Department of Defense-relevant
chemical and biological sensing
applications: state-of-the-art and
future outlooks
Rohith Chandrasekar
Zachary J. Lapin
Andrew S. Nichols
Rebecca M. Braun
Augustus W. Fountain, III
Rohith Chandrasekar, Zachary J. Lapin, Andrew S. Nichols, Rebecca M. Braun, Augustus W. Fountain,
, III“Photonic integrated circuits for Department of Defense-relevant chemical and biological sensing
applications: state-of-the-art and future outlooks,” Opt. Eng. 58(2), 020901 (2019),
doi: 10.1117/1.OE.58.2.020901.
Abstract. Photonic integrated circuits (PICs), the optical counterpart of traditional electronic integrated circuits,
are paving the way toward truly portable and highly accurate biochemical sensors for Department of Defense
(DoD)-relevant applications. We introduce the fundamentals of PIC-based biochemical sensing and describe
common PIC sensor architectures developed to-date for single-identification and spectroscopic sensor classes.
We discuss DoD investments in PIC research and summarize current challenges. We also provide future
research directions likely required to realize widespread application of PIC-based biochemical sensors.
These research directions include materials research to optimize sensor components for multiplexed sensing;
engineering improvements to enhance the practicality of PIC-based devices for field use; and the use of syn-
thetic biology techniques to design new selective receptors for chemical and biological agents. © The Authors.
Published by SPIE under a Creative Commons Attribution 4.0 Unported License. Distribution or reproduction of this work in whole or in part requires
full attribution of the original publication, including its DOI. [DOI: 10.1117/1.OE.58.2.020901]
Keywords: biochemical sensing; photonics; sensing; spectroscopic; interferometer; ring resonator; photonic crystal; spectroscopic
sensor.
Paper 181162SS received Aug. 10, 2018; accepted for publication Jan. 8, 2019; published online Feb. 5, 2019.
Fig. 2 Analyte detection using single-identification and spectroscopic sensors. (a) Single-ID sensor
waveguides are functionalized with a biochemical receptor, which selectively binds to a specific analyte.
These sensors operate at a single wavelength that is chosen based on the absorption of the analyte.
When the analyte successfully binds to the receptor, the evanescent field interacts with the analyte, and
accordingly a reduction in the signal transmitted through the waveguide is detected. (b) For spectroscopic
sensors, no receptor is used and a broadband optical source is used to probe the analytes. The resulting
evanescent field probes all analytes near the surface of the waveguide. The output spectrum may have
multiple features, which need to be deconvolved with known analytes in the environment in order to
ascertain their concentrations. These spectra may change rapidly based on the complexity of the
environment.
is also important to note that comparing sensitives across 3.1 Integrated Interferometers
reports can be challenging: sometimes units such as mass Interferometers are historically established as highly sensi-
per volume are given, which is difficult to extend to other tive sensors based on the phase-sensitive detection of minute
analytes, and even if a normalized unit such as refractive optical path differences; a striking recent example is the
index units (RIU) is given, it is not normalized to the physi- detection of gravitational waves.11 Interferometric PIC-
cal size of the sensor.
based sensors, like their traditional free-space counterparts,
Below, we give a brief overview of the state-of-the-art for
split light from one waveguide into two separate arms,
single-ID and spectroscopic sensors, principles of detection
a reference arm and a sensing arm, functionalized for analyte
unique to each type, and relative advantages and disadvan-
capture. If a target analyte is present in a sample, it will bind
tages of each type.
to receptors on the sensing arm, and, as a result, light passing
through the sensing arm will be phase shifted relative to the
3 Single-ID PIC-Based Biochemical Sensors reference light. The interference pattern formed when light
As discussed above, single-ID sensors use functionalized from both arms is recombined can then be measured to deter-
waveguide surfaces to capture and detect target analytes. mine the relative optical phase delay and calculate the
This type of detection is performed at a single wavelength amount of analyte present. Two standard configurations
because the optical properties of the analyte are known exist for interferometers12 integrated on a chip: the Mach–
a priori and changes in the transmitted optical signal can Zehnder interferometer (MZI)13 [Fig. 3(a)] and the Young
be directly attributed to that analyte. Three PIC-based sensor interferometer (YI) [Fig. 3(b)], which is similar but out-cou-
designs using evanescent field detection have been the most ples the light in both arms allowing the light to interfere in
prevalent over the past two decades: (i) integrated interfer- free space, instead of using a junction on-chip.
