Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise
authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service pro-
cedures.
Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Remarks Regarding Startup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Required Aids and Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Required Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Product-specific Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Checking the System Emergency Off (if configured) . . . . . . . . . . . . . . . . . . . . . . . . 13
Testing the Ground Wires . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Power Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
System Voltage and Frequency Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Measuring the Power Line Voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Checking the Connection for Correct Phase. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Measuring the Internal Line Resistance (Power Quality) . . . . . . . . . . . . . . . . . . . . . 15
Checking for Inadmissible Ground Loops. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
For POLYDOROS LX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
For POLYDOROS SX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Switching on the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Function Check of the Emergency Stop Switches . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Checking the Rotational Field at the Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Service Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Connecting the Service PC to the POLYDOROS LX . . . . . . . . . . . . . . . . . . . . . . . . 20
Connecting the Service PC to the POLYDOROS SX . . . . . . . . . . . . . . . . . . . . . . . . 21
Installation of the XCS SSW and the ICONOS R100 SW on the Service PC . . . . . 21
Checking the Software Versions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Check and Adjustment of Basic Unit Data, Site Structure . . . . . . . . . . . . . . . . . . . . 22
Downloading the Control Console Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
List of IQ Documents. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Function Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Data Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Customer-specific Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Changing the Tilt End Position (+88 to + 90° Position) . . . . . . . . . . . . . . . . . . . . . . . 75
Only at the customer's request: Programming the Grid Movement . . . . . . . . . . . . . 75
Programmiong the Light Key on the Remote Control Console . . . . . . . . . . . . . . . . . 76
Changing the Fluoroscopy Field Limitation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Required Action . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Cover Panels, Covers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Check of the Ground Wires. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Other Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Clearing the Exposure Counter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Erasing the XCS Error Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Completing the Certificates, Reporting and Archiving . . . . . . . . . . . . . . . . . . . . . . . 80
Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
All tools, measuring equipment and aids, with the exception of the "standard installation
tool kit", are listed and specified in the Service Tools Catalogue.
Only calibrated test equipment may be used.
Required Documents 0
Notes 1.1
Safety Information 0
• To switch off power to all parts of the system (generator and system cabinet). set the
system switch in the OFF position.
• Install or remove components only with the generator switched off; when doing this,
observe ESD guidelines.
• ESD guidelines must be observed during startup.
WARNING With the generator switched off, there is still line power present at
the T1 transformer D160 switch-on circuit (see Wiring Diagram
X2206-10). After switching off the generator, there is still approx.
600 V DC voltage present for the inverter!
This is indicated by LEDs V35 and V36 on the D110 board (see also
X2206-16).
¹ The voltage dissipates within approx. 2 minutes to 0 V,
the LEDs go off at approx. 30 V. The warning note (label)
in the generator cabinet must be observed.
Product-specific Information 0
Illustrations
Depending the version, the illustrations and drawings can appear slightly different from
what has been shipped.
• All texts labeled with p must be entered in the Test Certificate that is listed in each
instance:
- Quality Certificate
- Test Certificate for Image Quality without DFR.
The tables in the corresponding certificates are marked with the note "Checked during
startup".
Manual Fluoroscopy
Setting appropriate X-ray tube load factors (kV or mA) to check the radiation safety mea-
surement values.
Necessary only if a radiation safety measurement must be performed.
• Start the POLYDOROS SSW:
- < Diagnostics >
- < Manual Fluoroscopy >
In this menu, the kV and mA can be set manually and fluoroscopy can be switched on.
General Information 0
NOTE These measurement values and the other values determined dur-
ing startup can be transcribed from the Test Certificate to the
Abnahmeprüfprotokoll or to the Acceptance Test Certificate.
Tolerance Data
General tolerances for linear dimensions according to ISO 2768
Limit values for over 3 mm over 6 mm over 30 over 120 over 400 over over
the nominal to 6 mm to 30 mm mm to 120 mm to 400 mm to 1000 mm 2000 mm
range mm mm 1000 mm to 2000 to 4000
mm mm
Allowable tol- ± 0,5 mm ± 1 mm ± 1,5 mm ± 2,5 mm ± 4 mm ± 6 mm ± 8 mm
erance
These tolerances apply to all dimensions given in these instructions, unless another toler-
ance is expressly indicated after the value.
Torque Values
A tolerance of ± 10 % is permitted for torque values.
OT = Overtable
SID = Film-Focus Distance
I.I. = Image Intensifier
DL = Fluoroscopy
SSW = Service Software
SPC = Service PC
BWS = Bucky wall stand
S = Sensitivity of a film-screen system
KS = Manufacturer specification of the dose requirement for a
film-screen system
RGr = Resolution power of the screen depending on S; KS
RG = optical resolution power
KT = Dose, dose rate measured on the tabletop
KB = Dose, dose rate measured in the image receptor plane
fK = Correction factor for dose detector
m = Equipment attenuation factor
D = Optical density for the film
DN = Optical net density (D - base fog)
SHa = Numerically displayed tomo height on the control console
BST = Exposure steps
EP = Exposure points
IQ = Image quality
XCS = X-ray Communication System
XCU = X-ray Control Unit
• When the system emergency off switch is pressed, power must be switched off to the
entire system.
