Professional Documents
Culture Documents
FACULTY OF P HYSICS
C HAIR OF M EDICAL P HYSICS
M ASTER T HESIS
Author: Supervisor:
Rangoli Saxena Prof. Dr. Katia Parodi
M.Sc. Juliana Martins
Submitted on
December 3, 2018
L UDWIG -M AXIMILIANS -U NIVERSITÄT M ÜNCHEN
FACULTY OF P HYSICS
C HAIR OF M EDICAL P HYSICS
M ASTER T HESIS
Author: Supervisor:
Rangoli Saxena Prof. Dr. Katia Parodi
M.Sc. Juliana Martins
Submitted on
December 3, 2018
iii
Abstract
Monte Carlo simulations offer powerful techniques in the field of radiotherapy for
estimating precisely and efficiently the dose delivered to the patient. Accurate Monte
Carlo models of medical linear accelerators (linacs), based on these techniques can be
utilized in creating treatment plans. To implement such a model, detailed technical
information of almost all the internal components of the linac is needed. This infor-
mation is often a commercial secret and is not disclosed by vendors. Optionally, a
commissioned Phase Space (PhSp) file available on the International Atomic Energy
Agency (IAEA) portal can be used as the photon beam source instead of modeling
the entire linac. These PhSps are characteristic to the linac model they were gen-
erated for. In the event of an unavailability of the PhSp of a specific linac, which
is intended to be modeled, this study offers an alternative approach. An available
PhSp of a linac model is optimized for a different linac model. Initially, the informa-
tion in the PhSp is perturbed randomly, and is then simulated iteratively to identify
correlations of this information with the simulated dose profiles. Once a correlation
has been identified, the PhSp is optimized under supervision, aiming at reducing
the quantified discrepancy, referred to as cost values, between the simulated and
measured dose profiles. This optimization requires several iterations, each involv-
ing numerous simulations depending on the parameter being optimized. In this
study, the energy and momentum of the particles were optimized. The optimal fix
for energy was identified and the full PhSp was validated. Considerable reduction in
cost values was observed for percentage depth dose, as well as, inline lateral profile.
The optimization of momentum was relatively more challenging. Promising results
were obtained when the momentum optimization was performed for crossline and
inline profiles individually. The merging of the two optimal momentum fix in the
same PhSp proved to be a challenge. Multiple approaches were formulated, imple-
mented and analyzed. Three such approaches are presented in this work. The best
performing approach was used to determine the optimal perturbation factors which
produced the lowest cost values. These factors were then used to perturb the full
PhSp and the corresponding cost values of the lateral profiles were determined for
all considered field sizes. The optimization of momentum secured a collective gain
in improving the lateral profiles to a considerable degree. Further improvement can
be achieved by acquiring additional measured lateral dose profiles in different lat-
eral directions between crossline and inline.
The second part of this thesis is in the realm of dosimetric validations using Elec-
tronic Portal Imaging Device (EPID). EPID is a flat panel 2D detector attached at the
bottom of the linac gantry. Although traditionally employed for only fast verifica-
tion of patient position, it poses exciting usability in determining the 3D dose distri-
bution as part of pre-treatment patient specific treatment plan verification. For this
purpose, a clinical tool, Novel EPID-based Audit Tool (NEAT), was implemented to
audit a reconstruction algorithm, which had produced promising initial results. The
tool offers a compact interface for the complete reconstruction of a 3D dose distri-
bution from frames acquisition to generation of DICOM files. 22 VMAT treatment
plans were audited using the tool. These reconstructions were then compared with
the ones computed by the Treatment Planning System, using gamma index evalu-
ation with a criteria of (3 mm, 3%). The results show that 13 out of the 22 recon-
structions passed with passing rates higher than 95%. The passing rates of the other
9 failing reconstructions improve considerably when further correction factors are
employed.
v
Acknowledgements
I would first like to thank Prof. Dr. Katia Parodi who gave me the opportunity to
work on this project at the Chair of Medical Physics. Throughout the course of work-
ing on my Master Thesis she was supportive and understanding of the problems we
were facing. Despite of her busy routine, she responded very quickly to all queries
and kept track of the thesis progress. I felt very secure under her guidance. I am also
grateful to her for giving us the chance to present a summary of this project at the
DGMP Jahrestagung and in European Congress of Medical Physics. Thank you!
Next, I would like to extend a huge thanks to Juliana Cristina Martins. She
helped me at every step. Each time I was stuck with a complex problem I would
just turn back and call for her assistance. Be it late night or early morning, she re-
sponded almost instantly. A proper researcher with incredible writing skills. She
guided me greatly while I was writing this thesis. Without her support, guidance
and a myriad of comments, this thesis would never have been in the form as it is
now. She taught me from scratch how to write scientifically, a skill I would use from
now and forever. She is an amazing teacher. I would miss her dearly in my next
projects.
A huge thanks to Sebastian Neppl, whom I nagged a lot in the last weeks of the
project. I am forever indebted to his patience and helpful nature. I am grateful for
his assistance with the linac and also for setting up my workstation at least 15 times
in the past months. Thank you for everything.
I would also like to thank Abdulaziz Alhazmi, without whose support I would
never have been able to finish the second part of my thesis. I am grateful to him for
guiding me with the complete back-end code and for always being supportive.
Thank you Carmen Seller Oria, a colleague and a great friend, who listened to
all my cribbing over project problems and dead-ends. She has been a great support
throughout my Masters.
Neha Das and Sumit Dugar! My friends from India who are now pursuing mas-
ters at TUM. Thank you for feeding me countless dinners and lunches when I was
too busy to cook anything decent! I could always count on you guys shamelessly.
Masters would have very lonely without the both of you.
Lastly, I would like to thank my parents who supported me in all ways. They are
the biggest support in my life and my loudest cheerleaders who believe in me more
than I believe in a photon being the fastest (even in vacuum).
vii
Contents
Abstract iii
Acknowledgements v
1 Introduction 1
1.1 External Beam Therapy in Cancer Treatment . . . . . . . . . . . . . . . 2
1.2 MC Techniques in External Beam Therapy . . . . . . . . . . . . . . . . . 3
1.2.1 Complications with Linac MC modeling . . . . . . . . . . . . . 3
1.2.2 Optimization of PhSp . . . . . . . . . . . . . . . . . . . . . . . . 4
1.3 EPID Dosimetry in External Beam Therapy . . . . . . . . . . . . . . . . 5
1.3.1 Novel EPID-based Audit Tool . . . . . . . . . . . . . . . . . . . 5
2 Scientific Background 7
2.1 Introduction to Radiation Therapy . . . . . . . . . . . . . . . . . . . . . 7
2.1.1 Discovery of X-rays . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.1.2 Origin of Dosimetry . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.1.3 Biological Effects . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.2 X-rays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.2.1 Production of X rays . . . . . . . . . . . . . . . . . . . . . . . . . 10
A historical perspective of X-rays production for Radiation
Therapy . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.2.2 Interaction of X-ray photons with matter . . . . . . . . . . . . . 15
Compton effect . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Photoelectric effect . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Pair production . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Rayleigh scattering . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Photonuclear interactions . . . . . . . . . . . . . . . . . . . . . . 21
2.2.3 X-ray beam attenuation . . . . . . . . . . . . . . . . . . . . . . . 22
2.2.4 Brief Overview of Dosimetric Quantities . . . . . . . . . . . . . 25
2.3 Modern Linear Accelerators . . . . . . . . . . . . . . . . . . . . . . . . . 25
2.3.1 Linac Waveguide . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
2.3.2 Linac Head . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
The X-ray target . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Beam Collimators . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Flattening filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Ion Chamber . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Wedge Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
2.3.3 Electronic Portal Imaging Devices (EPID) . . . . . . . . . . . . . 30
2.3.4 Clinical Workflow in Conformal Radiotherapy . . . . . . . . . . 30
2.4 Monte Carlo Simulation . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
2.4.1 The Monte Carlo Linac Model . . . . . . . . . . . . . . . . . . . 32
2.4.2 Geant4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
viii
4 EPID-based QA Tool 75
4.1 Materials and Methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
4.1.1 Frame Acquisition . . . . . . . . . . . . . . . . . . . . . . . . . . 76
3D Dose Reconstruction . . . . . . . . . . . . . . . . . . . . . . . 76
Gamma Evaluation . . . . . . . . . . . . . . . . . . . . . . . . . . 80
4.2 Final Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
4.3 Discussion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
Bibliography 91
Declaration of Authorship 97
ix
List of Figures
2.1 (A) The first X-ray radiograph: a picture of Anna Bertha Roentgen’s
hand (Deutsches Röntgen-Museum, Remscheid, Germany, Mould, 1995).
(B) Photograph in The Windsor Magazine of April 1896 of a lecture
demonstrating the production of an X-ray radiograph at The Royal
Photographic Society, London, UK (Mould, 1995). . . . . . . . . . . . . 8
2.2 The top plot indicates the behavior of tumor cure or control and nor-
mal tissue necrosis (tissue death) with increasing radiation dose. The
bottom plot indicates a rough estimate of the therapeutic window
which represents the probability of complication free control of dis-
ease (Easson and Pointon, 1985; Steel, 2007). . . . . . . . . . . . . . . . . 9
2.3 A typical X-ray energy spectrum produced in an X-ray tube. The
different curves represent the spectrum produced by varying peak
operating voltage of the X-ray tube, with added filtration of 1-mm
aluminum. The dashed line represents the corresponding unfiltered
energy spectrum. The sudden peaks in the curves are representative
of the characteristic radiation, possessing discrete energies depending
on the orbital transition (Khan and Gibbons, 2014). . . . . . . . . . . . . 11
2.4 X-ray therapy in 1900 using an X-ray tube (Widder, 2014). . . . . . . . 12
2.5 The 8 MeV linear accelerator at Hammersmith hospital, London, UK
with the angle of the beam outlet adjusted to produce a beam directed
vertically downwards. The first linear accelerator to be built specially
for medical use (Howard-Flanders, 1954) . . . . . . . . . . . . . . . . . 13
2.6 (A) The IMRT technique of delivering intensity modulated radiation
from different direction to get a resultant conformal dose at the tumor
volume, as explained in the paper by Brahme, 1988. (B) Multi leaf
collimator manufactured using high attenuation material like tung-
sten. These leaves can be programmed to collimate the beam differ-
entially in different directions, thus making it possible to deliver com-
plex shapes (Bortfeld, 2006). . . . . . . . . . . . . . . . . . . . . . . . . . 14
x
3.1 (A) Visual construction of the MC Model used for PDD profiles. (B)
Visual construction of the MC Model used for lateral profiles. The
labeled components in both the models: (a) Detector; (b) Water Phan-
tom; (c) Inline Diaphragm; (d) Crossline MLC; (e) Photon Beam Di-
rection. The model is not drawn to scale. . . . . . . . . . . . . . . . . . . 36
3.2 An example of cropping the original PhSp (on the left) to generate a
cropped PhSp (on the right). The cropping radius is 20 mm which
reduces the simulation time by a factor of 20. . . . . . . . . . . . . . . . 37
3.3 The blue histogram represents the energy spectrum of the original
PhSp. The green histogram represents the energy spectrum of the
manipulated PhSp for which the energy of all the particles has been
increased by 0.2 MeV. The annihilation peak at ∼0.511 MeV remains
undisturbed. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
3.4 The measured depth dose profiles for all field sizes. All measurements
were performed using a 6 MeV Elekta Synergy
R
linac coupled to
Elekta Agility TM collimator (Elekta, Stoockholm, Sweden) in a Blue
Phantom (IBA Dosimetry, Schwarzenbruck, Germany), installed at
Klinikum der Universität München - Campus Großhadern, Munich,
Germany . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
3.5 Well fitted curve using the formula given by Eq. 3.4 (left). The same
plot was erroneously fitted +on using the formula given by Eq. 3.5
(right). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
3.6 The effect of large energy perturbation to the energy of the particles
within distinct radially symmetric regions of the PhSp is presented.