ometers, (ii) ring resonators, and (iii) photonic crystal Several demonstrations of MZI-based PIC biosensors are
(PhC)-based sensors. We briefly discuss the architectures reported in the literature, with LODs reaching as low as
of each device type and compare their relative performance 1 × 10−7 RIU or, equivalently, LOD of 10 pM of DNA.15
characteristics in the context of biochemical sensing, when Various waveguide designs, such as slot and bimodal wave-
applicable. guides (BiMW), have also been explored to improve sensor
Fig. 3 Interferometric PIC sensors. A standard (a) MZI (a) and (b) YI configurations. Light enters the
interferometers on the left and is split into the sensing and reference arms. In the MZI case, the
arms are recombined in a Y-junction and the interference measured in the output is used to quantify
the presence of an analyte in the sensing arm. In the YI configuration, the light is not coupled, but is
outcoupled from sensing and reference arms and interfered in free space to create interference pattern.
The presence of an analyte and its concentration is identified based on the interference pattern, repro-
duced from Ref. 14.
accuracy. In the slot design, the sensing arm is split into two considerably enhances sensor sensitivity.21 Analytes cap-
waveguides separated by a nanometer-scale, low-refractive tured on the functionalized surface of the ring resonator
index “slot” region for light confinement. This design greatly shift the resonant wavelength only slightly, but this change
enhances the light–matter interaction yielding higher sensi- is detected as a relatively large decrease in output intensity at
tivity to the refractive index of the analyte; an LOD of a specific wavelength [Fig. 4(a)].10
5.4 × 10−6 RIU and 1 pg∕mL of streptavidin was recently The number of revolutions light takes around the ring res-
demonstrated.16 The same slot design was also recently inte- onator yields an optical path length (OPL) orders of magni-
grated into an LOC-type point-of-care medical device that tude larger than the sensor’s physical footprint, which is
used a multiplexed approach to detect specific microRNAs generally described by the resonator’s quality (Q) factor.21
in human urine with an LOD of 1 nM.17 In the BiMW con- High Q-factors—considered to be at least 106—indicate
figuration, a step junction in a single-mode waveguide low optical loss and correspondingly long photon lifetimes,
excites two transverse modes, in a single waveguide, which which is translated into narrow linewidths and often high
interact with the sample before being measured by a photo- peak resolution (i.e., high sensitivity). Q-factors in the
detector. The sensitivity level of the BiMW sensor is com- 106 range have been reported for resonators around 50 to
parable to other integrated interferometers; LODs down to 200 μm in circumference, which has an OPL equivalent
5.9 × 10−4 experimentally were recently demonstrated for to a linear waveguide measuring several centimeters. In prac-
bovine serum albumin (BSA), with a theoretical LOD tice, a Q-factor of 108 effectively means a molecule will be
limit of 2.5 × 10−7 RIU.18 sampled more than 100,000 times, given by Leff ¼ 2πn Qλ
,
To the best of our knowledge, the most sensitive bio- where Leff is the effective free-space path length, Q is the
chemical interferometric sensor is a YI device with two Q-factor, λ;is the free-space wavelength, and n is the refrac-
Ta2 O5 slab waveguides, which achieved a sensitivity of tive index of the resonator.21
9 × 10−9 RIU and a surface coverage of 13 fg∕mm2 for both Other important characteristics of resonators are the free
bulk solution and surface sensitivities to immunoglobulin spectral range (FSR) and the finesse. The FSR is the inverse
G.19 However, the YI configuration requires off-chip detec- of the round-trip time of an optical pulse around the resona-
tion and analysis, which may impact operationally relevant tor and defines the optical frequency range over which the
considerations for DoD applications. resonator can be utilized. The finesse, which is determined
by the resonator losses and in independent of the resonator
circumference, is defined as the FSR divided by the full-
3.2 Ring-Resonator Sensors width at half-maximum bandwidth of the resonance. Hence,
Ring-resonator transducers are becoming more common for Leff can also be defined as the finesse times the ring circum-
biochemical sensing applications due to their high sensitivity ference, which is a useful metric as the sensitivity generally
and potential for multiplexing; the detection of multiple scales with finesse for most sensing applications.
