Professional Documents
Culture Documents
Introduction
Spectroscopy is the use of the absorption, emission, or scattering of electromagnetic
radiation by atoms or molecules (or atomic or molecular ions) to qualitatively or quantitatively
study the atoms or molecules, or to study physical processes. The interaction of radiation with
matter can cause redirection of the radiation and/or transitions between the energy levels of the
atoms or molecules. A transition from a lower level to a higher level with transfer of energy from
the radiation field to the atom or molecule is called absorption. A transition from a higher level
to a lower level is called emission if energy is transferred to the radiation field, or nonradiative
decay if no radiation is emitted. Redirection of light due to its interaction with matter is called
scattering, and may or may not occur with transfer of energy, i.e., the scattered radiation has a
slightly different or the same wavelength.
Absorption
When atoms or molecules absorb light, the incoming energy excites a quantized structure
to a higher energy level. The type of excitation depends on the wavelength of the light. Electrons
are promoted to higher orbitals by ultraviolet or visible light, vibrations are excited by infrared
light, and rotations are excited by microwaves.
An absorption spectrum is the absorption of light as a function of wavelength. The
spectrum of an atom or molecule depends on its energy level structure, and absorption spectra
are useful for identifying of compounds.
Measuring the concentration of an absorbing species in a sample is accomplished by
applying the Beer-Lambert Law.
The Beer-Lambert law (or Beer's law) is the linear relationship between absorbance and
concentration of an absorbing species. The general Beer-Lambert law is usually written as:
A=λbc
where A is the measured absorbance, λ is a wavelength-dependent absorptivity
coefficient, b is the path length, and c is the analyte concentration. When working in
concentration units of molarity, the Beer-Lambert law is written as: A = ε b c
where ε is the wavelength-dependent molar absorptivity coefficient with a unit of M-1 cm-1.
Experimental measurements are usually made in terms of transmittance (T), which is defined as:
T = I / Io
where I is the light intensity after it passes through the
sample and Io is the initial light intensity. The relation
between A and T is: A = -log T = - log (I / I o).
Modern absorption instruments can usually display
the data as transmittance, % transmittance, or absorbance.
An unknown concentration of an analyte can be determined Absorption of light by a sample
by measuring the amount of light that a sample absorbs and
applying Beer's law. If the absorptivity coefficient is not known, the unknown concentration can
be determined using a working curve of absorbance versus concentrations derived from
standards.
Electromagnetic Spectrum
Type of Radiation Frequency Range (Hz) Wavelength Range Type of Transition
γ-rays 1020-1024 <10-12 m nuclear
x-rays 1017-1020 1 nm-1 pm inner electron
ultraviolet 1015-1017 400 nm-1 nm outer electron
visible 4-7.5x1014 750 nm-400 nm outer electron
near-infrared 1012-4x1014 2.5 um-750 nm outer electron, molecular
vibrations
infrared 1011-1012 25 um-2.5 um molecular vibrations
microwaves 108-1012 1 mm-25 um molecular rotations,
electron spin flips*
radio waves 100-108 >1 mm nuclear spin flips*
Emission
Atoms or molecules that are excited to high energy levels can decay to lower levels by
emitting radiation (emission or luminescence). For atoms excited by a high-temperature energy
source this light emission is commonly called atomic or optical emission, and for atoms excited
with light it is called atomic fluorescence. For molecules it is called fluorescence if the transition
is between states of the same spin and called phosphorescence if the transition occurs between
states of different spin. In simple terms, phosphorescence is a process in which energy absorbed
by a substance is released relatively slowly in the form of light.
The emission intensity of an emitting substance is linearly proportional to analyte
concentrations (at low concentration range), and is useful for quantitating emitting species.
Scattering
When electromagnetic radiation passes through matter, most of the radiation continues in
its original direction, but a small fraction is scattered in other directions. Light that is scattered at
the same wavelength as the incoming light is called Rayleigh scattering. Light that is scattered in
transparent solids due to vibrations (phonons) is called Brillouin scattering. Brillouin scattering
is typically shifted by 0.1 to 1 cm-1 from the incident light. Light that is scattered due to
vibrations in molecules or optical phonons in solids is called Raman scattering. Raman scattered
light is shifted by as much as 4000 cm-1 from the incident light.
Raman Effect
The Raman effect occurs
when light impinges upon a
molecule and interacts with the
electron cloud of the bonds of that
molecule. The incident photon
excites one of the electrons into a
virtual state. For the spontaneous
Raman effect, the molecule will be
excited from the ground state to a
virtual energy state, and relax into
a vibrational excited state, which
generates Stokes Raman scattering.
If the molecule was already in an
elevated vibrational energy state,
the Raman scattering is then called anti-Stokes Raman scattering.
A molecular polarizability change, or amount of deformation of the electron cloud, with
respect to the vibrational coordinate is required for the molecule to exhibit the Raman effect. The
amount of the polarizability change will determine the Raman scattering intensity, whereas the
Raman shift is equal to the vibrational level that is involved
Spectrofluorometer
Two general types of instruments exist:
Filter fluorometers use filters to isolate the incident light and fluorescent light.
