Professional Documents
Culture Documents
Senographe DS
Senographe DS
5391271-3-1EN
Revision 1
© 2010-2014 by General Electric Company
All Rights Reserved.
Contrast Enhanced Spectral Mammography 5391271-3-1EN
IMPORTANT...X-RAY PROTECTION
IMPORTANT...X-RAY PROTECTION
CAUTION
If not properly used, x-ray equipment may cause injury. Accordingly, it is your obligation to
confirm that the instructions herein contained are thoroughly read and understood by everyone
who will use the equipment before you attempt to place this equipment in operation. The General
Electric Company, Healthcare Technologies, will be glad to assist and cooperate in placing this
equipment in use.
Although this apparatus incorporates a high degree of certain protections against x-radiation
other than the useful beam, no feasible design of equipment can provide complete protection
from all potential injury. Nor can any feasible design force the operator to take adequate
precautions to prevent the possibility of any persons carelessly exposing themselves or others to
radiation.
It is important that anyone having anything to do with x-radiation be properly trained and fully
knowledgeable about the recommendations of the National Council on Radiation Protection and
Measurements as published in NCRP Reports available from NCRP Publications, 7910
Woodmont Avenue, Room 1016, Bethesda, Maryland 20814, and of the International
Commission on Radiation Protection. It is your obligation and responsibility to take adequate
steps to protect against injury.
The equipment is sold with the understanding that the General Electric Company, Healthcare
Technologies, its agents, and representatives have no responsibility for injury or damage, which
may result from improper use of the equipment. Various protective materials and devices are
available. It is urged that such materials or devices be used in accordance your site’s clinical
practice.
IMPORTANT...X-RAY PROTECTION
Table of Contents
Table of Contents
IMPORTANT...X-RAY PROTECTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Table of Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Foreword . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1. Applicability of Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2. How to order a paper version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3. How to access the electronic version of a manual on a website . . . . . . . . . . . . . . . . . . . 7
4. Legal information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
4-1. Copyright information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
4-2. Trademark information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
5. Regulatory considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
6. Scope of this Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
7. Overview of this Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Chapter 3. Guidance
1. Flat field Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
2. CESM AUTO Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
3. Annual Physicist Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
3-1. Beam Quality Assessment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
4. Breast Entrance Exposure and Reproducibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Table of Contents
Revision History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Foreword
Foreword
1 Applicability of Manual
This document “Quality Control Manual” is an element of the Quality Assurance Plan (QAP) of the
mammographic facility. It describes Quality Control (QC) Tests for Contrast Enhanced Spectral
Mammography application, which constitutes an option of the acquisition system Senographe DS or
Senographe Essential. This Quality Control Manual is an addendum to the Senographe DS Quality
Control Manual or Senographe Essential Quality Control Manual.
Application of this manual is MANDATORY for facilities subject to the rules of the United States
Mammography Quality Standards Act (MQSA).
Application of this manual is RECOMMENDED for facilities in locations not subject to the rules of the
MQSA.
WARNING
Quality control tests must be performed regularly to ensure that CESM operates properly.
search field
Foreword
4 Legal information
4-1 Copyright information
All licensed software is protected by the copyright laws of the United States and by applicable
international treaties.
5 Regulatory considerations
In facilities subject to the MQSA, the procedures in the document “Contrast Enhanced Spectral
Mammography QC Manual” must be followed. Failure to follow these quality assurance procedures can
result in loss of MQSA certification at facilities subject to U.S. regulations.
WARNING
This manual provides appropriate user instructions on Quality Control procedures.
1 Introduction
QC tests are simple checks which ensure that the CESM application is operating to its design standards.
They are designed to detect any changes in settings which might compromise image quality, as well as
any deterioration in equipment performance over time.
QC tests for CESM are described in the following sections:
• Section 4 Flat field Test on page 10. Checks for consistency of image quality.
• Section 5 CESM AUTO Mode on page 12. Checks for correct operation of CESM AUTO mode.
• Section 6 Test Results Record Forms on page 15 provides charts for use in recording the results of
the various checks. It is recommended that you make copies of the chart pages and use them to
record the test results.
For further analysis, data generated on the Acquisition Workstation (AWS) for Flat Field tests can be
exported as text files to a CD-R.
