DICOM MPPS + SC
AX

Installation and Startup


DICOM MPPS + SC

AXIOM Artis Systems and Systems with


FLUOROSPOT T.O.P., POLYTRON T.O.P.,

05916791

© Siemens 2001
The reproduction, transmission or use
of this document or its contents is not
permitted without express written
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.

Print No.: AX41-000.814.01.02.02 English


Replaces: AX41-000.814.01.01.02 Doc. Gen. Date: 08.01
2 Revision / Disclaimer
1Revision / Disclaimer

Document revision level


The document corresponds to the version/revision level effective at the time of system
delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.

Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise autho-
rized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with
or authorized by Siemens or one of its affiliates are directed to contact one of the local
offices of Siemens or one of its affiliates before attempting installation or service proce-
dures.

DICOM MPPS + SC AX41-000.814.01.02.02 Page 2 of 26 Siemens


08.01 TD SD 21
Table of Contents 3
0Table of Contents

1 _______ General Information______________________________________________ 4

Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Applicability and Purpose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
System Requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Required Aids and Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Additionally Required Documents. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Room Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Remark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Product-specific Remarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

2 _______ Installation _____________________________________________________ 7

3 _______ Image System Setup _____________________________________________ 8

FLUOROSPOT TOP Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8


Configuring the POLYTRON TOP. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Configuring the AXIOM Artis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

4 _______ ExamWorks+ Setup _____________________________________________ 11

General Remarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Network Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Accessing the Network Configurator. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Service Login . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Configuration Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Application Language Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Network Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Config in the System Configuration Time and Date . . . . . . . . . . . . . . . . . . . . . . . . . 19
Input Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Workflow Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Output Configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Reboot ExamWorks+ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Configure Time and Date . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Save the Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
ExamWorks+ Shutdown from GUI (Graphic User Interface). . . . . . . . . . . . . . . . . . . 23

5 _______ Final Work Steps _______________________________________________ 25

Function Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25

6 _______ Changes to Previous Version _____________________________________ 26

Siemens AX41-000.814.01.02.02 Page 3 of 26 DICOM MPPS + SC


08.01 TD SD 21
4 General Information
1-
Requirements
1General Information

Applicability and Purpose 0

Installation and startup of the DICOM MPPS + SC (ExamWorks+) option in the FLUO-
ROSPOT T.O.P., POLYTRON T.O.P. systems and systems with AXIOM Artis are described
in this document.
With the DICOM Modality Performed Procedure Step (MPPS), progress of the examination
in the modality is reported to the RIS and data such as the dose area product can be trans-
ferred to the RIS. With the DICOM Storage Commitment (SC) it is assured that DICOM
images will be reliably saved in the image archive.
Images to be archived are sent from the image system to ExamWorks+. ExamWorks+ han-
dles archiving and saves the images until archiving has been successfully completed.
Additional patient data (e.g. dose data) can be entered in the ExamWorks+ via the touch-
screen monitor.
The keyboard that is included in the shipment is connected exclusively for service and con-
figuration purposes. After completing these work steps, the keyboard must be removed.

System Requirements 0

• Systems with FLUOROSPOT T.O.P.


- DICOM Worklist Option
- DICOM Send Option
- Software Version ≥ VD10A
- Service CD ≥ 6
• Systems with POLYTRON T.O.P.
- DICOM Worklist Option
- DICOM Send Option and/or Udig Option
- Software Version ≥ VF00R
• AXIOM Artis Systems
- DICOM Worklist Option

Required Aids and Tools 0

All tools, test equipment and aids, with the exception of the "standard installation tool kit",
are listed and specified in the Service Tools Catalogue.

• Standard installation tool kit

DICOM MPPS + SC AX41-000.814.01.02.02 Page 4 of 26 Siemens


08.01 TD SD 21
General Information 5

Additionally Required Documents 0

• Project Plan
• FLUOROSPOT T.O.P. Installation and Setting Instruction HCC, SIENET,
DICOM, RIS, Paper Print, SW ≥ VD10A
• POLYTRON T.O.P. Configuration Guide
• AXIOM Artis Configuration System

Room Preparations 0

• A power line outlet must be installed in a radius of 1.5 m from the installation location.
Since the ExamWorks+ is switched off only in a service situation, connection to line
power may not be made in the control room distributor.
• Have the RJ45 Ethernet connection cable ready at the setup location.

