You are on page 1of 1

CALIBRATION PLAN

MEASURING INSTRUMENT
Pressure recorder Pressure gauge Digital thermometer AC/DC Clamp tester Flaw detector (UT Equipment) Digital Thickness gauge Dumpy level Theodlite-Laser beam Surface profile gauge Hardness tester Oxygen indicator Dial micrometer Torque wrench Densitometer Baking oven Holding oven Temp gauge (Oven) Portable oven (Quiver) Welding machines

CALIBRATION INTERVAL(Max)
6 months 6 months 12 months 12 months 12 months 12 months 12 months 12 months 12 months 12 months 12 months 12 months 12 months 12 months 12 months 12 months 12 months 6 months 6 months

ACCEPTANCE LEVEL
1.25% 1.25% 1% PASS ALL FUNCTION TESTS PASS ALL FUNCTION TESTS PASS ALL FUNCTION TESTS PASS ALL FUNCTION TESTS PASS ALL FUNCTION TESTS 0.002mm 3%
2% 2%

CALIBRATED BY
EXTERNAL EXTERNAL EXTERNAL EXTERNAL EXTERNAL EXTERNAL EXTERNAL EXTERNAL EXTERNAL EXTERNAL EXTERNAL EXTERNAL EXTERNAL EXTERNAL INTERNAL INTERNAL INTERNAL INTERNAL INTERNAL

4% 7% 15 C 15 C 15 C 5 C 10 % (Amps)

Note: 1 The basis used for calibration acceptance level is obtained from the manufacturer's Recommendations & Standards. 2. External calibration shall be AWE approved supplier and certified by Third party. 3.The internal calibration/inspection shall be performed by the use of digital temp gauge as a Standard equipment.