2 QC Intervals
As a minimum, the Quality Assurance Procedures described must be performed at the intervals
specified in the test descriptions and summarized below.
Minimum Procedure Section
Frequency
Weekly Flat field Test 4
Monthly CESM AUTO Mode 5
Refer to Chapter 3 Guidance for additional information regarding the QC Tests results.
3 Data Export
To export the QC data files:
• From the Browser, click the QAP button, then select Extract Data.
• A pop-up is displayed instructing you to insert a blank CD-R. Insert the CD-R and click OK.
This action saves the data generated since the last data export.
Exported files can be easily opened on a computer using a word processing application. They contain
the data displayed on the screen after the test is completed.
Note:
Files can only be exported once.
8. If any test fails, check the test conditions and repeat the test after correction.
- Check that the detector has been allowed to warm up for at least 10 minutes before acquiring the
test images.
- Check that the compression paddle has been removed.
- Check that no object except the Flat field Test Object is present in the field of view.
- Check that the collimator is open to the largest available field size.
- Check that the tube arm angle is at zero degrees.
- Check that the surface of the image receptor is clean.
- Check that the Flat field Test Object fully covers the field of view of the image receptor.
- Check that the Flat field Test Object is clean and free from scratches or other imperfections:
- Clean or replace the test object as necessary.
- If there is a scratch or defect near an edge of the test object, attempt to orient the test object
so that the imperfection is outside the field of view of the image receptor. After re-orienting the
test object, ensure that it still fills the field of view of the image receptor.
• Action Limit:
All Flat field tests must pass.
• Use of Test Results:
If the system fails the test, the source of the problem must be identified, and corrective action taken,
before any further mammographic images are acquired using the CESM application. Refer to
Chapter 3 Guidance section 1 Flat field Test on page 37.
• Action Limit:
- The Auto test is successful if the exposure parameters are within the ranges specified in the
following table:
Acrylic Exposure Parameters for Auto mode
Thickness
Low energy High energy
(mm)
Track/Filter mAs kV Track/Filter mAs kV
25 Mo/Rh 32-56 27 Mo/Cu 56-100 46
50 Rh/Rh 50-89 29 Rh/Cu 128-224 45
60 Rh/Rh 56-100 31 Rh/Cu 144-252 47
• Use of Test Results:
If the system fails the test, the source of the problem must be identified, and corrective action taken,
before any further mammographic images are acquired using the CESM application. See
Chapter 3 Guidance section 2 CESM AUTO Mode on page 37.
Chapter 1
Frequency: Weekly
Room: ......................... Unit: .........................
Year
Date
Initials
Flat field MoCu Result Pass/ Result Pass/ Result Pass/ Result Pass/ Result Pass/ Result Pass/
Fail Fail Fail Fail Fail Fail
brightness non-
uniformity
High Frequency
Modulation
SNR non-uniformity
Flat field RhCu Result Pass/ Result Pass/ Result Pass/ Result Pass/ Result Pass/ Result Pass/
Fail Fail Fail Fail Fail Fail
brightness non-
uniformity
High Frequency
Modulation
SNR non-uniformity
Year
Month
Date
Initials
CESM AUTO Mode Check; mAs or F/n. See Chapter 3 Guidance section 2 CESM AUTO Mode on
page 37 regarding a method to record these results.
Track/ mAs kV Pass/ Track/ mAs kV Pass/ Track/ mAs kV Pass/
Filter Fail Filter Fail Filter Fail
Low energy:
25 mm acrylic
50 mm acrylic
60 mm acrylic
High energy:
25 mm acrylic
50 mm acrylic
60 mm acrylic
Notes:
1 Introduction
The QC tests listed in this section must be performed by the Medical Physicist to ensure that the CESM
application provides an acceptable level of mammographic image quality. These tests also form the
basis of the mammography equipment evaluation (MEE) that must be performed following the
installation of a new mammography system or the repair or replacement of a major component of the
system.
The QC tests to be performed by the Medical Physicist are listed in the following three tables.
- The following tests are described in Chapter 1 QC Tests for the Radiologic Technologist:
Radiologic Technologist’s QC section Minimum Section
Frequency
1. Flat field Test Annually Chapter 1 section 4 Flat field Test, page 10
2. CESM AUTO Mode Annually Chapter 1 section 5 CESM AUTO Mode, page
12
- Additional tests that must be performed by the Physicist are listed below.
