2010-05 Index: 01
Service booklet
ATMOS S 61 Servant
Introduction, care
and safety instructions
You have bought an innovative medical product with the latest state-of-the-art
technology.
You should have untroubled pleasure working with your treatment system
ATMOS S 61 Servant for a long time.
Preconditions for this are safety and reliability.
In order to achieve this, regular maintenance and care is necessary.
Protect your investment:
Care and safety controls, together with the professional application,
guarantee the operational reliability and availability for use of the ATMOS
S 61 Servant. This, as well as regular cleaning is essential.
Repair work and safety controls may only be performed by an authorized
ATMOS technician. With the use of the original spare parts you will be
guaranteed that the operational reliability and availability for use as well as
the value of your ATMOS S 61 Servant is maintained.
Please pay attention to the following:
n
n
The service technician should check the display (compressed air and
vacuum indicator) annually.
Please note: Please keep this service booklet at hand in case of a warranty
claim. You should provide us with the corresponding data in case of a warranty
claim.
Type plate
workstation
Type plate
vision
Type plate
instruments
Type plate
others
Type plate
others
Verantwortlicher /
responsible
1. Operator self check; Annex A
2. Final inspection;
Annex B
3. Order picking according
to consignment list
ok
Unterschrift / signature
FMS
FMS
FV
115 Vac [ ]
127 Vac [ ]
230 Vac [ ]
6. Service
Checkup performed
completely
7. Customer
- delivery obtained according to order
bergabe
Delivery
bergabe bestanden
Delivery passed
red LED
yellow LED
The yellow LED shows several safety circuits (see operating instructions).
The red LED signalizes the service according to the index.
Once the device is switched on the status is indicated on the display!
first light up
second light up
third light up
fourth light up
fifth light up
sixth light up
seventh light up
see operating
instructions
red LED
Possible service case:
Runtime pressure pump
Runtime suction pump
Runtime suction pump 2
Runtime hose pump
Runtime drain pump
Annual maintenance
Battery voltage
yes
yes
yes
yes
yes
yes
yes
Remark
runtime critical
runtime critical
runtime critical
runtime critical
runtime critical
annual maintenance
low battery voltage
The monitoring of the burning life of the halogen bulb on the ATMOS S 61 Servant vision:
The burning life of the halogen bulb is monitored by the control. If 90% of the bulbs burning life is
achieved (nominally indicated by the manufacturer) the LED on the activation button is no longer
green but turns orange.
If one of the above service cases occurs, please contact your local ATMOS service partner or contact
the ATMOS service centre, phone +49 7653 689-222.
Confirmation and
reference to
warranty
Confirmation
The ATMOS service technician performs all the services on schedule and confirms the proceedings
in this booklet.
Your ATMOS S 61 Servant receives a seal of approval stating the date for the next service
appointment.
The scope of the service does not include any repair work, for this a separate order must be given.
Reference to warranty
Please follow the tips in this service booklet and comply with the same. This is the only way to ensure
your warranty claim is maintained.
During the warranty if the prescribed services are not performed by an authorized ATMOS service
technician within the specified time, then a warranty claim may only be settled when the
manufacturers inquiry results are presented.
Service Report
Maintenance
contract:
Basic
Premium
None
Remarks
Functional test
passed
not passed
_________________________
passed
not passed
_________________________
Replaced:
Ear irrigation bowl
REF 505.0353.0
Yes No
_________________________
Drainage hose
REF 510.0381.0
Yes No
_________________________
Tecalan hose
_________________________
_________________________
REF 502.0844.0
Yes No
_________________________
REF 530.1180.0
Yes No
_________________________
Yes No
_________________________
Double-barrelled hose
REF 530.1181.0
Yes No
_________________________
Sound absorber
_________________________
Back-up battery
REF 011.1274.0
Yes No
_________________________
Suction hose
REF 530.1144.0
Yes No
_________________________
Sound absorber
REF 069.0131.0
Yes No
_________________________
REF 530.1110.0
Yes No
_________________________
Consumables replenished:
Cleaning agent
Quantity
REF 080.0005.0
Yes No
_________________________
Yes No
_________________________
Halogen bulb 15 V
REF 506.5227.0
Yes No
_________________________
Halogen bulb 24 V
REF 507.2083.0
Yes No
_________________________
Waste bags
REF 505.0515.0
Yes No
_________________________
Bacterial filter
REF 340.0054.0
Yes No _________________________