analytes is vital for testing complex fluids (e.g., blood, Therefore, despite the small physical size of the ring res-
saliva, and urine) in diagnostics, monitoring, and toxicity onator, it can achieve higher sensitivities than straight wave-
screening.20 PIC ring-resonator sensors use at least one linear guides while using orders of magnitude less surface area. Li
waveguide to couple light to a closed-loop waveguide (the and Fan23 have demonstrated LOD as low as 2.5 × 10−7 in
ring resonator) and, through evanescent-wave coupling, RIU, or equivalently detected 1.6 pg∕mm2 of biotin, using
excite resonant modes in the loop waveguide. On resonance, a ring resonator with a diameter of just 70 μm. Smaller wave-
constructive interference is generated in the multiple round- guide sizes also facilitate the development of more complex,
trips over the ring circumference; however, off resonance integrated devices. For example, Iqbal et al. demonstrated an
the transmission rapidly decreases. This resonance effect arrayed device using 32 ring-resonator sensors to detect
Fig. 4 Ring-resonator sensor. (a) Schematic of a ring-resonator sensor. The linear waveguide provides
input light to the loop waveguide at the coupling zone and captures output light from the loop for meas-
urement. The presence of analyte will shift the resonance wavelength within the loop, which is detected in
the output spectra, adapted from Ref. 10. (b) Micrograph of a fabricated ring resonator, adapted from
Ref. 22.
material platforms for on-chip integration but also yield which may also enable devices operating at longer wave-
higher sensitivities by probing fundamental rovibrational lengths, such as the mid- and long-wave IR. For example,
modes rather than the weaker overtones in the NIR. We dis- photonic systems based on III–V materials operating in
cuss some of these current challenges, as well as current and the mid-IR such as GaAs, InP, and GaN already have a com-
near-term DoD investments aimed at addressing these chal- plete set of photonic components, including on-chip mono-
lenges, in the next section. chromatic lasers, amplifiers, and modulators.49 These
platforms are continually being developed to create hybrid
5 Current PIC-Based Research Areas for the DoD platforms with silicon photonics.
The DoD is investing in the development of mission-critical
PIC-based sensor technologies through basic and applied
research programs spanning a diverse set of DoD organiza- 5.2 High-Volume Manufacturing of PIC Devices
tions. In this section, several of these programs are high- The prospect of single-chip, multitechnology integration
lighted in the fields of fully integrated devices, long- opens an array of opportunities. While on-chip PIC compo-
wavelength broadband sources, and data analysis. nents are still under development and being optimized
using several material platforms, the National Network for
5.1 Fully Monolithic Photonic Chips Manufacturing Innovation has founded the American
Institute for Manufacturing Integrated Photonics (AIM
Although significant advancements have been made in the Photonics) to advance large scale manufacturing of CMOS-
CMOS compatible fabrication of PIC technologies, fully compatible PICs. This institute serves to leverage existing
integrated monolithic silicon photonic/electronic chips CMOS electronics infrastructure and fabrication technolo-
with on-chip sources, detectors, and amplifiers have yet to gies to develop a complete PIC ecosystem within the
be demonstrated. This is primarily a material limitation United States; this includes foundry access, automated
because silicon is an indirect bandgap semiconductor with design tools, resources for packaging, assembling and testing
extremely low light emission efficiency. Given the vast devices, and a domestic workforce.50
knowledge and infrastructure for silicon device fabrication AIM Photonics is currently focusing on achieving
and the extensive role of silicon in electronics, it is desirable a manufacturing readiness level 7, as defined by the DoD
to find ways to use a silicon platform for all-optical devices. Manufacturing Technology program, for CMOS-compatible
To that end, several efforts have been made to enhance sil- integrated electronic and photonic devices, such as the device
icon emission efficiency to demonstrate on-chip lasers, such shown in Fig. 6.51 CMOS devices based on both silicon and
as growing successive layers of indium arsenide on silicon to indium phosphide would span applications in telecom, RF
form quantum dots as demonstrated through the DARPA analog circuits, and PIC sensors and array technologies.
