AXIOM Iconos R100


AX

Startup
System

System Shipments beginning 05/2007

Valid beginning with System Part Number: 10272401


10272401
05895151

© Siemens AG 2006
The reproduction, transmission or use
of this document or its contents is not
permitted without express written
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.

Print No.: AXD3-350.815.04.03.02 English


Replaces: AXD3-350.815.04.02.02 Doc. Gen. Date: 05.07
2 Revision / Disclaimer
1Revision / Disclaimer

Document revision level


The document corresponds to the version/revision level effective at the time of system
delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.

Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise
authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service pro-
cedures.

AXIOM Iconos R100 AXD3-350.815.04.03.02 Page 2 of 82 Siemens AG


05.07 AX SCM W II Medical Solutions
Table of Contents 3
0Table of Contents

1 _______ General Information______________________________________________ 6

Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Remarks Regarding Startup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Required Aids and Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Required Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Product-specific Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

2 _______ General Startup ________________________________________________ 13

Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Checking the System Emergency Off (if configured) . . . . . . . . . . . . . . . . . . . . . . . . 13
Testing the Ground Wires . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Power Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
System Voltage and Frequency Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Measuring the Power Line Voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Checking the Connection for Correct Phase. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Measuring the Internal Line Resistance (Power Quality) . . . . . . . . . . . . . . . . . . . . . 15
Checking for Inadmissible Ground Loops. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
For POLYDOROS LX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
For POLYDOROS SX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Switching on the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Function Check of the Emergency Stop Switches . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Checking the Rotational Field at the Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

3 _______ Service Software _______________________________________________ 20

Service Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Connecting the Service PC to the POLYDOROS LX . . . . . . . . . . . . . . . . . . . . . . . . 20
Connecting the Service PC to the POLYDOROS SX . . . . . . . . . . . . . . . . . . . . . . . . 21
Installation of the XCS SSW and the ICONOS R100 SW on the Service PC . . . . . 21
Checking the Software Versions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Check and Adjustment of Basic Unit Data, Site Structure . . . . . . . . . . . . . . . . . . . . 22
Downloading the Control Console Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27

4 _______ System Startup ________________________________________________ 28

Starting Up the Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28


Checking the Safety Limit Switches . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Room Dimensions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Patient Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Installing and Starting Up Compression (Option) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Compression Cabling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Starting Up Compression . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Unit Movements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Starting Up the Generator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Checking the Generator without High Voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36

Siemens AG AXD3-350.815.04.03.02 Page 3 of 82 AXIOM Iconos R100


Medical Solutions 05.07 AX SCM W II
4 Table of Contents

Checking the Generator with High Voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36


OPTITOP Tube Unit and Idle Times of < 2 Weeks: Conditioning the X-ray Tube . . . 37
Fluoroscopy Attachment (only in Combination with Polydoros SX) . . . . . . . . . . . . . 38
Check of the max. Generator Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Correction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39

5 _______ Beam Geometry ________________________________________________ 40

I.I. / TV and Central Beam. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40


Checking the Collimator Position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Determining the Center of the Spotfilm Device / Center of Monitor. . . . . . . . . . . . . . 40
Camera Rotation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Central Beam. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Adjusting the Centering Cross (Dose Measurement Chamber) . . . . . . . . . . . . . . . . 45
Checking the fluoro field limitation and centering . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Checking the ICONOS Spotfilm Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Cassette Position. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Coincidence of Center of Radiation Field and Center of Film . . . . . . . . . . . . . . . . . . 51
Accuracy of Automatic Format Collimation (reference to nominal formats). . . . . . . . 51
ICONOS Tube Unit on the Bucky Wall Stand (Option) . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Coincidence of Radiation Field and Center of Film . . . . . . . . . . . . . . . . . . . . . . . . . . 55

6 _______ IONTOMAT ____________________________________________________ 57

Procedure only for Systems in which DLR Cassettes are Used . . . . . . . . . . . . . . . . . . . 57


IONTOMAT Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
IONTOMAT Sensitivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
IONTOMAT Voltage Response Correction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Plani-IONTOMAT(for DLR cassettes) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Cutoff Dose and Resolution. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Function of the Measurement Fields . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Film-Screen Combinations Used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
IONTOMAT Sensitivity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Checking the Drift (only in a malfunction situation) . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Checking the Hum Voltage (only in a malfunction situation) . . . . . . . . . . . . . . . . . . . 63
IONTOMAT Voltage Response Correction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Cutoff Dose and Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Function of the Measurement Fields. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69

7 _______ Tomo (Direct Technique) _________________________________________ 70

Checking the Tomo Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70


Tomo Height / Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70

8 _______ VIDEOMED DIC_________________________________________________ 72

List of IQ Documents. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72

9 _______ Startup of Other Options_________________________________________ 73

Starting up the KermaX Dose Measurement Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73

AXIOM Iconos R100 AXD3-350.815.04.03.02 Page 4 of 82 Siemens AG


05.07 AX SCM W II Medical Solutions
Table of Contents 5

Function Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Data Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74

10 ______ Customer-specific Settings ______________________________________ 75

Customer-specific Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Changing the Tilt End Position (+88 to + 90° Position) . . . . . . . . . . . . . . . . . . . . . . . 75
Only at the customer's request: Programming the Grid Movement . . . . . . . . . . . . . 75
Programmiong the Light Key on the Remote Control Console . . . . . . . . . . . . . . . . . 76
Changing the Fluoroscopy Field Limitation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76

11 ______ Final Work Steps _______________________________________________ 78

Required Action . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Cover Panels, Covers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Check of the Ground Wires. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Other Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Clearing the Exposure Counter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Erasing the XCS Error Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Completing the Certificates, Reporting and Archiving . . . . . . . . . . . . . . . . . . . . . . . 80
Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81

12 ______ Changes to Previous Version _____________________________________ 82

Siemens AG AXD3-350.815.04.03.02 Page 5 of 82 AXIOM Iconos R100


Medical Solutions 05.07 AX SCM W II
6 General Information
1-
Requirements
1General Information

Remarks Regarding Startup 0

The system must be completely cabled.

Required Aids and Tools 0

All tools, measuring equipment and aids, with the exception of the "standard installation
tool kit", are listed and specified in the Service Tools Catalogue.
Only calibrated test equipment may be used.

• Standard installation tools n.a.


• Pliers for zipper hose n.a.
• Water pail ( > 5 ltr) and measuring cup n.a.
• Service PC per the specifications of ARTD-001.719.06.... n.a.
• Service cable, 5 m 99 00 440 RE 999
• Digital multimeter, e.g. Fluke 8060 A 97 02 101 Y4290
• Field rotation meter 97 02 713 Y7933
• Internal line resistance meter 84 28 104 RE999
• mAs meter 81 60 400 RE999
• Precision radiation filter (2.1 mm Cu) 99 00 598 XE999
• Set of Cu filters (10 pieces, each 0.3 mm Cu) 44 06 120 RV090
• Centering cross 96 60 051 RE999
• Ground wire test meter 44 15 899 RV090
• ESD equipment, Type 8501 - 3M 97 02 606 Y3121
• Oscilloscope, 100 MHz
• Resolution test, type 41/42 28 71 820 RE999
• Focus test for tomo (tomo block) 44 06 054 RV090
• Aperture diaphragm 72 52 067 G2080
• 17 µm Cu strips 11 67 662 G5247
• PTW DIADOS 97 17 612 Y0388
Required only where DHHS Regulations apply for the mea-
surement of skin dose rate.
(otherwise not needed for checks during startup, only if there is
a difference to required values and a new adjustment is
required).

AXIOM Iconos R100 AXD3-350.815.04.03.02 Page 6 of 82 Siemens AG


05.07 AX SCM W II Medical Solutions
General Information 7

• Mavo monitor 97 02 432 Y0526


(not required for checks, only if there is a difference to required
values and a new adjustment is required)
• Densitometer, e.g. X-Rite 331 97 02 416 Y1996
• Sensitometer, e.g. X-Rite 333 97 02 424 Y1996
• Magnifier (8x) 44 14 850 RH090

Required Documents 0

• AXIOM Iconos Operating Instructions


• POLYDOROS Operating Instructions
• AXIOM Iconos System Wiring Diagram
• POLYDOROS Generator Wiring Diagram
• AXIOM Iconos R100 Quality Assurance
(Test Procedure for Image Quality without DFR)
• AXIOM Iconos R100 Quality Assurance
(Test Certificate for Image Quality without DFR)
• AXIOM Iconos Quality Certificate and Test Certificate 1

Siemens AG AXD3-350.815.04.03.02 Page 7 of 82 AXIOM Iconos R100


Medical Solutions 05.07 AX SCM W II
8 General Information

Notes 1.1

Safety Information 0

General Safety Information


When carrying out the work steps and checks, the product-specific safety information con-
tained in the documents must be observed.

CAUTION Dangerous radiation!


¹ Checks or adjustments that must be performed with radi-
ation switched on are labeled with the radiation warning
symbol x . While performing adjustments labeled this
way, radiation protective clothing must be worn.

WARNING Dangerous voltage!


¹ As a rule, measurements made using the oscilloscope
may be performed only with the ground wire connected.
Disconnecting the ground wire without an isolation
transformer is not permitted under any circumstances.
The ground wire may be made ineffective by means of an
isolation transformer only if no voltages greater than
25 V AC or 60 V DC can occur on metal parts that can be
touched. For measurements in which any ground loops
can have a disturbing effect on the measurement result,
an isolation amplifier (e.g. GOULD Contec 9000 or TEK-
TRONICS A6902B) must be used, see the Service Tools
Catalogue.

• To switch off power to all parts of the system (generator and system cabinet). set the
system switch in the OFF position.
• Install or remove components only with the generator switched off; when doing this,
observe ESD guidelines.
• ESD guidelines must be observed during startup.

Product-specific Safety Information


• Prior to performing any activity in the generator, switch it off at the S2 power OFF switch
on the D160 board.

AXIOM Iconos R100 AXD3-350.815.04.03.02 Page 8 of 82 Siemens AG


05.07 AX SCM W II Medical Solutions
General Information 9

WARNING With the generator switched off, there is still line power present at
the T1 transformer D160 switch-on circuit (see Wiring Diagram
X2206-10). After switching off the generator, there is still approx.
600 V DC voltage present for the inverter!
This is indicated by LEDs V35 and V36 on the D110 board (see also
X2206-16).
¹ The voltage dissipates within approx. 2 minutes to 0 V,
the LEDs go off at approx. 30 V. The warning note (label)
in the generator cabinet must be observed.

WARNING Frequency Converter


After disconnecting the drive current converter in front of the
power supply, the frequency converter voltage-conducting unit
parts and power connections may not be touched immediately
because of capacitors that may possible still contain a charge.
Because of the intermediate circuit capacitors, there is still dan-
gerous voltage present at the frequency converter for up to
30 seconds after switching off power.
¹ Working on the motors or on the intermediate circuit ter-
minals of the frequency converter is permitted no sooner
than after this waiting period.

• Switching Off High Voltage/Radiation


- To prevent unintentional triggering of high voltage or radiation, set the SS switch (S1)
on the D100 to OFF (no control of the inverter).
• Preventing Unit Movements
- Press the unit emergency stop switch when performing all service and maintenance
work on the unit. The emergency stop switch is located on the front of the unit and on
the remote control console.

Product-specific Information 0

Illustrations
Depending the version, the illustrations and drawings can appear slightly different from
what has been shipped.

Remarks regarding Startup


The system was prestaged, programmed and tested as a system at the factory. The Test
Certificates are filed in the binder "Techn. Documents / System Reg. 9".
• During startup, it must be assured by means of checks or measurements that the test
area settings have not changed.

Siemens AG AXD3-350.815.04.03.02 Page 9 of 82 AXIOM Iconos R100


Medical Solutions 05.07 AX SCM W II
10 General Information

• All texts labeled with p must be entered in the Test Certificate that is listed in each
instance:
- Quality Certificate
- Test Certificate for Image Quality without DFR.
The tables in the corresponding certificates are marked with the note "Checked during
startup".

Manual Fluoroscopy
Setting appropriate X-ray tube load factors (kV or mA) to check the radiation safety mea-
surement values.
Necessary only if a radiation safety measurement must be performed.
• Start the POLYDOROS SSW:
- < Diagnostics >
- < Manual Fluoroscopy >
In this menu, the kV and mA can be set manually and fluoroscopy can be switched on.

General Information 0

Determining Veiling Luminance


To avoid any problems during the Acceptance Test for monitors per DIN 6868 - 57, the
lighting conditions in the control and examination rooms must be tested as soon as possi-
ble.
For this, the veiling luminosity of the image playback devices on which diagnosis is made
must be measured using a luminous density meter (e.g. Mavo monitor) in the position in
which the devices are normally operated.
• The room lighting is adjusted to the operating conditions (it may be necessary to ask the
user) that are planned for later using a dimmer with a scale.
• The diagnostic monitor must remain switched off.
• Measure the veil luminosity of the diagnostic monitor with the luminous density meter at
a distance in front of the screen that is specified by the manufacturer.
• If the veiling luminance is less than 2 cd/m2, the minimum requirements of the Accep-
tance Test for Image Playback Devices can be considered as met.
• If, however, the veiling luminance exceeds the above value, the operator and/or site
manager must be informed to that effect immediately to ensure that remedial action
(additional defined reduction of the illumination, darkening of windows etc.) can be
taken in time. For more information, please refer to the Project Manager Handbook or
PG Planning Guide.
These measurements are necessary so that the minimum requirements for the Accep-
tance and Constancy Test for Image Playback Devices per DIN 68 68 - 57 can be
ensured.

AXIOM Iconos R100 AXD3-350.815.04.03.02 Page 10 of 82 Siemens AG


05.07 AX SCM W II Medical Solutions
General Information 11

Country-specific Acceptance Certificates


The following tests applicable to the prescribed acceptance test in Germany in accor-
dance with §16 of the X-ray ordinance or the acceptance test in the USA, were performed
during factory testing and were documented in the test protocols:
• Visual test of filter values
• Coincidence of Radiation Field and Center of Film
• Accuracy of Automatic Formatting for Cassette Formats
• Fluoroscopic field limit
• Centering of radiation field and center of spotfilm device
• Fluoroscopy dose rate
• Function check of the IONTOMAT measuring fields
• Maximum skin dose rate
• Resolution and minimum contrast
• Check of kV Accuracy
• Accuracy of the X-ray tube voltage
• Accuracy of the voltage display for fluoroscopy
• Shortest cutoff time for automatic exposure
• Reproducibility of radiation

NOTE These measurement values and the other values determined dur-
ing startup can be transcribed from the Test Certificate to the
Abnahmeprüfprotokoll or to the Acceptance Test Certificate.

The tests required by DHHS Regulations must be performed:


1. Required documents for the customer
2. Radiation Safety
3. Check of DHHS and identification labels

Tolerance Data
General tolerances for linear dimensions according to ISO 2768

Limit values for over 3 mm over 6 mm over 30 over 120 over 400 over over
the nominal to 6 mm to 30 mm mm to 120 mm to 400 mm to 1000 mm 2000 mm
range mm mm 1000 mm to 2000 to 4000
mm mm
Allowable tol- ± 0,5 mm ± 1 mm ± 1,5 mm ± 2,5 mm ± 4 mm ± 6 mm ± 8 mm
erance

These tolerances apply to all dimensions given in these instructions, unless another toler-
ance is expressly indicated after the value.

Siemens AG AXD3-350.815.04.03.02 Page 11 of 82 AXIOM Iconos R100


Medical Solutions 05.07 AX SCM W II
12 General Information

Torque Values
A tolerance of ± 10 % is permitted for torque values.

Conventions and Abbreviations


P • The measurement results corresponding to the measurements labeled with the symbol
must be transcribed into the corresponding certificates.
Abbreviations:

OT = Overtable
SID = Film-Focus Distance
I.I. = Image Intensifier
DL = Fluoroscopy
SSW = Service Software
SPC = Service PC
BWS = Bucky wall stand
S = Sensitivity of a film-screen system
KS = Manufacturer specification of the dose requirement for a
film-screen system
RGr = Resolution power of the screen depending on S; KS
RG = optical resolution power
KT = Dose, dose rate measured on the tabletop
KB = Dose, dose rate measured in the image receptor plane
fK = Correction factor for dose detector
m = Equipment attenuation factor
D = Optical density for the film
DN = Optical net density (D - base fog)
SHa = Numerically displayed tomo height on the control console
BST = Exposure steps
EP = Exposure points
IQ = Image quality
XCS = X-ray Communication System
XCU = X-ray Control Unit

AXIOM Iconos R100 AXD3-350.815.04.03.02 Page 12 of 82 Siemens AG


05.07 AX SCM W II Medical Solutions
General Startup 13
2-
Checks
2General Startup

Checking the System Emergency Off (if configured) 0

• When the system emergency off switch is pressed, power must be switched off to the
entire system.

Testing the Ground Wires 0

The following ground wire check is used for the safety of the startup personnel.
The final ground wire check is performed and documented at the conclusion of the star-
tup.

Geraet / Unit
Monitorampel / Monitor rack
Monitore / Monitors
Schraenke / Cabinets
Trafos / Transformers
Pulte / Consoles
BV / I. I.
Roentgenroehre / X-ray tube unit
Tiefenblende / Collimator
Sonstige Komponenten / other Components

Generator
Netzverteiler / Power distbutor

Fig. 1: Ground wire check with fixed power connection


• Switch off power to the system.
• Test procedure:
- Measure between all conductible parts of the system that can be touched and the
ground wire bus rail (e.g. in the generator, power distributor).

¹ Maximum value: 0,2 Ω (Observe country-specific regulations!)


Exception: Control consoles, accessory rail on the collimator and tabletop.

Siemens AG AXD3-350.815.04.03.02 Page 13 of 82 AXIOM Iconos R100


Medical Solutions 05.07 AX SCM W II
14 General Startup

NOTE After completing all adjustment work steps and check measure-
ments, the responsible technician must sign and date the certifi-
cates (system, generator and IQ test certificate for direct
exposures) to confirm that all values have been correctly deter-
mined and recorded.

AXIOM Iconos R100 AXD3-350.815.04.03.02 Page 14 of 82 Siemens AG


05.07 AX SCM W II Medical Solutions
General Startup 15

Power Connection 2.1

System Voltage and Frequency Check 0

Using the customer order list, check the connections and voltage / frequency toggles (side
of unit / generator side).
Generator Voltage / Frequency
• In systems with 440V or 480V line power, a pre-transformer must be installed in the
generator and connected per the Wiring Diagram.
• Check of the Frequency on the Generator D160
for 50 Hz --> jumpers X20-21; X23-X24
for 60 Hz --> jumpers X21-22; X24-X25
Unit Voltage
• Voltage on the T1 unit transformer:
for 400V: 1U1 --> 1W3; 1V1 --> 1U3; 1W1 --> 1V3
for 440V: 1U1 --> 1W2; 1V1 --> 1U2; 1W1 --> 1V2
• With 480V, a pre-transformer (M20) must be installed and connected. The voltages on
the unit transformer remain at 400V.

Measuring the Power Line Voltage 0

• POLYDOROS: Remove the F1, F2 and F3 blade-type power line fuses.


• Switch off the F456 trip breaker for PL SX / F4 and F5 for PL LX.
• Switch the main system switch ON.
V • Measure the line voltage. The voltage must agree with the data in the "Test Certificate
1" (Techn. Doc. / System Reg. 9).
• Switch the main system switch OFF.

Checking the Connection for Correct Phase 0

• To check the connection for correct phase, see the chapter "Switching on the System".
(Checking the Rotational Field at the Unit / p. 19)

Measuring the Internal Line Resistance (Power Quality) 0

Measure with the internal line resistance meter connected to M16.L1, L2, L3 (POLY-
DOROS Wiring Diagram) (one after the other, phase to phase):
• Connect the internal line resistance meter each time between two phases of L1, L2, L3.
• Switch the main system switch ON.
• Perform the measurement.
• Switch the main system switch OFF.

Siemens AG AXD3-350.815.04.03.02 Page 15 of 82 AXIOM Iconos R100


Medical Solutions 05.07 AX SCM W II
16 General Startup

• Disconnect the test meter and perform the measurement for the other phases.
• Reinstall the M16 F1, F2 and F3 blade type fuses and switch on the trip breaker(s).

NOTE To achieve the full generator power, the measured internal line
resistance may not exceed the following values.

ma. RLine for POLYDOROS LX ma. RLine for POLYDOROS SX


ULine
50 KW 65 KW 80 KW
400 V 0.17 Ohms 0.17 Ohms 0.11 Ohms
440 V 0.20 Ohms 0.20 Ohms 0.14 Ohms
480 V 0.24 Ohms 0.24 Ohms 0.16 Ohms

P • Record the highest measurement value in "Test Certificate 1".


• If the internal line resistance is higher, the tube current must be limited in the XCS SSW.

AXIOM Iconos R100 AXD3-350.815.04.03.02 Page 16 of 82 Siemens AG


05.07 AX SCM W II Medical Solutions
General Startup 17

Checking for Inadmissible Ground Loops 2.2

• Switch the main system switch OFF.

For POLYDOROS LX 0

Fig. 2: D160 “E3”


• After disconnecting the connection between Gnd and 0VL in the generator power cabi-
net, remeasure to determine whether any inadmissible ground loops are present.
- To do this, disconnect the connection cable (Fig. 2 / p. 17) on the D160 board
between X110 (0V LOAD) and X112 (Gnd) at one end.
- If the touchscreen console is connected, unplug connector X9/D160, otherwise the
measurement value will be < 30 kΩ.
- Use the multimeter to measure between the X110 and X112 connectors:
If no ground loops are present, the measured resistance must be >30 kΩ.
- Plug the connection cable (Fig. 2 / p. 17) D160.PE to 0VL back in.
- Plug the X9/D160 connector back in.

