Professional Documents
Culture Documents
Fall, 2000
Chemistry 382/378
Role-Playing Lab in
Instrumental Analysis
Experiment #05
Polarography
Perhaps the most fascinating thing about solution electrochemistry is the fact that you literally stick
one end of a wire into the solution you want to analyze, and connect the other end to your
electronics and computer. There hardly could be a more graphic illustration of the interfacing of
computers, electronics, and chemistry than this. Of course, the wire is not simply a wire. It must
have certain properties in the solution, such as resistance to attack and the ability to conduct
electricity.
Mercury Fill
Leveling Bulb
Aluminum
Plate
Tygon
Tubing
Tight Fit
Tri-Pod Stand
Tight Fit
Capillary
Cell
Solution
Mercury Drop
3
FIRST AID MEASURES
Eyes:
Immediately flush eyes with plenty of water for at least 15 minutes, occasionally lifting the
upper and lower lids. Get medical aid immediately.
Skin:
Get medical aid. Immediately flush skin with plenty of soap and water for at least 15
minutes while removing contaminated clothing and shoes.
Ingestion:
Never give anything by mouth to an unconscious person. Get medical aid immediately.
Wash mouth out with water.
Inhalation:
Remove from exposure to fresh air immediately. If not breathing, give artificial respiration.
If breathing is difficult, give oxygen. Get medical aid.
Notes to Physician:
Treat symptomatically and supportively. The use of DMSA or BAL as antidotal treatment
should be determined only by qualified medical personnel (Medical Toxicology, 1988).
ACCIDENTAL RELEASE MEASURES
Spills/Leaks:
Vacuum or sweep up material and place into a suitable disposal container. Wear a self
contained breathing apparatus and appropriate Personal protection.
HANDLING and STORAGE
Handling:
Wash thoroughly after handling. Remove contaminated clothing and wash before reuse. Use
with adequate ventilation. Minimize dust generation and accumulation. Avoid breathing
dust, vapor, mist, or gas. Avoid contact with eyes, skin, and clothing. Keep container
tightly closed. Avoid ingestion and inhalation.
Storage:
Store in a cool, dry, well-ventilated area away from incompatible substances. Keep away
from metals (such as wedding or engagement rings).
PRODUCT NAME:
MERCURY (METAL)
FORMULA:
HG
FORMULA WT:
200.59
CAS NO.:
07439-97-6
NIOSH/RTECS NO.: OV4550000
HEALTH
FLAMMABILITY
REACTIVITY
CONTACT
4
0
1
3
EXTREME (POISON)
NONE
SLIGHT
SEVERE (LIFE)
4
Clearly, mercury is toxic. But, as long as there are no spills, as long as the whole electrode to
deliver the mercury to the solution has been constructed, assembled, and tested prior to the time you
have to use it, and provided there are safe methods available to you to fill and empty the cell
containing the solution into which the mercury flows, then the polarography experiment can be done
with only normal safe lab practices. That is the case here. When you come into this lab, you will
find two polarographic systems set up for you. You will have the opportunity to use them, and to
change the chemistry of the solutions being analyzed. But their assembly will have all been done for
you.
The Apparatus Pictured
The complete apparatus is pictured in Figure 2.
At the top of the apparatus is a leveling bulb
partially filled with mercury (Figure 3).
5
The bottom of the cell into which the DME is inserted
holds the solution being analyzed. This is shown in
Figure 5. Because the glass capillary and the glass cell
walls are equally clear, the DME as such is hard to see in
this picture.
There is a danger that in handling the solutions, the cell
parts (top and bottom) might become accidentally
separated. Since the bottom of the cell has a rather large
puddle of mercury in it after even a few hours of use, this
could be a disastrous happening. Thus, the cell parts are
firmly held together with a cam type of clamp, and the
cam is held in an engaged position by a plastic cable tie.
This means that the cell can only be filled and emptied by
a combination of syringe (to fill it) and vacuum aspirator
(to empty it). In this way, it is never necessary to take it
apart and come into any kind of contact with the mercury
it may contain.
Figure 5 The bottom of the DME cell
The vacuum pump that is used to aspirate out the contents
of the cell is shown in Figure 6. This pump pulls a vacuum
sufficient not only to draw the solution out of the cell into
the waste bottle (Figure 7), but also excess mercury that
accumulates over time in the bottom of the cell.
