Professional Documents
Culture Documents
Core Lab New PDF
Core Lab New PDF
2010
Mohsen Masihi
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 1
Reservoir Rock Properties Analysis, Mohsen Masihi
Table of contents
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 2
Reservoir Rock Properties Analysis, Mohsen Masihi
Course overview
This course provides an introduction to reservoir rock properties determination by core analysis. Part of
this course introduces the laboratory equipments as well as the procedures used for the core analysis.
Moreover, the theoretical aspects of the parameters used in the core analysis are briefly described.
Hence, the aim of this is to get familiar with of the main rock properties parameters, the way they can
be measured and the sources of errors in the results obtained from the laboratory measurements. At the
end of the course, you will learn what the main properties of rocks are and also you will learn how to
measure them and how to analyze the range of the uncertainty in the results.
These prepared course notes follow closely the sequence of material that will be presented. However,
these are further suggested materials:
1-J. S. Archie and Wall C. G., Petroleum engineering principles and practice. Graham & Trotman,
London, 1986.
2-J. W. Amyx, Bass D. M. and Whiting R. L., Petroleum reservoir engineering (physical properties).
McGraw Hill, Inc., 1988.
3- X. D. Jing, Rock properties course, Imperial College London, 2003
4- M. Masihi, Reservoir Rock Properties course, Sharif University of Technology, 2007.
5-T. Ahmed, Reservoir engineering handbook, Gulf professional publishing, 2001
6- M Ala, Jing, X. D. and Worthington P., Petrophysics course, Imperial College London, 2003
7- M Araujo Fresky, Rock properties, Imperial College London, 2004
Syllabus
1. Core and plug preparation: Introduction to the machines for the various steps of core preparation
including core slabbing, core pluging and trimming. Also sand/shale analysis and depth
matching through the use of core gamma logger will be introduced.
2. Cleaning and saturation determination. The available instruments for core cleaning as well as
saturation determination will be introduced. This includes extraction/distillation method for
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 3
Reservoir Rock Properties Analysis, Mohsen Masihi
core cleaning (Dean Stark), drying and heating for saturation determination (Retort Oven) and
core saturator for saturation of core before other experiments.
3. Porosity. An introduction to core porosity and various methods of its measurements. Describing Hg
porometer and KeyPhi instruments for porosity measurements.
4. Resistivity. Introduction to the EPS-A instrument for the rock conductivity measurements at surface
pressure and to the EPS-700 at overburden pressure.
6. Capillary pressure. Introduction to capillary measurement methods under drainage and imbibition
processes. Description of CAPRI instrument for measuring capillary pressures.
8. Rock mechanical Properties. An introduction on how to measure the rock properties like Young’s
modulus and Poisson’s ratio using AVS-700 instrument by application of the sonic waves.
Course structure
The course will consist of approximately 12 sessions (each one 3 hours) of laboratory works including
lectures. The course will be assessed by a final term examination and by laboratory reports.
The notes in some places cover more material than can reasonably be covered during the course and in
other places have deliberate gaps for more discussions. Please fell free to ask questions during the
course.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 4
Reservoir Rock Properties Analysis, Mohsen Masihi
Introduction
Knowledge of the physical properties of the rock and interaction between hydrocarbon system and the
formation rock is crucial in understanding and evaluating the performance of a given reservoir. These
data are usually obtained from two main sources: core analysis and well logging. In this course book
we describe the analysis of cores. A core which is a solid cylinder of rock about 3 inches in diameter
and would usually be about 30 feet long is taken by replacing the drill bit by a “core bit” which is
capable of grinding out and retrieving the heavy cylinder rock. Once the cores are retrieved it is crucial
to properly handle (avoid damaging) and preserve them by avoiding exposure to air. When the core
arrives in the laboratory plugs are usually drilled 20-30 cm apart throughout the reservoir interval. Then
the plugs are analyzed with respect to porosity, permeability, saturation, grain density and lithology.
This analysis, which is performed at high sampling frequency and low cost, is called routine core
analysis. The results from routine core analysis are used in interpretation and evaluation of the
reservoir. Examples are prediction of gas, oil and water production, definition of fluid contacts and
volume in place, definition of completion intervals. There are other important measurements with the
aim of obtaining the detailed information about the multiphase flow behavior. This analysis, which is
performed at low sampling frequency due to high cost and more time due to gathering the data, is
called special core analysis. Special core analysis gives information about the rock wettability, the
distribution of oil, gas, and water in the reservoir (capillary pressure data), residual oil saturation and
multiphase flow characteristics (such as capillary pressure and relative permeability). Measurements of
electrical and acoustic properties, which are mainly used in the interpretation of well logs, are
occasionally included in special core analysis.
The outline of this handout is organized as follows: We first describe the main pre-processing steps that
are considered on the core samples to be prepared before experiments. In chapter two we describe the
core cleaning methods, which are required before core analysis tests, as well as the saturation
determination methods. In the next chapter, the porosity measurement techniques are described and the
instruments available in the laboratory for the determination of the core porosity are described. In
subsequent chapters we then describe the techniques to measure and the available machines to
determine respectively the core sample resistivity, interfacial tension, capillary pressure and
permeability.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 5
Reservoir Rock Properties Analysis, Mohsen Masihi
(i) slabbing; the cores need to be slabbed with the aim of making the rock structure visible. This can
simply be done by sawning down the middle of the core and cuting it at desired length.
(ii) plugging: one can drill the plugs at every one foot of the core.
(iii) trimming: both end of the plugs and core samples can be trimmed in order to produce high quality
thin sliced samples without disturbing the structure of the sample.
Moreover, core scanner and core gamma logger may be used for depth matching as well as describing
the core lithology and texture. The detail description of these per-processing steps and the equipments
to do these are described in this section.
1-This machine must only be operated by qualified personnel specially those who has had the
operation training and the safety regulations.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 6
Reservoir Rock Properties Analysis, Mohsen Masihi
Machine Description:
The machine has a motor which supplies the rotational motion for sawing. The main machine parts are:
• Radial saw
• Electro motor
• Shaft, pulleys and rim
• Adjustable tilting table
• Cooling system, including pumps, water supply, flexible pipes
The rotational motion transmitted by two pulleys from electro motor. At either side, there are three
different stepped pulley couples. The speed of motion can be adapted by positioning the compound
diamond rim on the suitable stepped pulley couple on the electro motor and saw system. The dimension
of rock sample should be smaller than the max height of saw from table to be cut electively. Bigger
rocks should be broken to a proper part with other methods. One draw back of this machine is its noise.
The machine noise level is around 70 dBa. During the cutting process, the noise level depends on the
material to be cut, the type of disc used, the rotation speed and the cooling.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 7
Reservoir Rock Properties Analysis, Mohsen Masihi
Experiment operation
The procedure for cutting a given sample is as follows:
• If the rock is big then it should first be cut by another method
• Put the sample on the table
• Check the tension on the pulleys
• Adjust the direction of water line on both saw and sample
• Check water tanks and fill them if they are empty
• Start the water pump and check the direction of water and check the flow
• Start the saw by pressing the start bottom
• Cut the sample by screwing up and down of the saw and pushing the sample table of machine
• Press the stop button
• Unclamp the core sample
However, in working with the machine consider the following safety issues: (i) prevent touching of saw
with the machine sample table, because of damaging of saw, (ii) protect yourself from any splash and
water drops by wearing overall and safety glasses.
You should consider the appropriate rotational speed. The motor shaft rotates at 1,440 rpm. When the
belt is on the central pulley either on motor side and disc side, the disc rotates at 1,440 rpm. When the
belt is on the small pulley on motor side and on the large pulley on the disc side, the disc rotated at
2,160 rpm and finally when the belt is on the large pulley on motor side and on the small pulley on the
disc side, the disc rotated at 960 rpm. The schematic diagram of pulleys at both motor side and disc
sides are illustrated in Fig. 2.
Figure 2: Pulleys
Cutting linear speed in meters/seconds versus disc diameter and speed are shown in Table 1.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 8
Reservoir Rock Properties Analysis, Mohsen Masihi
For cooling of the saw and sample during the operation, water is used. Water is pumped from water
tank and poured directly to the sawing position on the sample. Then, the used water goes to the water
tank. Water tank is composed of to different tank. The recycled water first enters to the first tank, and
after precipitating of cuttings into this tank, the fresh water overflow to the main water tank.
Maintenance
Keep the machine clean.
Do not let water and waste in the tray.
Check carefully the v-belt and its strain.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 9
Reservoir Rock Properties Analysis, Mohsen Masihi
Machine Description:
The machine can accommodate cores measuring from ½” to 3”. A swivel joint with a tap allows
internal irrigation of the core drill and a hose fitted with a tap allows external irrigation of the core drill.
The speed can be adjusted by repositioning the belt (1800 rpm – 2500 rpm – 3500 rpm) like core
slabbing machine.
The machine comprises of:
• clamping stand
• column
• « Spindle-Motor » unit
• adjustable tilting table
• protective housing
• clamping unit
• recycling tank
Table of machine has a rotating capability up to 45º. This capability causes that user can make plugs
from any part of slab even deviated sides.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 10
Reservoir Rock Properties Analysis, Mohsen Masihi
Experiment operation
• Check the tension on the pulleys
• Check the direction of the spindle
• Fill the recycling tank
• Screw down the core drill and lock it in place
• Mount a core sample and lock it firmly in the clamping unit
• Adjust the lower stop on the core drill. 1-2 mm before the end of slab is sufficient for
prevention of plugging the sample in plug driller. Touching of the driller with table causes
severe damage to the driller.
• Close the core drill protective housing
• Press the Start button
• Open the irrigation taps
• Check the flows
• Cut the core sample
• Press the « Stop » button
• Unclamp the core sample
Speed of rotation can be adjusted by changing pulleys. The procedure for this operation is as:
• open the pulley cover
• loosen the 2 locking screws and pulley tension lever
• pinch together the two sides of the spindle belt
• change the position of the belt
• tighten the belts with the lever and lock the two screws in place
• close the cover
If the tension of pulleys is not sufficient then the belt should be changed. The procedure for changing
the belt is as follow:
• make sure that the machine is disconnected
• open the pulley cover
• loosen the two clamping screws and the belt tensioning lever
• pinch together the two sides of the spindle belt
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 11
Reservoir Rock Properties Analysis, Mohsen Masihi
Maintenance
• keep the unit and the protective housings clean
• remove debris and core particles
• remove sludge
• clean all moving parts
• change the cooling fluid as soon as it is dirty
• make sure that the machine is disconnected
• remove the pump and the waste pipe from the recycling tank
• empty the tanks
• clean the tanks and the separators
• fill the tank with cooling fluid
• refit the pump and the waste pipe to the recycling tank
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 12
Reservoir Rock Properties Analysis, Mohsen Masihi
Machine Description:
The basic model can work either in manual feed or with an optional hydraulic automatic feed which is
driven by the domestic water supply (Minimum pressure 1.5 bars). In the automatic mode, user can
determine the speed of rotation of saws. The machine consists of two radial saw that can work together
and cut both end of pugs simultaneously. Each cuts needs nearly 0.2 litter cooling water. This machine
is designed to work with all types of cutting wheel (resinous - diamond - boron carbide) and various
accessories and adaptations enable samples or core sections to be cut lengthways. These include cradles
or devices for holding the samples configurations using two wheels which allow parallel-sided sections
of continuous length to be cut in a single operation. The machine is fitted with a safety cut-out switch
which can be reset, or rewound should there be no power, as well as a gradual starting device. When
the cover is open this safety switch open the electric current and the machine don’t work. The use of
passivated water is strongly recommended to avoid corrosion.