Siemens AG AXD3-350.815.04.03.02 Page 17 of 82 AXIOM Iconos R100


Medical Solutions 05.07 AX SCM W II
18 General Startup

For POLYDOROS SX 0

X41 X13 X11

X12
CPU X5
X6 X121 X14

X60

X88
X9
X65
X7 X110 X77
X3
X64
D190
D160 X8 X10
D100
X63
X55
T1 LS
X62 X4
D165 PS
X88
X61 X45
X1
D111 X2 X44

D220
1
H1

KKI M15

EH111509-200202-00-01

Fig. 3: Jumper on D160


After disconnecting the connection between Gnd and 0VL in the generator power cabinet,
remeasure to determine whether any inadmissible ground loops are present.
• Remove the X6---X7 wire jumper (1/Fig. 3 / p. 18) on D160.
• Measure the resistance between X6 (0VL) and X7 (Gnd).
- The measured resistance in a cabled system must be > 30 kΩ.
• Reinstall the X6---X7 wire jumper on D160.

AXIOM Iconos R100 AXD3-350.815.04.03.02 Page 18 of 82 Siemens AG


05.07 AX SCM W II Medical Solutions
General Startup 19

Switching on the System 2.3

• Switch the main system switch ON.


The red LEDs on D160 in the generator (LX: V23; SX: V78) must go on and with the LX,
the H1 and H2 lamps must go on.

NOTE If there is a malfunction, check the fuses on the T1 transformer


(under the D160).

• Switch the system ON at the generator console.


Display in the Generator:
• The generator initializes up to “ 1” (see the V14 7-segment display on the D100).
¹ The system is now ready to trigger radiation.

NOTE If an error is indicated on the 7-segment displays, see also the


XCS Service Application, Main Program, Help, Index.

Function Check of the Emergency Stop Switches 0

• Press the EMERGENCY STOP switch on the remote control console and on the unit.
¹ The safety breaker(s) K1 (K2) is (are) audibly deenergized.
"Emergency stop" or "ERR 241/032 Iconos" is displayed on the remote control
console.
• Disengage the EMERGENCY STOP switches.
¹ The safety breaker(s) K1 (K2) is (are) audibly energized.

Checking the Rotational Field at the Unit 0

Checking the Direction of Movement


• Control tabletop movements in the head-end and foot-end directions and check for cor-
rect direction of movement.
Correction:
• Check the power line connection or the cabling from the generator to the system and, if
needed, change it (rotational field).

Siemens AG AXD3-350.815.04.03.02 Page 19 of 82 AXIOM Iconos R100


Medical Solutions 05.07 AX SCM W II
20 Service Software
3-
Service Software
3Service Software

Connecting the Service PC to the POLYDOROS LX 0

Option
D110
Touchscreen-
control console D115

D100

D200

D220
SN
H1
X5
D291

K20
D160
T2
X9 K30
F1 F2 F3

M16

ON

Com 1

Com 2

X5

SPC
Important!
The system controller (D200.X5)
must be connected to Com 1.

509-100100-0

Fig. 4: Connecting the Service PC

AXIOM Iconos R100 AXD3-350.815.04.03.02 Page 20 of 82 Siemens AG


05.07 AX SCM W II Medical Solutions
Service Software 21

Connecting the Service PC to the POLYDOROS SX 0

Com 1 (serial) to XCU


D320 CPU HD
D 320. SPC Mode (X5) M9 DIA UNIT

D101
SNT
D190 D100
D160

F8
F7
T1 D165 LS

PS
Z1
D110 C3, C5

D115
DLS K5
D111 N93
K3

D220

H1

KKI M16

Fig. 5:
• Connect the Service PC as shown in the above illustration.

Installation of the XCS SSW and the ICONOS R100 SW on the Service PC 0

NOTICE All Iconos R100 systems have the TOMO function as standard.
The distinction between “with Tomo” or “w/o Tomo” is made later
by the settings in the Site Structure of the SSW.
¹ The software "with Tomo" should ALWAYS be installed.

NOTE The XCS SSW is located on the system software CD included in the
shipment. Following installation, check for the correct version
(per the Compatibility List).

• Rename the previously installed XCS SSW on the Service PC or delete it.
• Insert the current ICONOS R100 SW CD into the Service PC.

Siemens AG AXD3-350.815.04.03.02 Page 21 of 82 AXIOM Iconos R100


Medical Solutions 05.07 AX SCM W II
22 Service Software

• Start Setup.exe from the CD.


• Select "for Serial No. 2001 and above..." in the "Iconos R100 overall installation" menu.
• Following installation, close the window.
• Remove the diskette from the drive.

Checking the Software Versions 0

• Select "Info/ SW Firmware Version" in the XCS SSW.


• Compare the software versions to the Compatibility List included in the shipment.
• If there are differences, perform an update.
P • Sign the Compatibility List.

Check and Adjustment of Basic Unit Data, Site Structure 0

NOTE Options, e.g. Tomo, Compression or KermaX are not pre-set at the
factory and must be checked in the basic unit data according to
the customer order and, if needed, configured.

Component Option Group Option Setting


Iconos R100 Tomo not available
available, only with the Schichten (Tomo)
option
Bucky wall stand (BWS) available / not available (depending on CO)
SID 150
Motorized compression not available
available only with the Compression option
Reversible Grid Not available
For DLR cassettes: available
Iconos R100 Grids enabled 17/70
--> Edit
Filter available Filter 1: Cu 0.1mm
(Only for
Polydoros Filter 2: Cu 0.2mm
SX) Filter 3: Cu 0.3mm

AXIOM Iconos R100 AXD3-350.815.04.03.02 Page 22 of 82 Siemens AG


05.07 AX SCM W II Medical Solutions
Service Software 23

Component Option Group Option Setting


Polydoros SX maximum power 65 kW or 80 kW (depending on CO)
fluoroscopy available
Planiiontomat available
max. tube current limita- 800 mA
tion
max. mAs limitation 800 mAs
SID tracking Not available
SDM Flip Not available
max. tube current < 100 800 mA (at 65 kW)
ms 1000 mA (at 80 kW)
DCM Not available
Twin mode Not available
Spine Reconstruction Not available
Value for exp. pt. 63 mAs For China: 63mAs/99kV
Rest of World: 65mAs/100kV
Polydoros LX maximum power 30 kW or 50 kW (depending on CO)
N
continuous fluoroscopy available
max. tube current limita- 500 mA (at 30 kW)
tion 800 mA (at 50 kW)
Desk PLX tube load buzzer on
Organ program no restriction for DFR
FLC emulation no
Videomed after flash no
DIC Dose reduction no/Supervision (depending on CO)

Siemens AG AXD3-350.815.04.03.02 Page 23 of 82 AXIOM Iconos R100


Medical Solutions 05.07 AX SCM W II
24 Service Software

Component Option Group Option Setting


Data Printer print selection print after fluoro and exp.
(option) Printer Type Star DP 8340S
Only for Poly- (depending on CO)
doros SX1
Print behavior (DFR,SFD, no printing
Bucky,...)
Printer settings
• Baud rate: 9600
• Parity: not selected
• Stop bit: 1 Bit
• Data bit: 8 Bit

1. With the LX, do not configure the data printer in the Site Structure.

Workplace Configuration

form Peri position


System Selection Displayed system
name
Default System

Dose Q-factor
Detector Type
Iont. Channel
Fluoro avail.

# onwards
Iontomat
GNo.
WS

ICONOS Spotfilm ICONOS Spotfilm X X - 1 A 2p 6mm - -


3 fields
ICONOS Film Tomo ICONOS Film Tomo - X - 1 - - - -
ICONOS Free Exp. ICONOS Bed - - - 1 - - - -
Bucky wall stand Bucky wall stand - - - 1 C 2p 6mm - -
(with BWS/without 3D 3 fields
TOP)

WS Tube type of ICONOS Diamentor High voltage Available


[kV] max. Focus
1 Optitop 150/40/80/HC-100 (3~) option 150 Small + Large
Opti 150/30/50/HC-100 (3~)

AXIOM Iconos R100 AXD3-350.815.04.03.02 Page 24 of 82 Siemens AG


05.07 AX SCM W II Medical Solutions
Service Software 25

Fluoroscopy Configuration

Configure: Site Structure: Fluoro Details


Tube focus for fluoro small
Detector type for fluoro B signal CCD

Fluoro/DFR image intensifier


Type 23 cm HDR
Type 33 cm HDR

Fluoro time max. settings in minutes


Cumulative fluoro time until alarm 5.0
Max. fluoro time without interrupt 10.0 (for Canada and USA on disable)

Skin entrance dose (only in connection with Diamentor)


No display not selected
Absolute display selected
skin to focus distance = 78 cm

Configuration of the Fluoroscopy Programs

Configure: Site Structure: Edit Fluoro Programs


(without Supervision)
Fluoro Program Mode Selection Curve Selection Dose Rate *[nGy/s]
Auto 1 Cont. fluoro Antiisowatt -
Auto 2 Cont. fluoro Dose-reduced -

Configure: Site Structure: Edit Fluoro Programs


(with Supervision option)
Fluoro Program Mode Selection Curve Selection Dose Rate *[nGy/s]
Auto 1 Cont. fluoro and Antiisowatt -
SV Auto 1 Supervision Antiisowatt SV -
Auto 2 Cont. fluoro and Dose-reduced -
SV Auto 2 Supervision Dose-reduced SV -

Siemens AG AXD3-350.815.04.03.02 Page 25 of 82 AXIOM Iconos R100


Medical Solutions 05.07 AX SCM W II
26 Service Software

Additional Settings

Configure: Site Structure: Site Adjustments


Control console and DR settings
kV/mAs/density step width 1/2 exposure point
Iontomat mode kV (not configurable for PL LX)
mAs mode mAs (not configurable for PL LX)
Error message display Display error message (not for PL LX)
Screen Saver turn on after 5 (min)
Room light switch off manual (not for PL LX)
Language Language see production control list (in
German only for LX gen.)
DSA dose incr. below 0 cm water equ. (not for PL LX)
min. DCM time (PL LX) Default (not for PL LX)
min. DR pulse wth. (PL SX)
Screen sensitvitiy H(400):400, U(200):200, D(100):100
(for display on console)

Site Options
Option group selection organ programs
Option selection available

Organ selection after system change


select prev. organ in system selected

Format-dependent correction
Adaption value [Exp. Points] 1.0

Spotfilm device
Eject cassette after last exposure selected

Collimator
Zoom 1 < Zoom 2: Set to full intensifier selected (not configurable for PL LX)
field

AXIOM Iconos R100 AXD3-350.815.04.03.02 Page 26 of 82 Siemens AG


05.07 AX SCM W II Medical Solutions
Service Software 27

Downloading the Control Console Language 0

Configure the language of the control console per the country or customer requirements.
• Select "Components / Desk PLX" in the XCS SSW.
• Select the appropriate language in the "Desk language to be loaded" box.
• Download the selected language with <Load language>.
• Close the window <Done>.
• Switch the system OFF and back on.

Siemens AG AXD3-350.815.04.03.02 Page 27 of 82 AXIOM Iconos R100


Medical Solutions 05.07 AX SCM W II
28 System Startup
4-
Starting Up the Unit
4System Startup

Checking the Safety Limit Switches 0

CAUTION Risk of crushing to hands!


¹ In the following check of the safety limit switches, make
sure that no unintentional unit movements are con-
trolled.

509-100101-02

Fig. 6: ICONOS R100

Safety Limit Switches on the Head-end/Foot-end Unit Strut


• Manually actuate the safety devices (1 und 2/Fig. 6 / p. 28).
¹ The safety breaker(s) audibly deenergize.
• Manually activate the safety devices, the switch strip affects the spotfilm device drive
and column drive (3/Fig. 6 / p. 28).
¹ The drives switch off.

