Professional Documents
Culture Documents
A novel approach to estimating the Doppler shift frequency from quadrature mixer output
Review of Scientific Instruments 88, 073503 (2017); https://doi.org/10.1063/1.4991018
Low noise current preamplifier for qPlus sensor deflection signal detection in atomic force
microscopy at room and low temperatures
Review of Scientific Instruments 88, 073702 (2017); https://doi.org/10.1063/1.4993737
© 2017 Author(s).
REVIEW OF SCIENTIFIC INSTRUMENTS 88, 074704 (2017)
(Received 30 January 2017; accepted 6 July 2017; published online 25 July 2017)
A 20 MHz Class D amplifier with an output of 100 W of RF power has been developed. The compact
size printed circuit board area of 50 cm2 and efficiency of 73% make it suitable for mobile nuclear
magnetic resonance (NMR) systems. Test results show that the rise and ring down times of the ampli-
fier are less than 0.2 µs, and it is capable of producing constant amplitude pulses as short as 2 µs.
Experiments using a Carr Purcell Meiboom Gill pulse sequence with a NMR MOUSE sensor confirm
that the Class D amplifier is suitable for mobile NMR applications. Published by AIP Publishing.
[http://dx.doi.org/10.1063/1.4994734]
D. Performance
The performance of our Class D amplifier was compared
to the existing Class A amplifier. Typical pulse lengths used
with the NMR MOUSE are between 2 µs and 10 µs.8 Figure 8
shows the two amplifiers outputting a short 100 W pulse into
a 50 Ω load. It can be seen that the Class A amplifier takes
about 1 µs to settle, whereas the Class D amplifier reaches the
maximum amplitude after the second cycle and stays constant
for the pulse length. Normally amplifiers with a resonant tank
output have the problem of long rise and ring-down times; this
problem was overcome by using a lower Q tank circuit and the
intelligent switching of the MOSFETs where at the end of the
pulse, the low side MOSFETs are turned on to reduce the ring
down time.
One of the NMR pulse sequences used with the NMR
MOUSE is the spin echo sequence15 and consists of two RF
pulses with the first pulse, a 90° pulse, being half the energy
FIG. 7. Existing Class A amplifier (a) with large heat-sink and the new Class
D amplifier (b). The Class A amplifier has a PCB area of 120 cm2 and weighs of the later second pulse known as the 180° pulse. The 180°
480 g; the Class D amplifier has a PCB area of 50 cm2 and weighs 120 g. pulse is either double the amplitude or double the length of
the 90° pulse. With classic (Class A) amplifier systems, the
weighs 480 g.14 A photo comparison of the two amplifiers is
shown in Fig. 7. One big difference between the amplifiers is
the heat-sink (bottom side of PCB) on the Class A amplifier.
FIG. 9. Spin echo sequence RF pulses from Class A amplifier (a) and Class
D amplifier (b) into a 50 Ω load. Note in this case, the Class D amplifier output
power level is fixed so pulse modulation is performed by altering the pulse
FIG. 8. 100 W pulses from Class A amplifier (a) and Class D amplifier (b) length. The Class D trace also shows the amplifier gating control signal that
into a 50 Ω load. The same input waveform was used in both cases. is typically applied to NMR pulse amplifiers.
074704-5 Zhen et al. Rev. Sci. Instrum. 88, 074704 (2017)
double amplitude method is used as it provides constant excita- TABLE I. Class A and D amplifier parameters including power and efficiency
tion bandwidth; however, for the Class D amplifier, the double measurements.
pulse length method is easier and is therefore used. Amplitude Parameter Magritek Class A Home built Class D
modulation using a Class D amplifier could be provided by
either modulating the conduction time of the transistors or by Rated DC supply voltage (V) 24 24
changing the bridge voltage and this will be the focus of some Maximum output power (W) 100 100
future work. Input impedance (Ω) 50 50
A spin echo pulse output from the two amplifiers is shown Output impedance (Ω) 50 50
in Fig. 9. It is interesting to note that the Class A amplifier is Input drive level (mW) <=1 1
still settling before the end of the pulse, whereas the Class D Bandwidth (MHz) 1-50 16-18
amplifier pulses show rapid rise time and very constant ampli- Module size (PCB area) (cm2 ) 160 50
tude even when doubling the pulse length. The Transistor Module weight (g) 480 120
Transistor Logic (TTL) pulse on the same figure shows the Vin (V)a 23.3 23.3
turn on and off times of the RF pulse. Iin (A)b 2.08 0.6
Another useful NMR pulse sequence for the MOUSE is Vout,p-p (V)c 190 202
the Carr Purcell Meiboom Gill (CPMG) pulse sequence16,17 η (%)d 19 73
which generates multiple echoes by using more than one 180° a V is the actual DC supply voltage to the amplifier.
in
pulse. This experiment is typically used to measure the sample b I is the difference in input current between the amplifier in the idle state and running
in
relaxation time T 2 . The efficiency of the two amplifiers was state.
cV
compared using their average input and output power over a out,pk-pk is the peak to peak output voltage measured on the 50 Ω load.
d η is the power efficiency, calculated using peak output RMS power at the duty cycle of
long CPMG pulse sequence consisting of 128°, 180° pulses at 10% verses DC input power.