Specific Medical Physicist tests Minimum Section
Frequency
3. Artifact Evaluation; Flat Field Annually Job Card VF-CESM02 - Artifact Evaluation
Uniformity and Flat field Uniformity on page 27
4. Breast Entrance Exposure and Annually Job Card VF-CESM01 - Breast Entrance
Reproducibility Exposure and Reproducibility on page 21
5. Beam Quality Assessment Annually Job Card VF-CESM03 - Half-Value Layer
(Half–value Layer Measurement) Measurement on page 31
Tests 3, 4 and 5 are described in Job Cards VF-CESM01 through VF-CESM03.
Charts for recording the results of the tests are provided as Charts 0-3.
Facility Name:
Address:
Date of Installation
Date of Survey
Room ID
Mammographic Unit Serial Number
Medical Physicist
Signature
• Objective:
To measure the typical entrance exposure for an average patient (approximately 4.2 cm compressed
breast thickness, 50% adipose, 50% glandular composition), and to assess short-term exposure
reproducibility.
• Required Test Equipment:
- Ionization chamber and electrometer calibrated at mammographic X-ray beam energies
(calibration factor constant to within ± 1% over the Half-Value Layer (HVL) range from 0.2 to
5 mm Al)
Note:
Not all dosimeters are compatible with such a large range. In particular, when using a semi-
conduction sensor, ensure that it can maintain its accuracy over that HVL range.
Note:
CESM acquisitions are made of two consecutive exposures: Low Energy and High Energy. This
procedure requires the total entrance dose to be measured, which includes both Low Energy and
High Energy exposures. Check if the dosimeter is capable of integrating both exposures in a single
measurement. If the dosimeter is resetting the dose measurement automatically after each
exposure, only the entrance dose corresponding to high energy exposure will be considered. In
that case, the corresponding low energy entrance dose can be deduced from a separated manual
exposure.
- Mammographic phantom (equivalent to approximately 4.2 cm compressed breast tissue 50/50
composition at mammographic energies). Acceptable phantoms are listed in Chapter 3 Guidance
section 4 Breast Entrance Exposure and Reproducibility on page 39.
• Procedure:
1. Prepare the mammographic imaging system as follows:
- Install the Bucky for contact mode.
- Select 24 cm x 31 cm field of view for Senographe Essential or 19 cm x 23 cm field of view for
Senographe DS.
2. Position the mammographic phantom on the breast support surface of the image receptor
assembly, laterally centered in the X-ray field with one edge coincident with the chest wall edge of
the breast support surface. When viewed from the patient’s position, i.e., facing the
mammography unit, the cut-out corner of the wax insert of the phantom must be at the chest wall
side and toward the right side of the detector (see Illustration 1).
3. Position the ionization chamber in the X-ray field with its center approximately 4 cm in from the
chest wall edge of the image receptor and with the top surface of the chamber near the top
surface of the phantom. The interior edge of the chamber must be no more than 3.5 cm from the
lateral edge of the X-ray field as indicated by the light localizer (See Illustration 1 and Illustration
2).
4 cm
3.5 cm
Ion chamber Ion chamber must not be
placed further inside the field
Chest wall edge of of view
image receptor
30.7 cm for Essential / 23 cm for DS
Phantom
Illustration 1 - Set up to monitor output reproducibility in CESM AUTO mode (top view)
Paddle
Ion chamber (5 daN compression)
Phantom
Bucky
30.7 cm
6. Select the image acquisition mode, most commonly used clinically, i.e., CESM AUTO or CESM
Manual, to image the standard breast equivalent phantom using the CESM application. For
Manual mode select the clinically used settings of kVp, anode track, X-ray beam filter, and mAs.
Record the settings on the data form.
7. Make an exposure in CESM mode.
8. Record the mAs for Low and High energy, and the reading of the ion chamber E (Entrance Skin
Exposure), and the selections of kVp, anode track, and filter.
9. Make additional exposures and record the results until a total of four exposures have been
recorded.
10. Compute the averages and standard deviations for mAs and exposure for the four exposures
acquired with identical technique settings. Record the values. Determine the coefficients of
variation (standard deviation divided by the mean).
