You are on page 1of 7

Electromechanical Instruments

(Continue)
6. RECTIFIER VOLTMETER
6.1 PMMC Instrument on AC
As discussed earlier, the PMMC instrument is polarized, that is, its terminals are
identified as + and , and it must be connected correctly for positive (on-scale) deflection to
occur. When an alternating current with a very low frequency is passed through a PMMC
instrument, the pointer tends to follow the instantaneous level of the ac. As the current grows
positively the pointer deflection increases to a maximum at the peak of the ac. Then as the
instantaneous current level falls, the pointer deflection decreases toward zero. When the ac goes
negative, the pointer is deflected (off-scale) to the left of zero. This kind of pointer movement
can occur only with ac having a frequency of perhaps 0.1 Hz or lower. With the normal 60 Hz or
higher supply frequencies, the damping mechanism of the instrument and the inertia of the meter
movement prevent the pointer from following the changing instantaneous levels. Instead, the
instrument pointer settles at the average value of the current flowing through the moving coil.
The average value of purely sinusoidal ac is zero. Therefore, a PMMC instrument connected
directly to measure 60 Hz ac indicates zero. It is important to note that although a PMMC
instrument connected to an ac supply may be indicating zero, there can actually be a very large
rms current flowing its coils.
6.2 Full-Wave Rectifier Voltmeter
Rectifier instruments use silicon or germanium diodes to convert alternating current to a
series of unidirectional current pulses, which produce positive deflection when passed through a
PMMC instrument. The full-wave bridge rectifier circuit in Figure 3-15 passes the positive halfcycles of the sinusoidal input waveform and inverts the negative half-cycles. When the input is
positive, diodes D1 and D4 conduct, causing current to flow through the meter from top to
bottom, as shown. When the input goes negative, D2 and D3 conduct, and current again flows
through the meter from the positive terminal to the negative terminal.

Figure 3-15 An ac voltmeter may be constructed of a PMMC instrument, a multiplier resistor,


and a full-wave bridge rectifier. The instrument scale is correct only for pure sine waves.

The resulting current waveform is a series of positive half-cycles without any intervening
spaces (see Figure 3-15). As in the case of a dc voltmeter, the rectifier voltmeter circuit in Figure
3-17 uses a series-connected multiplier resistor to limit the current flow through the PMMC
instrument. The meter deflection is proportional to the average current, which is 0.637 peak
current. But the actual current (or voltage) to be indicated in ac measurements is normally the
rms quantity, which is 0.707 of the peak value, or 1.11 times the average value. Since there are
direct relationships between rms, peak, and average values, the meter scale can be calibrated to
indicate rms volts.
A rectifier voltmeter as discussed above is for use only on pure sine-wave voltages. When
other than pure sine waves are applied, the voltmeter will not indicate the rms voltage.
Example 3-6
A PMMC instrument with FSD = 100 A and Rm = 1 k is to be employed as an ac voltmeter
with FSD =100 V (rms). Silicon diodes are used in the bridge rectifier circuit of Figure 3-15.
Calculate the multiplier resistance value required.
Solution

At FSD the average current following through the PMMC instrument is


Iav = 100 A
I
100A
peak current Im = av =
157 A
0.637 0.637
(applied peak voltage ) (rectifier volt drop )
Im =
total circuit resistance
rectifier volt drops = 2VF (for D1 and D4 or D2 and D3)
applied peak voltage = 1.414Vrms
total circuit resistance = Rs + Rm
1.414Vrms 2VF
Im =
Rs Rm
1.414Vrms 2VF
Rs =
Rm
Im
(1.414 100V ) (2 0.7V )
=
1k
157A
= 890.7 k

Example 3-7
Calculate the pointer indications for the voltmeter in Example 3-6, when the rms input voltage is
(a) 75 V and (b) 50 V.
Solution
(a)

(b)

1.414Vrms 2VF
Rs Rm
(1.414 75V ) (2 0.7V )
= 0.637
890.7k 1k
75A = 0.75 FSD
(1.414 50V ) (2 0.7V )
= 0.637
50A = 0.5 FSD
890.7k 1k

Iav = 0.637Im = 0.637

Iav

Example 3-8 Calculate the sensitivity of the voltmeter in Example 3-6.


