User’s Manual
Electrical Safety Type: Class type for the ability of electric shock protection is I, B,-type,
permanent installation of common equipment.
Work system in standby mode and in accordance with a continuous load for the grid
connection.
Manufacturer is responsible for the safety, reliability and performance only with following
conditions:
I
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
TABLE OF CONTENTS
CHAPTER 1 General Description ...................................................................................................................1
1.1 Usage of the Equipment .................................................................................................................... 1
1.2 Product Standard ............................................................................................................................... 1
1.3 Main Structures ................................................................................................................................. 1
1.4 Feature of structure ........................................................................................................................... 2
i
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
7.2.1 Description of symbols on remote console and local control panel ..................................... 39
7.2.2 Power on/off Operation ........................................................................................................ 42
7.2.3 Spot film operation............................................................................................................... 42
7.2.4 Table body control operations .............................................................................................. 42
7.3 Operation Instructions for high-voltage generator .......................................................................... 43
7.3.1 Structure and panel Diagram ................................................................................................ 43
CHAPTER 10 Maintenance...........................................................................................................................59
10.1 General Rules ................................................................................................................................ 59
10.2 Maintenance for equipment .......................................................................................................... 59
10.3 Maintenance and Repair for collimator......................................................................................... 60
10.4 maintenance and repair for X-ray tube components ..................................................................... 60
10.5 Maintenance and repair for high-voltage generator ...................................................................... 61
10.6 Maintenance and repair for tables ................................................................................................. 62
10.7 Maintenance Cycle........................................................................................................................ 62
ii
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
FOREWORD
At the first, thank you and your company for choosing our products!
Before the usage of this equipment, Please be carefully read the "User Manual",in particular, the
following precautions, in order to guide you through the proper use of this device.
Safety Note:
◆ In order to properly use this product please read the "safety considerations" carefully
before usage.
◆The following advisory symbols will be used throughout this manual. Their application and
ADVISORY SYMBOLS
iii
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
Please carefully read the following notes to the proper use of the device.
Use of this device, please fully understand the performance and operation of this
The device settings and storage areas should pay attention to the following matters:
·Do not place where inflammable and explosive materials used preserved;
·Do not use and preserve when the temperature, humidity, air pressure exceed the allowable
range of environments;
·Please note that the frequency and voltage of power supply and allow the current;
·Make sure that equipment and power connections, to check equipment is working
properly;
·Failure, do not casually dismantled move, should read, "there are failures!" Note the
word .Attached to the device, approved by the Company to send the designated
Maintenance, examination:
·Before operate equipment after it lays idle for long time, be sure to confirm whether the
iv
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
Figure1-1
1
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
installer will remove short wiring between the 4 and 5 on the terminal plate X12, connect the
normally closed port of the temperature control switch of X-ray tube assembly.
3
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
Power condition:
a) Three phase four line, AC 380V ± 38V, 50Hz ± 1Hz
b) Power inner resistance no more than 0.17Ω
c) Power content no less than 65kVA
Using environment:
Environment temperature: 10℃-40 ℃
Relative humidity: 30%-75%
Atmosphere air pressure range: 70KPa-106KPa
Attention:
● X-ray tube assembly equipped to this equipment (hereinafter referred to as assembly) only
● Using the assembly in the environment with inflammable or poisonous gas is not allowed.
The operation environment temperature is 10℃~40℃, relative humidity is 30%~85% and
the atmospheric pressure is (70~106)kPa.
As for transportation and storage environment conditions, the temperature range is -20℃~
70℃, the relative humidity is 20%~90% and the atmospheric pressure is (50~106)kPa.
● Inside the assembly, there’s glass X-ray tube mounted, hence it shall be handled gently
● As per the national regulations, installation, adjustment and using of the assembly shall be
5
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
● 6. Insulation resistance of stator between outgoing line and tube case: 2MΩ
(Min)
● 7. Total filtration of assembly: ≥2.4 mmAl (75kV)
● 8.Radiation protection of the assembly shall meet GB9706.12 III. General
Requirements on Radiation Protection of X-ray Diagnosis Equipment of Part I: General
Safety Requirements of Medical Electric Device (idt IEC601-1-3), i.e. 150kV 3.4mA.
● 9. Radiation field of assembly: when SID1000mm, 430mm×430mm
● 10.Assembly weight (approximate): 16kg
● 11.High voltage cable interface of the assembly shall comply with IEC 60526.
● 12.Cooling medium: transformer oil with temperature ≤70℃
● 13. Cooling method of assembly: natural cooling
● 14.Thermal capacity of assembly: 900kJ (1250kHU)
● 15. Max. cooling velocity of assembly (natural cooling): 180W(16HU/min)
● 16. Refer to Figure2- 1 for emission characteristics of X-ray tube.
Focal Spot:1.2mm
FILAMENT VOLTAGE (V)
TUBE CURRENT (mA)
6
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
Focal Spot:0.6mm
● 18. Refer to Figure 2-3 for anode thermal characteristics of X-ray tube.
7
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
NC: No connection
ET: Grounding terminal
M: Main terminal of stator
A: Auxiliary terminal of stator
9
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
10
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
11
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
Figure 3-1
12
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
Figure 3-2
3. Console nameplate (Figure 3-3)
Figure 3-3
5. X-RAY Tube assembly nameplate (Figure 3-4)
Figure 3-4
13
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
Figure 3-5
7. Electric Collimator nameplate (Figure 3-6)
Figure 3-6
8. Workstation nameplate (Figure 3-7)
Figure 3-7
14
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
Figure 3-8
The location of diagnostic table nameplate
15
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
focus mA kV S
16
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
CHAPTER 5 Collimator
1 - Display of SID and X-ray field size 2 - Lighting delay button 3- Vertical lead leaf adjust knob
4 Horizontal lead leaf adjust knob
Figure 5-1
1. Radiation field
a. Radiation field shape: Rectangle;
b. MIN:0×0(When Blade overlap)
c. Max:345mm×345mm(SID:835mm)
2.Visible light illuminate Brightness:
Centre average illuminate Brightness: >100Lux
Than the degree according to: >4
3. Light: halogen lamp, 24V, 100W
4. Irradiation time : <45S
5. Filtration :
Inherent filtration:1.5mmAl
6. Leaf Shield lead equivalent:3mmpb(intermediate blade 3mmpb)
7. Leaf blade Drive method:
17
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
Remote console control button operation and motor drive set by the closed-loop control
8. Contour Size: 225mm×262mm×210mm (LXWXH)
9. Weight: about 9.0kg;
10. Power Supply: DC24V
1. Temperature range:+10℃~+40℃
2. Relative Humidity: Range :30%-75%
3. Atmospheric pressure: 70kPa~106kPa
The Packed products can be put in the follow environment term within 15 weeks:
1. Temperature range :-40℃~+70℃
2. Relative Humidity: Range:10%-100%
3. Atmospheric pressure: 50kPa~106kPa
4. Stored in the No corrosive gas and well ventilated room.
5.3 Use and installation instructions
1. With four M6×25 countersunk head bolt to fix the collimator flange plate on the tube
assembly, then fix the beam applicator on the plate, when fixing, please make sure good matching
to avoid the collimator fall down.
