Professional Documents
Culture Documents
COLUMN FLOTATION
GLENN DOBBY SGS
ABSTRACT
Industrial application of column flotation began in earnest in the early 1980s, and is now accepted as a
conventional application. For example, most of the worlds major copper producers use column flotation as the
final stage of cleaning. In addition, the technology is common for final cleaning of zinc, lead and molybdenum
sulfides, and for processing of iron ores and phosphates. This article includes a description of the key features
and concepts of column flotation, and then reviews recent industrial applications of column flotation, covering
pilot testing and scale-up, circuit design of selected applications, and instrumentation and control.
INTRODUCTION Today column flotation has become KEY FEATURES AND CONCEPTS
an accepted means of froth flotation
The invention of the flotation column for a fairly broad range of applications, A schematic of a flotation column is
by Pierre Boutin and Remi Tremblay in particular the cleaning of sulfides shown in Figure 1.
(patented in 1962) came about from (copper, zinc, lead and molybdenite) and
earlier work conducted on solvent the flotation of iron ore, phosphate and Industrial flotation columns are 6-14 m
extraction at the research laboratory of coal. in height (from the bottom discharge
Eldorado Mining, in which they applied to the top lip), and range in diameter
the column as it is known today to Flotation columns differ dramatically from 0.5 to 5 m. Recent large scale
solvent-in-pulp processing of uranium from mechanical flotation machines in applications have typically been 4 to 4.5
ore. Slurried ore was fed near the top several ways: m diameter and approximately 12 m tall.
of a laboratory column and solvent [There are many installations of columns
droplets were generated at the bottom there is no mechanical agitation/shear that are rectangular. The fabrication cost
of the column. The density difference the cell is relatively tall and narrow for rectangular columns is significantly
caused a counter- current flow of solvent gas bubbles are generated by sparging higher than for circular columns, as
droplets and slurry, and a solvent phase froths typically are deeper, and wash approximately twice the mass of steel is
was created at the top of the column, water typically is liberally applied to the required; however, there is no evidence
with a distinct solvent-pulp interface. surface of the froth. that rectangular columns will provide
An aqueous diluent was introduced better metallurgy than from circular
near the interface in order to minimize There has been a wealth of research columns.]
contamination of the solvent with ore and fundamental studies on aspects
particles. Slurry was drawn out the of column flotation during the past 15 As with mechanical cells, two distinct
bottom of the column and solvent years or so (some of which has been zones are evident: the collection zone
overflowed the top lip. (Boutin 2002) described by Finch and Dobby, 1990 and (extending from the spargers to the
by Finch, Uribe-Salas and Xu, 1994.). froth:pulp interface) and the froth zone.
Boutin and Tremblay then applied the However, the fundamental aspects is Interaction between these two zones,
concept to ore flotation, replacing the not the focus here; rather, this article will especially the recycle of collected
solvent droplets with air bubbles and review recent industrial applications of particles from the froth to the pulp, is
replacing the diluent with water. The first column flotation, covering pilot testing particularly important in understanding
successful applications were on amine and scale-up, circuit design of selected design and operation of a flotation
flotation of silica from iron ore samples. applications, and instrumentation and column; this is addressed in the Scale-up
The first notable industrial success control. We begin with a description of section.
of flotation columns was by Column the key features and concepts of column
Flotation Company of Canada on moly flotation.
cleaning at Norandas Les Mines Gasp
(Cienski and Coffin, 1981; Coffin and
Miszczak, 1982).
SGS MINERALS SERVICES TECHNICAL BULLETIN 2002-23 2
Bubble generation in flotation columns PILOT TESTING AND SCALE-UP A key factor in column design is the
was originally achieved via sparging froth recovery, which directly controls
through pierced rubber or through woven Metallurgical testing of column flotation the extent of re- circulation within the
fabric. That approach is no longer very has generally been through continuous column. As a general observation,
common, as the column must be shut pilot plant operation. For retrofitting at the froth recovery of pilot columns
down and drained to assess or replace existing plants, the approach has been is considerably higher than in large
sparger units. Today there are two to conduct pilot testing on site, feeding industrial columns, due to the stability
principle bubble generation technologies. a bleed stream from the plant circuit to imparted to the froth by the walls in a
the pilot plant circuit (Kosick et. al. 1991) pilot column.