The following ground wire check is used for the safety of the startup personnel.
The final ground wire check is performed and documented at the conclusion of the star-
tup.
Geraet / Unit
Monitorampel / Monitor rack
Monitore / Monitors
Schraenke / Cabinets
Trafos / Transformers
Pulte / Consoles
BV / I. I.
Roentgenroehre / X-ray tube unit
Tiefenblende / Collimator
Sonstige Komponenten / other Components
Generator
Netzverteiler / Power distbutor
NOTE After completing all adjustment work steps and check measure-
ments, the responsible technician must sign and date the certifi-
cates (system, generator and IQ test certificate for direct
exposures) to confirm that all values have been correctly deter-
mined and recorded.
Using the customer order list, check the connections and voltage / frequency toggles (side
of unit / generator side).
Generator Voltage / Frequency
• In systems with 440V or 480V line power, a pre-transformer must be installed in the
generator and connected per the Wiring Diagram.
• Check of the Frequency on the Generator D160
for 50 Hz --> jumpers X20-21; X23-X24
for 60 Hz --> jumpers X21-22; X24-X25
Unit Voltage
• Voltage on the T1 unit transformer:
for 400V: 1U1 --> 1W3; 1V1 --> 1U3; 1W1 --> 1V3
for 440V: 1U1 --> 1W2; 1V1 --> 1U2; 1W1 --> 1V2
• With 480V, a pre-transformer (M20) must be installed and connected. The voltages on
the unit transformer remain at 400V.
• To check the connection for correct phase, see the chapter "Switching on the System".
(Checking the Rotational Field at the Unit / p. 19)
Measure with the internal line resistance meter connected to M16.L1, L2, L3 (POLY-
DOROS Wiring Diagram) (one after the other, phase to phase):
• Connect the internal line resistance meter each time between two phases of L1, L2, L3.
• Switch the main system switch ON.
• Perform the measurement.
• Switch the main system switch OFF.
• Disconnect the test meter and perform the measurement for the other phases.
• Reinstall the M16 F1, F2 and F3 blade type fuses and switch on the trip breaker(s).
NOTE To achieve the full generator power, the measured internal line
resistance may not exceed the following values.
For POLYDOROS LX 0
For POLYDOROS SX 0
X12
CPU X5
X6 X121 X14
X60
X88
X9
X65
X7 X110 X77
X3
X64
D190
D160 X8 X10
D100
X63
X55
T1 LS
X62 X4
D165 PS
X88
X61 X45
X1
D111 X2 X44
D220
1
H1
KKI M15
EH111509-200202-00-01
• Press the EMERGENCY STOP switch on the remote control console and on the unit.
¹ The safety breaker(s) K1 (K2) is (are) audibly deenergized.
"Emergency stop" or "ERR 241/032 Iconos" is displayed on the remote control
console.
• Disengage the EMERGENCY STOP switches.
¹ The safety breaker(s) K1 (K2) is (are) audibly energized.
Option
D110
Touchscreen-
control console D115
D100
D200
D220
SN
H1
X5
D291
K20
D160
T2
X9 K30
F1 F2 F3
M16
ON
Com 1
Com 2
X5
SPC
Important!
The system controller (D200.X5)
must be connected to Com 1.
509-100100-0
D101
SNT
D190 D100
D160
F8
F7
T1 D165 LS
PS
Z1
D110 C3, C5
D115
DLS K5
D111 N93
K3
D220
H1
KKI M16
Fig. 5:
• Connect the Service PC as shown in the above illustration.
Installation of the XCS SSW and the ICONOS R100 SW on the Service PC 0
NOTICE All Iconos R100 systems have the TOMO function as standard.
The distinction between “with Tomo” or “w/o Tomo” is made later
by the settings in the Site Structure of the SSW.
¹ The software "with Tomo" should ALWAYS be installed.
NOTE The XCS SSW is located on the system software CD included in the
shipment. Following installation, check for the correct version
(per the Compatibility List).
• Rename the previously installed XCS SSW on the Service PC or delete it.
• Insert the current ICONOS R100 SW CD into the Service PC.
NOTE Options, e.g. Tomo, Compression or KermaX are not pre-set at the
factory and must be checked in the basic unit data according to
the customer order and, if needed, configured.
1. With the LX, do not configure the data printer in the Site Structure.
Workplace Configuration
Dose Q-factor
Detector Type
Iont. Channel
Fluoro avail.
# onwards
Iontomat
GNo.
