You are on page 1of 2

How to use the Pressure Transmitter Selection Chart

The chart offers general support to pressure instrument selection. It presents information regarding available features and capabilities of the ST 3000 family*. When using the chart for pressure transmitter selection, it is suggested you keep the following 5 questions in mind. 1. What type of pressure is to be measured? (Or, what reference point is desired?) 2. What process connection(s) are required? 3. What wetted materials are suitable for the application? 4. What is the measurement span? 5. What level of instrument performance is required and/or useful? Use the resulting Key Model number(s) plus the Honeywell Smartline Software Selection Guide (or Model Selection Guide) to finalize your selection and specify a complete model number.
*Always review the appropriate Honeywell Specification and Technical Data publication prior to making a final instrument selection.

CAUTION ! Reference Accuracy Performance


External pressure & temperature effects are often of greater influence on installed instrument performance than reference accuracy specifications. Under actual operating conditions, pressure transmitters with identical reference accuracies can yield varying levels of measurement performance. A 0.1% transmitter can, in fact, outperform a 0.075% of span transmitter over a significant portion of their span. ST 3000s continuous, active pressure and temperature compensation deliver superior installed performance.
3.0 Total Probable Error 2.5
Error in % of Span

1000 psig Static Pressure 50 deg F (28C)

2.0

Honeywell Enhanced Brand R Honeywell Standard

1.5

1.0

0.5

0 10 25 50 100 250 400 Calibrated Span (*H2O)

Honeywell Standard Compensation Reference Accuracy Relative Price


Active Pressure & Temperature 0.1% of Span $$$

Honeywell Enhanced
Active Pressure & Temperature 0.075% of span $$$$

Brand R
Temperature 0.075% of span $$$$

Pressure Transmitter Selection Guide

Industrial Automation and Control


Honeywell Inc. http://www.iac.honeywell.com
U.S.A.: Honeywell Industrial Automation and Control, 16404 North Black Canyon Hwy., Phoenix, AZ 85053 Canada: The Honeywell Centre, 155 Gordon Baker Rd., North York, Ontario M2H 3N7 Latin America: Honeywell Inc., 480 Sawgrass Corporate Parkway, Suite 200, Sunrise, Florida 33325 Japan: Industrial Operations Tokyo, 4-28-1 Nishi-Rokugo Ohtu-ku, Tokyo 144, Japan Asia: Honeywell Asia Pacific Inc., Room 3213-3225, Sun Hung Kai Centre, No. 30 Harbour Road, Wanchai, Hong Kong Pacific Division: Honeywell Pty Ltd., 5 Thomas Holt Drive, North Ryde NSW Australia 2113 Northern Europe and Southern Africa: Honeywell Ltd., Honeywell House, Arlington Business Park, Bracknell, RG 12 1EB, U.K. Central Europe: Honeywell A.G., Kaiserleistrae 39, 63067 Offenbach, Germany Western and Southern Europe: Honeywell S.A., Avenue du Bourget 1, 1140 Brussels, Belgium Eastern Europe: Honeywell Praha, s.r.o., Budejovicka 1, 140 00 Prague 4, Czech Republic Middle East: Honeywell Middle East Ltd., Khalifa Street, Sheikh Faisal Building, Abu Dhabi, U.A.E.

SL-53-638 (5M) 1/99

Honeywell Inc.

Printed in U.S.A.

Printed on Recycled Paper

HONEYWELL PRESSURE TRANSMITTER SELECTION CHART


Model Range Covered
0 to 10 inH2O 0 to 400 inH2O 0 to 600 inH2O 0 to 100 psi 0 to 3000 psi 0 to 400 inH2O 0 to 100 psi 0 to 3000 psi (25 mbar) (1000 mbar) (1500 mbar) (7 bar) (210 bar) (1000 mbar) (7 bar) (210 bar)

Minimum Span
0.4 inH2O (1 mbar) 1 inH2O (2.5 mbar) 25 inH2O (62.2 mbar) 5 psi (0.35 bar) 100 psi (7 bar) 10 inH2O (25 mbar) 5 psi (0.35 bar) 100 psi (7 bar) 0.5 inH2O (2.5 mbar) 15 psia (1 barA) 1 inH2O (2.5 mbar) 15 psia (1 barA) 1 inH2O (2.5 mbar) 100 psi (7 bar)

Active Pres Active Temp Comp Comp


                     

Diaphragm Material*
S S, H, M, S S, H, M, S, H S, H, M, S, H, M, S, H S S, H, M, T S, H

Remote Seals

Flange Level

Temp Input

Performance Level
Enhanced Enhanced Enhanced Enhanced Enhanced Standard Standard Standard

Comm. Protocol**
DE, DE, DE, DE, DE, DE, DE, DE, HC, HC, HC, HC, HC, HC, HC, HC, DE DE DE FF FF FF FF FF FF FF FF

DP/Flow
STD110 STD120 STD125 STD130 STD170 STD924 STD930 STD974 SMA110 Multivariable**** SMA125 SMG170 T T T T Consult factory Consult factory Consult factory   

Single Variable

Differential Pressure
STD110 STD120 STD125 STD130 STD170 STD924 STD930 STD974

Multivariable Flow
SMA110 SMA125 SMG170

DP: 0 to 25 inH2O (62.2 mbar) AP: 0 to 100 psia (7 barA) DP: 0 to 400 inH2O (1000 mbar) AP: 0 to 750 psia (50 barA) DP: 0 to 400 inH2O (1000 mbar) GP: 0 to 3000 psi (210 bar)