AXIOM Iconos R100 AXD3-350.815.04.03.02 Page 28 of 82 Siemens AG


05.07 AX SCM W II Medical Solutions
System Startup 29

Room Dimensions 0

Fig. 7: Room limitation

NOTE The safety distance between the unit and the wall of ≥ 50 cm
applies only for Germany.
Observe country-specific safety regulations.
The footrest is always included in the head-end and foot-end colli-
sion calculation.

A: Ceiling Height
• Measure the distance from the floor to the ceiling (Fig. 7 / p. 29).

NOTE The SSW automatically limits unit movement and takes attaching
the footrest at the head-end into consideration.
The safety distance to the ceiling must be at least 5 cm!

B: Head-end Wall Distance


• Move the unit into the 0° position.
• Move the tabletop so it is flush with the frame at the head end.
• Measure the free space beginning at the head end of the table frame (Fig. 7 / p. 29).

Siemens AG AXD3-350.815.04.03.02 Page 29 of 82 AXIOM Iconos R100


Medical Solutions 05.07 AX SCM W II
30 System Startup

NOTE The safety distance (50 cm or 20 inches) is automatically calcu-


lated.

C: Foot-end Wall Distance


• Move the unit into the 0° position.
• Move the tabletop so it is flush with the frame at the head end.
• Measure the free space beginning at the foot end of the table frame (Fig. 7 / p. 29).

NOTE The safety distance (50 cm or 20 inches) is automatically calcu-


lated.

ICONOS SSW, "Room Dimensions"


• In the XCS SSW, start the program under "Components/ ICONOS/ Adjustments/ Room
Dimensions"

Fig. 8: Screen snapshot of the room dimensions


• Enter the ceiling height, head-end and foot-end wall distance as well as the floor dis-
tance in the SSW menu "Room Dimensions" and confirm with "OK" (Fig. 8 / p. 30).

AXIOM Iconos R100 AXD3-350.815.04.03.02 Page 30 of 82 Siemens AG


05.07 AX SCM W II Medical Solutions
System Startup 31

NOTE If no lifting base is configured, the default value in the "Distance


from floor to I.I." box may not be changed.

Patient Table 0

Checking lowering to the foot end


• Move the tabletop so it is approx. flush with the frame.
• Move the longitudinal carriage to approx. the middle of the movement range.
• Tilt the unit up all the way.
• Move the tabletop down until the end position is reached.
• Measure the distance from the floor to the bottom edge of the tabletop.
¹ Required: Distance to the floor at least 12 cm
• Again move the tabletop in the direction of the floor and measure the distance.
¹ Required: Distance to the floor at least 4 cm

Checking the Room Limits


• Move into the end positions with the tabletop or the tube carriage to determine whether
the safety distances are being maintained. If necessary, again enter the room dimen-
sions.

Siemens AG AXD3-350.815.04.03.02 Page 31 of 82 AXIOM Iconos R100


Medical Solutions 05.07 AX SCM W II
32 System Startup

Installing and Starting Up Compression (Option) 4.1

NOTICE If compression is ordered (per the customer order) and it is not


installed, it must be mechanically installed and started up.
¹ Compression is only mechanically installed on the tomo
rod during system installation.
¹ Cabling and startup are performed by the startup techni-
cian, because the unit must be tilted into the 90° position.
When doing this, pay attention to the cable length from
the unit support to the base.
¹ See also AXD3-350.814.01.xx (Compression Installation
and Startup).

Compression Cabling 0

• Install the cable holder bracket (1/Fig. 9 / p. 32) on the spotfilm device with 3 Allen
screws.

Fig. 9: Spotfilm device cable holder

Fig. 10: Laying cables for Compression

• Lay the cable under the plate per the above illustration and secure it to the frame with
an adhesive-backed clip.(1/Fig. 9 / p. 32) When doing this, pay attention to the cable
lengths when in an oblique position.
• Connect the ground wire in the spotfilm device electronics.
• Lay the compression cable along with the I.I. cable through the appropriate zipper
hoses.(1/Fig. 10 / p. 32)

AXIOM Iconos R100 AXD3-350.815.04.03.02 Page 32 of 82 Siemens AG


05.07 AX SCM W II Medical Solutions
System Startup 33

• The cable length of


(X/Fig. 11 / p. 33) from the unit sup-
port to the base, is 1600mm..

Fig. 11: Cable length from support to base

• Connect the cables in the electronics frame as follows:


- A4.D31 X23
- A4.D31 X32
- A4.D31 X51 (remove jumper)
- A4.D41 X9

Starting Up Compression 0

Adapting the SSW


• Start the XCS SSW.
• Select Configure -> Site Structure.
• Select <ICONOS>.
• Start with <Option>.
• Select <Motorized Compression> and set it to available.
• Confirm several times with "OK" and save with “Save”.
• End the XCS SSW.
• Switch the system OFF and back on.

Siemens AG AXD3-350.815.04.03.02 Page 33 of 82 AXIOM Iconos R100


Medical Solutions 05.07 AX SCM W II
34 System Startup

Function Test

• Move compression up and down:


3 - Check for reliable switch-off in the end positions.
Correction: Adjust the switches (M6.S103 and
M6.S102) accordingly.
4
- Check the tension between the friction wheel
1
and the tomo rod profile. To do this, move com-
pression down and while doing so, block com-
pression with your hand.
The friction wheel may not continue turning.
Correction: Adjust the friction wheel by turning
2 the adjustment screw (5/Fig. 12 / p. 34) counter-
clockwise.
5 • Secure the friction wheel adjustment screw with
the safety plate. (5/Fig. 12 / p. 34)

Fig. 12: Replacing the compression motor

• Check potentiometer adjustment in the SSW.


• Install the cover panels and stop studs.
• Move the unit into the 0° position.
• Push the centering button on the table-side or remote control panel to position the col-
umn vertically.
Checking the Speed
• Move the compression cone from up to down or from down to up and check the move-
ment time.
¹ Nominal value: 8 sec, +/- 2 sec
Checking the Force of the 50 N Monitoring
• Insert the compression cone.
• Place a suitable test object on the tabletop (do not scratch the tabletop!).
• Compress to force step ≥ 5 (display on the remote control console).
¹ All basic unit movements, except for unit tilt movement, must be blocked.
• Continue to increase the compression force until compression switches off.
¹ The S108 safety limit switch has responded.
¹ 12...15 must be displayed on the control console.

AXIOM Iconos R100 AXD3-350.815.04.03.02 Page 34 of 82 Siemens AG


05.07 AX SCM W II Medical Solutions
System Startup 35

Unit Movements 4.2

Check all unit movements for their function and for problem-free movement, i.e.:
• Perform a function check of the movements using the System User Instructions.
• No scraping noises.
• No hitting at the mechanical end stops.

Siemens AG AXD3-350.815.04.03.02 Page 35 of 82 AXIOM Iconos R100


Medical Solutions 05.07 AX SCM W II
36 System Startup

Starting Up the Generator 4.3

Checking the Generator without High Voltage 0

Checking the Starter Unit (Workstation Selection and rpm)


• Set the switch to SS “OFF” (S1 on the D100 in the generator).
• Set the ZK switch to “ON” (S2 on the D100 in the generator).
• Select "Free Exposure (on the table exposure)" at the generator control console.
• Press the exposure switch at the generator control console to prep:
¹ The rotating anode will start.
• Release the exposure switch at the generator control console:
¹ The rotating anode will be braked.

Checking the Generator with High Voltage 0

Preheating the X-ray Tube


Warm-up:
• Start the XCS SSW.
• Close the collimator and cover the I.I. with lead (e.g. lead aprons).
• Set the S1 switch (D100 in the generator) to “ON”.
• Set the S3 switch (D100 in the generator) to Service.
• Remove the T1 4U1 fuse in the electronics frame for the I.I. power supply.
• In the XCS SSW program, select "Components/ Polydoros LX N or SX/ Adjustment/
Warmup”.
• Select the tube unit.
• A message will prompt you to close the collimator (Please close the collimator).
• Close the message window with "OK”.
• Switch fluoroscopy ON. Fluoroscopy will turn on and will automatically be turned off
again after approx. 10 minutes.
• Confirm the menu with "OK".
Pre-Heating:
• Switch fluoroscopy ON.
• Fluoroscopy will switch back to OFF after a few seconds.
• Confirm the menu with "OK".
Filament Correction:
• The load values are selected automatically.
• Confirm the message "will now release exposure at..." with "OK". The exposure is per-
formed automatically.

AXIOM Iconos R100 AXD3-350.815.04.03.02 Page 36 of 82 Siemens AG


05.07 AX SCM W II Medical Solutions
System Startup 37

• Confirm the message "release next exposure" with "OK" and switch exposure back ON.
• Repeat this routine until the "OK" radio button is activated.
• Confirm the menu with "OK".
Perform the Pre-Heating and Learn Filament settings for small and large focus; the menus
for both focuses will be displayed automatically.
Filament Push current:
• Confirm the message "will now release exposure at..." with "OK".
• Confirm the message "release next exposure" with "OK" after 3 s and switch exposure
back "ON".
• Repeat this routine until the "OK" radio button is activated.
• Confirm the menu with "OK".
• Confirm the following window with “OK”.
Control of Adjustment Parameters:
• Select the "Diagnostics / Adjustment result / Tube Adjustment Params" menu item.
• Check the values for the push factors; the table shows the permissible ranges.

Tube Focus Pre-heating Filament Filament Filament


Unit [mA] Correction, Correction, Push Cur-
MUL ADD rent P-Factor
1 Small focus 2200 to 2900 0.9 to 1.2 -700 to +700 < 1,09
Large focus 1800 to 2300 1,09 ... 1,6

• Switch the system off and back on.


¹ If the above-mentioned values are not maintained, contact Process Planning!

OPTITOP Tube Unit and Idle Times of < 2 Weeks: Conditioning the X-ray Tube 0

• Select "Free Exposure (on-the-table exposure)" at the ICONOS control console.


X • Trigger the following exposures quickly and without braking the rotating anode between
the two exposures.

70 kV / 100 mAs / large focus / 80% 2X Pause 1 minute


90 kV / 400 mAs / large focus / 80% 2X Pause 3 min-
utes
109 kV / 400 mAs / large focus / 80% 2X Pause 3 min-
utes

NOTE If the tube tends to repeatedly and strongly "arc", abort the proce-
dure and, if needed, replace the tube unit (-> risk to sensitive
parts of the system electronics).

Siemens AG AXD3-350.815.04.03.02 Page 37 of 82 AXIOM Iconos R100


Medical Solutions 05.07 AX SCM W II
38 System Startup

Fluoroscopy Attachment (only in Combination with Polydoros SX) 0

NOTE Perform the adjustment of the fluoroscopy attachment in the XCS


SSW without interruption.
Carry out the adjustment only after the generator has been in
operation for 3 hours.

• Set the S3 switch to Service.


• In the XCS SSW program, select “Components/ Polydoros SX/ Adjustment/ FC Attach-
ment“.
• Switch Auto-Calibration to “Auto“ and confirm the message window with “OK“. The
adjustment is performed for several work points (kV, mA). When this is done, fluoros-
copy is switched on and off automatically.
• Exit the menu with “OK“.

