Standard operating procedure to calibrate the karl fischer apparatus using disodium tartrate.
1. Objective
2. Scope
This procedure is applicable to calibrate the Karl Fischer apparatus,installed in the quality control
department.
3. Responsibility
4. Accountability
5. Procedure
Frequency Once in a month
5.3.3. The performance of the instrument is satisfactory if the difference between the obtained
values is within the tolerance limit of 15.66 ± 5% i.e. 14.87% -16.44.
6. ABBREVIATIONS
6.1. % = Percentage
6.2. mg = milligram
6.3. W/W = Weight by Weight
6.4. SO P = Standard operating procedure
6.5. ml= milliliter
6.6. KFR = Karl fischer reagent
ANNEX 1
Q. C. Department
INSTRUMENT DETAILS
OBSERVATION
2 14.87% to 16.44%
BURETTE CALIBRATION :
Sr.no. Displayed Wt. of beaker+ Wt. of water Actual volume Difference Tolerance ±
volume (ml) water (g) (g) (ml) F = (E B) 0,5% of
B C D = (C A) displayed
E= (Factor X volume
D)
1
10