Professional Documents
Culture Documents
Introduction
Physical Principles
Liquid penetrant inspection depends mainly on a penetrant's effectively wetting the surface of a
solid workpiece orspecimen, flowing over that surface to form a continuous and reasonably uniform
coating, and then migrating into cavitiesthat are open to the surface. The cavities of interest are
usually exceedingly small, often invisible to the unaided eye. Theability of a given liquid to flow
over a surface and enter surface cavities depends principally on the following:
· Cleanliness of the surface
· Configuration of the cavity
· Cleanliness of the cavity
· Size of surface opening of the cavity
· Surface tension of the liquid
· Ability of the liquid to wet the surface
· Contact angle of the liquid
The cohesive forces between molecules of a liquid cause surface tension. An example of the
influence of surface tensionis the tendency of free liquid, such as a droplet of water, to contract into
a sphere. In such a droplet, surface tension iscounterbalanced by the internal hydrostatic pressure of
the liquid. When the liquid comes into contact with a solidsurface, the cohesive force responsible
for surface tension competes with the adhesive force between the molecules of theliquid and the
solid surface. These forces jointly determine the contact angle, θ, between the liquid and the surface
(Fig.1). If θ is less than 90° (Fig. 1a), the liquid is said to wet the surface, or to have good wetting
ability; if the angle is equalto or greater than 90° (Fig. 1b and c), the wetting ability is considered
poor.
Fig. 1 Wetting characteristics as evaluated by the angle, θ, between a droplet of liquid and a solid surface.Good wetting
is obtained when θ< 90° (a); poor wetting, when θ 90° (b) and (c).
Closely related to wetting ability is the phenomenon of capillary rise or depression (Fig. 2). If the
contact angle, θ,between the liquid and the wall of the capillary tube is less than 90° (that is, if the
liquid wets the tube wall), the liquidmeniscus in the tube is concave, and the liquid rises in the tube
(Fig. 2a). If θ is equal to 90°, there is no capillarydepression or rise (Fig. 2b). If θ is greater than
90°, the liquid is depressed in the tube and does not wet the tube wall, andthe meniscus is convex
(Fig. 2c). In capillary rise (Fig. 2a), the meniscus does not pull the liquid up the tube; rather,
thehydrostatic pressure immediately under the meniscus is reduced by the distribution of the surface
tension in the concavesurface, and the liquid is pushed up the capillary tube by the hydraulically
transmitted pressure of the atmosphere at thefree surface of the liquid outside the capillary tube.
Fig. 2 Rise or depression in small vertical capillary tubes, determined by the contact angle θ, between a liquidand the
wall of a capillary tube. (a) θ< 90°: capillary rise. (b) θ= 90°: no capillary depression or rise. (c) θ>90°: capillary
depression
Developer Application: A thin layer of developer is then applied to the sample to draw penetrant
trapped in flaws back to the surface where it will be visible. Developers come in a variety of forms
that may be applied by dusting (dry powdered), dipping, or spraying (wet developers).
Indication Development: The developer is allowed to stand on the part surface for a period of time
sufficient to permit the extraction of the trapped penetrant out of any surface flaws. This
development time is usually a minimum of 10 minutes. Significantly longer times may be
necessary for tight cracks.
Inspection: Inspection is then performed under appropriate lighting to detect indications from any
flaws which may be present.
Clean Surface: The final step in the process is to thoroughly clean the part surface to remove the
developer from the parts that were found to be acceptable.
Advantages and Disadvantages of Penetrant Testing
Advantages
The method has high sensitivity to small surface discontinuities.
The method has few material limitations, i.e. metallic and nonmetallic, magnetic and
nonmagnetic, and conductive and nonconductive materials may be inspected.
Large areas and large volumes of parts/materials can be inspected rapidly and at low cost.
Parts with complex geometric shapes are routinely inspected.
Indications are produced directly on the surface of the part and constitute a visual
representation of the flaw.
Aerosol spray cans make penetrant materials very portable.
Penetrant materials and associated equipment are relatively inexpensive.
Disadvantages
Only surface breaking defects can be detected.
Only materials with a relatively nonporous surface can be inspected.
Precleaning is critical since contaminants can mask defects.
Metal smearing from machining, grinding, and grit or vapor blasting must be removed prior
to LPI.
The inspector must have direct access to the surface being inspected.
Surface finish and roughness can affect inspection sensitivity.
Multiple process operations must be performed and controlled.
Post cleaning of acceptable parts or materials is required.
Chemical handling and proper disposal is required.
Ref.:
http://www.nde-ed.org/EducationResources/CommunityCollege/PenetrantTest/Principles/liquidpi.htm
https://eis.hu.edu.jo/ACUploads/10526/Liquid%20Penetrant%20Testing.pdf
Hands on Experiments (Ultrasonic NDT)
Calibration of the Reference Block
Figure: 1a
series of echoes are obtained at the screen positions of 25, 50, 75 and 100 scale divisions. These
movements of the two instrument controls automatically set the zero point. If the echoes will not
coincide with appropriate scale divisions, the time base is not linear and a graphical calibration
should be prepared.
Now place the probe in the position B as shown in figure 1b and obtain the backwall echo at a
position as shown in figure 1c. Screen display represents 100mm steel thickness.
Figure: 1b
Figure: 1c
Exercise 2: Determine the probe index and the angle of refraction of the transverse wave probe
The probe index is given on the probe; in order to check that the position marked on the probe is
correct, place the probe in position H (Fig. 2a).
Figure: 2a
Move the probe until maximum amplitude is received from the 100 mm curved surface. The central
mark on the graduated scale will be the position at which the beam leaves the Plexi- glass and
enters the steel, i.e. the probe index.
Move the probe until maximum signal amplitude is obtained from the Plexiglass cylinder (Fig. 2b).
Figure: 2b
The reference block has calibrated scales engraved at 40-70° and the relevant angle of refraction
marked on the probe should coincide with the correct scale position. The probe index which has
been previously determined should be marked on the probe in order to obtain the correct results.
Exercise 3: Determine the radius of curvature of V1 block using the angle beam probe.
Figure: 2c
The angle beam probe placed at the position shown in figure 2c. When the angle-beam probe is
exactly coupled at the centre of the circle segment, a first echo is exactly received from 100 mm out
of V1 block. Because the radius of the circle segment is exactly 100 mm you will regularly receive
an echo sequence with distances of 100 mm, 200 mm, 300 mm etc. (get the echo sequences).
Exercise 4: Determine the radii of curvatures of the V2 block using angle beam probe
The V2 block has two circle segments of different radius with a common centre point. Place the
angle beam probe as shown in figure 2d and 2e and check the echo sequences at the values
mentioned in the figures respectively.
Figure: 2d Figure: 2e
Exercise 5: Locating of Reflectors with an Angle-beam Probe
Figure: 2f
Figure: 2g
Place the angle beam probe on the V2 block (fig 2f) and move the probe in the horizontal direction
to find the position of maximum echo amplitude. Mark the position, then use the triangle law (figure
2g) to locate the position of the drill hole.
Exercise 6: Find the extreme boundary of the existing flaw in the supplied reference block.
B A
Figure 3a
Place the straight beam probe on the surface of the test block at some position (A) where from you
will get the flaw echo. Note the amplitude of the flaw echo. Now scan the test object and find the
position (B) where the echo amplitude becomes just half of the flaw amplitude. That position (B)
corresponds to the boundary of the flaw.
Ref.
https://www.ndt.net/article/v05n09/berke/berke3.htm
Phased Array Ultrasonic Testing
Phased array calibrations shall be performed for velocity, wedge delay, sensitivity and
DAC or TCG.
Any changes in search unit, wedges, instrumentation, aperture, focus etc. shall be cause
for re-calibration.
Calibration for sound velocity may be done using radius, depth, or thickness.
According to the type of probe and the calibration block used, calibration can be
performed using radius, depth, or thickness.
Generally, sound velocity calibration is performed before the wedge delay calibration
because most of the systems use the sound velocity calibration result for the wedge delay
calibration.
Sensitivity calibration may be done for an angle-beam probe, using a calibration block
with a side drilled hole or using the radius of a calibration block.
Time Corrected Gain (TCG) or Distance Amplitude Correction (DAC) may be used by
continuing the same process over a wide series of side drilled holes to calibrate reflectors
at different depths or metal paths so they are all detected at the same amplitude (typically
80% amplitude).
The same instrument, contact wedge, focal laws and focusing used during calibration
should be used for the examination.
The phased array beam should be focused just beyond the component thickness. The
beam may be re-focused at the depth of an indication and evaluated in accordance with
the requirements of this procedure.
Sectorial S- Scans may be used for the appropriate angles that ensure complete coverage
of the weld and heat affected zone.
Electromagnetic Sensors for damage detection and failure analysis
MagStar
MagStar is a multi-parameter, non-invasive magnetic sensing system developed at CSIR-NML
and commercialized in collaboration with M/s Techno four, Pune. The “MagStar” is a portable
Magnetic nondestructive testing equipment aimed to analyze both bulk and surface features of
ferromagnetic structural materials like steels, cast iron etc.
MagStar generates multiple Electromagnetic(EM) parameters based on the principles of
1. Magnetic Hysteresis Loop (MHL) and
2. Magnetic Barkhausen Emission (MBE)
Fig.1. The photographic view of MagStar sensing system and its industrial field trails on boiler
pipes and steel rolls.
Principle of technique
Sensor
MagStar sensor probe is used for magnetic excitation and flux pick up from the test object, and
consequently generating MBE and MHL signals (as shown in Fig.3). Sensor probe head is
customized as per the requirement of shape and size of test objects, e.g., flat sheet, pipe/ tube,
curvature surfaces of rolls or rails etc.
Technical specifications
MHL measurement:
Frequency Range: 50mHz to 200Hz,
Excitation magnetic field: 1500 Oe (max.)
Wave Shape: Sinusoidal / Triangular
MBE measurement:
Frequency Range: 10Hz to 200 Hz
Excitation magnetic field: 1500 Oe (max.)
Number of cycles 3 to 10
Gain 0 to 20 dB in steps of 1 dB
Filter setting 10-300KHz independently variable low pass and high pass.
Power Requirement:230V, 50Hz, 200VA
Weight:4.25Kg
Connectivity: Ethernet
Control, Display and Analysis:External laptop /notebook/ personal computer
Operating System: Windows 7
Typical Applications
1. Microstructural phase evaluation of ferritic steels
Residual stresses (RS) are self-balanced / locked in stresses that remain in a component in the absence
of external loading or thermal gradients. RS remains in a material during fabrication and processing.
Main concern lies in service inspection.
Coercivity increases with shot peening time and Almen intensity due to enhancement of compressive
stresses in the steel.
Conversely, Magnetic Barkahusen emission (MBE) signal decreases with increase of shot peening
time and Almen intensity for generating more compressive stresses (Fig. 7).
MagSys
MagSys is a Giant Magneto Impedance (GMI) principle based magnetic sensing device
developed at CSIR-National Metallurgical Laboratory, Jamshedpur. Giant Magneto-Impedance
(GMI) based sensing device for Structural Health Monitoring (SHM) of steels.The MagSysis a
portable device measuring GMI change in magnetic field. The device is capable of detecting
very low magnetic field and magnetic phases. The system is suitable for monitoring magnetic
field variation within a narrow region due to smaller core diameter (~100µm) of sensing element.
Fig. 8. The photographic view of MagSys system(left) and in-house prepared micro wire sensing
element (right)
Sensor
MagSys sensor comprises a nanostructured wire prepared by rapid solidification route through in-
rotating water quenching technique. The sensor operates on the large impedance change of the
sensing material at high frequency (250- 400 kHz) by the influence of local change in material
magnetic properties.
GMI Measurement
The strain induced martensite phase formed in austenitic stainless steels are magnetic in nature.
The strain induced martensite phases exerts residual magnetic field, which MagSys sensor
detects and corresponding sensor response will be obtained.
Fig. 11. The sensor output distinguishes the usage of carbon electrodes and SS electrodes in the
weld root pass.
Application Sectors
Power and petrochemical industries