You are on page 1of 4

Cd

The Chrom
Doctor

Ghost Peaks in Gas Chromatography Part 1:


The Carrier Gas and Carrier Gas Lines
Jaap de Zeeuw, Restek Corporation, Middelburg, The Netherlands.

We always get a number of inquiries from customers that see peaks show up in their chromatogram when they do not inject them.
There are several sources for these “ghost peaks” which we will discuss in several articles.

What are Ghost Peaks? • memory effects due to back flash Table 1

“A ghost peak is a peak that appears • contamination of injection port Solvent Expansion Maximum Injection Volume
at a position where we do not expect • Injection port reactivity factor 50% liner volume [μL]
a peak”. It is a component that shows • Septum, O-rings 250 ºC, 73 kPa 1 mm liner 2 mm liner 4 mm liner
up in the system and it may show up • Analytical column bleed
everywhere in the chromatogram. • Column installation Isooctane 153 0.2 0.8 3.3
Sometimes it is a sharp peak, • Sample contamination from syringe, Toluene 237 0.12 0.5 2.0
sometimes it’s a broad peak or a rinse vials, vial-septa or gloves. Methanol 645 0.05 0.20 0.8
“hump” and sometimes it is a rising In this part we will focus on the Water 1399 0.02 0.09 0.34
baseline. carrier gas and the gas lines.
Some of the key origins for ghost Table1: Maximum injection volume.
peaks are: Carrier Gas Purity
• the purity of the carrier gas The carrier gas must be of high content of the carrier gas. Oxygen higher temperatures. Water initiates
• using non-GC grade carrier gas purity. On average attention is paid can oxidize surfaces of liners and hydrolization that can also start at
tubing to the residual oxygen and moisture columns but this usually happens at lower temperatures and often relates

2 Chrom Doctor: The Carrier Gas and Carrier Gas Lines separation science — volume 5 issue 2 www.sepscience.com
Figure 1
at column connections and also the
septum.
Purification of Carrier Gas
One can choose ultrapure carrier
gas and assume a perfect sealed
gas distribution system. There is
practically always a risk that carrier
gases get contaminated. Each time
a cylinder is changed some air can
enter the system. Also there are many
fittings/couplings used in gas lines
that are sources for leaks. If systems
are not used, there is often back
diffusion of air into the gas lines.
Concentration gradients are very
strong drivers for diffusion.
It is highly recommended to use
filtration devices to clean carrier gas.
Figure 1: Formation and reactivity of silanol groups. There are different systems available
to the stability of the deactivations varying from in-line filters to cartridge
used. Hydrolization of liner and glass filters, indicating and non-indicating.
Figure 2
wool surfaces will create adsorption The cartridge filters have become
sites that interact with polar analytes quite popular as filter exchange is
when doing an injection. The result is very fast and there is no down time
lower response, peak broadening and because the main gas lines remain
even discrimination and reactivity can pressurized. An example system is
be enhanced. Hydrolization of the shown in Figure 2. Such systems are
stationary phase is a bigger problem available for filtering carrier gas for
as terminal silanol groups are formed oxygen and moisture. Also charcoal
that will initiate phase degradation filters can be beneficial, especially
(see Figure 1). This reaction is also for filtering FID gases. Charcoal
temperature dependent. Most Rtx/ may also be used for carrier gas line
Rxi phases can deal very well with purification in case hydrocarbon
water, if analysed as compound or contamination is suspected. This can
introduced as the matrix. Water will be the case when gas lines are used
elute at low oven temperatures. If that are not cleaned thoroughly.
water is in the carrier gas, it will also Although molsieve filters have a huge
be introduced when the column is trapping effect for water and linear
at high temperature, initiating the hydrocarbons, they do not have high
Figure 2: Carrier gas purification for O2, H2O and hydrocarbon removal. Most popular are the hydrolization. For the same reason, retention for branched hydrocarbons.
cartridge filters: System remains pressurized; they have indicators and filter exchange is very fast. one must check carefully for leaks Branched hydrocarbons will only be

3 Chrom Doctor: The Carrier Gas and Carrier Gas Lines separation science — volume 5 issue 2 www.sepscience.com
Figure 3 Figure 4

Figure 3: FID bake out of four different batches of copper tubing. Figure 4: A drift of baseline due to overlapping compounds that are moving to the column caused
by back-flash contaminated injection port.