Spectrofluorometers use diffraction grating monochromators to isolate the incident light
and fluorescent light.
Both types utilize the following scheme: The light from an excitation source passes through a
filter or monochromator, and strikes the sample. A proportion of the incident light is absorbed by
the sample, and some of the molecules in the sample fluoresce. The fluorescent light is emitted in
all directions. Some of this fluorescent light passes through a second filter or monochromator
and reaches a detector, which is usually placed at 90° to the incident light beam to minimize the
risk of transmitted or reflected incident light reaching the detector.
Various light sources may be used as excitation sources, including lasers, photodiodes,
and lamps; xenon arcs and mercury vapor lamps in particular. A laser only emits light of high
irradiance at a very narrow wavelength interval, typically under 0.01 nm, which makes an
excitation monochromator or filter unnecessary. The disadvantage of this method is that the
wavelength of a laser cannot be changed by much. A mercury vapor lamp is a line lamp,
meaning it emits light near peak wavelengths. By contrast, a xenon arc has a continuous
emission spectrum with nearly constant intensity in the range from 300-800 nm and a sufficient
irradiance for measurements down to just above 200 nm.
Filters and/or monochromators may be used in fluorimeters. A monochromator transmits
light of an adjustable wavelength with an adjustable tolerance. The most common type of
monochromator utilizes diffraction grating, that is, collimated light enters a grating and exits
with a different angle depending on the wavelength. The monochromator can then select which
wavelengths to transmit. For allowing anisotropy measurements the addition of two polarization
filters are necessary: One after the excitation monochromator or filter, and one before the
emission monochromator or filter.
As mentioned before, the
fluorescence is most often measured at a
90° angle relative to the excitation light.
This geometry is used instead of placing the
sensor at the line of the excitation light at a
180° angle in order to avoid interference of
the transmitted excitation light. No
monochromator is perfect and it will
transmit some stray light, that is, light with
other wavelengths than the targeted. An
ideal monochromator would only transmit
light in the specified range and have a high wavelength-independent transmission. When
measuring at a 90° angle, only the light scattered by the sample causes stray light. This results in
a better signal-to-noise ratio, and lowers the detection limit by approximately a factor 10000
when compared to the 180° geometry. Furthermore, the fluorescence can also be measured from
the front, which is often done for turbid samples.
The detector can either be single-channeled or multichanneled. The single-channeled
detector can only detect the intensity of one wavelength at a time, while the multichanneled
detects the intensity at all wavelengths simultaneously, making the emission monochromator or
filter unnecessary. The different types of detectors have both advantages and disadvantages.
The most versatile fluorimeters with dual monochromators and a continuous excitation
light source can record both an excitation spectrum and a fluorescence spectrum. When
measuring fluorescence spectra, the wavelength of the excitation light is kept constant,
preferably at a wavelength of high absorption (i.e., maximum absorbance, ex), and the emission
monochromator scans the spectrum. For measuring excitation spectra, the wavelength passing
through the emission filter or monochromator is kept constant and the excitation monochromator
is scanning. The excitation spectrum generally is identical to the absorption spectrum as the
fluorescence intensity is proportional to the absorption.
Excitation sources
A xenon arc lamp is an artificial light source. Powered by electricity, it uses ionized
xenon gas to produce a bright white light that closely mimics natural daylight. Xenon arc lamps
can be roughly divided into three categories:
Continuous-output xenon short-arc lamps
Continuous-output xenon long-arc lamps
Xenon flash lamps (which are usually considered
separately)
Each consists of a glass or fused quartz arc tube with tungsten
metal electrodes at each end. The glass tube is first evacuated
and then re-filled with xenon gas. For xenon flashtubes, a third
"trigger" electrode usually surrounds the exterior of the arc
tube.
Our spectrofluorometer is equipped with Continuous- 15 kW xenon short-arc lamp
output xenon short-arc lamps where the spectrum is between
200 nm and 650 nm.
Monochromator designs
A typical monochromator design is
shown on the right. It consists of a
diffraction grating (dispersing element),
slits, and spherical mirrors. The light source
emits a broad spectrum of radiation as
represented by the multi-colored line from
the lamp to the grating. (The yellow color of
the light source represents all colors.) The
diffraction grating disperses light by
diffracting different wavelengths at different
Schematic of a Czerny-Turner monochromator
angles. The grating is positioned so that
green light passes through the exit slit and
all other colors are blocked. The particular wavelength that passes through the monochromator is
selected by rotating the angle of the grating. The mirror and slit positions remain fixed. If this
grating was rotated clockwise slightly, what color light would pass through the exit slit?
Scanning a spectrum is accomplished by rotating the grating with a motor. The detector
measures the power of the light that strikes it, converting the light power to an electrical signal.
Monochromator parameters
Bandpass: The wavelength range that the monochromator transmits.
