ERA AUTOMATION & LAB (PVT) LTD

Calibration Certificate
CERTIFICATE NO. OGDC-065

HISTORY OF INSTRUMENT TO BE CALIBRATED: DATE: 15-07-2018


CUSTOMER NAME : OGDCL Nashpa.
EQUIPMENT NAME :digital earth resistance tester
PARAMETER : Signal
MANUFACTURER : Kyoritsu
MODEL/ TYPE : 4105A
RANGE :MULTI
LAST CAL/TEST : -------

CALIBRATION EQUIPMENT:
CALIBRATION INSTRUMENT: 1. Fluke calibrator, Model # 5100B, S # 4520006

CALIBRATION TRACEABILITY:
INTERNATIONAL: 1. Traceable to NIST.
2. ISO 10012-1 ANSI/NCSL Z540-1

CALIBRATION CONDITION
TEMPERATURE : 20 ± 4 C
HUMIDITY : 50 ± 10% RH.

CALIBRATION RESULT: (All result in MΩ)


READ VALUE TRUE VALUE DEVIATION
1.0 MΩ 1.00000 0.00000
5.06 MΩ 5.00000 +0.06000
10.50 MΩ 10.000 +0.5000
50.00 MΩ 50.0000 0.0000
100.00 MΩ 100.000 0.000
These are test rest result, not a specification
Remarks:
1.CONDITION: OK
2.RECOMMENDED DATE FOR THE NAXT CALIBRATION: 14-07-2019

CALIBRATION ENGINEER: CALIBRATION INCHARGE