6
Nitrogen is used to deaerate the solution. This is done by
first bubbling nitrogen through the solution (called
sparging) to displace the dissolved oxygen. After a short
time, most of the oxygen has been removed and replaced
with nitrogen, which produces no electrochemical signal.
After that, a flowing layer of nitrogen is directed over the top
of the solution to keep it from absorbing oxygen from the air.
A two pronged gas delivery tube is used to control the
nitrogen. It is shown in Figure 8. Two hoses come to it.
When nitrogen flows through the front hose, it is directed
down into the solution. When it flows through the back
hose, it is directed over the top of the solution.
The nitrogen pathway is selected with a small three way
toggle valve as shown in Figure 9.
7
The DME apparatus is on a
cart, and the potentiostat is
placed in a plasticware box
on the back of the cart. A
digital voltmeter is used to
monitor the potentiostat input
and output voltages. A large
ribbon cable connects the 50
pin terminal block on the side
of the cart to a companion
Mac, 640AV computer
carrying LabVIEW and a
single NB-MIO-16L A/D
board. Solutions to be
analyzed are placed on a
small shelf at the front of the
cart. All of this is shown in
Figure 12.
This completes the
apparatus.
Figure 12 The potentiostat and computer interface cable
The Potentiostat
As shown in Figure 13,
the potentiostat is made
Voltage Follower
from 3 operational
with Gain
amplifiers. The top
amplifier receives the
+
Input from D/A
output from the D/A
9
Converter
converter as a voltage and
applies it to the counter
+ 9
electrode in the
Counter Electrode
electrochemical cell. The
Reference Electrode
Voltage Follower
reference electrode probes
+
Working Electrode
- 9
the voltage dropped
between the counter
electrode and the working
+ 9
electrode, picking off a
fraction of it close to the
Current Follower
working electrode. This
+
voltage is applied to the
- 9
second amplifier. The 2nd
Output to A/D
amplifier plus the
Converter
resistance between the
Figure 13 Basic 3 Electrode Potentiostat Circuit
counter electrode and the
reference electrode form
the feedback loop for the first amplifier. This sets the voltage applied to the working electrode to be
essentially the same as the output of the D/A converter, no matter what the solution or reference
electrode resistances are. Current in the working electrode is measured with the current follower, as
described on the next page.
+ 9
8
The key operating principle of the three
amplifier potentiostat shown in Figure 13 is the
virtual ground.
In the potentiostat, the potential is that applied by the voltage follower with gain to the working
electrode and reference electrode. The resistance is complex and depends on the diffusion that
occurs in solution. The current however, still is determined by just the applied potential and the
effective resistance. The current follower holds the potential of the working electrode at virtual
ground to ensure that this is so. This is illustrated in Figure 15 below.
D
C
NI NB-MIO-16L A/D,
D/A and DIO Board to
Output to Potentiostat
LabVIEW Potentiostat
Driver and Scan
Selection/Controls
Voltage
Follower
w/Gain
Current
Follower
CE RE WE
Electrochemical Cell
NI NB-MIO-16L A/D,
D/A and DIO Board to
Input from Potentiostat
LabVIEW Potentiostat
V/t, i/t, and i/V
Graphic Displays
Figure 15 The 3 amplifier potentiostat connected to the three electrode electrochemical cell
9
1
4
Reference Electrode
Reference Electrode
Reference Electrode
Reference Electrode
+
E
(0) -
4
Reference Electrode
Reference Electrode
Reference Electrode
Reference Electrode
+
E
(0) -
E
-
E
-
E
-
10
IN from
Working
Electrode
(DME)
IN from
Reference
Electrode
(SCE)
+9 vDC
Battery
OUT to
Counter
Electrode
(Pt)
Ground
Connector
In from D/A
Converter
Channel 0
8 7 6 5
8 7 6 5
8 7 6 5
2 3 4
2 3 4
2 3 4
Ground
Bus
(green
colored)
Ground
Connector
OUT to A/D
Converter
(Channel 0)
Ground
Connector
-9 vDC Power Bus (black colored all connected)
Battery
On-Off
Switch
-9 vDC
Battery
11
(+9 vDC)
(in)
(-)
8 7 6 5
2 3 4
(in)
(+)
(out)
(-9 vDC)
Pinouts for the type 741 Operational Amplifier
+ 9
Voltage Follower
with Gain
+
- 9
+ 9
Counter Electrode
Reference Electrode
Voltage Follower
+
- 9
Working Electrode
+ 9
-
Current Follower
+
- 9
Output to A/D
Converter
Figure 18 Wiring the 741 operational amplifiers to make the potentiostat in the plasticware box.
