Service booklet
ATMOS S 61 Servant

ATMOS Service booklet

Introduction, care
and safety instructions

You have bought an innovative medical product with the latest state-of-the-art
technology.
You should have untroubled pleasure working with your treatment system
ATMOS S 61 Servant for a long time.
Preconditions for this are safety and reliability.
In order to achieve this, regular maintenance and care is necessary.
Protect your investment:
Care and safety controls, together with the professional application,
guarantee the operational reliability and availability for use of the ATMOS
S 61 Servant. This, as well as regular cleaning is essential.
Repair work and safety controls may only be performed by an authorized
ATMOS technician. With the use of the original spare parts you will be
guaranteed that the operational reliability and availability for use as well as
the value of your ATMOS S 61 Servant is maintained.
Please pay attention to the following:
n



n

A safety-related control should be performed annually according to the


accident prevention regulation. In addition we recommend the annual
service, as per service booklet.

Only place clean instruments on the deposit.

Clean and disinfect the instrument deposit at regular intervals!

Only use the approved disinfectants stated in the operating


instructions.

The service technician should check the display (compressed air and
vacuum indicator) annually.

Please note the ATMOS hygiene programme.


For further information please call:
Phone +49 7653 689-222

ATMOS Service booklet

Test protocol and


test instructions

Please note: Please keep this service booklet at hand in case of a warranty
claim. You should provide us with the corresponding data in case of a warranty
claim.

Type plate
workstation

Type plate
vision

Type plate
instruments

Test protocol and


test instructions

Type plate
others

Type plate
others

Verantwortlicher /
responsible
1. Operator self check; Annex A
2. Final inspection;
Annex B
3. Order picking according
to consignment list

ok

Unterschrift / signature

FMS

FMS

FV

Test protocol and


test instructions

5.1 Packaging undamaged

5.2 Correct Voltage of the unit


Crosscheck of data plate to actual voltage
100 Vac [ ]

115 Vac [ ]

127 Vac [ ]

230 Vac [ ]

5.3 Lieferung gem Lieferschein vollstndig

5.4 Miscellaneous, remarks:

6. Service
Checkup performed
completely

7. Customer
- delivery obtained according to order

bergabe
Delivery

to be filled in by customer service

bergabe bestanden
Delivery passed

4. Installation of the unit at customers side

ATMOS Service booklet

Service indicator on the


ATMOS S 61 Servant
workstation
Your ATMOS S 61 Servant is equipped with a service indicator! It helps you to identify when a
service is recommended
The indicator has a red and a
yellow LED

red LED
yellow LED

The yellow LED shows several safety circuits (see operating instructions).
The red LED signalizes the service according to the index.
Once the device is switched on the status is indicated on the display!

first light up
second light up
third light up
fourth light up
fifth light up
sixth light up
seventh light up

yellow LED Beep

see operating
instructions

red LED
Possible service case:
Runtime pressure pump
Runtime suction pump
Runtime suction pump 2
Runtime hose pump
Runtime drain pump
Annual maintenance
Battery voltage

yes
yes
yes
yes
yes
yes
yes

Remark
runtime critical
runtime critical
runtime critical
runtime critical
runtime critical
annual maintenance
low battery voltage

The monitoring of the burning life of the halogen bulb on the ATMOS S 61 Servant vision:
The burning life of the halogen bulb is monitored by the control. If 90% of the bulbs burning life is
achieved (nominally indicated by the manufacturer) the LED on the activation button is no longer
green but turns orange.
If one of the above service cases occurs, please contact your local ATMOS service partner or contact
the ATMOS service centre, phone +49 7653 689-222.

Confirmation and
reference to
warranty
Confirmation
The ATMOS service technician performs all the services on schedule and confirms the proceedings
in this booklet.
Your ATMOS S 61 Servant receives a seal of approval stating the date for the next service
appointment.
The scope of the service does not include any repair work, for this a separate order must be given.

Reference to warranty
Please follow the tips in this service booklet and comply with the same. This is the only way to ensure
your warranty claim is maintained.
During the warranty if the prescribed services are not performed by an authorized ATMOS service
technician within the specified time, then a warranty claim may only be settled when the
manufacturers inquiry results are presented.

We wish you a lot of pleasure with your new treatment system!