The PDD profile simulated using a perturbed PhSp is represented by
the purple curve. The PDD profile simulated using the original full
PhSp is represented by the orange curve. Different plots signify the
different radially symmetric regions within which the energy of the
particles was increased by 3 MeV. These regions are: (A) 0 to 5 mm, (B)
5 to 10 mm, (C) 10 to 15 mm, (D) 15 to 20 mm, (E) 20 to 40 mm and for
(F) 40 to 60 mm. The corresponding cost (%) consistently decreased
from ∼466% in (A) to ∼0.73% in (F). . . . . . . . . . . . . . . . . . . . . 43
3.7 Comparison of the PDD profile generated using a cropped PhSp (drawn
in purple) to the PDD profile generated using the full PhSp (drawn in
orange). The cropping radius was 20 mm and the field size used was
10x10 cm 2 . The final cost (%) is determined to be 0.82%. . . . . . . . . 44
3.8 The PDD profiles generated using cropped PhSps with different en-
ergy perturbations. The corresponding cost values (%) between the
measured PDD profile (green curve) and the simulated PDD profile
(red curve) is presented. . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
3.9 Comparison of the measured PDD profiles to simulated PDD profiles
using the original unperturbed PhSp (left) and the optimized PhSp
(right), for field sizes of 2x2, 5x5, 10x10 and 20x20cm2 . The corre-
sponding cost values (%) are presented together with the curves. All
profiles have been generated using the full PhSp for validation. . . . . 48
3.10 3D visualization of the photon beam. The blue plane represents the
PhSp in the XZ plane. A random particle is located at A on the PhSp
with the total momentum in the direction AB making an angle α with
the beam direction. The beam is traveling in the -Y direction. . . . . . . 49
3.11 Measured inline lateral profiles for field size 10x10 cm2 . The fitting
function used is a sigmoid which will be explained in section 3.4.2. . . 50
xiii
3.24 The variation of absolute cost with different combinations of µ0cr and
0 . Iteration sequence are just integers denoted to every such combi-
µin
nation. A trade-off between the crossline and the inline cost values is
evident. Iteration sequence 3 represents the last combination enumer-
ated in Table 3.12. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
3.25 The variation of µop with increasing angular distance, φ as presented
in this section, represented by the thick blue line. The thick pink line is
the variation of µop with increasing angular distance, φ, as presented
in section 3.4.7. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
3.26 (A) Inline dose profile for a field size 2x2 cm2 before the optimization
of energy. (B) Inline dose profile for a field size 2x2 cm2 after the opti-
mization of energy and momentum. The profile after the optimization
of energy has a lower cost value. . . . . . . . . . . . . . . . . . . . . . . 72
3.27 (A) Percentage Depth Dose profile for a field size 10x10 cm2 before the
optimization of momentum. (B) Percentage Depth Dose profile for a
field size 10x10 cm2 after the optimization of momentum. . . . . . . . . 73
4.1 The accurate positioning of the EPID before the treatment plan is per-
formed as required. The center of the EPID is positioned along the
isocenter axis at a depth of 160 cm from the X-ray source. The green
field is the laser guide which assists in estimating the photon beam
field position. The isocenter axis passes through the intersecting point
of the faint red cross. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
4.2 (A) The Elekta iViewGT computer, enclosed in the blue rectangle, re-
ceives the frames during the treatment. (B) The frame acquisition ca-
ble is redirected to the EPID computer, shown in the picture, from the
Elekta iViewGT computer, thus effectively redirecting the acquired
frames to the self-contained EPID computer. . . . . . . . . . . . . . . . 78
4.3 The start interface of NEAT. . . . . . . . . . . . . . . . . . . . . . . . . . 78
4.4 Once the reconstruction is complete the DICOM file can be viewed by
clicking on Show Reconstruction. . . . . . . . . . . . . . . . . . . . . . . . 79
4.5 Once the reconstruction is complete the DICOM file can be saved by
clicking on Save DICOM. . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
4.6 Comparison of the TPS calculated and a NEAT reconstructed dose
distribution of an exemplary treatment plan. The top left and bot-
tom left plots are the coronal plane of the TPS calculated and NEAT
reconstructed dose distributions, respectively. The differently colored
contours stand for different dose percentages. The top right plot is the
1D lateral profile of the corresponding vertical central line in the top
left plot. The bottom right plot is the visual plot of gamma passing
rates for the coronal plane. The red colored regions stand for failing
points with dose values higher than the reference dose values. Here
reference dose value is the TPS calculated dose distribution. The blue
colored regions (not present in the plot) stand for failing points with
dose values lower than the reference dose values. . . . . . . . . . . . . 81
xv
4.7 (A) A comparison of the 1D dose profile of the TPS calculation (drawn
in orange) and the reconstruction (drawn in blue) in the coronal plane
for an exemplary treatment plan. (B) A misalignment in the TG direc-
tion is corrected and the blue plot shifts to the right by 3.6 mm. This
represents the TG factor. (C) The misalignment in the absolute dose
values is corrected and the blue plot is raised by 3%. This represents
the NF. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
xvii
List of Tables
3.1 The initial and final cost values for the optimization of energy. The
simulated PDD profiles were generated using the full PhSp. . . . . . . 46
3.2 The initial and final cost values for the inline lateral profile after the
optimization of energy for field size of 2x2, 5x5, 10x10 and 20x20 cm2 . . 46
3.3 The initial and final cost values for the crossline lateral profile before
and after the optimization of energy for field size of 2x2, 5x5, 10x10
and 20x20 cm2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
3.4 The Sigmoid Fitting Parameters . . . . . . . . . . . . . . . . . . . . . . . 52
3.5 The Linear Fitting Parameters . . . . . . . . . . . . . . . . . . . . . . . . 52
3.6 The arithmetic progression of the perturbation factor with respect to
the sequence of simulation. . . . . . . . . . . . . . . . . . . . . . . . . . 57
3.7 Comparing the inline fitting parameters determined using a cropped
PhSp and the full PhSp. . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
3.8 Comparing the crossline fitting parameters determined using a cropped
PhSp and the full PhSp. . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
3.9 Results of the crossline profile (corresponding to fitting parameter k)
following the methodology, explained in the previous section 3.4.6.
∆Cost written in green represents a drop in cost and in red represents
a gain in cost . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
3.10 Initial cost values for inline and crossline profiles as reference. . . . . . 64
3.11 Results of the crossline profile for φ2 approach. ∆Cost written in green
represents a drop in cost and in red represents a gain in cost. . . . . . . 65
3.12 Cost values for different combinations of µ0cr and µin 0 . ∆Cost written
3.14 Final cost values of the crossline profiles for the best performing ap-
proach, compensated φ6 , generated for field sizes 2x2, 5x5, 10x10 and
20x20 cm2 using the full PhSp as compared to the initial cost values
before the optimization of energy and momentum. The initial cost
values before the optimization of energy can be obtained from Table
3.3. ∆Cost written in green represents a drop in cost and in red repre-
sents a gain in cost. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
3.15 Final cost values of the inline profiles for the best performing ap-
proach, compensated φ6 , generated for field sizes 2x2, 5x5, 10x10 and
20x20 cm2 using the full PhSp as compared to the initial cost values
before the optimization of energy and momentum. The initial cost
values before the optimization of energy can be obtained from Table
3.2. ∆Cost written in green represents a drop in cost and in red repre-
sents a gain in cost. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
3.16 Optimal perturbation factors determined using PhSp cropped with
different cropping radii. . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
4.1 (3%, 3 mm) gamma index evaluation for the NEAT reconstructed dose
distribution for 22 treatment plans. NF stands for normalization fac-
tor. TG corrected implies that the reconstruction has been shifted by
3.6 mm in the -Z direction to correct for the EPID positioning discrep-
ancy. Passing rates highlighted in green represent values greater than
95%. Yellow represents passing rates between 80% and 95%. Red
represents passing rates below 80%. TP stands for Treatment Plan.
The reconstructions which had achieved a passing rate higher than
the threshold was not corrected further using NF 8%, as a higher NF
produced a worse passing rate. . . . . . . . . . . . . . . . . . . . . . . . 83
4.2 (2%, 2 mm) and (1%, 1 mm) gamma index evaluation for the NEAT
reconstructed dose distribution for the the 5 best performing recon-
structions. Passing rates highlighted in green represent values greater
than 95%. Yellow represents passing rates between 80% and 95%. Red
represents passing rates below 80%. . . . . . . . . . . . . . . . . . . . . 84
xix
Physical Constants
Chapter 1
Introduction
Cancer is the uncontrolled growth and spread of cells that arises from a gene muta-
tion. This mutation may have been caused by dietary habits, genetic factors and/or
environmental factors. The transformation of a healthy cell into a cancerous cell can
eventually lead to tumors which can metastasize to other parts of the body (Cassidy
et al., 2015). Such an abnormal growth if not controlled can lead to death. A study
performed by the International Agency for Research on Cancer (Global Cancer Facts
& Figures 3rd Edition. 2015) reveals that cancer is the second leading cause of death
in high income countries and the third leading in low and middle income countries,
as can be seen in Table 1.1. There will be an estimated 18.1 million new cancer cases
and 9.6 million cancer deaths in 2018 (Bray et al., 2018). Globally, about 1 in 6 deaths
has been due to cancer (Cancer, Key facts 2018). Thus, cancer treatment has become
the focus of many growing areas of research.
clinical practices. The reconstruction algorithm used in NEAT has been provided by
Alhazmi et al., 2018.
In this study 22 Volumetric Arc Therapy (VMAT) treatment plans were per-
formed on a 6 MeV Elekta Synergy
R Linac coupled to Elekta AgilityTM collima-
tor (Elekta, Stoockholm, Sweden) installed at Klinikum der Universität München -
Campus Großhadern, Munich, Germany. A PerkinElmer XRD 1640 AL5 P aSi EPID
(Elekta iVewGT) mounted at source to surface distance of 160 cm was used for in air
frame acquisitions of the treatment plans. The frames acquired were then used to
reconstruct 3D dose distributions as imparted to a water equivalent medium. These
reconstructions were compared to the planned dose distributions, calculated by the
Treatment Planning System using a gamma evaluation with a (3%, 3 mm) criteria for
the purpose of testing the tool and auditing the reconstruction algorithm.
F IGURE 1.2: The start interface of NEAT: a clinical tool for EPID-
based dosimetry as part of pre-treatment patient specific QA.
7
Chapter 2
Scientific Background
(A) (B)
also documented when Bequerel carried a tube of pure radium in his pocket. Af-
ter 15 days, a severe burn underneath this particular pocket developed which was
deduced to be due to the effect of radium. These events indicated the need of analyz-
ing the biological effects of the radiation in human tissue. Radiation dose was thus
defined as the quantified energy deposited by radiation to matter. Dosimetry deals
with formation and validation of methods and procedures to measure the radiation
dose and its correlation with biological effectiveness. Over the years, efforts were
made to introduce suitable dosimetric concepts and conventions associated to strict
physical definitions. The current SI dosimetric quantity for dose is Gray, abbreviated
as Gy. Its equivalent relationship in energy units is Joules per kilogram (Ma, 2014;
Parodi, 2017). This is directly correlated to the biological effectiveness of ionizing
radiation.
This difference creates a therapeutic window as presented in the bottom plot of Fig.
2.2. It is a bell-shaped curve which represents the probability of complication free
control of disease. This therapeutic window enables the use of ionizing radiations
in cancer treatment (Steel, 2007).
F IGURE 2.2: The top plot indicates the behavior of tumor cure or con-
trol and normal tissue necrosis (tissue death) with increasing radia-
tion dose. The bottom plot indicates a rough estimate of the thera-
peutic window which represents the probability of complication free
control of disease (Easson and Pointon, 1985; Steel, 2007).
2.2 X-rays
The term radiation is referred to the propagation of energy through space or a medium.
Radiations which are characterized by their pronounced capability of ionizing the
atoms of the interacting medium are referred to as ionizing radiation. The ICRU (In-
ternational Commission on Radiation Units and Measurements, 1971) recommends
two broad classifications of ionizing radiation, on account of the grossly distinct in-
teractions of charged and uncharged particles with matter (Attix, 1986):
10 Chapter 2. Scientific Background
If no filtration is added in the path of the generated beam, the energy spectrum
for the bremsstrahlung radiation will be a continuous straight line, as represented
by the dotted line in Fig. 2.3. It is given by the Kramer’s equation, Eq.(2.1).
IE = KZ ( Em − E) (2.1)
Here, IE is the intensity of the photon with energy E, Z is the atomic number of
the target, Em is the maximum photon energy and K is a constant. The maximum
possible energy a photon can acquire is equal to the energy of the incident electron
which indirectly or directly produces it. The maximum energy of the electron is
determined by the peak operating voltage of the X-ray tube. Thus, on increasing
the potential difference between the two electrodes, the energy of the electron beam
can be increased, effectively increasing the energies of the photons produced. This
changes the shape of the spectrum as can be seen in Fig. 2.3. The upper curves cor-
respond to the X-ray energy spectrum produced by a relatively higher peak voltage,
or excitation voltage, applied to the electrodes of the X-ray tube (Khan and Gibbons,
2014).
It also marked a transitional period when giant particle accelerators were beginning
to be constructed (Slater, 2012; Robison, 1995; Hefner, 1996).
One of the earliest recorded treatment using a 1 MV X-ray beam was performed
in 1937. It was delivered by a 1 MV tube and a Van De Graaff generator unit, which
was available for treatment at the Collis P. Huntington Memorial Hospital in Boston.
A Van de Graaff generator is based on the concept of accumulating charge on the in-
side of a hollow metal sphere, insulated from the surroundings, using a moving belt.
This was capable of generating very high potentials which then produced high en-
ergy X-rays (Graaff et al., 1933). The patient who was treated with the generator was
recorded to have outlived the Boston hospital, which closed in 1941 (Robison, 1995).
Although, treatments with a Megavoltage X-ray beam was principally possible us-
ing such generators, it escalated with the advent of Cobalt 60 (Co-60) as a radiation
source which was comparatively much more stable than the other isotopes of Co and
had a long half-life of approximately 5.3 years. It produces γ-rays in the range of 1 to
1.3 MV. In 1951 the first clinical treatment using Co-60 γ-rays was performed, at the
Victoria Hospital in London, Ontario, Canada (Robison, 1995; London Health Sciences
Centre 2011). A survey done in 1986 by the American College of Radiology shows
the notable presence of Co-60 therapy units in cancer treatment (Robison, 1995), as
presented in Table 2.1.
The first clinical linac was installed for the Medical Research Council at Ham-
mersmith Hospital, London, UK. It was an 8 MeV machine attached to the ceiling
with exceedingly limited degree of freedom as can be seen in Fig. 2.5.
By mid 1950s, high energy X-rays were routinely employed in clinical treatments
for deep seated tumors owing to their increased penetration depth. However, as
high energy meant a higher dose delivery, the need of conforming the radiation to
the site of the tumor while sparing the surrounding healthy tissues emerged. This
brought the advent of three dimensional conformal radiation therapy techniques
(3D CRT). In this technique the radiation dose was delivered from multiple beam
angles, using stationary collimators to produce different beam shapes, thus deliver-
ing a relatively more conformal dose at the site of the tumor. The paper by Brahme,
Roos, and Lax, 1982 proved to be a cornerstone for Intensity Modulated Radiation
Therapy (IMRT), an advanced 3D CRT technique. The paper presented a mathemat-
ical solution of an integral, which resulted in the absorbed dose distribution during
a single 360-degree rotation of the beam delivery angle.