Solution
Im = 157 A
Irms = 0.707Im = 0.707 157 A
111 A (at FSD)
Vrms = 100 V (at FSD)
100V
total R =
= 900.9k
111A
R 900.9k
sensitivity = =
/V
100V
V
= 9.009 k/V
9 k/V
Examples 3-6 and 3-7 demonstrate that the rectifier voltmeter designed to indicate 100 V
rms at full scale also indicates 0.75 FSD when 75 V rms is applied, and 0.5 FSD for 50 V rms.
Therefore, the instrument has a linear scale. At low levels of input voltage the rectifier current is
also low, and this can result in errors due to variations in the diode voltage drop. The effect can
be countered by using a shunt resistor across the meter, as discussed next for a half-wave rectifier
voltmeter.
6.3 Half-Wave Rectifier Voltmeter
Half-wave rectification is employed in the ac voltmeter circuit shown in Figure 3-16. RSH
shunting the meter is included to cause a relatively large current to flow through diode D1 (larger
than the meter current) when the diode is forward biased. This is to ensure that the diode is
biased beyond the knee and well into the linear range of its characteristics. Diode D2 conducts
during the negative half-cycles of the input. When conducting, D2 causes a small voltage drop
(VF) across D1 and the meter, thus preventing the flow of any significant reverse leakage current
through the meter via D1. Diode D2 also protects the meter against reverse voltages.
The waveform of voltage developed across the meter and RSH is a series of positive halfcycles with intervening spaces, as illustrated. In half-wave rectification, Iav= 0.5(0.637Im). This
must be taken into account in the circuit design calculations.

Figure 3-16 Half-wave rectification may be used with a PMMC instrument and a multiplier
resistor for ac voltage measurements. A shunt resistor (RSH) is included to ensure a satisfactory
rectifier forward current level. The additional rectifier (D2) minimizes reverse leakage current
through D1.
3

Example 3-9
A PMMC instrument with FSD = 50 A and Rm 1700 is used in the half-wave rectifier
voltmeter circuit illustrated in Figure 3-16. The silicon diode (D1) must have a minimum (peak)
forward current of 100 A when the measured voltage is 20% of FSD. The voltmeter is to
indicate 50 V rms at full scale. Calculate the values of RS and RSH.
Solution

At FSD, Iav = 50 A.
Meter peak current,

I av
50A
=
157A
0.5 0.637 0.5 0.637
At 20% of FSD, diode peak current IF must be at least 100 A; therefore, at 100% of FSD,
100%
IF(peak) =
100 A = 500 A
20%
IF(peak) = Im + ISH
IF(peak) = Im + ISH
= 500 A 157 A = 343 A
Vm(peak) = ImRm = 157 A 1700
= 266.9 mV
V
266.9mV
RSH = m ( peak ) =
= 778
I SH ( peak )
343A

Im =

IF(peak) =
RS =

(applied peak voltage) Vm ( peak ) VF


RS
1.414Vrms Vm ( peak ) VF

I F ( peak )

(1.414 50V) 266.9mV 0.7V


= 139.5 k
500A

7. RECTIFIER AMMETER
Like a dc ammeter, an ac ammeter must have a very low resistance because it is always
connected in series with the circuit in which current is to be measured. This low-resistance
requirement means that the voltage drop across the ammeter must be very small, typically not
greater than 100 mV. However, the voltage drop across a diode is 0.3 to 0.7 V, depending on
whether the diode is made from germanium or silicon. When a bridge rectifier circuit is
employed, the total diode volt drop is 0.6 to 1.4 V. Clearly, a rectifier instrument is not suitable
for direct application as an ac ammeter.
The use of a current transformer (Figure 3-17) gives the ammeter a low terminal
resistance and low voltage drop. The transformer also steps up the input voltage (more secondary
turns than primary turns) to provide sufficient voltage to operate the rectifiers, and at the same
time it steps down the primary current to a level suitable for measurement by a PMMC meter.
Since the transformer is used in an ammeter circuit, the current transformation ratio Ip/Is = Ns/Np
is very important.
A precise load resistor (RL in Figure 3-17) is connected across the secondary winding of
the transformer. This is selected to take the portion of secondary current not required by the
meter. For example, suppose that the PMMC instrument requires 100 A (average) for FSD, and

the current transformer has Ns = 2000 and Np = 5. If the rms primary current is 100 mA, the
secondary rms current is
Is =

5
100 mA = 250 A
2000

or an average of
Is(av) =

1
250 A = 225.2 A
1.11

Since the meter requires 100 A for FSD, the value of RL is calculated to pass the
remaining 125.2 A.
The range of the instrument can be changed by switching-in different values of load
resistance. Another method of range changing involves the use of additional terminals (or taps)
on the primary winding to alter the number of primary turns, as shown in Figure 3-17.

Figure 3-17 Ac ammeter circuit consisting of a current transformer, full-wave bridge rectifier,
and a PMMC instrument.

