Professional Documents
Culture Documents
Appendix 1
s
w
o
I
a
i
d
v
rss
rw X 100
Procedure
M. S
Determination of Conductivity of
Water for Pharmaceutical Use
This method is used for the monitor of total amount of
electrolyte present in water for pharmaceutical use by
measure of water conductivity. Conductivity is a physical
quantity which represents a materials ability to conduct an
electric current. The conductivity of a solution is, by
definition the reciprocal of resistivity. The unit of
conductivity is S/cm (Siemens) or pS/crru
For purified water, to some extent, water molecules dissociate
into hydrogen ions and hydroxide ions, and the conductivity
could be detected irrespective of pure waters poorly ability to
Appendix V A -83
conduct electricity. Water conductivity is closely correlated to
the purity of water, the purer the water, the lower the
conductivity, and vice versa. Some atmospheric gases, such as
carbon dioxide, readily dissolve in water and interact to form
ions, which predictably affect conductivity. Water conductivity
is also affected by the presence of extraneous ions, which raises
the conductivity. Additionally, the water conductivity varies as
a function of pH and temperature.
Instrument specifications and operating parameters
Water conductivity must be measured accurately using
calibrated instrumentation. The conductivity cell contains 2
parallel electrodes which are generally protected by a glass
tube. Other types of cells may also be used. The frequency
of recalibration is a function of instrument design, degree of
use, etc. The cell constant can be verified directly by using
a solution of known conductivity, or indirectly by comparing
the instrument reading taken with the cell in question to
readings from a cell of known or certified cell constant The
conductivity cell constant must be known within 2% .
Each scale on the meter may require separate calibration prior
to use. The instrument must have a minimum resolution of
0 .1 f j.S / c m on the lowest range and the instrument accuracy
must be 0. 1 ^tS/cm.
Because temperature has a substantial impact on conductivity
readings, many instruments automatically correct the actual
reading to display the value theoretically. This temperature
compensation algorithm may not be accurate and is not applied
for the water conductivity. Conductivity values used in this
method are
nontemperatur^compensated
measurements.
Accuracy of the temperature measurement must be 2C .
Procedure
1. Purified Water
Determine the conductivity off-line or in-line and record the
temperature. The corresponding conductivity value at the
recorded temperature in Table 1 is the limit. The water to be
examined meets the requirements if the measured
conductivity is not greater than the value in Table 1. For
temperatures not listed in Table 1 calculate the maximal
permitted conductivity by interpolation between the next
lower and next higher data points in the table. If the
measured conductivity is not greater than the conductivity
requirements, the water to be examined meets the
requirements of the test for conductivity. If the measured
conductivity is greater than this value the water to be
examined does not meet the requirements of the test for
conductivity.
Table 1 Temperature and conductivity
requirements (purified water)
T emperature/C
Conductivity/pS cm' 1
2.4
0
10
3.6
20
4,3
5. 1
25
30
40
5.4
6
.5
50
7. 1
60
8 .1
70
9. 1
75
9.7
80
9.7
90
100
9.7
.2
10
Where
= ( { ) X (K- K
)+
Kq
0 .6
0 .8
10
0.9
15
1 .0
.1
20
25
1. 3
30
1.4
35
1.5
40
1.7
45
1 .8
50
1.9
55
60
2 .2
65
2.4
70
2.5
75
2.7
80
2.7
85
2.7
90
2.7
95
2.9
100
3. 1
.1
<
A - 84
Appendix
continue
pH
Conductivity/fxS crrf1
5.0
4.7
2.4