Experiment operation
The machine can work in both manual and automatic mode. By setting two lever taps on the body of
machine, three situations are achieved. The lever taps allow the wheel (saw) to advance or return.
• quick back mode: in this mode the saws go back quickly and positioned at the start point. This
mode can be achieved by setting both taps down.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 13
Reservoir Rock Properties Analysis, Mohsen Masihi
• stop manual: in this mode user should handle the position of saws for trimming the plug
manually and can be achieved by setting top tap to up and bottom tap to dawn
• automatic feed: in this situation samples are trimmed automatically by the machine. The
rotational speed of the saws can be adjusted by the “Movement regulator” beside the lever taps.
Manual mode
Steps for operating in manual mode are as follows:
• Press the start bottom
• Set the taps to " Quick back " position, at the end of the race, the lever is independent of jack,
• Set the taps to the "Manual stop" position.
• Adjust the direction of water line on saw and sample
• Check water tanks and fill them if they are empty
• Start the water pump and check the direction of water and check the flow
• Close the protective housing of machine
• By moving the saws to front start trimming of the sample
• Press the Start button
• Press the « Stop » button
• Unclamp the core sample
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 14
Reservoir Rock Properties Analysis, Mohsen Masihi
Maintenance
Apart from keeping the machine properly clean, no specific maintenance is required. Make sure that
any sediment or waste matter is removed from the tank before starting. Change the fluid according to
the frequency of use and its deterioration over time (shelf life).
Machine Description:
The core samples are carried over the detector by a motor driven belt conveyor. The conveyor is 15 cm
wide and 3 meters long (belt length is 6 meters). The belt is marked in increments of 0.25 meters to aid
in placement of the core along the belt (Fig. 5). Speed and motion of the conveyor belt are controlled
with a stepper motor drive from the SmartLog software. Belt speed is set in the software as 3 cm/min,
but can be adjusted in Factory Setup.
The gamma radiation is detected by a 3x3" Sodium Iodide crystal and photomultiplier located below
the belt in the center of the conveyor. The detector is protected from ambient radiation with low
activity virgin lead shielding.
Calibration
The process of calibration is as follows:
• Belt speed:
1. From the main APPLILAB screen, set the speed to the minimum value of 0.150 ft/min and
validate with OK.
2. Have a stop watch ready. Start the belt and the stop watch. Once the belt has moved by a
minimum of 1 foot, write down the exact belt displacement and the time elapsed. Repeat these
steps for various speeds e.g. for 0.450 ft/min to the maximum speed of 0.750 ft/min.
3. In the file: C:\AppliLab\Project\NGL200 \GammaLogger.xls, open the tab Calibration.
4. Report the set point, distance and time in related columns (Fig 6). The effective "measure"
speed is calculated in the last column. The graph, the gain and the offset are automatically
updated. Note: if the coefficient R2 is less than 0.9990, repeat the measurement. Save and close
the Excel file.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 16
Reservoir Rock Properties Analysis, Mohsen Masihi
• Probe:
In Fig. 7, gain refers to the probe amplifier gain. The amplifier gain can be adjusted in either automatic
or manual mode. Gain calibration must be refreshed from time to time to fit the probe response drift
with aging.
After a gain calibration, you must calibrate the "zero" (background). Natural radioactivity
(background) is usually of low level, especially apparent background accessible to the probe set in the
lead shield. However, this apparent background is not negligible compared to the natural radioactivity
in core sample. Natural radioactivity (background) depends on various factors such as local geology,
climate (rain, temperature) etc...
Zero calibration:
During zero calibration, the apparent background is scanned during a set time (Standard duration is
set to 10,000s to suit 99% of cases).Then this "zero" spectrum is automatically memorized. During
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 17
Reservoir Rock Properties Analysis, Mohsen Masihi
next core scan campaign; this "zero" will be automatically subtracted from all the spectrum
measurements. Note that zero calibration must be refreshed from time to time. Calibration of zero
must follow a gain calibration. The zero (spectral) needs to be experimentally recorded (no manual
edition is possible).
The calibration for U, Th and K can be operated in either automatic or manual mode. Channels
calibration must be refreshed from time to time to fit the probe response drift with aging. In order
to calibrate U, Th and K in sequence:
− Install the standard U over the detector in the lead tunnel, and then click on the button
accordingly to start a spectrum scan.
− The running process title and the time countdown are displayed in the top right corner of the
trend window.
− After completion of the spectrum acquisition, the countdown is reset and the probe amplifier is
automatically adjusted to match the region of interest of the standard.
− Repeat theses steps in sequence similarly for Th and then K standards.
Experiment operation
After completing the calibrations (belt, gain, zero and U, Th and K channels), you can start the
measurement of your core samples. Lay the cores in order of depth with the shallow end nearest to the
lead tunnel. Align the first core with the groove each side of the belt. Avoid installing the cores far from
the longitudinal axis (Fig 8). The operation can be done manually or automatically.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 18
Reservoir Rock Properties Analysis, Mohsen Masihi
Manual operation:
Manual operation is possible for specific uses disconnected from spectral logging. In case
you run the spectrum acquisition manually, the duration is free and the conveyor can be
stopped. This function can be used for checking the level of cores before selecting the
conveyor speed for instance.
The conveyor manual operation is basically used during of speed calibration. If you run the
conveyor at the same time as manual spectral acquisition, you cannot link the position with
the depth. Therefore, a spectral gamma logging should be conducted in automatic mode.
-To run the conveyor from the interface application, you can access to the belt speed set point
(bottom left) and to the ON / OFF switch (bottom right). Click in the window to pop up input
value dialogue box (Fig. 9).
-Type a value in the range from 0.15 to 0.75 ft/min and validate with OK.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 19
Reservoir Rock Properties Analysis, Mohsen Masihi
-Click on button labeled “Click to start the belt” to start the conveyor. When facing the electrical
cabinet, the belt moves from left to right. To stop the conveyor, click on button labeled “Click to
stop the belt”.
-You can start the spectral acquisition by clicking on GO. The spectral radioactivity (counts in Y axis,
energy in X axis) is refreshed continuously. Equivalent values of Uranium, Thorium and K
(potassium) and total activity are directly displayed at the bottom of the window, according to
calibration data. The Y axis is auto ranged and the scale is displayed in the top left corner.
-Click in the spectrum trend window (Fig 10): a vertical line is drawn at the mouse position. The
energy level (X value) is displayed and the total counts (Y value) is refreshed continuously.
-You can stop the spectral acquisition by clicking on the button Stop. The spectral trend remains
displayed as the relevant concentration of U, Th, K and the total activity.
Note: When you click on clear, the curve, the relevant concentration of U, Th, K and the total
activity are cleared out from the memory.
Automatic operation:
-To start the spectral acquisition, click on the button “Start Measure”. After a few seconds, you are
requested to name the file that will be used for your report. Here we select "MyReport". If this
name does not exist yet, this file will be created as a copy of the master GammaLogger.xls. The
file is opened. If a file with this name has already existed, it will be open. You are prompted to
check the parameters and to press start. It is time to edit the core identification and check the
belt speed as follow (Fig 11).
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 20
Reservoir Rock Properties Analysis, Mohsen Masihi
-You can edit the yellow cells only (column B). For instance, you cannot edit the belt speed directly.
The belt speed is calculated from the Sample Increment and the Sample Duration that you can
edit. In this example the current depth is 1,004 ft, and the initial depth is 1,000. This illustrates
how to go on with a file used earlier. In this case, the previous data from 1,000 to 1,003 will be
kept and the core scan starts from 1,004. Therefore, the Active Line (calculated by Excel) is 17;
in front of the Sample Depth 1,004. Note: In case you set the current depth to 1,002 ft, and the
core scan starts from 1,004. The data from 1,000 to 1,002 (excluded) will be kept, and the data
from 1,002 (inclusive) will be overwritten with new data. Moreover, changing the sample
increment or duration, will change the belt speed and the sample depth (automatic
recalculation).
-In our example the Sample Increment is 1 foot. We can skip 1, 2, 3 feet or any multiple of the
Sample Increment. To skip a missing core section, just enter "1" in the suitable cell(s) in the
column B in regard of the relative depth. In our XLS illustration, we skip the samples l,006
and 1,007. Finally we scan the sample beginning at depth 1,005, then we go on with the
sample beginning at 1,008 depth.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 21
Reservoir Rock Properties Analysis, Mohsen Masihi
-Once the previous steps are completed, click on start to run the experiment.
-Confirm the sample depth read at current depth in the XLS file. If you want to change, click on NO.
-On the experiment is completed, the XLS file is automatically saved and closed. Open the
MyReportxls (in our example, the report file was named by the operator "MyReport"). The path
is: C:\AppliLab\Project\NGL200\Excel Files\MyReport.xls.
-Open the tab GammaLog. We can see the values of equivalent U, Th, K and total gAPI interpreted
from the calibration versus the sample depth (Fig 12).
Figure 12 : Illustration of the values of equivalent U, Th, K and total gAPI interpreted from the
calibration versus the sample depth in the report
-Open the tab Graph. We can see the values of equivalent U, Th, K and total gAPI interpreted from the
calibration versus the sample depth (Fig 13).
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 22
Reservoir Rock Properties Analysis, Mohsen Masihi
Figure 13: Illustration of the variation of equivalent U, Th, K and total gAPI interpreted from the
calibration versus the sample depth
Maintenance
• Check that nothing can obstruct the belt motion.
• Check that drums and rollers are aligned to drive the belt in straight line.
• A lateral drift of 2 - 3 cm is not a problem if the belt comes back to the initial position.
• Avoid to load the cores far from the longitudinal axis
• Check the belt and chain tension once a month.
• Check that the motor fan is still efficient to cool the motor once a year.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 23
Reservoir Rock Properties Analysis, Mohsen Masihi
where Vw , Vo , Vg and Vp are water, oil, gas and pore volumes respectively and S w , So , S g are water, oil
and gas saturations. Note that fluid saturation may be reported either as a fraction of total porosity or as
a fraction of the effective porosity. Since fluid in pore spaces that are not interconnected can not be
produced from a well, the saturations are more meaningful if expressed on the basis of the effective
porosity. The weight of water collected from the sample is calculated from the volume of water by:
Ww = ρ wVw
Where ρ w is water density in g/cm3. The weight of oil removed from the core may be computed as,
Wo = Wl − Ww
Where Wl is the weight of liquids leaving the core sample in gr. Oil volume may then be calculated as
Wo / ρ o . Pore volume Vp is determined by a porosity measurement, and oil and water saturation may be
calculated by the above equation. Gas saturation can be determined using the summation equation.
There are several methods to clean and evaluate the fluid saturation within a core plug sample. Here we
briefly review some of them:
2-Centrifuge Flushing
A centrifuge which has been fitted with a special head sprays warm solvent onto the sample. The
centrifugal force then moves the solvent through the sample. The used solvent can be collected and
recycled
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 24
Reservoir Rock Properties Analysis, Mohsen Masihi
4-Soxhlet Extraction
A Soxhlet extraction apparatus is the most common method for cleaning sample, and is routinely used
by most laboratories. As shown in Fig. 14a, toluene is brought into a slow boil in a Pyrex flask, its
vapors move upwards and the core becomes engulfed in the toluene vapors (at approximately 110 C).
