Professional Documents
Culture Documents
General
Specifications
GS 01C25F01-01E
STANDARD SPECIFICATIONS
SPAN AND RANGE LIMITS
(For EJX510A, values are in absolute and lower
range limits are 0.)
Measurement
Span/Range
A
B
C
D
MPa
psi (/D1)
bar (/D3)
kgf/cm2 (/D4)
0.08 to 2
Span
8 to 200 kPa
1.16 to 29
0.08 to 2
Range
14.5 to 29
1 to 2
1 to 2
Span
0.04 to 2
5.8 to 290
0.4 to 20
0.4 to 20
Range
0.1 to 2
14.5 to 290
1 to 20
1 to 20
Span
0.2 to 10
29 to 1450
2 to 100
2 to 100
Range
0.1 to 10
14.5 to 1450
1 to 100
1 to 100
Span
1 to 50
145 to 7200
10 to 500
10 to 500
Range
0.1 to 50
14.5 to 7200
1 to 500
1 to 500
T01E.EPS
PERFORMANCE SPECIFICATIONS
Zero-based calibrated span, linear output, wetted
parts material code S and silicone oil, unless
otherwise mentioned.
Specification Conformance
EJX series ensures specification conformance to at
least 3.
Reference Accuracy of Calibrated Span
(includes the effects of terminal-based linearity,
hysteresis, and repeatability)
Measurement span
Reference Span X
accuracy Span X
X
URL
(Upper range limit)
0.1% of Span
(0.01+0.009 URL/Span) % of Span
20 kPa
(2.9 psi)
200 kPa
(29 psi)
0.2 MPa
(29 psi)
2 MPa
(290 psi)
1 MPa
(145 psi)
5 MPa
(720 psi)
10 MPa
50 MPa
(1450 psi) (7200 psi)
T02E.EPS
Vibration Effects
Less than 0.1% of URL when tested per the requirements of IEC60770-1 field or pipeline with high
vibration level (10-60 Hz, 0.21 mm peak to peak
displacement/60-2000 Hz 3g)
Mounting Position Effects
Rotation in diaphragm plane has no effect. Tilting up
to 90 degree will cause zero shift up to 0.21 kPa
(0.84 inH2O) which can be corrected by the zero
adjustment.
Response Time (All capsules)
95 msec
When software damping is set to zero and including
dead time of 45 msec (nominal)
FUNCTIONAL SPECIFICATIONS
Output
Two wire 4 to 20 mA DC output with digital communications, linear or square root programmable. BRAIN
or HART FSK protocol are superimposed on the 4 to
20 mA signal.
Output range: 3.6 mA to 21.6 mA
Output limits conform to NAMUR NE43 can be preset by option code C2 or C3.
Failure Alarm
Output status at CPU failure and hardware error;
Up-scale: 110%, 21.6 mA DC or more (standard)
Down-scale: 5%, 3.2 mA DC or less
Note: Applicable for output signal code D and E
GS 01C25F01-01E
Copyright Jun. 2004
1st Edition Jun. 2004
<<Contents>> <<Index>>
Update Period
Pressure: 45 msec
SIL Certification
EJX series transmitters are certified by RWTV
Systems GmbH in compliance with the following
standards;
IEC 61508: 2000; Part1 to Part 7
Functional Safety of Electrical/electronic/programmable electronic related systems; Type B;
SIL 2 (single use)
Pressure
A and B
EJX530A
10(75)
2.7(20)
Working
pressure
1(7.5)
kPa abs
(mmHg abs)
Applicable range
0.13(1)
0.1(0.75)
0.013(0.1)
0.01(0.075) -40
(-40)
0
(32)
40
(104)
80
120 (248)
(176) 85
(185)
Process temperature C(F)
F01E.EPS
Atmospheric
pressure
100(14.5)
Working
pressure
kPa abs
(psia)
Applicable range
10(1.4)
1(0.14)
-40
(-40)
EJX510A
2.7(0.38)
Capsule
Pressure
Capsule
EJX530A
0
(32)
40
(104)
80
(176)
120
(248)
GS 01C25F01-01E
<<Contents>> <<Index>>
Supply & Load Requirements
HART
Communication Distance
Up to 1.5 km (1 mile) when using multiple twisted pair
cables. Communication distance varies depending on
type of cable used.