The IMRT technique uses multiple beams in different directions. The intensity
of the beam is modulated during the irradiation of the tumor thus delivering a very
conformal dose at the tumor volume as presented in Fig. 2.6 (A). The intensity is
14 Chapter 2. Scientific Background
modulated using a Multi Leaf Collimator (MLC) as shown in Fig. 2.6 (B) which
collimates it into complex shapes (Bortfeld, 2006).
(A) (B)
By 2000, IMRT was considered to be one of the most advanced radiation tech-
niques for high energy X-ray photon therapy. Around 2008, Volumetric Arc Ther-
apy (VMAT), a novel delivery technique, was introduced. This technique was an
improvement over IMRT delivering the radiation with a 360-degree continuous ro-
tation of the gantry around the patient. The conventional IMRT treatments, how-
ever required the rotation of the gantry several times and/or with repeated starts
and stops, thus increasing the delivery time. One of the earliest study using VMAT
was done by Palma et al., 2008. In this study, the conformity of the treatment plans
of a VMAT was compared to that of an IMRT and a conventional 3D CRT. The re-
sults showed that IMRT and VMAT were significantly much more conformal in their
dose deliveries as compared to 3D CRT. The work published in the paper by Teoh
et al., 2011 compares two dose distributions achieved using VMAT and conventional
IMRT showing that VMAT has a superior conformity. Fig. 2.7 shows a commercial
Elekta Synergy
R
linear accelerator, installed at Klinikum der Universität München -
Campus Großhadern, Munich, Germany. It is capable of performing treatment using
advanced delivery techniques including IMRT and VMAT.
2.2. X-rays 15
1. Compton effect
2. Photoelectric effect
3. Pair production
4. Rayleigh scattering
5. Photonuclear interactions
The photon first transfers energy to secondary particles, electrons, of the medium
with which they are interacting via any of the first three interactions. These electrons
then deposit some or all of its energy to the medium via Coulomb force interactions
along their path of travel. The first three interactions are thus crucial for the use of
X-ray in therapy. In Rayleigh interaction, a photon trajectory is deflected without
any change in its energy, therefore, no energy is release in the material. Photonu-
clear interactions are observed for photons with energy above few MeV and are not
relevant for the photon energies dealt in this study. In such interactions, the high
energy photon is absorbed by the nucleus and a proton or a neutron is emitted. The
resulting nucleus is radioactive (Attix, 1986).
Compton effect
In Compton effect, an unbound electron is struck by an incoming photon, scattering
it at an angle. The photon transfers some of its energy to the electron and is deflected
by an angle from its initial path. The scattering angles and the final energies of the
electron and the photon can be solved using the kinematic law of conservation of
energy and momentum. These final quantities are dependent on the initial energy of
16 Chapter 2. Scientific Background
the photon. The resulting solutions are presented in Eq. (2.2), (2.3) and (2.4), where
hν and hν’ is the initial and final energy of the photon respectively and T is the final
kinetic energy of the scattered electron. θ and φ are the scattering angles for the
electron and the photon respectively. mo c2 is the electron rest energy. A schematic of
the Compton interaction is depicted in Fig. 2.8.
hν
hν0 = (2.2)
1 + ( mhνc2 )(1 − cosφ)
0
0
T = hν − hν (2.3)
hν φ
cotθ = (1 + 2
)tan (2.4)
m0 c 2
There are two quantities of interest. The first is the probability of the occurrence
of a Compton interaction which is given by the scattering cross section. The second
is the average fraction of the incident energy transferred to the electron. The dif-
ferential cross-section for an interaction between a photon and a single free electron
de σ
per unit solid angle ( dΩ φ
) is given by the Klein-Nishina (K-N) formula, as can be
referred from Eq. (2.5). Here, hν is the initial photon energy, φ is the scattering angle
of the photon, r0 is the classical electron radius, which is equal to 2.818 × 10−13 cm
(Attix, 1986; Podgorsak, 2010) and dΩφ is the differential solid angle for the photon
scattering angle φ.
2
r2 hν0 hν0
de σ hν
= 0 0
+ − sin2 φ (2.5)
dΩφ 2 hν hν hν
The total electronic cross-section, e σ, for the photon interaction with a single elec-
tron is calculated by integrating this differential equation for all photon scattering
angles. The final value is dependent on the initial energy of the photon hν. The
variation of the cross-section with respect to different photon energies is presented
2.2. X-rays 17
graphically as the upper curve in Fig. 2.9 (A). Furthermore, this cross-section is in-
dependent of Z, thus, the total atomic cross-section is proportional to the atomic
number, Z, of the interacting matter.
aσ ∝Z (2.6)
The proportionality presented in Eq. 2.6 will be revisited later in this chapter when
X-ray beam attenuation is explained (section 2.2.3).
The fraction of the photon energy transferred to the electron is of prime interest
as it indirectly effects the amount of dose deposited in the medium. For a single
T
electron, this fraction can be written as hν . Thus, the differential cross-section for
energy transfer is given by Eq. (2.7). The total cross-section for energy transfer to a
single electron (e σtr ) is then calculated by integrating it over all scattering angles. It
is graphically shown as the bottom curve in Fig. 2.9 (A). Finally the average fraction
of the total photon energy imparted to the electrons is calculated by using eq. (2.8).
The variation of this fraction with respect to different hν values is shown in Fig. 2.9
(B).
de σtr de σ T
= · (2.7)
dΩφ dΩφ hν
T̄ e σtr
= (2.8)
hν eσ
(A) (B)
F IGURE 2.9: (A) The variation of e σtr and e σ with respect to the initial
photon energy hν. (B) The variation of the mean fraction of the total
photon energy transferred to the electron for all scattering angles with
respect to the initial photon energy hν (Attix, 1986).
Thus, it can be observed that the fraction of the energy imparted to the electrons
increases with increasing initial photon energy, hν for given cross-sections, e σtr and
e σ.
Photoelectric effect
The photoelectric effect is essentially an absorption of a photon by a bound electron.
The photon instead of transferring only a part of its energy, like in Compton effect,
imparts all of its energy to a bound electron and ceases to exist. The electron is then
ejected from the atom at an angle and the nucleus absorbs the recoil momentum as
18 Chapter 2. Scientific Background
shown in Fig. 2.10 (Attix, 1986). The final kinetic energy of the nucleus is negligible
owing to its higher mass.
The photoelectric effect only occurs if the energy of the incoming photon, hν, is
greater than the binding energy of the electron, Eb . The difference of the two energies
is the resultant kinetic energy, T, of the ejected electron.
Zn
a σph ∝ (2.9)
hνm
The sudden surge in the cross-section, as can been seen in the curves in Fig. 2.11
(denoted by the letters ’K’, ’L’ and ’M’), occurs when the photon energy is very close
to the binding energy of an atomic shell. In such cases, resonance occurs and the
probability of the absorption of the photon by an electron from that specific shell
escalates abruptly (Khan and Gibbons, 2014). These effects are referred to as shell
absorption edges in the plot and is denoted by ’K’, ’L’ and ’M’ for their respective
shells.
2.2. X-rays 19
A vacancy in an inner shell of the atom, caused by any process such as the pho-
toelectric effect or shell transition effect, can produce fluorescence X-rays. These
X-rays are emitted when this vacancy is filled by an electron from a higher shell.
Furthermore, the fluorescence X-rays can be absorbed by another bound electron
of the same atom subjecting it to an ejection. Such an effect is classified in the cat-
egory of Auger emission which contributes to the final dose deposited in the sur-
rounding matter. Under Auger emission, the electrons can also be emitted by other
mechanisms such as the Coster-Kronig effect or the super Coster-Kronig effect. In
these mechanisms the electron transitions to a sub-shell with a higher binding en-
ergy within the same shell. The emitted radiation is then absorbed by either an
electron from a higher shell (Coster-Kronig effect) or an electron in the same shell
with a lower binding energy (super Coster-Kronig effect) (Podgorsak, 2010).
Photoelectric effect is the dominant interaction for X-ray imaging and is not of
much consequence in the therapeutic energy range. This will be reviewed later,
along with sufficient explanations, in X-ray beam attenuation, section 2.2.3.
Pair production
When a photon with energy higher than 1.02 MeV is in the vicinity of a Coulomb
force field, there is a reasonable probability of it getting absorbed giving rise to an
electron and a positron. Such an absorption process is called a pair production, the
electron and the positron being the pair produced (Fig. 2.12). This Coulomb force
field, which triggers such an event, is generally that of an atomic nucleus. However,
a similar event can also occur near the Coulomb force field of an atomic electron for
even higher photon energies. The photon is absorbed giving rise to two electrons
and a positron at the site of absorption. Such an event is called a triplet production
and will be discussed briefly later in this section (Attix, 1986).
The relationship for conservation of energy in a pair production can be given by
Eq. (2.10). Here the initial energy of the photon is hν, the final energies of the electron
and the positron are T− and T+ respectively, and the rest mass energy of an electron
20 Chapter 2. Scientific Background
is mo c2 which is approximately 0.51 MeV. The average kinetic energy received by the
electron and the positron is given by Eq. (2.11). It is not necessary that they receive
equal energy.
hν = ( T− + mo c2 ) + ( T+ + m0 c2 ) (2.10)
hν − 2m0 c2
T̄ = (2.11)
2
The differential atomic cross-section of nuclear pair production for the creation of
a positron of energy T+ (and a corresponding electron of energy of hv − 2m0 c2 − T+ ),
is derived by the Bethe-Heitler theory as presented in Bethe and Maximon, 1954. On
integrating the differential atomic cross-section over all values of T+ , the total atomic
cross-section, a σn , is given by the proportionality (2.12).
a σn ∝ Z2 (2.12)
The atomic cross-section for a triplet production is relatively small when com-
pared to the atomic cross-section of pair production. The ratio of the two cross-
sections is given by Eq. (2.13). Here, a σe is the atomic cross-section of a triplet pro-
duction and a σn is the atomic cross-section of a pair production. C is a parameter
dependent entirely on hν, which increases as the photon energy decreases. Thus
for lower energies the probability of a triplet production decreases for a specific Z.
The threshold energy below which triplet production does not occur is 4m0 c2 . The
different values of the aforementioned ratio for different gaseous medium has been
presented in the work by Evans, 1955 as can be referred from Fig. 2.13. The photon
energy used was ranging from 6-7 MeV. It can be inferred from the same that even
at such a high energy the number of pairs produced are much higher than the num-
ber of triplets produced. Triplet production is not a dominant event for the photon
energies dealt in this study.
a σe 1
= (2.13)
a σn CZ
2.2. X-rays 21
Rayleigh scattering
Rayleigh scattering is a coherent elastic scattering where, coherent means that the
photon interacts with the entire atom as a whole, and elastic means that the photon
does not lose any of its energy in the process. It is merely deflected by an angle from
its original path.
The Rayleigh scattering atomic cross-section is dependent on the Z and hν ac-
cording to the proportionality (2.14).
2
Z
a σR ∝ (2.14)
hν
Photonuclear interactions
In photonuclear interactions, an energetic photon, with energy higher than a few
MeV, enters an atomic nucleus. The excited nucleus then emits either a neutron or a
proton.
The emitted proton contributes directly to the dose imparted to the surround-
ing matter. Although, the relative contribution is less than 5% because of the high
probability that the photon undergoes pair production instead, being in the close
proximity of the nucleus (Attix, 1986).
The emitted neutron generates a need of better radiation protection. Neutrons
are hard to shield as they are neutral and can pass through dense materials unat-
tenuated. Furthermore, the resultant daughter nuclei may be radioactive adding to
the radiation hazard. The threshold energy, which denotes the binding energy of a
neutron with the nucleus, is of the order of 8 MeV or higher for most nuclei, except
for Deuteron and Beryllium-9 where it is 2.22 MeV and 1.66 MeV, respectively (Pod-
gorsak, 2010). Thus, it is a concern for facilities in which machines are capable of
producing photon beams of 8 MeV or higher. The beam energy used in this study is
around 6 MeV and hence such an interaction is not a concern.
22 Chapter 2. Scientific Background
µo = µc + τ + κ (2.15)
I = I0 exp(−µc x − τx − κx ) (2.16)
Mass attenuation coefficient for any type of interaction is the linear attenuation
coefficient divided by the density of the material. It is fundamentally more valu-
able than the linear coefficient because of its independence from the density and the
physical state of the medium material. The mass attenuation coefficients for differ-
ent interactions are related to the corresponding atomic cross-sections by Eq. (2.17).
Here µ is the linear attenuation coefficient, ρ is the density of the material, a σ is the
atomic cross-section of the interaction, NA is Avogadro’s constant,
Z is the atomic
Z
number and A is the atomic mass of the material. The ratio A written separately
in the parenthesis is roughly constant, equal to 0.45 ± 0.05 (Evans, 1955).
µ N Z
=a σ · A (2.17)
ρ Z A
The dependence of the mass attenuation coefficient for the different interactions
can now be given as Eq. (2.18), (2.19) and (2.20).