12
+9 v
-
-9 v
D/A
ground
OUT to
Counter
Electrode
(Pt)
IN from D/A
Converter
(Channel 0)
IN from
Reference
Electrode
(SCE)
DME
+9 v
2
D/A
ground
SCE
Pt
6
-9 v
OUT to
Counter
Electrode
(Pt)
A/D
ground
IN from
Working
Electrode
(DME)
+9 v
2
7
4
-9 v
ground
OUT to A/D
Converter
(Channel 0)
A/D
ground
Figure 19 Physical layout of the connections between the potentiostat and the electrochemical cell.
A/D Channel 0
Input Twisted Pair
13
D/A Channel 0
Twisted Pair
System
Ground
2 4
10 12 14 16 18 20 22 24 26 28 30 32 34 36 38 40 42 44 46 48 50
11 13 15 17 19 21 23 25 27 29 31 33 35 37 39 41 43 45 47 49
Reference
Jumper - do not
remove
Drop
Knocker
Line
Vac.
Line
N2
Line
Digital
Ground
D/A Channel 1
Twisted Pair
1
3
5
7
9
11
13
15
17
19
21
23
25
27
29
31
33
35
37
39
41
43
45
47
49
2
4
6
8
10
12
14
16
18
20
22
24
26
28
30
32
34
36
38
40
42
44
46
48
50
A/D Channel 0
Input Twisted Pair
D/A Channel 0
Twisted Pair
Drop
Knocker Vacuum
Line
Line
N2
Line
Digital
Ground
System
Ground
14
The connections shown in Figure 20 are of three types. The A/D twisted pair receives the output of
the current follower in the potentiostat and carries it to the NB board as an input signal. The D/A
twisted pair carries the DAC-0 channel as an output to the potentiostat. These leads are color coded,
and can be easily traced between the connector and the potentiostat. This should be done prior to
using the apparatus.
The drop knocker, nitrogen gas on/off vale, and aspiration vacuum pump are all switched on or off
by solid state relays controlling the primary 110 vAC power to them. One set of solid state relays
used for this are shown in Figure 20. The digital lines from the NB board are then used as outputs
to turn on or off the solid state relays. This is an excellent way to control devices that draw
significant amounts of AC power.
When you are in the lab, trace the lines that run from the 50 pin terminal block to the devices shown
in Figures 20 and 21. You will have to stretch to do this, since the solid state relays are underneath
the cart that holds the polarograph. Also trace out the ribbon cable that connects the terminal block to
the NB board in the Mac 640AV computer. The rest of the work involves the LabVIEW software on
that computer, so it is important that you have a good physical picture of what is being controlled
before you start building the LabVIEW potentiostat controller.
Circuit
Breaker
AC Line
Switch
Nitrogen Line
Switch 1
AC out
AC out
AC in
SSR 1
+
- black
+ red
Vacuum Pump
Switch 2
AC in
SSR 2
+
- black
Figure 21 Solid state relays used to control nitrogen valve and vacuum pump devices
+ red
15
The LabVIEW potentiostat is a VI that is best viewed in terms of how the DME is operated.
Consider the sequence portrayed in Figure 22 below.
Read Current
at DME vs. Ground
6 Apply Current to
Drop Knocker
Solenoid
Repeat Cycle
9 Until Done
Set Potential
to Initial Value
Change Potential
16
The experiment starts when the pinch clamp on the mercury flow line to the capillary is opened and
the mercury drops start free falling (1). Then, solution is added to the cell (2), and the potential set
to the desired starting point (3). Time is then allowed to pass while the mercury drop grows to some
size less than what it would take to freely detach (4), when the current is read (5). After this, the
solenoid controlling the drop knocker is powered (6), and the drop knocker plunger taps the side of
the cell, causing the drop to fall off (7). The potential then is set to a new value (8), and the whole
cycle is repeated (9) for as many times as it takes to complete the experiment.
One of these cycles takes 2 to 3 seconds. The number of cycles used to record the data depends on
the size of the potential change between drops.
