Professional Documents
Culture Documents
PQ Protocol - Final
PQ Protocol - Final
0/A-06
Revision No : 0.0
PROCESS VALIDATION - ETO STERILIZER
Effective Date : 01-06-2018
F88 PERFORMANCE QUALIFICATION OF STERILIZER
EO STERILIZER- MSPL/ETO/01
<<<X<
The protocol has been prepared, checked, and approved by the concerned personnel for
implementation by the undersigned:
CHECKED
BY Utility Jayesh Pandya In-Charge
<
2. OBJECTIVE:
The objective of performance qualification of EO sterilizer is as under:
The objective of validation is to documents the evidence required to provide a high degree of assurance
that a specific process will consistently produce product meeting the required sterility assurance level (SAL).
Product sterilized in the validated process should be shown to meet predetermined specification and quality
characteristics related to product functionality and safety.
To determine that Ethylene oxide sterilization process consistently performs as intended by running the
system and recording all relevant information. The Data and Test results must demonstrate that the
process meets pre-determined specifications under normal conditions as well as worst case conditions.
Performance qualification of rigorous microbiological and physical testing, beyond routine monitoring, to
demonstrate the efficacy and reproducibility of the sterilization process.
The results of the equipment validation shall meet pre-determined acceptance criteria(s) under normal
conditions of operations, and where appropriate for worst-case situations. If any variation is observed
between the set acceptance criteria and the actual data obtained, then same should be explained
reasonably.
On the basis of validation data, the Work Instructions (WI) for routine monitoring of sterilization shall be
evaluated for its suitability.
In case any deviation is observed during validation, it shall be recorded in Deviation Section.
The site Quality Assurance In-charge shall decide whether deviation is acceptable or not. If deviation is
acceptable and it does not impact the quality and Performance of sterilizer, continue the validation. If
deviation is not acceptable and it has impact on the quality and Operational performance of sterilizer,
necessary corrective action shall be taken before qualification.
The equipment conforms to the complete qualification requirements and shall permit a formal “RELEASE” of
the equipment for the routine use in sterilization process.
<<<<X<
3. SCOPE:
3 Way stop cock with/without extension tube Infusion (IV) Set / Transfusion (BT) set
I.V. Flow regulator Measured Volume Infusion Set
I.V. Cannula Extension Line/ Pressure Monitoring Line
Hypodermic Syringes Hypodermic Needle
Other similar packaged products
4. RESPONSIBILITY
<X<
S. Name of Material
No.
Three way Stop Cock
1. Packed in blister (unit container) made of medical grade paper and PP+PE film. Further
unit pack is packed in duplex of 50 Blister Packed three-way stop cock and finally these
duplex boxes are packed in Corrugated box.
2. Calibrated Temperature & RH sensors
11. EO Sterilizer
ISO 11135:2014/ EN ISO 11135-1: 2015, Sterilization of health care products- Ethylene oxide- - Requirements
for development, validation and routine control of a sterilization process for medical devices
ISO 10993-7:2008, Biological evaluation of medical devices- Part 7: Ethylene oxide sterilization residuals
BS EN ISO 11737-2: 2009, Sterilization of medical devices - Microbiological methods - Part 2: Tests of
sterility performed in the validation of a sterilization process
ISO 14161 : 2009, Sterilization of health care products —Biological indicators — Guidance for the
selection, use and interpretation of results
<<<<X<
The Ethylene oxide sterilization process uses ethylene oxide as a sterilant to destroy viable microorganisms.
The ethylene oxide sterilization is suitable for the products which are heat and moisture sensitive.
Sterilizer Chamber: The sterilization chamber having total internal volume is 20 m3 & max. usable volume
15.8 m3 which have double walled, corrosion and gas resistant and is constructed of stainless steel and is
designed to withstand the extreme pressures and elevated temperatures utilized in the EO sterilization
process. The inner surface of sterilizer is smoothly finished to minimize gas deposits. The chamber is
insulated against heat emission and the jacket is connected to the warm water circulation arrangement.
The sterilizer doors have a quick release locking arrangement with door opening. Suitable safety locking
arrangement has provided for the door.