Eventually the amount of water within the core sample in the thimble will be vaporized. The toluene
and water vapors enter the inner chamber of the condenser; the cold water circulating around the inner
chamber condenses both vapors to immiscible liquids. Recondensed toluene together with liquid water
falls from the base of the condenser onto the core sample in the thimble; the toluene soaks the core
sample and dissolves any oil with which it conic into contact. When the liquid level within the Soxhlet
tube reaches the top of the siphon tube arrangement, the liquids within the Soxhlet tube are
automatically emptied by a siphon effect and flow into the boiling flask. The toluene is then ready to
start another cycle. A complete extraction may take several days to several weeks in the case of low
API gravity crude or presence of heavy residual hydrocarbon deposit within the core. Low permeability
rock may also require a long extraction time
Figure 14: Schematic diagram of (a) Soxhlet and (b) Dean- Stark apparatus
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 25
Reservoir Rock Properties Analysis, Mohsen Masihi
5-Dean-Stark Distillation-Extraction
The Dean-Stark distillation provides a direct determination of water content. The oil and water area
extracted by dripping a solvent, usually toluene or a mixture of acetone and chloroform, over the plug
samples. In this method, the water and solvent are vaporized, recondensed in a cooled tube in the top of
the apparatus and the water is collected in a calibrated chamber (Fig. 14b). The solvent overflows and
drips back over the samples. The oil removed from the samples remains in solution in the solvent. Oil
content is calculated by the difference between the weight of water recovered and the total weight loss
after extraction and drying.
6-Vacuum Distillation
The oil and water content of cores may be determined by this method. As shown in Fig. 15, a sample is
placed within a leak-proof vacuum system and heated to a maximum temperature of 230oC. Liquids
within the sample are vaporized and passed through a condensing column that is cooled by liquid
nitrogen.
7-Retort Oven
The retort oven is used to determine the porosity and total saturation of water (Sw) and residual
saturation of oil (Sro) in core samples, either fresh or preserved ones.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 26
Reservoir Rock Properties Analysis, Mohsen Masihi
In each of these methods, the number of cycles or amount of solvent which must be used depends on
the nature of the hydrocarbons being; removed and the solvent used. Often, more than one solvent must
be used to clean a sample. The solvents selected must not react with the minerals in the core. The
commonly used solvents are: acetone/benzene, benzes/methol alcohol, carbon/tetrachloride,
chloroform, methylene dichloride, mexane, naphtha, tetra chloroethylene, toluene, trichloro ethylene
and xylene. Toluene and benzene are most frequently used to remove oil and methanol and water is
used to remove salt from interstitial or filtrate water. The cleaning procedures used are specifically
important in special core analysis tests, as the cleaning itself may change wettabilities. The core sample
is dried for the purpose of removing connate water from the pores, or to remove solvents used in
cleaning the cores. When hydratable minerals are present, the drying procedure is critical since
interstitial water must be removed without mineral alteration. Drying is commonly performed in a
regular oven or a vacuum oven at temperatures between 50°C to 105oC. If problems with clay are
expected, drying the samples at 60°C and 40 % relative humidity will not damage the samples.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 27
Reservoir Rock Properties Analysis, Mohsen Masihi
Machine Description
This consists of a distillation / extraction glassware unit and a heating mantle with thermostatic
controller. The glassware for one sample is composed of boiling flask with extractor, a sample support
screen, volumetrically graduated water receiving tube of 5 or 10 ml, condenser and desiccant drying
tube. A soft tubing is also used to connect the condenser to the water cooling unit. All these devices are
mounted on a frame rack (Fig.16).
Figure 16: Dean Stark apparatus model (left) DS_1.5”x6 and (right) DS_4”x3
Experiment operation:
The objective of the experiment is to determine the oil, water and gas saturation of a core sample. The
procedure is as foolow:
1. Weigh a clean, dry cellulose thimble. Use tongs to handle the thimble.
2. Place the cylindrical core plug inside the thimble, then quickly weigh the thimble and sample.
3. Fill the extraction flask two-thirds full with toluene. Place the thimble with sample into the long
neck flask.
4. Tighten the ground joint fittings. Start circulating cold water in the condenser.
5. Turn on the heater and adjust the rate of boiling so that the reflux from the condenser is a few
drops of solvent per second. The water circulation rate should be adjusted so that excessive
cooling does not prevent the condenser solvent from reaching the core sample.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 28
Reservoir Rock Properties Analysis, Mohsen Masihi
Worg Wdry ρw ρo Vw Wo Vo Vp
So Sw Sg
gr gr gr/cc gr/cc cc gr cc cc
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 29
Reservoir Rock Properties Analysis, Mohsen Masihi
Machine Description
The retort oven is composed of following items:
• Insulated oven with heat control with a maximum operating temperature of 700 °C.
• 12 traps and stainless steel screens
• stainless steel tank for condensed water
The procedure of the test is as follow:
• Install water supply to each end of bath which has a baffle in centre, ie each end independent of
other end.
• Set required temperature using UP and DOWN arrows buttons only.
• Heat the chamber (cover closed) to 540-650 °C (1000–1200 °F). When retort cups are placed in
the heated enclosure, the temperature will fall down about 400 °C, and then will go up again.
• The fresh sample is roughly crushed, weight = 125 grams, and placed in the retort cup.
• Volume of the crushed sample = VT with VT = Weight / Natural density
Note : measurement must be realized on an another piece of core whatever is its form, about 30
cm3, weighed with its fluids, water, oil, gas and with total volume will have been determined
with a volumetric pump.
• Place all the retort cups (12 max) in the retort, even there is no sample to analyze in each of
them and close the retort which is under voltage. The receiving graduated 20 cc tubes clean are
in place.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 30
Reservoir Rock Properties Analysis, Mohsen Masihi
• Read the volume of water “INI.W” collected after “T” minutes. This time “T” corresponds
exclusively to the water contained into the pores of the sample and not to the water contained
into the clays. It is determined in pointing out according to the time (or to the temperature) the
volume of water collected into the receiving tubes. It is the value of the time which corresponds
with the middle of the plateau, or of the inflection point (in case of clayey samples).
• Read the volume of water “FINAL W” collected when temperature reaches 650 °C.
• Read the volume of oil “OBS.O” collected when the temperature reaches 650 °C. This quantity
of oil collected is by default because oil is lost by cracking during heating, from which a
calibration is needed.
Calibration
In sands samples contained in retort cups, one places growing known quantities of oil from Occ to
10cc. Using normally the retort oven, one notes the “OBS.O” volumes. Then plot working curves of
correction: see correction table (for example a retort oven and crude oil 45° API)
For reading higher then 1.0 cc, add 0.4 to determine the corrected volume. Correction of the volume of
oil collected. Use the precedent table. It is determined “CORR.O”.
Calculation
Amount of oil in the sample:
CORR.O
Ob = × 100
VT
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 31
Reservoir Rock Properties Analysis, Mohsen Masihi
Porosity : φ = Ob + Wb + Gb
Wb
STW = 100
Total saturation of water : φ
Residual saturation of oil :
Ob
S RO = 100
φ
Machine Description
This machine can either work on automatic mode and semi-automatic mode. Although standard
parameters fulfil most of applications, the operator can easily change the duration steps of the
automated mode. Furthermore, it is possible to shift to semi-auto or manual mode for any specific
process. In the semi automated mode the operator controls the Electro-Valves, as in the manual mode
where the operator work with the manual valves. In semi-auto and manual mode the steps duration and
valve sequence is custom. For very low permeability samples, you can increase the "additional sample
evacuation" step to evacuate the samples longer than the saturant (and avoid vaporization and salt
concentration change or precipitation). This system is composed of the following items as shown in
Fig. 18 and its main feature is summarized in Table 4.
• A console with injection pump, valves and PLC (Programmable Logical Controller)
• A pressure vessel
• A vacuum pump
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 32
Reservoir Rock Properties Analysis, Mohsen Masihi
To achieve a good quality saturation on a low permeability sample, thoroughly evacuation (sample and
test brine) is followed by saturation at pressure controlled via a turn-key unattended program.
Experiment operation:
To do the experiment, some pre-steps to prepare the system are necessary before using either
automatic, semi-auto or manual modes. These are four main steps to prepare the system:
i)Setting the target pressure. For setting the target pressure you can use the knob on the pressure gauge.
You can change the set point by rotating the knob to move the red tip needle to the requested value.
(Fig. 19)
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 33
Reservoir Rock Properties Analysis, Mohsen Masihi
ii)Setting the pump. The pump panel is located at the right low corner of the console (Fig. 20left).
Screw the knob to reduce the flow rate. The setting must be at least higher than 30% of the range (Fig.
20right). Before starting an automatic process, switch “On” the pump. If you plan to run a semi-
automatic mode or manual mode, you will switch on the pump on request.
Figure 20: Illustration of: (left) setting the pump and (right) pump flow rate
iii)Sample loading. Discharge the pressure in the unit by setting the manual valves MV3 to by-pass and
MV2 to vent. Disconnect the tubing from the lid. Then open the cell by unscrewing the lid. After you
are checking the sample identification, install the sample with the tray into saturator cell. Eeventually
complete the volume with some plain plugs in order to reduce the volume of brine to inject (Fig. 21).
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 34
Reservoir Rock Properties Analysis, Mohsen Masihi
Check the lid seal. Screw the lid to the pressure vessel. Connect back the tubing. Install the tray with
samples into the saturator cell. Check the lid seal. Figure 22 shows a home-made hook to lift the tray.
iv)Jars preparation. Poor the saturant (brine) into the tank jar (2 ports model). Let enough free space for
bubbling during degassing. Empty the water trapped in the jar (1 port) connected to the vacuum pump.
Close the two jars. (Fig. 23)
1-Automatic mode
• Switch the selector in horizontal position to Automatic option. The green bulb is lightening.
• Press on the button START, the green bulb "IN PROGRESS" is lightening.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 35
Reservoir Rock Properties Analysis, Mohsen Masihi
• There are six steps (sample and saturant evacuation, sample additional saturation, saturant to atm
pressure, by-pass opens, pump priming, saturation under pressure) that are processed automatically.
• You can abort the cycle at any time by pressing on the STOP button. Check the pressure and the
pumps status.
Figure 24 shows the switch time table with default time span. The global duration is about 3 1/2 hour.
The long durations are respectively T6 (saturation under pressure), T2 (additional saturation), T1
(sample and saturant evacuation) and for 2, 1½ and 1 hours. Others durations are similar to a
commutation time. Note that any of these durations can be modified. The sample and saturant
evacuation step start simultaneously with T2 but as soon as T2 is elapsed, the saturant is isolated while
the sample goes on to be evacuated until T1 is elapsed. This allows evacuating the saturant for a shorter
time T2 than the sample evacuated during T1 and to prevent from potential saturant evaporation (in
case of brine etc). Make sure that no manual valve is set open during semi-auto process.
Saturant to atm. press.
Saturation @ pressure
End of Step # 1
End of Step # 2
End of Step # 3
End of Step # 4
End of Step # 5
End of Step # 6
Sample additional
Sample and
after
evacuation
Saturant
Pump Priming
By-pass open
about
evacuation
3 1/2 h
Start Stop
T2 : 60 '
EV4 ON
EV4
EV5 ON
EV5
EV6 ON EV6
EV7 ON EV7
EV8 ON EV8
Pump run Pump
In progress Completed
Figure 24: The time duration of six steps used in the automatic mode
Step1: Samples and brine evacuation with T2 = 60 min (note that T1 starts simultaneously)
• The valves EV4 and EV5 are open to the vacuum pump during T2 (not T1)
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 36
Reservoir Rock Properties Analysis, Mohsen Masihi
• Check that the vacuum pump is running and bubbles appear immediately in the saturant jar (as for
boiling water). If no bubble comes after the first minute elapsed, stop the process and fix the leak.