retained by the outer surface of the replace all filters during a period not removed, these hydrocarbons in the next issue. By keeping the
molsieve 5A particles, which means when workload is not that big. will accumulate on the columns and injector cold, the contribution will
they will elute very fast ( for example: cause background and ghost peaks. be minimal. When the oven is set at
on molecular sieve 5A iso-butane Carrier Gas Tubing For new GC lines, make sure that low temperature, the column will act
even elutes before nitrogen at 50 °C, Normally gas lines do not contain GC‑grade tubing is used. as a “trapping device”, meaning that
while n-butane elutes at 300 °C. “hydrocarbons” but that is only true any component that is present or
Filters are best placed as close if GC-cleaned tubing is used. In the How to Test if a Ghost Peak Comes formed in the gas system before the
as possible to the instrument. Try process of drawing metal tubing, a from Carrier Gas Contamination? column, will be trapped (and focused)
always to maintain a positive flow lot of hydrocarbon-based ”greases” The easiest way to test the carrier on the GC column. The amount of
through the filters. Instruments that are used. As a result metal tubing is gas for hydrocarbon contamination component will relate to the time and
are “shut-off” sometimes show back- strongly contaminated and cleaning is to trap the components on the GC flow that is used during the trapping
diffusion of air through regulators, is required. Recently a study was done column and look at the response. In period. After 10 minutes of “trapping”
contaminating the gas lines. Filter on this as a new source for copper order to do this, the injection port the oven is programmed to the upper
life time is usually about one year. tubing became available. Figure 3 and the GC oven plus the column is temperature of the application for
Although filters can be “indicating” via shows a comparison of a “bake cooled down to room temperature. that GC and the baseline is observed.
a change of colour, this often is not out” of similar lengths of copper The injector must be cold as this After this, the oven is cooled again
checked due to the lab’s workload. tubing using FID signal. Note that is another big source for ghost and the trapping time is repeated
If applicable, it’s a good habit to the scale was 7.000.000 counts. If peaks, which will be discussed but now the oven is kept for a

4 Chrom Doctor: The Carrier Gas and Carrier Gas Lines separation science — volume 5 issue 2 www.sepscience.com
longer time at a low temperature, very recognizable spectra for • If the ghost peaks still show up,
such as 30 minutes. Again the oven hydrocarbons. This injection port had you may extend the flushing time.
is programmed and the baseline is to be cleaned and for future work, a
observed. If there are visible peaks smaller amount had to be injected. Jaap de Zeeuw studied six years of chemistry and
on the baseline and these peaks Depending on the injection graduated in 1979. Jaap has 33 years’ experience in
are 3x higher in the second run, port liner, the maximum injected GC capillary technology and has developed many
contaminants are coming from before amount should not be exceeded. PLOT columns as well as bonded-phase columns.
the injection system. The vapour volume of the sample He is also the originator for simple concepts
Filters have to be checked and depends mainly on the amount and for fast GC–MS using a high vacuum inside the
maybe also the tubing from the filter the molecular weight of the solvent capillary column. He has published more than
to the GC inlet. used. Table 1 shows a few maximum 100 publications in the field of GC on column
There may also be another reason amounts related to a 2 and a 4 mm technology and application. He worked for 27 years
for contamination in the carrier gas ID liner. The maximum amount is for Chrompack/Varian and for the last six years
line: this is called back-flash and set to 50% of the liner volume of has served as an international specialist on gas
here often the full range of (heavier) that solvent at given temperature/ chromatography for Restek in The Netherlands.
components show up that are pressure. The biggest risk for back-
typically analysed on this GC. This is flash is with using the splitted
also often called “memory-effect”. injection.
Cleaning a contaminated injection
Back-flash port is best done using the next steps:
Back-flash occurs when the sample • Remove the analytical column
injected exceeds the volume of and replace it for a short length
the liner. The excess sample will be of fused silica capillary (2–3 m
pushed out the liner and can get into from an old column with the same
the carrier gas line. diameter);
Because the carrier gas line • Set the carrier gas at a high
is outside the heated zone, the flow‑rate for the column and split
sample components will condense line. Also make sure the septum
and slowly evaporate. This causes purge is open; (If hydrogen is
“memory-effects” and also impacts used, the column end is best
quantification for single analytes. positioned outside the oven);
If complex samples are analysed, • Heat the injector to a high
the ghost peaks all merge together temperature (325–350 °C) and
and result in an increase of baseline. flush the inlet for 30–60 minutes;
Figure 4 shows an example where Indirectly the carrier gas lines will
the injection port was contaminated also be warmer and back-flashed
with hydrocarbon oils. Usually when residues will evaporate;
observing such a baseline, the first • Cool injector and replace seals,
reaction is that the column bleeds, liner, O-ring and septum.
but in this case the detector was • Connect the column again and
a MS and the TIC signal showed also condition the column;

5 Chrom Doctor: The Carrier Gas and Carrier Gas Lines separation science — volume 5 issue 2 www.sepscience.com

You might also like