Learning Objectives • By the end of this session the learner, should be able to : • Monitor the different values and settings of ventilator. • Detect and response properly to ventilator’s alarms. • Know the proper troubleshooting for each alarm and how to response. • Provide immediate and suitable management. Monitoring the ventilator • The ventilator should be checked systemically on a scheduled, institution-specific basis, but usually no less often than every 4 hours . A check should also be performed • Before ABG values • Bedside pulmonary function data are obtained • After any change in ventilator setting • After an event of patient deterioration • At any time when the function of the ventilator is questionable . • Monitor information found on control panel – Mode – VT – Set rate – FIO2 – I:E ratio – Level of PS – Level PEEP – Sensitivity • Monitor the following information found on the display panel : – Peak airway pressure – Plateau pressure – Mean airway pressure – RR, Vt of the patient – Exhaled tidal volume ( mandatory and spontaneous breath). • Adequacy of oxygenation and ventilation can be monitored by arterial blood gases and pulse oximetry, while patient-ventilator synchrony is best monitored by clinical observation. Response to ventilator alarms • In the ICU there are many alarms • A ventilator alarm, however may be the one with the highest priority. • When it sounds, it may indicate a problem with the patients airway or breathing, the two highest priorities in the ABCs . • When any ventilator alarm sounds, the first thing to do is look at the patient . • If the patient is disconnected from the ventilator, then reconnect the patient to the machine. If the patient is connected : • Is the patient in distress ? • Not in distress . Identifying the patient in distress ? • When patients are described as fighting the ventilator, this means that they were doing well, but suddenly developed distress The sudden onset of dyspnea can be visually recognized by the following physical signs . • Tachypnea • Nasal flaring • Diaphoresis • Accessory muscle use • Retraction of the intercostal spaces • Paradoxical or abnormal movement of the thorax and abdomen . • If the patient is connected to the ventilator and is in distress and you can not readily identify the cause, then disconnect the patient from the machine and provide manual ventilation . • If the patients distress immediately goes away, the problem is with the ventilator . • If the distress does not go away, then the problem is with the patient . Ventilator-related problems • Leaks • Inadequate oxygen supply. • Inadequate ventilatory support • Inappropriate Trigger sensitivity • Inadequate flow setting Trigger sensitivity The lack of ventilator response to a patients inspiratory effeorts may be because of • Inadequate sensitivity settings • Poorly responsive internal demand valve • The use of an external nebulizer that is blocking the machine ability to sense a patients breath • Water in the inspiratory line • The presence of auto-PEEP . • Inappropriate sensitivity is easy to adjust; simply increase the sensitivity setting . Inadequate flow setting • A low inspiratory gas flow can be corrected by increasing the flow setting or by changing flow pattern by using a descending ramp instead of a rectangular wave . Patient-related problems • Airway problems : secretions, bronchospasm, • ETT related problems . • Dynamic hyperinfaltion (auto-PEEP) . • Pneumothorax • Pulmonary edema • Atelectasis • Pulmonary embolism . Pressure Alarms • High pressure alarm • Low pressure alarm • High minute volume Remember: Airway pressure = flow x resistance + alveolar pressure High Pressure Limit • The high pressure limit is usually set 10 cm above the patients average PIP • Normal PIP for a patient on a ventilator is between 20 and 30 cmH2O • Usual pressure set alarm : 40 cmH2O • When this alarm is activated, the ventilator terminates the inspiratory phase. Peak and plateau pressures • When the inflation volume is constant, an increase in the PIP is due to changes in pulmonary compliance or airway resistance or both . • Because no flow is present when the plateau pressure is created, this pressure is not a function of flow resistance in the airways . • Instead, the plateau pressure is directly proportional to the elastance of the lungs and chest wall . • If the peak pressure is increased but the plateau pressure is unchanged, the problem is an increase in airway resistance . In this situation, the major concerns are • Obstruction of the endotracheal tube • Migration and displacement of ETT • Airway obstruction from secretions • Acute bronchspasm . • Kinks or water in tubing (circuit) • Biting of ETT by patient • If the peak and plateau pressures are both increased, the problem is a decrease in distensibility of the lungs and chest wall(decreased compliance) . In this situation, the major concerns are • Pneumothorax, • Lobar atelectasis • Auto-PEEP • Acute pulmonary edema • Worsening pneumonia • ARDS . Extrapulmonary restriction as may arise from : • Abdominal distension (abdominal compartment syndrome, ascitis) • Tidal volume too large for the amount of lung being ventilated (eg, a normal tidal volume being delivered to a single lung because of malpositioning of the ETT. • Asynchronus breathing, increases both peak and plateau pressures . Response (management) to high PIP
• Suction airway secretions