Electronic-Photonic Heterogeneous Integration (E-PHI) pro- Integrated devices should serve to increase both the capacity
gram.44,45 This demonstration is now serving as a foundation and reliability while decreasing the cost, size, weight, and
for the development of other photonic components such as power (SWaP) consumption of devices.52 AIM also receives
optical amplifiers, modulators, and detectors. funding from the Navy, to advance their multiproject wafer
Current efforts are now focused at large-scale integration assembly and packaging service, which provides access to
of electronics and photonics, such as the recent DARPA the most advanced 300-nm semiconductor fabrication facili-
Photonically Optimized Embedded Microprocessors ties and a suite of passive and active PIC wafer processing
(POEM) program.46 This program is developing chip-scale technologies. This infrastructure, largely funded by the
integrated photonic technology to enable seamless intrachip Government, will ensure that the necessary manufacturing
and off-chip photonic communications. This program has resources are available for mission-critical technologies.
already had several groundbreaking demonstrations, includ- AIM Photonics is already making an impact on mission-
ing the first single-chip microprocessor with 70 million tran- relevant technologies, including the successful development
sistors and 850 integrated photonic elements, all fabricated of a chip-scale Sarin gas photonic sensor.53 The team
using the standard microelectronics foundry process, to work achieved the necessary high sensitivity by developing
together to provide logic, memory, and interconnect
functions.47 This is significant because the silicon fabrication
requirements for electronics and photonics are orthogonal—
while electronics requires very thin silicon materials (order
of single nm) to quickly dope and transition the silicon
between conducting and insulating states, photonics requires
much thicker silicon (order of 100 nm) to confine light effi-
ciently. Earlier this year, the same group demonstrated, for
the first time, integrated photonics (optical waveguides and
resonators, transceivers, high-speed optical modulators, and
sensitive avalanche photodetectors) with silicon nanoelec-
tronics in polycrystalline silicon operating at 10 Gbits∕s,
all on the same chip.48 These developments are pushing
the technological limits of integrated photonics, but they
do not yet include on-chip optical sources.
While the on-chip integration of sources, detectors, and
electronics, may be possible in the near-IR, there is still Fig. 6 PIC sensor device with integrated photonics and electronics
much interest in integrating different material platforms, manufactured by the AIM Photonics teams.
a long-path on-chip waveguide and coating it with a polymer There is also a strong push to move sources for chemical
to concentrate the simulated Sarin agent. Their next steps are detection to longer wavelengths in the mid-IR and long-wave
to develop a fully integrated sensor with on-board spectrom- IR ranges as it allows for probing much stronger fundamental
eters and lasers and to apply such monolithic sensors for non- molecular vibrations, as opposed to the overtones excited in
defense applications. the near-IR. Programs such as SCOUT have specifically pur-
In collaboration with the MIT Microphotonics Center and sued the development of mid-IR frequency combs for DCS-
the International Electronics Manufacturing Initiative, AIM based biochemical sensing and have demonstrated signifi-
released their 2017 Integrated Photonics Systems Road Map cant enhancement in sensitivity and limits of detection.
in March 2018.54 The roadmap, which serves to drive tech- Earlier this year, Ycas et al.40 at the National Institute of
nology development, identifies several grand challenges, Standards and Technology demonstrated a mid-IR dual-
including: the drastic improvement in power efficiency, comb system spanning 2.6 to 5.2 μm with sub-MHz fre-
the development of tools for large-scale PIC system design, quency precision. While this is an open-path-based system,
new manufacturing and material technologies, and standards recently developed chip-scale battery-operated frequency
for interoperability to broaden market potential. AIM has combs have the potential to revolutionize the SWaP of
already solicited and funded several proposals for R&D DCS systems.60 Just this year, Gaeta et al. have demonstrated
efforts to overcome these challenges, including a new effort high-SNR absorption spectroscopy with an on-chip DCS for
to develop a CMOS-compatible waveguide platform for inte- the first time in the NIR, with a 170-nm bandwidth and
grating mid-IR and long-wave IR on-chip laser sources a 20-μs acquisition time. Further broadening of the spectral
(spanning 3 to 14 μm). The large wavelength range would coverage of this on-chip system and shifting to longer wave-
enable several DoD-relevant applications including detecting lengths could enable truly disruptive chemical and biological
atmospheric trace gases. We discuss the importance of these sensing technologies relevant to DoD applications. However,
integrated sources further below. moving to the mid-IR does introduce several operational
challenges including limited achievable Q-factors and
increased water vapor absorption, which, respectively,
5.3 Broadband, On-Chip IR Sources for Biochemical reduce sensitivity and increase background signatures,
Detection making the detection and quantification of analytes in native
environments more difficult. We discuss the challenges with
Looking specifically toward spectroscopy-based biochemi- detection in native environments in further detail below.