WS
Fluoroscopy Configuration
Additional Settings
Site Options
Option group selection organ programs
Option selection available
Format-dependent correction
Adaption value [Exp. Points] 1.0
Spotfilm device
Eject cassette after last exposure selected
Collimator
Zoom 1 < Zoom 2: Set to full intensifier selected (not configurable for PL LX)
field
Configure the language of the control console per the country or customer requirements.
• Select "Components / Desk PLX" in the XCS SSW.
• Select the appropriate language in the "Desk language to be loaded" box.
• Download the selected language with <Load language>.
• Close the window <Done>.
• Switch the system OFF and back on.
509-100101-02
Room Dimensions 0
NOTE The safety distance between the unit and the wall of ≥ 50 cm
applies only for Germany.
Observe country-specific safety regulations.
The footrest is always included in the head-end and foot-end colli-
sion calculation.
A: Ceiling Height
• Measure the distance from the floor to the ceiling (Fig. 7 / p. 29).
NOTE The SSW automatically limits unit movement and takes attaching
the footrest at the head-end into consideration.
The safety distance to the ceiling must be at least 5 cm!
Patient Table 0
Compression Cabling 0
• Install the cable holder bracket (1/Fig. 9 / p. 32) on the spotfilm device with 3 Allen
screws.
• Lay the cable under the plate per the above illustration and secure it to the frame with
an adhesive-backed clip.(1/Fig. 9 / p. 32) When doing this, pay attention to the cable
lengths when in an oblique position.
• Connect the ground wire in the spotfilm device electronics.
• Lay the compression cable along with the I.I. cable through the appropriate zipper
hoses.(1/Fig. 10 / p. 32)
Starting Up Compression 0
Function Test
Check all unit movements for their function and for problem-free movement, i.e.:
• Perform a function check of the movements using the System User Instructions.
• No scraping noises.
• No hitting at the mechanical end stops.
• Confirm the message "release next exposure" with "OK" and switch exposure back ON.
• Repeat this routine until the "OK" radio button is activated.
• Confirm the menu with "OK".
Perform the Pre-Heating and Learn Filament settings for small and large focus; the menus
for both focuses will be displayed automatically.
Filament Push current:
• Confirm the message "will now release exposure at..." with "OK".
• Confirm the message "release next exposure" with "OK" after 3 s and switch exposure
back "ON".
• Repeat this routine until the "OK" radio button is activated.
• Confirm the menu with "OK".
• Confirm the following window with “OK”.
Control of Adjustment Parameters:
• Select the "Diagnostics / Adjustment result / Tube Adjustment Params" menu item.
• Check the values for the push factors; the table shows the permissible ranges.
OPTITOP Tube Unit and Idle Times of < 2 Weeks: Conditioning the X-ray Tube 0
NOTE If the tube tends to repeatedly and strongly "arc", abort the proce-
dure and, if needed, replace the tube unit (-> risk to sensitive
parts of the system electronics).
X • Trigger exposure and use the oscilloscope to measure the kV and mA.
• Calculate the max. power from the measured kV and mA:
• Recorde the measured mA value and the calculated power value (kW) in the "Quality
Certificate or Test Certificate 1”.
Correction 0
If the values are not within tolerance, the tube currents must be reduced; to do this:
• In the XCS SSW, select <Configure/ Site Structure/ Component selection/ Poly-
doros/ Option>.
EH111509-120302-00-01
NOTE The adjustment screws must be up against the brackets and the
counter nuts must be tightened.
X • Switch on fluoroscopy.
If the I.I. circle is not in the center of the blanking circle, the optics must be readjusted. To
do this,
• Loosen the screw on the I.I. optics.
X • Under fluoroscopy, move the optics until the center is reached.
• Readjust the eccentric bearing for the optics.
• Set switches S2.1, 2, 7 and 8 on the D111 camera board back to OFF.
• Press the S1 reset button.
Camera Rotation 0
NOTE The D111 board can slip out of the camera mount by loosening
the clamp on the camera mount! There may be no gap between
the board and the camera mount.
Central Beam 0
Determining the Drift of the Central Beam between SID min. and SID max.
NOTE If only one SID (115 cm) is available, set the center of the radiation
field to the center of the spotfilm device.
Head-end/foot-end ≤2 mm
direction:
Operator's side / ≤2 mm
wall side direction:
Loosen
4 screws
Adjustment
• Following adjustment, carefully tighten the four tube unit mounting screws; when doing
this, use the light localizer to make sure that the tube unit does not come out of adjust-
ment.
X • Perform a check under fluoroscopy and if needed, repeat the procedure.
Preparation
• Use the centering button on the unit console to center the column.
• Move the unit into the 0° position.
• Set an SID of 115 cm.
Serviceabdeckung
Service cover
Kunststoffabdeckung
Plastic cover
EH111509-110402-00-01
• Adjust the collimator so that the plates are visible on the monitor.
