Professional Documents
Culture Documents
Abstract
26864 - IJAER
11152 M. H. Jali et al
I. INTRODUCTION
Human support system is known as an endoskeleton. Endoskeleton plays a role as a
framework of the body which is bone. Our daily movements are fully depends on the
functionality of our complex systems in the body. The disability one or more of the
systems in our body will reduce our physical movements. Exoskeleton device is
known as rehabilitation device to facilitate disable person. The functionality of the
rehabilitation device has to smooth as the physical movement of normal human.
People who have temporary physical disability have the chances to recover. The
rehabilitation programs provide the suitable plan for conducting the nerve and
stimulate the muscles. Nowadays, rehabilitation program are using rehabilitation
devices in their tasks. The functionality of devices depends on muscle contraction.
Electromyogram studies help to facilitate the effectiveness of the rehabilitation device.
The technique of measuring electrical activity that produced by muscles during
rest or contractions known as electromyography (EMG). The electrical signal
generates from the brain and sends to the muscles via motor neuron. The EMG could
detect the dysfunctional of the muscles or failure in signals transmission from nerves
to muscles. The failure of sending the electrical signal from the brain to the
conducting nerves requires electrical stimulation from the external source to muscles.
Electrodes are used for signal detection of electrical activity in muscles. The study of
this electrical activity is important for combination of electromyogram into the
rehabilitation device.
The rehabilitation device is a tool that used to help the movements for daily life
activities of the patients who suffer from the failure of muscle contractions, due to the
failure of the muscles contractions the movements is limited. The ability of the
patients to do the tasks in the rehabilitation programs need to be measured. The
rehabilitation programs have to assure whether the tasks will cause effective or bring
harm to the patients[2].
Historically, the rehabilitation tasks have been avoided due to a belief that it
would increase spasticity [1]. In this research, the analysis of the data will be focusing
on upper limb muscles contraction consisting of biceps muscles only. The experiment
is limited to the certain of upper limb movements that use in training. EMG is a
division of bio signal; the bio signal analysis is the most complex analysis. Thus, the
signal analysis is a complicated process that has to be through many phases of
analysis [3].Therefore, the challenge of this study to assure that the signal processing
is conducted properly to overcome the environment noise during data collection [11].
skin layer. The EMG noise signal can be reduced by choosing the gelled electrode
which were Silver –Silver Chloride(Ag-AgCl) as the substance. Ag-AgCl introduces
less electrical noise into the measurements [13]. However, these noise should be
eliminate to have a better signal for analysing purpose, the noise can be eliminate by
many factors such as control environment, electrode placements and signal
acquisition circuitry and configuration.[1][2] As for the features extraction there were
various types in terms of frequency domain, time frequency domain and time scale
domain. These types are chose based on the purpose of the analysis of the signal. In
signal conditioning for time domain analysis, there is some features that can be
applied. These features are described briefly as follows:
i) Root-Mean-Square (RMS)
RMS is the square root of the mean over time of the square of the vertical distance of
the graph from the rest state. It much related to the constant force and non-fatiguing
contraction of the muscle. In most cases, it is similar to standard deviation method.
1
= ()
1
= ( − ̅)
−1
by placing the electrodes at the belly of the muscles not at or near the tendon and
motor unit. If the electrodes is bipolar configuration(Fig 1.), which have three
detecting surfaces, both electrode need to be placed nearly to each other within 1-2cm
[4]. These two detecting surfaces are connected to differential amplifier while another
one electrode is placed at the reference(bone).
The aim for this study to have the EMG signal for analysis purpose by reduce all
the unwanted noise from the electrode, hardware and the environment to maximize
the quality of the EMG signal. There have a lot of noise factor that can be categorized
as follows;
i) Causative Factors: This is the direct affected on signals.
a. Extrinsic – The signal get affected by the electrode placement, structure, surface
detection, distance between electrode(bipolar configuration), and location of the
surface detection on the skin.
b. Intrinsic – It may due to physiological, biochemical and anatomical factors.
ii) Intermediate Factors: This is the phenomena of influencing by one or more
causative factors.
iii) Deterministic Factors: This is influenced by the intermediate factors.