Remark 0

This document was originally written in German.

Safety Information 0

When carrying out the work steps and checks, the general safety information for medical
products must be observed.

Product-specific Remarks 0

Checking the Cabling and Power Connection


• For safety reasons, the power connection may only be made by a SIEMENS techni-
cian/electrician.
• Cabling must also be checked by the SIEMENS technician prior to switching on the sys-
tem.

Siemens AX41-000.814.01.02.02 Page 5 of 26 DICOM MPPS + SC


08.01 TD SD 21
6 General Information

Service 1.1

• The “Installation and Service Manual” from the Merge Company is included in the ship-
ment as a service document.
• In an error situation, the complete ExamWorks+ must be replaced.
• Contact with Merge should be made only through HSC.

DICOM MPPS + SC AX41-000.814.01.02.02 Page 6 of 26 Siemens


08.01 TD SD 21
Installation 7
2Installation 2-

1
2

Fig. 1: ExamWorks+ Fig. 2: ExamWorks+ cable connections

• Install the desktop stand included in the shipment according to the accompanying
installation instructions.
• Set up the ExamWorks+ per the Project Plan. As a rule, the ExamWorks+ is set up in
the control room within easy reach of the operator. Make sure that there is sufficient
space left for the ventilation slots (> 15 cm).
• Provide a power supply (1/Fig. 2 / p. 7).
• Plug in the Ethernet cable (2/Fig. 2 / p. 7).
• Snap the rear cover panel onto the desktop stand.

Siemens AX41-000.814.01.02.02 Page 7 of 26 DICOM MPPS + SC


08.01 TD SD 21
8 Image System Setup
3-
FLUOROSPOT TOP Configuration
3Image System Setup

• Configure the ExamWorks+ as a new network node in the image system (Create new
record) and add the ExamWorks+ in the group members in “my nodes” (FLUO-
ROSPOT T.O.P.)n.
• For this, see:
FLUOROSPOT T.O.P.
Installation and Setting Instructions
HCC, SIENET, DICOM, RIS, Paper Print, SW ≥ VD10A

DICOM MPPS + SC AX41-000.814.01.02.02 Page 8 of 26 Siemens


08.01 TD SD 21
Image System Setup 9

Configuring the POLYTRON TOP 3.1

• Configure the ExamWorks+ as a new network node in the image system (Create new
record) and add the ExamWorks+ in the group members in “my nodes” (POLYTRON
T.O.P.).
• For this, see:
Configuration Guide
System software ≥ VF0..

Siemens AX41-000.814.01.02.02 Page 9 of 26 DICOM MPPS + SC


08.01 TD SD 21
10 Image System Setup

Configuring the AXIOM Artis 3.2

The ExamWorks+ is configured as a new DICOM network node in the AXIOM Artis Sys-
tem.
• For this, see:
AXIOM Artis
Configuration
System

DICOM MPPS + SC AX41-000.814.01.02.02 Page 10 of 26 Siemens


08.01 TD SD 21
ExamWorks+ Setup 11
4-
General Remarks
4ExamWorks+ Setup

The ExamWorks+ Service Script is a configuration interface with help files to assist you in
understanding the information required for a given field. The Service Script configures the
following:
• User interface language
• Network
• Time and Date
• DICOM inputs and outputs
• Workflow, including MWL and MPPS

Network Configuration 0

This section explains how to configure the ExamWorks+ network parameters such as static
IP address and netmask. Contact the on-site system administrator for IP Address, Net-
mask, Host name, and related network information.

Accessing the Network Configurator 0

To access the Network Configurator, select Perform Network Configuration from the Ser-
vice Script.