The most common method of bubble Typical pilot columns are 10 cm diameter
generation is jet sparging of air through by 6 m tall, although pilot testing has Therefore, in sizing large columns from
orifices. Air is forced under high been conducted with columns as small pilot plant data the designed retention
pressure (30-100 psig) through annular as 5 cm and as large as 30 cm. There time will be longer for two reasons:
or circular orifices, creating a jet of air have been some attempts to conduct
through the slurry. The high shear that pilot testing and scale-up though a) short-circuiting Since the large column
exists at the gas-slurry interface results operation of a batch laboratory column has a fluid flow pattern that is closer to
in generation of gas bubbles. Bubble (Flint, 2002). a stirred tank than a plug flow vessel
diameter is affected by the jet velocity; (opposite to that of a pilot column),
the higher the jet velocity the smaller Since the mechanism of particle a longer retention time is required
the resulting bubbles. The spargers are capture differs significantly between a to achieve the same collection zone
fit into the column through a nipple mechanical cell and a flotation column, recovery as in the pilot column
and valve assembly, extending only kinetic data obtained from bench b) lower froth recovery Since the
several centimeters into the pulp. With mechanical flotation machines generally plant column will have a lower froth
this configuration, the spargers can be is not easily applied to column sizing. recovery (i.e. higher internal circulating
removed and inserted while the column This sometimes poses a problem load) it is necessary to have a higher
is full of slurry and under operation. A for design and sizing of columns for collection zone recovery than the pilot
large column will use 15-20 spargers. cleaning on green-fields design. The column in order to attain the same
quantity of rougher feed required to overall column recovery as with the
The second principle method of bubble generate sufficient rougher concentrate pilot column.
generation employed today is through for column pilot testing is generally too
pumping of a portion of the underflow large to justify a pilot plant. However, Three issues deserve further elaboration:
through static mixers back into the pulp, the selectivity that is obtained from the carrying capacity, required bias flow and
employed by the Microcel technology bench testing can be used for helping column height.
(Yoon and Luttrell, 1994). Air is injected design the number of stages of cleaning.
at the static mixer, and the resultant high New tools have been developed to Carrying Capacity Operation of full
shear through the mixer generates very assist in the design of circuits using scale columns has shown that carrying
small gas bubbles; hence, the bubble flotation columns (Dobby et al. 2002), capacity (CM) for a large column can be
generation is external to the column. A and the design of sulfide cleaner circuits substantially lower than that of a pilot
typical column will have several static for greenfield plants is entirely feasible column. For example, Espinosa and
mixers fed by a common pump and without the need for pilot testing (aside Johnson (1991) reported a CM for 2 m
header. from some uncommonly complex ores). and 2.5 m diameter plant columns on
lead and zinc cleaning that was about
An advantage of the static mixer Scale-up of flotation columns has 50% of the CM for a 5 cm pilot column.
approach is that it can produce smaller been well documented (Finch and However, the trend of higher carrying
gas bubbles. However, this is at the Dobby, 1990). The general approach capacity for higher k80 is still valid for
expense of pumping a very high slurry is to apply a first-order kinetic model plant columns.
flow through the mixers. The Microcel to particle collection, and select froth
approach has been applied primarily to zone recoveries appropriate for the An important and often neglected issue
coal flotation, with a large and successful size of the columns and the froth when quoting CM, whether for a pilot
installation at BHP Coals Peak Downs mobility. Particle retention time may be column or a full scale column, is the
coal preparation plant (Brake, 1998). significantly different from that of the value for the operating bias. Figure 2 is
water, especially for coarse particles, an example of measured concentrate
Whatever method of air addition so the retention time of particles needs solids flux Ca for a copper cleaning pilot
employed, the pressure drop of a column to be calculated and applied. Hydraulic column, plotted as a function of bias.
will require that the air is supplied via a entrainment of fine particles is accounted In this case feed grade to the column
compressor (not a blower). Selection of for by quantifying the recovery of water is high, so Ca is reasonably close to CM,
the correct compressor capacity clearly to the froth; for columns operating and CM would be expected to follow the
is important. with a bias close to zero, the effect of same trend as in Figure 2. A full scale
entrainment is usually minor. column circuit was installed subsequent
SGS MINERALS SERVICES TECHNICAL BULLETIN 2002-23 4
Figure 7 Zinc flotation circuit at the BHP-Billitons Les Mines Selbaie (Wright, 1995).