The sterilizer has been provided with an automatic programmable panel with memory for preset operating
sequence of all programs of operation. Review the temperature of each channel. T1 is selected as a master
channel to control the cycle.
Independent system such as pressure gauge, temperature gauge, has been provided with the sterilizer for
proper operation and monitoring of sterilizing process to cross check between control and monitoring which
identifies any discrepancies and indicates fault so that an ineffective process doesn’t appears effective.
The ETO gas sterilizer is fully automatic controlled by computer. The control system layout for the PC based
system for ETO sterilizer is as under:
Main
Anal Relay Powe Temperat
mother Digital Control
og Outp r ure
board I/O Panel
Inpu ut Suppl Transmitt
with Intel Cards
t Cards y er Cards
Pentium
Card Card
Processor
s s
CPU & Control Box
If the fresh (Non-sterile) load is not available then we will take sterile load for fractional &
half cycle to performed the validation & testing will be performed after validation cycle (fractional & half
cycle)
For Physical performance qualification testing, product packaging & functional testing will
performed.
Biological indicators (BI) shall be placed inside the articles as defined in diagram-1.0,
product for sterility testing, BET test & product functioning testing shall be placed at different locations in
sterilization chamber.
Each new validation cycle shall be taken after getting the results of biological indicators of
previous validation cycles.
The 3 way stop cock product load used for PQ is representative of that to be sterilized routinely/ most
frequently and shall be defined based upon the most challenging routine load among the family. Product
three way stop cock which have 288 corrugated boxes during sterilization load for validation of EO
sterilization process. Loading pattern for three way stop cock has maximum density & occupies max. volume
of sterilizer among the family which is sterile in EO sterilizer (ETO-04) as mention in Table -1. Factors, which
may be considered in this load, are those likely to affect penetration of heat, humidity and gas.
Table – 1
Three way sop cock as a Product load used for PQ should be at least as difficult to sterilize as the most
challenging load. For validation cycles, the maximum quantity of product representing the densest load
configuration that will be processed in full routine cycles must be loaded in the EO sterilizer chamber.
The transit containers are placed inside the chamber as per the loading pattern presented in Diagram 1.1.
A range of product is to be sterilized together in one cycle. The load used for sterilization process
qualification reflects the most difficult combination based on the product design of different sizes, length or
thickness, design complexity & amount and type of packaging as mention below:
Physical comparison:
5 A.V. Fistula Needle 450 x 65 mm 500 x 110 x 140 575 x 525 x 305 mm
. mm
6 Polyspike 105 x 55 mm 215 x 205 x 160 850 x 450 x 225 mm
. mm
The following requisite must be fulfilled before the ethylene oxide sterilization process validation.
Certification of EO gas
Ethylene oxide gas use as sterilant in the ratio of EO gas (30 %) + CO 2 (70 %). Concentration of EO gas
shall be verified by supplier’s certificate.
Calibration
All critical measuring devices e.g. temperature sensors, humidity sensor, pressure gauges, utilized on
ethylene oxide sterilizers which may impact the quality of the process are calibrated to traceable national or
international standards.
Population of viable microorganisms on finished product shall be determined before the sterilization process.
It is collected to quantify the amount of product contamination. At least 5 unsterilized samples from the
representative EO sterilized product families to be tested for bio-burden. The product bio burden is tested as
per GTP/QC/12 and record is maintained in F/QA /40.
Product Bio- Burden is tested as per ISO 11737-1. To ensure the product with low bio- burden the product is
manufactured in class 7 at rest and class 8 in dynamic condition.
For the validation study, biological indicators (BIs) of Bacillus Atrophaeous with a minimum population of
spore of 106 are used to challenge EO cycles. The population label claim from each lot of biological
indicators used for validation testing must be verified as per in-house specification. A supplier certificate can
be supported to verify the population of biological indicator count.
A minimum of 40 biological indicators (Spore strips/ or ampoules), 20 temperature sensors and 8 humidity
sensors are to be placed in one validation cycle (Table –2). Highlighted areas are the locations of spore
strips (BIs) as per the Diagram - 1.1. Temperature sensors and humidity sensors kept inside the corrugated
box. Except from these, two temperature sensors & two humidity sensors are placed in to the sterilizer
chamber for monitoring. One temperature sensor (T-3) is kept inside the gas supply line to record the EO
gas inlet temperature.