10% duty cycle. A current probe was used to measure the cur-
rent drawn from the power supply of the entire spectrometer
system while the amplifier was running. This was then com-
pared to the steady state current demand when the amplifier
was not operational, the difference being the power required
for the RF generation (Fig. 10). It should be noted that both
amplifiers were expected to have similar and small standby cur-
rents, therefore justifying a relative efficiency measurement as
a good indicator of amplifier absolute efficiency. The DC input
voltage was measured to be 23.3 V, and the same 50 Ω load
was used to test both amplifiers. The corresponding RF gen-
erating current component for the Class A amplifier is 2.08 A
and 0.6 A for Class D. The RF output voltage for the Class A is
190 Vp-p and 202 Vp-p for the Class D. Based on those values,
the efficiency of the Class A amplifier was calculated to be
19% and 73% for Class D. A summary of the measurement
parameters as well as other amplifier parameters is shown in
Table I.
FIG. 11. Spin echo results from the Class A amplifier (a)
and Class D amplifier (c). CPMG results from the Class
A amplifier (b) with T2 determined to be 5.5 ms and the
Class D amplifier (d) with T2 = 4.7 ms.
amplitudes suggest that the behavior of the Class D amplifier multiple primaries in a single transformer. This would allow
is close to that of the Class A amplifier. the doubling of the output power with no expected increase in
the circuit board area.
1 M. Albulet, RF Power Amplifiers (Noble Publishing Corporation, Atlanta,
IV. CONCLUSION
GA, 2001).
2 M. K. Kazimierczuk, RF Power Amplifiers (John Wiley & Sons, West
The similarity of the NMR test results between the Class
Sussex, 2008).
A and Class D amplifiers suggests that the latter is capable 3 H. Koizumi, T. Suetsugu et al., “Class DE high-efficiency tuned power
of producing the output power and pulse sequences necessary amplifier,” IEEE Trans. Circuits Syst. I: Fundam. Theory Appl. 43(1), 51
for NMR experiments. The efficiency of 73% at the operating (1996).
4 T. Hopper, Y. Q. Song et al., “Low-frequency NMR with a non-resonant
frequency of 20 MHz with 100 W output power, combined
circuit,” J. Magn. Reson. 210, 69–74 (2011).
with the fast rise time and short turn off time, is a significant 5 N. Dye, H. Granberg et al., Radio Frequency Transistors—Principles and
improvement when compared with the existing Class A ampli- Practical Applications, 2nd ed. (Newnes, USA, 2001).
fier used with the NMR MOUSE sensor. However, it should 6 F. Schwierz and J. J. Liou, “Development of RF transistors: A historical
be noted that the Class D amplifier described here is a tuned prospect,” in Proceedings of International Conference on Solid-State and
Integrated-Circuit Technology (IEEE, 2001), Vol. 2, pp. 1314–1319.
amplifier and therefore does not have the benefit of the Class 7 X. Zhang, N. Schemm et al., “A novel power amplification scheme for
A amplifier of wide bandwidth and the ability to operate at nuclear magnetic resonance/nuclear quadrupole resonance systems,” Rev.
multiple frequencies. Nevertheless, the compact size and light Sci. Instrum. 82, 034707 (2011).
8 G. Eidmann, R. Savelsberg, P. Blümler, and B. Blümich, “The NMR
weight make the Class D amplifier a key enabler for portable
MOUSE, a mobile universal surface explorer,” J. Magn. Reson., Ser. A
NMR systems. 122, 104–109 (1996).
Some applications will require more than 100 W of 9 W. J. Chudobiak and D. F. Page, “Frequency and power limitations of Class-
RF power and or greater than 20 MHz operating frequency. D transistor amplifiers,” IEEE J. Solid-State Circuits 4(1), 25–37 (1969).
10 See http://www.ti.com/product/CSD17313Q2 for Texas Instruments.
There are several possible solutions to this need. First, new 11 See http://www.intersil.com/en/products/amplifiers-and-buffers/all-amplifi
power transistors with improved voltage and current handling ers/powerfet-ccd-drivers/ISL55110.html for Intersil.
capability as well as reduced gate capacitance are continu- 12 A. Grebennikov and N. O. Sokal, Switchmode RF Power Amplifiers
ally entering the market so it is expected that higher output (Elsevier, Inc., MA, 2007).
13 See http://www.linear.com/designtools/software/ for LTSpice.
powers and operating frequencies will be obtained in the 14 See http://www.magritek.com for Class A RF amplifier produced by
future by simply replacing the power devices and increas- Magritek Ltd.
ing the supply voltage. Second, several of the described 15 E. L. Hahn, “Spin echoes,” Phys. Rev. 80(4), 580–594 (1950).
16 S. Meiboom and D. Gill, “Modified spin-echo method for measuring nuclear
100 W modules could have their output powers combined
relaxation times,” Rev. Sci. Instrum. 29, 688–691 (1958).
through a power combiner such as a lumped element quar- 17 H. Y. Carr and E. M. Purcell, “Effects of diffusion on free precession
ter wave (Wilkinson) power combiner. Third, a new design in nuclear magnetic resonance experiments,” Phys. Rev. 94(3), 630–638
could be implemented where multiple driver stages drive (1954).