Blanking circle
head end
operator side
wall side
Center of radiation field foot end
EH111509-310102-02-04
Fig. 16: Determining center of radiation field / center of spot film device
• Determine the difference per the above illustration.
¹ Maximum admissible difference: ≤3 mm
Adjusting the Tube Unit in the Operator's Side/Wall Side Direction (only if neces-
sary)
• Switch on the collimator light localizer.
• Mark the position of the light field on the tabletop in the head-end/foot-end directions.
• Correct the position of the tube unit by moving it in the operator's side/wall side direc-
tion.
• Tighten one of the two set screws again and when doing so make sure that the position
in the head-end/foot-end directions does not change (watch the mark on the tabletop).
• Loosen the clamping screw on the top part of the tomo rod and align it so it is vertical
using a spirit level.
X • If needed, repeat the procedure using control measurements with fluoroscopy.
• If the value is within tolerance, tighten both set screws.
T • Tighten the screw on the tomo rod to a torque of 50 Nm.
Fig. 18:
• Loosen the 4 screws (1/Fig. 18 / p. 45) on the tube unit mount.
• Move the tube unit in the transverse direction.
• Retighten all 4 screws.
X • Repeat the measurements under fluoroscopy.
• If a dose measurement chamber with centering cross was installed in the collimator at
the customer's location, the 4 Allen-head mounting screws must be loosened and the
centering cross adjusted so it is centered to the collimator blanking.
NOTE Centering of the TV camera and the fluoro field limitation has
been adjusted at the factory and is documented in the "Quality
Certificate". Make sure that the setting has not changed during
transport and installation.
• Move the unit into the 35° position (for units with the lifting base, move into the 0° posi-
tion).
• Set the max. SID.
• Select image intensifier full format.
X • Switch fluoroscopy ON
• Open the collimator all the way.
NOTE To ensure that the settings have not changed during system
transport, test exposures must be made and compared to the
exposures that are included in the shipment. If these test expo-
sures are within the listed tolerances, the factory set values
should not be changed.
Cassette Position 0
Preparation
• Use the centering button on the unit console to center the column.
• Move the unit into the 0° position.
• Set an SID of 115 cm.
• Place on the centering template (Fig. 15 / p. 43).
Film
Serviceabdeckung
Service cover
382 mm
Fig. 20:
• Place the marked film on the spotfilm device so that it is flush at the head end with the
outside edge of the spotfilm device and lines up with the tableside controller on the
operator's side (Fig. 20 / p. 47).
X • Under fluoroscopy, check whether the center mark on the marked film lines up with the
mark on the template.
Correction: If needed, correctly position the template or slightly bend the template.
Perform the check again!
Lichtvisier
Light localizer
• Measure the dimensions “c” and “d”: These values should be identical.
c and d = max. admissible difference≤2 mm
If there are differences that are outside of the tolerance:
Correction: Mechanically adjust the symmetry of the collimator plates.
See the instructions, Replacement of Parts”, Spotfilm Device
• Measure the dimension “e”.
e = 72 mm, max. admissible difference ± 1 mm
If there are differences that are outside of the tolerance:
Correction: Collimator plate opening
• Insert the cassette.
Adjust using the service software.
Adjustment/ Calibration/ Spotfilm Device/ Diaphragm plates: Basic setting
Default value approx. 85; a lower figure means that "e" becomes smaller.
• In the range “f”, the distance between the center markers is equal.
f = max. admissible difference ± 5 mm
If there are differences that are outside of the tolerance:
Mechanical cause: The entire tensioning system is not symmetrical.
Check the tensioning drive mechanism.
• In the range “g”, the distance between the center markers is equal.
g = max. admissible difference 3 mm ± 2mm
If there are differences that are outside of the tolerance:
Mechanical cause: The entire tensioning system is not symmetrical.
Inform the process planner.
• In the range “h”, the distance between the center markers is equal.
h = max. admissible difference 3 mm ± 2 mm
If there are differences that are outside of the tolerance:
Mechanical cause: The entire tensioning system is not symmetrical.
Inform the process planner.
Fig. 23:
• Insert a 24 x30 cm or 10 x 12 inch cassette sideways.
• Select the direct technique, full format and semi-automatic lengthwise and trans-
versely.
• Set the max. SID.
• Move the unit into the 35° position (for units with the lifting base, move into the 0° posi-
tion).
• Manually set a format of approx.18 x 24 cm (display on collimator).
• Place a side marker on the tabletop.
X • Release an exposure.
• Develop the films and label them (exposure data, date).
Evaluation:
• Draw in the center of the radiation field and the center of the film as shown in the illustra-
tion.
• Measure the difference between the radiation field center and film center.
¹ Max. admissible difference ≤8 mm
Correction:
• Check the centering of the tube unit/collimator to the center of the spotfilm device.
