Definium AMX 700 Handswitch Holder Upgrade
Page 2
GE HEALTHCARE DEFINIUM AMX 700 HANDSWITCH HOLDER UPGRADE
REVISION 1 DIRECTION 5198244-100
Revision History
Page 5
GE HEALTHCARE DEFINIUM AMX 700 HANDSWITCH HOLDER UPGRADE
REVISION 1 DIRECTION 5198244-100
Make sure the support brackets for the top cover lock into place.
4.) Remove the current handswitch holder assembly by removing the (2) cap screws from the
backside of the top cover.
5.) Drill new mounting holes as follows:
Note: If the top cover already has four mounting holes, proceed to Step 6.
a.) Place the drill template in the existing holes in the top cover as shown. See Figure 3.
b.) Drill two holes using the #20 drill bit. See Figure 4.
Page 6
GE HEALTHCARE DEFINIUM AMX 700 HANDSWITCH HOLDER UPGRADE
REVISION 1 DIRECTION 5198244-100
7.) Re-assemble the handswitch holder and cover (Figure 6 and Figure 7) and tighten the screws
(spare screws are provided in kit). Ensure that the handswitch assembly is tight.
Page 7
GE HEALTHCARE DEFINIUM AMX 700 HANDSWITCH HOLDER UPGRADE
REVISION 1 DIRECTION 5198244-100
8.) Close the top cover and re-install the cast collar.
Page 8