To better see the timing, refer to Figure 23
3 Read Current
for Drop #1
Read Current
for Drop #2
13
Area of Drop
14
2
Allow Drop #1 to
Grow for time t
12
Allow Drop #2 to
Grow for time t
Step Potential
Cathodic by E
E
Allow Drop #3 to
Grow for time t
10
Step Potential
Cathodic by E
Applied Potential
Detach Drop #3
Detach Drop #2
Detach Drop #1
1
Set Potential at E
Read Current
for Drop #3
6
E + E
Time
11
E + 2E
t
t
Time
17
To compensate for small variations in the surface geometry of the drop when it is large, several
thousand current readings are usually taken in a fraction of a second and averaged.
Then, after the drop has been detached, the potential is stepped by an amount E. This usually is a
small potential, so that there is a need for many drops to be repeated to scan the total voltage from 0
to -1.7 or so.
13
14
15
16
17
18
19 20
21
22
12
11
10
6
9
8
24
23
2
18
The front panel also contains the controls that are needed to cause the potential to scan as shown,
plus other functions. For example, a switch is provided (6) to determine if the potential should be
scanned (as shown) or simply held at some initial value (11). The scan polarity can be set to be
cathodic or anodic with switch #5. The range of potential scan is determined by the final voltage
setting (12). When the VI is used with solid electrodes (instead of mercury) the scan usually starts at
some anodic initial voltage. This is set by the slider (10) as an initial anodic offset voltage.
The scanning voltage waveform is made using a cyclic function generator. To get a wave such as
shown here, only one half of a complete cycle is needed, and this is entered in control #14. There
are other possible waveforms to use for the scan. All waveforms are selected from the menu in
control #13.
The output potential does not scan smoothly. The D/A output is discrete, i.e., it occurs in steps. The
size of the step, E, and the time for which it is constant, t, is determined by the settings in
controls #15 and #16. This determines how long the drop is allowed to grow before the current is
read. Shortly after the current is read, the drop is detached.
The switches needed to turn on the drop knocker (17), the vacuum pump to aspirate out the cell
contents (18) and to turn on the nitrogen gas for deaeration (19) also are on the front panel of the
VI. These switches are queried only when the VI is first started, so it usually is best to turn the scan
switch #6 off, turn on the other switches desired, and then start and stop the VI. A bit of practice in
the lab will make this more clear.
Saving data to a spreadsheet for subsequent plotting and other manipulations is important. A button
is available (#20) to allow this. When it is depressed, the data shown in the XY plot (23-24) will be
written out to a file as named in path control #21, but not until after the scan has been completed. If
the scan is interrupted before it is plotted on XY graph 3, then nothing is saved. If the file name is
not changed between runs, then the new data are appended to the old, i.e., the file is not
overwritten, but just becomes longer.
The controls for smoothing (8,9) will be mentioned later. Half wave potentials and related currents
can be read by adjusting the positions of the cursors with the fine position diamond control (22), or
by simply dragging the cursor with the mouse. Readout (4) will show the cursor intersections.
The data shown on graph (3) in Figure 24 are for a solution containing 0.001 molar Cu+2 and 0.001
molar Zn+2 in 0.1 molar KCl and 5 drops of 0.2% Triton X-100. Wave 23 is for copper, and 24 for
Zinc.
19
The VI Diagram
The diagram for the potentiostat VI as set up for polarography is shown in Figure 25 below. There
are four parts to it. At the left are shown the sub-VIs used to generate the voltage scanning
waveform. In the middle is the FOR loop containing the frames used to do the timing sequence and
data acquisition shown previously in Figure 23. The right hand side contains the file writing and
smoothing routines. To build the VI, or understand it better, it is the central FOR loop that needs to
be understood first. The other functions follow from that.
20
A
Solenoid
P
I
S
AC
B
P
AC
21
Once the old drop has been knocked off,
and before the new drop has had an
appreciable chance to grow to a fragile
geometry, one step of the potential scan
either is, or is not, applied to the electrodes
through the NB-MIO-16L D/A converter.
Figure 28 shows the case where we wish
the potential to increase cathodically. The
output of the function generator is simply
passed through the TRUE case and
presented to the AO sub-VI. Channel 0 has
been selected on device 5. The voltage out
is read by the AI MULT PT Sub VI and
averaged before being displayed. Channel 1
is used.
22
flowing as the drop grows. A single reading would not be at all reproducible from drop to drop.
The result is multiplied and offset for display,
and passed out of the frame.
The manner in which the different waveforms are
generated to make the scan is interesting. This is
shown in Figure 31.