Dispersion: The wavelength dispersing power, usually given as spectral range / slit width
(nm/mm). Dispersion depends on the focal length, grating resolving power, and the grating
order.
Resolution: The minimum bandpass of the spectrometer, usually determined by the
aberrations of the optical system.
Acceptance angle (f/#): A measure of light collecting ability, focal length / mirror diameter
Blaze wavelength: The wavelength of maximum intensity in first order.
Emission spectra
Timescale
1. Fluorescence
S0 → Sn absorption
Sn → S1 internal conversion (10-11 - 10-14 sec)
Characteristics of Excited States
S1 → S0 + h fluorescence (10-7 - 10-9 sec)
a. Energy
S1 → Tn intersystem crossing (10-8 sec)
b. Lifetime
S1 → S0 internal conversion (10-5 - 10-7 sec)
c. Quantum Yield
d. Polarization T1 → S0 + h phosphorescence (10 - 10-3 sec)
T1 → S0 internal conversion (10 - 10-3 sec)
2. Phosphorescence occurs at longer wavelength than does fluorescence.
Often, the emission band is red-shifted
relative to the absorption band: "Stokes shift"
Quantum yield is the ratio of photons emitted to photons absorbed by the system: FF = kF / kF +
kISC + knr + kq + kr
Fluorescence decay of a pure sample follows a single exponential decay. The dark line
shows the excitation pulse. Time correlated single photon counting was used to obtain this data.
This technique counts the number of emitted photons hitting a detector at times, t, following
excitation.
One critical difference between steady-state and kinetic measurements of fluorescence is
that the value of tF is not a function of concentration of the sample while the value of FF is
concentration dependent. Only at low concentration is the value of FF linearly dependent on
concentration. The reason is the so-called inner filter effect.
b. Keep the excitation wavelength ((λex) constant and scan the emission wavelength
between λ > than λex and < 2 λex. This operation is to avoid the incident light going
directly into the photomultiplier.
Excitation spectrum
a. Select the wavelength at the maximum emission and keep constant at λem.
Conclusion
a. Compare the maxima in the emission spectrum and in the excitation spectrum
b. Calculate the values of those maxima in nm-1 (Dr. Rima: please clarify this questions or
show it how you do it?)
c. Mention the Rayleigh and Raman diffusion
d. Plot the fluorescence intensities in function of the concentrations
e. Determine limit of detection: Follow the intensity of the fluorescence with the dilution
until the signal cannot be detected. It is often defined by 3-fold of background noises.
References
1. A.J. Pesce, C.G. Rosen, and T.L. Pasby (eds.), “Fluorescence Spectroscopy: An Introduction
for Biology and Medicine”. Dekker, New York, 1971.
2. H.H. Bauer, G.D. Christian, and J.E. O’Reilly (eds.), “Instrumental Analysis”. Allyn &
Bacon, Boston, 1978, chap. 9.
3. H. H. Willard, L.L. Merritt, Jr., J.A. Dean, and F.A. Settle, Jr., “Instrumental Methods of
Analysis”. 6th ed., Van Nostrand, New York, 1981, Chap. 4.
4. G. W. Ewing, “Instrumental Methods of Chemical Analysis”. 4th ed., McGraw-Hill, New
York, 1975, chap. 4.
5. J.E. O’Reilly, J. Chem. Ed., 53, 191 (1976).
6. J.E. O’Reilly, J. Chem. Ed., 52, 610 (1975).
7. M.W. Legenza and C. J. Marzzacco, J. Chem. Ed., 54, 183 (1977).
8. J.D. Winefordner, S.G. Schulman, and T.C. O’Haver, “Luminescence Spectrometry in
Analytical Chemistry”. Wiley, New York, 1972, p. 293.
9. S. Udenfriend. “Fluorescence Assay in Biology and Medicine”. Academic Press, New York,
vol. I, 1962; vol II, 1969.
10. G.G. Guilbault, “Practical Fluorescence: Theory, Methods, Techniques”. Dekker, New
York, 1973.
11. C. E. White and R.F. Argauer, “Fluorescence Analysis. A Practical Approach”. Dekker,
New York, 1970.
12. E.D. Olsen, “Modern Optical Methods of Analysis”. McGraw-Hill, New York, 1975, chap.
8.
13. W. R Seitz in “Treats on Analytical Chemistry”. P.J. Elving, E.J. Meehan, and I.M. Kolthoff
(eds.), 2nd ed., part 1, vol. 7, Interscience, New York, 1981, chap. 4.
14. P. Atkins, J. de Paula, “Physical Chemistry”. 7th Ed., Oxford Univ. Press, New York, 2002.
15. A. Gilbert, J. Baggott, “Essentials of Molecular Photochemistry”. 1st Ed., Blackwell, Oxford,
1991.
16. D. Skoog, F. Holler, T. Nieman, “Instrumental Analysis”. 5th Ed., Harcourt, Philadelphia,
2002