1.0 OBJECTIVE
To lay down a procedure for operation and calibration of melting point apparatus.
2.0 SCOPE
This SOP Covers the operation and calibration procedure of melting point apparatus
and is applicable to Quality Control Department.
3.0 RESPONSIBILITY
Officer / Executive.
4.0 ACCOUNTABILITY
Department Head
5.0 PROCEDURE
5.1.1 Ensure that the power supply to the instrument is switched ‘ OFF’ before
cleaning
5.1.2 De-dust the instrument daily externally with a clean dry cloth.
5.1.3 Once in a week remove adhered dust by wet mopping using detergent solution.
Afterwards wipe the surface with a clean dry cloth to remove traces of detergent.
5.1.4 Precaution has to be taken to clean the instrument immediately with dry cloth to
remove the moisture.
5.2.2 Switch on the Main power supply. Switch on the ‘MAINS’ of the instrument.
5.2.3 The display will show some random figure, then press the ‘RESET’ push button
to display the actual temperature of the furnace.
5.2.4 Keep the melting point tube in silicone oil containing glass container. . Check
through the lense type mirror about the capillary is correctly seated in its cavity.
5.2.5 Turn the STIRRER knob slowly towards MAX. Press the START push button.
5.2.6 Turn the heater knob to MAX and put HEATER HIGH/LOW switch to HIGH. (
This switch is provided on right hand side of the instrument ). The raise of
temperature can be controlled by using this switch and the HEATER knob.
5.2.7 When the temperature reaches within about 30 o C, of the expected melting
point, reduce the rate of heating and adjust the rate of rise of temperature to about 1 o
C, per minute.
5.2.8 Wait till the melting point is reached which is indicated by sample is melted in
the tube. The digital read-out will display the exact temperature at which the sample is
melted. Record this temperature. Turn off the ’HEATER’ Knob.
5.3 PRECAUTIONS
5.3.1 Care should be taken while inserting the capillary into its cavity or taking out
from the cavity to avoid breaking, take it out vertically.
5.3.2 Check always colour of the silicone oil. Normally it is colourless. If it observed
as yellow or brown, discard the oil. And take fresh oil in that container.
5.4.1 Follow the cleaning procedure mentioned in the steps 5.1 to 5.4.
5.4.2 Check the melting points of the following Reference Standards as per the steps
of operating procedure 5.2.1 to 5.2.9.
5.4.3 Record all the results in the calibration record as per Annexure – I.
5.4.5 Report any discrepancy observed during calibration or operating the instrument
to Department Head and notify the defect to utility department to rectify the defect.
Affix “BREAK DOWN” label on the instrument till it get rectified.
The melting point of reference standard should be within limit mentioned in the
calibration procedure as melting range.
7.0 REASON FOR REVIEW: