You are on page 1of 10

GE Healthcare

gehealthcare.com

Technical
Publication

Direction 5198244-100
Revision 1
GE Healthcare
Definium AMX 700 Handswitch Holder Upgrade

Copyright © 2007 by General Electric Company, Inc.


All rights reserved.
GE HEALTHCARE DEFINIUM AMX 700 HANDSWITCH HOLDER UPGRADE
REVISION 1 DIRECTION 5198244-100

THIS PAGE LEFT INTENTIONALLY BLANK

Published by the GE Company, Inc.


Engineering/Technical: Dan Beetcher

Page 2
GE HEALTHCARE DEFINIUM AMX 700 HANDSWITCH HOLDER UPGRADE
REVISION 1 DIRECTION 5198244-100

Revision History

Revision Date Reason for change


1 9MAR2007 Initial release of document.

Revision History Page 3


GE HEALTHCARE DEFINIUM AMX 700 HANDSWITCH HOLDER UPGRADE
REVISION 1 DIRECTION 5198244-100

THIS PAGE LEFT INTENTIONALLY BLANK

Page 4 List of Effected Pages


GE HEALTHCARE DEFINIUM AMX 700 HANDSWITCH HOLDER UPGRADE
REVISION 1 DIRECTION 5198244-100

Section 1.0 - Personnel Requirements


Required Persons: 1

Section 2.0 - Preliminary Requirements


2.1 Tools and Test Equipment
Allen wrenches and Drill are required for this procedure.

2.2 FRU Kit Contents (5212522)

PART DESCRIPTION QUANTITY


NUMBER
5212885 Handswitch Holder Spacer 1
5213361 Drilling Template for Handswitch Spacer 1
5198309 #20 Drill Bit 1
5198244-100 Definium AMX 700 Handswitch Upgrade instructions 1
(this document)
46-170498P69 8-32 Socket Head Cap Screw (extra) 2
Table 0-1 FRU Kit Contents

Figure 1 Kit Contents

Page 5
GE HEALTHCARE DEFINIUM AMX 700 HANDSWITCH HOLDER UPGRADE
REVISION 1 DIRECTION 5198244-100

Section 3.0 - Handswitch Replacement Procedure


1.) Power down the system. Flip main breaker to the OFF position. Verify that the system is
unplugged (not charging). Follow standard LOTO (Lock Out Tag Out) procedures.
2.) Use 7/64" allen wrench to remove the cast collar (around tube park latch). See Figure 2.

Figure 2 Remove Cast Collar

3.) Raise the top cover of the AMX unit.

Make sure the support brackets for the top cover lock into place.

4.) Remove the current handswitch holder assembly by removing the (2) cap screws from the
backside of the top cover.
5.) Drill new mounting holes as follows:

Note: If the top cover already has four mounting holes, proceed to Step 6.

a.) Place the drill template in the existing holes in the top cover as shown. See Figure 3.

Figure 3 Place Drill Template

b.) Drill two holes using the #20 drill bit. See Figure 4.

Page 6
GE HEALTHCARE DEFINIUM AMX 700 HANDSWITCH HOLDER UPGRADE
REVISION 1 DIRECTION 5198244-100

Figure 4 Drill Holes

c.) Remove the template.


6.) Confirm the fit of the aluminum spacer (p/n 5212885). It should be a snug/tight fit into the top
cover. See Figure 5.

Figure 5 Install Aluminum Spacer

7.) Re-assemble the handswitch holder and cover (Figure 6 and Figure 7) and tighten the screws
(spare screws are provided in kit). Ensure that the handswitch assembly is tight.

Page 7
GE HEALTHCARE DEFINIUM AMX 700 HANDSWITCH HOLDER UPGRADE
REVISION 1 DIRECTION 5198244-100

Figure 6 Attach Handswitch Holder

Figure 7 Handswitch Holder Completed

8.) Close the top cover and re-install the cast collar.

Page 8