Table-2:
Minimum no. of BI’s / Temperature sensors/ Humidity
Sterilizer Size of sensors to be used for PQ
No. sterilize
BI’s Temperature Humidity
r
sensor sensor
ETO-04 20 m3 40 Nos. 20 8
A process challenge device is a device used to assess the effective performance of a sterilization process by
providing a challenge to the process that is equal to or greater than the challenge posed by the most
difficult item routinely processed.
The selection of process challenging device is based on product design which is influence the sterilization
process. The three way stop cock with extension tube is designed in such a way that product is fully closed
from all side of device & tubing product with longest lumen, having most convoluted passageways on basis
of design. Average Product bio-burden is higher than the other product from the family. We compared to
various product into the family at various aspect for selection of PCD as mention in table -3 & 4 which
deemed difficult to sterile product among the family & will be used as the BI’s (Bacillus atropheous) carrier.
Table-3
Overall
Average
S Name of density Unit
Raw material used product
. Product (kg/m3) packaging
Bio-
N of load material
burden
o used
.
Three way stop HDPE, Paper &
1 117.14 5 CFU/ Device
cock Polycarbonate, Tyvek/PVC Film
kg/m3
Polypropylene
Three way stop
HDPE,
2 cock with 87.07 kg/m3 Paper & PE Film 10 CFU/
Polycarbonate,
extension Device
Polypropylene, ABS
tube
IV Cannula HDPE, Polycarbonate,
Paper &
3 with 76.55 kg/m3 Polypropylene, 6 CFU/ Device
Tyvek/PVC Film
three way Poly Vinyl
Page No Prepared By Reviewed By Approved By
Gaurav Joshi Alpesh Ghoghari Viral Sheth
12 of 34
Medivac Surgical Pvt. Ltd.-Kutch, Gujarat India. Doc No : MVS/QP/19.0/A-06
Revision No : 0.0
PROCESS VALIDATION - ETO STERILIZER
<<<<X<
Overall
Average
S Name of density Unit
Raw material used product
. Product (kg/m3) packaging
Bio-
N of material
burden
o load used
.
A.V. Fistula Poly Vinyl
4 106.89 Paper & PE 7 CFU/
Needle Chloride, LDPE, Film
kg/m3 Device
HDPE, ABS
Polycarbonate,
Rigid Polyvinyl
5 IV. Flow 99.45 kg/m3 Paper & PE 7 CFU/
chloride, TPR, Rigid Film
regulator Device
Polyvinyl chloride,
HDPE, ABS
Acrylonitrile
6 Polyspike 73.50 kg/m3 Butadiene Paper & PE 5 CFU/ Device
Styrene (ABS), Film
Polypropylene
Table-4
Having a opening
Less Less closure
6 Polyspike from cover & from 6 cm
internal present
the thread
surface
of luer lock.
area
To create a challenge to the sterilization process, PCD, Bacillus atropheous having minimum 10 6 population,
by placing biological indicators in the product at location where sterilizing condition are most difficult to
achieve. Although the product are designed in such a way that EO gas reaches to all locations in the product
for example the passage from luer cover help EO gas to reach inside the channel &, handle, screw cap,
tubing etc. But still biological indicators are kept at the locations indicated below diagram 1.0.
Spore Strips
Diagram – 1.0
Location of Data logger (Sensors), Bi’s & samples for physical &
microbiological testing in to the product load volume
Three way stop cock is choosing as product load for validation. Temperature and humidity sensors should be
located within the sterile barrier system or amongst the unit packages in the sterilization load different
location i.e trolley centers, trolley edges & surfaces. A minimum of 20 numbers of the sensors for
temperature and 8 numbers of sensors for humidity measurement are located at different locations into the
product load volume. Samples for physical testing (like product functional testing & packaging testing) as
well as microbiological testing (like product sterility, BET, EO etc.) are placed different location. Following
are the location for the placement of temperature sensors, Humidity sensors & Biological indicator are given
below:
Chemical indicator tape is pasted on all containers kept inside the sterilizer chamber. All boxes containing the
biological indicators shall be marked for proper identification.