2. Radiation vision field is determined by collimator splitter radiation field, adjust the vision
field by press the lead blades, also there are knobs on console to control the radiation field. In the
spot film operation, radiation field is controlled by closed loop through the controller, its parameters
are pre-set in the controller.
3. There are two air plugs in collimator, 19-pin for control line and 2-pin for the lights, ensure
both plugs are in good installation before usage. Collimator parameters have been adjusted before
leaving factory. After installation, before usage, need to calibrate to eliminate the errors caused
by improper installation. Debugging methods are as follows:
1) Connect Collimator with the X-ray tube assembly, turn on collimator lights, and make
sure the pre vision field cross-line axis coincides with the centerline of the table.
2) Adjust the cross center line of both radiation field and pre vision field and get them
overlap.
3) Take out one cassette and install one X-ray film, place it on the table vision area.
4) Press “parameters” on touch screen of the console, find "exposure debug" on top of
the page, press it to change it into " debug Cancel ", then put an empty film cassette to the
Spot-film device , put metal wire (diameter thicker than 1 mm is better) on both film cassette center
of the table and opposite angle, To the simulate center line and the radiography scope, carrying
out the entire film spot film exposure; get the x ray film developed, check if the cross line is in the
radiation field ( ie, exposure area).
4. When in Fluoroscopy mode, adjust vertical and horizontal lead leaf to change vision size;
when it is in spot film mode, collimator can track automatically according to the selected slice
mode, Manual adjustment is not supported.
18
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
6.1 Summary
This equipment must be authorized by the Company or the company's professional installers
to install and adjustment.
Before opening the box ,the box should be stored in water-resistant and non-eolian
environment。 Open the box out of the box boars should be reasonable. Removal of all components
should be careful not to scratch the surface of parts, check packing before installation, for any
questions, and please notify the Company in time so we can deal with it.
6.2 Room Layout (Figure 6-1)
Figure 6-1
19
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
6.3 Installation
Power distribution box required the preparation of breaking current 100A, rated voltage 500V,
rated current 60A three-phase air-switch, as chapter switch with power mains, power distribution
box must be firmly fixed on the wall with no less than 1.5m distance from ground level.
1. Installation of diagnostic table
1) Room requirements
a) Equipment install room area at least 5000×4000mm;
b) Control room area 5000×2500mm, control room should be connected with equipment room
and should has lead glass window;
c) Equipment room clearance height at least 2800mm;
d) Equipment room door, width 1500mm, 2300mm high;
e) Diagnostic of table should be directly fixed on the level ground and can load 1500KGS;
f) With cable channel, cable channel cover flushed with the ground;
g) Room should be equipped with special grounding wire, grounding resistance less than 4Ω;
h) X-ray room protection capacity shall conform to the provisions of protection requirements.
2) Installation
a) Install diagnostic table, electric control cabinet, and the console in accordance with the
room layout requirement. Connect all wires with circuit guide. Remove the board which is used for
fixing film cassette, then tighten the screws from the bottom again.
b) Switch on equipment, get the table tilt to 90 °, remove mounting plate of image intensify
from the bottom of film cassette, install image intensifier with the M4 screw, fix image intensifier
with the installation plate on bottom of the film cassette, also install camera on image intensifier.
c) Install tube, collimator.
20
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
6.4 Adjustment
Press the key on console and power on,release until digital led display 0000A,
after 3 seconds digital led display EXXXX,the detail as show in table 1:
Table 1
Display codes Description
E0011 Communication fault between the spot film device and collimator
E0101 Communication fault between the spot film device and table
Press the key again, enter into service mode and the digital LED display 0100A。
2. The corresponding action function and buttons in the service mode are described as following table 2.
Table 2
Display
Function Description
codes
Vertical 2-frame format 4-frame format selection
0100A
Table tilt direction adjustment Table tilt up Table tilt down
0400A Tube stand column longitudinal movement Tube stand column move Tube stand column move
direction adjustment up down
0500A Large division mask transversal movement Large division mask Large division mask move
direction adjustment move outside inside
21
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
1. With four M6×25 countersunk head bolt to fix the collimator flange plate on the tube
assembly, then fix the beam applicator on the plate, when fixing, please make sure good matching
to avoid the collimator fall down.
2. Radiation vision field is determined by collimator splitter radiation field, adjust the vision
field by press the lead blades, also there are knobs on console to control the radiation field. In the
spot film operation, radiation field is controlled by closed loop through the controller, its parameters
are pre-set in the controller.
3. There are two air plugs in collimator, 19-pin for control line and 2-pin for the lights, ensure
both plugs are in good installation before usage. Collimator parameters have been adjusted before
leaving factory. After installation, before usage, need to calibrate to eliminate the errors caused
by improper installation. Debugging methods are as follows:
1) Connect Collimator with the X-ray tube assembly, turn on collimator lights, make sure
the pre vision field cross-line axis coincides with the centerline of the table;
2) Adjust the cross center line of both radiation field and pre vision field and get them
overlap.
3) Take out one cassette and install one X-ray film, place it on the table vision area.
4) Press “parameters” on touch screen of the console, find "exposure debug" on top of
the page, press it to change it into " debug Cancel ", then put an empty film cassette to the
Spot-film device , put metal wire (diameter thicker than 1 mm is better) on both film cassette center
of the table and opposite angle, To the simulate center line and the radiography scope , carrying
out the entire film spot film exposure; get the x ray film developed, check if the cross line is in the
22
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
or in local control panel, table will tilt up or down as same as show in the button direction,
if not, exchange the any two wires of tilt motor at terminal block X3 on the base of table body.
(2)Table tilt range limit adjustment: the optocoupler switches assembly mounted on the
base of the worm speed reducer is used as tilt limit device to control the movement range, as
shown in Figure 6-3.
