Professional Documents
Culture Documents
rays, gamma rays, X-rays, ultra-violet radiation, visible light, infrared radiation, microwaves and
radio waves in the increasing order of wavelength and decreasing order of frequency
This requires even less energy than a n * transition and usually occurs
in the near UV region (300-400)
This requires least amount of energy among the four types of transitions
and usually occurs in the visible region (400 – 800 nm)
(ii)Double beam spectrophotometer: In this the light is split into two beams by a
beam splitter and one passes through the sample while the other passes through
a reference. The absorbance is calculated by comparing the ratio of the
intensities of two beams.
(a) Light source: It provides two lamps as sources of UV & Visible light.
(i) Deuterium lamp [D2] is used for UV region [200 to 400 nm]
(ii) Tungsten lamp is used for visible region [400 to 700nm]
Note: The generated heat from tungsten lamp may warm up the sample
and other components of the instrument, so that the lamp is often cooled with
an air blower.
(b) Filter:
Commonly used for reflection, transmission, absorption, and blocking UV
light or wavelengths, which they do without affecting, light within the visible
region.
(c) Monochromator:
Most of the bands that indicate what functional group is present are found in the
region from 4000 cm-1 to 1300 cm-1.
Their bands can be identified and used to determine the functional group of an
unknown compound.
Bands that are unique to each molecule, similar to a fingerprint, are found in the
fingerprint region, from 1300 cm-1 to 400 cm-1.
These bands are only used to compare the spectra of one compound to another
Fundamental modes
(a) Symmetric stretching: Terminal atoms move away from or towards each
other from the central atom, simultaneously.
(b) Aymmetric stretching: One atom moves away while the other atom moves
towards the central atom, simultaneously.
(ii) Bending vibrations for a methylene group (-CH2-) can be represented as
follows:
(a) Scissoring: Two adjacent atoms of a central atom move towards and away from
each other (like scissors) in the same plane.
(b) Rocking: Two adjacent atoms of a central atom move to the same side relative to
the central atom, in the same plane.
Now both of these beams are reflected to pass through a splitter and then
through a detector.
Finally, the required reading is printed out after the processor deciphers the data
passed through the detector.
IR Spectroscopy Instrumentation
The main parts of the IR spectrometer are as follows:
1. Radiation source
2. Sample cells and sampling of substances
3. Monochromators
4. Detectors
5. Recorder
1. IR radiation sources
Various sources of IR radiations are: Nernst glower, Incandescent lamp,
Mercury arc, Tungsten lamp, Glober source, Nichrome wire
(ii) Liquid – Samples can be held using a liquid sample cell made of alkali
halides. Aqueous solvents cannot be used as they will dissolve alkali halides. Only
organic solvents like chloroform can be used.
4. Detectors
Detectors are used to measure the intensity of unabsorbed
infrared radiation.
Detectors like thermocouples, Bolometers, thermistors, Golay
cell, and pyro-electric detectors are used.
5. Recorders
Recorders are used to record the IR spectrum.