Check of the max. Generator Power 0

• Cover the I.I. (e.g. with a lead apron).


• In the XCS SSW program, select "Components/ POLYDOROS/ Diagnostics/ Nominal
Power".
- Tube 1
- Large focus
- 100 kV
- Depending on the tube unit / generator, set 50 mAs (50 kW), 65 mAs (65 kW) or
80 mAs (80 kW)
- 3-point technique
- 100 ms
• To check the tube voltage and tube current, connect the oscilloscope to the following
test points on the D100.

kV_IST (X61) 1V 20 kV / 0 V/A_GND (X61)


MA_IST (X64) 1V 200 mA for exposure

MA_IST (X64) 1V 1 mA for fluoroscopy


Trigger: SWR (X64) Start/Stop the inverter

X • Trigger exposure and use the oscilloscope to measure the kV and mA.
• Calculate the max. power from the measured kV and mA:

P • Calculate the power value (kW).

AXIOM Iconos R100 AXD3-350.815.04.03.02 Page 38 of 82 Siemens AG


05.07 AX SCM W II Medical Solutions
System Startup 39

• Recorde the measured mA value and the calculated power value (kW) in the "Quality
Certificate or Test Certificate 1”.

Correction 0

If the values are not within tolerance, the tube currents must be reduced; to do this:
• In the XCS SSW, select <Configure/ Site Structure/ Component selection/ Poly-
doros/ Option>.

EH111509-120302-00-01

Fig. 13: Tube current reduction


• Set a reduced value under "max. tube current limitation” and "max. tube current
<100 ms".
• Reinstall the I.I. power supply fuse.

Siemens AG AXD3-350.815.04.03.02 Page 39 of 82 AXIOM Iconos R100


Medical Solutions 05.07 AX SCM W II
40 Beam Geometry
5-
I.I. / TV and Central Beam
5Beam Geometry

Checking the Collimator Position 0

• Switch on the collimator light field.


• Select Bucky.
• Completely open the collimator.
¹ The collimator plates must be parallel to the edges of the table.
Correction:
• The position can be changed by turning the eccentric bearing for the collimator stop
lever.

Determining the Center of the Spotfilm Device / Center of Monitor 0

Checking Image Centering


• Move the unit into the 35° position.
• Select image intensifier full format.
X • Select all 4 possible image reversal positions at the unit control console and when
doing so, under fluoroscopy check whether the center of the spotfilm device (centering
template) remains within tolerance on the monitor when switching between normal dis-
play and image reversal.
¹ Maximum tolerance with the 23 cm I.I.: 5 mm and with the 33 cm I.I.: 7 mm.

Correcting Image Centering (only if necessary)


• Loosen the counter nuts for the adjustment screws on the I.I. mount.
X • Under fluoroscopy, move the I.I. until the adjustment template comes into admissible
tolerances with image reversal.
• Also perform these checks in the 90° position.
• Retighten the counter nuts.

NOTE The adjustment screws must be up against the brackets and the
counter nuts must be tightened.

¹ If the maximum difference is not reached, correct the I.I. optics.

Correcting the I.I. Optics (adjusted at the factory)


• Set switches S2.1, 2, 7 and 8 on the D111 camera board to ON.
- Press the S1 reset button.
¹ The blanking circle will be displayed (flickers).

AXIOM Iconos R100 AXD3-350.815.04.03.02 Page 40 of 82 Siemens AG


05.07 AX SCM W II Medical Solutions
Beam Geometry 41

X • Switch on fluoroscopy.
If the I.I. circle is not in the center of the blanking circle, the optics must be readjusted. To
do this,
• Loosen the screw on the I.I. optics.
X • Under fluoroscopy, move the optics until the center is reached.
• Readjust the eccentric bearing for the optics.
• Set switches S2.1, 2, 7 and 8 on the D111 camera board back to OFF.
• Press the S1 reset button.

Camera Rotation 0

Checking the Image Position under Fluoroscopy


• Move the unit into the 35° position.
• Place on the centering template (Fig. 15 / p. 43).
• Place a marker (metal washer) at the head end, left, on the tabletop.
• At the unit control console, select full format and normal image display.
¹ The unit head end must be displayed at the top on the monitor. The left patient
side must appear on the right on the monitor.
X • Switch on fluoroscopy and check whether the centering template is displayed vertical to
the monitor.

Correcting Camera Rotation (only if necessary)

NOTE The D111 board can slip out of the camera mount by loosening
the clamp on the camera mount! There may be no gap between
the board and the camera mount.

• Switch off the system and remove the camera housing.


• Loosen the mounting clamp for the BNC cable to the camera.
• Loosen the clamp (2.5 mm Allen screw) for the camera (D111) so that the camera can
just still be turned.
• Rotate the camera in the desired direction.
• After rotating it, use your thumb to press on the copper stud in the middle of the board
and tighten the clamp.
• Switch the system on.
X • Repeat the adjustment until the camera position is optimal.
• Retighten the clamp for the BNC cable.
• Remove the marker on the tabletop.

Siemens AG AXD3-350.815.04.03.02 Page 41 of 82 AXIOM Iconos R100


Medical Solutions 05.07 AX SCM W II
42 Beam Geometry

Central Beam 0

Determining the Drift of the Central Beam between SID min. and SID max.

NOTE If only one SID (115 cm) is available, set the center of the radiation
field to the center of the spotfilm device.

• Position the column in the 0° position.


• Move the unit into the +35° position.
• Set an SID of 115 cm.
• Select the 23 or 22 I.I. format (with the 33 cm I.I.).
X • Switch on fluoroscopy.
• Adjust the collimator so that the plates are visible on the monitor.
• Place a sheet of plastic film on the monitor and mark the center of the radiation field on
it.
• Repeat the procedure at SID max.
• Determine the drift by means of the center markings:
¹ Maximum admissible difference:

Head-end/foot-end ≤2 mm
direction:
Operator's side / ≤2 mm
wall side direction:

If there is a difference, adjust the tube unit as follows.

Adjusting the Tube Unit (only if necessary)

Loosen
4 screws

Adjustment

Turn in operatorís side or Turn in foot-end or head-end


wall side direction direction
EH111509-310102-02-03

Fig. 14: Adjustment of Image Centering


• Perform the adjustment per the above illustration; to do this,
- temporarily loosen screws for the tube unit or the flange.

AXIOM Iconos R100 AXD3-350.815.04.03.02 Page 42 of 82 Siemens AG


05.07 AX SCM W II Medical Solutions
Beam Geometry 43

• Following adjustment, carefully tighten the four tube unit mounting screws; when doing
this, use the light localizer to make sure that the tube unit does not come out of adjust-
ment.
X • Perform a check under fluoroscopy and if needed, repeat the procedure.

Preparation
• Use the centering button on the unit console to center the column.
• Move the unit into the 0° position.
• Set an SID of 115 cm.

Placing On the Template

Serviceabdeckung
Service cover

Kunststoffabdeckung
Plastic cover

EH111509-110402-00-01

Fig. 15: Centering template


• Place on the centering template included in the shipment as shown in the above sketch;
to do this,
- remove the paneling screw.
- Place the centering template on the head end of the spotfilm device and secure it in
place with the paneling screw. It is critical that the centering template is flat on the
spotfilm device frame.

Determining Center of Radiation Field to Center of Spotfilm Device


Determine the maximum difference between the center of the radiation field and the cen-
ter of the spotfilm device at SID min.
• Move the unit into the +35° position.
• Set an SID of 115 cm.
• Select the 23 or 22 I.I. format (with the 33 cm I.I.).
X • Switch on fluoroscopy.

Siemens AG AXD3-350.815.04.03.02 Page 43 of 82 AXIOM Iconos R100


Medical Solutions 05.07 AX SCM W II
44 Beam Geometry

• Adjust the collimator so that the plates are visible on the monitor.

Centering template Center of spotfilm device

Blanking circle

head end

operator side

wall side
Center of radiation field foot end

EH111509-310102-02-04

Fig. 16: Determining center of radiation field / center of spot film device
• Determine the difference per the above illustration.
¹ Maximum admissible difference: ≤3 mm

Adjusting the Tube Unit in the Operator's Side/Wall Side Direction (only if neces-
sary)
• Switch on the collimator light localizer.
• Mark the position of the light field on the tabletop in the head-end/foot-end directions.

Fig. 17: Tube unit rotation assembly


• Loosen the two set screws (1/Fig. 17 / p. 44) in the adjustment ring (on the top part of
the tomo rod); to do this, move the column into the +30° position or -30° position.

AXIOM Iconos R100 AXD3-350.815.04.03.02 Page 44 of 82 Siemens AG


05.07 AX SCM W II Medical Solutions
Beam Geometry 45

• Correct the position of the tube unit by moving it in the operator's side/wall side direc-
tion.
• Tighten one of the two set screws again and when doing so make sure that the position
in the head-end/foot-end directions does not change (watch the mark on the tabletop).
• Loosen the clamping screw on the top part of the tomo rod and align it so it is vertical
using a spirit level.
X • If needed, repeat the procedure using control measurements with fluoroscopy.
• If the value is within tolerance, tighten both set screws.
T • Tighten the screw on the tomo rod to a torque of 50 Nm.

Adjusting the Tube Unit in the Head-end/Foot-end Direction (only if necessary)


• Remove the cable holder from the tube unit mount.

Fig. 18:
• Loosen the 4 screws (1/Fig. 18 / p. 45) on the tube unit mount.
• Move the tube unit in the transverse direction.
• Retighten all 4 screws.
X • Repeat the measurements under fluoroscopy.

Adjusting the Centering Cross (Dose Measurement Chamber) 0

• If a dose measurement chamber with centering cross was installed in the collimator at
the customer's location, the 4 Allen-head mounting screws must be loosened and the
centering cross adjusted so it is centered to the collimator blanking.

Siemens AG AXD3-350.815.04.03.02 Page 45 of 82 AXIOM Iconos R100


Medical Solutions 05.07 AX SCM W II
46 Beam Geometry

Checking the fluoro field limitation and centering 5.1

NOTE Centering of the TV camera and the fluoro field limitation has
been adjusted at the factory and is documented in the "Quality
Certificate". Make sure that the setting has not changed during
transport and installation.

• Move the unit into the 35° position (for units with the lifting base, move into the 0° posi-
tion).
• Set the max. SID.
• Select image intensifier full format.
X • Switch fluoroscopy ON
• Open the collimator all the way.

Fig. 19: Fluoroscopy limitation


Evaluate the display of the collimator/iris diaphragm plates on the monitor for all SIDs and
zoom formats (Fig. 19 / p. 46).
For collimators without iris:
¹ At least 2 collimator plates must be visible (one for height and one for width).
For collimators with iris:
¹ At least 2 collimator leaves (one for height and one for width) and at least 3 iris
diaphragm leaves must be visible.
Correction:
• In the XCS SSW program, select
"Components/ ICONOS/ Adjustments/ Calibrations/ Collimator/ Indirect technique, flu-
oroscopy field limitation".
¹ A lower offset value corresponds to a smaller collimation field
• Exit the ICONOS SW.