µc
∝ Z0 (2.18)
ρ
τ
∝ Z3 (2.19)
ρ
κ
∝ Z1 (2.20)
ρ
2.2. X-rays 23
(A)
(B)
dN
Φ= (2.21)
dA
The energy fluence is the ratio of the amount of radiant energy, dR, incident on
a sphere to the cross-sectional area of that sphere, dA and is defined by (2.22). It is
related to the particle fluence by (2.23) where E is the energy of the monoenergetic
photons. (Seuntjens, Strydom, and Shortt, 2005).
dR
Ψ= (2.22)
dA
Ψ = ΦE (2.23)
Kinetic energy released per unit mass, also known as kerma, is defined as the
mean energy, Ētr , transferred by the electrically neutral radiation to charged parti-
cles in the medium per unit mass, m. This quantity is only applicable to indirectly
ionizing radiations and is defined by (2.24). The unit of kerma is joule per kilogram
(J/kg) and is called as the gray (Gy) (Seuntjens, Strydom, and Shortt, 2005) .
d Ētr
K= (2.24)
dm
In an indirectly ionizing radiation the energy is deposited in the medium in a two
step process. In the first step, the energy is imparted to the charged particles of the
medium as kinetic energy. The charged particles then transfer some of this kinetic
energy to the medium. This absorbed energy is the dose imparted to the medium.
It is defined as the mean energy, ē deposited in a finite volume of the material, with
mass m and is given by (2.25). The unit of absorbed dose is joule per kilogram (J/kg)
and is called as the gray (Gy) (Seuntjens, Strydom, and Shortt, 2005).
dē
D= (2.25)
dm
into complex shapes but rather had simple rectangular collimation. In modern tech-
nologies, emphasis is on shaping the beam using Multi Leaf Collimator (MLC). MLC
modulates the intensity of the beam in different directions thus conforming the dose
at the site of the tumor while minimizing the dose to the healthy tissues surrounding
it. This requires complex programming and sophisticated computational models to
solve for the inverse treatment plans.
2.18. It can rotate around the patient table thus delivering the beam from any di-
rection in the rotation plane. Modern techniques of delivering the beam like VMAT
(as introduced in section 2.2.1) utilizes the gantry’s capability of achieving variable
rotational speed along with a continuous delivery unlike conventional delivery tech-
niques. The linac in Fig. 2.18 depicts an Elekta Synergy
R
linac (Elekta 2018).
Beam Collimators
In Elekta Synergy
R
, the beam is collimated at several steps, the first of which is
done by the primary collimator. It is a cone shaped collimator positioned close to
the target (X-ray source) to limit the field to a circular shape (as can been in Fig.
2.19). Afterwards, the beam is collimated by further collimating devices. The con-
ventional collimators can only shape the radiation in rectangular fields, which offers
a limited conformity. With the development of Multi Leaf Collimator (MLC) it be-
came possible to produce highly complex radiation fields, as they are composed
of many thin leaves which move independent from one another. In the present
study, an Elekta Synergy
R
linac coupled to an Elekta AgilityTM collimator are used.
The Agility MLC is composed by 80 pairs of inter-digitating leaves and two block-
diaphragms that move perpendicular to the leaves. The leaves can move indepen-
dently. This movement is controlled by miniature direct current motors according
to a pre-uploaded MLC plan. Each leaf is fitted with a reflector illuminated by an
28 Chapter 2. Scientific Background
F IGURE 2.18: A visualization of the linac head and the patient table.
The X-ray beam (yellow rays) is modified in the linac head. The linac
head or the gantry can rotate around the patient table delivering the
beam in an arc like movement. The patient table also has multiple
degrees of freedom to further aid in the conformity of the dose (How a
Linear Accelerator Works, Elekta 2010). (The X-ray beam is represented
as a yellow radiation only for illustration purposes). The Electronic
Portal Imaging Device, EPID, will be explained in the next section.
optical beam. The reflection from them are captured by a video camera. This infor-
mation is used to continuously estimate and validate the position of the leaf. In case
of an error, the delivery is stopped (Loverock, 2007).
Flattening filter
The X-ray beam produced via Bremsstrahlung is primarily directed forward. This
produces a beam which has a lower profile in the edges, in comparison to the central
part of the profile as shown in Fig. 2.20 (left). The filter is designed in a way that it
flattens the beam profile as shown in Fig. 2.20 (right). It is usually radially symmetric
(Loverock, 2007).
Ion Chamber
A transmission ionization chamber is placed right above the MLC, which is the clos-
est practical position to the patient. Its primary purpose is monitoring the delivered
dose and the dose rate along with the beam uniformity and flatness. It also auto-
matically terminates treatment if any of the mentioned quantities are outside the
acceptable tolerances (Loverock, 2007).
Wedge Filter
In cases where a wedge-shaped beam profile is needed instead of a uniform one, the
wedge filter is employed. This filter can be placed either above or below the moving
collimator assembly. It changes the beam profile according to the length, the time
and the position of its insertion into the beam during the delivery (Loverock, 2007).
2.3. Modern Linear Accelerators 29
1. The patient is immobilized and a high quality three dimensional medical im-
age is acquired to anatomically locate the disease. These images can be any
high quality imaging modality like X-ray CT (Computed Tomography), PET
(Positron Emission Tomography), MRI (Magnetic Resonance Imaging) images
or any combination thereof.
2. Then the target volumes, such as the PTV and the OAR are identified and con-
toured. Once the desired contours of the dose delivery have been identified,
2.3. Modern Linear Accelerators 31
6. To asses the response of the treatment, medical images are acquired again after
a certain amount of time has passed and then analyzed.
This study is involved with the step 2 and 3 of this clinical workflow.
2.4.2 Geant4
Geant4 (Geometry And Tracking) is an open source simulation toolkit, written in
C++, which can be customized vigorously for accurately simulating passage of par-
ticles through matter. It was developed by The European Organization for Nuclear
Research (CERN) and is based on an object-oriented framework to enhance the ease
of use (CERN, 2018).
The simulation model can be configured by defining almost every intricate de-
tail of the setup. This includes the geometry, the spatial position and relative move-
ment, the density and the material composition. The physics can also be governed
by defining the physical processes to be included, the number and the type of par-
ticles involved and their statistical weight among others. The detector position can
be defined precisely including the resolution and spatial position and dimension.
Multiple detectors can be used at the same time and the response format can also be
customized. Furthermore, the complete information of events and tracks during the
simulation can be stored. The toolkit at the fundamental level includes abundant
34 Chapter 2. Scientific Background
F IGURE 2.23: The mirror is part of an optical system used for reflect-
ing the information received from the reflectors on the leaves as a
verification measure. Mylar is a highly reflective material (Loverock,
2007).
F IGURE 2.24: Visualizing the setup model using OpenGl. The inter-
face comes with a dialog box which can be used to edit the model and
run the simulations in real-time.
All simulations done in this study have been performed using Geant4 10.01.p02.
35
Chapter 3
A Phase Space (PhSp) File is a record of all or a part of the information of the particles
passing through an imaginary plane in the path of a particle beam. It is a powerful
tool used in decreasing redundancy in MC simulations by employing it as the beam
source instead of modeling the entire linac. Optimizing a commissioned PhSp, vali-
dated for a specific linac model, to a different linac model requires a set procedure to
determine the progress of the optimization. This procedure involves comparing the
simulation and the measured dose profiles in every optimization cycle. The infor-
mation in the PhSp file, used as the beam source in the model, is then manipulated
according to the quantified discrepancies between measured and simulated profiles.
Such discrepancies from hereon will be referred to as cost values. This process is
repeated until a minimum cost value is reached. Furthermore, dosimetric commis-
sioning of an IMRT planning system requires a systematic sequence to be followed,
for example, the guidelines mentioned in The American Association of Physicists in
Medicine (AAPM) report (Ezzell et al., 2003). Keeping in line with these guidelines,
the simulated lateral and percentage depth dose (PDD) profiles were compared to
the corresponding measurements as a reference for the optimization. The PDD pro-
file is the dose distribution at different depths in the absorption medium and lateral
profile is the dose distribution in a plane perpendicular to the beam central axis and
at a certain depth in the absorption medium (Ezzell et al., 2003). The dimension and
spatial position of the detectors for the two profiles are mentioned in section 3.1.
In this study, a commissioned PhSp for the MC model of an Elekta Precise for a
6 MeV photon beam was used, available at the International Atomic Energy Agency
(IAEA) website (Capote, 2011) and optimized it for a 6 Mev photon beam of an
Elekta Synergy. The information optimized is the energy and the momentum of the
particles.
(A) (B)
F IGURE 3.1: (A) Visual construction of the MC Model used for PDD
profiles. (B) Visual construction of the MC Model used for lateral
profiles. The labeled components in both the models: (a) Detector; (b)
Water Phantom; (c) Inline Diaphragm; (d) Crossline MLC; (e) Photon
Beam Direction. The model is not drawn to scale.
and 4-mm elsewhere. The PDD profiles were measured from the surface of the wa-
ter phantom to a depth of 300 mm, with 1-mm step size in the build-up region. The
step size then increased to 2-mm up to the 50th dose measuring point and further
increased to 4 mm for the rest of the measurement.
A virtual water phantom was defined in the existing MC model of the equip-
ment. A command-based scoring method, with 1 mm step size, was used for dose
recording in the phantom for both PDD and profile simulations. The visual schematic
of the MC model, including the collimator assembly, is presented in Fig. 3.1.
The commissioned IAEA Phase Space (Capote, 2011) was recorded at a distance
of 27.21 cm from the X-ray source. A new PhSp positioned 2.79 cm below the origi-
nal IAEA PhSp was generated, from hereon referred to as the LMU PhSp, recording
only the relevant information for the simulations performed for the present investi-
gation. This LMU PhSp has been optimized in this study.
The linac was modeled following the "phase-space" approach introduced in chap-
ter 2. The LMU PhSp was used as a substitute for the static part of the linac head.
The dynamic part of the model, namely the AgilityTM collimator, labeled as (c) and
(d) in Fig. 3.1 was modeled and commissioned in Martins et al., 2017.
PhSp is generated, cropping the original PhSp into a smaller radius as represented in
Fig.3.2. This radius has been referred to as the cropping radius or the cut-off radius.
The location of the particles, the dimension and spatial position of the detector and
the field size of the photon beam in the MC Model determine the smallest appropri-
ate cropping radius. An example of cropping is presented in Fig. 3.2, in which the
cropped PhSp only has the information of the particles within 0 to 20 mm.
F IGURE 3.2: An example of cropping the original PhSp (on the left)
to generate a cropped PhSp (on the right). The cropping radius is 20
mm which reduces the simulation time by a factor of 20.
38 Chapter 3. Optimization of Phase Space File
µ
TE (r ) = EφE (r ) (3.1)
ρ
Z Z Z Z
D (r ) = TE (s)hρ ( E, r − s)dEd3 s (3.2)
This shows that the PDD profile of a poly-energetic photon beam depends sig-
nificantly on its energy. Thus, PDD profiles can be taken as a measure to optimize
the energy spectrum of the PhSp.
Furthermore, the mathematical manipulation to the energy was chosen as a sim-
ple addition or subtraction to keep it uncomplicated. This meant that an addition of
0.2 MeV to the PhSp would be performed by adding 0.2 MeV to the energy of every
photon in the PhSp as presented in Fig. 3.3. This manipulation shifts the mean and
the median of the energy spectrum by the same amount. The only constraint iden-
tified in this manipulation was the annihilation peak at approximately 0.511 MeV
which was independent of the energy distribution and thus was not affected by the
manipulation. The annihilation peak comes from an abundance of photons with en-
ergy very close or equal to 0.511 MeV. This is indirectly related to the production
of positrons during pair-production. These positrons usually travel a small distance
and then annihilate with an interacting electron producing two photons of approxi-
mately 0.511 MeV energy.
F IGURE 3.4: The measured depth dose profiles for all field sizes. All
measurements were performed using a 6 MeV Elekta Synergy
R
linac
coupled to Elekta AgilityTM collimator (Elekta, Stoockholm, Sweden)
in a Blue Phantom (IBA Dosimetry, Schwarzenbruck, Germany), in-
stalled at Klinikum der Universität München - Campus Großhadern,
Munich, Germany
The χ2 function given by Eq. (3.3) is the chosen cost function for PDD (Aljarrah
et al., 2006). It was selected because of its simplicity and accurate reflection of the
disagreement between the simulation and the measurement profiles. Furthermore,
it is reasonably insensitive to statistical uncertainty as shown in the extensive study
presented in the paper Aljarrah et al., 2006. In Eq. (3.3), χ2 is the cost, Dim is the
normalized dose at a dose deposition point i for the measured profiles and Dis is the
normalized dose, normalized to the mean of the dose deposited at points near or at
the point of maximum dose, at a dose deposition point i for the simulated profiles.
n is the total number of points used in the calculation of the cost.
1 n
∑
(m) (s)
χ2 = ( Di − Di ) 2 (3.3)
n i =1
The measured and the simulated depth dose profiles consists of approximately
300 dose values normalized to the maximum as can be seen in Fig. 3.4 and 3.5 (left).
The measured dose values are positioned at a step size of 1 mm in the buildup re-
gion. The step size then increases to 2 mm up till the 50th dose deposition point and
then further increases to 4 mm elsewhere. The corresponding step size in the simu-
lated profiles is consistent throughout the depth at 1 mm. Due to this difference in
the total number of measured and simulated dose deposition points, the direct use
of (3.3) was not possible. To overcome this problem, both measured and simulated
scatter plots were fitted to a continuous function.
This fitting function used to plot continuous curves using the measured and sim-
ulated profiles is given by (3.4). This fitting formula used is an improvement over
the proposed formula in Wierzbicki, Sixel, and Podgorsak, 1993, given by Eq. (3.5)
which did not have a linear coefficient (cd).
3.3. Optimization Of Energy 41
F IGURE 3.5: Well fitted curve using the formula given by Eq. 3.4 (left).
The same plot was erroneously fitted +on using the formula given by
Eq. 3.5 (right).