The accuracy of automatic format collimation is checked for two formats with minimum
and (optional) maximum SID.
NOTE The upper tolerance limit (0.8% … 1.2%) has to be utilized for the
image quality of the subdivided exposures.
Preparation
• Set the unit out of the 0° position.
• Set the SID to 115 cm; set the focus-image intensifier distance rB-min = 115 cm.
• Select the ICONOS Spotfilm workstation.
• Place the 35 x 35cm cassette with film (or film envelope) in the center of the tabletop.
• Enter a preferred value (see adjustment).
Procedure
• Insert the following cassettes (without film) and release an exposure with the indicated
exposure data, one after the other:
18 x 24 cm cassette lengthwise; 44 kV; 3.20 mAs (70 kV; 40 mAs)
24 x 30 cm cassette lengthwise; 44 kV; 1.40 mAs (70 kV; 50 mAs)
• Set the SID to 150 cm (optional); set the focus-image intensifier distance rB-min = 150
cm.
• Place the 35 x 35cm cassette with film (or film envelope) in the center of the tabletop.
• Insert the following cassettes (without film) and release an exposure with the indicated
exposure data, one after the other:
18 x 24 cm cassette lengthwise; 44 kV; 4,50 mAs (70 kV; 40 mAs)
24 x 30 cm cassette lengthwise; 44 kV; 3.20 mAs (70 kV; 50 mAs)
• Develop the film, measure the radiation field width (b’) and length (l’) and record the val-
ues.
2
1
Fig. 24:
Pos. 1 The 2 format sizes to be checked
Pos. 2 Film
Evaluation
• Measure the width (b) and length (l) of the radiation fields in the image receptor plane
and record the values.
b = b’ x M; l = l’ x M; calculate the magnification factor M according to the formula
(Fig. 25 / p. 53) separately for min. and max. SID.
“a“ = distance between the measuring plane and the film plane.
“rb“ = SID in cm.
r
Magnification factor M = r ba
b
a = 8 cm
rB
Front panel
a
Cassette to be checked in film plane
(catapult bucky)
Fig. 25:
• Determine the percent difference between the radiation field (b) and the nominmal for-
mat (bN) and record this value.
Width (%) = [(b - bN)/rB] x 100
Length (%) = [(l - lN)/rB] x 100
• maximum permissible difference for width and length in %: ≤1,2%
P • Enter the value in the Quality Certificate or Test Certificate 1.
Adjustment
NOTE If there is a difference, perform the adjusment with only one SID
(e.g. 115 cm).
• Enter the overframing of the film format for the width and height in mm (0 ... 25mm);
always enter the same correction value for height and width. The value of e.g. 10 mm
entered corresponds to the displayed figure of 100.
An empirical value = 90
• Confirm with OK.
• Afterwards a window with the old and the new value is shown.
• Confirm by clicking on OK. The new values are now stored only in the unit.
• Exit the XCS SSW and switch the system off and on again to store the change perma-
nently in the XCU.
• After the adjustment, repeat the steps described under preparation, procedure and
evaluation.
Fig. 26:
• Select "Bucky Wall Stand (BWS)”.
• Raise the unit into the 90° position or tilt it down in the direction of the Bucky wall stand.
• Set the unit to the min. SID (max. SID to the BWS).
• Adjust the tube unit so that it is centered to the BWS cassette using the light localizer.
• Mark the center of the film with a washer.
• Insert a 24x30 cm or 10x12 inch cassette sideways into the Bucky wall stand.
• Manually set a format of 18 x 24 cm (display on collimator).
• Place a side marker on the Bucky wall stand.
• Release an exposure.
Exposure parameters (e.g. for a 200 screen system): 42 kV, 8 mAs, 80% kW, large
focus.
• Develop the films and label them (exposure data, date).
Evaluation
• Mark the center of the radiation field and the center of the film (Fig. 26 / p. 55).
• Measure the difference between the radiation field center and film center.
¹ Max. admissible difference ≤1.2% of the SID
P • Enter the difference from the center in the Quality Certificate or Test Certificate 1.
• Label and archive the film.
Correction:
• If there is a difference in height:
The tube unit has not been correctly set to the center of the Bucky wall stand by the
operator. Repeat the exposure.
• If there is a difference in the transverse direction:
Correct the BWS position (move laterally).
The following values can be entered or transcribed from the data sheet for the DLR cas-
settes into Test Certificate 1 or the "Quality Certificate” only after discussing this with the
manufacturer.
• Screen Types and Film Manufacturers
• Color Sensitivity
• Sensitivity, S
• Dose Requirement, KS
• Minimum resolution
P Enter the values into Test Certificate 1.
IONTOMAT Adjustment 0
Unlike with normal film-screen cassettes, the IONTOMAT can be adjusted only with the
cutoff dose in DLR systems.
