You are on page 1of 2

CLASS

TERMNATON
NSULATON
MOUNTNG
Denote RTD
NSULATON
NSULATON LENGTH
Code Termination
FL
SA
LA
PT
SL
PM
BY
TS
Flying leads no potseal. (Specify length - mm.)
Small aluminium WP head
Large aluminium WP head
Potseal & tails (100mm). (Specify
extra cable length.)
Spring loaded DN block
Two pin plug
Bayonet adaptor 1/4 insert.
Transition pot with spring.
(Specify length in mm.)
45
60
80
Code Sheath dia.
4.5mm OD 316 S/S
6.0mm OD 316 S/S
8.0mm OD 316 S/S
SHEATH DAMETER
ELEMENT LENGTH
0100
0200
0300
0400
0500
0600
0700
0800
0900
1000
Code
EIement
Length
100mm
200mm
300mm
400mm
500mm
600mm
700mm
800mm
900mm
1000mm
Code Wiring System
2
3
4
2 Wire Output
3 Wire Output
4 Wire Output
WRNG SYSTEM
AS
BS
CS
DS
AD
BD
CD
DD
Code Resistance
PT 50 Simplex
PT 100 Simplex
PT 500 Simplex
PT 1000 Simplex
PT 50 Duplex
PT 100 Duplex
PT 500 Duplex
PT 1000 Duplex
RESSTANCE
No Fitting
1/8 Fixed Brass
1/4 Fixed Brass
1/2 Fixed Brass
3/4 Fixed Brass
1/8 Compression Brass
1/4 Compression Brass
1/2 Compression Brass
3/4 Compression Brass
1/8 Fixed Stainless Steel
1/4 Fixed Stainless Steel
1/2 Fixed Stainless Steel
3/4 Fixed Stainless Steel
1/8 Compression Stainless Steel
1/4 Compression Stainless Steel
1/2 Compression Stainless Steel
3/4 Compression Stainless Steel
0
01FB
02FB
03FB
04FB
01AB
02AB
03AB
04AB
01FS
02FS
03FS
04FS
01AS
02AS
03AS
04AS
Code Mounting
Code InsuIation*
None
PVC/PVC
Silicone/Silicone
FB/FB SS Braided
Teflon/Teflon
0
PVC
SL
SSB
FEP
InsuIation
Lengths
0
0100
0200
0300
0400
0500
0600
0700
0800
0900
1000
1100
*
60 AS SL 3 100 P 02FB R PVC 0300
*
* = Only applicable when selecting
options FL and PT
*
Code InsuIation
P
M
MgO Powder Filled
Mineral nsulated
RTD - PT100
Resistance Temperature Detector

All temperatures from cryogenic to 600C where fast


response and accuracy are required
Pharmaceutical
Laboratory
General industry
The resistance that electrical conductors exhibit to the flow of an
electrical current is related to their temperature.
A PT-100 is a precision platinum resistor that exhibits 100 at
0C.
t has a positive temperature co-efficient so as the temperature
rises, so does the resistance.
ORDERING CODE
SEE BELOWFOR ORDERING OPTIONS
Resistance Temperature Detector (RTD) Ordering Code
AppIication ExampIes
Description
For more info: www.rhomberg.co.za
6B (C)
-200
-100
-0
+0
100
200
300
400
500
600
700
800
900
1000
IndustriaI PT Resistance Thermometer to BS 1904 TabIe 1. 1970 Revision to IPTS 68 and DIN 43760
-0
18.56
60.28
100.00
+0
100.00
138.50
175.83
212.02
247.08
375.6
313.72
345.4
375.6
404.8
432.7
-10
14.40
56.21
96.09
+10
103.90
142.29
179.50
215.58
250.52
284.31
316.93
348.4
378.7
407.6
435.4
-20
10.45
52.12
82.16
+20
170.79
146.06
183.16
219.12
253.95
287.63
320.12
351.5
381.6
410.5
438.1
-30
48.01
88.23
+30
111.67
149.82
186.82
222.66
257.37
290.93
232.31
354.5
384.6
413.2
440.8
-50
39.72
80.32
-50
119.40
157.31
194.07
229.69
264.17
297.50
329.6
360.6
390.4
418.8
446.3
-60
35.54
76.34
-60
123.24
161.04
197.69
233.19
267.56
300.76
332.8
363.6
393.3
4421.7
-70
31.34
72.35
+70
127.07
164.76
201.29
236.68
270.94
304.02
335.9
366.6
396.2
424.4
-90
22.83
64.32
+90
134.70
172.16
208.46
243.61
277.64
310.51
342.2
372.7
401.9
430.0
-100
18.56
60.28
+100
138.50
175.83
212.02
247.08
280.98
313.72
345.4
375.6
404.8
432.7
-40
43.88
84.29
+40
115.54
153.57
190.45
226.18
260.77
294.22
326.5
357.7
387.5
416.1
443.6
-80
27.11
68.34
+80
130.89
168.46
204.88
240.88
274.29
307.27
339.1
369.7
399.0
427.2
Resistance - Ohms
Resistance - Ohms
Two-Wire:
Provides one connection to each end of the
element. This construction is suitable where the
resistance of the lead wire may be considered as
and additive constant in the circuit and
particularly where the changes in lead resistance
due to ambient temperature changes may be
ignored.
Four-Wire:
Provides two connections to each end of the
element to completely compensate for leadwire
resistance and temperature change in leadwire
resistance. This configuration is used where
highly accurate temperature measurment is vital.
Three-Wire:
Provides one connection to one end of the
element and two to the other end of the element.
Connected to an instrument designed to accept
three wire input, sufficient compensation is
usually achieved for leadwire resistance and
temperature change in leadwire resistance. This
is the most commonly used configuration.
ELEMENT CONNECTIONS
RTD sensor assemblies are available with two, three and four wire leads. Two wire connected elements do not provide
lead resistance compensation for the measuring device. Three and four wire connected elements provided a means for
compensating for lead resistance between the sensor and the measuring device.
White
Red
Red
White
White
Red
White
Red
Black
Green
White
Red
White
Red
Red
Black
Red
Black
Green
White
Red
White
Red
Red
Black
White
Green
Length Below Thread
Element Length
(See below)
Precision Measures Perfection
T
S
-
L
I
N
E
T
E
M
P
E
R
A
T
U
R
E



S
E
N
S
O
R
S
DimensionaI Diagram
Wiring and Connection
PT100 Conversion TabIe