Professional Documents
Culture Documents
SPECTROSCOPY
Spectroscopy
Visible Wavelength
Violet 400-420 nm
Indigo 420-440 nm
Blue 440-490 nm
Green 490-570 nm
Yellow 570-585 nm
Orange 585-620 nm
Red 620-780 nm
Principles of spectroscopy
Absorption spectroscopy
An analytical technique which concerns with the measurement of absorption of
electromagnetic radiation
Example
UV spectroscopy 185-400 nm
Visible spectroscopy 400-800 nm
IR Spectroscopy 0.76-15 µm
Principles of spectroscopy
Emission spectroscopy
An analytical technique in which emission (of a particle or radiation) is dispersed
according to some property of the emission and the amount of dispersion is
measured
Mass spectroscopy
Interaction of electromagnetic
radiation with matter
𝑘𝑐𝑎𝑙
∆𝐸 = 35 𝑡𝑜 71
𝑚𝑜𝑙𝑒
Interaction of electromagnetic
radiation with matter
σ σ* Transition
σ electron from orbital is excited to corresponding anti-bonding orbital σ*
The energy required is large for this transition
Methane (CH4) has C-H bond only and can undergo σ σ* transition and
shows absorbance maxima at 125 nm
The organic compounds in which all the valence sheel electrons are involved in
the formation of σ bond do not show absorption in normal UV region (200-400
nm)
The possible electronic transitions are
π π* transition
π electron in a bonding orbital is excited to corresponding antibonding orbital
π*
compounds containing multiple bonds like alkenes, alkynes, carbonyl, nitriles,
aromatic compounds, etc undergo π π* transitions.
Alkenes generally absorb in the region 170 to 205 nm
The possible electronic transitions are
n σ* transition
Saturated compounds containing atoms with lone pair of electrons like O, N, S,
and halogens are capable of n σ* transition
These transitions usually requires less energy than σ σ* transitions
The number of organic functional groups with n σ* peaks in UV region is small
(150-250 nm)
The possible electronic transitions are
n π* transition
An electron from non-bonding orbital is promoted to anti-bonding π* orbital
Compounds containing double bond involving hetero atoms (C=O), N=O)
undergo such transitions
n π* transition require minimum energy and show absorption at longer
wavelength around 300 nm
σ π* transition and π σ* transition
These electronic transition are forbidden transitions and are theoretically
possible
Thus, n π* and π π* electronic transitions show absorption in region above
200 nm which is accessible to UV-Vis spectrophotometer
The UV spectrum is of only a few broad of absorption
Absorption and intensity shifts
In a single beam instrument, the light beam follows a single path from
the source, to the monochromator, to the sample cell and finally to the
detector
1- Sources of light
30
Sources used in UV-Vis Spectrophotometers are continuous sources.
• Continuous sources emit radiation of all wavelengths within the
spectral region for which they are to be used.
• Sources of radiation should also be stable and of high intensity.
Continuous
Sources
Visible and
Ultraviolet
near IR
radiation
radiation
Tungsten Deuterium
Lamp Lamp
320-2500 nm 200-400 nm
2. Wavelength Selectors
31
Ideally the output of a wavelength selector would be a radiation of a single
wavelength.
Wavelength
selectors
Filters Monochromators
i- Filters
32
• Filters permit certain bands of wavelength (bandwidth of ~ 50 nm) to pass
through.
• The simplest kind of filter is absorption filters , the most common of this
type of filters is colored glass filters.
• They are used in the visible region.
• The colored glass absorbs a broad portion of the spectrum
(complementary color) and transmits other portions (its color).
Disadvantage
• They are not very good wavelength selectors and can’t be used in
instruments utilized in research.
• This is because they allow the passage of a broad bandwidth which
gives a chance for deviations from Beer’s law.
• They absorb a significant fraction of the desired radiation.
ii- Monochromators
They are used for spectral scanning (varying the wavelength of 33
radiation over a considerable range ).
Bandwidth Choice
Selection of wavelength
Absorbance measurements are always carried out at fixed
wavelength (using monochromatic light). When a wavelength is
chosen for quantitative analysis, three factors should be considered
λmax
Absorbance
such wavelengths, it will be 10-4 M
5x10-5 M
sure that the lowest sample
10-5 M
concentration can be
measured with fair accuracy.
A Broad horizontal
bands
Band B
Absorbance
A shoulder
m wavelength
3- If the solution contains more than absorbing species, the 38
wavelength should be chosen, whenever possible, in region at which
the other species does not absorb radiation or its absorbance is
minimum. By this way, the second species does not interfere in the
determination.
Absorbance sample
Interfering
species
m wavelength
3- Sample compartment (cells)
39
For Visible and UV spectroscopy, a liquid sample is usually
contained in a cell called a cuvette.
Glass is suitable for visible but not for UV spectroscopy because it
absorbs UV radiation. Quartz can be used in UV as well as in visible
spectroscopy
Transparent
Face
1 cm 1 cm
Short pathlength (b)
4- Detectors
The detectors are devices that convert radiant energy into electrical 40
signal.
A Detector should be sensitive, and has a fast response over a
considerable range of wavelengths.
In addition, the electrical signal produced by the detector must be
directly proportional to the transmitted intensity (linear response).
i- Phototube
light dynodes
anode
electrons
e-
photochathode
high voltage
The components of a single beam
spectrophotometer 42
Light source - white light of constant intensity
slits
Grating
slits Separates white light
Phototube
into various colors
detects light &
measures intensity Rotating the grating
Sample
changes the wavelength
When blank is the sample going through the sample
Po is determined,
otherwise P is measured
Double Beam Spectrophotometer 43
Mirror
44
Double Beam Spectrophotometer
B B B B B represents P0
S
Signal %T 100
B
S S S S represents P
Time
45
When light passes through the sample, the detector measures the
P. When the chopper diverts the beam through the blank solution, the detector
measures P0.
The beam is chopped several times per second and the electronic
circuit automatically compares P and P0 to calculate absorbance
and Transmittance.
Advantages of double beam instruments over single beam instruments