• Action Limit: mAs and Air Kerma Reproducibility
The maximum acceptable coefficient of variation for both mAs and air kerma in the reproducibility test
is 0.05.
• Use of Test Results:
If these conditions are not met, the source of the problem must be identified, and corrective action
taken, within 30 days of the test date.
Exposure #1
Exposure #2
Exposure #3
Exposure #4
Mean Values
Standard Deviations (SD)
Coefficients of Variation (CV)
Action Limit:
The Coefficient of Variation for both air kerma and mAs must not exceed 0.05.
Use of Test Results:
If these conditions are not met, the source of the problem must be identified, and corrective action
taken, within 30 days of the test date.
• Objective:
To assess the degree and source of artifacts visualized in full field digital mammograms or phantom
images.
To assure that the Flat field image is uniform.
• Required Test Equipment:
- A 25 mm thick uniform sheet of acrylic (PMMA) that is clean and free from defects. The acrylic
sheet must be sufficiently large to cover the entire image receptor.
Note:
To allow for temperature stabilization of the detector, the system must be powered on for at least
10 minutes before performing any measurements related to detector image quality. If any test is
not passed after allowing a 10-minute warm–up period, see Chapter 3 Guidance section 5 Artifact
Evaluation and Flat field Uniformity on page 39.
• Procedure:
Note:
Since it is not possible to perform a single energy acquisition, the tester shall perform a CESM dual
energy acquisition and take into account only the high energy image.
1. Select technique factors normally used clinically in the CESM application, choosing the lowest kVp
setting (to be most sensitive to artifacts) for high energy (any parameters for low energy since it is not
assessed in this test). Record the high energy technique factors on the data form.
2. Install the Bucky. Place the attenuator sheet on top of the image receptor. Set collimation that
permits irradiation of the entire image receptor.
3. Make a CESM exposure.
4. Repeat steps 1 through 3 for the Mo/Cu and Rh/Cu target/filter combinations.
5. Select the High energy Raw image for review.
6. Set the window width in the range of about 100 to 150 and the window level to a setting that allows
visualization of artifacts if present. Examine the image for gridlines, plus or minus signal variations,
streaking in the horizontal or vertical directions, residual images left over from repeated test or clinical
exposures, acrylic sheet artifacts, or bad pixels.
7. With a window width in the range of about 100 to 150, examine the image for Flat Field uniformity.
8. Review the images for the two target/filter combinations.
9. Review at least one of the images on all display devices, e.g., acquisition workstation, laser printer.
10. If any artifacts or non-uniformities are present, rotate the acrylic sheet 90° and repeat the exposure
and acquisition procedure.
11. If any artifacts or non-uniformities keep a fixed orientation relative to the image receptor in both
images and are expected to simulate or mask visualization of breast structures or breast pathology,
determine whether the artifact or non-uniformity is due to the detector or the display.
If and then
The artifact or non- the artifact can be related a. Remove the PMMA sheet and the
uniformity is present on all to a latent image from Bucky.
monitors or the artifact previous exposures, for b. Remove the compression paddle.
moves as the image is example the artifact has c. Make sure that no object is present in
moved using the Roam a breast shape, or looks the X-ray field of view.
tool, it is most likely like collimator blades
d. Make five exposures using the following
associated with the image corresponding to a
technique: Rh/Rh, 30kV, 125mAs.
acquisition system collimated view
The artifact appears only - Clean the faceplate of the display (See
on one monitor Senographe DS Quality Control Manual or
Senographe Essential Quality Control Manual)
The artifact appears only - Follow the directions provided by the
on printed images manufacturer of the printer
12. If any corrective actions in the table above were taken, repeat the artifact evaluation and review the
image.
• Action Limit:
There must be no artifact or non-uniformity that could mimic or obscure clinical information.
• Use of Test Results:
If the artifact or non-uniformity is still present, and can mimic or obscure clinical information, the
source of the problem must be identified and corrective actions taken within 30 days of the test date.
See Chapter 3 Guidance section 5 Artifact Evaluation and Flat field Uniformity on page 39.
Mo/Cu Rh/Cu
kVp (high energy)
mAs (high energy)
Artifact visible?