• Check that no saturant (brine) is going out from the trap jar to the vacuum pump.
Step 2: Relieving vacuum in the brine jar with T1 = 90 min, ends 30 min after T2
• The valve EV5 is closed to isolate saturant brine jar from the vacuum pump. T2 (1 minute default
value) is devoted to these operations (see Fig. 25 for the steps 1 & 2)
Vacuum Vacuum
5(open) 5
8 8
4 4
Vent Vent
7 (closed)
7
Pump
(stopped)
6 6
By-Pass By-Pass
Step#1 Sample & Saturant evacuation T2=60 min Step#2 Sample additional evacuation T1-T2=30 min
T1=90min starts at same time as T2
Figure 25: Illustration of the schematic connections used in the steps 1 and 2
• The valve EV8 is open: the vacuum is relieved in the desaturant jar. The brine pressure builds up to
atmospheric but stay degassed.
• The by-pass EV6 is open, thus the saturant invades the sample vessel under push-pull action of
atmospheric pressure at valve EV8 and vacuum at valve EV4 (see Fig. 26 for the steps 3 & 4)
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 37
Reservoir Rock Properties Analysis, Mohsen Masihi
Vacuum Vacuum
5 5
8 8
4 4
Vent Vent
7 7
6 6
By-Pass By-Pass
Figure 26: Illustration of the schematic connections used in the steps 3 and 4
Vacuum Vacuum
5 5
8 4 8
4
Vent Vent
7 7
Pump run
6 6
By-Pass By-Pass
Figure 27: Illustration of the schematic connections used in the steps 5 and 6
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 38
Reservoir Rock Properties Analysis, Mohsen Masihi
• The injection pump hold on (stops temporarily) as the pressure set point ("target") is reached.
If the pressure decreases before T6 is elapsed, the pump restarts to build-up the pressure up back to the
target value. This ensures the saturation of low permeability samples (see Fig. 27 for the steps 5 & 6)
Figure 28: button of valves panel from which the operating mode can be changed
Note that you run a single mode at once (e.g. semi-auto, automatic or manual). Also when an automatic
mode is running, just press the STOP button before starting a manual mode.
3-Manual mode
Manual mode is a custom process where operator adjusts the EV valves manually. Make sure that no
electro-valve is set open during manual process. Switch the selector to Manual. You can open as many
valves as you want at the same time and run any step of yours as long as required. Figure 29 shows the
various status of tank.
The valves and pump sequence to reproduce the automatic process in the manual mode is as follow:
• Check that the vacuum pump is running and bubbles appear immediately in the saturant jar (as if
the water was boiling). If no bubble comes within the first minute, stop the process and fix the leak.
• Check that no desaturant (brine) is going out from the trap jar to the vacuum pump.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 39
Reservoir Rock Properties Analysis, Mohsen Masihi
Figure 29: Illustration of tank status by setting it to: the vent, the vacuum pump, be isolated
There are 7 steps with various the ellipse(s), point(s) and the valve(s) actions. You can set custom
duration for each step. The default duration is summarized in Table 5:
1 30 Global evacuation,
4 2 By-pass open
5 1 Pump priming
These are the description of various steps. Also Fig. 30 show the connections at each step.
step 2: set V2 to the close and set V1 to the vacuum (sample additional evacuation).
step 6: stop the vacuum pump and set V1 to the close (pump brine into core vessel for 120 min)
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 40
Reservoir Rock Properties Analysis, Mohsen Masihi
Vacuum Vacuum
(Action )
Vent
2 Vacuum 1 Vent
2 Closed 1
Stop
3 3
Closed
Vacuum Vacuum
Vent
2 Vent 1 Vent
2 1
3 By-pass
3
Vacuum
Vacuum Vacuum
2 1 2 1 Vent
2 1
Vent Vent
Run Stop
Pump 3 3 By-pass
3
Figure 30: Illustration of the schematic connections used in the steps 1-7
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 41
Reservoir Rock Properties Analysis, Mohsen Masihi
3-Porosity
From the viewpoint of petroleum engineers one of the most important property of a reservoir rock is
porosity. Porosity is a measure of storage capacity of a reservoir. It is defined as the ratio of the pore
volume to bulk volume, and it may be expressed as either a percent or a fraction,
PoreVolume Bulk Volume − Grain Volume
φ= =
Bulk Volume Bulk Volume
Two types of porosity are total or absolute porosity and effective porosity. Total porosity is the ratio
of all the pore spaces in a rock to the bulk volume of the rock while the effective porosity φe is the ratio
of interconnected void spaces to the bulk volume. Thus, only the effective porosity contains fluids that
can be produced from wells. For granular materials such as sandstone, the effective porosity may
approach the total porosity, however, for shales and for highly cemented or vugular rocks such as some
limestones, large variations may exist between effective and total porosity.
Porosity may be classified according to its origin as either primary or secondary. Primary or original
porosity developed during deposition of the sediment. Secondary porosity is caused by some geologic
process subsequent to formation of the deposit. These changes in the original pore spaces may be
created by ground stresses, water movement, or various types of geological activities after the original
sediments were deposited. Fracturing or formation of solution cavities often will increase the original
porosity of the rock.
Figure 31: Cubic packing (a), rhombohedral (b), cubic packing with two grain sizes (c), and
typical sand with irregular grain shape (d).
Grain size distribution and sorting can influence the porosity. For a uniform rock grain size,
porosity is independent of the size of the grains. A maximum theoretical porosity of 48% is achieved
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 42
Reservoir Rock Properties Analysis, Mohsen Masihi
with cubic packing of spherical grains, as shown in Fig. 31a. Rhombohedral packing, which is more
representative of reservoir conditions, is shown in Fig. 31b; the porosity for this packing is 26%. If a
second, smaller size of spherical grains is introduced into cubic packing Fig. 33c, the porosity
decreases from 48 to 14%. Thus, porosity is dependent on the grain size distribution and the
arrangement of the grains, as well as the amount of cementing materials. Not all grains are spherical,
and grain shape also influences porosity. Typical reservoir sand is illustrated in Fig. 31d.
Moreover, the rock compaction can affect the porosity. This is the process of volume reduction due
to an externally applied pressure. For extreme compaction pressures, all materials show some
irreversible change in porosity. This is due to distortion and crushing of the grain or matrix elements of
the materials, and in some cases, recrystallization. The variation of porosity with change in pressure
can be represented by,
C f ( P2 − P1 )
φ2 = φ1e
From the definition of porosity, it is evident that the porosity of a sample of porous material can be
determined by measuring any two of the three quantities: bulk volume, pore volume or grain volume
from core plugs (Fig. 32).
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 43
Reservoir Rock Properties Analysis, Mohsen Masihi
i) Bulk volume:
Although the bulk volume may be computed from measurements of the dimensions of a uniformly
shaped sample, the usual procedure utilizes the observation of the volume of fluid displaced by the
sample. The fluid displaced by a sample can be observed either volumetrically or gravimetrically. In
either procedure it is necessary to prevent the fluid penetration into the pore space of the rock. This
can be accomplished by:
(1) coating the sample with paraffin or a similar substance,
(2) saturating the core with the fluid into which it is to be immersed, or
(3) using mercury.
Gravimetric determinations of bulk volume can be accomplished by observing the loss in the weight
of the sample when immersed in a fluid or by change in weight of a pycnometer with and without
the core sample.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 44
Reservoir Rock Properties Analysis, Mohsen Masihi
(4) It has a high diffusivity so affords a useful mean for determining porosity of low
permeability rocks.
The schematic diagram of the helium porosimeter shown in Fig. 33 has a reference volume V1, at
pressure P and a matrix cup with unknown volume V2, and initial pressure P2. The reference cell and
the matrix cup are connected by tubing; the system can be brought to equilibrium when the core holder
valve is opened, allowing determination of the unknown volume by measuring the resultant equilibrium
pressure p. (pressure P1 and P2 are controlled by the operator; usually P1 = 100 and P2=0 psig). When
the core holder valve is opened, the volume of the system will be the equilibrium volume V, which is
the sum of the volumes V1 and V2. Boyle's law is applicable if the expansion takes place isothermally.
Thus the pressure-volume products are equal before and after opening the core holder valve:
2 2 = P (V1 + V2 )
1 1 + PV
PV
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 45
Reservoir Rock Properties Analysis, Mohsen Masihi
system, including the changes in tubing and fittings caused by pressure changes during equalization. A
correction factor may be introduced to correct for the composite system expansion. The correction
factor is determined for porosimeters before they leave the manufacturer, and this correction is built
into the gauge calibration in such a way that it is possible to read the volumes directly from the gauge.
Another method of pore volume determination is to saturate the sample with a liquid of known density,
and noting the weight increase (gravimetric method).
When a rock has a small fraction of void space, it is difficult to measure porosity by the mentioned
methods. In this case, mercury injection (Fig. 34a) is used which has the principle of forcing mercury
under relatively high pressure in the rock pores. A pressure gauge is attached to the cylinder for reading
pressure under which measuring fluid is forced into the pores. Fig. 34b shows a typical curve obtained
from the mercury injection method. The volume of mercury entering the core sample is obtained from
the device with accuracy up to 0.01 cm3.
Figure 34: Mercury injection pump (a) and porosity through mercury injection (b).
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 46
Reservoir Rock Properties Analysis, Mohsen Masihi
3-1 Hg porometer
The Mercury Pump is designed to measure the bulk volume of a sample and also pore volume in a dry
core sample. By measuring these two parameters, porosity of the sample can be calculated. The
Mercury Pump has the advantage to allow very rapid measurements of regularly or irregularly shaped
core samples. However, students should consider the following saftey notices:
• Any doubt must be resolved before performing a test or maintenance.
• Protect yourself from any mercury contact by wearing overall.
• The pycnometer lid has a rapid lock closure with O'ring seal. This O'ring ensures the tightness
of the system. Make sure it is not damaged and installed in the groove correctly.
• Do not let other liquids in the system.
• Make sure the sample is removed after each experiment
Machine Description:
The Hg porometer is based on a volumetric pump attached to a sample vessel. The vessel is used as a
pycnometer. The pycnometer lid has a rapid lock closure with O'ring seal. A needle valve (also named
drain plug) in the lid opens the chamber to atmosphere. The movement of the pump metering plunger
is indicated on a volume gauge. The pressure in the pycnometer is read on the vacuum gauge or on the
pressure gauge (according to the sequence).
The main machine parts are shown in Fig. 35.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 47
Reservoir Rock Properties Analysis, Mohsen Masihi
The hand wheel dial is graduated in 0.01 of rotation subdivisions and permits estimation of plunger
displacement to 0.005 rotation. Due to mechanical transmission, each rotation is equivalent to 1.375 cc.
Fig. 36 shows an example of reading rounded to 0.01 rotation. The red figure of the counter is reported
on the clock dial. When you can read 56.1 on the counter ("56" in black color, "1" in red color) and the
clock dial displays 1.3, the global reading is 56.13 rotations.
The reading on the volume gauge multiplied by 1.375 gives the actual volume. Some features of the
system are mentioned in Table 6.