Use the following formula to determine cable length
for specific applications:
65 106 (Cf + 10,000)
L=
C
(R C)
600
R=
E-10.5
0.0244
External
load
resistance
Where:
L = length in meters or feet
R = resistance in (including barrier resistance)
C = cable capacitance in pF/m or pF/ft
Cf = maximum shunt capacitance of receiving devices
in pF/m or pF/ft
Digital
Communication
range
BRAIN and HART
250
R ()
42
PHYSICAL SPECIFICATIONS
Wetted Parts Materials
F03E.EPS
Supply Voltage
10.5 to 42 V DC for general use and flameproof type.
10.5 to 32 V DC for lightning protector (option code A).
10.5 to 30 V DC for intrinsically safe, type n, nonincendive or non-sparking type.
Minimum voltage limited at 16.6 V DC for digital
communications, BRAIN and HART
Load (Output signal code D and E)
0 to 1290 for operation
250 to 600 for digital communication
Communication Requirements
(Approval codes may affect electrical requirements.)
BRAIN
Communication Distance
Up to 2 km (1.25 miles) when using CEV polyethylene-insulated PVC-sheathed cables. Communication
distance varies depending on type of cable used.
Load Capacitance
0.22 F or less
Load Inductance
3.3 mH or less
Input Impedance of communicating device
10 k or more at 2.4 kHz.
Connections
Refer to MODEL AND SUFFIX CODE.
GS 01C25F01-01E
<<Contents>> <<Index>>
Suffix Codes
-D
-E
Measurement
A
span (Capsule)
B
C
D
Wetted parts
material
S
H
Process connections
4
7
8
9
-0
Amplifier housing
1
Electrical connection
0
2
4
5
7
9
Integral indicator
D
E
N
Mounting bracket
F
N
Description
Absolute pressure transmitter
Gauge pressure transmitter
4 to 20 mA DC Output with digital communication (BRAIN protocol)
4 to 20 mA DC Output with digital communication (HART protocol)
8 to 200 kPa
(1.16 to 29 psi)
0.04 to 2 MPa
(5.8 to 290 psi)
0.2 to 10 MPa
(29 to 1450 psi)
1 to 50 MPa
(145 to 7200 psi)
Process connection
Diaphragm
316L SST
Hastelloy C-276 *1
Hastelloy C-276 *1
Hastelloy C-276 *1
1/2 NPT female
1/2 NPT male
G1/2 DIN 16 288 male *2
M201.5 DIN 16 288 male *2
Always N
Always 0
Cast alluminum alloy
G1/2 female, one electrical connection without blind plugs
1/2 NPT female, two electrical connections without blind plugs
M20 female, two electrical connections without blind plugs
G1/2 female, two electrical connections with a blind plug
1/2 NPT female, two electrical connections with a blind plug
M20 female, two electrical connections with a blind plug
Digital indicator
Digital indicator with the range setting switch
None
304 SST 2-inch pipe mounting
None
T03.EPS
The marks indicates the most typical selection for each specification. Example: EJX530A-DAS4N-012NN/.
*1:
Hastelloy C-276 or N10276.
*2:
Not applicable for combination of capsule code D and wetted parts material code H.