cal sensing platforms, there is a clear need for broadband
sources with extremely high-spectral resolution, different
from the on-chip monochromatic lasers mentioned above: 5.4 Analyte Identification in Native Environments
increasing the bandwidth will allow for more properties of As PIC-based sensors transition from the lab to native envi-
each analyte to be measured, resulting in improved specific- ronments for field use, metrics for assessing utility must go
ity and multiplexing. One promising technology is optical beyond sensitivity alone to also include concepts of device
frequency combs, which have emerged in the past decade robustness and trustworthiness, such as the probability of
as powerful and precise tools with coherent, equidistant correct detection, false alarm rate (FAR), and the response
spectral lines spanning a broad range with unprecedented time.61 FAR is defined as the number of false alarms per
precision in frequency in timing.55 Several programs across unit time and defines how well the sensor identifies analytes
the DoD have been pushing the development of frequency amidst a background and interferents. For biochemical detec-
combs spanning the visible and IR ranges for applications tion, false negatives are far more dangerous than false pos-
in biochemical sensing, time-frequency transfer, and metrol- itives, though a high false positive rate undermines the
ogy, such as DARPA Spectral Combs from UV to THZ detection system trustworthiness (a challenge to combat
(SCOUT),56 DARPA program in Ultrafast Laser Science operational and response fatigue). The sensor’s receiver
and Engineering (PULSE),57 and the DARPA Direct On- operating characteristic (ROC) curve captures the perfor-
Chip Digital Optical Synthesizer (DODOS)58 programs. mance trade-off between sensitivity, probability of correct
These programs continue to push the limits of frequency detection, and FAR and dictates an implicit associated detec-
combs in terms of power generation, bandwidth, spectral res- tion confidence.61 Additionally, metrics for long-term stabil-
olution, and detection. ity of devices, such as the Allan variance or Allan deviation,
One major challenge of optical frequency combs is their will be required to understand device stability to noise proc-
requirement for large spectrometers to spectrally resolve esses and drift and assess their ability to transition to
each individual comb tooth. However, by interfering two DoD-relevant missions. We direct the reader to Ref. 61, a
combs with slightly different repetition rates, one can gen- government-issued study with in-depth analyses of chemical
erate an RF comb composed of heterodyne beat pairs, and biological sensor metrics, testing, and recommendations.
which contain all the relevant spectral information and is The DoD is actively using machine learning algorithms
easily accessible with RF electronics.59 This technique, that are trained with experimental data to reduce FAR by
termed dual-comb spectroscopy (DCS), has several advan- identifying analyte signatures amidst a strong background
tages over conventional spectrometers, including a broad and interferents; this includes water absorption in the mid-
spectral coverage, high-frequency resolution, high-sensitiv- IR, which obscures real-time analyte identification. In
ity/signal-to-noise ratio, and fast acquisition speed; as addi- terms of fielded devices, DHS S&T first responders group
tionally, DCS performance is not limited by the instrument is actively looking at machine learning and artificial intelli-
OPL, which may enable ultracompact systems. Hence, DCS gence technologies for wearable sensor technologies.62 Their
is proving to be an invaluable tool for biochemical detection new platform, assistant for understanding data through rea-
in the near- and mid-IR. soning, extraction, and synthesis, collects data from various
sensors and notifies first responders of any necessary actions reported74 and demonstrated to function as an ultracompact
to be taken, as well as any contextual insight of similar events circular dichroism (CD) spectrometer; traditional CD spec-
that occurred elsewhere in the country.63 trometers were large, mechanically slow, and expensive
In addition to machine learning algorithms, networking instruments with no on-chip counterpart. The integration
orthogonal sensing platforms to reduce the overall FAR of metasurface-based spectrometric methods on a chip plat-
across a network are a capability being developed at DARPA form could enable the ultrafast detection of complex, mis-
through the SIGMA+ program.64 SIGMA+ is focused on sion-relevant molecules.