Diameter 1"½
Core sample maximum size
Length 3"
Wetted part Stainless steel
Experiment operation
The following steps must be performed:
1. Make sure the pycnometer is empty (no sample inside).
2. Lock the lid and open the drain plug by one turn.
3. Inject the mercury (turn the wheel anticlockwise) until the first bead of mercury appears at the drain
outlet.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 48
Reservoir Rock Properties Analysis, Mohsen Masihi
6. Make sure to allow enough void space to prevent from overflow when closing the lid with the
sample in place.
9. Install the sample into the pycnometer. Make sure that the sample dimensions are compatible with
the specification.
10. Close and lock the lid, leaving the drain plug open.
11. Inject the mercury until the first bead of mercury appears at the drain outlet.
12. Write down the dial read at the volume gauge. The pycnometer is full: the void space is
occupied by mercury and the sample. The pressure is atmospheric; the pore volume is not invaded.
14. Turn the hand wheel in order to build up the pressure up to the reference value (typically 750 psi).
Vb = c (V1 − V0 )
φ = c (V1 − V2 ) where c is the volume coefficient
The porometer internal volume expands by a volume called correction factor when building the
pressure up to the reference pressure. The correction factor must be calibrated before any measurement
campaign. It is easy to determine this by calculating the difference of the two volume reading:
1) at atmospheric pressure and,
2) at reference pressure with the empty porometer (no sample inside).
1-Direct measurements: Core length and diameter/ Pore volume/ Gas permeability
2-Calculated parameters: Sample bulk volume / Grain volume/ porosity / Grain density (if sample is
weighed)/ Klinkenberg slip factor "b"/ Klinkenberg corrected permeability/ Inertial coefficient
The gas permeability determination is based on the unsteady state method (pressure falloff ) whereas the
pore volume is determined using Boyle’s law technique. Length and diameter of the core sample are
measured with digital caliper and subsequently bulk volumes are determined automatically. The
instrument is also provided with a unique automatic core sample loading mechanism which can process
up to 20 samples. A summary of operational conditions of this instrument is given in Table 8.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 50
Reservoir Rock Properties Analysis, Mohsen Masihi
Moreover, the initial and final conditions are shown in Fig. 37.
At the initial condition, total numbers of gaseous moles are:
n(total) =n (helium) + n(air)
ntotal =
P1 (Vr + Vv )
+
(
Pa Vd + V p )
Z He1RT Z a RT
At the final condition, total numbers of gaseous moles are:
n(total )= n(helium& air)
ntotal =
(
nHe P2 Vr + Vv + Vd + V p
+
)
nair P2 Vr + Vv + Vd + V p
ntotal Z He 2 RT ntotal Z a 2 RT
Now assuming the assumptions: P1>> Pa , and: n He >> n air thus it comes as
nHe n
≅ 1 and air ≅ 0
ntotal ntotal
Hence, the final material balance that gives after simplifications:
ntotal =
(
P2 Vr + Vv + Vd + V p )
Z He 2 RT
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 51
Reservoir Rock Properties Analysis, Mohsen Masihi
P × Z He 2
Vr 1 − 1 − Vv
P2 × Z He1
Vp = − Vd
Pa × Z He 2
1−
P2 × Z He 2
And the porosity is given by:
φ = 100Vp / BV
(
BV = π D 2 / 4 L)
Experiment operation:
1. Connect the keyphi to the main power supply. Connect the air to the air inlet. Connect the N2 or
He cylinder to the gas inlet.
2. Connect the 3 way valve of the confining pump to refill position. Start the pump in Start/Empty
mode, piston is in the top position. Connect a reservoir containing the driving oil to the valve Start
the pump in Refill mode until the pump is full.
3. Connect the 3 way valve of the confining pump to confining position. The pump is ready to
transfer driving fluid into the coreholder.
4. Select the carousel corresponding to the core diameter. Put the core samples in the carousel up to
20 max. Mount the carousel on its position and screw the crank of the carousel and then close the
door.
5. Open the calibrated file you want to use to make your measures in the directory
C:\APPLILAB\PROJECT\KEYPH[\EXCEL FILE\
7. Fill the fields Sample ID, Weight, Atmospheric Pressure, Sample Position (in the Sample Holder),
Confining Pressure, Inlet pressure, Stability Set Point and Vacuum Time Set Point
8. Choose the measures you want to do on each sample (porosity or permeability or both)
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 52
Reservoir Rock Properties Analysis, Mohsen Masihi
9. In case for a sample you only want to perform a permeability measure you have to fill the field
pore volume.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 53
Reservoir Rock Properties Analysis, Mohsen Masihi
4-Resistivity
Porous rocks, which are comprised of solid grains and void space, with the exception of certain clay
minerals are nonconductors. The electrical properties of a rock depend on the geometry of the voids
and the fluid with which those voids are filled. The fluids of interest in petroleum reservoirs are oil,
gas, and water. Oil and gas are nonconductors while water is a conductor when it contains dissolved
salts, such as NaCl, MgCl2, KCl that are normally found in formation water. Current is conducted in
water by movement of ions and can therefore be termed electrolytic conduction. The resistivity of a
porous material is defined by,
rA
R= with: r = resistance (Ohm-meter), A = cross-sectional area (m 2 ), L = length (m)
L
For a complex material like rock containing water and oil, the resistivity of the rock depends on various
factors including water salinity, temperature, rock porosity, composition, pore geometry and formation
stress.
The resistivity of an electric current in porous rock is due primarily to the movement of dissolved ions
in the brine that fills the pore of the rock. The resistivity varies with temperature due to the increased
activity of the ions in solution as temperature increases. Due to the conductivity properties of reservoir
formation water, the electrical well-log technique is an important tool in the determination of water
saturation versus depth and thereby a reliable resource for in situ hydrocarbon evaluation. An empirical
relation for this was developed by Archie in 1942, the so called Archie’s equation for clean water-wet
sandstones over a reasonable range of water saturation and porosities. In practice, Archie’s equation
should be modified according to the rock properties: clay contents, wettability, pore distribution, etc.
The first equation to express the electrical properties of rocks is the formation factor F, defined as,
Obviously, the factor depends on the pore structure of the rock. The second equation to express the
electrical properties of rocks is the resistivity index I, defined as,
Rt
I= Rt = rock resistivity when saturated partially with water, Ω.m
Ro
Wyllie has developed a relation between the formation factor and other rock properties, such as
porosity φ and tortuosity τ defined as ( La / L ) where L is the length of the core and La represents
2
the effective path length through the pores. Based on simple pore models the following relationship has
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 54
Reservoir Rock Properties Analysis, Mohsen Masihi
been derived, F = τ / φ . Moreover, Archie has suggested a slightly different relation between the
formation factor and porosity by introducing the Archie’s cementation factor, m as: F = φ with the
−m
cementation factor in the range [1.8-2.0] for consolidated sandstones and about 1.3 for clean
unconsolidated sands. The famous Archie’s equation gives the relationship of resistivity index with
water saturation of rocks through saturation exponent, n as:
Rt
I= = Sw− n
Ro with the saturation exponent in the range [1.4-2.2].
Rt and Ro can be obtained from well logging data while the saturation exponent n is experimentally
determined in laboratory. Therefore, the in situ water saturation can be calculated by using Archie’s
equation. Based on the material balance equation, Sw + So + Sg = 1.0, the hydrocarbon reserve in place
may then be calculated.
Machine Description:
The system includes:
♦ An atmospheric Electrical Core Holder
♦ An ambient Brine Resistivity cell
♦ A RFL meter (Fluke make)
Experiment operation
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 55
Reservoir Rock Properties Analysis, Mohsen Masihi
Then slide the platen until the sample touches the conductive pad at left side. Notice the locking pin
(Fig. 40). Topple the lid over the sample. Next slide the pin to lock the lid until the ball thrust is
engaged. If you cannot slide easily the pin, unscrew the knob by one or two turns and lock the lid.
Finally, tight the knob to ensure a good contact of the sample with the inner electrodes.
Figure 40: Illustration of system under locked pin and tighten the knob
Close the cover to prevent evaporation of the liquid from the sample. Rotate the valve in the
appropriate direction to actuate the piston jack (Fig. 41). This will cause the core sample to be firmly
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 56
Reservoir Rock Properties Analysis, Mohsen Masihi
held between the 2 pads. The piston is now held against the sample. Allow equilibrium conditions to be
reached and record the experiment temperature.
After the measurement do the sample unloading and retract the conductive pad by rotating the valve
command in the appropriate direction. Open the cover and the lid over the sample. If necessary, you
can help the pads to open out by sliding the piston by hand to the far right, and then remove the core
sample. Dry the pads electrode and wetted parts thoroughly of all brine with a cloth. Repeat theses
steps for each sample.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 57
Reservoir Rock Properties Analysis, Mohsen Masihi
Figure 42: Illustration of power supply cord and the temperature display
3- Rinsing of cell
The actual brine can be incompatible with previous brine and leads to contamination or chemical
precipitation in case some previous brine remained in the cell. Hence, you must rinse thoroughly the
cell and the syringe with distilled water and empty the tubing from liquid before performing calibration
or measurement. Set the flexible tubing into the calibration brine contained in a beaker. Open the valve
(handle vertical), drive the piston down to the lower position, then lift slowly the piston to draw some
brine into the cylinder (Fig. 43). If some bubbles are generated in the cylinder, empty and refill the
cylinder until getting clear liquid. When all the electrodes are immersed in clear liquid, close the valve
(handle horizontal).
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 58
Reservoir Rock Properties Analysis, Mohsen Masihi
4- Phase angle
In addition to resistance, the phase angle will be measured to check if there was a poor contact between
the core sample and the electrodes. Good contact between the core sample and the electrodes is usually
achieved for phase angle less than 2 degrees. However, abnormal phase angle can also indicate unusual
effects such as pyrite in the core. To fix elevated phase angle and ensure good contact between the core
sample and the electrodes: reload the core and then re-measure the core sample.
5- Frequency
In addition by varying the frequency, you can achieve the lower phase angle, i.e. the closer the circuit
is to a pure resistive component. The resistance value at the lowest phase angle should be
used as the actual measurement. Proceed by starting the frequency as low as possible and
increase the frequency until getting the lowest phase angle.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 59
Reservoir Rock Properties Analysis, Mohsen Masihi
Figure 45 shows a spherical cap which is subjected to interfacial tension σ around the base of the cap
and two normal pressures P1, and P2 at each point on the surface. The effect of the interfacial tension is
to reduce the size of the sphere unless it is opposed by a sufficiently great difference between
pressures, P1, and P2.
1 1
P2 − P1 = σ +
r1 r2
Where r1, and r2, are the principal radii of curvature. Introducing the mean radius of curvature defined
phase on the concave side of the surface must have pressure P2 which is greater than the pressure P1,
on the convex side. The surface tension of a liquid surface in contact with its own vapor or with air
is found to depend only o n the nature of the liquid, and on the temperature. Usually, surface
tensions decrease as temperature increases.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 60
Reservoir Rock Properties Analysis, Mohsen Masihi
Where θ is contact angle between liquid and glass tube and g is acceleration of gravity.
gρhr r∆P
Hence the value of σ calculated by, σ = = where ∆P is the hydrostatic pressure of
2 cos θ 2 cosθ
the column of liquid in the capillary.
F = WP + 2( x + y )σ
Where 2(x+y) is the contact area between the liquid and the plate, and F is the weight of the plate.
Figure 47: Wilhelmy plate methods: (a) dynamic, and (b) static method
In the static method the plate is held at the position shown in Fig. 47b and the equation will be
F = WP − b + 2( x + y )σ cosθ
Where b is buoyancy force of immersed part of the plate in the liquid and θ is contact angle.
This instrument can be calibrated such that the interfacial tension reads directly.