ATMOS MedizinTechnik

ATMOS Service booklet

Service Report
Maintenance
contract:

Basic

Premium

None

Service report No. ___________________________


Remarks

Functional test

passed

not passed

_________________________

Safety check (DIN EN 62353)

passed

not passed

_________________________

Replaced:
Ear irrigation bowl

REF 505.0353.0

Yes No

_________________________

Drainage hose

REF 510.0381.0

Yes No

_________________________

Tecalan hose

REF 005.0025.0 Yes No

_________________________

Irrigation adapter hose irrigation REF 506.2228.1 Yes No

_________________________

Splash protection/hose tip

REF 502.0844.0

Yes No

_________________________

Irrigation handle and hose

REF 530.1180.0

Yes No

_________________________

Irrigation handle and hose REF 530.1190.0

Yes No

_________________________

Double-barrelled hose

REF 530.1181.0

Yes No

_________________________

Sound absorber

REF 069.0027.0 Yes No

_________________________

Back-up battery

REF 011.1274.0

Yes No

_________________________

Suction hose

REF 530.1144.0

Yes No

_________________________

Sound absorber

REF 069.0131.0

Yes No

_________________________

Suction canister lid

REF 530.1110.0

Yes No

_________________________

Consumables replenished:
Cleaning agent

Quantity

REF 080.0005.0

Yes No

_________________________

Irrigation adapter hose irrigation REF 506.2228.1

Yes No

_________________________

Halogen bulb 15 V

REF 506.5227.0

Yes No

_________________________

Halogen bulb 24 V

REF 507.2083.0

Yes No

_________________________

Waste bags

REF 505.0515.0

Yes No

_________________________

Bacterial filter

REF 340.0054.0

Yes No _________________________

_____________________________
_________________________________________
Date
signature ATMOS service
MedizinTechnik

Service Report
Maintenance
contract:

Basic

Premium

None

Service report No. ___________________________


Remarks

Functional test

passed

not passed

_________________________

Safety check (DIN EN 62353)

passed

not passed

_________________________

Replaced:
Ear irrigation bowl

REF 505.0353.0

Yes No

_________________________

Drainage hose

REF 510.0381.0

Yes No

_________________________

Tecalan hose

REF 005.0025.0 Yes No

_________________________

Irrigation adapter hose irrigation REF 506.2228.1 Yes No

_________________________

Splash protection/hose tip

REF 502.0844.0

Yes No

_________________________

Irrigation handle and hose

REF 530.1180.0

Yes No

_________________________

Irrigation handle and hose REF 530.1190.0

Yes No

_________________________

Double-barrelled hose

REF 530.1181.0

Yes No

_________________________

Sound absorber

REF 069.0027.0 Yes No

_________________________

Back-up battery

REF 011.1274.0

Yes No

_________________________

Suction hose

REF 530.1144.0

Yes No

_________________________

Sound absorber

REF 069.0131.0

Yes No

_________________________

Suction canister lid

REF 530.1110.0

Yes No

_________________________

Consumables replenished:
Cleaning agent

Quantity

REF 080.0005.0

Yes No

_________________________

Irrigation adapter hose irrigation REF 506.2228.1

Yes No

_________________________

Halogen bulb 15 V

REF 506.5227.0

Yes No

_________________________

Halogen bulb 24 V

REF 507.2083.0

Yes No

_________________________

Waste bags

REF 505.0515.0

Yes No

_________________________

Bacterial filter

REF 340.0054.0

Yes No _________________________

_____________________________
_________________________________________
Date
signature ATMOS service
MedizinTechnik

Service Report
Maintenance
contract:

Basic

Premium

None

Service report No. ___________________________


Remarks

Functional test

passed

not passed

_________________________

Safety check (DIN EN 62353)

passed

not passed

_________________________

Replaced:
Ear irrigation bowl

REF 505.0353.0

Yes No

_________________________

Drainage hose

REF 510.0381.0

Yes No

_________________________

Tecalan hose

REF 005.0025.0 Yes No

_________________________

Irrigation adapter hose irrigation REF 506.2228.1 Yes No

_________________________

Splash protection/hose tip

REF 502.0844.0

Yes No

_________________________

Irrigation handle and hose

REF 530.1180.0

Yes No

_________________________

Irrigation handle and hose REF 530.1190.0

Yes No

_________________________

Double-barrelled hose

REF 530.1181.0

Yes No

_________________________

Sound absorber

REF 069.0027.0 Yes No

_________________________

Back-up battery

REF 011.1274.0

Yes No

_________________________

Suction hose

REF 530.1144.0

Yes No

_________________________

Sound absorber

REF 069.0131.0

Yes No

_________________________

Suction canister lid

REF 530.1110.0

Yes No

_________________________

Consumables replenished:
Cleaning agent

Quantity

REF 080.0005.0

Yes No

_________________________

Irrigation adapter hose irrigation REF 506.2228.1

Yes No

_________________________

Halogen bulb 15 V

REF 506.5227.0

Yes No

_________________________

Halogen bulb 24 V

REF 507.2083.0

Yes No

_________________________

Waste bags

REF 505.0515.0

Yes No

_________________________

Bacterial filter

REF 340.0054.0

Yes No _________________________

_____________________________
_________________________________________
Date
signature ATMOS service
MedizinTechnik

Service Report
Maintenance
contract:

Basic

Premium

None

Service report No. ___________________________


Remarks

Functional test

passed

not passed

_________________________

Safety check (DIN EN 62353)

passed

not passed

_________________________

Replaced:
Ear irrigation bowl

REF 505.0353.0

Yes No

_________________________

Drainage hose

REF 510.0381.0

Yes No

_________________________

Tecalan hose

REF 005.0025.0 Yes No

_________________________

Irrigation adapter hose irrigation REF 506.2228.1 Yes No

_________________________

Splash protection/hose tip

REF 502.0844.0

Yes No

_________________________

Irrigation handle and hose

REF 530.1180.0

Yes No

_________________________

Irrigation handle and hose REF 530.1190.0

Yes No

_________________________

Double-barrelled hose

REF 530.1181.0

Yes No

_________________________

Sound absorber

REF 069.0027.0 Yes No

_________________________

Back-up battery

REF 011.1274.0

Yes No

_________________________

Suction hose

REF 530.1144.0

Yes No

_________________________

Sound absorber

REF 069.0131.0

Yes No

_________________________

Suction canister lid

REF 530.1110.0

Yes No

_________________________

Consumables replenished:
Cleaning agent

Quantity

REF 080.0005.0

Yes No

_________________________

Irrigation adapter hose irrigation REF 506.2228.1

Yes No

_________________________

Halogen bulb 15 V

REF 506.5227.0

Yes No

_________________________

Halogen bulb 24 V

REF 507.2083.0

Yes No

_________________________

Waste bags

REF 505.0515.0

Yes No

_________________________

Bacterial filter

REF 340.0054.0

Yes No _________________________

_____________________________
_________________________________________
Date
signature ATMOS service
MedizinTechnik

Service Report
Maintenance
contract:

Basic

Premium

None

Service report No. ___________________________


Remarks

Functional test

passed

not passed

_________________________

Safety check (DIN EN 62353)

passed

not passed

_________________________

Replaced:
Ear irrigation bowl

REF 505.0353.0

Yes No

_________________________

Drainage hose

REF 510.0381.0

Yes No

_________________________

Tecalan hose

REF 005.0025.0 Yes No

_________________________

Irrigation adapter hose irrigation REF 506.2228.1 Yes No

_________________________

Splash protection/hose tip

REF 502.0844.0

Yes No

_________________________

Irrigation handle and hose

REF 530.1180.0

Yes No

_________________________

Irrigation handle and hose REF 530.1190.0

Yes No

_________________________

Double-barrelled hose

REF 530.1181.0

Yes No

_________________________

Sound absorber

REF 069.0027.0 Yes No

_________________________

Back-up battery

REF 011.1274.0

Yes No

_________________________

Suction hose

REF 530.1144.0

Yes No

_________________________

Sound absorber

REF 069.0131.0

Yes No

_________________________

Suction canister lid

REF 530.1110.0

Yes No

_________________________

Consumables replenished:
Cleaning agent

Quantity

REF 080.0005.0

Yes No

_________________________

Irrigation adapter hose irrigation REF 506.2228.1

Yes No

_________________________

Halogen bulb 15 V

REF 506.5227.0

Yes No

_________________________

Halogen bulb 24 V

REF 507.2083.0

Yes No

_________________________

Waste bags

REF 505.0515.0

Yes No

_________________________

Bacterial filter

REF 340.0054.0

Yes No _________________________

_____________________________
_________________________________________
Date
signature ATMOS service
MedizinTechnik

Service Report
Maintenance
contract:

Basic

Premium

None

Service report No. ___________________________


Remarks

Functional test

passed

not passed

_________________________

Safety check (DIN EN 62353)

passed

not passed

_________________________

Replaced:
Ear irrigation bowl

REF 505.0353.0

Yes No

_________________________

Drainage hose

REF 510.0381.0

Yes No

_________________________

Tecalan hose

REF 005.0025.0 Yes No

_________________________

Irrigation adapter hose irrigation REF 506.2228.1 Yes No

_________________________

Splash protection/hose tip

REF 502.0844.0

Yes No

_________________________

Irrigation handle and hose

REF 530.1180.0

Yes No

_________________________

Irrigation handle and hose REF 530.1190.0

Yes No

_________________________

Double-barrelled hose

REF 530.1181.0

Yes No

_________________________

Sound absorber

REF 069.0027.0 Yes No

_________________________

Back-up battery

REF 011.1274.0

Yes No

_________________________

Suction hose

REF 530.1144.0

Yes No

_________________________

Sound absorber

REF 069.0131.0

Yes No

_________________________

Suction canister lid

REF 530.1110.0

Yes No

_________________________

Consumables replenished:
Cleaning agent

Quantity

REF 080.0005.0

Yes No

_________________________

Irrigation adapter hose irrigation REF 506.2228.1

Yes No

_________________________

Halogen bulb 15 V

REF 506.5227.0

Yes No

_________________________

Halogen bulb 24 V

REF 507.2083.0

Yes No

_________________________

Waste bags

REF 505.0515.0

Yes No

_________________________

Bacterial filter

REF 340.0054.0

Yes No _________________________

_____________________________
_________________________________________
Date
signature ATMOS service
MedizinTechnik

ATMOS Service booklet

Service Report
Maintenance
contract:

Basic

Premium

None

Service report No. ___________________________


Remarks

Functional test

passed

not passed

_________________________

Safety check (DIN EN 62353)

passed

not passed

_________________________

Replaced:
Ear irrigation bowl

REF 505.0353.0

Yes No

_________________________

Drainage hose

REF 510.0381.0

Yes No

_________________________

Tecalan hose

REF 005.0025.0 Yes No

_________________________

Irrigation adapter hose irrigation REF 506.2228.1 Yes No

_________________________

Splash protection/hose tip

REF 502.0844.0

Yes No

_________________________

Irrigation handle and hose

REF 530.1180.0

Yes No

_________________________

Irrigation handle and hose REF 530.1190.0

Yes No

_________________________

Double-barrelled hose

REF 530.1181.0

Yes No

_________________________

Sound absorber

REF 069.0027.0 Yes No

_________________________

Back-up battery

REF 011.1274.0

Yes No

_________________________

Suction hose

REF 530.1144.0

Yes No

_________________________

Sound absorber

REF 069.0131.0

Yes No

_________________________

Suction canister lid

REF 530.1110.0

Yes No

_________________________

Consumables replenished:
Cleaning agent

Quantity

REF 080.0005.0

Yes No

_________________________

Irrigation adapter hose irrigation REF 506.2228.1

Yes No

_________________________

Halogen bulb 15 V

REF 506.5227.0

Yes No

_________________________

Halogen bulb 24 V

REF 507.2083.0

Yes No

_________________________

Waste bags

REF 505.0515.0

Yes No

_________________________

Bacterial filter

REF 340.0054.0

Yes No _________________________

_____________________________
_________________________________________
Date
signature ATMOS service
MedizinTechnik

Service Report
Maintenance
contract:

Basic

Premium

None

Service report No. ___________________________


Remarks

Functional test

passed

not passed

_________________________

Safety check (DIN EN 62353)

passed

not passed

_________________________

Replaced:
Ear irrigation bowl

REF 505.0353.0

Yes No

_________________________

Drainage hose

REF 510.0381.0

Yes No

_________________________

Tecalan hose

REF 005.0025.0 Yes No

_________________________

Irrigation adapter hose irrigation REF 506.2228.1 Yes No

_________________________

Splash protection/hose tip

REF 502.0844.0

Yes No

_________________________

Irrigation handle and hose

REF 530.1180.0

Yes No

_________________________

Irrigation handle and hose REF 530.1190.0

Yes No

_________________________

Double-barrelled hose

REF 530.1181.0

Yes No

_________________________

Sound absorber

REF 069.0027.0 Yes No

_________________________

Back-up battery

REF 011.1274.0

Yes No

_________________________

Suction hose

REF 530.1144.0

Yes No

_________________________

Sound absorber

REF 069.0131.0

Yes No

_________________________

Suction canister lid

REF 530.1110.0

Yes No

_________________________

Consumables replenished:
Cleaning agent

Quantity

REF 080.0005.0

Yes No

_________________________

Irrigation adapter hose irrigation REF 506.2228.1

Yes No

_________________________

Halogen bulb 15 V

REF 506.5227.0

Yes No

_________________________

Halogen bulb 24 V

REF 507.2083.0

Yes No

_________________________

Waste bags

REF 505.0515.0

Yes No

_________________________

Bacterial filter

REF 340.0054.0

Yes No _________________________

_____________________________
_________________________________________
Date
signature ATMOS service