GS 01C25F01-01E
<<Contents>> <<Index>>
CENELEC ATEX
Canadian Standards
Association (CSA)
Description
FM Explosionproof Approval *1
Explosionproof for Class I, Division 1, Groups B, C and D
Dust-ignitionproof for Class II/III, Division 1, Groups E, F and G
in Hazardous locations, indoors and outdoors (NEMA 4X)
Temperature class: T6
Amb. Temp.: 40 to 60C (40 to 140F)
FM Intrinsically safe Approval *1
Combined FF1 and FS1 *1
CENELEC ATEX (KEMA) Flameproof Approval *1
II 2G,1D EExd IIC T4, T5, T6
Amb. Temp. (Tamb) for gas-proof : T4 ; 50 to 75C (57 to 167F),
T5; 50 to 80C (57 to 176F), T6; 50 to 70C (57 to 158F)
Max. process Temp.(Tp): T4; 120C (248F), T5; 100C (212F), T6; 85C (185F)
Max. surface Temp. for dust-proof :
T80C (Tamb: 40 to 40C, Tp:80C), T100C (Tamb: 40 to 60C, Tp:100C),
T120C (Tamb: 40 to 80C, Tp:120C)
CENELEC ATEX (KEMA) Intrinsically safe Approval *1
Combined KF2, KS2 and Type n *1
CSA Explosionproof Approval *2
CSA Intrinsically safe Approval *2
Combined CF1 and CS1 *2
Code
FF1
KF2
T04.EPS
GS 01C25F01-01E
<<Contents>> <<Index>>
OPTIONAL SPECIFICATIONS
Description
Item
Color change
Painting
Code
P
PR
*1
X2
Lightning protector
Status output *2
AL
Low level: 0 to 2 V DC
K1
Oil-prohibited use
K2
K3
Output limits
and failure operation *4
D1
(See Table for Span and
Range Limits.)
*6
D3
D4
Failure alarm down-scale: Output status at CPU failure and hardware error is 5%,
3.2mA DC or less.
C1
C2
C3
N4
CA
Software damping
CB
Process Connector
M15
kgf/cm2) *8
kgf/cm2) *9
T05
Nitrogen(N2) Gas or Water *11
Retention time: one minutes
T06
T07
T08
T05.EPS
*1:
*2:
*3:
*4:
*5:
*6:
*7:
*8:
*9:
*10:
*11:
GS 01C25F01-01E
<<Contents>> <<Index>>
DIMENSIONS
Model EJX510A and EJX530A
With process connections code 7
95(3.74)
188(7.4)
60(2.36)*2
Zero
adjustment
110(4.33)
Conduit
connection
12
(0.47)
Electrical connection
for code 5 and 9.
77(3.03)
163(6.42)
Integral
indicator
(optional)
169(6.65)
129(5.08)
6
(0.23)
39
(1.54)
69
(2.72)
54
(2.13)
110(4.33)
External indicator
Conduit connection
(optional)
Ground
terminal
Open to
atmosphere*1
2-inch pipe
(O.D. 60.5mm)
Mounting
bracket
(optional)
129(5.08)
12*3
(0.47)
158(6.22)
5(0.20)
174(6.85)
20(0.78)
6(0.23)
F04.EPS
Terminal Wiring
Terminal Configuration
Communication
terminals (BT200 etc.)
connection hook
Check meter
connection hook*1*2
+
SUPPLY
CHECK +
or
PULS
ALARM
SUPP
LY
Ground terminal
K
CHEC M
ALAR
SUPPLY +
SUPPLY
CHECK + or ALARM +
CHECK or ALARM
GS 01C25F01-01E
<<Contents>> <<Index>>
Description
Connection
Analog output
Distributor
CHECK
Analog and
status output
(when /AL is specified)
If shield cable is not
used, communication is
not possible.
250
Shielded cable
ALARM
24V DC
24V DC
Distributor
250
+
Magnetic
valve
AC power supply
F06.EPS
As specified in order
2.00 sec or as specified in order
Calibration range
lower range value As specified in order
Calibration range As specified in order
upper range value
[EJX530A]
Selected from mmH2O, mmH2O(68F),
mmAq, mmWG, mmHg, Pa, hPa, kPa,
MPa, mbar, bar, gf/cm2, kgf/cm2, inH2O,
inH2O(68F), inHg, ftH2O, ftH2O(68F) or
psi. (Only one unit can be specified)
Calibration range
[EJX510A]
units
Torr, Pa abs, hPa abs, kPa abs, MPa abs,
mbar abs, bar abs, kgf/cm2 abs,
mmH2O abs, mmH2O abs(68F),
mmHg abs, inH2O abs, inH2O abs(68F),
inHg abs, ftH2O abs, ftH2O abs(68F),
psia, atm.
Display setting
GS 01C25F01-01E