AXIOM Iconos R100 AXD3-350.815.04.03.02 Page 46 of 82 Siemens AG


05.07 AX SCM W II Medical Solutions
Beam Geometry 47

Checking the ICONOS Spotfilm Device 5.2

NOTE To ensure that the settings have not changed during system
transport, test exposures must be made and compared to the
exposures that are included in the shipment. If these test expo-
sures are within the listed tolerances, the factory set values
should not be changed.

Cassette Position 0

Preparation
• Use the centering button on the unit console to center the column.
• Move the unit into the 0° position.
• Set an SID of 115 cm.
• Place on the centering template (Fig. 15 / p. 43).

Checking the Template Position


• Place a marker, e.g. using soldering wire on a 35x43 cm film as shown in
(Fig. 20 / p. 47). When doing this, work very exactly!

Zentrierschablone
Center matrix
305 mm

Film

Serviceabdeckung
Service cover
382 mm

Markierung auf dem Film


Marker on the Film

Fig. 20:
• Place the marked film on the spotfilm device so that it is flush at the head end with the
outside edge of the spotfilm device and lines up with the tableside controller on the
operator's side (Fig. 20 / p. 47).

Siemens AG AXD3-350.815.04.03.02 Page 47 of 82 AXIOM Iconos R100


Medical Solutions 05.07 AX SCM W II
48 Beam Geometry

X • Under fluoroscopy, check whether the center mark on the marked film lines up with the
mark on the template.
Correction: If needed, correctly position the template or slightly bend the template.
Perform the check again!

Checking the Cassette Position


• Insert a 24x30 cm or 10x12 inch cassette sideways. Set an SID of 115 cm
• Place a marker (e.g. washer) for the 1st exposure on the tabletop at the foot end of the
central beam using the light localizer.

Lichtvisier
Light localizer

Markierung für 1. Aufnahme


Mark 1st exposure

Fig. 21: Test exposure marker


• Select the 4-part subdivision and full automatic.
• Exposure data: Select 42 kV, 8 mAs, large focus, 80% kW (the exposure data are
applicable for 200 film-screen combinations).
X • Trigger the first exposure and remove the mark from the tabletop.
X • Trigger the 2nd exposure.
• Remove the cassette, turn it 180° and reinsert it (wall side becomes the operator's
side).
X • Trigger the 3rd and 4th exposures.
• Develop the film.

AXIOM Iconos R100 AXD3-350.815.04.03.02 Page 48 of 82 Siemens AG


05.07 AX SCM W II Medical Solutions
Beam Geometry 49

Evaluating the Film:

Fig. 22: Film evaluation and cassette position

NOTE The dimensions c, d and e can also be measured in Step 2 or 3.

• Measure the dimension “a”.


a = 75 mm max. admissible difference ± 2 mm
If there are differences that are outside of the tolerance:
Mechanical cause: The cassette has slipped to the belt.
Perform a function check per:
Diagnostic/ Tests/ Movement tests/ Movement test: Transverse cassette drive in
park position direction
Diagnostic/ Tests/ Movement tests/ Movement test: Transverse cassette drive in
loading position direction
• Measure the dimension “b”.
b = 75 mm, max. admissible difference ± 4 mm
If there are differences that are outside of the tolerance:
• Insert the cassette.
Cassette exposure position: Adjust using the service software.
Adjustment/ Calibration/ Spotfilm Device/ Cassette drive, transverse: Exposure
position
Default value approx. 85; a higher figure means that dimension "b" becomes
smaller.

Siemens AG AXD3-350.815.04.03.02 Page 49 of 82 AXIOM Iconos R100


Medical Solutions 05.07 AX SCM W II
50 Beam Geometry

• Measure the dimensions “c” and “d”: These values should be identical.
c and d = max. admissible difference≤2 mm
If there are differences that are outside of the tolerance:
Correction: Mechanically adjust the symmetry of the collimator plates.
See the instructions, Replacement of Parts”, Spotfilm Device
• Measure the dimension “e”.
e = 72 mm, max. admissible difference ± 1 mm
If there are differences that are outside of the tolerance:
Correction: Collimator plate opening
• Insert the cassette.
Adjust using the service software.
Adjustment/ Calibration/ Spotfilm Device/ Diaphragm plates: Basic setting
Default value approx. 85; a lower figure means that "e" becomes smaller.
• In the range “f”, the distance between the center markers is equal.
f = max. admissible difference ± 5 mm
If there are differences that are outside of the tolerance:
Mechanical cause: The entire tensioning system is not symmetrical.
Check the tensioning drive mechanism.
• In the range “g”, the distance between the center markers is equal.
g = max. admissible difference 3 mm ± 2mm
If there are differences that are outside of the tolerance:
Mechanical cause: The entire tensioning system is not symmetrical.
Inform the process planner.
• In the range “h”, the distance between the center markers is equal.
h = max. admissible difference 3 mm ± 2 mm
If there are differences that are outside of the tolerance:
Mechanical cause: The entire tensioning system is not symmetrical.
Inform the process planner.

P • Enter all results in the Quality Certificate or Test Certificate 1.


• Label and archive the film.
• Remove the centering template.

AXIOM Iconos R100 AXD3-350.815.04.03.02 Page 50 of 82 Siemens AG


05.07 AX SCM W II Medical Solutions
Beam Geometry 51

Coincidence of Center of Radiation Field and Center of Film 0

Fig. 23:
• Insert a 24 x30 cm or 10 x 12 inch cassette sideways.
• Select the direct technique, full format and semi-automatic lengthwise and trans-
versely.
• Set the max. SID.
• Move the unit into the 35° position (for units with the lifting base, move into the 0° posi-
tion).
• Manually set a format of approx.18 x 24 cm (display on collimator).
• Place a side marker on the tabletop.
X • Release an exposure.
• Develop the films and label them (exposure data, date).

Evaluation:
• Draw in the center of the radiation field and the center of the film as shown in the illustra-
tion.
• Measure the difference between the radiation field center and film center.
¹ Max. admissible difference ≤8 mm
Correction:
• Check the centering of the tube unit/collimator to the center of the spotfilm device.

Accuracy of Automatic Format Collimation (reference to nominal formats) 0

The accuracy of automatic format collimation is checked for two formats with minimum
and (optional) maximum SID.

Siemens AG AXD3-350.815.04.03.02 Page 51 of 82 AXIOM Iconos R100


Medical Solutions 05.07 AX SCM W II
52 Beam Geometry

NOTE The upper tolerance limit (0.8% … 1.2%) has to be utilized for the
image quality of the subdivided exposures.

Preparation
• Set the unit out of the 0° position.
• Set the SID to 115 cm; set the focus-image intensifier distance rB-min = 115 cm.
• Select the ICONOS Spotfilm workstation.
• Place the 35 x 35cm cassette with film (or film envelope) in the center of the tabletop.
• Enter a preferred value (see adjustment).

Procedure
• Insert the following cassettes (without film) and release an exposure with the indicated
exposure data, one after the other:
18 x 24 cm cassette lengthwise; 44 kV; 3.20 mAs (70 kV; 40 mAs)
24 x 30 cm cassette lengthwise; 44 kV; 1.40 mAs (70 kV; 50 mAs)
• Set the SID to 150 cm (optional); set the focus-image intensifier distance rB-min = 150
cm.
• Place the 35 x 35cm cassette with film (or film envelope) in the center of the tabletop.
• Insert the following cassettes (without film) and release an exposure with the indicated
exposure data, one after the other:
18 x 24 cm cassette lengthwise; 44 kV; 4,50 mAs (70 kV; 40 mAs)
24 x 30 cm cassette lengthwise; 44 kV; 3.20 mAs (70 kV; 50 mAs)
• Develop the film, measure the radiation field width (b’) and length (l’) and record the val-
ues.

2
1

Fig. 24:
Pos. 1 The 2 format sizes to be checked
Pos. 2 Film

AXIOM Iconos R100 AXD3-350.815.04.03.02 Page 52 of 82 Siemens AG


05.07 AX SCM W II Medical Solutions
Beam Geometry 53

Evaluation
• Measure the width (b) and length (l) of the radiation fields in the image receptor plane
and record the values.
b = b’ x M; l = l’ x M; calculate the magnification factor M according to the formula
(Fig. 25 / p. 53) separately for min. and max. SID.
“a“ = distance between the measuring plane and the film plane.
“rb“ = SID in cm.

r
Magnification factor M = r ba
b

a = 8 cm

rB

Cassette with testfilm in measuring plane

Front panel
a
Cassette to be checked in film plane
(catapult bucky)

Fig. 25:
• Determine the percent difference between the radiation field (b) and the nominmal for-
mat (bN) and record this value.
Width (%) = [(b - bN)/rB] x 100
Length (%) = [(l - lN)/rB] x 100
• maximum permissible difference for width and length in %: ≤1,2%
P • Enter the value in the Quality Certificate or Test Certificate 1.

Adjustment

NOTE If there is a difference, perform the adjusment with only one SID
(e.g. 115 cm).

• Start the XCS service software.


• Select Components -> ICONOS R100 -> Adjustment -> Calibrations.
• Select the function Collimator -> Direct technique collimation, overframing on film.
• Close the message window with OK.

Siemens AG AXD3-350.815.04.03.02 Page 53 of 82 AXIOM Iconos R100


Medical Solutions 05.07 AX SCM W II
54 Beam Geometry

• Enter the overframing of the film format for the width and height in mm (0 ... 25mm);
always enter the same correction value for height and width. The value of e.g. 10 mm
entered corresponds to the displayed figure of 100.
An empirical value = 90
• Confirm with OK.
• Afterwards a window with the old and the new value is shown.
• Confirm by clicking on OK. The new values are now stored only in the unit.
• Exit the XCS SSW and switch the system off and on again to store the change perma-
nently in the XCU.
• After the adjustment, repeat the steps described under preparation, procedure and
evaluation.

AXIOM Iconos R100 AXD3-350.815.04.03.02 Page 54 of 82 Siemens AG


05.07 AX SCM W II Medical Solutions
Beam Geometry 55

ICONOS Tube Unit on the Bucky Wall Stand (Option) 5.3

Coincidence of Radiation Field and Center of Film 0

Fig. 26:
• Select "Bucky Wall Stand (BWS)”.
• Raise the unit into the 90° position or tilt it down in the direction of the Bucky wall stand.
• Set the unit to the min. SID (max. SID to the BWS).
• Adjust the tube unit so that it is centered to the BWS cassette using the light localizer.
• Mark the center of the film with a washer.
• Insert a 24x30 cm or 10x12 inch cassette sideways into the Bucky wall stand.
• Manually set a format of 18 x 24 cm (display on collimator).
• Place a side marker on the Bucky wall stand.
• Release an exposure.
Exposure parameters (e.g. for a 200 screen system): 42 kV, 8 mAs, 80% kW, large
focus.
• Develop the films and label them (exposure data, date).

Evaluation
• Mark the center of the radiation field and the center of the film (Fig. 26 / p. 55).
• Measure the difference between the radiation field center and film center.
¹ Max. admissible difference ≤1.2% of the SID
P • Enter the difference from the center in the Quality Certificate or Test Certificate 1.
• Label and archive the film.