LOCATION PATTERN FOR PRODUCT
TROLLEY-1 TROLLEY-2
2 15 10 5 68 63 58 53
0
19 14 9 4 67 62 5 52
7
18 13 8 3 66 61 56 51
17 12 7 2 65 60 55 50
16 11 6 1 64 5 54 49
9
28 24 76 72
27 23 75 71
26 22 74 70
25 21 73 69
48 43 38 33 96 91 86 81
47 4 37 32 9 90 85 8
2 5 0
Page No Prepared By Reviewed By Approved By
Gaurav Joshi Alpesh Ghoghari Viral Sheth
17 of 34
Medivac Surgical Pvt. Ltd.-Kutch, Gujarat India. Doc No : MVS/QP/19.0/A-06
Revision No : 0.0
PROCESS VALIDATION - ETO STERILIZER
<<<<X<
46 41 36 31 94 89 84 79
45 40 35 30 93 88 83 78
44 39 34 2 92 87 82 77
9
TROLLEY-3 TROLLEY-4
11 1 10 10 1 15 15 14
6 1 6 1 6 9 4 9
1 4
11 11 10 10 16 15 15 14
5 0 5 0 3 8 3 8
11 10 10 99 16 1 15 14
4 9 4 2 5 2 7
7
1 10 10 9 16 15 15 14
1 8 3 8 1 6 1 6
3
11 10 10 97 16 15 15 1
2 7 2 0 5 0 4
5
14 13 13 12 19 18 18 17
4 9 4 9 2 7 2 7
1 13 13 1 19 18 18 17
4 8 3 2 1 6 1 6
3 8
14 13 13 12 1 18 18 17
2 7 2 7 9 5 0 5
0
14 13 13 12 18 18 17 1
1 6 1 6 9 4 9 7
4
14 13 1 12 18 18 17 17
0 5 3 5 8 3 8 3
0
TROLLEY-5 TROLLEY-6
21 20 20 19 2 25 25 24
2 7 2 7 6 5 0 5
0
2 20 20 1 25 25 24 24
1 6 1 9 9 4 9 4
1 6
21 20 20 19 25 2 24 24
0 5 0 5 8 5 8 3
3
20 2 19 19 25 25 24 2
9 0 9 4 7 2 7 4
4 2
20 20 19 19 25 25 24 24
8 3 8 3 6 1 6 1
24 23 2 22 28 28 27 2
0 5 3 5 8 3 8 7
0 3
23 23 22 22 28 2 27 27
9 4 9 4 7 8 7 2
2
2 23 22 22 28 28 27 27
3 3 8 3 6 1 6 1
8
23 23 22 22 28 28 27 27
7 2 7 2 5 0 5 0
23 23 22 22 28 27 2 26
6 1 6 1 4 9 7 9
4
Diagram -1.1
Schematic Diagram Showing Location of RTD Sensors & humidity sensor inside chamber
EO gas pipeline
UNLOADING SIDE
T3
RH 2
T1
T2
RH 1
Location of Sensors:
That the specified acceptance criteria are met throughout the load for the duration of the proposed
routine process specification, and
The PPQ shall include following planned qualification cycles in table-3, in which all the specified acceptance
criteria shall be met. PPQ is performed in parallel with MPQ runs; therefore a minimum of one additional
PPQ run (280 minutes) shall be performed using the full routine process specification.
Table-3
Sr. Cycle no. Exposure Time Cycle Description
No.
1 1st cycle 20 Minutes Fractional cycle
2 2nd cycle 40 Minutes Fractional cycle
3 3rd cycle 60 Minutes Fractional cycle
4 4th cycle 80 Minutes Fractional cycle
5 5th cycle 100 Minutes Fractional cycle
6 6th cycle 120 Minutes Half Cycle
7 7th cycle 140 Minutes Half Cycle
8 8th cycle 140 Minutes Half Cycle
9 9th cycle 280 Minutes Full cycle
10 10th cycle 280 Minutes Full cycle
The EO sterilizer is equipped with computer control software to meet the above requirement. EO Sterilization process
steps are described below:
Preconditioning: Treatment of product prior to the sterilization cycle in the chamber to attain specified
limits for temperature and relative humidity. The sterilization chamber is used as pre conditioning area. The
load is kept on pallets as per loading pattern.