A sub-VI is used to generate the waveform. It
can output a triangle wave,
, a sine wave,
Figure 31 The waveform generator sub-VI
, a sawtooth,
, or a square wave,
23
24
How the Potentiostat Gathers and Presents Data to its Display
To see how the VI works to give a smooth display from a dropping electrode, where the drop area
changes continuously as a function of time, some expanded scale recorder tracings were prepared to
accompany the brief explanation given previously in Figure 23. These follow here.
25
26
Analytical Lab Work
You will be working as a Software/Chemist team when doing this lab. There are two DME
polarograph stands, and each team will have access to one of them.
There are several analytical projects that can be explored in this lab. They are:
1.
2.
3.
4.
Determine the effects of variable amounts of acid in an unbuffered solution on the expected
accuracy analyzing a set of unknowns.
5.
Determine the effects of other cations present in a natural sample matrix on the expected
accuracy analyzing a set of unknowns.
Project 1 - Determination of Unknown Concentration
This is the classic application for which the Nobel Prize was awarded. It is based on the fact that
under diffusion control, in an unstirred solution, the limiting current is directly proportional to
concentration of the reducible species.
To help Chemist with meeting this objective, a set of cadmium solutions and an unknown have
been prepared ahead of time for you. This means that all that Chemist has to do is load the cell
with the appropriate solution, add Triton X-100 maximum suppresser, deaerate, and let Software
take the scan and save the data. After that, Chemist would empty the cell using the aspiration pump,
refill it with another solution, and repeat the process. Software would compile a working curve
from the polarograms, and use it to determine the concentration of the unknown.
The recipe for the solutions is as follows:
Concentration
0 in KCl blank
0.0004 M Cd+2
0.0008M Cd+2
0.0012 M Cd+2
0.0016 M Cd+2
Ml. to add
~ 50
~ 50
~ 50
~ 50
~ 50
X-100 to add
6 drops
6 drops
6 drops
6 drops
6 drops
N2 time
10 minutes
10 minutes
10 minutes
10 minutes
10 minutes
Aspirate to
waste bottle
waste bottle
waste bottle
waste bottle
waste bottle
The unknown solution uses the same recipes as the standard solutions.
It will take some time to run these standards. During this time, all of the conditions in the
experiment must remain constant. One of the things that you should attempt to determine is over
what length of time the experimental parameters do in fact remain constant. I did some work this
summer and found that the data shown in Figures 35 and 36 occurred. This convinced me that you
should be able to hold your experimental conditions constant enough to do this determination for at
least a period of 2 to 3 hours.
27
Oxygen Reduction
Residual
Current
28
Figure 37 Cd working curve, taken from polarograms run over several hours
When the data are taken properly, using either a spreadsheet or the cursor method, the results can be
impressive. For example, in Figure 37 is shown one working curve that was prepared from
polarograms taken both in the morning and the afternoon. A straight line with good regression data
resulted, even though the DME was stopped and the system shut down for lunch, and restarted
from scratch after that. The DME characteristics, and all of the electronic settings, obviously held
over this long a period. Such performance cannot be guaranteed, but with careful techniques it
should be possible to determine several unknown concentrations over a full days work when the
standardization is done at the start of the day.
29
Project 2 - Effects of a Surfactant on Analytical Results
In preparing the solutions to use in meeting the first objective, you added 6 drops of a surfactant
called Triton X-100 to each solution. In the past, other compounds like gelatin also were added.
The function of these compounds is to coat the surface of the mercury drop as it forms with a
substance that will stabilize the flow that it undergoes while expanding. Mercury flows into the drop
from the inside, and then circulates inside of it. It swirls outward as the drop grows, and causes the
surface itself to flow. This leads to mixing in the immediate vicinity of the drop surface, and eddies
and cavities form in the solution that produces irregular streaming along the drop. All of this can
produce strange, non-diffusion effects when the potential is sufficiently cathodic that the surface
concentration of the analyte is approaching zero.
When a surfactant is added to the solution, the streaming effects at the drop surface are minimized,
and the polarographic wave approaches the limiting current smoothly. We expect then that there will
be better analytical results.
Consider the data shown in Figure 38 below.
