Professional Documents
Culture Documents
Lecture Notes
6. Experimental
Techniques
Sampling Data! 3
Resolution! 5
Windowing! 9
Aliasing! 13
Derivation of Equations! 16
Calibration! 18
Pistonphone! 21
Static Calibration! 22
Design Issues! 23
Microphone spacing! 25
Measurement of waves! 26
Summary! 30
This form for acoustic signals is more useful than expressions in terms
of time, because so much of the generation and response depends
upon the frequency of the wave. These include our ears (see figure
6.1), transmission across walls, sound-absorbing materials,
resonators, etc.
-1
-2
-3
0
-50
-40
-30
-20
-10
10
20
30
40
50
3 3
2 2
1 1
0 0
-1 -1
-2 -2
-3 -3
=
0
-50
-40
-30
-20
-10
10
20
30
40
50
0
-50
-40
-30
-20
-10
10
20
30
40
50
-1
-2
-3
0
-50
-40
-30
-20
-10
10
20
30
40
50
That just leaves the time step dt. Since this is just a constant it is
normally ignored (just as the total time is ignored in going from a
Fourier series to a Fourier transform). Thus our discrete Fourier
Transform is defined as
N
�
DFT(k) = fn e−ikω0 nT
0
6.1.3. Resolution
� �
e−iω 2 � iω T � e−iω T2
T
T ωT
= e 2 −e−iω 2 = 2 sin
iω ω 2
� � ωT � �
T sin 2
= T e−iω 2 (6.2)
ωT /2
Equation (6.2) has introduced the factor T on both the numerator and
the denominator so that the term in parentheses has the same
argument for the sine function as on the denominator. The term in
parentheses is a spherical Bessel function usually referred to as a sinc
function in the context of spectrum analysis. It is plotted in figure 6.6.
3
Secondary
2 peak spread
-1
-2
-3
0
-50
-40
-30
-20
-10
10
20
30
40
50
f (t) = sinω0 t
�� � � ��T
ei(ω0 −ω)t e−i(ω0 +ω)t
= −
2(ω − ω0 ) 2(ω + ω0 ) 0
� � � �
(ω−ω0 )T (ω+ω0 )T
T
sin 2 T
sin 2
= −iei(ω0 −ω) 2 + ie−i(ω0 +ω) 2 (6.3)
ω − ω0 ω + ω0
spread
Secondary
peak
6.1.3.1. Windowing
0.75
0.5
0.25
0
3
0.5
1.5
2.5
�T � �� �
eiω0 t − e−iω0 t eiπt/T − e−iπt/T
e−iωt dt
2i 2i
0
�T � �
1
=− ei(ω0 +π/T −ω)t
−e i(ω0 −π/T −ω)t −i(ω0 −π/T +ω)t
−e +e −i(ω0 +π/T +ω)t
dt
4
0
� �T
ei(ω0 +π/T −ω)t ei(ω0 −π/T −ω)t
= −
4i(ω − ω0 − π/T ) 4i(ω − ω0 + π/T ) 0
� �T
e−i(ω0 −π/T +ω)t e−i(ω0 +π/T +ω)t
+ − +
4i(ω + ω0 − π/T ) 4i(ω + ω0 + π/T ) 0
� �
iT i(ω0 −ω)T /2 cos((ω − ω)T /2) cos((ω − ω)T /2)
= e −
4 T /2(ω − ω0 − π/T ) T /2(ω − ω0 + π/T )
� �
iT i(ω0 −ω)T /2 cos((ω + ω)T /2) cos((ω + ω)T /2)
+ e −
4 T /2(ω + ω0 + π/T ) T /2(ω + ω0 − π/T )
(6.4)
Rectangular
window
Reduced
peak Half-sine
window
Increased
spreading
Reduced
secondary
peak
Hamming window
0.53836-0.46164cos(2πt/T)
zero
gradient
Hamming window
0.53836-0.46164cos(2πt/T)
RORRIM MIRROR
ω0 - ωs ωs/2 ω0 ω1
ωs - ω1
Figure 6.14 Aliasing
Data logger
Charge
amplifier
sample
sample holder
Data logger
Charge
amplifier
sample
sample holder
p1 = p̂1 eiωt
and
p2 = p̂2 eiωt
A = Â1 ei(ωt−kx)
B = B̂1 ei(ωt+kx)
x$ p$ x% p%
Aei!
" t"kx#
Bei!
" t!k x#
Equations (6.9) and (6.10) are extremely useful for determining the
acoustic response of materials and acoustic devices to acoustic
excitation. However, since it depends upon the simultaneous
measurement of two different signals by two microphones, it is
important that the microphones are calibrated both for their individual
responses and their relative responses.