The load in pre-conditioning is kept for 60 minutes to attain a minimum temperature of 45ºC ± 5ºC and
humidity between 30% to 90% at the end of pre- conditioning.
Conditioning: Treatment of product within the sterilization cycle, but prior to sterilant admission, to attain
a predetermined temperature and relative humidity. This part of the sterilization process is carried out under
vacuum.
Since the pre-conditioning & conditioning area are same so all parameters like loading pattern and location
of temperature & humidity sensors remain same.
package’s sterile barrier properties are maintained, thus ultimately protecting the sterility of the product
once the process is complete. The rate of vacuum up-to -0.75kg/cm2 is approximate 30 ± 15 minute.
After conditioning time (RH dwell phase), the gas injection is started. The gas, which is admitted inside the
chamber, is preheated and is in vapor form with a temperature between 20ºC~60ºC. If the temperature
comes below 20ºC, the injection of EO will be automatically stopped.
The amount of gas added was calculated to a particular weight and weighed using a calibrated balance.
However, the evaluation of the EO gas concentration injected can be correlated with the pressure of
sterilizer chamber.
Since between preconditioning / conditioning are performed in sterilizer chamber, therefore maximum
elapsed time between condition and sterilization is NIL.
Exposure phase: After the gas has been injected, the exposure phase of the process is performed. This is
the phase in which the product is exposed to heat, relative humidity, and gas for a predetermined amount of
time. The amount of exposure time is determined after careful analysis of the product, load configuration,
and desired level of sterility. During EO exposure period, sterilizer chamber temperature should have
45±5°C, temperature of sterilization load should be 35±5°C.
Aeration
After completion of required exposure time, Ethylene Oxide gas is removed/ flushed from the chamber and
package (product) up to -0.75 kg/cm² with the rate approximate 40±20 minute and the Aeration is started.
Flushing is done by removing the air from chamber and package by creating a vacuum deep to -0.65
kg/cm². The rate of air wash phase is approximate 30±15 minutes. Air flushing is done two times. During
aeration temperature within the chamber/ area (should be 35°C - 60°C) & Temperature of the sterilization
load (30°C – 50 °C) shall be recorded.
The parameters for standard process used for routine sterilization cycle, which is being validated are as
follows:
Acceptance
Sr. CYCLE PHASE Specificati
Criteria/ Limit
No on
.
Minimum Temperature of product to enter the Sterilization
a) NLT 20 °C NLT 20 °C
Process
Pre-Humidification Phase (Steam Injection)
1
Time 5 minutes 1 minute
b) Preconditioning (if used)
1 Time in Chamber/Area 60 minutes 2 minute
2 Temperature of Chamber/Area 45°C 5°C
3 Humidity of Chamber/Area 60% 30% ~ 90%
4 Temperature of sterilization load 35°C 5C
5 Humidity of sterilization load 60% 30% ~ 90%
c) Vacuum levels and rate of evacuation (if used)
1 Vacuum levels -0.75 0.01Kg/cm2
Kg/cm2
2 Vacuum Rate 30 minutes 15 minutes
3 Holding time under vacuum 10 minutes ± 1 minutes
e) Conditioning (RH Dwell Phase)
Chamber Pressure after insertion of steam (Relative NLT -0.70
1 kg/cm2 NLT -0.70 kg/cm2
Humidity)
One time &
2 Number of steam pulses & Duration 2-5 minutes
2
minutes
3 Conditioning time (RH Dwell Time) 10 minutes ± 1 minutes
Acceptance
Sr. CYCLE PHASE Specificati
Criteria/ Limit
No on
.