Siemens AX41-000.814.01.02.02 Page 11 of 26 DICOM MPPS + SC


08.01 TD SD 21
12 ExamWorks+ Setup

Configuration 4.1

Preparations 0

• The keyboard included in the shipment must be connected for configuration and ser-
vice purposes.
• The following configuration data must be requested from the customer’s network
administrator:

Configuration Data: ExamWorks+

ExamWorks+ IP address:
Subnet mask:
Host Name: EWPlus1
Domain Name:
Default Gateway:

1. Suggestion, can be changed.

Configuration Data: Storage SCP ExamWorks+

Storage SCP Local AE Title1:


Port No:

1. This data’s also have to be configured in the modality (AX System)

Configuration Data: Modality Worklist Server Information

Modality Worklist Server Infor- MWL Server IP address:


mation
MWL Server Port No:
Called AE Title 1.:
Local AE Title :
Local AE Title for Modality
Unit (Imaging System2).:

1. The CALLED_AE title is the AE title of the MWL server in the network.
2. The LOCAL_AE Title used for Modality Specific Worklist query. The value is the same as that configured
in the modality (AX System).

DICOM MPPS + SC AX41-000.814.01.02.02 Page 12 of 26 Siemens


08.01 TD SD 21
ExamWorks+ Setup 13

Configuration Data: MPPS Information

MPPS Information MPPS Server IP adress:


MPPS Server Port No:
Called AE Title 1.:
Local AE Title .:

1. The Called_AE Title is the AE Title of the MPPS Server in the Network.

Configuration Data: Storage Commitment Information

Storage Commitment Informa- Local AE Title1:


tion
Local Listening Port2:

1. The local AE title is the ExamWorks+ Storage Commitment AE title.


2. ExamWorks+ Storage Commitment Port Number

Configuration Data: Storage Target Information

Storage Target Information


Target 1 Target Name1:
Target IP address:
Target AE Title.:
Target Port Number:
Storage Commit called
AE Title2:
Storage Commit Target
IP adress:
Storage Commit Target
Port Number:

Siemens AX41-000.814.01.02.02 Page 13 of 26 DICOM MPPS + SC


08.01 TD SD 21
14 ExamWorks+ Setup

Target 2 Target Name1:


Target IP address:
Target AE Title.:
Target Port Number:
Storage Commit called
AE Title2:
Storage Commit Target
IP adress:
Storage Commit Target
Port Number:

Target 3 Target Name1:


Target IP address:
Target AE Title.:
Target Port Number:
Storage Commit called
AE Title2:
Storage Commit Target
IP adress:
Storage Commit Target
Port Number:

Target 4 Target Name1:


Target IP address:
Target AE Title.:
Target Port Number:
Storage Commit called
AE Title2:
Storage Commit Target
IP adress:
Storage Commit Target
Port Number:

1. The target name is the user readable target name.


2. This is the AE title, the ExamWorks+ uses to send storage commitment request for this target. Only
needed if the target AE title and storage commit AE title are different.

DICOM MPPS + SC AX41-000.814.01.02.02 Page 14 of 26 Siemens


08.01 TD SD 21
ExamWorks+ Setup 15

Service Login 0

• Apply power to the ExamWorks+ unit.


• Select the <Utils> tab.
• Select the <Exit to Service> button.
• Type the following password.
- Password: <m0rg01>
• Select the <Done> button.
• Press <Ctrl> + <Alt> + <Backspace> on the keyboard to exit to the LINUX command
line prompt.
• At the prompts, enter:
- Login: service ↵
- Password: m0rg01 ↵

Fig. 3: Main menu for ExamWorks+ service interface


The Main Menu for the ExamWorks service interface displays. The ExamWorks+ is now
ready to be configured.
• From the Main Menu for the ExamWorks Service Interface, select <1> ↵ Configure
System.
• Type the following password.
- Password: m0rg01 ↵

NOTE To navigate the Network Configurator, use the keyboard


<Tab> key to move among the fields and functions. Use the
<Up Arrow> and <Down Arrow> keys to move through
options in a group or list. Press the <Enter> or <Return> key
to select the highlighted field, function, or option.
Enter a question mark (?) to view help for any given prompt
in the Service Script.

Siemens AX41-000.814.01.02.02 Page 15 of 26 DICOM MPPS + SC


08.01 TD SD 21
16 ExamWorks+ Setup

Configuration Menu 0

Fig. 4: Configuration Menu


• In the Configuration menu, select <1> ↵ for Perform Full Configuration.

Application Language Configuration 0

• Select the desired language.

Network Configuration 0

Fig. 5: Network configurator

Basic Host information


• Im Network Configurator unter Client tasks <Basic host information> ↵ auswählen.