Flat Field Non-Uniformity
Equipment Artifact
Detector
Grid
Attenuator Defect
Other
Description of Artifacts:
• Objective:
The purpose of this procedure is to measure the Half-Value layer (HVL) of the X-ray beam in High
Energy (HE) configuration. An acceptable value of HVL will ensure patient safety by demonstrating
the combination of kVp and filtration. A follow up of the HVL value with time will ensure the stability of
the equipment in the HE configuration.
• Required Test Equipment:
- Standard 24 x 31 compression paddle for Essential systems and Standard 19 x 23 compression
paddle for DS systems
- Dosimeter calibrated over the HVL range 1.5 to 5 mm Al.
Note:
Mammographic ionisation chamber dosimeters generally cover both usual mammographic energy
range (LE) and the extended High Energy range. The latter may require using a general
radiography dosimeter instead of a mammographic dosimeter. In particular, when using a semi-
conduction sensor, ensure that it can maintain its accuracy over that HVL range.
Note:
CESM acquisitions are made of two exposures, Low Energy and High Energy. Check if the
dosimeter is capable to integrate both exposures, or gives provision to select first or second, or to
measure both independently.This Job Card addresses only the second exposure (High Energy).
In regular use, the High Energy (HE) exposure (kV>45, Cu filter) of a CESM acquisition cannot be
performed separately, but only following the Low Energy (LE) exposure. When using a dosimeter
integrating the two exposures (in particular ionization chamber dosimeters) the HVL in High
Energy conditions can be measured by:
- minimizing the Low Energy contribution by adjusting the related parameters to the minimum
possible (22 kV, 4 mAs for Molybdenium, 25 kV, 4 mAs for Rhodium);
- measuring separately the air kerma relative to the Low Energy exposure in Routine (non
CESM) mode;
- substracting this Low Energy contribution from the total air kerma measured in CESM mode.
Note:
The Low Energy contribution affects mainly the exposure without added aluminium filter. It
becomes more and more negligible when additional aluminium filtration is placed in the beam.
- Plates of aluminum (type 1100 aluminium alloy, 99.0% Pure), allowing to reach a total thickness
of, 2.0, 3.0 and 4.0 mm, of length and width sufficient to fully cover the ionization chamber. The
stated thickness should be accurate within ±1%.
- A regular HVL measuring stand as provided by the dosimeter manufacturer.
Alternatively, the standard compression paddle can be used to hold the aluminium filters.
- It is recommended to protect the detector from raw high level radiation during HVL measurement
by placing a high attenuation metallic plate on the breast support such as the radiation protection
plate.
• Procedure:
1. Install the Bucky on the image receptor.
2. Place the radiation protection plate on the surface of the Bucky. Each side of the radiation protection
plate must extend about 4 cm beyond each edge of the light field.
Note:
The X-ray doses used in this procedure can saturate the image receptor if the radiation protection
plate is not used and not correctly positioned. If the radiation protection plate does not completely
cover the field of view, a ghost image may be imposed on the digital image receptor at the edge of
the radiation protection plate.
Note:
It is recommended to place a mechanical protection (paper, cardboard, cloth, etc.) between the
radiation protection plate and the breast support to avoid possible damaging.
3. Position a HVL filter holder. If none is available, the Aluminium plates will be placed on the
compression paddle.
4. Install the 19 x 23 cm compression paddle for Senographe DS or the 24 x 31 cm compression paddle
for Senographe Essential, and position it approximately half way between the tube head and the
dosimeter.
5. Ensure that paddle decompression at end of exposure is set to “NO”. From the X-ray Console menu,
press the SET UP menu key , then select MEDICAL/DECOMP/DECOMP/NO. Position the
compression paddle plane at approximately 30 cm from the focal spot.
6. Select the 9 x 9 collimator format.
7. Using the light beam, position the dosimeter sensor in the center of the X-ray field.
If the sensor is a semi-conductor, it can be placed directly on the radiation protection plate. If an
ionization chamber is used, it is recommended to place it at approximately 10 cm above it, using a
block of light foam (e.g. expanded Polystyrene) or a holder out of the X-ray beam. Secure the
chamber in position and do not change the position of the chamber during the following
measurements.