Temperature of the sterilization load at the end of
5 conditioning 35 °C 5C
6 Humidity of the sterilization load at the end of conditioning 60% 30% ~ 90%
f) EO Injection and Exposure
1 EO injection pressure rise (ΔP) 0.25 Kg/cm2 ± 0.05kg/cm2
2 EO injection time 40 min 20 minute
3 Terminal pressure of EO injection phase 0.25 Kg/cm 2
± 0.05 Kg/cm2
4 EO gas Inlet Temperature NLT 20C 20C ~ 60C
5 Mass of EO used 36.7 kg 35 kg ~ 38.3 kg
6 Concentration of EO 550 mg/lt ± 25mg/lit
7 Sterilizer chamber temperature 45°C 5 C
8 Exposure time As point ± 1 minutes
no. 6.0
9 Temperature of the sterilization load 35°C 5 C
h) Aeration
-0.75 0.01 kg/cm2 @
1. Gas Discharge phase
kg/cm2 @ 20 minutes
40 Minutes.
2. Air Inlet phase Up to 0 ± 0.05kg/cm2
kg/cm2
-0.65 0.01 kg/cm2 @
3. Air washing phase and Rate
kg/cm2 @ 15 minutes
30 Minutes.
4. Air inlet phase Up to 0 ± 0.05kg/cm2
kg/cm2
5. Temperature within the Chamber/Area 45°C ± 15 C
6. Temperature of the sterilization load 40°C ± 10 C
7. Total Aeration 2 Times 2 Times
The total inactivation of microorganisms using EO is attained when sterilizing conditions are met within the
chamber. The four active ingredients required to deliver a successful process are:
• Heat/Temperature
• Moisture
• Gas concentration
• Exposure Time
The microbiological Performance Qualification (MPQ) shall demonstrate that, on application of the
sterilization process, the specified requirements for sterility are met. Exposure time is the key parameter that
is varied during microbiological qualification to provide assurance that MPQ delivers less lethality than the
normal production process. Determination of lethal rate of sterilization process (Bi/ bioburden approach) is
carried out by fraction negative method {Annex A, clause A.1.3. (b)} & conservative determination of lethal
The Overkill Approach is most commonly used to validate ethylene oxide cycles used in the sterilization of
medical devices. The overkill method is based on demonstrating that the sterilization of a microbial
challenge/ biological indicator (Bacillus atropheous) exceeds the challenge posed by the bio-burden of the
product.
MPQ shall include at least three half cycles for determination of the minimum time of exposure to EO, with
all other process parameters except time remaining constant, at which biological indicators are not showing
any growth. The specified exposure time should be at least double this minimum time.
Following cycles to be taken during the process qualification & calculation of D- Value by Limited Holcomb-
Spearman-Karber Procedure (LHSKP)
Table-2 — calculation of D- Value by LHSKP with constant time intervals and constant number
of samples
D value calculation: Formula to be used for fraction negative method for D value calculation as per clause
C.3.2 Limited Holcomb-Spearman-Karber Procedure (LHSKP) of ISO 14161:2009 (E).
After exposure, the test samples are inoculated into Soyabean Casein Digest Medium and observe the
growth for 7 days.
ACCEPTANCE CRITERIA FOR PERFORMANCE QUALIFICATION-MICROBIOLOGICAL
At least 5 unsterilized samples from the representative EO sterilized product families to be tested for bio-
burden. The product bio burden is tested as per ISO 11737-1 & GTP/QC/12 and record is maintained in
F/QA /40. Product Bio- Burden should be NMT 100 CFU/device.
Exposed biological indicator (spore strips) in 40 nos. for each cycle are
inoculated in Soyabean Casein Digest Medium Previously sterilized at 121ºC for 15 minutes and kept at
30-35ºC in BOD incubator for 7 days. . The media is monitored daily for any growth of microorganism.
Exposed biological indicator (ampoules) (if used) are crush & incubate processed
and control BI’S for 48 hours at 30-35ºC in BOD incubator.
The result shall comply test plan as per clause 7.0 and record shall be maintained F/QA/35. BI’s should be
incubated within 24 hours after cycle completion (unloading).
The sample (minimum 5 in nos.) would be drawn out for EO residual testing as mention below:
For fractional cycle- Sample would be collected after execution of full routine cycle
For half cycle- Sample would be collected after execution of full routine cycle.