1-Tilt emergency stop protection switch 2-90°optocoupler limit switch 3--25°optocoupler limit switch
Under normal operating conditions, inching the buttons and table tilt up or down, then insert
the thin shading plate into the appropriate optocoupler infrared transmission slot, the table will
stop tilting, else the table control circuit is not normal.
Finally, adjust the position of each optocoupler switch, in order to meet the need of the tilt
angle.
(3)Table tilt function check: After completion of adjustment, carry out the function test of
23
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
table tilting:
Press the button or on remote control panel or local control panel, the table will
tilt down or up and stop immediately when release the button .
The table will stop automatically, when tilt up to +90 degrees or tilt down to -25 degrees,
including: If start tilt down from above level position, when reach 0 degrees(horizontal position)
the table will stop, to continue tilt down to any negative angle , you must first release the button
and press the tilt down button again(especially operating in the remote console, the
start button must be pressed simultaneously), then the table continue tilt down until reach
negative angle protection limit position or -25 degrees. At this point, if the tube stand column
move upward to the camera system protection limit position, the table can’t reach -25 ° and will
stop at negative angle protection limit position to protect Image Intensifier camera system.
2) Tube stand column longitudinal movement adjustment
(1)Longitudinal movement direction adjustment: when the Tube stand column is not at
upward or downward limit position, jog the tilt button or in local control panel. Tube
stand column will move up or down as same as show in the button direction, if not, exchange the
any two wires of Tube stand column longitudinal drive motor at terminal block X3 on the base of
table body.
(2)Longitudinal movement range limit adjustment: the optocoupler switches assembly
mounted on the body of the table is used as longitudinal limit device to control the movement
range, as shown in Figure 6-4.
1-upward limit switch 2-camera system protection limit switch 3-downward limit switch
Under normal operating conditions, tilting the table to horizontal position, inching the
longitudinal movement buttons and Tube stand column move upward or downward, then insert
the thin shading plate into the appropriate optocoupler infrared transmission slot, the Tube
stand column will stop moving, else the table control circuit is not normal.
Finally, adjust the position of each optocoupler switch, in order to meet the need of the
longitudinal movement range(720 ㎜).
(3)longitudinal movement function check:After completion of adjustment, carry out the
24
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
Press the button or on remote control panel or local control panel, the tube stand
column will move upward or downward and stop immediately when release the button.
The tube stand column will stop automaticlly when reach up to 720mm or down to 0mm of
movement range, else: If table tilt down to negative angle protection limit position, the tube
stand column can’t move up to the upward limit position and just to camera system protection
limit position, so the movement range be less than 720mm, to protect Image Intensifier camera
system.
3) Tabletop transversal movement adjustment
(1) Tabletop transversal movement direction adjustment: when the Tabletop is not at left or
right limit position, jog the tilt button or in local control panel, Tabletop will move
outside or inside as same as show in the button direction, if not ,exchange the any two wires of
Tabletop transversal drive motor at terminal block X2 on the base of table .
(2)Tabletop transversal movement range limit adjustment: the optocoupler switches
assembly mounted on the foot end of the table body is used as transversal limit device to control
the movement range, as shown in Figure 6-5.
Under normal operating conditions, tilting the table to horizontal position, inching the Tabletop
transversal movement or buttons and Tabletop move outside or inside, then insert
the thin shading plate into the appropriate optocoupler infrared transmission slot, the Tabletop will
stop moving, else the table control circuit is not normal.
Finally, adjust the position of each optocoupler switch, in order to meet the need of the
transversal movement range (220mm).
(3)Tabletop transversal movement function check:After completion of adjustment, carry out
the function test of transversal movement:
Press the button or on remote control panel or local control panel, the tabletop
will move outside or inside and stop immediately when release the button.
25
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
The tabletop will stop automatically, when reach to outside or inside limit position.
4) Compressor movement adjustment
(1) Compressor movement direction adjustment: when the Compressor is not at limit position,
jog the tilt button or in remote control panel, Compressor will move backward or
forward as same as show in the button direction, if not ,exchange the any two wires of
Compressor drive motor at terminal block X2 on the base of table .
(2) Compressor movement range limit adjustment: the micro switch mounted on the inside of
the compressor is used as backward limit device to control the movement range, as shown in
Figure 6-6.
Under normal operating conditions, tilting up the table to vertical position, inching the
button or in remote control panel and compressor move backward or forward, then
press the compressor backward limit switch by finger, the compressor will stop moving backward,
else the table control circuit is not normal.
5) Foot stand rotate direction adjustment
Twist left or right the multi-function switch handle on remote control panel, Foot stand will
rotate left or right, if not,exchange the any two wires of foot stand drive motor at terminal block X1
on the base of table .
6) Tube Lifting adjustment:
Tube lifting direction adjustment: That must be under the debugging status.
26
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
Press the button" " enter into the debug interface, loosen the button " " when
buzzer rang, then Press the button" ", and enter into the debug state.
In the working status, set to save the value of the debug state,
the Collimator must be sure to be connected first, or display error.
Then we can't see whether out film is accurate or not,
No display.
First, press the button " ", or the button " ", the digital tube display “1000A”, that’s
27
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
the Small Light bar debugging, the button " "and " " is for open bigger and close smaller.
Otherwise, need to change the two drive line of motor.
3) Adjustment for the parameter (AD)
When adjust the open small light bar, the potentiometer AD values increase. When close,
decrease. Then make the small light bar opening and closing positions to the middle position,
adjust the potentiometer value to 2.5 V, and Install the potentiometer well.
First, press the button " ", enter into the Small Light bar parameter debugging, now,
digital tube display "1001A", and then press the button " " or " " to select or change into
the needed Small Light bar parameter position;
Second, press the button " " to enter into the selection debugging. If the button
" " has not been pressed, in other words, when the button " " go out, the digital tube
display " the AD value of the Small Light bar position. Manually turn the Small Light bar to probably
needed position, press the button " ", buzzer rang, save the corresponding AD value;when
press the button " ", in other words, when the button " " bright, press the button
" ", Save the modified AD value. Now, press the button " "or the button " " bright.
Then, each time you press the button " ", the saved AD value increase "1" or reduce "1",
After modify the needed AD value,press the button " ". Now, when the button " " and
the button " " or the button " " all go out,Once again, press the button " ", mow,
the button " " bright, press the button " ", buzzer rang, Save the modified AD value.
Otherwise, don’t save the modified AD value, save the original AD value. In other words, No press
Finally, press the button " " exit the option debugging. Repeat the above operation,
debug another option.