DICOM MPPS + SC AX41-000.814.01.02.02 Page 16 of 26 Siemens


08.01 TD SD 21
ExamWorks+ Setup 17

Fig. 6: Basic host configuration


• Enter the Host name (Configuration Data: ExamWorks+ / p. 12).
• Set Adaptor 1 to Enabled. An X in the bracket [ ] field to the left of Enabled indicates
that the adaptor is enabled.
• Set Config mode to Manual. An o in the bracket (o) field indicates that the Config mode
is Manual.
• Enter the Primary Name + Domain (Configuration Data: ExamWorks+ / p. 12). The
Primary Name and the Host name must be the same (e.g.: acomtdf77.med.sie-
mens.de).
• Enter an Aliases name (Option)
• Enter IP adress in the IP address field (Configuration Data: ExamWorks+ / p. 12).
• Enter a netmask in the Netmask field (Configuration Data: ExamWorks+ / p. 12).
• The Net Device field is pre-configured. Do not change this field.
• Select <Accept> ↵ to complete the settings.
- Use the keyboard <Tab> key to move among the fields.

Name Server Specification (Optional)


This fields needs to be filled in only if a DNS server is configured.

Siemens AX41-000.814.01.02.02 Page 17 of 26 DICOM MPPS + SC


08.01 TD SD 21
18 ExamWorks+ Setup

Fig. 7: Name server specification


• DNS usage, if a DNS server is available, enter an X in the bracket [ ] field.
• Enter IP address of name server 1.
• Enter the IP address of name server 2 (optional).
• Enter the IP Adress of name server 3 (optional).

Routing and Gateways


• Under Client Tasks, select <Routing and Gatways> ↵ in the Configurator.
• Select <set Defaults> ↵

Fig. 8: Default gateway


• Enter in the field “Default gateway“ the IP number of the main gateway (Configuration
Data: ExamWorks+ / p. 12).

DICOM MPPS + SC AX41-000.814.01.02.02 Page 18 of 26 Siemens


08.01 TD SD 21
ExamWorks+ Setup 19

• Leave the field of Enable routing disabled.


• Select <Accept> ↵ to complete the settings.
- Use the keyboard <Tab> key to move among the fields.
• Select <Quit> ↵ to complete the settings

Host name search path


• In the Network Configurator, select <Host name search path> under client tasks.

Fig. 9: Name service access


• Enter an <X> ↵ in the bracket [ ] field of Multiple IPs for one Host.
• Select <Accept> ↵ to complete the settings.
- Use the keyboard <Tab> key to move among the fields.
• Leave the Network Configurator with <Quit> ↵ .
• Activate the changes with <Quit> ↵ .

Config in the System Configuration Time and Date 0

• Time Zone Selection.


Select your time Zone.
• Enter <1> ↵ for accept changes in System configuration.

Siemens AX41-000.814.01.02.02 Page 19 of 26 DICOM MPPS + SC


08.01 TD SD 21
20 ExamWorks+ Setup

Input Configuration 0

• Enter the ExamWorks+ Local Storage SCP AE Title (Configuration Data: Storage SCP
ExamWorks+ / p. 12).
• Enter the ExamWorks+Storage Listening Port Number (Configuration Data: Storage
SCP ExamWorks+ / p. 12).
• Enter <1> ↵ for accept changes in System configuration.

Workflow Configuration 0

• Siemens Modality Model.


- Enter <1> ↵ for AXIOM Artis Systems.
- Enter <2> ↵ for Other Siemens System.

Modality Worklist Server Information (MWL)