SGS MINERALS SERVICES TECHNICAL BULLETIN 2002-23 7
Table 1 Example of column and plant performance in zinc cleaning, showing the difference between two The solution to the problem described
stage and three stage column circuits.
above is to use several stages of
Circuit Feed Rate (L/ Feed %Zn Conc %Zn Zn Recovery flotation. A good example of this is from
min) CVRDs Samitri circuit in Brazil, Figure
Plant mechanical 42.6 55.5 57 8. All of the four columns are 4.6 m
cleaners diameter. The first (rougher) rougher and
the cleaner-scavenger column are both
Two Columns 3.3 41.4 57.2 82
water washed and only 8 m tall, while
Three Columns 3.3 38.2 59.1 84 the cleaner and re-cleaner columns are
not water washed and are 15 m tall.
Froth from the two tall columns feeds
MOLYBDENITE CLEANING the target level of silica in the
the scavenger column by gravity flow.
As mentioned in the introduction, the underflow is usually <1%, so when the
The feed grade to this circuit is typically
first notable industrial application of feed grade is high (15-25 %silica) the
20 % silica or higher. Most of the silica is
column flotation was on molybdenite recovery of silica to the froth product
removed in the front end column circuit,
cleaning at Les Mines Gasp, where must be higher than 90 %, at which
and the mechanical secondary circuit
three columns replaced approximately point short-circuiting of slurry in the
ensures removal of coarse silica.
10 stages of mechanical cells. collection zone plays a dramatic role;
Since then, column flotation has and
PHOSPHATE FLOTATION
become the standard approach for as discussed earlier, the plant column
Flotation columns have been applied
molybdenite cleaning circuits. Typically will provide a considerably lower
successfully for phosphate flotation in
threestsequential column stages conduct degree of stability to the froth, which
both Brazil and Florida. A good example
the final cleaning, treating either rougher means that there will much higher
is the Serrana concentrator (Brazil),
concentrate or 1 cleaner concentrate dropback from the froth to the pulp
where six columns are in operation
generated in mechanical cells. An Hence, the increased short-circuiting
(Guimares and Peres 2000, Guimares
approach to determining the number that occurs in the industrial column
et al. 1999). Each (rectangular) column
of stages required has been described and the significantly higher degree of
is 3 m by 4.5 m by 14.5 m tall. Th3e
by Amelunxen (1990). At the Endako internal circulation combine to prevent
columns were all installed as retrofits to
concentrator in British Columbia, Canada successful separation in one (or even
the existing mechanical cells; ultimately
five columns are used, two at 2.7 m two) column stages. In addition, the
66 8.4 m cells were replaced by the
diameter, two at 1.5 m diameter and one lower froth recovery in the industrial
six columns. The flowsheet for the
at 1.1 m diameter, all 10 m tall. column is usually magnified for coarse
plant is shown in Figure 9, and typical
silica, often resulting in difficulties with
operating conditions for each column are
An additional advantage of using removal of coarse silica.
summarized in Table 2.
columns for molybdenum cleaning
from bulk copper-moly concentrate is
that the columns are well suited to the
use of nitrogen in place of air (which
is sometimes practiced in order to
significantly reduce the consumption of
sodium hydrosulfide, used for copper
sulfide depression).
Table 2 Typical operating conditions for the phosphate flotation columns at Serrana (Guimares et al. 1999)
INSTRUMENTATION AND
CONTROL
ITEM UNITS APATITE: APATITE: APATITE: APATITE: BARITE;
NATURAL FINES COARSE REGROUND COARSE
The degree of instrumention on an
FINES FROM
GRINDING industrial flotation column can range
from very basic to reasonablely complex.