For Full cycle – Sample would be collected.
This test is used to determine the concentration of Ethylene Oxide in compliance with the EN ISO 10993-
7:2008 standard. EO residual testing will be performed by external NABL certified Lab. Acceptable residual
limits for Ethylene oxide is NMT 4mg/ 24 hrs, & for Ethylene Chlorohydrin is NMT 9 mg/ 24 hrs. Simulative
used extraction method is using for testing of EO residual.
Note- if the fresh load is not available then we will take sterile load for fractional & half cycle & performed
the EO residual test after the fractional & half cycle execution.
Take 3 sterilized samples from the EO sterilized product to be tested for bacterial endotoxin test. The
sterilized product shall comply the Bacterial Endotoxin Testing using Gel-clot method as per GTP/QC/10.
20 sterilized samples from the EO sterilized product load to be tested for sterility test. The sterilized
product shall comply the Sterility test as per GTP/QC/09, revision no.-02.
Product is tested randomly for the effect of sterilization on packaging condition, printing matter, sealing or
any other effect due to temperature & humidity as well as functioning of product. The effect of sterilization
shall not have any detrimental effect on the package.
Data logger, internal PCD & samples shall be collected after the completion of cycle for further evaluation.
For process parameter analysis, collect the data from the routine graph as well as data logger cycle wise.
All the data will be carefully reviewed & should be met as per specification/ acceptance criteria.
Any deviation from the acceptance criteria of the specific check point shall be reported in document
F/QA/168 Revision no. - 00 and decision should be taken for the rejection, replacement or rectification of the
equipment/component.
Fractional cycle & half cycle are followed by Full sterilization cycle during PPQ & MPQ. So product will be
verified product packaging, functioning test & EO residual test after repeatability sterilization process. If the
product failed to achieve the acceptance criteria in any test like product functioning test, EO residual test &
Product packaging test, then it should be disposed off or discarded in presence of QA & recorded in
validation report.
DATA RECORDING
Data shall be recording in following annexure & format during performance qualification.
For report for effect of EO sterilization on product package shall be recorded in Annexure-I
Annexure-I
EFFECT ON EFFECT
MODE PACKING
OF
OF MATERIA
SEALIN PRINTI DEFORMATIO COLOR HUMIDI
PACKIN L
G NG N TY ON
G
PAPER
Medical
UNIT grade
PACKING paper /
Tyvek &
rigid
PVC
film
DUPLEX Card
CARTON board
paper
Page No Prepared By Reviewed By Approved By
Gaurav Joshi Alpesh Ghoghari Viral Sheth
31 of 34
Medivac Surgical Pvt. Ltd.-Kutch, Gujarat India. Doc No : MVS/QP/19.0/A-06
Revision No : 0.0
PROCESS VALIDATION - ETO STERILIZER
Effective Date : 01-06-2018
F88 PERFORMANCE QUALIFICATION OF STERILIZER
CORRUGA Semi
TED BOX craft
paper
TRAINING GIVEN BY
Name
Designation
Signature/ Date
Venue
Date of training
Duration of training
15.0 EVALUATION
The EO sterilizer shall be tested for its killing efficiency by review of the data for biological indicator and RTD
sensors on different locations.
Chamber leak test of EO sterilizer shall be performed (Integrated in Standard process test) and pressure
drop during vacuum hold time shall be observed.
The RTD sensors and biological indicators shall be placed in to the product load volume at different
positions. At any time interval of sterilization hold time no RTD sensor shall show the temperature difference
± 5°C from the mean temperature of all the temperatures. The chemical indicators shall show the colour
changes as per the instructions of the manufacturer.
The validation report shall include but not limited to - date of study initiated, date completed, name and
signature of execution team, material and instruments used, observations made, problems encountered,
completeness of information collected, summary of deviation report, results of tests (conclusion), do results
meet acceptance criteria, other information relevant to the study and conclusions on the validity of the
equipment/system.
REQUALIFICATION CRITERIA
Once in a year
Major sterilizer repairs and changes
Changes to construction or relocation
Unexplained sterility failures in routine sterilization
Changes to product or packaging
Modification to the sterilizing agent