Benets
Comprehensive visualisation
SmartPilot View uses widely accepted patient models to calculate and visualise pharmacokinetic and
pharmacodynamic information and displays the calculated anaethesia level. At the same time it displays the
most important hemodynamic parameters to support the intuitive and fast compilation of patient status at a
glance.
SmartPilot View features an extensive database of pharmacokinetics and pharmacodynamic patient models.
Thanks to the visualisation of this knowledge base on SmartPilot View, the anaesthesiologist has instant access
to the information.
Process optimisation
SmartPilot View addresses the main challenges of the anaesthesia process by helping you to perform smooth
induction, maintain adequate anaesthesia levels and a well-timed, painless wake-up of the patient. Thanks to
SmartPilot View, already during drug administration, the anaesthesiologist can visualise the calculated current
anaesthesia level and the expected course of anaesthesia at a glance. This can facilitate the optimization of
anaesthesia application and analgesia management. SmartPilot view simplies the estimation of individual drug
dosing.
Designed for Primus and Zeus Anaesthesia Workstations Platforms and Perseus A 500
The SmartPilot View is designed to be combined with Primus, Primus Innity Empowered, Perseus A500,
Zeus and Zeus Innity Empowered and can automatically transfers various data from these anaesthesia
workstations. When used in conjunction with Zeus Innity Empowered, the "what if" function will display the
calculated effect of setting change of anaesthesia level even before the setting is conrmed.
Common features
Primus Innity Empowered: Automatic data transfer of measured ventilation and gas concentration values
Perseus A500: Automatic transfer of Vapor 3000 hand dial settings as well as measured ventilation and
gas concentration values
Drger SmartPilot View | 03
Benets
Zeus Innity Empowered: Prediction based on automatically transferred settings, measured values and
drug congurations; What if function
System Components
Transform your clinical workow with Innity Acute Care System. Its
multiparameter monitor integrates with its networked medical-grade
workstation, giving you real-time vital signs, access to clinical hospital
systems and data management applications for a comprehensive range
of patient information and powerful analysis tools at the point-of-care.
D-19739-2009
Technical Data
SmartPilot View
Software version 3.0n
Classication
Classication in accordance with Directive 93/42/EEC Annex IX Class IIb
General
SmartPilot View displays pharmacokinetic, pharmacodynamic (PK/PD) modelling information for use with total intravenous (TIVA),
balanced anaesthesia and volatile anaesthesia.
Display of haemodynamic measurements and processed EEG (BIS Aspect).
Drug and dosage data entry for intravenous drugs: manually or automatically from the syringe pumps. For inhaled anaesthetics:
automatically from anaesthesia device.
Event markers allow the patients individual status to be documented with regard to the observed anaesthesia level. The markers may
be deleted or repositioned if needed.
Support of USB stick for case recording.
Editable drug database.
PK/PD models
Interaction between volatile and intravenous hypnotics and opioids visualised through two-dimensional graph with 15 minute prediction.
PK models for each single drug visualised in a time-based graph with 40 minute trend and 20 minute prediction.
Manual entry of drug bolus and rates at current time or retrospectively followed by recalculation of the case, incl. prediction.
PK models
Propofol Schnider
Remifentanil Minto
Alfentanil Scott
Fentanyl Scott
Sufentanil Gepts
Volatiles Bailey
Pancuronium Rupp
Rocuronium Wierda
Vecuronium Cronnelly
Mivacurium Laurin
PD models
Sequential model Bouillon
*Supported drugs
Hypnotics:
Volatile drugs Desurane, Isourane, Sevourane
Intravenous drugs Propofol
Opioids (intravenous) Fentanyl, Remifentanil, Alfentanil, Sufentanil
Muscle relaxants* Pancuronium, Rocuronium, Mivacurium, Vecuronium
*The impact of muscle relaxants on the drug interaction is not taken into account.