Enhanced Enhanced Enhanced

Pressure
GP STG140 STG170 STG180 STG944 STG974 STG14L STG17L STG18L STG94L STG97L STG98L STR14G STR17G STR94G STG93P STG14T STF14T STA122 STA140 STA922 STA940 STR14A 0 to 500 psi 0 to 3000 psi 0 to 6000 psi 0 to 500 psi 0 to 3000 psi 0 to 500 psi 0 to 3000 psi 0 to 6000 psi 0 to 500 psi 0 to 3000 psi 0 to 6000 psi (35 bar) (210 bar) (415 bar) (35 bar) (210 bar) (35 bar) (210 bar) (415 bar) (35 bar) (210 bar) (415 bar) 5 psi (0.35 bar) 100 psi (7 bar) 100 psi (7 bar) 20 psi (1.4 bar) 300 psi (21 bar) 5 psi (0.35 bar) 100 psi (7 bar) 100 psi (7 bar) 20 psi (1.4 bar) 300 psi (21 bar) 500 psi (35 bar) 5 psi (0.35 bar) 100 psi (7 bar) 20 psi (1.4 bar) 5 psi (0.3 bar) 0.9 psi (63 mbar) 0.9 psi (63 mbar) 50 mmHgA (67 mbarA) 5 psia (0.35 barA 50 mmHgA (67 mbarA) 20 psia (1.4 barA) 5 psia (0.35 barA)                                   S, H, M S, H, M S, H S, H S, H S, S, S, S, S, S, H H H H H H    Enhanced Enhanced Enhanced Standard Standard Enhanced Enhanced Enhanced Standard Standard Standard Enhanced Enhanced Standard Standard Enhanced Enhanced Enhanced Enhanced Standard Standard  Enhanced             DE, DE, DE, DE, DE, DE, DE, DE, DE, DE, DE, HC, HC, HC, HC, HC, HC, HC, HC, HC, HC, HC, FF FF FF FF FF FF FF FF FF FF FF

GP In-Line

Flush Mount Gauge


STG93P

DualHead Gauge
STG944 STG974

In-Line Gauge
STG14L STG17L STG18L STG94L STG97L STG98L

Gauge & Absolute


STG140 STG170 STG180 STA122 STA140

High Temp. Gauge


STG14T STF14T
GP Remote

0 to 500 psi (35 bar) 0 to 3000 psi (210 bar) 0 to 500 psi (35 bar) 0 to 100 psi (7 bar) 0 to 500 psi (35 bar) 0 to 500 psi (35 bar) 0 to 780 mmHgA 0 to 500 psia 0 to 780 mmHgA 0 to 500 psia (1040 mbarA) (35 barA) (1040 mbarA) (35 bar)

Depends on seal Depends on seal Depends on seal S S S S, H, M S, H, M S, H S, H Depends on seal

DE, HC, FF DE, HC, FF DE, HC, FF DE, FF DE, FF DE, FF DE, DE, DE, DE, HC, HC, HC, HC, FF FF FF FF

GP Specialty

AP

AP Remote

0 to 500 psia (35 barA)

DE, HC, FF

Level
DP Pseudo STF12F STF13F STF14F STF92F STF93F STF128 STF132 STF924 STF932 STR12D STR13D STR93D 0 to 400 inH2O 0 to 100 psi 0 to 600 inH2O 0 to 400 inH2O 0 to 100 psi 0 to 400 inH2O 0 to 100 psi 0 to 400 inH2O 0 to 100 psi (1000 mbar) (7 bar) (1500 mbar) (1000 mbar) (7 bar) (1000 mbar) (7 bar) (1000 mbar) (7 bar) 1 inH2O (2.5 mbar) 5 psi (0.35 bar) 25 inH2O (62.2 mbar) 25 inH2O (62.2 mbar) 5 psi (0.35 bar) 10 inH2O (25 mbar) 5 psi (0.35 bar) 25 inH2O (62.2 mbar) 5 psi (0.35 bar) 10 inH2O (25 mbar) 5 psi (0.35 bar) 0.9 psi (62.2 mbar)             S, H, M S, H, M S S, H S, H S, H, M*** S, H, M*** S, H S, H Depends on seal Depends on seal Depends on seal Enhanced Enhanced Enhanced Standard Standard Enhanced Enhanced Standard Standard Enhanced Enhanced Standard DE, DE, DE, DE, DE, DE, DE, DE, DE, HC, HC, HC, HC, HC, HC, HC, HC, HC, FF FF FF FF FF FF FF FF FF

Flange Mount Liquid Level


STF128 STF132 STF924 STF932

DP w/ Pseudo Flange
STF12F STF13F STF92F STF93F STF14F
DP Flange

DP Remote

0 to 400 inH2O (1000 mbar) 0 to 100 psi (7 bar) 0 to 100 psi (7 bar)

DE, HC, FF DE, HC, FF DE, HC, FF

Remote Seals
STR12D STR13D STR93D

* S = 316 LSS H = Hastelloy C M = Monel T = Tantalum

**DE = Choice of 4-20mA or Honeywell Digitally Enhanced HC = HART FF = Foundation Fieldbus

*** May not be available in all variations **** Measures DP, Static P, Temperature and calculates Compensated Mass or Volumetric Flow

Low Ranges (inH2O / mbar) Med Ranges (psi / bar) High Ranges (psi / bar)