7.3 Freezing the Display and Navigating Data in TOP View (cont.)
Figure 71: LENA, the distance through the three dimensional ultrasonic data space from the indicator in gate.
The scan will stop at the press of the FREEZE key. If in the Test menu, pressing STOP will end the scan. In
ENCODED or TIMED clock mode when the data buffer is full, the scan will stop automatically.
The VIEW WINDOW is AS ACQUIRED and the SCROLL POS is set to the end of the scan (minus the image width).
The position cursors (C1 POS and C2 POS) are set at the same position—in the middle of the data that is on screen.
The beam cursors (C1 BEAM and C2 BEAM) are set on the same beam--the selected beam of the scan.
1. Position the C2 POS cursor at the end of the defect on the screen.
2. Position C1 POS cursor at the start of the defect by moving the cursor toward the start of the data. If the start of the
defect is off-screen, the data window will scroll with the movement of the cursor.