Professional Documents
Culture Documents
TELETHERAPY MACHINES
1
1. Kilovoltage Units
2. Van de Graaff Generator
3. Linear Accelerator
4. Betatron
5. Microtron
6. Cyclotron
7. Machine Using Radionuclides
8. Heavy Particle Beams
2
X-ray machines for radiotherapy
4
The main components of
a typical therapy x-ray
tube are:
• Water or oil cooled
target (anode)
• Heated filament
(cathode)
5
X-ray machines for radiotherapy
6
Kilovoltage Units
• Up to above 1950
• X-rays generated at voltages up to 300 kVps
• Still some use in the present era, esp. treatment of
superficial skin lesions
• Kilovoltage Therapy
– Grenz-Ray Therapy
– Contact Therapy
– Superficial Therapy
– Orthovoltage Therapy or Deep Therapy
– Supervoltage Therapy
7
Kilovoltage Units
• Grenz-Ray Therapy
– Energy : < 20 kV
– Very low depth of penetration
– No longer used in R/T
• Contact Therapy
– Energy: 40 – 50 kV
– Short SSD (< 2 cm)
– Produces a very rapidly decreasing depth dose
– Max irradiated tissue : skin surface
– Application: Tumor not deeper than 1 – 2 mm
8
Kilovoltage Units
• Superficial Therapy
– Energy: 50 – 150 kV
– HVLs: 1.0- – 8.0-mm Al
– Applicator or cone attached to the diaphragm
– SSD: 15 – 20 cm
– Tube current: 5 – 8 mA
– Application: tumors confined to about 5-mm depth
9
Kilovoltage Units
• Orthovoltage Therapy or
Deep Therapy
– Energy: 200 – 300 kV
– Tube current: 10 – 20 mA
– HVLs: 1 – 4 mm Cu
– Cones or movable diaphragm
(continuous adjustable field
size)
– SSD: 50 cm
– Application: tumor located < 2
–3 cm in depth
– Limitation of the treatment:
• skin dose
• Depth dose distribution
10
• Increase absorbed dose in bone
Kilovoltage Units
• Supervoltage Therapy
– Energy: 500 – 1000 kV
– Technical problem
• Insulating the high-voltage transformer
• Conventional transformer systems were not suitable
for producing potential > 300 kVp
– The problem solved by invention of resonant
transformer
11
Kilovoltage Units
At resonant frequency
1. Oscillating potential attains very high amplitude
2. Peak voltage across the x-ray tube becomes very large
12
Megavoltage Therapy
13
Clinical x-ray beams
14
1. Kilovoltage Units
2. Van de Graaff Generator
3. Linear Accelerator
4. Betatron
5. Microtron
6. Cyclotron
7. Machine Using Radionuclides
8. Heavy Particle Beams
15
Van de Graaff Generator
• Electrostatic accelerator
• Energy of x-rays: 2 MV
(typical), up to 10 MV
• Limitation:
– size
– high-voltage insulation
• No longer produced
commercially
– Technically better machine
(e.g. Co-60 units & linear
accelerators)
16
1. Kilovoltage Units
2. Van de Graaff Generator
3. Linear Accelerator
4. Betatron
5. Microtron
6. Cyclotron
7. Machine Using Radionuclides
8. Heavy Particle Beams
17
LINACS
18
Linac generations
19
Linear Accelerator
20
Linear Accelerator
21
The Magnetron
22
The Magnetron
24
The Klystron
25
Schematic diagram of a modern
fifth generation linac
26
Electron beam transport
27
The Linac X-Ray Beam
• Production of x-rays
– Electrons are incident on a target of a high-Z material
(e.g. tungsten)
– Target – need water cooled & thick enough to absorb
most of the incident electrons
– Bremsstrahlung interactions
• Electrons energy is converted into a spectrum of x-rays energies
• Max energy of x-rays = energy of incident energy of electrons
• Average photon energy = 1/3 of max energy of x-rays
• Designation of energy of electron beam and x-rays
– Electron beam - MeV (million electron volts,
monoenergetic)
– X-ray beam – MV (megavolts, voltage across an x-ray
tube, hetergeneous in energy) 28
Linac treatment head
Components of a modern linac
treatment head:
• Several retractable x-ray targets (one for
each x-ray beam energy).
