Professional Documents
Culture Documents
CLINICAL CHEMISTRY I
Analytic Techniques
SPECTOPHOTOMETRY
Beer-Lambert Law
Percent of light transmitted (%T) can be calculated by multiplying the ratio of the intensity of
transmitted light (I) to the intensity of absorbed light (Io) by 100:
%T = (I / Io) x 100
Absorbance (A), the amount of light absorbed, is inversely proportional to %T:
A 1/α log(%T)
To compute for the absorbance, the log(%T) must be subtracted from the log of 100% (which
is equal to 2)
A = 2 – log(%T)
Absorbance (A) can also be computed by multiplying the molar absorptivity (ε), length of
light path through the solution (b), and concentration of absorbing molecules (C):
A = εbC
Since path length and molar absorptivity are constant for a given wavelength, it can be
deduced that absorbance is directly proportional to concentration:
A~C
In manual spectrophotometer, only %T can be displayed by the machine whereas in digital
spectrophotometers, A can also be selected to be displayed rather than %T.
STANDARDIZATION: To compute for the unknown concentration of the compound of
interest (Cu), the absorbance of compound of interest (Au) is divided by the absorbance of a
standard solution (Astd – the standard solution then must be tested first before the patients
sample) and then the quotient is multiplied with the concentration of the standard solution
(Cstd – this is provided by the manufacturer of the std. sol’n):
Cu = Au x Cstd
Astd
Components of a Spectrophotometer:
Light source – Important factors like range, spectral distribution within the range, the
source of radiant production, stability of radiant energy, and
temperature.
- The most common light source for work in the visible and near infrared region
is the incandescent tungsten or tungsten-iodide lamp.
- In incandescent tungsten or tungsten-iodide lamp, only 15% of emitted ER
falls in the visible region with the rest emitted as near-infrared.
- The most common light source used for UV work are the deuterium-discharge
lamp and mercury arc lamp.
Page |3
and converts the signal into digital signals which can then be displayed as
absorbance readings.
- Photodiode: uses a reverse-biased +/- junction diode that produces a
photocurrent that is proportional to the transmitted light; though not as
sensitive as PM, its speed and small size make them useful in applications
where light levels are adequate.
Blanking – the use of a reference solution that does not contain the compound of interest; this is
for the calibration of the spectrophotometer. The steps to blanking are:
1. Water reagent blanking – the use of distilled water as the sample to be tested; if the
machine’s meter is not zeroed, then the meter is adjusted so that it points to zero. This is
to make sure that the spectrophotometer reads a 100% transmittance.
2. Reagent blanking – the use of a colored reagent that will be used as a diluting medium
for the samples with the compound of interest; the reagent is put in the cuvet and then
tested in the spectrophotometer; it is expected that there will be a reading in the meter,
this is the A or %T of the reagent – this adds to the amount of light absorbed or
transmitted by the compound of interest; to remove this, simply adjust the meter reading
until it reads zero; this functions in the same way as the “TARE” option in a electronic
balance which removes the weight of the container from the reading and adjusts the
reading back to zero and the balance will now measure only the weight of the substance
being added.
Spectrophotometer QA
Wavelength accuracy – means the wavelength indicated on the control dial is the actual
wavelength being passed by the monochromator; this is commonly checked using standard
absorbing solutions or filters with absorbance maxima of known wavelength; the solution is
tested and the control dial is then rotated to the wavelength of maximum absorbance and, if
working properly, will result in a 0%T; if not, the optics must be adjusted to calibrate the
monochromator correctly. If the wavelength indicator control is the one with a problem, a
mercury vapour lamp is used as light source and the spectrum is scanned to locate mercury
emission lines; the wavelength indicated on the control is then compared to the known
mercury emission peaks.
Stray light – refers to any wavelengths outside the band of the monochromator; commonly
caused by scratched and dust particles anywhere on the light path; checked using cutoff
filters which eliminate wavelengths except the one of interest; the reading in a quality
assured machine while performing this QC must be 0%T; if reading is greater, there is
presence of stray light.
Linearity – demonstrated when a change in concentration results in a straight-line calibration
curve; checked using colored solutions with known concentration of constituents; less than
expected absorbance is an indication of stray light or bandpass that is wider than specified
(wavelength accuracy issues).
Source of Variation
Transfer from testube into the cuvet
Contamination of sample to be tested
FLAME PHOTOMETRY
The following are the step-by-step process in the use of a flame photometer:
1. Sample is atomized into small droplets by an atomizer
or nebulizer.
2. The droplets are mixed with gas and air in the mixing
chamber, forming an aerosol mixture.
3. The aerosol mixture is then led up to burner, where the
droplets containing the element of interest are reduced
and excited.
4. After the element’s electrons are excited, they go back
to their ground state and release ER.
5. Desired ER is filtered from other wavelengths by a
monochromator and then detected by a photodetector
which will give a reading according to the intensity of
light produced.
The more intense the light emitted, the higher the concentration of the element of interest.