Technical Data
System compatibility
Anaesthesia devices Zeus, Zeus IE, Perseus A500, Primus, Primus IE
PC-platform Drger C700 for IT with or without Innity Explorer or other
medical-grade PC (system requirements:
W7 32/64bit
1-GHz-CPU (x86x64)
1 GB RAM (at least 500 MB available RAM at startup of SPV)
1 GB available Hard Disk Drive space
Screen resolution: 1,280 x 768
Monitoring Drger Innity Delta, Drger Innity Delta XL, Drger Zeus/
Zeus Innity Empowered with integrated haemodynamic
monitoring, IACS patient monitor, Philips monitor (MP50, MP70,
MX800)
Display of patient measurements and processed EEG (BIS Aspect) provided by the basic device:
Heart rate 1/min
Blood pressure (ART M or NIBP) mmHg or kPa
etCO2 mmHg or kPa
BIS Index and SQI
Infusion pumps
Drger IVenus**
B. Braun Perfusor Space with SpaceCom**
Fresenius Orchestra Modul DPS with Base A
Fresenius Orchestra Modul DPS with Base Primea*
Fresenius Injectomat Agilia with Link Agilia
Alaris GH, PK*, TIVA with Gateway
Terumo Terumo Terufusion TE-332, TE-371, TE-372
Terumo Terumo Terufusion TE-SS800 with TE-RS800
Range of supported devices (e.g. Patient Monitors, Anaesthesia devices and infusion pumps) is expanding: please contact your
Drger representative for the current list of compatible devices.
*TCI pumps are supported
**up to four infusion pumps
Perfusor Space with SpaceCom is the registered trade mark of Company B.Braun.
Orchestra Modul DPS with Base A, Orchestra Modul DPS with Base Primea and Injectomat Agilia with Link Agilia are registered
trade mark of Company Fresenius.
Terumo Terufusion TE-332, TE-371, TE- 372 and Terumo Terufusion TE-SS880 with TE-RS800 are registered trade mark of Company
Terumo.
06 | Drger SmartPilot View
Notes
Drger SmartPilot View | 07
Notes
08 | Drger SmartPilot View
Notes
90 66 500 | 16.03-6 | Communications & Sales Marketing | PP | LE | Chlorine-free environmentally compatible | Subject to modications | 2016 Drgerwerk AG & Co. KGaA
CORPORATE HEADQUARTERS REGION EUROPE CENTRAL REGION MIDDLE EAST, AFRICA REGION CENTRAL
Drgerwerk AG & Co. KGaA AND EUROPE NORTH Drgerwerk AG & Co. KGaA AND SOUTH AMERICA
Moislinger Allee 5355 Drgerwerk AG & Co. KGaA Branch Oce Drger Panama Comercial
23558 Lbeck, Germany Moislinger Allee 53-55 P.O. Box 505108 S. de R.L.
www.draeger.com 23558 Lbeck, Germany Dubai, United Arab Emirates Complejo Business Park,
Tel +49 451 882 0 Tel +971 4 4294 600 V tower, 10th oor
Fax +49 451 882 2080 Fax +971 4 4294 699 Panama City
info@draeger.com contactuae@draeger.com Tel +507 377 9100
Manufacturer: Fax +507 377 9130
Drgerwerk AG & Co. KGaA REGION EUROPE SOUTH REGION ASIA / PACIFIC contactcsa@draeger.com
Moislinger Allee 53-55 Drger Mdical S.A.S. Draeger Medical
23558 Lbeck, Germany Parc de Haute Technologie South East Asia Pte Ltd.
d'Antony 2 25 International Business Park
25, rue Georges Besse #04-27/29 German Centre
92182 Antony Cedex, France Singapore 609916, Singapore
Tel +33 1 46 11 56 00 Tel +65 6572 4388
Fax +33 1 40 96 97 20 Fax +65 6572 4399
dlmfr-contact@draeger.com asia.pacic@draeger.com