both developing highly sensitive detectors and advanced Usability advances could be enabled by transitioning PIC
intelligence analytics to detect minute traces of substances sensors from a hard silicon substrate to flexible, biocompat-
related to CBRNE threats. By fusing CBRNE data, as ible material platforms that would enable PIC-based
well as local weather and external contextual data across wearable and implantable sensors for both physiological
a network infrastructure, SIGMA+ aims to develop a persis- monitoring and the continuous interrogation of the immedi-
tent, real-time early detection system to monitor a city-wide ate environment. Several reports demonstrate PIC sensors
region (km2 ) with maximum sensitivity and minimum FAR. embedded in polymers such as poly(dimethyl)siloxane
(PDMS) or photoresists such as SU-8 via device transfer
methods.75 However, the mechanical and thermal durability
6 PIC-Based Sensor Research Horizons
of these sensors is not yet fully understood. PDMS-only sen-
Current and future research initiatives across a range of dis- sors may offer an ideal platform in terms of combined robust-
ciplines provide opportunities to improve the functionality of ness and flexibility, however, current devices suffer from
PIC-based biochemical sensors. Advances could enable the low-sensitivity attributable to challenges refractive index
development of novel synthetic receptors, the exploitation of control during fabrication. Recently, Cai et al.76 developed
new material platforms for on-chip detection, the develop- a fabrication method for tuning the refractive index of
ment of protocols for higher precision waveguide surface PDMS by adjusting the ratio of base and curing agent and
functionalization, and more generally identify methods to demonstrated waveguides with a transmission loss of about
design sensors with improved operational utility. 1.1 dB∕cm at 460 nm. Independently, methods to mass-
New synthetic biology tools may enable the development manufacture polymer-based PIC devices with high Qs,
of receptors with both high affinity and high selectivity, such as roll-to-roll nanoimprint lithography,77 need to be fur-
capable of filling the existing technological void of differen- ther developed.
tiating chemically similar small DoD-relevant toxic chemi- Beyond optimizing fundamental design and detection
cals. For example, detecting chemical warfare agents approaches, PIC-based sensors for DoD applications still
using acetylcholinesterase or butyrylcholinesterase as the need to achieve practical field and operational utility.
receptor will not provide selectivity across a variety of Research to improve shelf-life and sensor reusability without
OPs.65 Gene editing tools such as CRISPR-Cas9 and directed a degradation in precision or sensitivity4 is required.
evolution approaches could be used to quickly and efficiently
Reporting ROC curves and Allan deviation plots to under-
construct chimeric receptors or receptor complexes with var-
stand FAR and long-term drift of PIC-sensors will be
ied allosteric properties and substrate binding kinetics to
required to assess operational utility. Fielded sensors also
functionalize waveguide surfaces. Ancillary research
require a simple sample preparation procedure at a minimum
would likely include the development of surface chemistries
and ideally will work with native samples. Research into
for immobilization of new receptor types.66
microfluidics may provide sample preparation solution;
Along with the development of novel receptors and sur-
face chemistries, the spatial selectivity of waveguide surface for example, magnetic and digital microfluidics are of con-
functionalization must also be improved. Traditionally, siderable interest for field-based LOC applications due to
waveguide functionalization for multiplexed sensing4,20 is their high precision, power-free operation, and utility as
performed following photonic structure fabrication; how- a functional substrate for molecular capture.78,79
ever, in situ functionalization may improve spatial control
and reduce contamination of other photonic components,
7 Conclusions
increasing the sensors multiplexed capacity.4,67 The capabil-
ity of click chemistry and silanization to attach binding sites Integrated photonics technology is paving the way toward
to silicon surfaces may provide an additional benefit.68–70 truly portable LOC platforms and highly accurate sensors
These approaches, however, have not been demonstrated for applications outside the laboratory. Significant efforts
or developed for PIC sensor specific technologies.4 have been made to develop ultrasensitive PIC-based sensors,
New material platforms, such as porous silicon, could be but limitations of integrating all components into sample-to-
exploited to enhance detection via larger transducer area and answer microsystems, and the challenges associated with the
molecular size selectivity in PIC sensors, as demonstrated in real-world implementation of such devices, need to be over-
a DNA sensor and an interferometer-based nerve agent come. The DoD need for such technologies has led to several
sensor.71,72 Additional on-chip optical control could be research programs spanning basic research to integrated
enabled by nanoscale structured surfaces. Referred to as devices, as discussed in this paper, to drive the development
metasurfaces,73 such surfaces have demonstrated an unprec- of disruptive PIC sensors. To have a significant impact in
edented control of light in wavelength and polarization, advancing PIC-based sensors toward field-ready technolo-
offering a route not only to improve devices but offer gies, future research can develop reliable functionalization
novel on-chip functionality. For example, an ultrathin meta- protocols for chemical and biological agents, use synthetic
surface that efficiently separates left-hand and right-hand cir- biology techniques to design new selective receptors, and
cular polarizations at visible wavelengths was recently optimize sensors for multiplexing.