Here d is the depth, f (d) is the dose at depth d and a1 , a2 , b1 , b2 and c are the
fitting coefficients.
The formula described by Eq. (3.5) was unable to determine acceptable fitting
parameters for numerous simulated and measured profiles as shown in Fig. 3.5.
The "curve_fit" function in the "optimize" class of Python 2.7 library Scipy was used
for all curve fittings.
The initial cost values for field sizes 2x2, 5x5, 10x10 and 20x20 cm2 are presented
in the results section (Fig. 3.9).
cm2 is presented in Fig. 3.7 where the cost was calculated to be a relatively small
value of 0.82%.
3.3. Optimization Of Energy 43
(A) (B)
(C) (D)
(E) (F)
( A ) Adding 0.1 MeV to the energy of all the ( B ) Adding 0.2 MeV to the energy of all the
particles particles
( C ) Adding 0.3 MeV to the energy of all the ( D ) Adding 0.4 MeV to the energy of all the
particles particles
F IGURE 3.8: The PDD profiles generated using cropped PhSps with
different energy perturbations. The corresponding cost values (%)
between the measured PDD profile (green curve) and the simulated
PDD profile (red curve) is presented.
3.3.4 Methodology
Finally, a methodology containing sequential steps to optimize the energy of the
PhSp was developed as part of the complete optimization. This methodology has
been validated for the entire PhSp. The steps are presented below.
1. Add large energy perturbations to all the particles in radially symmetric re-
gions of the original PhSp and generate the corresponding PDD profiles. Gen-
erate a cropped PhSp using an acceptable cropping radius as explained in sec-
tion 3.3.2.
2. Add/subtract energies in steps of 0.1 MeV to all the particles in the cropped
PhSp. Simulate the corresponding PDD profiles as presented in section 3.3.3.
3. Identify the energy perturbation which produces the lowest cost locally by
analyzing the simulation results of all field sizes. The simulations should be
performed for all field sizes to avoid over-fitting for a specific field size. An
over-fitted fix in the context of this study implies that it is producing low cost
values for a specific field size while producing relatively high cost values for
other field sizes. This concept has been explained comprehensively in Babyak,
2004 within the paradigm of regression models.
46 Chapter 3. Optimization of Phase Space File
4. Add/subtract energies in even smaller steps of 0.02 MeV to all the particles in
the cropped PhSp. Simulate the corresponding PDD profiles as presented in
section 3.3.3.
5. Identify the energy perturbation which produces the lowest cost locally by
analyzing the simulation results of all field sizes
6. Steps 4 and 5 can be repeated multiple times, using even smaller perturbations
to identify an optimal energy fix of desired accuracy.
7. Add/subtract the identified optimal fix to the entire PhSp and validate for all
field sizes using the full PhSp.
TABLE 3.1: The initial and final cost values for the optimization of
energy. The simulated PDD profiles were generated using the full
PhSp.
Furthermore, it was observed that the optimization of the energy also improved
the inline lateral profiles. Inline profiles were simulated for the original PhSp and the
optimized PhSp, with optimized energy information for field sizes 2x2, 5x5, 10x10
and 20x20 cm2 . The results are presented in Table 3.2. A different cost calculation
technique was used in case of lateral profiles which is explained in section 3.4.2.
TABLE 3.2: The initial and final cost values for the inline lateral profile
after the optimization of energy for field size of 2x2, 5x5, 10x10 and
20x20 cm2 .
A similar comparison for the crossline profiles showed that the optimization of
energy did not improve the cost values considerably. Moreover, the cost values in-
creased instead of decreasing in case of 5x5 and 20x20 cm2 field size. The results are
presented in Table 3.3.
TABLE 3.3: The initial and final cost values for the crossline lateral
profile before and after the optimization of energy for field size of
2x2, 5x5, 10x10 and 20x20 cm2 .
s
Px2 + Pz2
div = tan α = (3.6)
Py2
F IGURE 3.11: Measured inline lateral profiles for field size 10x10 cm2 .
The fitting function used is a sigmoid which will be explained in sec-
tion 3.4.2.
The dose in lateral profile within the water phantom has a distinct steep climb as
can be seen in Fig. 3.11. This is called the penumbra of the lateral dose profile which
is typically taken as the dose profile between 20% and 80% of the maximum dose. It
is the dose profile at the geometric penumbra of the radiation region. The geometric
penumbra, p geom , is defined as the width of the radiated region where the source is
partly shielded by the collimator, as shown in Fig. 3.12.
The dose profile in this region depends largely, but not exclusively, on the photon
source size, the distance from the source and the source to collimator distance (SCD).
This photon source size in X-ray beam therapy strongly depends on the electron spot
3.4. Optimization of Momentum 51
size which the electron beam makes on the target. Experimental studies on the de-
pendence of the dose profile in the penumbra region and the electron spot size have
been presented in Sterpin et al., 2011; Wang and Leszczynski, 2007. Furthermore,
any change in the divergence of the particles in the PhSp reflects as a change in the
virtual photon source size when traced back along the beam upstream as can be seen
in Fig. 3.13. This also effectively changes the virtual electron spot size. Lastly, as the
divergence is essentially a function of the momentum of the particle, the penumbra
region in the lateral profiles were taken as a measure to optimize the momentum of
the particles inside the PhSp.
1
y= (3.7)
1 + exp(−k ( x − x0 ))
Here k is the slope and x0 is the position of the center of the sigmoid. The cost
functions are defined by (3.8) and (3.9),
k − k̂
cost f or k = × 100 (3.8)
k
x0 − x̂0
cost f or x0 = × 100 (3.9)
x0
Here k and x0 are the fitting parameters for measurements, and k̂ and x̂0 are the
fitting parameters for the simulations.
The difference in the fitting parameters were chosen as the cost function instead
of a simple χ2 function (Eq. (3.3) in section 3.3.1) as the latter only provides an
absolute quantified value of disagreement. However, the difference in the fitting
52 Chapter 3. Optimization of Phase Space File
parameters are more informative. They indicate distinctly the quantified positive or
negative disagreement in the slope and the position of the center of the sigmoid.
The "curve_fit" function in the "optimize" class of Python 2.7 library Scipy was
used for all sigmoid fittings. It takes the set of Cartesian coordinates of the curve as
input and returns fitting parameters (k and x0 ) of a sigmoid function optimized to
follow this curve as shown in Fig. 3.14.
F IGURE 3.14: Two cases of sigmoid fitting with different fitting peri-
ods. The corresponding cost values for k and x0 are shown.
F IGURE 3.15: The slope of the beam divergence line, shown in green,
is perturbed by a factor of µ, transforming it to the beam divergence
line shown in red. The Y-intercept value is maintained at c.
m̂ = µm (3.10)
Here µ is the perturbation factor for the slope of the beam divergence line. The
individual transformation identities for the momentum of the particles to carry out
a perturbation factor µ are presented in this section.
The original quantities are div(r ), m, c, Px , Py and Pz , where div(r ) is the diver-
gence of a particle at a radial distance r, m and c is the slope and the y-intercept
of the linear relationship between the divergence and the radial distance (as can be
seen in Fig. 3.15). Px , Py and Pz are the momentum of the particle, whose divergence
is being transformed in the x, y and z direction. div ˆ (r ), k̂, ĉ, P̂x , P̂y and P̂z are the re-
spective transformed quantities. The beam direction is in -Y and the PhSp is placed
at the XZ plane as presented in Fig. 3.10
3.4. Optimization of Momentum 55
div(r ) = mr + c (3.11)
ˆ (r ) = m̂r + ĉ
div (3.12)
m̂ = µm (3.13)
ĉ = c (3.14)
P̂x Px
= (3.15)
P̂z Pz
2 2 2 2
Px + Py + Pz = P̂x + P̂y2 + P̂z2 = 1 (3.16)
Eq. (3.11) and (3.12) are the straight line function for the original and the trans-
formed divergence with varying radius, r. Eq. (3.13) is the transformation of the
slope of the beam line divergence from m to m̂. Eq. (3.14) maintains the divergence
of the particles in the central region of the PhSp. Eq. (3.15) is the constraint identified
which assures that the momentum of the particles do not twist around the central
axis by maintaining the ratio between Px and Pz . Eq. (3.16) is the normalization
condition of the momentum which needs to be maintained.
The goal is to find the transformation identities to be applied to P̂x , P̂y and P̂z ,
given a momentum perturbation factor µ. Starting with Eq. (3.11) and (3.12)
div(r ) − c
m= (3.17)
r
ˆ
div(r ) − c
m̂ = (3.18)
r
Replacing (3.17) and (3.18) in (3.13)
ˆ = µ(div − c) + c
div (3.19)
Using (3.19) and the original definition for the divergence (3.6)
s
P̂x2 + P̂z2 Px2 + Pz2
= ( µ ( 2
− c ) + c )2 (3.20)
2
P̂y Py
Using (3.21) and (3.22), the equation describing the transformed Py is obtained as
1
P̂y2 = (3.23)
γ
s
Px2 + Pz2
γ( P̂x2 + P̂z2 ) = (µ( − c ) + c )2 (3.25)
Py2
Finally, using (3.24) and (3.26) the equation describing the transformed Px is ob-
tained as
ω2 1
P̂x2 = (1 − ) (3.27)
1 + ω2 γ
The final transformation identities for Px , Py and Pz (as referred from (3.23), (3.26)
and (3.27)) for a transformation m → µm are summarized below
2 ω2 1
P̂x = 2
(1 − ) (3.28)
ω +1 γ
2 1
P̂y = (3.29)
γ
2 1 1
P̂z = 2 (1 − ) (3.30)
ω +1 γ
where
Px
ω=
Pz
3.4. Optimization of Momentum 57
and
s
Px2 + Pz2
γ = (µ( − c ) + c )2 + 1
Py2
Sequence µ
0 1.00000
1 1.00001
2 1.00002
3 1.00003
.... ....
800 1.008
The results are presented in Fig. 3.16 and 3.17. The correlation was identified by
fitting a straight line through the scatter plot using linear regression as represented
by the red line in the aforementioned figures. It can be inferred from the two plots
that the fitting parameters k and x0 have a negative linear correlation with µ. This
implies that as the slope of the beam line divergence is increased, the corresponding
fitting parameters decrease in value. This was a crucial detection in this study. The
fitting parameters could now be steered using different perturbation factors.
58 Chapter 3. Optimization of Phase Space File
F IGURE 3.16: The sigmoid fitting parameter, k, for different slope per-
turbation factors (blue dots). The red line represents the linear fit to
the data.
blocked. This marks the radius of the LMU PhSp within which the particles are of
relevance. The relevance of the particles beyond this radius drops substantially. The
minimum cropping radius should be much greater than this relevant radius as a cau-
tionary measure. The relevant radius gives a close approximation of the minimum
cropping radius, however multiple simulations using the cropped PhSp should be
performed and validated against the profiles generated using full PhSp.
The sigmoid fitting parameters of crossline and inline profiles generated using
the full PhSp are presented in Table 3.7 and 3.8 for field sizes 2x2 and 5x5 cm2 re-
spectively. These values have been compared with the corresponding fitting param-
eters generated using a cropped PhSp of cropping radius 20 mm. The percentage
differences (Diff%), as can be seen in the same table, in all of the cases is acceptable.
This proves that the cropping radius of 20 mm for the field size 2x2 and 5x5 cm2 is
acceptable.
Field Size (cm2 ) (k, x0 )Full PhSp (k, x0 )Cropped PhSp (k, x0 )Diff%
2x2 0.9782,9.418 0.9712, 9.379 0.71%, 0.41%
5x5 0.8724,23.196 0.8742,23.184 0.20%, 0.04%
Field Size (cm2 ) (k, x0 )Full PhSp (k, x0 )Cropped PhSp (k, x0 )Diff%
2x2 0.7892, 9.5179 0.7696, 9.5010 2.48%, 0.17%
5x5 0.7392, 23.335 0.7432, 23.3261 0.54%, 0.03%
3.4. Optimization of Momentum 61
3.4.6 Finding the Optimal Slope Perturbation Factor for Crossline (µcr ) or
Inline (µin )
In this section, the sequential steps followed to find the optimal slope perturbation
factor for particles in the crossline ,µcr or inline direction, µin are presented. The two
perturbation factors µcr and µin are determined using the same technique. For the
purpose of this demonstration, µcr has been determined and the final cost values are
presented. The field size is 10x10 cm2 and a cropped PhSp, with a cropping radius
of 50 mm, has been used.
The technique follows 10 steps which are presented below.
5. A straight line passing through the two medians was then plotted (the diago-
nal red line in Fig. 3.21). This signifies the statistical correlation between the
fitting parameters and the perturbation factor. It is analogous to the linear cor-
relation presented in Fig. 3.16 and 3.17. The µcr , for the field size 10x10 cm2 , is
the intersection between the red line and the green line, where the latter rep-
resents the k value for the measured curve. The intersection is highlighted by
the green dot in Fig. 3.21.
6. The steps (1-5) were repeated for the field size 2x2, 5x5 and 20x20 cm2 and the
respective µcr specific to every field size was determined.
7. The final µcr , which in this example was determined to be 1.00605, was taken
as the average of the four µcr values, specific to each field size, as mentioned
in step 6.
8. A third full PhSp was then perturbed by the determined µcr . 24 instances of
simulations were performed to generate crossline profiles.
costs (after the optimization of energy) for inline profiles were already low, as pre-
sented in Table 3.10, thus the focus was kept on optimizing the crossline profiles
while maintaining the low cost values for inline. ∆Cost is given by Eq. (3.32). ∆Cost
written in green represents a drop in cost and in red represents a gain in cost.
