USBR 1040-89
PROCEDURE FOR
Scope
5.3 Tubing.-High-pressure
tubing of appropriate size
and length to connect pressure gauges and the primary
pressure standard to the pressure source.
5.4 Pressure Regulator.-A
pressure regulator..capable
of maintaining the desired pressure to the gauges and
primary pressure standard.
5.5
Valves.-Appropriate
size valves, two required;
bleed and pressure cutoff.
6.
Applicable Documents
7.
Summary of Method
8.
and Standardization
Conditioning
Procedure
Calibration
Precautions
Apparatus
5.1 Pressure
Gauge.-A pressure gauge meeting
requirements of Federal Specification GG-G-76E and the
American National Standard ANSI B40.1-1980.
5.2 Primary Pressure Standard (fig. 1).-A high
accuracy, deadweight tester designed for precision measurement of applied gas pressure. The device consists of
a weighted piston which becomes suspended in static
equilibrium when a prescribed gas pressure is introduced.
Calibrated masses are used for a range of pressures.
70
USB, R l,I], 0
i,LH( I
Il!I
9.21,
iit lc'
]',,-'Clll:
:,a:1,:_
crr,:iq '!lle;
<t,.2,
mi).!,
':::;.!,.::
i rm
'!' ::,.2,,
II'ILSt'F
( :{:l,]l,l{.,L [
t]IIU
',zllLJe
('{
M i, :1, ] h_
I, I;I 11
, ,tll 11
,; )11(
i ,r,,
,: m I
,: ) I
m),
,/,,
,:1 :,
(i 2: ::
.)
._ (m
[.!:,
2()
]:!,
:,tin
2.i ;,
].i;,
-I)
,.( l,
S :m
),
IYYC'NF'(IF:IL] i FII,,
,:, L,',],:>,
mn]s,{r
,;!,Llg2
i:]d iL:te:113'.!
L{, ,,,,,r
,1i_
,IibItv:li :ll kL{:{:( :,l LJK] ,1L,.' 1,1, ] { t'1%11L.pa]!;i.g i.1 ]RII ';. 2
,L[,',,
:'11 I 1:1 i yi ' i I: t { 'L i i'( I, I, I "( I' (' r il I: :, " > I' L I I I
,,L ,L L I l ; I ,k :!
]',],:'Sll
A,:,.u,,',
i"z
lir::vmzcc
i / / L:FL']: IIJll'[.
i:,,
,,,I
>:<::,LI,,,
,. ,I I,I:,:,
:1 1:1.
:,:).]
1 ,
:,=<I
,,L=m.,,
,E
,l,<,i:!i,,l
:,,,
!:,',
:I<
L:L,,],II',
i,
,m,, :,i,,
<:,
fl ,: ;,
:],,,,,,l,,,m,,,:
,,,i,:hh
f,,r
2',,,
:n<,,[
,lili,.,,:,:,
Ca]lcuh:/i.:,,rJ,s
I::a:l< t]Lii[:.i-,-, r_.q,iir{,.I:,e,.r :,. >J<:.',,. ,* ,:,u ri.,,r" 2.
Report
] I ]
T!i>.' re]l><:,l = i.. [,x u, :,,nsii>:a, ,:,f :* c<,.n]l,./c:,:Ja,,{I .I:>=:.Lk,:d
P',: ,:..:r.iLII'C (.] t,.gc' (:::[il:,.r:Ll:i,.r," fl :,11:,1 : fi.z. 2 )
112:
All ,,]i:t,Ld,:*,,>i.
:,v I,,/1<:,,.,. ] ,:l],,.:k,:r,arh
USBR 1040
7-2328 (10-85)
Bureau of Reclamation
MANUFACTURER
DesignationUSBR 1040
89
EXAMPLE
GAUGE RANGE
ACCURACY GRADE
SERIAL NO.
ACCURACY GRADE
SERIAL NO.
0-1400 kPa
3A
O-1500kPa
3A
CALIBRATED BY
DATE
1500
7
45787
ICHECKED BY,
DATE
kPa
i
Dt
%error -(1)-IMGn'
(FCM)
Trial
(1)
(A)
(B)
(C)
X 100
(A)
(B)
0.0
0.0
0.0
200
0.0
0.0
0.0
402
-0.07
0.0
-0. 13
600
599
O. 13
0.0
0.07
801
801
801
-0.07
1000
1000
1000
I000
0.0
0.0
1200
1202
1201
1200
0.0
-0.13
-0.07
0.0
1400
1400
1398
1399
0.0
O. 13
0.07
1500
1502
1500
1500
-0. 13
0.0
0.0
200
200
200
400
401
400
600
598
800
1400
-0.07
(C)
-0.07
kPo
Gauge calibration
Master gauge
reading
[] Ibf/in2
[] kPa
Trial
(1)
(A)
(B)
(C)
(A)
200
202
-0. 14
400
600
402
602
- 0. 14
800
802
-0. 14
-0.14
I000
I000
0.0
1200
1200
0.0
1400
1400
0.0
Remarks
Accept []
Reject
[]
(B)
(C)
0.0
GPO 849-232
72
USBR 1040
atmosphere
Bleed valve (
Primary Pressure
Standard
Pressure
cutoff valve
Pressure
regulator
Pressure source
(a)
Gauge to be calibrated
I
Bleed valve ()
Pressure
regulator
Pressure source
(b)
73
Pressure