does not represent the actual anatomy and often misleads/ interferes with the diagnosis. Classification
• Magnetic and RF field Distortion
Artifacts • Reconstruction Artifacts • Noise-Induced Artifacts Magnetic and RF field Distortion Artifacts Ferromagnetic Material
Produces a local signal loss and a
warping distortion of the surrounding areas. Ferromagnetic Material
Metal contains no hydrogen, resulting
in a signal void on that location. A small band of metal in the patient’s ponytail. A metallic zipper in the patient’s shirt. Ferromagnetic Material
Utilizing lower field strength can
reduce metal artifacts Body shape, Conductivity, and Extension
The patient’s shape, electrical
conductivity, and filling of the rf coil all become factors in creating inhomogeneity of both the primary static magnetic field and the transmitted RF pulse. Chemical Shift
Chemical-shift artifacts are present
wherever tissues sharing a common border have considerably different molecular organization. Chemical Shift
The most prominent examples seen are
at interfaces of fat and the other body tissues. Chemical Shift
The artifact is seen as a bright rim of
signal at one interface and a dark rim on the opposite side of the particular organ. Reconstruction Artifacts Aliasing
One of the most commonly
encountered artifact. Also known as the wrap-around artifact. Aliasing
Occurs when portions of the patient’s
body are outside of the FOV but within the area of RF excitation. Aliasing
When hydrogen nuclei outside the area
of interest are excited, the signal they return is interpreted to have originated from within the imaging FOV. Aliasing
Aliasing artifact can be avoided by
using a larger FOV. Partial Volume Averaging
Occurs whenever the particular
structure of interest is contained within two tissues next with each other. Partial Volume Averaging
These artifact is worse with thick slices
and large voxels. Partial Volume Averaging
PVA can be reduced by using thinner
slices. Truncation or “ringing artifact”
Truncation artifact appears as multiple
rings of regular periodicity or duplication at transitions between high- and low-intensity signals. Truncation
Truncation is a geometric term that
describes the lessening by cutting off, or lopping, the vertex of a cone or cylinder. Truncation
Occurs when the reconstruction matrix
is asymmetric. (128x256 instead of 256x256) Truncation
Sometimes called as Gibbs
phenomenon Quadrature Detection
A zero line or zipper artifact is caused
by RF feed through from the RF transmitter along the frequency- encoding direction at the central or reference frequency of the imaging sequence. Quadrature Detection
Can also be produced by RF noise from
extraneous sources and is sometimes called an FID line. Quadrature Detection
The FID or zero-line artifact occurs
because of the remnant FID that occurs after the 180-degree pulse. Noise-Induced Artifacts Motion
An additional group of artifacts is
caused by voluntary, involuntary, and even microscopic physiologic motion. Motion
Repetitive motion can produce the
typical “ghosting,” or irregular wavelike lines of increased and decreased signal. Misregistration
If patient motion occurs between a
mask image and a subsequent image, the subtracted image contains misregistration artifacts. Off-resonance artifact
Degradation of the image as a result of
inexact tuning of the RF transmitter and/or receiver to the larmour frequency, resulting in an overall noisy image.