_____________________________
_________________________________________
Date
signature ATMOS service
MedizinTechnik
Service Report
Maintenance
contract:
Basic
Premium
None
Remarks
Functional test
passed
not passed
_________________________
passed
not passed
_________________________
Replaced:
Ear irrigation bowl
REF 505.0353.0
Yes No
_________________________
Drainage hose
REF 510.0381.0
Yes No
_________________________
Tecalan hose
_________________________
_________________________
REF 502.0844.0
Yes No
_________________________
REF 530.1180.0
Yes No
_________________________
Yes No
_________________________
Double-barrelled hose
REF 530.1181.0
Yes No
_________________________
Sound absorber
_________________________
Back-up battery
REF 011.1274.0
Yes No
_________________________
Suction hose
REF 530.1144.0
Yes No
_________________________
Sound absorber
REF 069.0131.0
Yes No
_________________________
REF 530.1110.0
Yes No
_________________________
Consumables replenished:
Cleaning agent
Quantity
REF 080.0005.0
Yes No
_________________________
Yes No
_________________________
Halogen bulb 15 V
REF 506.5227.0
Yes No
_________________________
Halogen bulb 24 V
REF 507.2083.0
Yes No
_________________________
Waste bags
REF 505.0515.0
Yes No
_________________________
Bacterial filter
REF 340.0054.0
Yes No _________________________
_____________________________
_________________________________________
Date
signature ATMOS service
MedizinTechnik
Service Report
Maintenance
contract:
Basic
Premium
None
Remarks
Functional test
passed
not passed
_________________________
passed
not passed
_________________________
Replaced:
Ear irrigation bowl
REF 505.0353.0
Yes No
_________________________
Drainage hose
REF 510.0381.0
Yes No
_________________________
Tecalan hose
_________________________
_________________________
REF 502.0844.0
Yes No
_________________________
REF 530.1180.0
Yes No
_________________________
Yes No
_________________________
Double-barrelled hose
REF 530.1181.0
Yes No
_________________________
Sound absorber
_________________________
Back-up battery
REF 011.1274.0
Yes No
_________________________
Suction hose
REF 530.1144.0
Yes No
_________________________
Sound absorber
REF 069.0131.0
Yes No
_________________________
REF 530.1110.0
Yes No
_________________________
Consumables replenished:
Cleaning agent
Quantity
REF 080.0005.0
Yes No
_________________________
Yes No
_________________________
Halogen bulb 15 V
REF 506.5227.0
Yes No
_________________________
Halogen bulb 24 V
REF 507.2083.0
Yes No
_________________________
Waste bags
REF 505.0515.0
Yes No
_________________________
Bacterial filter
REF 340.0054.0
Yes No _________________________
_____________________________
_________________________________________
Date
signature ATMOS service
MedizinTechnik
Service Report
Maintenance
contract:
Basic
Premium
None
Remarks
Functional test
passed
not passed
_________________________
passed
not passed
_________________________
Replaced:
Ear irrigation bowl
REF 505.0353.0
Yes No
_________________________
Drainage hose
REF 510.0381.0
Yes No
_________________________
Tecalan hose
_________________________
_________________________
REF 502.0844.0
Yes No
_________________________
REF 530.1180.0
Yes No
_________________________
Yes No
_________________________
Double-barrelled hose
REF 530.1181.0
Yes No
_________________________
Sound absorber
_________________________
Back-up battery
REF 011.1274.0
Yes No
_________________________
Suction hose
REF 530.1144.0
Yes No
_________________________
Sound absorber
REF 069.0131.0
Yes No
_________________________
REF 530.1110.0
Yes No
_________________________
Consumables replenished:
Cleaning agent
Quantity
REF 080.0005.0
Yes No
_________________________
Yes No
_________________________
Halogen bulb 15 V
REF 506.5227.0
Yes No
_________________________
Halogen bulb 24 V
REF 507.2083.0
Yes No
_________________________
Waste bags
REF 505.0515.0
Yes No
_________________________
Bacterial filter
REF 340.0054.0
Yes No _________________________
_____________________________
_________________________________________
Date
signature ATMOS service
MedizinTechnik
Service Report
Maintenance
contract:
Basic
Premium
None
Remarks
Functional test
passed
not passed
_________________________
passed
not passed
_________________________
Replaced:
Ear irrigation bowl
REF 505.0353.0
Yes No
_________________________
Drainage hose
REF 510.0381.0
Yes No
_________________________
Tecalan hose
_________________________
_________________________
REF 502.0844.0
Yes No
_________________________
REF 530.1180.0
Yes No
_________________________
Yes No
_________________________
Double-barrelled hose
REF 530.1181.0
Yes No
_________________________
Sound absorber
_________________________
Back-up battery