Furthermore, the methodology explained in section 3.4.6 optimizes the PhSp sep-
arately for crossline and inline. This meant that in the end, there were two PhSps,
one was optimized for crossline profiles and the other for inline profiles. Multiple
approaches were formed and tested to merge the two factors. Among them, the
three best performing approaches have been presented in the coming sections.
Initial cost values for all the field sizes are given in Table 3.10 as reference for the
results presented in the coming sections.
TABLE 3.10: Initial cost values for inline and crossline profiles as ref-
erence.
Field Size (cm2 ) Initial Cost for Crossline Initial Cost for Inline
2x2 -22.37% -1.81%
5x5 -24.71% -1.09%
10x10 -15.23% -1.30%
20x20 -12.69% -1.57%
The φ2 approach
This approach uses Eq. 3.33 to find the optimal perturbation factor, µop for every par-
ticle as a function of its radial position in the PhSp plane. It was initially assumed
that the crossline dose profiles are largely dependent on the divergence of the parti-
cles near the crossline. Here φ is the angular position of the particle with respect to
the crossline direction as can be seen in Fig. 3.22 (A). The PhSp is in the XZ plane,
represented by the blue circle.
The technique to find the corresponding µin and µcr remains the same as in sec-
tion 3.4.6. However, in step 8 (3.4.6) when a third full PhSp is simulated, the per-
turbation is carried out using Eq. (3.33) instead of a flat perturbation, as presented
in Fig. 3.22 (B) (the red and the green horizontal line). The final perturbation µop is
represented by the thick blue line in the same figure.
The result is presented in Table 3.12. It can be seen that the cost values for
crossline and inline profiles are much higher than the cost values achieved in sec-
tion 3.4.6. The inline cost increases even when the divergence of the particles within
a local angular distance from the inline direction is not considerably perturbed. This
can be seen by the behavior of the thick blue line in Fig. 3.22 (B) which represents
the different perturbation factors with varying angular position from crossline to
the inline direction. Furthermore, the crossline profiles failed in maintaining their
low cost values even when the slope of the beam line divergence was considerably
perturbed locally. Thus, it is perceived from the results that the dose profile in the
inline does not strictly depend only on the divergence of the local particles. It also
depends on the divergence of the particles located far away. This dependence how-
ever decreases on going further away from the inline direction. The same could be
inferred for the crossline. This proves that the initial assumption was wrong. A
more intuitive explanation is presented in Fig. 3.22 (A). The PhSp is represented as
the blue circle in the XZ plane. The relevance of the divergence of the particles for
the crossline profile is represented as different shades of pink. The more intense the
color the higher the relevance.
(A)
(B)
TABLE 3.12: Cost values for different combinations of µ0cr and µin 0 .
∆Cost written in green represents a drop in cost and in red represents
a gain in cost.
TABLE 3.13: Final cost values of the best performing approach, com-
pensated φ6 , for field sizes 2x2, 5x5, 10x10 and 20x20 cm2 using the
full PhSp. ∆Cost written in green represents a drop in cost and in red
represents a gain in cost.
TABLE 3.14: Final cost values of the crossline profiles for the best per-
forming approach, compensated φ6 , generated for field sizes 2x2, 5x5,
10x10 and 20x20 cm2 using the full PhSp as compared to the initial
cost values before the optimization of energy and momentum. The
initial cost values before the optimization of energy can be obtained
from Table 3.3. ∆Cost written in green represents a drop in cost and
in red represents a gain in cost.
TABLE 3.15: Final cost values of the inline profiles for the best per-
forming approach, compensated φ6 , generated for field sizes 2x2, 5x5,
10x10 and 20x20 cm2 using the full PhSp as compared to the initial
cost values before the optimization of energy and momentum. The
initial cost values before the optimization of energy can be obtained
from Table 3.2. ∆Cost written in green represents a drop in cost and
in red represents a gain in cost.
3.5 Discussion
A PhSp is essentially an enormous matrix of interconnected numerical values. These
values define the physical properties of the particles it stores. The PhSp is character-
istic to the linac MC model which generates it. In this study an attempt was made to
manipulate a PhSp of one linac model to optimize it for another linac model.
The complete optimization of the PhSp involved optimizing 4 physical quan-
tities, the energy and the momentum in the three directions. A technique named
cropping was introduced in which the phase space file was cropped into smaller ra-
dially symmetric circles. It was observed that this technique was powerful to fasten
the estimation of the optimal perturbation factors for both energy and momentum.
The optimization of energy was straightforward and less challenging than the opti-
mization of momentum. For the energy optimization, a simple addition/subtraction
operation is necessary. In this study, it was found that an optimal addition of 0.32
MeV should be applied to all photons inside the original PhSp (correspondent to the
IAEA Elekta Precise PhSp), in order to adapt this for modeling the Elekta Synergy
linac. This optimal perturbation decreases the PDD cost values to a maximum of
1.39% for all field sizes. Moreover, the agreement between measured and simulated
lateral profiles in the inline directions increased after the energy of the particles was
optimized. Fig. 3.26 shows lateral profiles for the 2x2 cm2 field in the inline direc-
tion, as simulated with the original PhSp (left) and with the energy-optimized PhSp
(right), in comparison to the measured lateral profile. The absolute cost values for k
and x0 decreased from 7.86% to 1.09% and from 1.41% to 0.93%, respectively.
(A) (B)
F IGURE 3.26: (A) Inline dose profile for a field size 2x2 cm2 before the
optimization of energy. (B) Inline dose profile for a field size 2x2 cm2
after the optimization of energy and momentum. The profile after the
optimization of energy has a lower cost value.
It was also observed that the percentage depth dose profiles retained their low
cost values after the momentum was optimized as can be seen in Fig. 3.27.
The optimization of the momentum, however, was very demanding. A distinct
mathematical relationship between the sigmoid fitting parameters of the lateral pro-
files and the slope of the beam line divergence was detected. This provided a di-
rection on how to manipulate the divergence of the particles. Due to statistical un-
certainties, a minimum number of simulation instances for the same configuration
was estimated, and in the present work 24 instances were used. The median of
the fitting parameters was then taken as the final result of that simulation. Accept-
able cropping radii were determined for all field sizes to improve the efficiency of
3.5. Discussion 73
(A) (B)
F IGURE 3.27: (A) Percentage Depth Dose profile for a field size 10x10
cm2 before the optimization of momentum. (B) Percentage Depth
Dose profile for a field size 10x10 cm2 after the optimization of mo-
mentum.
It can be inferred from the results that a smaller cropping radius of 30 mm and
lateral profiles for 3 field sizes were sufficient to determine the optimal perturbation
74 Chapter 3. Optimization of Phase Space File
factor with acceptable precision. The time taken for estimating the perturbation fac-
tors was reduced by half when a cropping radius of 30 mm was taken. Furthermore,
it was observed that a cropping radius of 20 mm was acceptable for determining
the cost values of the profiles for field size 2x2 and 5x5 cm2 but was not sufficient
in finding out the optimal perturbation factors. A cropping radius of 5 mm is not
acceptable in calculating absolute cost values for 2x2 cm2 . It was however, useful in
testing theories quickly, similar to the study presented in section 3.4.4. To conclude
the discussion, it is crucial to mention that measurements of the dose profiles in mul-
tiple angular directions between the crossline and the inline on the lateral plane are
required for further investigation. These measurements taken from the linac will
assist in determining the slope of the beam line divergence as it transitions from the
crossline direction to the inline.
75
Chapter 4
EPID-based QA Tool
Microsoft Visual Studio was used for all C++ developmental needs. It is an inte-
grated development environment from Microsoft Corporation, Redmond, Washing-
ton, US (Microsoft, 2018).
Two applications written in C++ were developed by the research team of the par-
ent paper as part of the complete project, the automatic FrameGrabber and the iCOM
logger. Once initialized, the FrameGrabber continuously acquires frames from the
EPID regardless of whether the EPID is being radiated or not. Each frame is tagged
with a timestamp in nanosecond resolution. The iCOM logger once initialized, gen-
erates a separate time log file containing many details including the gantry angle,
the dose rate value, the position of the collimator assembly and the time in nanosec-
onds as soon as the beam is turned on. The information in this log file is used to
associate the gantry position to the corresponding EPID frame acquired using the
time-stamp as the stitching link.
All measurements were performed on Elekta Synergy
R
linac coupled to Elekta
AgilityTM collimator (Elekta, Stockholm, Sweden) installed at Klinikum der Uni-
versität München - Campus Großhadern, Munich, Germany, using 6 MeV photon
beam. A PerkinElmer XRD 1640 AL5 P aSi EPID (Elekta iViewGT) mounted at
source to surface distance (SSD) of 160 cm was used for all frame acquisitions. A
self-contained computer system, from hereon referred to as the EPID computer, was
used for the EPID frame acquisitions.
A total of 22 VMAT treatment plans were audited using NEAT. These included
treatment plans for tumors located in different anatomical regions of the body such
as head and neck, prostate, esophagus, lymph nodes, cervix, scalp, vagina and
bronchi. The complete process of each audit consisted of three sequential steps
broadly, which are explained in this section.
3D Dose Reconstruction
Once the frames of the treatment plan are acquired, the 3D dose reconstruction is
performed using NEAT as the interface, and the reconstruction algorithm provided
by the research team of the parent paper, as the back-end code base. The first step
4.1. Materials and Methods 77
F IGURE 4.1: The accurate positioning of the EPID before the treat-
ment plan is performed as required. The center of the EPID is posi-
tioned along the isocenter axis at a depth of 160 cm from the X-ray
source. The green field is the laser guide which assists in estimating
the photon beam field position. The isocenter axis passes through the
intersecting point of the faint red cross.
in the reconstruction algorithm is to convert the pixel values in the frames to dose
values as imparted to a water equivalent medium, using predetermined conversion
factors and a set of correction procedures. The dose values thus obtained are back-
projected upstream at different depths. A series of mathematical operations are then
applied to this conglomeration of 2D planar distributions to obtain a 3D dose distri-
bution. The complete reconstruction technique has been explained comprehensively
in Alhazmi et al., 2018.
The interface of NEAT removes the need of dealing with the low level code. The
offset is loaded by clicking on Choose offset. This is usually the first frame in the treat-
ment plan, acquired without any radiation. It is used to filter out noise from the rest
of the frames. On clicking Choose offset a browse window of the file explorer pops
up. The offset frame is now chosen. Once the offset is successfully loaded the OFF-
SET in the STATUS wire-frame turns from EMPTY to LOADED. On clicking the Start
Reconstruction another browse window of the file explorer pops up. All the frames
required for reconstruction of the treatment plan are now selected by the user. These
include all the frames acquired while frames acquisition of a single treatment plan.
Once the frames are successfully loaded, the reconstruction starts. After the recon-
struction is complete, a Digital Imaging and Communications in Medicine (DICOM)
file is ready to be viewed or saved. DICOM is an international standard to transmit,
store, retrieve, print, process, and display medical imaging information (Sage et al.,
2007; DICOM Standard 2018). The DICOM in the STATUS wireframe turns from AB-
SENT to READY. It can be viewed through Show Reconstruction, as shown in Fig. 4.4.
Alternatively, the file can can be stored using Save DICOM. On clicking Save DICOM
a text field appears to type in the desired name of the new DICOM file, as shown in
Fig. 4.5. The interface also suggests a name, created using the current date and time.
The DICOM file is then saved at the same path as the Frames Acquisition Directory.
78 Chapter 4. EPID-based QA Tool
(A) (B)
F IGURE 4.2: (A) The Elekta iViewGT computer, enclosed in the blue
rectangle, receives the frames during the treatment. (B) The frame ac-
quisition cable is redirected to the EPID computer, shown in the pic-
ture, from the Elekta iViewGT computer, thus effectively redirecting
the acquired frames to the self-contained EPID computer.
F IGURE 4.4: Once the reconstruction is complete the DICOM file can
be viewed by clicking on Show Reconstruction.
F IGURE 4.5: Once the reconstruction is complete the DICOM file can
be saved by clicking on Save DICOM.
80 Chapter 4. EPID-based QA Tool
Gamma Evaluation
In the final step a 3D gamma index evaluation, with the dose distribution calculated
by the Treatment Planning System (TPS), was performed for all the reconstructed 3D
dose distributions. The gamma index method was proposed and evaluated by Low
et al., 1998 in details, though it is summarized in this section for clarity. The gamma
index is used in determining the degree of agreement between a calculated and a
reference dose distribution and is given by Eq. (4.1).
r
r2 (~rc ,~rr ) δ2 (~rc ,~rr )
Γ(~rc , ~rr ) = + , γ(~rr ) = min{Γ(~rc , ~rr )} ∀ {~rc } (4.1)
∆d2 ∆D2
r (~rc ,~rr ) = |~rc −~rr | (4.2)
δ(~rc ,~rr ) = D (~rr ) − D (~rr ) (4.3)
Here, ~rc and ~rr are the radial vectors of all the points in the volume for the cor-
responding calculated 3D dose distribution and the reference 3D dose distribution.
r (~rc ,~rr ) is given by Eq. (4.2) and δ(~rc ,~rr ) is given by Eq. (4.3), where D (~r ) is the dose
deposited at the point ~r. For a certain distance to agreement (DTA) criteria, ∆d, and
a certain dose difference criteria, ∆D, the pass-fail decision is given by:
Here, γ(~rr ) is calculated using (4.1). These values are determined for every point
in the dose volume for the calculated dose distribution against the reference dose
distribution. In this study the result is given as a passing rate of the gamma criteria
(3 mm, 3 %), where 3 mm is the DTA and 3% is the dose difference criteria. This
represents the cumulative passing rate of all the points in the dose volume.