Validation: enter the working state, verify that the debugging are accurate or not. If there is
deviation, perform corresponding fine-tuning or In combination with coarse and fine tuning. In
working status, view the corresponding film, the trip and openings of the small light bar are
accurate or not.
When the debugging is completed, fix motor tightly.
Each array of the small light bar opening size in the following table:
28
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
1001A
1004A
1005A
1006A
1007A
1008A
1009A
the LED of PCB go out, and the signal lamp bright when get out the shading film. Then we can be
sure that the test normal. Otherwise, need to check the corresponding line and optocoupler, until
the all is normal.
In the same way, to test close photoelectric coupler of the Small Light bar.
2) Adjustment for the Motor direction
First, press the button " ", or the button " ", the digital tube display “500A”, that’s
the Big Light bar debugging, the button " "and " " is for openings bigger and closing
smaller. Otherwise, need to change the two drive line of motor.
When the debugging is completed, fix motor tightly.
First, press the button " ", or the button " ", the digital tube display “3000A”, that’s
the car Y to debugging, the button " "and " " is for Y to up and down. Otherwise, need
to change the two drive line of motor.
3) Adjustment for the parameter (AD)
When adjust the Y to up, the potentiometer AD values increase. When down, decrease. Then
make the Y to positions to the middle position, adjust the potentiometer value to 2.5 V, and Install
the potentiometer.
First, press the button " ", enter into the Y to parameter debugging, now, digital tube
display "3001A", and then press the button " " or" " to select or change into the needed
the Y to parameter position;
Second, press the button " " to enter into the selection debugging. If the button
" " has not been pressed, in other words, when the button " " go out, the digital tube
display " the AD value of the Y to position. Manually turn the Y to position to probably needed
30
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
position, press the button " ", buzzer rang, save the corresponding AD value;when press the
button " ", in other words, when the button " " bright, press the button " ", Save
the modified AD value. Now, press the button " "or the button " " bright. Then, each
time you press the button " ", the saved AD value increase "1" or reduce "1", After modify the
needed AD value,press the button " ". Now, when the button " " and the button
" " or the button " " all go out,Once again, press the button " ", mow, the
button " " on light, press the button " ", buzzer rang, Save the modified AD value.
Otherwise, don’t save the modified AD value, save the original AD value. In other words, No press
Finally, press the button " " exit the option debugging. Repeat the above operation,
debug another option.
Validation: enter the working state, verify that the debugging are accurate or not. If there is
deviation, perform corresponding fine-tuning or in combination with coarse and fine tuning. In
working status, if you want to view the second film of the 1/2 Division is right or not, you must
exposure the first film. In the same way, if you want to view the fourth film of the 1/4 Division is right
or not, you must exposure the first, the second, and the third film.
Array instructions
3001A Y middle position
3002A Y up position
3003A Y down position
3004A Y to Motor speed setting(1~64)
3005A Y to position precision setting(0~15)
31
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
First, press the button " ", or the button " ", the digital tube display “2000A”, that’s
the car X to debugging, the button " "and " " is for X to IN and OUT. Otherwise, need to
change the two drive line of motor.
3) Adjustment for the parameter (Pulse value)
Adjust the car X to encoder pulse direction. Ensure that A connect in A of PCB, B connect in B
of PCB.
When adjust the X to OUT, the encoder values increase. When IN, decrease. Install the
encoder well.
First, press the button " ", enter into the X to motor debugging, now, digital tube display
"2001A", and then press the button " " or" " to select or change into the needed the X
to position;
Second, press the button " " to enter into the selection debugging. If the button
" " has not been pressed, in other words, when the button " " go out, the digital tube
display " the pulse value of the X to position . Press the button " " or " ", turn the X to
motor to the needed position, press the button " ", buzzer rang, save the corresponding
pulse value;when press the button " ", in other words, when the button " " bright,
press the button " ", Save the modified pulse value. Now, press the button " "or the
button " " on light. Then, each time you press the button " ", the saved pulse value
increase "1" or reduce "1", After modify the needed pulse value,press the button " ". Now,
32
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
when the button " " and the button " " or the button " " all go out,Once again,
press the button " ", now, the button " " bright, press the button " ", buzzer
rang, Save the modified pulse value. Otherwise, don’t save the modified pulse value, save the
original pulse value. In other words, No press the button" " no save.
Finally, press the button " " exit the option debugging. Repeat the above operation,
debug another option.
Special instructions: in the working state,
In the debugging, the 2018A、2019A and 2020A of X to adjustment, presses the button
" ", when the button " " is on light, and when the light is bright, and button " "
and " ", or button " " is bright , press button " " each time, , the value
Array instructions
2001A 8×10 full film(Vertical 1/2 Division)
2002A 12×10 The third film of 1/3 Division
2003A 12×10 full film
2004A 10×8、10×12 full film
2005A 10×12 The second film of 1/2 Division
2006A 10×12 The first film of 1/2 Division
2007A 12×10 The first film of 1/2 Division
The first and third film of 1/4 Division
2008A 12×10 The second film of 1/2 Division
The second and forth film of 1/4 Division
2009A 11×14 full film
2010A 11×14 The first film of 1/2 Division
2011A 11×14 The second film of 1/2 Division
2012A 14×11、14×14、14×17 full film
2013A 14×11、14×14、14×17 The second film of 1/2 Division
2014A 14×11、14×14、14×17 The first film of 1/2 Division
2015A 12 X 10 The first film of 1/3 Division
2016A 14×11、14×14、14×17 The first film of 1/3 Division
2017A 14×11、14×14、14×17 The third film of 1/3 Division
2018A X to motor fast speed setting (400 ~ 3000)
2019A X to motor Slow down speed setting (400 ~ 3000)
33
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
34
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
L1 L1(A) A
L2 L2(B) B
L3 L3(C) C
N N N
2, Diagnostic Table Power Cable:
Diagnostic table Diagnostic Table Power Cable Generator(Control Cabinet)
(PCB20.81) J4&F1 (XJZ20.01) AC Output XC1
F1-1 L1(031)A XC1—1
F1-2 L2(002)B XC1—2
F1-3 L3(003)C XC1—3
J4-1,2 N(102) XC1—4
3, Console Panel Power Cable:
Console Diagnostic table
Console Panel Power Cable (XJZ20.10)
X1 (PCB60.12) J6
X1—2203 L J6—2125
X1—2204 N J6—2124
XC10-1 MAIN1
XC10-2 SHIFT1
XC10-3 RCOM1
XC10-4 T1-1
XC10-5 TI-2
FG Protect Ground
35
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
1 517(+12V) J18-6
5 519(A) J18-5
2 825(GND) J18-4
6 521(B) J18-3
3 826(PGND) J18-2
4 520(+32V) J18-1
FG Protect Ground FG
36
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
Final Check: Radiography table positioning installation of the mast base, probe frame assembly fixed
without looseness, detector installed correctly, confirms the accuracy of equipment cable connection.