Siemens AG AXD3-350.815.04.03.02 Page 55 of 82 AXIOM Iconos R100


Medical Solutions 05.07 AX SCM W II
56 Beam Geometry

Correction:
• If there is a difference in height:
The tube unit has not been correctly set to the center of the Bucky wall stand by the
operator. Repeat the exposure.
• If there is a difference in the transverse direction:
Correct the BWS position (move laterally).

AXIOM Iconos R100 AXD3-350.815.04.03.02 Page 56 of 82 Siemens AG


05.07 AX SCM W II Medical Solutions
IONTOMAT 57
6-
Procedure only for Systems in which DLR Cassettes are Used
6IONTOMAT

The following values can be entered or transcribed from the data sheet for the DLR cas-
settes into Test Certificate 1 or the "Quality Certificate” only after discussing this with the
manufacturer.
• Screen Types and Film Manufacturers
• Color Sensitivity
• Sensitivity, S
• Dose Requirement, KS
• Minimum resolution
P Enter the values into Test Certificate 1.

IONTOMAT Adjustment 0

Unlike with normal film-screen cassettes, the IONTOMAT can be adjusted only with the
cutoff dose in DLR systems.

IONTOMAT Sensitivity 0

• Select the work station


• Set the standard SID or the default SID.
• Place 20 cm, ± 1 cm water in the beam path.
• Collimate to the water phantom.
• Insert the dose measurement chamber into the cassette holder and collimate slightly
larger than the IONTOMAT chamber size.
• Select the exposure parameters:
- 81 kV, 80%, small focus, middle IONTOMAT measuring field.

Possible Adjustment of the Sensitivity (Cutoff Dose):


In the XCS SSW program, select "Components/ Polydoros/ Adjustment/ Iontomat Sensi-
tivity".

NOTE The data can be updated in the XCU with "Transfer Value" and
control exposures can be made without having to exit the SSW.

• Depending on the exposure step (H, U, D), trigger an exposure.


Required: Set the cutoff dose to the manufacturer's value.
The following values must be used only if there are no dose values from the manufac-
turer:

Siemens AG AXD3-350.815.04.03.02 Page 57 of 82 AXIOM Iconos R100


Medical Solutions 05.07 AX SCM W II
58 IONTOMAT

Exposure step 1 (H) -> KB = 2.0 µGy, Tol. ± 10%


Exposure step 2 (U) -> KB = 3.7 µGy, Tol. ± 10%
Exposure step 3 (D) -> KB = 7.5 µGy, Tol. ± 10%

KB = cutoff dose in the image receptor plane

IONTOMAT Voltage Response Correction 0

Enter the following values in XCS SSW program "Competent/ Polydoros/ Adjustment/
Voltage Response Correction”.

kV values < 81 kV, curve 3


kV values > 81 kV, curve 10

Plani-IONTOMAT(for DLR cassettes) 0

No correction needs to be entered.

Cutoff Dose and Resolution 0

For the procedure, see the IONTOMAT PN

Function of the Measurement Fields 0

For the procedure, see the IONTOMAT PN

AXIOM Iconos R100 AXD3-350.815.04.03.02 Page 58 of 82 Siemens AG


05.07 AX SCM W II Medical Solutions
IONTOMAT 59

Film-Screen Combinations Used 6.1

Three different exposure steps (BSt) corresponding to three different film-screen combi-
nations (H, U, D) can be programmed at the generator.
For each exposure step, the screen and the film provided for it must be documented with
the following data from the manufacturer in the "Quality Certificate":
- Color sensitivity of the films and screens
- Sensitivity, S
- Theoretical dose requirement, KS
- Minimum resolution, RGr
Make sure that only films for the light color for which the screen is sensitivity has been
determined are used (green sensitive film for green sensitive screen, blue sensitive film for
the blue screen).
There are no international, uniform regulations regarding use of the film-screen combina-
tions; national regulations (e.g §16 RöV) must be observed in the particular country.
Screens and Film Manufacturers/Types
P Enter the value in the "Quality Certificate”.
Color Sensitivity of the Films and Screens
P Enter the value in the "Quality Certificate”.
Sensitivity, S
P The sensitivity, S (speed), specified by the manufacturer must be recorded for each
film-screen combination.
As a rule (in some countries, this is specified), this is the S value on the cassettes (e.g. S
= 200, S = 400) or it must be taken from the data sheets for the film-screen combinations.
The sensitivity of a film-screen combination is defined according to ISO 9236 as the quo-
tient of 1000 µGy and the air kerma (dose) KS required to achieve an optical density
(blackening) of 1 above base fog.

1000
S = ------------ µGy
KS

Fig. 27:
Dose Requirement, KS
P With a known sensitivity, S, calculate and document the theoretical dose requirement KS.

Siemens AG AXD3-350.815.04.03.02 Page 59 of 82 AXIOM Iconos R100


Medical Solutions 05.07 AX SCM W II
60 IONTOMAT

1000
K S= ------------ µGy
S
Fig. 28:
Minimum Resolution, RGr
A minimum resolution RGr in LP/mm is assigned to each required dose KS, i.e. the higher
the required dose is for a film-screen combination, the higher the resolution that must be
reached in order to justify the increased dose.
P The particular minimum resolution, RGr must be taken from the illustration for each
film-screen combination used and must be documented.

Fig. 29:

KS = Dose requirement of the film-screen combination per DIN 6867


RGr = Minimum resolution of the film-screen combination per DIN 6868
S (Speed) = Sensitivity of the film-screen combination (ISO 9236)

AXIOM Iconos R100 AXD3-350.815.04.03.02 Page 60 of 82 Siemens AG


05.07 AX SCM W II Medical Solutions
IONTOMAT 61

IONTOMAT Sensitivity 6.2

• Select the workstation:


- ICONOS Spotfilm Device or Bucky W.S. (option)
• With "ICONOS Spotfilm", set an SID of 115 cm.
Set the corresponding SID for the Bucky W.S.
X • Place 20 cm ± 1 cm water in the beam path. Collimate to the water phantom.
• Exposure parameters:
81 kV, 80%, small focus, exposure step H, U or D, middle Iontomat measuring field.
• In the XCS SSW program, select "Components/ Polydoros/ Adjustment/ Iontomat Sen-
sitivity".
- Select the IONTOMAT chamber.
- Enter the sensitivity values for H, U, D according to the film-screen combinations
being used.
Recommended Values for the Basic Setting:

Screen Sensitivity
(H) 800 8
(U) 400 11
(D) 200 14

NOTE The data can be updated in the XCU with "Transfer Value" and
control exposures can be made without having to exit the SSW.

Adjustment with 20 cm Water:


X • Make test exposures for all screens and work stations using 20 ± 1 cm water (film for-
mat ≥ 18 x 24 cm) and check the film density of each.
- The films must have the optical density that the customer desires.
- If there is no specific customer requirement, adjust to a net optical density, DN = 1
(density of 1.0 + base fog).
- If necessary, correct the sensitivity value for H, U, D.
P • Enter the values into Test Certificate 1.
• Label and archive the films.

NOTE If the specified values are not reached, drift and hum of the Ion-
tomat chamber must be checked.

Siemens AG AXD3-350.815.04.03.02 Page 61 of 82 AXIOM Iconos R100


Medical Solutions 05.07 AX SCM W II
62 IONTOMAT

Checking the Drift (only in a malfunction situation) 0

NOTE The generator must have been in operation for at least 5 minutes
before the test is performed.

• Connect the oscilloscope to D100.X63.VION and D100.X63.AGND; trigger on


D100.X64.SWR
• Cover the collimator with lead and switch off the I.I. power supply.
• Load a blank cassette: 35 x 35 cm
• Set SS (S1) to ON
• Select Iconos spotfilm device.

NOTE If a digital oscilloscope is used for the measurement, do not mea-


sure in the "glitch detect mode" (mean value calculation, smooth-
ing), since this will distort the measuring results.

• In the XCS SSW program, select "Components/ Polydoros LX or SX/ Diagnostic/ Ion-
tomat Drift".
• Perform the drift measurement for the middle measuring field per the information in the
SSW.
Drift: 0, ± 2 V/0.5 s

example 1
Oscillogram 1:

Drift on
example 2 D100.X63.VION
Toleranz
1,5 V +/- 2V 1 V/T

example 3 0,1 s/T


0V
DC

Fig. 30:
• Close the window with "OK" and exit the XCS SSW.
• Remove the lead and switch the I.I. power supply back on.
P • Enter the value in the Quality Certificate or Test Certificate 1.

AXIOM Iconos R100 AXD3-350.815.04.03.02 Page 62 of 82 Siemens AG


05.07 AX SCM W II Medical Solutions
IONTOMAT 63

Checking the Hum Voltage (only in a malfunction situation) 0

• Connect the oscilloscope to: D100.X63.DLIN and D100.X63.AGND; trigger on


D100.X64.SWR
• Set the S3 switch to Normal.
• Set SS (S1) to OFF
• Measure the hum voltage for all measurement fields without fluoroscopy.
Max. permissible hum voltage ≤20 mVss

Oscillogram 2:

Uhum Hum voltage on


D100.X63.DL_IN
0V
10 mV/T

100 ms/T

Fig. 31:
P • Enter the value in the Quality Certificate or Test Certificate 1.

Siemens AG AXD3-350.815.04.03.02 Page 63 of 82 AXIOM Iconos R100


Medical Solutions 05.07 AX SCM W II
64 IONTOMAT

IONTOMAT Voltage Response Correction 6.3

• Select the workstation:


- ICONOS Spotfilm, Vertix PRO or 3D -Vertix PRO
• Set a 115 cm SID for "ICONOS Spotfilm". Set the corresponding SID for the Vertix
PRO.
X • Place 20 cm ± 1 cm water in the beam path. Collimate to the water phantom.
• Exposure parameters:
60 kV, 80%, small focus, exposure step H, U or D, middle Iontomat measuring field.
125 kV, 80%, small focus, exposure step H, U or D, middle Iontomat measuring field.
• In the XCS SSW program, select "Components/ Polydoros/ Adjustment/ Voltage
Response Correction".
- Select the detector type.
- Determine and enter the sensitivity curves for H, U, D according to the film-screen
combinations being used.

Value Ranges: < 81 kV: 0...13;


≥81 kV: -13...0

NOTE The data can be updated in the XCU with "Transfer Value" and con-
trol exposures can be made without having to exit the SSW.

X • Make test exposures at 60 kV and 125 kV using 20 cm, ± 1 cm water for each screen
used and check the film density of each exposure.
- The film density at 60kV and 125kV must match the film density at 81kV (Section
"IONTOMAT Sensitivity").
- The difference in density can be
Dopt (81 kV) - Dopt (60 kV) ≤ 0.2
Dopt (81 kV) - Dopt (125 kV) ≤ 0.2
of the optical density.
- If necessary, select other correction curves for the upper or lower kV range.
• If a Vertix PRO (option) is configured, the adjustment must also be performed for this
unit.
P • Enter the measured values in the "Quality Certificate or Test Certificate 1".
• Exit the SSW screen with "OK".
Correction curves 1...13 for the lower kV range (40...81 kV) and correction curves 1...13
for the upper kV range (81...150 kV) are displayed in the following illustration and the cor-
responding value table.
From this it is possible to read out what change in exposure points or in density results
when there is a switch to a different correction curve.