• Flattening filters (one for each x-ray
beam energy).
• Scattering foils for production of clinical
electron beams.
• Primary collimator.
• Adjustable secondary collimator with
independent jaw motion.
• Dual transmission ionization chamber.
• Field defining light and range finder.
• Retractable wedges.
• Multileaf collimator (MLC). 29
Multi Leaf Collimator (MLC)
30
31
Siemens Elekta Varian
Y MLC Y Y
Jaw
YY Jaw
X1 X2
39.2 cm
MLC X X
Jaw
55.0 cm 57.6 cm
X X X X
MLC
Accessory
Holder Accessory
Holder Holder
Accessory
100 cm
43 cm 32 cm 29.2 cm
1.0 cm
Isocenter 1.0 cm 1.0 cm
Resolution
Resolution Resolution
32
33
Lead or tungsten
Opening from 0 x 0
to 40 x 40 cm at SSD
100 cm
34
Production of clinical x-ray beams
Typical values:
Pulse height: 50 mA
Pulse duration: 2 μs
Repetition rate: 100 pps
(pulse per second)
Period: 104 μs
35
Collimation System
In modern linacs the x-ray beam collimation is achieved
with
three collimation devices:
• Primary collimator.
• Secondary adjustable beam defining collimator
(independent jaws).
• Multileaf collimator (MLC).
The electron beam collimation is achieved with:
• Primary collimator.
• Secondary collimator.
• Electron applicator (cone).
• Multileaf collimator .
36
37
38
Production of clinical electron beam
39
Narrow pencil about 3
mm in diameter
Uniform electron
fluence across the
treatment field
e.g. lead
Electron scatter
readily in air
Beam collimator
must be achieved
close to the skin
surface
40
Dose monitoring system
41
Dose monitoring system
42
Dose monitoring system
43
Dose monitoring system
44
1. Kilovoltage Units
2. Van de Graaff Generator
3. Linear Accelerator
4. Betatron
5. Microtron
6. Cyclotron
7. Machine Using Radionuclides
8. Heavy Particle Beams
45
Betatron
Betatron is a cyclic accelerator in which the electrons are
made to circulate in a toroidal vacuum chamber (doughnut)
that is placed into a gap between two magnet poles.
Conceptually, the betatron may be considered an analog of
a transformer:
• Primary current is the alternating current exciting the
magnet.
• Secondary current is the electron current circulating in the
doughnut.
46
Betatron
Energy of x-rays:
6 – 40 MV
Disadvantage:
low dose rate
Small field size
47
1. Kilovoltage Units
2. Van de Graaff Generator
3. Linear Accelerator
4. Betatron
5. Microtron
6. Cyclotron
7. Machine Using Radionuclides
8. Heavy Particle Beams
48
Microtron
49
Microtron
Advantage:
Easy energy selection, small beam energy spread
and small size 50
1. Kilovoltage Units
2. Van de Graaff Generator
3. Linear Accelerator
4. Betatron
5. Microtron
6. Cyclotron
7. Machine Using Radionuclides
8. Heavy Particle Beams
51
Cyclotron
52
Cyclotron
Structures
• Short metallic cylinder divided into two section (Ds)
• Highly evacuated
• Placed between the poles of a direct current magnet
• Alternating potential is applied between two Ds
53
Cyclotron
• In a cyclotron the particles are accelerated
along a spiral trajectory guided inside two
evacuated half-cylindrical electrodes (dees)
by a uniform magnetic field produced
between the pole pieces of a large magnet
(1 T).
54
Cyclotron
55
1. Kilovoltage Units
2. Van de Graaff Generator
3. Linear Accelerator
4. Betatron
5. Microtron
6. Cyclotron
7. Machine Using Radionuclides
8. Heavy Particle Beams
56
The important characteristics of radionuclides useful for
external beam radiotherapy are:
• High gamma ray energy (of the order of 1 MeV).
• High specific activity (of the order of 100 Ci/g).