65. G. Liu and Y. Lin, “Biosensor based on self-assembling acetylcholines- Rohith Chandrasekar is a lead scientist at Booz Allen Hamilton. He
terase on carbon nanotubes for flow injection/amperometric detection of is an expert in metamaterials, plasmonics, optics, nanophotonics, and
organophosphate pesticides and nerve agents,” Anal. Chem. 78, 835– spectroscopy. He received his PhD from Purdue University and his
843 (2006). bachelor’s and master’s degrees in electrical engineering from
66. Y. J. Jang et al., “Update 1 of destruction and detection of chemical
warfare agents,” Chem. Rev. 115, PR1–PR76 (2015). Cooper Union for the Advancement of Science and Art.
67. B. He, T. J. Morrow, and C. D. Keating, “Nanowire sensors for multi-
plexed detection of biomolecules,” Curr. Opin. Chem. Biol. 12, 522– Zachary J. Lapin is a senior consultant at Booz Allen Hamilton.
528 (2008). Previously, he was a scientific editor for nature communications.
68. C. Haensch, S. Hoeppener, and U. S. Schubert, “Chemical modification He is an expert in optics and nanophotonics. He received his doctor
of self-assembled silane based monolayers by surface reactions,” Chem. of sciences from ETH Zurich in Switzerland, master of sciences in
Soc. Rev. 39, 2323–2334 (2010). optics from the Institute of Optics at the University of Rochester,
69. M. A. C. Campos, J. M. J. Paulusse, and H. Zuilhof, “Functional mono- and a bachelor of sciences from Bates College.
layers on oxide-free silicon surfaces via thiol-ene click chemistry,”
Chem. Commun. 46, 5512–5514 (2010).
70. G. Mohammadi Ziarani et al., “Advances in click chemistry for silica- Andrew S. Nichols is a senior lead scientist at Booz Allen Hamilton.
based material construction,” RSC Adv. 6, 21979–22006 (2016). He is an expert in chemical and biological detection and defense. He
71. G. Rong et al., “Nanoscale porous silicon waveguide for label-free DNA received his PhD from the University of Miami School of Medicine and
sensing,” Biosens. Bioelectron. 23, 1572–1576 (2008). his BA degree in biology from Franklin and Marshall College.
72. S. Jang et al., “Detection of G-type nerve agent simulants based on
a double reflection DBR porous silicon interferometer,” J. Korean Rebecca M. Braun is a lead associate at Booz Allen Hamilton. She is
Phys. Soc. 52, 212–215 (2008). an expert in basic research program management. He received her
73. Z. Li et al., “Controlling propagation and coupling of waveguide modes
using phase-gradient metasurfaces,” Nat. Nanotechnol. 12, 675–683 master’s degree in public health from the University of Maryland—
(2017). College Park in 2013 and her BA degree in biological sciences
74. A. Shaltout et al., “Photonic spin Hall effect in gap–plasmon metasur- and international relations from the University of Delaware.
faces for on-chip chiroptical spectroscopy,” Optica 2, 860–863 (2015).
75. L. Li et al., “Integrated flexible chalcogenide glass photonic devices,” Augustus W. Fountain III is the U.S. Army’s senior research scientist
Nat. Photonics 8, 643–649 (2014). for chemistry at the Edgewood Chemical Biological Center. He is an
76. Z. Cai et al., “A new fabrication method for all-PDMS waveguides,” internationally recognized expert in electro-optics as it pertains to
Sens. Actuators, A Phys. 204, 44–47 (2013). chemical, biological, radiological, nuclear, and explosives (CBRNE)
77. A. V. Shneidman et al., “All-polymer integrated optical resonators by
roll-to-roll nanoimprint lithography,” ACS Photonics 5, 1839–1845 detection. He received his PhD from Florida State University, his mas-
(2018). ter’s degree in national strategic studies from the U.S. Army War
78. Y. Zhang and N.-T. Nguyen, “Magnetic digital microfluidics—a College, his bachelor’s in science from Stetson University. He is a
review,” Lab Chip 17, 994–1008 (2017). fellow of SPIE.
79. T. Kokalj, E. Pérez-Ruiz, and J. Lammertyn, “Building bio-assays with
magnetic particles on a digital microfluidic platform,” Nat. Biotechnol.
32, 485–503 (2015).