6.2.1.2. Calibration
Two types of calibration are essential for the two microphone
technique: relative calibration and absolute calibration. Of these
relative calibration is the most important to ensure that the
measurements by any of the microphones is the same as it would
have been if measured using the other microphone. We consider this
type of calibration first. There are two techniques commonly used for
this: the swapped microphone technique and the calibration box
technique.
where p̂1,ac and p̂2,ac are the calibrated readings (relative to the
reference microphone), and rerr is the relative error between the
second microphone and the reference microphone. If we then swap
and
p̂2
p̂2,s = p2,ac =
rerr
We want to find the equivalent of equations (6.9) and (6.10) for the
calibrated signals. We can use the swapped measurements to find the
calibrated ratio between the signals at positions 1 and 2. Multiplying
the ratios from each of the two measurements we have
� �� � � �2 � � � �2
p̂1 p̂1,s p̂1,ac rerr p̂1,ac
= = = T12
2
(6.11)
p̂2 p̂2,s p̂2,ac rerr p̂2,ac
Data
logger
microphones
Calibration
unit
Signal
generator
200
6.2.1.2.3. Pistonphone
Adaptor
Pistonphone
7
microphone
Data logger
1.5.3
Figure 6.21 Using a pistonphone
Absolute amplitude calibration
Both the previous two techniques calibrate the microphones relative to each other. In this
The
sectionstandard
we consider thefortwo calibrating the acoustic
techniques for calibrating the amplitude. amplitude measure by a
The standard is rather
microphone
poor and is rather poor because it relates to only oneas frequency.
relates to only one frequency. The official way to calibrate is using a pistonphone
illustrated in figure 5. This device has a small diaphragm which acts as a loudspeaker which
The official
oscillates atway
a knownto calibrate
frequency (120Hz) andis a using a pistonphone
known amplitude (around 115dB, butas illustrated in
specified
for each unit). Each microphone fits snuggly in the device (using an adaptor for different
figuresize6.21.
microphones).This device
The output has
is logged a small
and compared to thediaphragm which
reference. The accuracy of the acts as a
loudspeaker
amplitude which oscillates at a known frequency (120Hz) and a
generated by the pistonphone depends upon the atmospheric pressure
be measured. A calibration chart is also included with a pistonphone. A photograph of a
which should
Pistonphone
microphone Pressure
adaptor gauge
Figure 6: Photograph of a pistonphone.
duster microphones
data logger
Vacuum
cleaner
spacing) and the assumption that the acoustic waves in the pipe are
planar. This latter condition affects 10 the frequency range which can be
We can drive the pipe with axisymmetric waves and so the condition
reduces to:
3.83c̄
r0 < (6.14)
2πfmax
If we chose exactly this radius, all the high-order modes would, indeed,
be cut-off. Unfortunately, however, at frequencies near 1kHz the high-
order modes would only just be cut-off. As a result they would decay,
but the decay would be slow. This would require us to measure waves
some distance from the end to ensure that all these modes had
decayed sufficiently. As a result a safety margin would normally be
taken to ensure rapid decay. Thus a pipe radius of 50mm would be a
good compromise.
I.e. L = 31mm
In other words the matrix becomes singular, and the two microphone
technique fails to be useful, when the distance between the two
microphones is an integer multiple of a half wavelength. In practice, a
working range of between 5 and 95\% of a half wavelength is usually
specified. However, this is not strictly correct since it could be greater
than a half wavelength so long as a half wavelength is itself ignored. If
the higher range is taken, then this leaves gaps in the solutions unless
multiple pairs of microphones are used to provide redundancy.
microphone sample
loudspeaker
xmax xmin
Figure 6.25 Using the moving microphone technique
A = Âeiω(t−x/c̄)
B = B̂eiω(t+x/c̄)
The signals at the two positions will be the local superposition of these
two waves. Hence, writing the measured signals in terms of the two
waves yields:
Equations (6.17) and (6.18) are simultaneous equations for the two
unknowns. Eliminating B gives:
� �
Âeiωt
e−2iωxmax /c̄
−e−2iωxmin /c̄
= pmax e−iωxmax /c̄ − pmin e−iωxmin /c̄
(6.19)
and
or
|pmin | = |B| −| A| (6.25)
If |B| > |A| (for an absorbing device, so we solve equations (6.23) and
(6.25) rather than (6.23) and (6.24)) we have:
Fourier Transform
�∞
F (w) = f (t)e−iωt dt
−∞
Window functions
# Half-sine:# # # # sin(πt/NT)
Aliasing:# More than one signal can pass through measured points
|pmax | + |pmin |
|B| =
2
A = |A|eiω(t+xmax /c̄)
B = |B|eiω(t−xmax /c̄)