3-Ring Method
The ring (or Nouy) method of measuring surface and interfacial tension is commonly used and the
apparatus is called a ring tensiometer. To measure interfacial tension, a platinum ring is placed in the
test liquid. The force necessary to withdraw it front the liquid is determined (Fig. 48).
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 62
Reservoir Rock Properties Analysis, Mohsen Masihi
F = Wr − b + 2(2πrσ )
where F is measured force, r is radius of the ring at centre (the radius of the platinum thread is
negligible compared to r), Wr is weight of the ring in air and b is buoyancy force of the ring immersed
in the liquid. For interfacial measurements, the ring is placed in the interface and the force necessary to
break the interfacial film with the ring is determined.
The instrument can be regulated in such a way that the ring weight and buoyancy effect are taken care
of with a correction factor C given by,
F
σ =C
2(2πr )
The maximum amount of liquid W, which can hang from a capillary tube with radius r without falling
depends on the surface tension as
W = mg = 2πrσ
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 63
Reservoir Rock Properties Analysis, Mohsen Masihi
where m is the mass per drop. Observations of falling drops show that a considerable portion of the
drop (up to 40%) may remain attached to the capillary end. This effect will be compensated with a
correction factor f.
mg
σ= f
2πr
The correction factor f varies in the region of 0.5 to 1.0. The drop method can be used for the
determination of both gas-liquid and Iiquid-liquid interfacial tensions.
de is the equatorial diameter and ds is the diameter measured distance de, from the bottom of the chop.
The interfacial tension can be calculated by the following equation,
ρgd e2
σ=
H
where H is a shape determining variable. The relationship between the shape dependent quantity H and
the experimentally measured shape dependent quantity S is determined empirically. A set of 1/H versus
S values is obtained in form of tables (Table 9). The quantity of S is calculated after measuring de and
ds from shape of the pendant drop, and then 1/H can be determined from Table 9.
The pendant drop method is widely used and has good accuracy.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 64
Reservoir Rock Properties Analysis, Mohsen Masihi
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 65
Reservoir Rock Properties Analysis, Mohsen Masihi
6-Spinning Drop
In this method, a drop of a less dense fluid is injected into a container of the denser fluid, and the whole
system is rotated as shown in Fig. 51. In the resulting centrifugal field, the drop elongates along the
axis of rotation. The interfacial tension opposes the elongation because of the increase in area and a
configuration which minimizes system free energy is reached. The method is similar to that for the
pendant drop with the gravitational acceleration g replaced by the appropriate acceleration term for a
centrifugal field.
If the fluid densities are ρ a and ρ b , and the angular velocity ω of rotation are known, then interfacial
tension can be calculated from the measured drop profile. When drop length is much greater than the
radius rm the following approximate expression holds
( ρ A − ρ B )ω 2 rm3
σ =
4
The spinning drop device has been widely used in recent years to measure very low interfacial tensions.
Unlike the other methods, no contact between the fluid interface and a solid surface is required.
The flow sheet of this machine is illustrated in Fig. 52. The IFT 700LL includes basically:
• Two pressure generators (PG1 and PG2)
• Temperature indicator (TI)
• Pressure indicator (PI)
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 66
Reservoir Rock Properties Analysis, Mohsen Masihi
• Rupture disc and support (RD); the (safety) burst pressure is given on the plate attached to the
support. Actual burst pressure may vary by ± 5%.
• Two supply tanks for liquid (TL1, TL2).
• Capillary tubes (CT): dosage of liquid for pendant drop, type “CT-A” and another one for
standing bubble/drop, type “CT-B”, outer diameter 1/16” (1.59 mm), inner diameter 0.8 mm.
• Video system: 1 CCD camera
Gas-liquid systems:
For generating liquid drops in gas atmosphere, the view chamber needs to be filled with the desired gas
up to the designated pressure. Therefore, the gas supply needs to be connected to valve G.
• shut all valves.
• open all suitable valves of the gas supply line (not belonging to the IFT 700).
• open G shortly and shut right away (allowing a small amount of gas to enter).
• open D for venting the system and displacing the air inside the view chamber (Purging).
• shut valve D.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 67
Reservoir Rock Properties Analysis, Mohsen Masihi
• open G carefully. Listen for whistling noises indicating escaping gas. In case of leakage,
depressurize respective part of the equipment. Fasten connection. Eventually replace damaged
pieces.
Pressure increase must be taken into account when filling the chamber prior raising the temperature.
You can fill the capillary with liquid before pressurizing the chamber to ensure that no gas is trapped in
the capillary. For this purpose, follow the dedicated instructions for filling liquid (in a separate
chapter), then open valve B while valve C is closed. The pressure generator PG1 is operated clockwise
until liquid appears at the capillary tip inside the chamber. Then close valve B and proceed to fill the
chamber with gas. Note that previous filling of the capillary or liquid being present elsewhere in the
chamber may cause fog when gas enters the view cell.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 68
Reservoir Rock Properties Analysis, Mohsen Masihi
• operate pressure generator PG1 anti-clockwise to withdraw the piston until reaching the end of
stroke OUT. This load the cylinder with the test liquid.
• operate pressure generator clockwise until no bubbles rise anymore at the liquid supply tank.
• Eventually drive the piston IN and OUT until the tubing is completely filled with liquid and no
gas remains trapped. Finally, withdraw the piston to the end of stroke OUT to get the cylinder
completely filled with liquid.
• shut the valve A
• open the valve C
• operate the pressure generator clockwise until the liquid is discharged through valve C. Collect
the liquid with a small glass beaker or similar.
• shut valve C.
Liquid-liquid systems:
For measuring either liquid-liquid interfacial tension by the pendant-drop method or for measuring gas-
liquid interfacial tension by the standing-bubble method, the chamber must be filled with a second
liquid in the following named “liquid C”. The liquid forming the drop is called “liquid D”. Fill the view
chamber with liquid according to the following instruction:
• Shut all valves.
• Fill the tank TL2 with liquid C: open the valve E and operate the pressure generator PG2 anti-
clockwise until the piston reaches the end of stroke OUT.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 69
Reservoir Rock Properties Analysis, Mohsen Masihi
• Operate the pressure generator clockwise until no bubble rises anymore in the liquid supply
tank.
• Eventually drive the piston IN and OUT until the tubing is completely filled with liquid and no
gas remains trapped. Finally, withdraw the piston to the end of stroke OUT to get the cylinder
completely filled with liquid.
• Close the valve E.
• Open the valves F and D.
• Hold a container (glass beaker) at the outlet of the valve D.
• Operate PG2 clockwise to inject the liquid C into the view chamber until producing at valve D
outlet.
• Go on to operate PG2 clockwise, and follow liquid level in the view chamber.
• Watch for leakages inside the support.
• Close the valve D.
• Go on to operate the pressure generator PG2 clockwise until reaching the required pressure
controlled at the pressure indicator.
Note: If PG2 is easy going and no change in pressure is recorded; air may be trapped. Open the
valve D shortly for venting and restart at f.
If PG2 is increasingly hard-going but no change in pressure is recorded, look for closed valves,
or disconnected pressure indicators.
From the start window, general options can be set. For experiments using the high pressure equipment
all symbols “Check Opt.” and “Ctrl Options” should be deactivated (crossed out). This can be done
directly by clicking the mouse or by pressing the “Check/Ctrl” icon depending on the version released.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 70
Reservoir Rock Properties Analysis, Mohsen Masihi
Select the pendant drop mode in the Start-Window when working with captured bubbles. The
“Camera” window can be selected in "portrait" or "landscape" orientation. The drop icon “Exper.”
allows for selection pendant/sessile drop or standing bubble (see Fig. 53).
The “CALIB.” icon opens the “Device Parameters” window for checking calibration of the complete
camera system. This calibration is checked automatically when starting a measurement from the main
window. Notice that the aspect ratio should be closed to 1.
In the usual case the user carries out calibration by using the capillary (nozzle) diameter. For this
purpose, press the “PRGM” icon, choose “New/Edit” and select nozzle. Now, close the “Device
Parameters” window by clicking on “OK” and return to the start window. The position of the capillary
(nozzle) for calibrating is selected in the window opened by choosing the “FOCUS” icon.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 71
Reservoir Rock Properties Analysis, Mohsen Masihi
A new window called “Focus &Light” opens containing a life image of the actual drop at the capillary
tip. This window should be checked frequently for right adjustment of light, focus and nozzle
calibration. The graph “light intensity” shows the gray values of the image ranging between 0 and 255.
The focusing marker must be positioned at the drop surface. Then, the graph shows the transition from
outside (high value=light) to inside the drop (low value = dark). By adjusting the values given in the
boxes called “white” and “black” the graph must be adjusted in order to show values of 200 – 250
outside the drop (white) and 0 – 50 inside the drop once the adjustments are completed, press “OK” to
return to the start window.
Now, you can press the “START” button to display the main measurement window.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 72
Reservoir Rock Properties Analysis, Mohsen Masihi
Click the “New or Edit” icon to open the “Time Line” window:
The time program must be set. For example T = 0.00 sec …. dT=1.0, which means that starting at 0
seconds a measurement is carried out every second. With the high pressure equipment “CONTROL
Vol/Area” is not applicable. The “CALIBR.” cursor must be switched toward “NOZ.”
(nozzle/capillary).
Settings can be saved or loaded as ~.prg files. By choosing “Done” you can return to the parameter
window. After having chosen a file name to save the image data (~.prf), the parameter window is
closed and the main window appears once you click “OK”. Now, you can start the measurement by
selecting “START”. The calibration data is used during the interfacial tension determination. (Fig. 58)
The diagrams in the lower part of this window show the interfacial tension (as a rough estimate) and
the drop area in real time. These data are given for information only and are not saved.
A small picture at the right side of the life drop image named “Drop Dim.” Indicates whether the drop
has a suitable dimension (green) or is too small (red) for determination of the interfacial tension.
The symbol below named “Ctrl Loop” is not applicable with the high pressure equipment and should
be switched off.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 74
Reservoir Rock Properties Analysis, Mohsen Masihi
The data acquisition in the “Main Measurement” window is stopped by pressing the “ABORT” button.
Afterwards, press “EXIT”. In case “Fitting Auto-Start” was switched on in the “Experimental
Parameter” window, the fitting procedure starts automatically. Otherwise, press “FITTING” in the
start window.
Press “LOAD” in the appearing “Fitting” window and select the ~.prf file containing the data of an
experiment performed earlier. All the available measurement points are computed to match the Laplace
equation, and then to determine the interfacial tension as a function of time. Data of the interfacial
tension, the drop volume and area etc. are saved in a ~.fit file to the hard disk. This file can be imported
to an excel sheet.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 75
Reservoir Rock Properties Analysis, Mohsen Masihi
ΛρgDe2
σ= where g=981 cm/s2, and De (real size of de) = de (1.1/ d IMG), mm
H
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 76
Reservoir Rock Properties Analysis, Mohsen Masihi
6-Capillary pressure
The coexistence of two or more immiscible fluids within the voids of a porous medium, such as a
reservoir rock, gives rise to capillary forces. As interfacial tension exists on the boundary between two
immiscible fluids in a pore space, the interface is curved and there is a pressure difference across the
interface. The pressure difference is termed the capillary pressure and can be expressed as:
PC = Po - Pw
where:
Po is the oil pressure (non-wetting phase)
Pw is the brine pressure (wetting phase)
Capillary pressure as a function of surface tension and radii of curvature r (for spherical cap) is,
2σ wo cos θ
Pc = ( ρ w − ρo ) gh =
r
Note that the free water level, FWL is the point where Pc=0. Oil-water contact OWC above FWL has
Sw=100% due to the capillary action. The zone of varying water saturation with height above OWC is
called the transition zone (Fig. 61).