30
Project 3 - Effects of Deaeration Time
Dissolved oxygen in test solutions has always been a headache in polarography, and in its
variations. Oxygen is reducible in two steps according to the follow half-reactions:
O2 + 2H 2 O + 2e- H2 O2 + 2OH-
E1/2 = -0.1
H2 O2 + 2e- 2OH-
E1/2 = -1.1
These two waves are irreversible, and drawn out over virtually the entire cathodic range. They also
are pH dependent, and will change height with mechanical agitation of the solution as oxygen is
mechanically displaced.
Our book, unfortunately, is not clear on what the oxygen waves will look like at a DME in just a
solution of supporting electrode. Figure 39, below, was scanned from Bard and Faulkner to give
you an idea of what to expect in our lab.
31
before passing it into the solution, how much to disperse it, how long to pass it through the
solution, and whether to blanket the solution with it after the deaeration has been done. Another
important question is whether to deaerate the solutions to be analyzed in a separate part of the
analysis (say, in a set of bottles before adding them to the cell), or to wait until the actual analytical
aliquot is in the polarographic cell and deaerate it there.
An example of one way to explore this objective is shown in Figure 40. Here, a Cd+2 test solution
was scanned both with and without deaeration. While there was little control over the design of the
study, it did show where the greatest obvious effect of deaeration would be on the Cd wave.
Clearly, you can come up with much better approaches.
32
Project 4 - Effects of Solution Acidity
Consider the half reaction :
2H+ +2e- H2
E1/2 = - 1.5
This implies that small and variable amounts of hydrogen ion could have an effect on the Cd+2
analysis. The fact that the half-wave potential is significantly larger than that for Cd+2 /Cd does
however suggest that it might not be a problem. Most chemists know though that running
unbuffered solutions raises problems in general with CO2 absorption. So, we need some ideas of
what the effects of [H+] will be on the analysis.
To gain some experimental ideas, I suggest that you study the poster outside SC328. Some strip
chart recordings that indicate at least one effect of [H+] are shown there. Also, since the supporting
electrolyte for the Cd+2 standards is KCl, the addition of small (0.001 M) amounts of HCl seems
sensible.
Again, the idea here is not just to observe the reduction of H+. Instead, we need to know if control
of the amounts of [H+] in a series of natural samples is important. And, always remember, there
may be ways to measure the diffusion current that eliminate the need for such control.
Objective 5 - Effects of a Natural Sample Matrix
In Figure 41 above is shown the polarogram of a mixture of cations. This, in one step, shows what
33
If the sample were collected downstream from the source, the chances are excellent that it would
have dissolved some of the soil over which it was flowing. This would contribute transition metals
to the sample, among other things. We thus could expect variable, and largely unpredictable,
contamination of the natural sample with Cu+2 , Ni +2 and Zn+2 cations, since these are common to
soils. Since we would hope that there would not be significant Cd+2 in the sample, it might even
occur that the contaminants would be present at a higher concentration than the analyte. Such is the
nature of a matrix effect.
The central question is what effect variable amounts of these contaminants would have on the
analysis. Note in Figure 41 that the 0.1 M NH4Cl supporting electrolyte has buffered the solution to
be basic. This extends the range of the DME cathodic scan enough that all of the cations produce a
detectable wave.
As you explore this objective, keep in mind that you can measure the diffusion current from these
ions separately. You also can transfer the data to a spreadsheet, and in a series of columns literally
subtract out the current from one cation from the net current produced by more than one. These kind
of manipulations presuppose a linearity between diffusion current and concentration for all of the
cations, but, given the excellent behavior for Cd+2 , this does not seem and unrealistic assumption.
Again, keep in mind the central objective. How does the presence of unpredictable amounts of
contaminant cations, such as those shown in Figure 41, effect the accuracy with which Cd+2 can be
determined?
Project 5 - A Class Project
The entire class can do an interesting project that relates to mixtures. Each lab section will have to
make up a set of solutions that constitutes one part of a larger mixture. The section will then take the
polarograms of each set of solutions, and pass the data on, via the Analytical Chemistry server, to
the next section. At the staff meeting in the following week, the data can be assembled to complete
the project.
As an example of this kind of project, view the figures on the following pages. Solutions were
prepared that had a base concentration of copper, and to each were added increasing amounts of
zinc. The question that was to be answered was, Could varying amounts of copper be determined
in solutions that had varying amounts of zinc in them, and could varying amounts of zinc be
analyzed in solutions that had varying amounts of copper in them?. While the solutions were not
hard to make, they took at least 20 minutes each to deaerate, and about 10 minutes to scan, so each
set of data took an afternoon to gather.
