Professional Documents
Culture Documents
Nasal Catheters
measured from nose to ear, lubricated and inserted to just above the uvula
**deep insertion can cause air swallowing and gastric distension **must be
repositioned every 8 hours to prevent breakdown
Transtracheal Catheters
Delivers low flow rates (1 to 3 L/min) directly to the trachea through a
small incision
Requires less oxygen (lower flows) because the upper airway deadspace
is bypassed
**Can develop infection and irritation at site and there is a risk of accidental removal
**If pt becomes SOB or has increased **Flush with saline**
**Cleaning Rod **reposition
Cannulas
better tolerated
may be humidified but often not if run @ less than 5 L/min
there are two types of O2-conserving cannulas
Nasal reservoir
reservoir just below the nose stores ~ 20 mL of O2
allows for lower flows because of increase O2 delivery
Pendent reservoir
pendent stores O2 must exhale through nose
Non-rebreathing Mask
Delivers 60 to 90% O2 at flow rates of 8 to 15 L/min (NBRC says 100%)
**Flow rate must be sufficient to keep bag 1/3 to 1/2 inflated at all times
Is equipped with a one-way valve that does not allow exhaled gas into the
reservoir
One way valves are located on both expiratory ports of the mask to prevent
RA entrainment
High flow O2 systems provide all of the inspiratory flow required by the patient at
consistent FiO2s
These devices are normally attached to nebulizers
(Venti-mask may or may not be)
Face Tent
21% to 40% depending on nebulizer setting
Flow rates of 8 to 15 L/min
Used mainly for patients who can not tolerate a mask
Trach Mask
35-60% O2 depending on nebulizer setting
Flow rates of 10 to 15 L/min
**Adequate flow shown by mist flowing out the exhalation port at all times
Oxygen Tent
21 -50% at flow rates of 10 to 15 L/min
Used mainly on children with croup or pneumonia
Problem with leakage
Fire hazard
Important Points
**High FiO2’s (>.60) may not meet the patient’s inspiratory flow demands
To insure adequate flow with stable FiO2’s, a special made high flow device should be
used or two flowmeters set up to provide at least 40 L/min total flow
**A restriction, such as a kink, or water in the tubing, causes back-pressure into the
nebulizer. This decreases the amount of RA entrained and INCREASES the delivered
FiO2.
**Increasing flow on a high flow device will not increase FiO2, only total flow