IONTOMAT Sensitivity 0
NOTE The data can be updated in the XCU with "Transfer Value" and
control exposures can be made without having to exit the SSW.
Enter the following values in XCS SSW program "Competent/ Polydoros/ Adjustment/
Voltage Response Correction”.
Three different exposure steps (BSt) corresponding to three different film-screen combi-
nations (H, U, D) can be programmed at the generator.
For each exposure step, the screen and the film provided for it must be documented with
the following data from the manufacturer in the "Quality Certificate":
- Color sensitivity of the films and screens
- Sensitivity, S
- Theoretical dose requirement, KS
- Minimum resolution, RGr
Make sure that only films for the light color for which the screen is sensitivity has been
determined are used (green sensitive film for green sensitive screen, blue sensitive film for
the blue screen).
There are no international, uniform regulations regarding use of the film-screen combina-
tions; national regulations (e.g §16 RöV) must be observed in the particular country.
Screens and Film Manufacturers/Types
P Enter the value in the "Quality Certificate”.
Color Sensitivity of the Films and Screens
P Enter the value in the "Quality Certificate”.
Sensitivity, S
P The sensitivity, S (speed), specified by the manufacturer must be recorded for each
film-screen combination.
As a rule (in some countries, this is specified), this is the S value on the cassettes (e.g. S
= 200, S = 400) or it must be taken from the data sheets for the film-screen combinations.
The sensitivity of a film-screen combination is defined according to ISO 9236 as the quo-
tient of 1000 µGy and the air kerma (dose) KS required to achieve an optical density
(blackening) of 1 above base fog.
1000
S = ------------ µGy
KS
Fig. 27:
Dose Requirement, KS
P With a known sensitivity, S, calculate and document the theoretical dose requirement KS.
1000
K S= ------------ µGy
S
Fig. 28:
Minimum Resolution, RGr
A minimum resolution RGr in LP/mm is assigned to each required dose KS, i.e. the higher
the required dose is for a film-screen combination, the higher the resolution that must be
reached in order to justify the increased dose.
P The particular minimum resolution, RGr must be taken from the illustration for each
film-screen combination used and must be documented.
Fig. 29:
Screen Sensitivity
(H) 800 8
(U) 400 11
(D) 200 14
NOTE The data can be updated in the XCU with "Transfer Value" and
control exposures can be made without having to exit the SSW.
NOTE If the specified values are not reached, drift and hum of the Ion-
tomat chamber must be checked.
NOTE The generator must have been in operation for at least 5 minutes
before the test is performed.
• In the XCS SSW program, select "Components/ Polydoros LX or SX/ Diagnostic/ Ion-
tomat Drift".
• Perform the drift measurement for the middle measuring field per the information in the
SSW.
Drift: 0, ± 2 V/0.5 s
example 1
Oscillogram 1:
Drift on
example 2 D100.X63.VION
Toleranz
1,5 V +/- 2V 1 V/T
Fig. 30:
• Close the window with "OK" and exit the XCS SSW.
• Remove the lead and switch the I.I. power supply back on.
P • Enter the value in the Quality Certificate or Test Certificate 1.
Oscillogram 2:
100 ms/T
Fig. 31:
P • Enter the value in the Quality Certificate or Test Certificate 1.
NOTE The data can be updated in the XCU with "Transfer Value" and con-
trol exposures can be made without having to exit the SSW.
X • Make test exposures at 60 kV and 125 kV using 20 cm, ± 1 cm water for each screen
used and check the film density of each exposure.
- The film density at 60kV and 125kV must match the film density at 81kV (Section
"IONTOMAT Sensitivity").
- The difference in density can be
Dopt (81 kV) - Dopt (60 kV) ≤ 0.2
Dopt (81 kV) - Dopt (125 kV) ≤ 0.2
of the optical density.
- If necessary, select other correction curves for the upper or lower kV range.
• If a Vertix PRO (option) is configured, the adjustment must also be performed for this
unit.
P • Enter the measured values in the "Quality Certificate or Test Certificate 1".
• Exit the SSW screen with "OK".
Correction curves 1...13 for the lower kV range (40...81 kV) and correction curves 1...13
for the upper kV range (81...150 kV) are displayed in the following illustration and the cor-
responding value table.
From this it is possible to read out what change in exposure points or in density results
when there is a switch to a different correction curve.
Fig. 32:
EP = exposure points
∆EP are listed in the table in steps of 1/8 EP.
The cut-off dose must be measured at each application unit for IONTOMAT P for all expo-
sure steps (screen) used.
Record in Test Certificate 1. The values can then be transcribed for the §16 RöV Accep-
tance Test.
Test Conditions:
• Select the workstation:
- ICONOS Spotfilm, Vertix PRO or 3D -Vertix PRO
• Set a 115 cm SID for "ICONOS Spotfilm".
Set the corresponding SID for the Vertix PRO.
P Record the SID that is set (Quality Certificate or Test Certificate 1).