37
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
450
Effective occupation
600
2110
Effective
600
occupation
800
38
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
39
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
Tilt up key:
Press this key, the Table body is tilt up to the 90° direction
(vertical).
Tilt down key:
Press this key, the Table body is tilt down to the horizontal
position or the negative angle direction
Longitudinal upward movement key:
Press this key, the Tube stand column longitudinal
move up to the head end
Longitudinal downward movement key:
Press this key, the Tube stand column longitudinal
move down to the foot end
Tabletop transversal right movement key:
Press this key, the tabletop move to inside direction
41
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
1) Power on: when the “Power on” button of generator on the remote console is pressed,
And wait for a moment, the diagnostic table power on too and Enters to the work status.
2) Power off: While the “Power off” button of generator on the remote console is pressed, the
diagnostic table is power off.
1) When the Cartridges is at loading position, put the Film cassettes into holder, press the
movement button, Cartridges move back to the parking position and ready for exposure. The
system will automatically identify and display the cassettes size/ allowed subdivision format in this
process.
2) When pressing the needed subdivision format selection key, the last of the digital LED will
display the exposure count number.
3) When pressing the exposure switch in handle for the film exposure, the exposure count
number of digital LED will be reduced by one. Continue to press the exposure switch for the rest
of the film exposure till the number of digital LED displays to zero.
4) Press the movement button Cartridges move to loading position, take out the exposured
film cassettes.
Serial radiography: press the speed button, the cassette leaving from the PARKING position
and trace to next exposure position, Collimator can be tracked in place automatically. At any time,
when press the exposure switch in handle for rapid film exposure, but fluoroscopy is not allow.
Press the speed perturbation button again, serial radiography mode is cancelled and cassette
return back to parking position.
1). The table body rotation: press rotation button of the remote control panel or control panel,
it will tilt, release the button to stop, positive direction up to +90 °, when rotating to limited
location, the table body will stop automatically : the opposite direction up to -25 °: if the opposite
direction rotation starting from above the horizontal level, when rotating to the horizontal level it will
stop, if continue to tilt to the opposite direction, must release the button first, then go on rotating to
negative direction, if the tube rack (tube column) moves up to protective limited location of the
sliding camera system, it will stop when the counter-rotation moves to protective limited location of
the rotating camera system, it can’t reach -25 ° to protect the security of camera system, if need to
reach -25 °, return the tube rack from the protective limited location of camera system.
42
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
2). The sliding of tube stand: operate handle or press the sliding button on control panel for
tube stand sliding, release the handle or the button for immediately stop, sliding up distance is
720mm, sliding down is 0mm, the total sliding distance is 720mm, when the table body is at the
negative protection angle or below, the sliding up distance of tube stand can’t reach 720mm, it can
only reach up to protective upwards limited location so as to protect the camera system, to reach
720mm, make the table body tilt to the positive direction until it leaves protective limited location of
negative angle from camera system.
3). Up and down for pressured cup: Press "pressured cup down" button of the remote control
console to control the downward of pressured Cup, Press "pressured cup up" button of the
remote control console to control the upward of pressured Cup.
Devices encountered in the use of sudden interruption of electricity network , When the
compressor is in the oppression of press patients , in order not to harm the patients by the long
duration of pressure ,please press button as shown in pressured cup labeled "PUSH" position to
rises upward pressure cup ,so that can relieve the pressure force to patients. Before pressing the
button, the direction of the pressure cup pressure to the patient, and then press and hold down the
button, remove pressure to the patient from the pressure cup, and then press cup automatically
pops up.
4). Table movement: control handle or press button of the control panel for controlling the left
and right movement of table, release handle or the button to stop moving, the total moving
distance is 220mm.
The console structure includes the main console, high-voltage generator display board and
some other parts.
The console panel as shown in Appendix B, the key function are as follows:
43
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
Figure 7-5
B Switch On M Pg Up/Dn
C Save N Return
clearing setting
selection setting
1, Radiography Indicator
Ready for exposure lamp: the lamp lights up when the generator is ready for exposure,
indicating that the technique selected is properly set, there are no interlock failures or system
faults, the anode is rotating and the X-ray tube is ready for exposure.
Radiography indicator: the indicator lights up when the X-ray exposure is in progress. At the
44
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
same time that radiographic exposures are being made, an audible signal sounds.
2, Switching the Generator On/Off
Pressing and holding Switch On button for one second will light up the lamp on the upper left
of the button, indicating the generator is switched on to working mode.
Pressing and holding both Switch On and Reset button at the same time for two seconds
starts up the generator to set-up mode. To enter to working mode, switch off the generator and
restart.
Press and hold Switch Off button for one second to switch off the generator.
When the generator is switched on, it performs a self-test:
· All the lamps and display segments light up briefly.
· There is an audible signal.
During this test you should check the following:
· Do all the lamps and display segments light up?
· Is there an audible signal?
If an error message appears, you will find help in the Error Message List. If you need more
help, please contact PNMS.
3, Radiographic Parameters
1) Tech select: these three lamps indicate the 3 radiography modes AEC, mAs and mA/ms
respectively.
Press Tech Select button to select a radiography mode. The lamp indicating the selected
mode will light up.
· AEC Mode: Only kV, mA can be adjusted; mAs Mode: Only kV, mAs can be
adjusted;
· mAs/ms Mode: kV, mAs, ms all can be adjusted.
2) Fields Select (only for AEC tech mode): three lamps indicate the left, middle and right field
respectively.
Press Fields Select button to select a field. The lamp indicating the selected field will light up.
3) Film Speed Select (only for AEC tech mode): three lamps indicate the low, medium and
high film speed respectively.
Press Film Select button to select a film speed. The lamp indicating the selected speed will
light up.
4) Focal Spot Select: there are two buttons for selecting the focal spot, Small Focal Spot and
Large Focal Spot.
Press either of the buttons to select the focal spot. The lamp indicating the selected focal spot
will light up.
5) Exposure Mode Select: there are three buttons to select an expo- sure mode, Bucky Table,
Wall-Stand and Free Cassette.
Press any of the buttons to select an exposure mode. The lamp indicating the selected
exposure mode will light up.