Gamma evaluation is not supported by NEAT and was computed on a sepa-
rate system using VeriSoft
R
Patient Plan Verification Software, PTW, Freiburg, Ger-
many. This software offers multiple tools for dose comparison ranging from basic
visual comparison to detailed quantitative evaluation (PTW, 2018b). The results are
presented in section 4.2.
4.2. Final Results 81
In this study 22 VMAT treatment plans were performed and evaluated. Before
the frame acquisition, the EPID is positioned manually such that its center point
intersects with the beam central line as shown in Fig. 4.1. This manual interven-
tion introduces errors in the alignment of the reconstructed dose positions as shown
in Fig. 4.7 (A). This shift is in the inline direction, also known as the target-gantry
(TG) direction. To determine a correction factor for this error, the first set of frame
acquisitions is performed for a square shaped beam with a field size of 10x10 cm2
without any linac rotation. These frames are reconstructed and compared with the
corresponding TPS calculations. If the 1D dose profile, for this reconstruction, in the
coronal plane is misaligned, then the reconstructed dose distribution is shifted by
a factor which aligns both the profiles. This factor is the aforementioned correction
factor in TG direction, hereby referred to as the TG factor. The coronal plane is the
82 Chapter 4. EPID-based QA Tool
plane normal to the beam central line. In this study, this TG factor is determined to
be 3.6 mm in the -Z direction as shown in Fig. 4.1. As a part of this audit process, it
was essential that such a correction factor is taken into consideration as an accurate
positioning of the EPID was a pre-requisite for the reconstruction algorithm. Fur-
thermore, the reconstructed dose distribution for the 10x10 cm2 field was also used
to determine a recalibration or normalization factor (NF). If in a certain case, the ab-
solute dose values of the reconstructed dose distribution are considerably higher or
lower than the corresponding TPS calculations, then the reconstructed dose values
are normalized using an appropriate NF. This factor, when applied to the recon-
structed dose distribution shifts the dose values, while retaining its complex shape
as shown in Fig. 4.7 (B) and (C). The NF determined using the 10x10 cm2 field size
reconstruction was 3%.
(A)
(B) (C)
F IGURE 4.7: (A) A comparison of the 1D dose profile of the TPS cal-
culation (drawn in orange) and the reconstruction (drawn in blue) in
the coronal plane for an exemplary treatment plan. (B) A misalign-
ment in the TG direction is corrected and the blue plot shifts to the
right by 3.6 mm. This represents the TG factor. (C) The misalignment
in the absolute dose values is corrected and the blue plot is raised by
3%. This represents the NF.
4.2. Final Results 83
The values for gamma criteria (3%, 3 mm) are presented in Table 4.1. "NF" stands
for the normalization factor used in the evaluation and "TG corrected" implies that
it has been fixed for the error in alignment. In this audit, a passing rate higher than
a threshold of 95% will be considered satisfactory, which is highlighted in green.
The color yellow implies a passing rate between 80% and 95% and the color red
implies a passing rate lower than 80%. It can be seen from the results that without
normalization the reconstruction has very low passing rates with only a third of the
total number of treatment plans achieving more than 95%. On employing the 3% NF,
the passing rates improve considerably, with 13 out of 22 treatment plans passing the
threshold. It was observed that for few reconstructions a higher normalization factor
produces much better passing rate.
TABLE 4.1: (3%, 3 mm) gamma index evaluation for the NEAT recon-
structed dose distribution for 22 treatment plans. NF stands for nor-
malization factor. TG corrected implies that the reconstruction has
been shifted by 3.6 mm in the -Z direction to correct for the EPID
positioning discrepancy. Passing rates highlighted in green represent
values greater than 95%. Yellow represents passing rates between
80% and 95%. Red represents passing rates below 80%. TP stands
for Treatment Plan. The reconstructions which had achieved a pass-
ing rate higher than the threshold was not corrected further using NF
8%, as a higher NF produced a worse passing rate.
4.3 Discussion
For this audit, a GUI, named Novel EPID-based Audit Tool (NEAT) has been de-
veloped. It was created as a clinical tool for reconstructing dose distributions as
imparted to a water phantom, using in air-frames acquired by EPID. Furthermore,
NEAT was used to reconstruct the 3D dose distributions of 22 VMAT treatment
plans. The reconstructed dose distributions were then compared with the TPS cal-
culated dose distributions using a gamma index evaluation with a criteria of (3%, 3
mm). The result has been presented in Table 4.1.
Along with the frame acquisitions of the treatment plans, 10x10 cm2 field size
frames were also acquired in air. The reconstructed dose distribution for this ra-
diation was compared to the corresponding TPS calculated dose distribution. Two
corrections factors were identified using this comparison. First was the TG factor, to
compensate for the EPID positioning error, introduced during the manual position-
ing of the EPID before the acquisition of the frames, as can be seen in Fig. 4.1. The
second was the normalization factor (NF). It can be expected that the dose response
of the EPID changes over time. This factor should compensate for any drop in the
response. It should be noted that the reconstruction algorithm, as presented in the
parent paper, Alhazmi et al., 2018, has its own extensive calibration matrix, deter-
mined by taking a myriad of readings from the linac, which are then processed to
determine the values in the calibration matrix. Although, this complete procedure
takes few hours and cannot be practically performed each day a dose distribution is
reconstructed. Thus, it is recommended in the audit to perform the 10x10 cm2 field
size radiation and reconstruct the dose distribution with the frames acquired. The
two factors should then be determined and employed during the gamma evaluation.
This acquisition and reconstruction takes only few minutes but the reconstruction
improves considerably.
The TG factor remains the same for all dose distributions, however the ideal
NF does not remain the same. This should be investigated further to improve the
consistency of the algorithm.
Furthermore, the 5 best performing reconstructions were tested against different
gamma criterias, (2%, 2 mm) and (1%, 1 mm). The results are presented in Table 4.2.
4 out of the 5 reconstructions were able to pass the strict criteria of (2%, 2 mm), how-
ever all were unable to reach the threshold in case of a stricter (1%, 1 mm) criteria.
TABLE 4.2: (2%, 2 mm) and (1%, 1 mm) gamma index evaluation for
the NEAT reconstructed dose distribution for the the 5 best perform-
ing reconstructions. Passing rates highlighted in green represent val-
ues greater than 95%. Yellow represents passing rates between 80%
and 95%. Red represents passing rates below 80%.
It can be concluded that, with the appropriate normalization factor, the recon-
struction algorithm was able to produce exceedingly high passing rates for all cases.
4.3. Discussion 85
However, this normalization factor must be determined using a more reliable tech-
nique which requires further investigation.
87
Chapter 5
5.1 Conclusion
Optimization of PhSp
In radiation dosimetry, Monte Carlo techniques offer a powerful tool for modeling
radiation transport. If accurately implemented, a MC model is exceedingly precise
in the determination of the absorbed dose. However, accurate modeling of medical
linear accelerators is often a daunting task. The internal components must be known
in details for an accurate MC modeling. If these details are not known, it becomes
almost impossible to model the linac.
In this study an alternative approach to this problem was researched. Commis-
sioned PhSp files of a validated linac model were optimized for another linac with-
out the need of detailed knowledge of the internal components. A technique referred
to as cropping was introduced. It turned out to be a powerful technique without
which fast experimental validations of theories would not have been possible. It
was discerned that the optimization works well with energy. An optimal fix of 0.32
MeV was found. This fix was then applied to the full PhSp and PDD profiles were
generated for all field sizes considered for validation. The cost values for PDD, using
the PhSp with optimized energy were all under 1.39%. The optimization of energy
also improved the cost values of inline lateral profiles. No significant improvement
was observed for crossline profiles.
The optimization of momentum proved to be relatively more challenging. A dis-
tinct relationship between the sigmoid fitting parameters of the lateral profiles and
the slope of the beam line divergence was detected and studied. This provided a
good guidance on how to manipulate the particle’s momentum. It was observed that
multiple simulation instances of the same configuration produced varying results.
An investigatory study was performed to estimate the least number of simulations
needed to determine a fairly stable fitting parameter. Utilizing the results from these
sub-studies, slope perturbation factors for crossline and inline were determined in-
dividually. Furthermore, a general methodology was developed to determine slope
perturbation factors. Individually the optimization of momentum for the crossline
and inline produced low cost values. Implementing the two fixes in the same PhSp
has proven to be difficult. Multiple approaches were formulated, implemented and
then analyzed. Three approaches were presented along with their results. The best
performing approach was then used to determine the optimal perturbation factors
which produced the lowest cost values. The full PhSp was perturbed using these
factors and lateral profiles were generated for all the considered field sizes and the
corresponding cost values were determined. The average drop in the cost values,
as compared to the initial cost values before the optimization of momentum and af-
ter the optimization of energy, was determined to be around 3.58% for all profiles.
88 Chapter 5. Conclusion and Outlook
The cost values were compared to the initial cost values, determined before the opti-
mization of energy. The average drop in the cost values was evaluated to be around
7.10%.
EPID-based QA Tool
EPID is conventionally used in patient positioning, owing to its sub-mm resolution
and fast frame acquisition capability. However, numerous works have been pub-
lished on utilizing the EPID in the pre-treatment dosimetric verification of treatment
plans. One such work has been published by Alhazmi et al., 2018 which presents
a robust algorithm to reconstruct 3D dose distributions. The EPID frames are ac-
quired in air, into a self-contained computer, while the treatment plan is being per-
formed using the linac without the patient or a phantom. These frames are then
used to reconstruct 3D dose distributions as imparted to a water equivalent cylin-
drical medium using the reconstruction algorithm, described comprehensively in
the paper.
In this study, a GUI named Novel EPID-based Audit Tool (NEAT) was devel-
oped as a clinical tool to reconstruct the 3D dose distributions using these algo-
rithms. It facilitates the implementation of the algorithm in clinical routine, as it
enables the user to easily perform the full process from acquisition to reconstruction
without dealing with the low level code. 22 VMAT treatment plans were acquired
and the corresponding 3D dose distributions were reconstructed using NEAT. These
dose distribution were then compared with the corresponding TPS calculated dose
distribution using a gamma evaluation of criteria (3%, 3 mm). Two correction fac-
tors, Target Gantry (TG factor) and normalization factor (NF), were determined us-
ing a reconstruction corresponding to a square shaped beam with a field size of
10x10 cm2 without any linac rotation. These factors were employed while perform-
ing the gamma evaluation. Out of 22 reconstructions, 13 passed the threshold with
the gamma criteria of (3%, 3 mm). It was observed that a higher NF improved the
pass percentages of the failing reconstructions. Further investigation is required to
form a reliable technique which determines the ideal NF for every reconstruction.
5.2 Outlook
In the past decades, MC Techniques have presented powerful numerical solutions,
superior to analytical ones in the field of X-ray photon therapy. Yet, to exploit these
techniques, an accurate MC model is required. The PhSp optimization techniques
introduced and analyzed in this study offers an alternative approach to model the
linac when detailed information of its internal components are not known.
Automated engine
An automated engine can be implemented for unsupervised optimization. It is con-
ceivable that all the individual steps followed in the optimization of the PhSp can
be stitched together and automated. Such an engine should be able to run indepen-
dently using correlations detected in this study. Currently, many of the steps are
initialized manually as the framework is still in its rudimentary form.
trained using the aforementioned training data set. Such a trained model should
be capable of identifying numerous parameters which are too subtle to be noticed
otherwise. This model should take the MLC/diaphragm plan as the input, in the ap-
propriate format, and should produce a 3D matrix containing the dose distribution.
This is an ambitious yet exciting field of research which might have the potential to
challenge the efficiency of MC simulations.
NEAT 2.0
The current version of NEAT is rudimentary and has a considerable scope of im-
provement. A gamma evaluation section can be added to the interface which com-
pares the reconstructed DICOM file to a reference file. The Frames Acquisition Direc-
tory text box can be changed into a drop down box displaying the last 5 selected file
paths. An optional auto stop button can be installed which automatically identifies
when the the linac beam is no longer on and stops the acquisition, at the same time
popping a notification. This would make the interface more independent. An op-
tional auto-start reconstruction button can be installed which starts the reconstruc-
tion automatically after the frames have been acquired. A log file generator for the
interface can be implemented which creates a log file, containing every information
of all the interface activities. This should also be helpful in archiving the treatment
informations which is a critical issue in clinical routines. Thus an improved version
of NEAT, NEAT 2.0 is a potential project for any future work.
The complete research presented in the first part of this thesis was an attempt at
contributing to the understanding of the gross and subtle correlations between the
properties of photons in the Phase Space and the resulting dose profiles. The second
part of the thesis was an attempt at improving the efficiency of the QA procedure in
a clinical routine of radiotherapy by developing a GUI. While this thesis concludes
here, the research on the understanding of the aforementioned correlations and on
the improvement of EPID dosimetry has much to offer.
91
Bibliography
Ahnesjö, Anders, Mikael Saxner, and Avo Trepp (1992). “A pencil beam model for
photon dose calculation”. In: Medical physics 19.2, pp. 263–273.
Ahnesjö, A., P. Andreo, and A. Brahme (1987). “Calculation and Application of Point
Spread Functions for Treatment Planning with High Energy Photon Beams”. In:
Acta Oncologica 26.1, pp. 49–56. DOI: 10.3109/02841868709092978. URL: https:
//doi.org/10.3109/02841868709092978.
Alhazmi, Abdulaziz et al. (2018). “A novel approach to EPID-based 3D volumetric
dosimetry for IMRT and VMAT QA”. In: Physics in Medicine and Biology 63.11,
p. 115002.
Aljarrah, Khaled et al. (2006). “Determination of the initial beam parameters in Monte
Carlo linac simulation”. In: 33, pp. 850–858. URL: https://doi.org/10.1118/1.
