Programat

Instructions for the software update via USB stick


(P310, P510, P710, EP 3010, EP 5010, CS2, CS3):

Software updates can be easily installed on the furnace by means of a USB stick. For the purpose, a USB memory stick is
required which contains a current software file (e.g. P510:V1.10.iv). The software version on the USB memory stick must be
higher than the one installed on the furnace.

The free software updates for Programat furnaces are available from www.ivoclarvivadent.com/downloadcenter.

A software update requires only a few steps:

1. Open the software update


Press the [Software update] button in the
home screen.

2. Conduct the software update


If the USB stick with the software file is already
connected, the furnace automatically searches for
a valid software file. If the USB stick has not yet
been connected with the furnace, do so now.
Press the [Execute] button.

3. The status bar shows the progress


of the update.

4. Complete the software update


The following messages are displayed:

Software update successful


Software update failed

The furnace has to be switched off and on again by means of the mains switch at the rear of the furnace to
complete the software update.

Important information
Please note that modified Ivoclar Vivadent programs may be overwritten during a software update. Individual
programs are not affected and will not be overwritten.