• Move the scatter radiation grid into the beam path.
• Set the density correction to ± 0 exposure points.
• Do not select any additional filtering at the collimator.
• Test object:
• Exposure parameters:
77 kV, small focus, 80%, exposure step H, U or D, middle Iontomat measuring field.
Record the kV value (Test Certificate 1).
Procedure:
• Connect the mAs meter
• Dali, Nomex:
Place the dose measurement chamber on the tabletop, centered to the middle Ion-
tomat measuring field.
Diados:
Place the dose detector on the tabletop, outside the middle Iontomat measuring field.
Set the filter on the Diados to 23.5 mm Al.
• Place resolution test, Type 42 on the tabletop, turned 45° to the patient axis, outside the
middle Iontomat measuring field.
X • Trigger an exposure for all H/U/D exposure steps and application units (spotfilm device,
Vertix PRO), trigger exposure and perform the following measurements:
- Measure the cutoff dose KT.
- Determine the resolution, Rg of the film.
Fig. 33:
• Resolution, Rg
¹ Rg must be ≥ RGr for each film-screen combination.
NOTE If the minimum resolution, RGr is not reached, check the screen
for damage and for soiling and the cassette for correct flattening
of the screen.
P • Record all values that have been determined in the "Quality Certificate or Test Certifi-
cate 1".
• Label and archive the films.
• Exposure parameters:
77 kV, small focus, 80%, exposure step D
X • Make an exposure for all application units (spotfilm device, Vertix PRO) using all three
Iontomat measuring fields.
• Determine the mAs value.
∆ mAs left measuring field = mAs middle measuring field mAs left measuring field
∆mAsright measuring field = mAsmiddle measuring field mAsright measuring field
¹ Tolerance: ∆mAs ≤25%
P • Record the values in the "Quality Certificate or Test Certificate 1".
• Label and archive the films.
NOTE For room heights of < 285 cm, Tomo can be triggered only in unit
positions between - 5° to + 5° and + 87° to + 90°.
It is not generally possible to perform tomography under -5°.
Evaluation
¹ Tomo height
Tomo height display on the console, SHa - tomo height on the film, SH = ∆
SHa (≤ 5 mm)
¹ Resolution of the Tomo Exposure
Difference between the Bucky and tomo exposure ≤ 1 line group
Resolution limit value for the tomo exposure ≥ 2.0 Lp/mm
Correction
• Start the XCS SSW.
• Start Components -> ICONOS R100 -> Adjustment -> Calibrations .
• Select the Tomo -> Tomo height console display function.
- Default value in the "Old offset value" box = 90.
• Enter a value for the correction.
- 1 mm corresponds to the displayed number 10; if the tomo height is too high, correct
with “+”, if the tomo height is too low, correct with “-”.
• Confirm with “OK”.
¹ A window with the new and the old values will be displayed.
• Confirm with “OK”.
¹ The value will be saved.
• Exit the XCS SSW.
• Switch the system off and back on.
X • Make a new tomo exposure.
• Remove the 0.6 mm Cu from the collimator.
P • Enter the values determined into Test Certificate 1.
• Label and archive the films.
NOTE The monitor values must be checked and, if needed, set and
adjusted usin the IQ Test Procedure.
Perform monitor adjustment and the image quality test according to the following docu-
ments:
• AXIOM Iconos R100, Quality Assurance, Test Procedure for Image Quality with-
out DFR, AXD3-350.820.01.01.xx
• AXIOM Iconos R100, Quality Assurance, Test Procedure for Image Quality with-
out DFR, AXD3-350.820.02.01.xx
P • Fill out the general data in the IQ Test Certificate.
P • Enter the System Serial Number on each page of the IQ Test Certificate to ensure there
is clear allocation of each certificate page.
• For detailed test descriptions, see the IQ Test Instructions.
NOTE If a KermaX dose measurement meter was retrofitted (per the cus-
tomer order), startup per the document: AX00-100.814.01.01.xx,
(AX00-100.814.01 / KermaX with external Display) must be per-
formed.
Function Check 0
Fluoroscopy
• Select the tube unit.
• At the generator console, reset the cGy/cm2 display with the "0 set button”.
• Insert 2.1 mm Cu.
• Completely open the collimator.
• Set the max. SID.
X • Trigger fluoro for 30 sec.
• The value of the area dose product (cGy/cm2) displayed on the console or monitor must
increase continuously.
• Remove the 2.1 Cu from the beam path.
Exposure
• Select the tube unit.
• At the generator console, reset the cGy/cm2 display with the "0 set button”.
• Set 70 kV and 10 mAs.
X • Release an exposure.
• Read and make a note of the area dose product (cGy/cm2) displayed on the console.
• Set 70 kV and 5 mAs.
X • Release an exposure.
¹ The value read on the console must be half of the above value.
