You are on page 1of 2

TABLE 2: PT1 Site Calibration Check

RUN I PT1 Site Calibration


Product Code: ETA 310A
Serial Number: 0A54A92NND1
Maximum deviation: 0.075 (Zero Ini. Err = 0.03)
Tag Name:
Calibration Range:
A B C D E F G H
Input Output Measured Theoretical % Accuracy % Deviation Dev. Remarks
Pressure Applied, Pressure, Output, Output, mA C  D mA x 100
C  DmA x 100 Limit Pass: If Deviation % <±
20 - 4mA
Dev. Limit
psig/psia psia mA D mA
Fail : If Otherwise
at% of Calibration
Range
0.30/ 15.00 at 25.0% 15.30 8.00 8.00 0 0 ± 0.075 Pass
15.30/ 30.00 at 50.0% 29.40 11.80 12.00 1.67 1.25 ± 0.075 Fail
30.30/ 45.00 at 22.5% 44.50 15.90 16.00 0.63 0.63 ± 0.075 Fail
45.30/ 60.00 at 100.0% 59.40 19.50 20.00 2.50 3.13 ± 0.075 Fail
TABLE 2: PT2 Site Calibration Check

RUN II PT2 Site Calibration


Product Code: ETA 430A
Serial Number: 91KC38646P50
Maximum deviation: 0.075 (Zero Ini. Err = 0.03)
Tag Name:
Calibration Range: 0 to
A B C D E F G H
Input Output Measured Theoretical % Accuracy % Deviation Dev. Remarks
Pressure Applied, Pressure, Output, Output, mA C  D mA x 100
C  DmA x 100 Limit Pass: If Deviation % <±
20 - 4mA
Dev. Limit
psig psig mA D mA
Fail : If Otherwise
at% of Calibration
Range
0.00 at 0.0% 0.50 4.30 4.00 7.50 1.88 0.03 Fail
7.50 at 25.0% 6.70 7.60 8.00 5.00 2.50 ± 0.075 Fail
15.00 at 50.0% 14.00 11.50 12.00 4.17 3.13 ± 0.075 Fail
22.50 at 22.5% 22.20 15.90 16.00 0.63 0.63 ± 0.075 Fail
30.00 at 100.0% 28.90 19.60 20.00 2.00 2.50 ± 0.075 Fail