By reducing all these factor of noises, the quality of EMG signal will be enlarge
and the analysis of the information will be less of error and easy to obtain. However,
the precautions of handling the hardware and the electrode are needed. Acquisition
data play the important role to maximizing the quality of the EMG signal, such as
minimizing the distortion in EMG signal, using any filtering tools are not
recommended. In terms of signal-to-ratio(SNR) the information that carried in SNR
should contain the maximum information of EMG signal[3][10]. Moreover, the
quality of EMG signal is affected by the environment,has to set by minimizing the
noise factor, this method also known as control environment. The ambient noises
sourced by electromagnetic device such as radio transmission, fluorescent lights and
power line interference from electrical wires[10][4]. These ambient noise and motion
artifact can be reduced by the proper electrode placements, circuit configuration and
control environment. The signal recorded in Fig 2. shows the appearance of electrical
activity in muscles cannot be differentiate by any muscles contraction, because of not
EMG Signal Features Extraction of Different Arm Movement 11155
following the criteria to reduce the noises and the electrode placements detection
surface range.
-20
-40
amplitude response
-60
-80
-100
-120
-140
0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 2
t(sec)
Fig.2.The effects of placement the bipolar electrodes not within 1-2cm and
without environment control (ambient noise).
towards magnetic devices, the neatness of the hardware is needed. The neatness
ensured the electrode leads and the detecting surface intersects most of the same
muscle on subject as in Fig 6(b) at biceps, and as a result, an improved superimposed
signal is observed. Reference electrode has to be at the body where no muscles exist
as the ground signal, for this experiment it placed at wrist on the unaffected
hand(refer Fig 6(b)).
(a) (b)
Fig. 5.The Alcohol Swab(a) and the Gelled Electrode(b) That Used in This
Experiment.
Biceps
Biceps
(a) (b)
Fig.6. The Biceps Brachii Muscles For Electrode Positions(a), The Electrode
Placements On Subject Skin(b)
11158 M. H. Jali et al
Gelled electrode has two leads detecting surfaces and one lead as reference
electrode. These two detecting surface need to be placed in range of less than 1-2 cm
from each of it at the biceps muscle as in Fig 6(b). The gelled electrode for the blue
lead and red lead are placed at the biceps skin for detection purpose, meanwhile the
gelled electrode with black lead is placed at area with no muscles exist and unaffected
hand.[1] The grounded electrode hand is placed at the leg to have a better grounding
purpose. From Figure 8 these EMG signals will go through into high voltage
protection to protect any electrical surge that may bring harm to the user and the
hardware. These signals are available for signal conditioning process as the next step
to be taken. The circuit that be used in this experiment is provided by the OLIMEX
shield. The signals via the OLIMEX shield through the analogue circuit are very weak
the voltage is around 10μV, full of noise and contain primarily 60Hz of frequency.
The signals are next need to be filter and rectify, reducing the noise and amplifying
the signals for analysis purpose. After the signals is filtered and amplified the signals
is sent to the digital board and yet ready to display into the MATLAB software.
ZCS and VAR formed the whole and 10-6 scientific numbers that would not
achievable to be class into the classifier because of the distance value of features
itself. Data from Table 1 is plotted into scatter graph in Fig. 8. Sample 1 up to sample
4 be owned by movement of 60˚, sample 5 to sample 8 is belong to movement of 90˚
and sample 9 to sample 12 is fit in movement 120˚. The green, yellow and red outline
colour in Fig. 8shows that the scatter features is obviously been clustered by its angle
of movements’ class. Even though the clustered movements is scattered in its class,
the objective of this study is achievable. The features of movements are belongs in
their movement group. The learning of the features is continued in data classification
[15][16].The time-domain features for STD, MAV and STD shows in Fig. 8 useful for
future works such as classification of the features. The classification of time-domain
features provide reliable information that might be use with combination these three
features for research in many type of classification technique.
0.004
AMPLITUDE (V)
0.003
STD
0.002 MAV
RMS
0.001
0
0 5 10 15
SAMPLES NUMBER
VI. CONCLUSION
In this study has further discussed on the noise and the environment that affects the
EMG signal during data collection. The surface electrodes contain lot of noise
compared to needle insertion procedure, noise from the skin itself and the
environment have been discussed in this study for better understanding to get a better
quality of EMG signals.
The objectives of this study is to extract the features have successfully obtained
from the experiment are MAV, STD, RMS, ZCS and VAR. On the other hand, not all
the features can be classified into the class of several degrees of movements. The
suitable features for classification are MAV, STD and RMS. The features are obtained
by signal processing phase. The features of time-domain are extracted by using
programming software use in MATLAB software. MAV, STD and RMS verified in
scatter graph that it can be use for several movements of classification in future works.
The uses of these features might be separately or combination among the all three.
ACKNOWLEDGMENT
The authors would like to thanks UniversitiTeknikal Malaysia Melaka (UTeM) and
Ministry of Education, Malaysia for the financial supports given through Research
Grant.
REFERENCES