REF 011.1274.0
Yes No
_________________________
Suction hose
REF 530.1144.0
Yes No
_________________________
Sound absorber
REF 069.0131.0
Yes No
_________________________
REF 530.1110.0
Yes No
_________________________
Consumables replenished:
Cleaning agent
Quantity
REF 080.0005.0
Yes No
_________________________
Yes No
_________________________
Halogen bulb 15 V
REF 506.5227.0
Yes No
_________________________
Halogen bulb 24 V
REF 507.2083.0
Yes No
_________________________
Waste bags
REF 505.0515.0
Yes No
_________________________
Bacterial filter
REF 340.0054.0
Yes No _________________________
_____________________________
_________________________________________
Date
signature ATMOS service
MedizinTechnik
Service Report
Maintenance
contract:
Basic
Premium
None
Remarks
Functional test
passed
not passed
_________________________
passed
not passed
_________________________
Replaced:
Ear irrigation bowl
REF 505.0353.0
Yes No
_________________________
Drainage hose
REF 510.0381.0
Yes No
_________________________
Tecalan hose
_________________________
_________________________
REF 502.0844.0
Yes No
_________________________
REF 530.1180.0
Yes No
_________________________
Yes No
_________________________
Double-barrelled hose
REF 530.1181.0
Yes No
_________________________
Sound absorber
_________________________
Back-up battery
REF 011.1274.0
Yes No
_________________________
Suction hose
REF 530.1144.0
Yes No
_________________________
Sound absorber
REF 069.0131.0
Yes No
_________________________
REF 530.1110.0
Yes No
_________________________
Consumables replenished:
Cleaning agent
Quantity
REF 080.0005.0
Yes No
_________________________
Yes No
_________________________
Halogen bulb 15 V
REF 506.5227.0
Yes No
_________________________
Halogen bulb 24 V
REF 507.2083.0
Yes No
_________________________
Waste bags
REF 505.0515.0
Yes No
_________________________
Bacterial filter
REF 340.0054.0
Yes No _________________________
_____________________________
_________________________________________
Date
signature ATMOS service
MedizinTechnik
Service Report
Maintenance
contract:
Basic
Premium
None
Remarks
Functional test
passed
not passed
_________________________
passed
not passed
_________________________
Replaced:
Ear irrigation bowl
REF 505.0353.0
Yes No
_________________________
Drainage hose
REF 510.0381.0
Yes No
_________________________
Tecalan hose
_________________________
_________________________
REF 502.0844.0
Yes No
_________________________
REF 530.1180.0
Yes No
_________________________
Yes No
_________________________
Double-barrelled hose
REF 530.1181.0
Yes No
_________________________
Sound absorber
_________________________
Back-up battery
REF 011.1274.0
Yes No
_________________________
Suction hose
REF 530.1144.0
Yes No
_________________________
Sound absorber
REF 069.0131.0
Yes No
_________________________
REF 530.1110.0
Yes No
_________________________
Consumables replenished:
Cleaning agent
Quantity
REF 080.0005.0
Yes No
_________________________
Yes No
_________________________
Halogen bulb 15 V
REF 506.5227.0
Yes No
_________________________
Halogen bulb 24 V
REF 507.2083.0
Yes No
_________________________
Waste bags
REF 505.0515.0
Yes No
_________________________
Bacterial filter
REF 340.0054.0
Yes No _________________________
_____________________________
_________________________________________
Date
signature ATMOS service
MedizinTechnik
Service Report
Maintenance
contract:
Basic
Premium
None
Remarks
Functional test
passed
not passed
_________________________
passed
not passed
_________________________
Replaced:
Ear irrigation bowl
REF 505.0353.0
Yes No
_________________________
Drainage hose
REF 510.0381.0
Yes No
_________________________
Tecalan hose
_________________________
_________________________
REF 502.0844.0
Yes No
_________________________
REF 530.1180.0
Yes No
_________________________
Yes No
_________________________
Double-barrelled hose
REF 530.1181.0
Yes No
_________________________
Sound absorber
_________________________
Back-up battery
REF 011.1274.0
Yes No
_________________________
Suction hose
REF 530.1144.0
Yes No
_________________________
Sound absorber
REF 069.0131.0
Yes No
_________________________
REF 530.1110.0
Yes No
_________________________
Consumables replenished:
Cleaning agent
Quantity
REF 080.0005.0
Yes No
_________________________
Yes No
_________________________
Halogen bulb 15 V
REF 506.5227.0
Yes No
_________________________
Halogen bulb 24 V
REF 507.2083.0
Yes No
_________________________
Waste bags
REF 505.0515.0
Yes No
_________________________
Bacterial filter
REF 340.0054.0
Yes No _________________________
_____________________________
_________________________________________
Date
signature ATMOS service
MedizinTechnik
MedizinTechnik
www.atmosmed.de