The size of the transition zone depends on pore size distribution (PSD), grain sorting, interfacial
tension, rock characteristics (porosity/permeability) and also difference in fluid density. The saturation
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 77
Reservoir Rock Properties Analysis, Mohsen Masihi
history of the rock can also affect the capillary pressure. Two processes usually considered are
drainage which is the replacement of the wetting phase (e.g. water) with non wet phase (saturation of
wet phase decreases). In this case, the saturation level is dictated by the capillary pressure associated
with the narrow pores and it is able to maintain water saturation in the large pore below. The second is
imbibition where wetting phase saturation increases by the expulsion of the non wet phase. The level
of saturation is determined by the large pore reducing the capillary pressure effect and preventing
water entering the larger pore. Figure 62 shows different process of this kind in a water wet system
where A, B, C shows respectively the primary, negative secondary and positive secondary drainage and
D and E shows positive and negative imbibition respectively.
Figure 62: Typical capillary pressures for in various stages of drainage and imbibition processes
Capillary pressure measurements are essential for the complete characterization of a hydrocarbon
reservoir. A plot of capillary pressure versus fluid saturation for a core plug can be used:
• to calculate reserves
• in reservoir simulation computer routines
• provides data on the irreducible water saturation of a reservoir rock
• shows the entry pressure of fluid into a water saturated reservoir or cap rock.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 78
Reservoir Rock Properties Analysis, Mohsen Masihi
conditions. Knowledge of the water-oil capillary pressure and resistivity index vs. saturation
relationship is necessary for many reservoir-engineering tasks:
(1) assess connate water saturation to calculate oil in place;
(2) calibrate resistivity logs;
(3) determine the height of the transition zone;
(4) model oil displacement either by free water imbibition and /or water injection.
The equipment measures the resistivity of core plugs at different brine saturations by a precise
resistivity meter having 4 electrode measurement system. The core plug is placed in a core holder
subjected to confining pressure and pore pressure equivalent to reservoir conditions. Temperature of
the core plug is raised to the reservoir condition. The brine saturation is changed from 100 % to lower
values by gradually increasing the pressure of an upstream metering pump which injects oil into the
core plug and displace the brine. The volume of the displaced brine is accurately determined by a
downstream metering pump which accumulates the displaced brine at a constant pressure. Both the
pumps are attached to the core holder through semi porous membrane saturated with respective fluid.
The resistivity of the pure brine used for the saturation of the core plug is determined with the help of a
Rw cell subjected to the reservoir pressure and temperature condition. The resistivity measurement of
a core plug at 100% brine allows for the evaluation of the Formation Factor (FF) whereas the
resistivity of a core sample at different saturations allows for the determination of the Resistivity Index
(RI) at simulated reservoir pressure and temperature conditions. The equipment has also the provision
for estimation of Resistivity of brine (Rw) at simulated reservoir temperature and pressure condition.
The user can run tests either in manual mode or in sequenced mode. In automatic mode, the operator
fills a table of pressure increments. Test data is graphically displayed on the screen and is logged on the
hard disk as a table file. Printouts of the input and test data are done using a printer.
Machine Description
Sample Diameter 1” or 1” ½ or 30 mm (according to customer’s selection)
Sample Length from 2" to 3"
Sample: Square edge, Parallel faces and consolidated
Pressure:
Overburden: up to 10,000 psi
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 79
Reservoir Rock Properties Analysis, Mohsen Masihi
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 80
Reservoir Rock Properties Analysis, Mohsen Masihi
Experiment operation:
During the test, the temperature of the core plug is raised to the reservoir condition. The brine
saturation is changed from 100 % to lower values by gradually increasing the pressure of an upstream
metering pump which injects oil into the core plug and displaces the brine. The volume of the displaced
brine is accurately determined by a downstream metering pump which accumulates the displaced brine
at a constant pressure. Both the pumps are attached to the core holder through semi porous membrane
saturated with respective fluid. For this the following is done.
After setting up the core holder, the first step is to fill the pumps. This can be done through the panel
available in the software. Note that the pumps should be calibrated so that the volumes shown on the
panel is the true one. At the next step it is time to give the core holder some confining pressure. An
starting pressure on 500 psi should be sufficient. It is highly recommended that before any process the
air probably contained within the core holder and the pumps are removed. This should be done for the
three pressures. The two filled pumps then should be started one at each time with almost the same
pressure in order to make sure the core is intimately saturated with the fluids. The pressure of oil pump
is a little less then the brine pressure so that no flow into core occurs.
When all the steps are performed correctly it is time to start the test. The brine pump is always on and
the run mode is pressure. In this case the brine pressure is set to some value. The set mode on oil pump
is delta pressure and the test begins by setting a small delta pressure. It is good to check to see if the
delta pressure shown by the software is calculated correctly. The first steps of changing the delta
pressure should be done by care cause it might affect the accuracy of the result curve. The steps are
changed if the resistivity has become constant or if the displaced volumes of oil and brine pumps are
showing the same value. At the end of the experiment decrease the pore pressure by setting the water
pump in constant pressure with 15 psi set value with a pressure ramp of -50 psi/min and the oil pump in
constant delta pressure at zero psi set value. As the pore pressure decreases, the confining pressure by
opening slowly the isolating confining valve. Keep a difference of 300 psi between the confining
pressure and the pore pressure and decrease the temperature to room condition.
The process body is schematically composed of 5 stages. The end of a step is determined from
stabilisation of electrical measurement and volume in the pumps. Water pump is pressure controlled at
pore pressure set point. Oil pump is DeltaP controlled at capillary pressure.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 81
Reservoir Rock Properties Analysis, Mohsen Masihi
1. Starting from a sample 100% saturated with brine, Primary drainage is applied by increasing
DeltaP in small steps. DeltaP is going positive. The maximum final DP can be 150 psi. Practically,
the drainage is completed when no more water production happens for an increase in DeltaP. Each
step can need from a few hours up to several days to reach equilibrium. When reaching equilibrium,
DP and pumps volume are recorded. From the reading of the pumps volume, the balance with the
sample content ("saturation" in oil and water) is determined. The DeltaP is the capillary pressure. A
graph can be plot of DeltaP vs water saturation.
2. Spontaneous imbibition is then conducted by reducing DeltaP in small steps, down to DeltaP
equal zero. There is an hysterisis, and the curve of DeltaP vs water saturation is different from the
curve obtained before. The saturation curve does not cross the pressure axis at for 100% water.
3. Going further, we run the Forced imbibition by reducing DeltaP in small steps, getting negative
values of DeltaP. The maximum final DeltaP can be -150 psi (theoretical value). Practically, the
imbibition is completed when no more water enters the sample for a change in DeltaP.
4. In the next step, we increase DeltaP by steps during Spontaneous drainage, until DeltaP is back
again to zero. There is an hysterisis, and the curve of DeltaP vs water saturation is different from
the curve obtained at previous stage.
5. Going on to increase DeltaP in positive value, we run the Secondary drainage. DeltaP is build up to
the previous maximum value during Primary drainage.
The process schematic to get whole pc curve is shown in Fig. 64.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 82
Reservoir Rock Properties Analysis, Mohsen Masihi
Figure 64: Schematic illustration of various stages of the process of getting whole pc curve
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 83
Reservoir Rock Properties Analysis, Mohsen Masihi
Figure 65: Summarized report of the formation factor results obtained from CAPRI
The type of report of Pc, Ir and n with core partially or totally desaturated is shown in Fig. 66.
Figure 66: A typical summarized report of Pc, Ir and n of a core partially or totally desaturated
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 84
Reservoir Rock Properties Analysis, Mohsen Masihi
The equation used to calculate the parameters are shown in Fig. 67.
Figure 67: Illustration of equations used to calculate the parameters in CAPRI instrument
PC curve
5000
4000
3000
Capillary Pressure (Po - Pw), mbar
2000
1000
0
0 10 20 30 40 50 60 70 80 90 100
-1000
-2000
-3000
-4000
-5000
Brine Saturation (Sw), %PV
Figure 68: Illustration of a typical capillary pressure curve obtained from CPAR instrument
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 85
Reservoir Rock Properties Analysis, Mohsen Masihi
7-Permeability
Permeability is a property of a porous medium which shows the ability of porous media to
transmit fluids. The reciprocal of permeability represents the viscous resistivity. The effective
permeability of a porous medium is a measure of the rock conductivity to a particular phase of a multi-
phase fluid system residing within the porous medium, where the saturation of each phase is specified.
Relative permeability is the ratio of the effective permeability of a particular fluid phase to some
arbitrary reference permeability (i.e. absolute permeability). Permeability has the unit of m2 in SI
system or Darcy in field unit with a conversion factor of 1D ≡ 0.986923 × 10 m . Note that a rock
-12 2
sample has a permeability of one meter squared when it permits 1 m3/s of fluid of 1 Pa.s viscosity
through an area of 1 m2 under a pressure gradient of 1 Pa/m. Permeability is calculated using the
following equation:
Where:
k = permeability (Darcy)
q = flow rate (cm3/S)
= viscosity (mPa.s)
x = length (cm)
A = cross sectional area (cm2)
p = differential pressure across flow section (atm).
One of the quick methods to measure permeability is the “transient method”, which is used in KeyPhi
device. Transient measurements employ fixed-volume reservoirs for the gas. These may be located
upstream of the sample from which the gas flows into the sample being measured. The pressure falloff
apparatus (Fig. 69) employs an upstream gas manifold that is attached to a sample holder capable of
applying hydrostatic stresses to a cylindrical plug of diameter D and length L. An upstream gas
reservoir of calibrated volume can be connected to the calibrated manifold volume by means of a valve.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 86
Reservoir Rock Properties Analysis, Mohsen Masihi
Multiple reservoir volumes are used to accommodate a wide range of permeabilities. The
downstream end of the sample is vented to atmospheric pressure. An accurate pressure transducer is
connected to the manifold immediately upstream of the sample holder. The reservoir, manifold and
sample are filled with gas. After a few seconds for thermal equilibrium, the outlet valve is opened to
initiate the pressure transient. When the upstream pressure has decayed to about 85% of the fill
pressure, data collection is started. Pressures and times are recorded. This technique has a useful
permeability range of 0.001 to 20,000 md.
The use of liquids for permeability measurements eliminates the problem of gas slippage, and at
reasonable and usual flow rates, inertial resistance is generally negligible. Thus, Darcy’s Law can be
used directly to calculate permeability from a single flow rate measurement. However, potential
permeability alteration from interaction of rock constituents and liquids (especially aqueous solutions),
fines movement, and microbial plugging requires special attention. Also, the liquid remaining in a
sample may have to be removed before other measurements can be performed. Because of these
problems, most routine permeability measurements have been made using gases. However, for some
samples, such as those sensitive to drying techniques, liquid permeability measurements are considered
to be the only acceptable alternative. BRP-350 is a device which can be used for absolute and relative
permeability measurement using both gas and liquids. The description of this instrument is given in
section 7.2.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 87
Reservoir Rock Properties Analysis, Mohsen Masihi
Klinkenberg effect:
Klinkenberg (1941) has reported that there are variations in the permeability determined by using gases
as the flowing fluid compared to those obtained when using liquids. This has the following relation,
kg
kL =
b
1 +
Pm
where Pm is the mean pressure, kg is the gas permeability and kL is the equivalent liquid permeability.