• Straighten airway and other tubing to eliminate links, • Drain water from circuit tubing • Place bite block (mouth gage) or sedate patient • Checking the centimeter markings of the tube can show if a tube is too deep or too far out of the airway . • Auscultate for equal air-entry • Auscultate breath sounds for wheeze, and administer bronchodilator . • Checking for signs of ETT obstruction. Acute obstruction of endotracheal tubes (ETT) • Increases peak airway pressure • Decreases tidal volume (during pressure ventilation) • Hypercapia • Rapid decrease of end-tidal CO2 • Oxygen desaturation that may have lagged behind the decrease of end-tidal CO2 • DON’T DEPEND ON PIP TO DETECT ETT OBSTRUCTION • Peak airway pressure increase is a late warning sign of partial endotracheal tube obstruction whereas change in expiratory flow is an early warning sign (decelerating effect on the high expiratory is pronounced) . Evaluation • Evaluation of an obstruction problem should include : • Disconnect and manually bagging the patient • passage of a suction catheter through the ETT to confirm its patency and then making deflation and re-inflation of the cuff to see the difference upon ventilation. Endotracheal tube cuff inflated with 20 ml of air after removal from the patient. Note herniation of the cuff on the Murphy eye's side. Estimation of auto-PEEP • It’s PEEP above the preset level on the ventilator • Result from the short expiratory time, and not completely allowing full exhalation, resulting in air trapping • Expiratory hold (End Expiratory Pause) Expiratory Hold : End Expiratory Pause
• Is a procedure performed to estimate pressure
in the patients lung and circuit from trapped air (auto-PEEP) . • It is performed at the end of exhalation, following a mandatory breath. Causes There are three common situations during which auto-PEEP develops: • High minute ventilation • Expiratory flow limitation, and • Expiratory resistance. lack of return of flow to baseline at the end of the breath • When a high minute ventilation is the presumed cause of auto-PEEP, the minute ventilation should be decreased by lowering the tidal volume or respiratory rate. • This frequently requires a strategy of permissive hypercapnia • When an expiratory flow limitation due to obstructive airways disease is the presumed cause of auto-PEEP, the duration of expiration should be prolonged. • This can be accomplished by increasing the inspiratory flow, decreasing the tidal volume, or decreasing the respiratory rate. • Treatment with bronchodilators, steroids, and antibiotics may also be beneficial. • When increased expiratory resistance is the presumed cause of auto-PEEP, the source of increased resistance should be identified and corrected. • This may require sedation, pharmacologic paralysis, or replacement of the endotracheal tube or ventilator tubing. • If the cause is airflow limitation, intrinsic PEEP can be reduced by shortening inspiratory time or reducing the respiratory rate, thereby allowing a greater fraction of the respiratory cycle to be spent in exhalation. Low pressure alarm • Alarm will be activated if pressure in system has fallen and is not reaching the level that has generally been required for adequate ventilation in a particular patient . • The low inspiratory pressure alarm is usually set 5 to 10 cmH2O pressure below the patients average PIP. The cause is : • Patient-ventilator disconnection, or • Leak in the system (ETT or ventilator-related) . • If the leak is not obvious (not related to ETT), then the patient must be manually ventilated with a resuscitation bag while the machine leak is checked . High minute ventilation Cause : • increase in respiratory rate or tidal volume • Ventilator self-cycling because of incorrect set sensetivity . • Assessment and treatment : anxiety, pain, hypoxemia, or metabolic acidosis • Treat the underline cause. • Don’t jump to sedation and paralysis (unless it’s compromising the patient), figure out what’s wrong Expiratory Tidal Volume • Regardless of mode of ventilation, the most accurate measure of the volume received by the patient is the exhaled tidal volume (EVT). • The volume actually received by the patient, in any mode of ventilation, must be confirmed by monitoring the EVT on the display panel of the ventilator . • Generally the low exhaled tidal volume alarm is set 100 mL less than the prescribed tidal volume . Low Oxygen Pressure • The low oxygen pressure alarm warns of inadequate pressure in the oxygen lines supplying the ventilator .
• Due to loss of oxygen source or loss of
adequate pressure within oxygen source . Low Air Pressure • The low air pressure alarm warns of inadequate pressure in the air lines supplying the ventilator. • If the air source is lost, most ventilators will provide 100% oxygen in an attempt to maintain an adequate source of fresh gas to the patient . Patient-ventilator dysynchrony • Detect the cause first before sedate the patient more heavily. • Causes : • Inappropriate mode • I:E ratio • Mode of triggering and trigger sensitivity • Inappropriate sedation Hypotension • Hypotension immediately after initiation of positive ventilation is usually due to : • Hypovolemia • Drugs used for induction of anaesthesia • Gas trapping • Tension pneumothorax (less common) management • If the cause hypovolemia give a fluid chelleng and check. • Immediately disconnect the patient from the ventilator, if the causes is trapping 10 – 30 sec the BP will rise. • Neither the disconnection or fluids helped, consider other causes (tension pneu) Apnea/Low Frequency Alarm • The apnea or low frequency alarm is triggered when the total frequency drops below the low frequency limit. • Disconnection of the ventilator • patient under respiratory depressants or muscle-paralyzing agents, and respiratory muscle fatigue, cardiopulmonary arrest. • Always set a backup ventilation mode. poor oxygenation and ventilation • If the patient is desaturating, and Pao2 is decreased, it can be improved by (in this order) : • Increase FiO2. • Increase PEEP. • Increase inspiratory time. • Increase tidal volume or respiratory rate. Remember the effect of each change you make • Poor ventilation and high CO2 level, can improve the elimination by (in this order): • Increase respiratory rate. • Increase tidal volume. • Increase the expiratory time. • Decreasing dead space. Remember the effect of each change you make Detect
Manage Monitor
Remember … ‘’We’re Treating Patients, Not Ventilators’’ Questions ?..
Respiratory Therapy: 66 Test Questions Student Respiratory Therapists Get Wrong Every Time: (Volume 2 of 2): Now You Don't Have Too!: Respiratory Therapy Board Exam Preparation, #2