•Verifying that the endotracheal tube is correctly
positioned.
•Ensuring the endotracheal tube cuff is adequately
inflated.
•Adjusting the ventilator settings as necessary.
Note: If a leak or disconnection is detected, the
nurse/respiratory therapist must promptly ensure that the
patient is adequately ventilated. Manual breaths should be
provided if necessary until the source of the problem is
identified and corrected.
High frequency ALARM
A high frequency alarm in mechanical ventilation is triggered
whenever the total breathing frequency exceeds a preset high
frequency limit. This alarm aids in preventing tachypnea and
hyperventilation.