Air cm/s 0.9 0.9 0.97 0.82 0.94
The minimum control requirement is
Wash water cm/s 0.19 0.19 0.19 0.23 0.13 that for pulp level. However, a typical
Bias cm/s 0.08 0.05 0.12 0.17 0.02 industrial column will have automatic
Residence min 39 61 50 33 24 control of level, air flow rate and wash
Time water flow.
Carrying t/h/m2 1.5 1.7 2.7 1.2 1.4
Rate The low ratio of surface area to volume
Lip loading kg/h/m 550 630 1000 440 510 inherent with a column means that the
Froth Depth cm/s 70 70 60 60 25 pulp level (which directly affects froth
depth of course) will change quickly
and significantly with changes in feed
flow rate. Hence, this puts emphasis
on good tuning of level controllers, but
more importantly on ensuring that a
reasonably stable feed flow is provided
to the column. Feed to a column may be
directly from a pump; however, multiple
columns in parallel are usually fed via a
slurry distributor.
REFERENCES Dobby, G., Kosick, G. and Amelunxen, Guimaraes, R.C., Takata, L.A., Peres,
R. 2002. A focus on variability within the A.E.C. and Wyslouzil, H. 1999. Column
Amelunxen, R.L. 1990. Moly upgrading orebody for improved design of flotation flotation applied to the production
profiles with flotation columns, CMP plants, CMP Proc. 34th Annual Meeting, of phosphate rock in Arax, Brazil,
Proc. 22nd Annual Meeting, p.258 p.77 Proc. 31st Annual Canadian Mineral
Processors Operators Conf.
Boutin, P. 2002. Private communication Espinosa-Gomez, R. and Johnson,
to the author N.W. 1991. Technical experiences with Kosick, G., Dobby, G. and Shimizu, P.
conventional columns at Mount Isa 1991. The use of column flotation pilot
Brake, I.R. 1998. The development Mines Limited, Proc. Column 91 (Agar, plants for column flotation circuit design,
and commissioning of a new Microcel Huls and Hyma, eds.), Sudbury, CIM Proc. Column 91, Canadian Inst. Min. &
column flotation circuit for BHP Coals Metall., Vol 2, p.359
Peak Downs coal preparation plant, XIIIth Flint, I. 2002. Batch flotation modeling,
Coal Preparation Congress, Brisbane, PhD thesis, UBC, Canada Werniuk, J. 2000. Getting down and dirty
Australia at Laronde, Canadian Mining J., August,
Finch, J.A. and Dobby, G.S. 1990. p.13 Wright, A. 1995. Flotation at Les
Chevalier, G. and Dobby, G. 1996. Column Flotation, Pergamon Press Mines Selbaie, CMP Proc. 27th Annual
Flotation column installations for zinc and Meeting, p.120
copper cleaning at Les Mines Selbaie, at Finch, J.A., Uribe-Salas, A. and Xu,
Copper 96, CIM Conf of Metallurgists, M. 1994. Column flotation: a selected Yianatos, J.B., Finch, J.A. and Laplante,
Montreal review Part III, in Flotation - Science and A.R. 1986. Holdup profile and bubble size
Engineering (ed. K.A. Matis), Marcel determination of flotation column froths,
Cienski, T. and Coffin, V.L. 1981. CMP Dekker Canadian Metall. Quarterly, 25(1), p.23
Proc. 13th Annual Meeting, p.240 Yianatos, J.B., Finch, J.A. and Laplante,
Furey, J.T. 1990. Rougher column A.R. 1987. Cleaning action in column
Coffin, V.L. and Miszczak, J. 1982. Proc. flotation of gold tellurides, CMP Proc. flotation froths, Trans. Inst. Min. &
14th Int. Mineral Processing Congress, 22nd Annual Meeting Guimaraes, R.C. Metall., Section C, 96, p.199
Toronto, paper 4- 21 andst Peres, A.E.C. 2000. Industrial
practice of phosphate ore flotation at Yoon, R.H. and Luttrell, G.H. 1994.
Dobby, G. and Kosick, G. 1995. Case Serrana- Arax, Proc. 21 Int. Mineral Microcel column flotation scale-up and
studies on circuit design using flotation Processing Congress, p.B9-17 plant practice, CMP Proc. 26th Annual
columns, CMP Proc. 27th Annual Meeting, paper 12
Meeting, p.171