• Relatively long half life (of the order of several years).
• Large specific air kerma rate constant.
Of over 3000 radionuclides known only 3 meet the
required characteristics and essentially only cobalt-60 is
currently used for external beam radiotherapy.
57
Machines Using
Radionuclides
• Radionuclides have been used as source of γrays
for teletherapy
• Radium-226, Cesium-137, Cobalt-60
• Co has proved to be most suitable for external
60
beam R/T
• Higher possible specific activity
• Greater radiation output
• Higher average photon energy
I- Value
Half-Life γRay Energy Specific Activity Achieved
Radionuclide (Years) MeV
( Rm2_)
Ci – h in Practice (Ci/g)
59
Teletherapy machines
60
Teletherapy machines
Cobalt-60 teletherapy machine, Theratron-780, AECL (now MDS Nordion), Ottawa, Canada
61
Cobalt-60 Unit
• Source
– From 59Co(n, γ) nuclear reactor
– Stable 59Co → radioactive 60Co
– In form of solid cylinder, discs, or pellets
• Treatment beam
60Co →60Ni + 0β(0.32 MeV) + γ(1.17 & 1.33 MeV)
• Heterogeneity of the beam
– Secondary interactions
– βabsorbed by capsule → bremsstrahlung x-rays (0.1MeV)
– scattering from the surrounding capsule, the source
housing and the collimation system (electron
contamination)
62
Teletherapy sources
To facilitate interchange of sources from one teletherapy
machine to another and from one radionuclide production
facility to another, standard source capsules have been
developed.
Teletherapy sources are cylinders with height of 2.5 cm
and diameter of 1, 1.5, or 2 cm.
• The smaller is the source diameter, the smaller is the physical
beam penumbra and the more expensive is the source.
• Often a diameter of 1.5 cm is chosen as a compromise between
the cost and penumbra.
63
Teletherapy sources
64
Teletherapy source housing
The source head consists of:
• Steel shell with lead for shielding purposes
• Mechanism for bringing the source in front of the collimator
opening to produce the clinical gamma ray beam.
Currently, two methods are used for moving the
teletherapy source from the BEAM-OFF into the BEAM-
ON position and back:
• Source on a sliding drawer
• Source on a rotating cylinder
65
Teletherapy source housing
66
Teletherapy source housing
67
Collimator and penumbra
68
Penumbra
1. Transmission penumbra
2. Geometric penumbra
69
Source
• Transmission penumbra
– The region irradiated by photons Collimator SDD SSD
which are transmitted through the
edge of the collimator block
– The inner surface of the blocks is
made parallel to the central axis of
the beam
– The extent of this penumbra will be
more pronounced for larger
collimator opening
– Minimizing the effect
• The inner surface of the blocks remains
always parallel to the edge of the beam
70
• Geometric penumbra
• Radiation source: not a point source
– e.g. 60 Co teletherapy → cylinder of diameter ranging
from 1.0 to 2.0 cm
Pd = s (SSD + d – SDD)
SDD
Parameters determine the width of
penumbra
71
• Geometric penumbra (con’t)
– Solutions
• Extendable penumbra trimmer
– Heavy metal bars to attenuate the beam in the penumbra
region
• Secondary blocks
– Placed closed to the patient for redifining the field
– Should not be placed < 15 – 20 cm, excessive electron
contaminants
– Definition of physical penumbra in dosimetry
• Lateral distance between two specified isodose
curves at a specified depth
– At a depth in the patient, dose variation at the field border
– Geometric, transmission penumbras + scattered radiation
produced in the patient
72
1. Kilovoltage Units
2. Van de Graaff Generator
3. Linear Accelerator
4. Betatron
5. Microtron
6. Cyclotron
7. Machine Using Radionuclides
8. Heavy Particle Beams
73
Heavy Particle Beams
• Advantage
– Dose localization
– Therapeutic gain (greater effect on tumor than on
normal tissue)
• Including
– neutrons, protons, deuterons, αparticles, negative pions,
and heavy ions
• Still experimental
• Few institutions because of the enormous cost
74
Neutrons
– Characteristic distribution of
dose with depth
76