These variations were considered to be due to slippage, a phenomenon well known with respect to gas
flow in capillary tubes. The phenomenon of gas slippage occurs when the diameter of the capillary
openings approach the mean free path of the gas. Note that the mean free path of a gas is a function of
molecular size and the kinetic energy of the gas. Therefore, permeability of gas depends on factors
which influence the mean free path, such as temperature, pressure and the molecular size of the gas.
Figure 70 is a plot of the permeability of a porous medium as determined at various mean pressures
using three different gases. Note that for each gas a straight line is obtained for the observed
permeability as a function of the reciprocal of the mean pressure of the test. All the lines when
extrapolated to infinite mean pressure (1/Pm = 0) intercept to the same point kL.
Figure 70: Variation in gas permeability with mean pressure and type of gas
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 88
Reservoir Rock Properties Analysis, Mohsen Masihi
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 89
Reservoir Rock Properties Analysis, Mohsen Masihi
Working Procedure
The Applilab software has been developed by the programmers from VINCI in order to control the
process parameters simultaneously from one central unit. The graphical interface of Applilab is shown
below.
D
E
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 90
Reservoir Rock Properties Analysis, Mohsen Masihi
1. Check that the caliber is mounted on the core holder. Saturate the tubing from the oil
accumulator to the core holder with oil. Then inject brine to saturate all the lines until the
burette with brine. Refer to the above picture to open/close the burette valves.
2. Inject oil (drainage) and simultaneously open the top burette valve and close the bottom burette
valve. When only oil is produced in the burette, measure the Volume of water produced. The
first drainage dead volume is equal to this volume minus the caliber void volume, i.e. 0.1 cc.
3. Inject brine (imbibition) and simultaneously close the top burette valve and open the bottom
burette valve. When only brine is produced in the burette, measure the Volume of oil produced.
The first imbibition dead volume is equal to this volume minus the caliber void volume, i.e. 0.1
cc.
4. Repeat the same operations until getting the same volumes (typically after 3 or 4 cycles). Dead
volume measurements for gas-liquid experiments are similar except that gas is injected instead
of brine.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 91
Reservoir Rock Properties Analysis, Mohsen Masihi
permeability, or oil in case of gas-liquid permeability. The saturation of the core can be done with a
saturator or with a vacuum bell connected to a vacuum pump. Then write down the wetted weight.)
2. Load the sample in the core-holder
3. Initiate the confining pressure
Close the hand pump release valve. Open the confining valve HV06 and close the buffer valve HV08.
Operate the handle to increase the pressure up to around 700 psi. If needed, you can decrease the
pressure by opening the hand pump release valve. Close the confining valve HV06. Wait for stability
and adjust if necessary.
4. Initiate the back pressure
Set the nitrogen pressure at your facility to around 250 psi, open back pressure valve HV07 and close
buffer valve HV08. Wait until back pressure reaches roughly 200 psi. If needed, you can decrease the
pressure by opening buffer valve HV08 and opening the hand pump release valve. When the set point
pressure is reached, close back pressure valve HV07 and close buffer valve HV08.
5. Purge the lines
Before performing a test, you have to ensure that the inlet and outlet lines are correctly saturated and
that no gas is remaining in these lines. For the following explanations, refer to the below picture
concerning the mentioned numbers.
6. Pressure Control
Open process valve HV01, open brine valve HV03 and close oil valve HV02. Start injection pump and
control that outlet pressure increases up to back pressure. Wait until inlet and outlet pressure are
stabilized.
At any time you must control the following relations true:
Back P < Outlet P < Inlet P < Confining P < 5000 psi
In the case Inlet P > Confining P, you can spoil the core by getting invasion of confining oil into the
core sample. Typically, Back P = 200 psi and Confining P = 700 psi.
I. Monophasic Permeability
Once the different steps to prepare a test are realized, you can start the monophasic permeability
measurement. Refer to Darcy’s law for permeability calculation. Darcy’s law can be applied under the
following assumptions:
Ø The core plug is 100% saturated with the flowing fluid.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 92
Reservoir Rock Properties Analysis, Mohsen Masihi
Ø The flow is horizontal, steady state and under the laminar regime.
Ø The flow of fluid through the porous medium takes place under viscous regime (i.e. the rate of
flow is sufficiently low so that it is directly proportional to the pressure differential or the
hydraulic gradient).
Ø The flowing fluid does not react with the porous medium because it may alter the characteristics
of the porous medium thereby changing its permeability as flow continues.
Once the system is stabilized you can run brine injections at different flow rates. For each flow rate,
wait stabilization of the ΔP and complete the excel file “Permeability-template” provided with the
Applilab software as in the below picture.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 93
Reservoir Rock Properties Analysis, Mohsen Masihi
Ø No capillary pressure
Ø No gravity
Ø Turn the ΔP valve HV05 on bypass position. Then offset the ΔP thanks to the Applilab
software. Make sure that the actual Delta P is within the transmitter range (0-500 psi), then turn
the ΔP valve HV05 on measure position.
Ø Start data logging. You can adjust the time interval to your needs.
Reset the camera volume thanks to the Applilab software. Run the injection pump at medium flow rate
(example: 0.5 cc/min) and simultaneously open oil valve HV02 and close brine valve HV03. Refer to
the below picture concerning the position of the burette valves. Wait until ΔP and Vw stabilization.
Then complete the excel file “Permeability-template” provided with the Applilab software, especially
the “Drainage” part. From the pump flow rate, ΔP and Vw, the excel file will calculate automatically
the initial water saturation (Swi) and the permeability of oil at Swi, i.e. Ko(Swi).
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 94
Reservoir Rock Properties Analysis, Mohsen Masihi
When the drainage is over, i.e. the ΔP and Vw are stabilized, you can prepare the imbibition. This is the
most important phase of the test. Before injecting brine, check the following points:
Ø Decrease the flow rate of the pump (example: 0.1 cc/min).
On Applilab trend curves, you can control the evolution of the ΔP and oil production. The typical shape
of these curves is shown in the below picture.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 95
Reservoir Rock Properties Analysis, Mohsen Masihi
Stabilization usually occurs after 5 pore volume injections. Now, we can proceed to the treatment of
data to calculate the relative permeability curves using CYDAR software.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 96
Reservoir Rock Properties Analysis, Mohsen Masihi
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 97
Reservoir Rock Properties Analysis, Mohsen Masihi
Where:
VP: Velocity of P wave (in m/s)
VS: Velocity of S wave (in m/s)
ρ: Density (in kg/m3)
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 98
Reservoir Rock Properties Analysis, Mohsen Masihi
ν: Poison’s ratio
E: Young’s modulus (in Pa)
K: Bulk modulus (in Pa)
μ: Shear modulus (in Pa)
λ: Lame’s constant (in Pa)
K: Compressibility (in Pa−1)
Al: Acoustic impedance (in kg·m−2·s−1)
Operational Parts
1. Core Holder
AVS is provided with two core holders one to accommodate plugs of 1’’ diameter and one to
accommodate plugs of 1.5’’ diameter. The core holders are tri-axial and can accommodate cores of 1 to
3’’ length. It is made of stainless steel. Radial pressure, Axial pressure and Pore pressure can be applied
independently. The design of the vessel is so that when Radial and Axial pressures are identical the
core holder is hydrostatic.
The different elements of the vessel are shown on the pictures below. The main features of the device
are as follows:
o Core diameter: 1 & 1.5”
o Core length: 1 to 3”
o Material: stainless steel
o Pore pressure: up to 10,000 psi
o Confining pressure: up to 10,000 psi
o Radial pressure: up to 10,000 psi
The scheme below shows the 1.5’’ core holder.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 99
Reservoir Rock Properties Analysis, Mohsen Masihi
2. Accumulator
AVS includes a 100 cc pore fluid accumulator. This accumulator is connected to the hand pump and
permits to apply a pore pressure up to 10,000 psi. The accumulator is located on the rear side of the
panel. The process fluid is on the top side of the accumulator.
The accumulator is isolated on the pump side by the valve V5 and on the process side by the valve V1.
3. Electrical Connections
The connections between the acoustic core holder, the switch box, the impulser and the oscilloscope
are shown below. The Oscilloscope is connected to the impulser via 2 BNC female / female cables. The
impulser is connected to the Switch box via 2 BNC female / female cables. The switch box is
connected to the core holder via 6 SMA male to BNC female cables.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 100
Reservoir Rock Properties Analysis, Mohsen Masihi
Measurement
1. Overburden Set-up
1.1. Building up step by step procedure
o Check the connection from the panel ports RADIAL and AXIAL to the core holder.
o Open Valve 3 “RADIAL INLET” and Valve 4 “AXIAL INLET".
o Check the oil level in the manual pump. Refill if necessary with specific oil (refer to specific
manual).
o Check the built-in valve of the manual pump: it must be closed.
o Operate the pump to build up the pressure in both radial and axial direction.
o Check the pressure on the 2 dedicated displays: upper central for RADIAL, upper right for
AXIAL.
o Optionally, close Valve 3 RADIAL INLET and operate the pump to build-up more axial
pressure.
o Once you have reached the axial and radial pressure, close Valve 3 and Valve 4 to isolate the
pump.
o When building up the process fluid pressure, it may interact with the axial and radial pressure.
Adjust them if necessary.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 101
Reservoir Rock Properties Analysis, Mohsen Masihi
2. Transducers’ Orientation
Whatever the relative positions of the transducers are, the signal of the P-waves displayed on the
oscilloscope should remain constant. The intensity of the signal of S waves should be maximum when
S1 transducers are parallel. If this is not the case, it means that the core is anisotropic (refer to the
“Theory” chapter). To adjust the relative positions of the transducers, the operator may rotate them
thanks to the tube guides H02-213 (long) and H02-224 (short).
The correct position is determined by a constant checking of the oscilloscope signal.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 102
Reservoir Rock Properties Analysis, Mohsen Masihi
3.1. P Waves
Using an oscilloscope, calculating the time of flight of P-waves is very easy. It is possible to trig the
signal of the impulse. The time of flight is the difference between the pulse and the first detected signal
following the pulse. On the following screen copies, the time of flight of P-waves is of about 28.6 μs.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 103
Reservoir Rock Properties Analysis, Mohsen Masihi
4. Calibration
The instrument is provided with some calibration plugs: Two aluminum plugs of 1.5’’ diameter (1 and
3’’ of length respectively) and two nylon plugs of 1.5’’ diameter (1 and 3’’ of length respectively).
Those plugs can be used to calibrate the AVS system and to get the propagation
speed of S and P waves via the time of flight of those waves. The table below gives the propagation
speed of P and S waves in aluminum and nylon at ambient temperature.
The propagation speed of P and S waves can be calculated via the following equations:
With:
VPwave & VSwave : Propagation speed of P/S waves (m/s)
TPwave & TSwave : Time of flight of P/S waves (s)
CFPwave & CFSwave : Correction factor of P/S waves (s)
The correction factors correspond to the time of flight of the waves without any plug. They can be
directly measured by pressing the two spreaders one against each other. To get an accurate
measurement of those correction factors it is possible to measure the time of flight of the waves in a
calibration plug. Thus the correction factor can be calculated using the equations below:
The calibration factors of the AVS system should be of about 10 μs for P waves and 16 μs for S waves.
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 104
Reservoir Rock Properties Analysis, Mohsen Masihi
Department of Chemical & Petroleum Engineering, Sharif University of Technology, Tehran, IRAN 105