2168433.
Attix, Frank Herbert (1986). Introduction to Radiological Physics and Radiation Dosime-
try. WILEY-VCH Verlag GmbH and Co. KGaA. ISBN: 978-3-527-61714-2. URL:
https://www.wiley.com/en- us/Introduction+to+Radiological+Physics+
and+Radiation+Dosimetry-p-9783527617142.
Babyak, Michael A. (2004). “What You See May Not Be What You Get: A Brief,
Nontechnical Introduction to Overfitting in Regression-Type Models”. In: Psy-
chosomatic Medicine 66 (3), pp. 411–421. URL: https : / / journals . lww . com /
psychosomaticmedicine/Fulltext/2004/05000/What_You_See_May_Not_Be_
What_You_Get__A_Brief,.21.aspx.
Baskar, Rajamanickam et al. (2012). “Cancer and radiation therapy: current advances
and future directions”. In: International journal of medical sciences 9.3, p. 193.
Bednarz, Bryan and X George Xu (2009). “Monte Carlo modeling of a 6 and 18 MV
Varian Clinac medical accelerator for in-field and out-of-field dose calculations:
development and validation.” In: Physics in Medicine and Biology 54 (4), N43–57.
URL : https://www.ncbi.nlm.nih.gov/pubmed/19141879.
Bencheikh, Mohamed et al. (2017). “Validation of Monte Carlo simulation of 6 MV
photon beam produced by Varian Clinac 2100 linear accelerator using BEAM-
nrc code and DOSXYZnrc code”. In: Physics of Particles and Nuclei Letters 14.5,
pp. 780–787.
Bethe, H. A. and L.C. Maximon (1954). “Theory of Bremsstrahlung and Pair Pro-
duction. I. Differential Cross Section”. In: Phys. Rev. 93 (4), pp. 768–784. DOI:
10.1103/PhysRev.93.768. URL: https://link.aps.org/doi/10.1103/PhysRev.
93.768.
Bidmead, Margaret and Jean-Claude Rosenwald (2007). Handbook of radiotherapy physics:
theory and practice. CRC Press. Chap. 35.
Bielajew, Alex F (1989). Improved angular sampling for pair production in the EGS4 code
system. Institute for National Measurement Standards, NRCC. URL: https : / /
www.researchgate.net/publication/44054016_Improved_angular_sampling_
for_pair_production_in_the_EGS4_code_system.
Bortfeld, Thomas (2006). “IMRT: a review and preview”. In: Physics in Medicine &
Biology 51.13, R363.
92 BIBLIOGRAPHY
0360-3016. DOI: https : / / doi . org / 10 . 1016 / 0360 - 3016(81 ) 90379 - 5. URL:
http://www.sciencedirect.com/science/article/pii/0360301681903795.
Lin, Sung-Yen, Tieh-Chi Chu, and Jao-Perng Lin (2001). “Monte Carlo simulation of
a clinical linear accelerator.” In: Applied Radiation and Isotopes 55 (6), pp. 759–765.
URL : https://doi.org/10.1016/S0969-8043(01)00130-0.
Little, John B. (2003). Principal Cellular and Tissue Effects of Radiation. 6th ed. BC Decker,
Canada. ISBN: 10: 1-55009-213-8. URL: https://www.ncbi.nlm.nih.gov/books/
NBK12344/.
Liu, HH., TR Mackie, and EC. McCullough (1997). “Calculating output factors for
photon beam radiotherapy using a convolution/superposition method based on
a dual source photon beam model.” In: Medical Physics 24 (12), pp. 1975–1985.
URL : https://www.ncbi.nlm.nih.gov/pubmed/9434980.
London Health Sciences Centre (2011). URL: http://www.lhsc.on.ca/About_Us/LHSC/
Publications/Features/Cobalt-60.htm.
Loverock, Les (2007). Handbook of radiotherapy physics: theory and practice. CRC Press.
Chap. 11.
Low, Daniel A et al. (1998). “A technique for the quantitative evaluation of dose
distributions”. In: Medical physics 25.5, pp. 656–661.
Low, Daniel A. et al. (2011). “Dosimetry tools and techniques for IMRT”. In: Medical
Physics 38.3, pp. 1313–1338. DOI: 10 . 1118 / 1 . 3514120. URL: https : / / aapm .
onlinelibrary.wiley.com/doi/abs/10.1118/1.3514120.
Ma, C-M Charlie (2014). Measurement of Radiation: Therapy Physics Review Course.
AAPM.
Martins, Juliana Cristina et al. (2017). “An equipment-specific Geant4 model for the
Elekta Agility collimator”. In: Physica Medica: European Journal of Medical Physics
42 (1), 30–31. URL: https://doi.org/10.1016/j.ejmp.2017.09.076.
MATLAB (2018). The MathWorks, Inc. URL: https://ch.mathworks.com/products/
matlab.html.
McCall, Richard C., Raymond D. McIntyre, and William G. Turnbull (1978). “Im-
provement of linear accelerator depth-dose curves”. In: The Journal of Applied
Clinical Medical Physics 5 (6), pp. 518–524. URL: https : / / doi . org / 10 . 1118 /
1.594487.
Microsoft (2018). Visual Studio. URL: https://visualstudio.microsoft.com/de/
?rr=https%3A%2F%2Fwww.google.de%2F.
Mohan, R., C. Chui, and L. Lidofsky (1985). “Energy and angular distributions of
photons from medical linear accelerators.” In: Medical Physics 12 (5), pp. 592–597.
URL : https://www.ncbi.nlm.nih.gov/pubmed/4046993.
Monaco (2018). Elekta. URL: https : / / www . elekta . com / software - solutions /
treatment-management/external-beam-planning/monaco/.
Mooney, Christopher Z (1997). Monte carlo simulation. Vol. 116. Sage Publications.
Mould, R F (1995). “The early history of X-ray diagnosis with emphasis on the con-
tributions of physics 1895-1915”. In: Physics in Medicine and Biology 40.11, p. 1741.
URL : http://stacks.iop.org/0031-9155/40/i=11/a=001.
Mubata, Cephas (2007). Handbook of radiotherapy physics: theory and practice. CRC
Press. Chap. 14.
Munro, P., J. A. Rawlinson, and A. Fenster (1988). “Therapy imaging: Source sizes of
radiotherapy beams”. In: The International Journal of Medical Physics Research and
Practice 15 (4), pp. 517–524. URL: https://doi.org/10.1118/1.596295.
Nagashima, Yorikiyo (2010). Elementary Particle Physics Volume 1: Quantum Field The-
ory and Particles. Wiley-VCH. ISBN: 9783-527-40962-4.
BIBLIOGRAPHY 95
Nascimento, Marcio Luis Ferreira (2014). “Brief history of X-ray tube patents”. In:
World Patent Information 37, pp. 48–53. URL: https://www.sciencedirect.com/
science/article/pii/S0172219014000374.
Nath, Ravinder et al. (1994). “AAPM code of practice for radiotherapy accelerators:
Report of AAPM Radiation Therapy Task Group No. 45”. In: Medical Physics 21.7,
pp. 1093–1121.
Nijsten, SMJJG et al. (2007). “A global calibration model for EPIDs used for transit
dosimetry”. In: Medical physics 34.10, pp. 3872–3884.
Nwankwo, Obioma et al. (2017). “A single-source photon source model of a linear
accelerator for Monte Carlo dose calculation”. In: PloS one 12.9, e0183486.
Palma, David et al. (2008). “Volumetric Modulated Arc Therapy for Delivery of
Prostate Radiotherapy: Comparison With Intensity-Modulated Radiotherapy and
Three-Dimensional Conformal Radiotherapy”. In: International Journal of Radia-
tion Oncology*Biology*Physics 72.4, pp. 996 –1001. ISSN: 0360-3016. DOI: https :
//doi.org/10.1016/j.ijrobp.2008.02.047. URL: http://www.sciencedirect.
com/science/article/pii/S0360301608003428.
Parodi, Katia (2017). Lecture notes in Medical Physics in Radiation Therapy. URL: https:
//www.physik.uni-muenchen.de/lehre/vorlesungen/wise_17_18/Medical-
physics-in-radiation-therapy/vorlesung/index.html.
PerkinElmer (2005). Brochure. URL: http://static6.arrow.com/aropdfconversion/
a071b1ff22ab513723dabf2aac4e256eacea2328/ds_pkiasi_1640_al_ag.pdf.
Podesta, Mark et al. (2014). “Time dependent pre-treatment EPID dosimetry for stan-
dard and FFF VMAT”. In: Physics in Medicine & Biology 59.16, p. 4749.
Podgorsak, Ervin B. (2005). Radiation Oncology Physics. Vienna: INTERNATIONAL
ATOMIC ENERGY AGENCY. URL: http://www-pub.iaea.org/books/IAEABooks/
7086/Radiation-Oncology-Physics.
— (2010). Radiation Physics for Medical Physicists. Springer. ISBN: 9783642008757. URL:
http : / / www . bookmetrix . com / detail / book / 33b356ea - ce7b - 4adc - acb8 -
3a74922cad06#citations.
PTW (2018a). microDiamond Synthetic Diamond Detector. URL: https://www.ptw.de/
2733.html?&cId=6851.
— (2018b). VeriSoft
R
. URL: https://www.ptw.de/2406.html.
Reed, Amy B (2011). The history of radiation use in medicine. URL: https://doi.org/
10.1016/j.jvs.2010.07.024.
Rezaei-Ochbelagh, D and S Azimkhani (2012). “Investigation of gamma-ray shield-
ing properties of concrete containing different percentages of lead”. In: Applied
Radiation and Isotopes 70.10, pp. 2282–2286.
Robison, Roger Frank (1995). “The Race For Megavoltage X-Rays Versus Telegamma”.
In: Acta Oncologica 34.8, pp. 1055–74. URL: https : / / www . researchgate . net /
publication/14592585_The_Race_For_Megavoltage_X-Rays_Versus_Telegamma.
Rodriguez, M et al. (2015). “A geometrical model for the Monte Carlo simulation of
the TrueBeam linac”. In: Physics in Medicine and Biology 60.11, N219–N229. URL:
http://stacks.iop.org/0031-9155/60/i=11/a=N219.
Rosenwald, Jean-Claude (2007). Handbook of radiotherapy physics: theory and practice.
CRC Press. Chap. 39.
Rosenwald, Jean-Claude, Ivan Rosenberg, and Glyn Shentall (2007). Handbook of ra-
diotherapy physics: theory and practice. CRC Press. Chap. 26.
Sage, John P. et al. (2007). Handbook of radiotherapy physics: theory and practice. CRC
Press. Chap. 26.
96 BIBLIOGRAPHY
Sarrut, David et al. (2014). “A review of the use and potential of the GATE Monte
Carlo simulation code for radiation therapy and dosimetry applications”. In:
Medical physics 41.6Part1.
Seco, Joao and Frank Verhaegen (2016). Monte Carlo Techniques in Radiation Therapy.
CRC Press. ISBN: 9781138199903. URL: https : / / www . crcpress . com / Monte -
Carlo-Techniques-in-Radiation-Therapy/Seco-Verhaegen/p/book/9781138199903.
Seuntjens, JP, Wynand Strydom, and KR Shortt (2005). “Dosimetric principles, quan-
tities and units”. In: Radiation oncology physics: a handbook for teachers and students.
Podgorsak EB, editor. Vienna, Austria: IAEA.
Slater, James M. (2012). “From X-Rays to Ion Beams: A Short History of Radiation
Therapy”. In: Ion Beam Therapy: Fundamentals, Technology, Clinical Applications. Ed.
by Ute Linz. Berlin, Heidelberg: Springer Berlin Heidelberg, pp. 3–16. ISBN: 978-
3-642-21414-1. DOI: 10.1007/978-3-642-21414-1_1. URL: https://doi.org/10.
1007/978-3-642-21414-1_1.
Steel, Gordon (2007). Handbook of radiotherapy physics: theory and practice. CRC Press.
Chap. 8.
Sterpin, Edmond et al. (2011). “On the relationships between electron spot size, focal
spot size, and virtual source position in Monte Carlo simulations.” In: Medical
physics 38 3, pp. 1579–86.
Teoh, M et al. (2011). “Volumetric modulated arc therapy: a review of current lit-
erature and clinical use in practice”. In: The British journal of radiology 84.1007,
pp. 967–996.
Wang, Lilie LW and Konrad Leszczynski (2007). “Estimation of the focal spot size
and shape for a medical linear accelerator by Monte Carlo simulation”. In: Medi-
cal physics 34.2, pp. 485–488.
Widder, Joachim (2014). “The origins of radiotherapy: Discovery of biological effects
of X-rays by Freund in 1897, Kienböck’s crucial experiments in 1900, and still it
is the dose”. In: Radiotherapy and Oncology 112.1, pp. 150 –152. ISSN: 0167-8140.
DOI : https://doi.org/10.1016/j.radonc.2014.06.010. URL : http://www.
sciencedirect.com/science/article/pii/S0167814014002680.
Wierzbicki, W, K E Sixel, and E B Podgorsak (1993). “An analytical representation of
radiosurgical depth-dose data”. In: Physics in Medicine and Biology 38.9, p. 1351.
URL : http://stacks.iop.org/0031-9155/38/i=9/a=014.
Yu, Shiqi, Sen Jia, and Chunyan Xu (2017). “Convolutional neural networks for hy-
perspectral image classification”. In: Neurocomputing 219, pp. 88–98.
97
Declaration of Authorship
I hereby declare that this thesis is my own work, and that I have not used any
sources and aids other than those stated in the thesis.
München
Author’s signature