4, APR Programs Display
APR programs display segment includes one display screen, three APR Select buttons, two Page Up/Dn
buttons, one Return button, three Body Size select button sand one Save button.
45
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
· Display Screen: Shows the different body regions, the generator status and the current time.
· Body Size select: These push-buttons are used to adapt the APR technique chosen according to patient size.
Three patient sizes are available, Large, Medium and Small. Default patient size for APR is Medium. When
APR is activated, the lamp indicating medium patient size lights up and relevant radiographic parameters
are shown in the Exposure Data Display segment. When you adjust the patient size selection, the lamp
indicating the selected patient sizelightsupandtheradiographicparameterswillchangeaccordingly.
Note: You can only save the kV value if you want to modify and
save the radiographic parameters when Large or Small patient size
is selected.
· Page Up/Dn buttons: The screen displays three body regions or APR programs at one time only. Use
these buttons to scroll through the display pages.
· APR Select: Press the corresponding button to select a body region or APR program shown on the
display. The selected body region or program will be highlighted on the display. Meanwhile, all the
radiographic parameters of the selected APR program will be shown in the Exposure Data Display
segment. Use these parameters directly or make any adjustment as you require.
· Save: Press and hold the Save button for two seconds to save adjusted radiographic parameters of any
APR program.
· kV setting: Press ―+‖ or ―-‖ button to change the kV value in step of 1 kV. The current kV value is
displayed in the LED.
· mA/mAs setting: Press ―+‖ or ―-‖ button to change the mA/mAs value in step of R'10 mA/mAs (mA for
AEC and mA/ms mode, mAs for mAs mode). The current mA/mAs value is displayed in the LED.
· ms/s setting: Press ―+‖ or ―-‖ button to change the ms/s value in step of R'10 ms/s (When the time
parameter is less than 10 seconds, the LED display the corresponding "ms"; when the time parameter is
greater than the at 10 seconds, the LED displays the corresponding "s" values. The current ms/s value is
displayed in the LED.
· DEN setting: Press ―+‖ or ―-‖ button to change the DEN value in step of 1 DEN ( DEN for AEC mode).
The current DEN value is displayed in the LED.
Note: If you do not select any APR program after selecting a new
body region, the data last shown in the radiography data display
segment will be used for the new exposure.
6, Fluoroscopy Indicator
46
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
Ready for fluoroscopy indicator: the lamp lights up when the generator is ready for fluoroscopy, indicating
that the technique selected is properly set, there are no interlock failures or system faults, the anode is rotating
and the X-ray tube is ready for fluoroscopy.
fluoroscopy indicator: the indicator lights up when the X-ray fluoroscopy is in progress. At the same time that
fluoroscopy exposures are being made, an audible signal sounds.
7, Fluoroscopy Parameters
PF fluoroscopy setting: the button and related indicator is for setting continuous fluoroscopy
or pulsed fluoroscopy.
Press PF Select button to select a continuous fluoroscopy or pulsed fluoroscopy mode. The
indicator indicating the selected mode will light up. When the indicator is lighted, it indicates that
the pulsed fluoroscopy is selected. When the indicator is not lighted, it indicates that a continuous
fluoroscopy is selected.PPS parameters can be adjusted when the pulsed fluoroscopy is selected.
The PPS parameter is not adjustable when the continuous fluoroscopy is selected.
ABS fluoroscopy setting: the button and related indicator is for setting ABS or non ABS.
Press ABS Select button to select a ABS or non ABS mode. The indicator indicating the
selected mode will light up. The indicator is lighted, indicates that the ABS is selected. The
indicator is not lighted, ndicates that the ABS is not selected. Fluoroscopy kV and fluoroscopy mA
parameters can be adjusted when the non ABS is selected.
The Fluoroscopy kV and fluoroscopy mA parameter is not adjustable when the ABS is
selected.
Dose parameter setting: Three indicators indicate the selection of ABS dose of Normal Dose,
Low Dose and High Dose. Only when ABS and continuous fluoroscopy are selected, the dose
parameter is adjustable.
Press "DOSE" key to switch and select the corresponding continuous fluoroscopy ABS dose.
Indicator light, Indicates the corresponding continuous fluoroscopy ABS dose was selected.
8, Fluoroscopy Data Display area
Fluoroscopy data displaying segment supports the setting and display of foure parameters,
Fluoroscopy kV, Fluoroscopy mA, Fluoroscopy time and PPS.
· Fluoroscopy kV setting: Press ―+‖ or ―-‖ button to change the kV value in steps of 1 kV. The current kV
value is displayed in the LE D. Non ABS Fluoroscopy is selected when the kV can be adjusted.
· Fluoroscopy mA setting: Press ―+‖ or ―-‖ button to change the mA value in steps of 0.1 mA. The current
mA value is displayed in the LED.
· PPS setting: Press ―+‖ or ―-‖ button to change the PPS value in steps of 3, 7.5, 15, 30 PPS. The current
mA value is displayed in the LED. PPS is only adjustable when PF is selected. The PPS parameter is not
adjustable when the continuous Fluoroscopy is selected.
47
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
48
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
8.4.1 Fuse
1 .Fuse location
The diagnostic table the fuse of the equipment: FU1, FU2, FU3 are in control cabinet.
The equipment's generator fuses: FU1, FU2, FU3, FU4, FU5 are in HV generator control
cabinet, they can be found when the lateral plate of cabinet is opened.
2. Fuse Replacement
Dangerous!
Cut off power before replace fuses!
If high voltage generator’s fuses need to be replaced, remove cover of HV generator with
screw knife to replace the fuse, after exchange, recover generator.
If diagnosed table fuses need to be replaced, open control cabinet of diagnostic table, at
side of the control cabinet and the electrical installation board have FU1, FU2, FU3, open the lid of
fuses need to be replaced to replace them, close the lid of fuse and control cabinet door.
3. Fuse parameters
Dismantle Collimator shell, dismantle the collimator lead sheet at collimator lamp side,
dismantle collimator lampshade, and take lamp off from lamp base by screwdriver. After
replacement, to test whether the lamp screw is collide with the shell in multimeter. Install collimator
3. Collimator lamp parameter:
Halogen lamp, AC 24V, 100W
8.5 X-ray protection safety.
Before each exposure, please confirm necessary precautions have been made. Please follow
below steps.
⑴ Put on protective clothing and wear protective glasses.
⑵ Stay far from X-ray source with hand-switch.
⑶ Adopts low-dose exposure as lower as possible, in the permission of projection,.
⑷ Increase SID distance properly in order to improve image resolution, also patients and
operator can reduce the effect from X-ray.
