Step Action
Step Action
Select Rgts/Cal
Proceed to next
maintenance
task
Reinstall covers
Proceed to next
maintenance
task
Select:
Utils from the menu bar
2 Maintenance
7 Syringe Rod Replacement
Continue
Wear the wrist Remove the old complete sample syringe assembly:
ground strap when Unscrew the syringe barrel to release it from the top
performing this Luer-lock fitting
procedure. Unscrew the round plunger end
Discard the complete sample syringe assembly
Select OK
Select Close
Proceed to next
Record completion of this procedure in the
maintenance
instrument Action Log
task