A class project that could be done involving the same kind of question would be to use cations other
than copper and zinc. You would refer to a table of half wave potentials (in the Science Library) to
select these ions.
34
0.200
0.175
0.150
0.125
0.100
0.075
0.050
0.025
0.000
-0.025
0.0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1.0
1.1
1.2
1.3
1.4
0.225
0 ppM Cu 0 ppM Zn
10 ppM Cu 0 ppM Zn
0.200
10 ppM Cu 10 ppM Zn
10 ppM Cu 20 ppM Zn
10 ppM Cu 30 ppM Zn
0.175
0.150
0.125
0.100
0.075
0.050
0.025
0.000
-0.025
0.0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1.0
1.1
1.2
1.3
1.4
Figure 43 These data would be used to determine if Zn can be quantitated in the presence
of 10 ppM Cu.
35
Variable Zinc, Fixed Copper, pH = 3.7
6/18/97 - John Walters - Bronze Analysis Development Study
0.325
0 ppM Cu 0 ppM Zn
0.300
20 ppM Cu 0 ppM Zn
0.275
20 ppM Cu 10 ppM Zn
0.250
20 ppM Cu 20 ppM Zn
20 ppM Cu 30 ppM Zn
0.225
0.200
0.175
0.150
0.125
0.100
0.075
0.050
0.025
0.000
-0.025
0.0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1.0
1.1
1.2
1.3
1.4
Figure 44 These data would be used to determine if Zn can be quantitated with the same working
curve slope and intercept in the presence of 20 ppM Cu as was observed with 10 ppM Cu.
Variable Zinc, Fixed Copper, pH = 3.7
6/17/97 - John Walters - Bronze Analysis Development Study
0.450
0.425
0 ppM Cu, 0 ppM Zn
0.400
0.375
0.350
0.325
0.300
0.275
0.250
0.225
0.200
0.175
0.150
0.125
0.100
0.075
0.050
0.025
0.000
-0.025
0.0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1.0
1.1
1.2
1.3
Figure 45 These data would repeat the Zn quantitation question with 30 ppM Cu.
1.4
36
Reporting the Results
Different teams in the class will select these objectives. Many of them will be done on different lab
days. To report out the results of all of them, a special Monday meeting will be scheduled to allow
short 10-minute presentations, with appropriate visual aids, of what the conclusions are.
Asking that an elected class representative present an answer to the question, Is polarography an
effective analytical method for the routine determination of sub-millimolar amounts of Cd+2 in
natural samples, will emphasize whole class communication?
This is one of several lab section laterals that we will do this semester. Success in an experiment
like this requires that the people in one lab section communicate ahead of time to those in the next
what has happened to them and how well the results met their expectations. Good communication
can be done by e-mail while the lab is being done. The analytical Chemistry server can be used to
drop ship results, or they may be attached to an Eudora message.
The reason for such efforts is shown on the next page in the form of a survey done by Dr. Robert
Gentry of the University of Minnesota. No matter what you have done, how you share it with
others is the most valuable skill you can have.
37
Dr. Ron Gentry, Chair of the Chemistry Department, reported the following material to Chemistry
faculty and students of the University of Minnesota on October 4, 1994. In an accompanying
memo, he stated:
A committee of the Council for Chemical Research this year polled major chemical
companies about the characteristics they look for when assessing M.S. or Ph.D. job
candidates in chemistry and chemical engineering. Attached are the responses, which should
be interesting and helpful for those preparing for industrial interviews.
When seeking a graduate degree level (M.S. or Ph.D.) candidate in Chemistry or Chemical
Engineering, which of the following are most important characteristics?
Thesis Topic
Thesis Advisor/Professor
Academic Institution Granting Advanced Degree
Grade Point Average - Graduate School
Grade Point Average - Undergraduate School
Undergraduate School
Undergraduate Discipline/Curriculum
Communications Skills
Extracurricular Activities
Foreign Language Skills
Professional/Honor Society Membership
Papers Presented (Number)
Publications (Number)
Computer Skills
Participation in co-op Programs
Full/Part-Time Employment in Major Field
II.
Independence
Self Motivation
Good Problem Solving Skills
Team Player
Enthusiasm
Decision Making
Professionalism
21
57
55
43
36
27
26
31
11
11
18
24
29
20
13
0
2
5
6
10
17
1
0
0
2
1
1
3
0
0
0
0
0
0
2