8. Select the CESM Manual image acquisition mode.
9. Select the Molybdenum anode.
10. Adjust the Low Energy parameters to the minimum possible (22kV, 4 mAs).
11. Adjust the High Energy high voltage to 47 kV. If the dosimeter is an ionization chamber integrating
the two successive LE and HE exposures, select 160 mAs so that the LE air kerma is small relative
to the HE air kerma.
12. Make an exposure and record the settings on the Mo table of the data form with the corresponding
measured air kerma (line Total CESM “E0” in Chart 3. Half Value Layer Test Record on page 35).
13. Position the 2 mm sheet of aluminium in the filter holder or on the compression paddle. Using the
light beam, check that the x-ray beam is entirely blocked.
14. Make an exposure and record the settings on the data form with the corresponding measured air
kerma (line Total CESM “E (2 mm)” in Chart 3. Half Value Layer Test Record on page 35).
15. Repeat steps 13. and 14. with respectively 3 and 4 mm aluminium sheets.
16. Go to regular non CESM (manual) mode.
17. Repeat steps 9. to 15., with the exception of 11. and write down the measured doses in the
corresponding LE lines of Chart 3. Half Value Layer Test Record on page 35.
18. Subtract the LE values from the Total values, and write down the difference in the HE lines of Chart
3. Half Value Layer Test Record on page 35.
19. Determine the aluminium thicknesses T1 and T2 that give dose values E1 and E2 which are respectively
just above and just below the half of E0 (in general T1 = 3 mm and T2 = 4 mm).
20. Compute the HVL using the following formula and record the value in the line HVL of Chart 3. Half
Value Layer Test Record on page 35.
E1 E2
T 2 ln 2 ------ – T 1 ln 2 ------
E 0 E 0
HVL = -----------------------------------------------------------------------------
E1
ln ------
E2
21. Select the rhodium anode; adjust the Low Energy parameters to the minimum possible (25 kV,
4 mAs).
22. Repeat steps 11. to 20. using the Rh table of the data form
• Action Limit:
The HVL at 47 kV should be larger than 3 mm for both Mo and Rh targets.
• Use of Test Results:
If these conditions are not met, the source of the problem must be identified and corrective action
taken.
Frequency: Yearly
Room: ......................... Unit: .........................
Date
Initials
HVL MoCu
Total
E0 LE
HE
Total
E (3mm) LE
HE
Total
E (4mm) LE
HE
Total
E (5mm) LE
HE
HVL MoCu Result Pass/ Result Pass/ Result Pass/ Result Pass/ Result Pass/
Fail Fail Fail Fail Fail
HVL (mm)
Date
Initials
HVL RhCu
Total
E0 LE
HE
Total
E (3mm) LE
HE
Total
E (4mm) LE
HE
Total
E (5mm) LE
HE
HVL RhCu Result Pass/ Result Pass/ Result Pass/ Result Pass/ Result Pass/
Fail Fail Fail Fail Fail
HVL (mm)
Guidance
Chapter 3 Guidance
United States regulations consider the image receptor manufacturer’s QC Test procedures as a
requirement for compliance with the MQSA regulations. This Guidance section is not considered an
element of those regulations. It is intended to provide guidance to mammography facilities and their
personnel. It represents the image receptor manufacturer’s current thinking on appropriate procedures
for conducting the QC tests. Procedures described in guidance represent supplementary information on
acceptable ways of doing a test but, unlike regulations, do not bind the facility to using that way only. An
alternative procedure may be used if such a procedure satisfies the requirements of the applicable
statute, regulations, or both. Mandatory language, such as “shall,” “must”, and “require”, is used when
referring to statutory or regulatory requirements. Non-mandatory language, such as “should”, “may”,
“can” and “recommend” is used when referring to guidance. It is the responsibility of the facility to read,
understand, and follow the final regulations.
Under its own authority, a state may impose more stringent requirements beyond those specified under
MQSA and its implementing regulations. A facility may want to check with the state or local authorities
regarding their requirements. Tests may be performed more frequently than specified in the QC tests if
required by local regulations or hospital policies.
QC tests may be used as a check for correct operation, for example, after a change of operating
parameters.
The following sections provide specific guidance for sections in the Contrast Enhanced Spectral
Mammography QC tests.
Refer to Senographe DS or Senographe Essential Quality Control Manual for a global Mammography
system guidance.
Guidance
Guidance
Guidance
Revision History