• Enter the ExamWorks Local MWL AE title (Configuration Data: Modality Worklist
Server Information / p. 12).
• Enter the MWL Server AE title (Configuration Data: Modality Worklist Server
Information / p. 12).
• Enter MWL Server IP address (Configuration Data: Modality Worklist Server
Information / p. 12).
• Enter MWL Server Port Number (Configuration Data: Modality Worklist Server
Information / p. 12).
• Modality type to be provide by MWL. Enter <2> ↵ for no.
• Patient/Procedure to be schedule by Specific Modality System? Enter <1> ↵ for yes.
• Enter the Local AE title for the Modality Unit (Image System) (Configuration Data:
Modality Worklist Server Information / p. 12).
• DICOM field to display in column 1 of the Procedure In-Progress Table. <1> ↵ for
Patient Name.
• DICOM field to display in column 2 of the Procedure In-Progress Table. <4> ↵ for
Study Date.
• DICOM field to display in column 1 of the MWL Information Table. <1> ↵ for Patient
Name.
• DICOM field to display in column 2 of the MWL Information Table. <6> ↵ for Scheduled
Procedure Step ID.
• DICOM field to display in column 3 of the MWL Information Table. <5> ↵ for Scheduled
Start Time.
• DICOM field to display in column 4 of the MWL Information Table. <7> ↵ Scheduled
Procedure Description.
• MWL Manual Entry Allowed: <1> ↵ for Yes.

DICOM MPPS + SC AX41-000.814.01.02.02 Page 20 of 26 Siemens


08.01 TD SD 21
ExamWorks+ Setup 21

Modality Performed Procedure Step (MPPS)


• Enter ExamWorks+ MPPS Local AE Title (Configuration Data: MPPS
Information / p. 13).
• Enter MPPS Server AE title (Called AE Title) (Configuration Data: MPPS
Information / p. 13).
• Enter MPPS Server IP address (Configuration Data: MPPS Information / p. 13).
• MPPS Server Port Number (Configuration Data: MPPS Information / p. 13).
• Performing physician. <2> ↵ for No.
• Performing technologist. <2> ↵ for No.
• Protocol name allowed to be edited on ExamWorks. <2> ↵ for No.
• AXIOM Artis and Systems with FLUOROSPOTT.O.P. <2> ↵ for No; Systems with
POLYTRON T.O.P.<1> for Yes.
• Dose Data. <1> ↵ for Yes.
• Study ID provided by Modality. <1> for Yes.
• Accept changes in Workflow Configuration? <1> ↵ for Yes.

Output Configuration 0

Storage Commitment (SC)


• Configure with Storage Commitment? <1> ↵ for Yes.

CAUTION StorageCommitment Service shall only be configured if the


destination really supports it.
¹ A wrong configuration will lead to a malfunction of the
ExamWorks+ which can be solved only by MERGE ser-
vice. Contact with Merge should be made only through
HSC.

• Enter ExamWorks+ Storage Commitment AE title (Local AE Title) (Configuration Data:


Storage Commitment Information / p. 13).
• Enter ExamWorks+ Storage Commitment Port Number (Local Listening Port) (Configu-
ration Data: Storage Commitment Information / p. 13).
• Currently Configured Targets. <n> ↵ for new target.
Enter all targets to which the ExamWorks+ sends images or other data to.
• New target name as displayed to the user (Configuration Data: Storage Target
Information / p. 13).
• Enter Target AE title (Configuration Data: Storage Target Information / p. 13).
• Enter Target IP address (Configuration Data: MPPS Information / p. 13).
• Enter Target Port Number (Configuration Data: MPPS Information / p. 13).

Siemens AX41-000.814.01.02.02 Page 21 of 26 DICOM MPPS + SC


08.01 TD SD 21
22 ExamWorks+ Setup

• Target Service Selection


1. Storage
2. Storage with Storage Commitment.

NOTE If Storage Commit called AE Title und Target AE Title are the
same (Configuration Data: Storage Target
Information / p. 13), the following configuration of the Stor-
age Commit Target
does not have to be performed.

• Enter AE Title of Storage Commitment Device (Storage Commit called AE title) (Config-
uration Data: Storage Target Information / p. 13).
• Enter Target IP Number of Storage Commitment Device (Storage Commit called IP
adress) (Configuration Data: Storage Target Information / p. 13).
• Enter Port Number of Storage Commitment Device (Storage Commit called AE Title)
(Configuration Data: Storage Target Information / p. 13).
• Default on this target being selected on startup.
1. Yes (the target selection is automatically selected)
2. No (images can be sent after manual selecting to the selected target)
• Configure the remaining targets in the same way.
• Exit Output Configuration with <quit> ↵ .
• <1> ↵ for accept
• Exit the Configuration Menu with <quit> ↵ .