Example 3-10
A rectifier ammeter with the circuit shown in Figure 3-17 is to give FSD for a primary current of
250 mA. The PMMC meter has FSD 1 mA and Rm=1700 . The current transformer has Ns= 500
and Np = 4. The diodes each have VF = 0.7 V, and the series resistance is Rs = 20 k. Calculate
the required value of RL.
Solution
I av
1mA
=
= 1.57mA
0.637 0.637
Transformer secondary peak voltage,
Em = Im(RS + Rm) + 2VF
= 1.57 mA (20 k + 1700 ) + 1.4 V
35.5 V
or secondary voltage
ES = (0.707 35.5 V) rms
25.1 V
and
rms meter current = 1.11 Iav
= 1.11 mA

Peak meter current

Im =

Transformer rms secondary current,


N
Is = Ip p
Ns
= 250 mA

4
= 2 mA
500

IL = Is Im
= 2 mA l.11 mA = 0.89 mA
E
25.1V
RL = s =
= 28.2 k
I L 0.89mA

8. DEFLECTION INSTRUMENT ERRORS


8.1 Reading Errors
Some sources of error in measurements made by deflection instruments are: bearing
friction, improperly adjusted zero, and incorrect reading of the pointer indication. Zero and
friction errors can be minimized by carefully adjusting the mechanical zero of an instrument
before use and by gently tapping the meter to relieve friction when zeroing and reading. Portable
instruments should normally be used lying flat on their backs. Care in deciding the exact position
of the pointer on the scale will reduce reading errors.
Even with an accurately marked scale and a sharp pointer, two observers may disagree
about the exact scale reading. This occurs because of parallax error: the uncertainty about the
eye of the observer being directly in line with the end of the pointer. Parallax error is eliminated
in good instruments by the use of a knife-edge pointer and a mirror alongside the scale. When an
observer lines up the pointer and the mirror image of the pointer, the observers eye is exactly in
the line with the pointer, and the scale can then be read accurately.
8.2 Specified Accuracy
High-quality instruments may have their accuracy specified as a percentage of the actual
scale reading, or measured quantity. However, for most deflection instruments, manufacturers
specify the accuracy as a percentage of FSD. This means, for example, that an instrument that
gives FSD for a coil current of 100 A, and which is specified as accurate to l%, has a l A
accuracy at all points on its scale. Thus, as demonstrated in Example 3-11, the measurement error
becomes progressively greater for low scale readings.

Example 3-11
An instrument that indicates 100 A at FSD has a specified accuracy of l%. Calculate the upper
and lower limits of measured current and the percentage error in the measurement for (a) FSD
and (b) 0.5 FSD.
Solution
(a) At FSD:

indicated current
Error
actual measured current
Error

= 100 A
= 1% of 100 A = 1 A
= 100 A 1 A = 99 A to 101 A
= 1% of measured current

(b) At 0.5 FSD indicated current = 0.5100 A = 50 A


Error = 1% of FSD = 1 A
6

actual measured current = 50 A 1 A


= 49 A to 51 A
1A
Error =
100%
50A
= 2% of measured current

Problems
1) An ac voltmeter uses a bridge rectifier with silicon diodes and a PMMC instrument with FSD
= 75 A. If the meter coil resistance is 900 and the multiplier resistor is 708 k , calculate the
applied rms voltage when the voltmeter indicates FSD. (60 V)
2) Determine the new multiplier resistance required for the voltmeter in Problem 1 to change its
range to 300 V FSD. (3.59M)
3) Determine the pointer position on the voltmeter in Problem 2 when the applied rms voltage is
(a) 30 V and (b) 10 V. (9.7%, 3%)
4) A PPMC instrument with a 900 coil resistance and an FSD of 75 A is to be used with a
half-wave rectifier circuit as an ac voltmeter. Silicon diodes are used, and the minimum diode
forward current is to be 80 A when the instrument indicates 0.25 FSD. Calculate the shunt and
multiplier resistance values required to give 200 V FSD. (880.9k)
5) Calculate the sensitivity of the ac voltmeter in Problem 1. (12k/V)
6) Calculate the sensitivity of the ac voltmeter in Problem 4. (4.4k/V)
7) A rectifier ammeter is to indicate full scale for a 1 A rms current. The PMMC instrument
used has a l200 coil resistance and 500 A FSD, and the current transformer has Ns =7000 and
Np = 10. Silicon diodes are used and the meter series resistance is Rs = l50 k. Determine the
required secondary shunt resistance value. (97.6k)
8) A rectifier ammeter has the following components: PMMC instrument with FSD = 200 A
and Rm =900 ; current transformer with Ns =600 and Np =5; diodes with VF = 0.3 V; meter
series resistance Rs = 270 k; transformer shunt secondary resistance RL =98.7 k. Calculate the
level of transformer primary current for instrument FSD. (100mA)
9) Calculate the sensitivity of the ac voltmeter in Example 3-9 when diode D2 is (a) included in
the circuit and (b) omitted from the circuit. (2.8 k/V, 4k/V)
10) A PMMC instrument with 250 A FSD has a specified accuracy of 2%. Calculate the
measurement accuracy at currents of 200 A and l00 A. (2.5%, 5%)
11) A deflection instrument with 100 A FSD has a 3% specified accuracy. Calculate the
possible error when the meter indication is (a) 50 A and (b) 10 A. (6%, 30%)
12) A 25 A current is measured on an instrument with 37.5 A FSD. If the measurement is to be
accurate to within 5%, determine the required instrument accuracy. (3.3%)

You might also like