• Set 70 kV and 20 mAs.
X • Release an exposure.
¹ The value read on the console must be double the above value.
Two printer models can be connected. EPSON LQ 570 and Star DP8340S.
For the Star DP8340S printer model, the DIP switches must be set as follows:
Fig. 35:
Polydoros SX:
• In "Configure:Site structure:Component selection", integrate the priner connection in
the system list with "Data printer com2" .
• Select the connected printer under "printer type" (e.g. Star DP8340S).
• Select when the data are to be printed in "print selection" (the print routine must be trig-
gered using the RESET button for the following check).
• Switch on the data printer and insert paper.
X • Trigger fluoro for 5 sec.
• Make a note of the kV, mA, sec. values displayed on the generator console.
• Activate the printer with the Reset button.
• Compare the values on the label with the values noted from the generator console.
X • Release an exposure.
• Compare the values on the label with the values displayed on the generator console.
• Make sure that both labels are printed in the vertical direction and are in the middle.
• For further descriptions, see the documents included in the data printer shipment.
Polydoros LX:
NOTE The data printer is connected directly to the external KermaX disi-
play and does NOT have to be configured in the Site Structure.
Only dose values are printed out.
NOTE If tomo is configured, the unit stop position may not be set < 88°,
otherwise tomo will be blocked with room heights of < 2.85 m or it
will not be possible to lower the tabletop.
Correction:
• In the XCS SSW,
- click on < Components >
- < ICONOS >
.
• In the ICONOS SSW,
- click on < Adjustment >
- < Calibrations >
- < Table tilt >
- < limitation of + 90° position >
- < Execute >
.
Additional steps are menu-prompted.
Depending on the customer's request: Switching on the collimator light localizer is also
possible from the ICONOS remote control console (light key). Switching off is not possi-
ble.
• To switch on the light localizer: In the ICONOS remote control console, move the
jumper on the D41 from XP4 to XP12
or
• To switch on the room light: In the ICONOS remote control console, move the jumper
on the D41 from XP4 to XP12.
P • If the collimator plates are adjusted larger at the customer's request, determine the
amount of over-framing and record this in Quality Certificate 1.
Workflow
• Unit in 0°position. Set an SID of 115 cm Select I.I. full format.
• Place the centering cross on the tabletop; when doing this, set the X axis in the tabletop
transverse direction.
• Switch fluoroscopy ON. Collimate a radiation field of approx. 4 x 4 cm and position the
centering cross so it is centered to the radiation field.
• Open the collimator all the way. Starting from the center of the centering cross, going in
the width and length directions, read the cm division of the four X1, X2, Y1, Y2 coordi-
nates displayed on the monitor and record the values. Perform this for all I.I. formats
and all SIDs.
• In pace of the centering cross, place a cassette with film (at least 35 x 35 cm or 14 x 14")
turned 45° on the tabletop and center it to the light field. Trigger fluoroscopy with the col-
limator open all the way for all I.I. formats until sufficient film density is achieved. Make a
test film for each SID, whereby all 4 I.I. formats are visible. Evaluation is possible
because of the different film blackening of the individual formats. Make a test film for
each SID.
• Measure the width (X) and length (Y) of the radiation fields on the test films and record
the values.
Admissible Overframing of the I.I. Formats in the Centering Cross Plane:
SID 115 cm = 2.1 cm
SID 150 cm = 2.8 cm
Fig. 36:
• Install all covers and cover panels, particularly the lateral I.I. cover panel
(1/Fig. 37 / p. 78).
Unit
Table
Ceiling rails
Monitor rackl
Less than 0.2 Ohm Monitors
Cabinets
Transformers
Consoles
I.I.
Generator / X-ray tube unit
Power Ground wire test instrument
distributor
Collimator
Fig. 38:
• Power to the system must be switched off.
• Measuring device:
- Ground wire meter
• Measurement value:
- Max. 0.2 Ohm (observe country-specific regulations!)
Exceptions: Accessory rails on the collimator, tabletop and control console.
• Test procedure:
- Measure between all exposed conductive parts of the system and the ground wire
bar in the generator.
• Recording:
P - Enter the measurement value into the "Quality Certificate".
Other Adjustments 0
• Enter the time and date in the XCU in the menu System /Realtime Clock.
NOTE After completing all adjustment work steps and check measure-
ments, the responsible technician must sign and date the certifi-
cates (System, Unit Certificates, Generator Certificate, IQAP,
Installation Certificate, Startup Certificate) and the country-spe-
cific certificates (e.g. DHHS, RöV §16) to confirm that all values
have been correctly determined and recorded.
SIEMENS Shanghai
Medical Equipment Ltd.
278 Jin Ha Rd. Shanghai
PRC, 201206
Labels 11.1
Training Requirements
Due to significant changes in this document / new publication of this document, user training is
required.
Due to slight changes in this document, user training is not required.