Reboot ExamWorks+ 0

• In Main Menu for Exam Works Service Interface (Fig. 3 / p. 15) enter <6> ↵ Reboot.
• Enter Password: m0rg01 ↵

Configure Time and Date 0

Accessing the Unix Command Prompt


• Log in to ExamWorks+ through the service login (Service Login / p. 15).
• Select menu item <0 Exit to Command Prompt> ↵ , from the Main Menu for Exam-
Works+ Service Interface.
• Press <Enter> ↵ to access the shell prompt.
• Enter <su> ↵
• Enter Password: m0rg01 ↵

DICOM MPPS + SC AX41-000.814.01.02.02 Page 22 of 26 Siemens


08.01 TD SD 21
ExamWorks+ Setup 23

Set Time and Date


The LINUX date command sets the time and date. For more information on the LINUX date
command, enter <man date> ↵ at a LINUX shell prompt.
Display or set the date UCT (universal) time. At command prompt.
• Enter <quit> ↵
• Enter <date MMddhhmmyyyy.ss> ↵

MM = Numeric month, a number from 1 to 12


dd = Day, a number from 01 to 31
hh = Hour, a number from 00 to 23
mm = Minutes, a number from 00 to 59
yyyy = Year, such as 2001
.ss = Seconds, a number from 00 to 61 (59 plus a maximum of two leap sec-
onds)

• Exit the Time and Date configuration with <quit> ↵ and <exit> ↵ .

Reboot ExamWorks+
• I the Main Menu for ExamWorks Service Interface (Fig. 3 / p. 15) enter <6> ↵ Reboot.
• Enter the password: m0rg01 ↵

Save the Configuration 0

• Enter the Configuration menu (Service Login / p. 15).


• Im Configuration menu, select <8> ↵ für Perform Configuration Backup.
• Insert an empty MS-DOS formatted floppy disk into the floppy drive. It is shipped along
with the system.
• Press enter to backup the configuration file.
• Store the backup configuration file in a safe place.
• Reboot the ExamWorks+.

NOTE You must reboot the ExamWorks+ after making changes to


the configuration. Rebooting the ExamWorks+ ensures
proper memory allocation and information distribution.

ExamWorks+ Shutdown from GUI (Graphic User Interface) 0

This section explains how to shut down ExamWorks+ from the GUI.

Siemens AX41-000.814.01.02.02 Page 23 of 26 DICOM MPPS + SC


08.01 TD SD 21
24 ExamWorks+ Setup

CAUTION Failure to follow shutdown procedures may result in lost or


corrupted data and files.
¹ Follow these shutdown procedures exactly.

• Select <System Shutdown> ↵ from the Utils Menu.


• Wait for the System confirm that it is safe to turn off ExamWorks+ power.

DICOM MPPS + SC AX41-000.814.01.02.02 Page 24 of 26 Siemens


08.01 TD SD 21
Final Work Steps 25
5-
Function Test
5Final Work Steps

HIS/RIS System
• Create a test patient in the HIS/RIS system.
AX System
• Request the patient list from the HIS/RIS system.
• Select a test patient and make an exposure.
• Send the images to the Examworks+ for archiving.
ExamWorks+

NOTE The name of the buttons depends on the application lan-


guage selected in the configuration.

• Select <Acquire> on the touchscreen monitor for the workstation.


• Select the test patient.
• The current worklist will be acquired from the HIS/RIS system by pressing the <done>
button.
• Select <Target>, the archiving destination must be online.
• Select <Acquire>.
• Select <check worklist>.
- The patient will appear under “Procedures in Progress”.
- The status of the patient must be: “Matches found”.
• Press <Use Exist>, the test patient will appear in the “Modality Worklist Information”
field.
• Enter the dose information.
• Press <completed>.
- The images will be sent to the selected target.
- The MPPS will be sent to the HIS/RIS.
• There will be no acknowledgement, and the Examworks+ is now ready for a new trans-
fer.

Siemens AX41-000.814.01.02.02 Page 25 of 26 DICOM MPPS + SC


08.01 TD SD 21
26 Changes to Previous Version
6Changes to Previous Version 6-

all Chapters: Document expanded to include Axiom Artis and Polytron T.O.P.
all Chapters: Rewritten
Chapter 5: Added

DICOM MPPS + SC AX41-000.814.01.02.02 Page 26 of 26 Siemens