Carestation 30
Features
GE Datex-Ohmeda family look and quality
One or Two gases: O2+ N2O or Air (optional)
5.7-inch color ventilator display with waveforms
and alarm message indicators
Lightweight and compact for easy maneuverability
One or two vap positions
Dove tails for mounting accessories (optional)
Advanced Ventilation
Ventilation Modes:
- VCV (Volume Control) Revolutionary New Breathing Circuit
WYSIWYG (What You Set Is What You Get) Tidal Easy to clean, fully autoclavable, latex-free
Voulme setting Easy removal-no tools required
- Compensation for breathing circuit
Quick Release for fast remove canister and refill
compliance
soda lime
- Compensation for Fresh Gas Flow
Integrated design-less parts and connections helps
(optional)
reduce potential for leaks and misconnects
Pressure waveform for visual reference on a
One step bag/vent switch turns ventilator on/off
breath-by-breath basis
Passive AGSS (optional)
Standby-mode
Auxilary Comment Gas Outlet (optional)
Physical Specifications Ventilator accuracy
Dimensions Delivery/monitoring accuracy
Height: 148 cm/58 in Volume delivery:
≥ 100 mL = better than 20%
Width: 90 cm/35 in
< 100 mL = better than 20 mL
Depth: 70 cm/28 in Volume monitoring:
Weight: 70 kg/220 lbs ≥ 100 mL = better than 20%
< 100 mL = better than 30 mL
Top shelf Alarm settings
Weight limit: 15 kg/33 lbs Tidal volume (TVexp):
Low: 0 to 800 mL
Width: 40 cm/16 in High: 100 to 1800 mL
Depth: 30 cm/12 in Inspired oxygen (FiO2):
Low: 18 to 99%
Work surface High: 19 to 100%
Apnea alarm:
Height: 74 cm/29 in
In Bag Mode, Apnea alarm happens
Size: 630 cm2/98 in2 under condition of no flow value, PAW
fluctuates less than 2cmH2O, and
Casters continue time exceeds 30 seconds.
12.5 cm/5 in, with brakes on the front casters Airway pressure (PAW):
Low: 4 to 20 cm H2O
Drawers High: 5 to 100 cm H2O
20 cm x 30 cm x 30 cm/8 in x 12 in x 12 in Sustained airway pressure:
Paw ≥ 10cmH2O continuously for 10
Ventilator display seconds
5.7 inch TFT, 640 x 480
Ventilator components
Ventilator Operating Specifications Flow transducer
Type: TVX Flow Transducer Cartridge
Ventilation operating modes Dimensions: 22 mm OD and 15 mm ID/22 mm ID
VCV Location: Expiratory Port
Pulse rate
Displayed range:
30 to 300 beats per minute (bpm)
First reading, full accuracy:
© 2012 General Electric Company - All Rights Reserved.
Healthcare Re-imagined
GE is dedicated to helping you transform
healthcare delivery by driving critical
breakthroughs in biology and technology. Our
expertise in medical imaging and information
technologies, medical diagnostics, patient
monitoring systems, drug discovery, and
biopharmaceutical manufacturing
technologies is enabling healthcare
professionals around the world to discover
new ways to predict, diagnose and treat
disease earlier. We call this model of care
“Early Health.” The goal: to help clinicians
detect disease earlier, access more
information and intervene earlier with more
targeted treatments, so they can help their
patients live their lives to the fullest. Re-think,
Re-discover, Re-invent, Re-imagine.
www.gehealthcare.com