50
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
9.1. Introduction
51
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
1. Please assure that the power line is connected with image processor and earthed reliably
before connection of all signal cables!
2. Connect each cable to enable it to insert naturally and correctly and switch on the other
powers of host finally.
3. Power on.
1. Please set up the LCD according to the following methods if an independent LCD is used:
1.1. Enter “Patient Files Management” interface, then click the “Add” button to enter the main
interface;
1.2. Press “Ctrl”, “Shift” and “F10” on the keyboard simultaneously and a correction dialogue
window will pop up as shown in the following figure 9-1.
52
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
Figure 9-1
1.3. Click the “Calibration” button. Then, press the automatic adjustment button on the LCD
and wait for the automatic adjustment of LCD;
1.4. The left margin of whole picture will be parallel to that of LCD via adjusting the
“Horizontal Position” option on the LCD if the picture does not cover the whole screen and then
enable the picture to cover the whole screen via adjusting the “Horizontal Width” or “Clock” option
on the LCD;
1.5. For ordinary colorful screen, it is suggested to adjust the brightness and contrast of
screen to 70% and 80%~90% respectively via “Contrast” and “Brightness” option.
2. This product supports the remote upgrading (release the updating once every year). The
corresponding upgraded package can be obtained via contacting manufacturer of the complete
machine or dealer to get “Service No.” in due course.
3. Network Setting (if necessary)
3.1. Click the “Network Setting” button and the following dialog will pop up:
Figure 9-2
3.2. Please let the website personnel set up such relevant information as IP site, gateway,
and subnet mask.
4. Press F11 on the keyboard to enter into the setting mode (the electric secret key from
manufacture is required to be inserted):
Figure 9-3
4.1. Set such information of user as company name and department name;
53
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
*Note: the information of user after setting will be displayed on the main interface and can be
overlaid into image report automatically.
4.2. Set current time;
4.3. After selecting the default mirror state of image (mirror in X direction, Y direction and
both X and Y direction and standard), please press the “SET” button after selecting;
4.4. Set the initial value of gage rod:
1) Press OK after the value of X ruler and Y ruler is set as 1000 uniformly (default value);
2) Take the photo of gage rod under the dynamic condition;
3) Select the 100mm * 100mm area by pulling the enhanced box and the length unit of x
and y coordinates mm will be displayed in the right corner.
4) Record the length of X and Y coordinates and let it multiply 10, and then remove the
decimal (it is supposed to be free of decimal). For example, if x: 56.70 mm is on the right corner,
record the 567;
5) Open F11 window to write the recorded value of x and y presented on the X ruler and Y
ruler respectively. Then, press OK.
4.5. Restart the image processor in about 5 seconds
4.6. The above-mentioned can also be adjusted on the full screen image via “Calibration”
button now.
Figure 9-4
1. The retrieval elements include “ID No. of Patient”, “Patient Name” and “Date Interval”,
where the data means when the image report is established.
2. Function keys
2.1. “Search” ——the search can be performed by filling one or more options in the search
elements and the qualified file of patient will be displayed in the primary table. Under the default
condition, the search result is the information about files of patients within recent one month.
2.2. “Add” ——it is used to add the file of patient. Please establish a new file for the file of
patient for initial use of this system or no selection for any file of patient.
54
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
Figure 9-5
2.3. “Modification” ——the image data can be read on the main interface via clicking the
“modification” button or clicking the selected file of patient twice and the image report can also be
modified.
2.4. “Export” ——all relevant information and data concerned to selected file of patient can
be exported to such storage medium as USB flash drive, mobile hard drive via USB 2.0 port after
clicking the “export” button. The relevant data can also be exported to the designated server or
terminal equipment if the internet port is set up or connected to the network.
2.5. “Patient in Emergency” ——the procedure for establishing of a new file of patient can be
skipped to enter the main interface and conduct the fluoroscopy directly via clicking the
“Emergency Patients”.
3. The file list used for displaying qualified retrieved elements in the file preview area includes
the patient ID, name, gender, age (the full age can be calculated automatically in the system
according to the birth date) and date when the report is established.
1. Interface layout:
Figure 9-6
55
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
1.1. There are eight storage places for images in each file of patient with No. of each storage
place displayed on the upper right in the preview area of saved picture. When the storage place is
selected, the green part will be on, or it will be off. It is in No. 1 storage place under the default
condition.
1.2. The master picture is on the left of interface
1) The patient ID, birth date, name and sex are displayed on the upper left of master picture
from top to bottom.
2) The organization description and department description of user are displayed on the
upper right of master picture.
3) The width and value of current window are displayed on the bottom right of master
picture.
1.3. The toolbar is in the center of interface.
2. Toolbar
2.1. “Info Manage (F1)” ——obtain the interface of “File Management for Patient” via clicking
this button or “F1” on the keyboard.
2.2. “Report (F2)” ——obtain the interface of “Radiologic Image Report List” via clicking this
button or “F2” on the keyboard.
2.3. “Denoise (F3)” ——the noise level can be reduced via clicking this button or “F3” on the
keyboard. The five levels of 0, 1, 2, 3, 4 can be selected.
1) The button in the center can become the green circular button after steeping on the foot
brake. Click it and enter into the continuous shooting condition for 4s; alternatively, it is suggested
to press “F4” on the keyboard.
2) The images group by continuous shooting is also displayed in the preview area as shown
in the following figure;
3) The play button for continuous shot images group is on after selecting the images group
by continuous shooting. For the play button in the center, the left one is the last frame and the right
one is the next frame;
4) The play button will be used as the pause button during playing;
Note: This function is mainly set for barium check of stomach and intestine and esophagus.
2.5. “Horizontal Mirror” ——the horizontal mirror of image can be performed via clicking this
button (turn over from left to right and vice versa).
2.6. “Vertical Mirror” ——the vertical mirror of image can be performed via clicking this button
(turn over from top to bottom and vice versa).
2.7. The “Left” and “Right” is used for functions selection of “Reverse (F5)”, “Clockwise”,
“Anticlockwise” and “Gamma”.
2.8. “Reverse (F5)” ——the conversion between positive and negative can be fulfilled via
56
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
Figure 9-7
1. Adding reference image: the required image can be added and the display mode of image
(positive or negative) can be selected via clicking the right button of mouse on the upper part of
this area. Meanwhile, the window width and window place for the image in the report can be
adjusted via pressing the left button of mouse to move from left to right and from top to bottom and
vice versa.
2. Template model: the preset medical template model is displayed with method of drop-down
57
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
folder on the bottom right of this area. Its content is shown on the boxes of “image” and “report” on
the bottom left by double-clicking the left button of mouse.
3. The text of report can be modified in the image performance bar and impression bar.
4. The default folders can be modified as needed: select the folder and click the right button of
mouse to obtain the drop-down menu, which includes renaming of the category (name of folder),
and renaming, adding and deletion of the item name, as well as saving the modification made on
the left side boxes.
Note (A): the content and format of report cannot be modified on the preview interface.
Note (B): the system support the use of HP Laser Jet P1005/1006/1007/1008 now, through
the USB 2.0 connection way.
1. Storage of fluoroscopy and end frame: enter the fluoroscopy condition after stepping on the
foot brake for fluoroscopy and the end frame image will be displayed after releasing the foot brake.
Meanwhile, the end frame image will be stored in the current storage place with No. 1default
storage place;
2. Image storage: please click the next storage place via mouse and conduct the fluoroscopy
if the current end frame image is thought to be stored;
3. Intelligent gray-scale adjustment of image
3.1. Under fluoroscopy condition: the gray-scale optimization can be conducted
automatically in the center of master picture area via clicking the right button of mouse twice and
the optimization can be canceled via further clicking of right button;
3.2. Storage and end frame image:
1) The interested part can be selected by right-clicking and dragging the mouse in the
master picture and the gray-scale optimization will be conducted automatically for the selected
area in the system;
2) If you are not satisfied with the optimization in the system, press the left button of mouse
to drag the mouse from top to bottom and modify the image gray-scale for the selected area;
3) Click the right button of mouse again in the master picture and recover the image
gray-scale.
4. Use of ruler
While clicking the right button of mouse to select the interested part, the horizontal and
longitudinal dimension of it will be displayed in the right corner of master picture.
5. Power off: Press the button of power without exiting the system after use.
58
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
CHAPTER 10 Maintenance
transformer with few numbers of turns, turn to turn has big current, check regularly if any damages,
debris, or dust accumulation and remove in order to avoid sparking caused by failure of the
expansion.
7. Check frequently protective earth terminal for each components is in good condition
(grounding resistance is very small).
8. Check power supply on a regular basis, check power supply resistance value (or the power
supply voltage value is changed), make sure it meet equipment’s requirement.
9. When don’t use controller, cut off the power.
10. Avoid striking of x-ray tube components during operation.
11.When X-ray tube assembly are unused for long time or when new x ray tube components
are used, load X-ray tube gradually to avoid damage for instantaneous loading.
12. When equipment works continuously, check allowed heat capacity of x-ray tube
components, or check surface temperature of X-ray tube assembly, it can’t exceed 70 ℃.
13. Clean equipment with a soft cloth with a little anhydrous alcohol to scrub; organic solvents
are not allowed.
14. Anti-bacterial requested equipment surface touched by patients, such as sterilization of
table panels, use ethanol for disinfection, mercurochrome, iodine are not allowed. Ultraviolet
radiation etc. is recommended on a regular basis, regular disinfection.
4. Considerations and measures, when appear the oil spill on the top of the high pressure
tank.
Check for add oil filled/ oil plane (HT tanks):
Under normal circumstances, does not require regular inspection of high pressure tanks in the
horizontal position of the insulating oil. However, if you have signs of oil spill, should check the oil surface,
refer to follow: high voltage generator oil filled /oil plane check
(1) Loosen the screw of the oil filled mouth, which is on the high pressure tank cover.
(2)After loosen the screws, remove the rubber stopper.
(3) With a clean ruler, cardboard, or other similar instrument measuring oil plane - began to measure
from the top of the tank lid.
The normal range is 1.0 - 1.25 inches (25 - 32mm)。
If the range is 1.25 - 1.6 inches (32 - 41mm), at this time, should add some clean oil.
If the range is more than 1.6 inches (41 mm),please contact the manufacturer.
61
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
device inside the tank. Then reinstall filling oil plug well, install plug
well, and tight the four screws after fix screw well. That’s be to
protect the oil charge plug. Processing is the object of oil pollution,
must be in According to the requirements of the government, and
conform to local laws and regulations.
Check if the signal, display and indicator are normal. Check if labels for
Daily warming, danger are good. Check if X-ray generator components has abnormal
Inspection noise during exposure. Check if all grounding are good. Check if there are any
loose, sticking buttons on console.
Check if any oil leakage of the X-ray generator component. Check flexibility of
table and column bearing, if any big friction, coat bearing orbit with lubricating
Weekly
oil. Check if any loose and shedding of fastener parts for all components, fix
Inspection
them if any. Check if any broken steel wire on column due to lots of wear and
tear., exchange them if any.
Check equipment grounding resistance (less than 0.2Ω). Check if any filler
coming out of head of HV cable due to heat. Clear silicone oil on high-voltage
outlet, rub plugs and sockets with ether or carbon tetrachloride, and then use
Check every
some new silicone grease, tighten the fixing ring of plug. Check if circuit boards
six month
are fixed, if any corrosion, use vacuum to remove dust, for powdered carbon on
carbon toner movement orbit of autotransformer, and clean it with eraser.
Check center of X-ray generator components and collimator.
62
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
E 001 Film Box Communication Error, and check 485 chip and communication lines.
E 010 Collimator Communication Error, and check 485 chip and communication lines.
E 100 Diagnostic table Communication Error, and check 485 chip and communication lines.
Film Box dolly X to Error, check whether something gets stuck, check the motor control circuit,
E 002
and check the rotary encoder circuit.
Film Box small light bar Error, check the motor control circuit, and check the angular
E 003
displacement sensor circuit.
E 004 Film Box big light bar Error, check the motor control circuit, check the optocoupler.
Film Box dolly Y to Error, check whether something gets stuck, check the motor control circuit,
E 005
and check the angular displacement sensor circuit.
11.2 HV Generator
63
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
64
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
65
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
66
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
67
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
68
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
69
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
70
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
71
PLD5600 USER’S MANUAL JY-5600-23 REV2.1E
1 .Manufacture Duty
We are responsible for the products’ security, reliability and performance under following
situations:
a. all the assembling, adjusting, modification and maintenance are made by our technical stuff
or our authorized people
a. all the electrical parts equipped on the equipment are qualified to the manual
b. the operators should be fully trained, and operate according to manual
2 .After-sales Service
1) 12 months started from the date of your purchasing. If any malfunction is caused by quality
related problems within warranty, the product can be repaired for free.
2) After warranty, we charge for spare part at cost.
72