AXIOM Iconos R100 AXD3-350.815.04.03.02 Page 64 of 82 Siemens AG


05.07 AX SCM W II Medical Solutions
IONTOMAT 65

NOTE A change of 1 exposure point results in a change in density of


approx. ∆ S = 0.25.

Recommended values: +/- 8 exposure points.

Fig. 32:

Siemens AG AXD3-350.815.04.03.02 Page 65 of 82 AXIOM Iconos R100


Medical Solutions 05.07 AX SCM W II
66 IONTOMAT

(≤81 kV) Corrections for the lower kV range (≤81 kV)


kV 13 12 11 10 9 8 7 6 5 4 3 2 1
40 48 42 35 32 28 24 20 16 13 11 8 6 3
41 47 41 34 31 27 23 19 15 13 10 8 6 3
42 46 40 33 30 26 23 19 15 12 10 8 5 3
44 44 38 31 28 25 21 17 13 11 9 7 5 2
46 41 35 30 26 23 20 16 12 10 8 7 4 2
48 39 33 28 25 22 18 15 11 10 8 6 4 2
50 37 31 26 23 20 17 14 10 9 7 5 4 2
52 34 29 24 21 18 15 13 9 8 6 5 3 2
55 30 26 21 18 16 13 11 8 7 6 4 3 1
57 28 23 19 17 14 12 9 7 6 5 3 2 1
60 24 20 16 14 12 10 8 6 5 4 3 2 1
63 20 17 13 11 10 8 6 5 4 3 2 2 1
66 16 13 10 9 8 6 5 4 3 2 2 1 1
70 11 9 7 6 5 4 3 3 2 2 1 1 0
73 8 6 5 4 3 3 2 2 1 1 1 0 0
77 3 3 2 2 2 1 1 1 1 0 0 0 0
81 0 0 0 0 0 0 0 0 0 0 0 0 0
85 -2 -1 -1 -1 -1 -1 -1 -1 0 0 0 0 0
90 -4 -3 -3 -2 -2 -2 -1 -1 -1 -1 -1 0 0
96 -7 -6 -6 -4 -4 -3 -3 -2 -2 -1 -1 -1 0
102 -10 -8 -8 -6 -5 -4 -4 -3 -2 -2 -1 -1 0
109 -13 -11 -11 -9 -7 -6 -5 -4 -3 -2 -1 -1 -1
117 -18 -15 -15 -11 -10 -8 -6 -5 -4 -3 -2 -2 -1
125 -24 -20 -20 -14 -12 -10 -8 -6 -5 -4 -3 -2 -1
133 -31 -25 -25 -17 -15 -12 -10 -7 -6 -5 -4 -2 -1
141 -40 -32 -32 -21 -18 -15 -11 -8 -7 -6 -4 -3 -1
150 -53 -40 -40 -26 -21 -17 -14 -10 -8 -7 -5 -3 -2
kV -13 -12 -11 -10 -9 -8 -7 -6 -5 -4 -3 -2 -1
(> 81 kV) Corrections for the Upper kV Range (> 81 kV)

EP = exposure points
∆EP are listed in the table in steps of 1/8 EP.

AXIOM Iconos R100 AXD3-350.815.04.03.02 Page 66 of 82 Siemens AG


05.07 AX SCM W II Medical Solutions
IONTOMAT 67

Cutoff Dose and Resolution 6.4

The cut-off dose must be measured at each application unit for IONTOMAT P for all expo-
sure steps (screen) used.
Record in Test Certificate 1. The values can then be transcribed for the §16 RöV Accep-
tance Test.
Test Conditions:
• Select the workstation:
- ICONOS Spotfilm, Vertix PRO or 3D -Vertix PRO
• Set a 115 cm SID for "ICONOS Spotfilm".
Set the corresponding SID for the Vertix PRO.
P Record the SID that is set (Quality Certificate or Test Certificate 1).
• Move the scatter radiation grid into the beam path.
• Set the density correction to ± 0 exposure points.
• Do not select any additional filtering at the collimator.
• Test object:

Area of applicability for §16 RöV Other countries


Place a 25 mm Al test phantom (mea- Insert 2,1 mm Cu in the collima-
surement stand) on the tabletop in the tor.
beam path.

• Exposure parameters:
77 kV, small focus, 80%, exposure step H, U or D, middle Iontomat measuring field.
Record the kV value (Test Certificate 1).
Procedure:
• Connect the mAs meter
• Dali, Nomex:
Place the dose measurement chamber on the tabletop, centered to the middle Ion-
tomat measuring field.
Diados:
Place the dose detector on the tabletop, outside the middle Iontomat measuring field.
Set the filter on the Diados to 23.5 mm Al.
• Place resolution test, Type 42 on the tabletop, turned 45° to the patient axis, outside the
middle Iontomat measuring field.
X • Trigger an exposure for all H/U/D exposure steps and application units (spotfilm device,
Vertix PRO), trigger exposure and perform the following measurements:
- Measure the cutoff dose KT.
- Determine the resolution, Rg of the film.

Siemens AG AXD3-350.815.04.03.02 Page 67 of 82 AXIOM Iconos R100


Medical Solutions 05.07 AX SCM W II
68 IONTOMAT

• Calculate the cut-off dose KB.

Fig. 33:

KT Cut-off dose measured on the table


KB Cut-off dose in the image receptor plane
m Equipment attenuation factor (see Test Certificate 1)
fK = 0.94 Correction factor for Dali, Nomex measuring chambers
fK = 1,0 Correction factor for Diados detector

• Resolution, Rg
¹ Rg must be ≥ RGr for each film-screen combination.

NOTE If the minimum resolution, RGr is not reached, check the screen
for damage and for soiling and the cassette for correct flattening
of the screen.

P • Record all values that have been determined in the "Quality Certificate or Test Certifi-
cate 1".
• Label and archive the films.

AXIOM Iconos R100 AXD3-350.815.04.03.02 Page 68 of 82 Siemens AG


05.07 AX SCM W II Medical Solutions
IONTOMAT 69

Function of the Measurement Fields 6.5

• Select the workstation:


- ICONOS spotfilm, Vertix PRO (option)
• With "ICONOS Spotfilm", set an SID of 115 cm.
Set the corresponding SID for the Vertix PRO.
• Set the density correction to ± 0 exposure points.
• Do not select any additional filtering at the collimator.
• Insert a cassette with a screen corresponding to exposure step D.
• Test object:

Area of applicability for §16 RöV Other countries


Place a 25 mm Al test phantom (mea- Insert 2.1 mm Cu in the collima-
surement stand) on the tabletop in the tor.
beam path.

• Exposure parameters:
77 kV, small focus, 80%, exposure step D
X • Make an exposure for all application units (spotfilm device, Vertix PRO) using all three
Iontomat measuring fields.
• Determine the mAs value.
∆ mAs left measuring field = mAs middle measuring field mAs left measuring field
∆mAsright measuring field = mAsmiddle measuring field mAsright measuring field
¹ Tolerance: ∆mAs ≤25%
P • Record the values in the "Quality Certificate or Test Certificate 1".
• Label and archive the films.

Siemens AG AXD3-350.815.04.03.02 Page 69 of 82 AXIOM Iconos R100


Medical Solutions 05.07 AX SCM W II
70 Tomo (Direct Technique)
7-
Checking the Tomo Device
7Tomo (Direct Technique)

Tomo Height / Resolution 0

Fig. 34: Tomo block

NOTE For room heights of < 285 cm, Tomo can be triggered only in unit
positions between - 5° to + 5° and + 87° to + 90°.
It is not generally possible to perform tomography under -5°.

• Set an SID of 115 cm.


• Insert 0.6 mm Cu prefiltering into the collimator.
• Move the grid into the beam path.
• Insert a 24x30 cm or 10 x 12 inch cassette.
• Select the 2-part subdivision and tomo at 40° / 2 sec.
• Exposure parameters:
- approx. 50 kV, 40 mAs, 80%, small focus.
• Set a tomo height of 116 mm (display on the ICONOS remote control console).
X • Set the tomo focus test in the tomo block to 116 mm.
• Center the tomo block in the beam path and collimator to the tomo block.
- The focus test must be positioned on the anode side, transversely to the table longi-
tudinal axis.
- The notch in the focus test must be in the central beam.
X • Trigger the 1st tomo exposure at 40°/2 sec.
• 2. Trigger the tomo exposure at 8°/0.8 sec.

AXIOM Iconos R100 AXD3-350.815.04.03.02 Page 70 of 82 Siemens AG


05.07 AX SCM W II Medical Solutions
Tomo (Direct Technique) 71

• Position the column in the 0° position.


• Select the system spotfilm device (deselect tomo).
X • Insert a new cassette and trigger exposure.
• Develop the films and label them (exposure data, date).

Evaluation
¹ Tomo height
Tomo height display on the console, SHa - tomo height on the film, SH = ∆
SHa (≤ 5 mm)
¹ Resolution of the Tomo Exposure
Difference between the Bucky and tomo exposure ≤ 1 line group
Resolution limit value for the tomo exposure ≥ 2.0 Lp/mm

Correction
• Start the XCS SSW.
• Start Components -> ICONOS R100 -> Adjustment -> Calibrations .
• Select the Tomo -> Tomo height console display function.
- Default value in the "Old offset value" box = 90.
• Enter a value for the correction.
- 1 mm corresponds to the displayed number 10; if the tomo height is too high, correct
with “+”, if the tomo height is too low, correct with “-”.
• Confirm with “OK”.
¹ A window with the new and the old values will be displayed.
• Confirm with “OK”.
¹ The value will be saved.
• Exit the XCS SSW.
• Switch the system off and back on.
X • Make a new tomo exposure.
• Remove the 0.6 mm Cu from the collimator.
P • Enter the values determined into Test Certificate 1.
• Label and archive the films.

Siemens AG AXD3-350.815.04.03.02 Page 71 of 82 AXIOM Iconos R100


Medical Solutions 05.07 AX SCM W II
72 VIDEOMED DIC
8-
List of IQ Documents
8VIDEOMED DIC

NOTE The monitor values must be checked and, if needed, set and
adjusted usin the IQ Test Procedure.

Perform monitor adjustment and the image quality test according to the following docu-
ments:
• AXIOM Iconos R100, Quality Assurance, Test Procedure for Image Quality with-
out DFR, AXD3-350.820.01.01.xx
• AXIOM Iconos R100, Quality Assurance, Test Procedure for Image Quality with-
out DFR, AXD3-350.820.02.01.xx
P • Fill out the general data in the IQ Test Certificate.
P • Enter the System Serial Number on each page of the IQ Test Certificate to ensure there
is clear allocation of each certificate page.
• For detailed test descriptions, see the IQ Test Instructions.

AXIOM Iconos R100 AXD3-350.815.04.03.02 Page 72 of 82 Siemens AG


05.07 AX SCM W II Medical Solutions
Startup of Other Options 73
9-
Starting up the KermaX Dose Measurement Meter
9Startup of Other Options

NOTE If a KermaX dose measurement meter was retrofitted (per the cus-
tomer order), startup per the document: AX00-100.814.01.01.xx,
(AX00-100.814.01 / KermaX with external Display) must be per-
formed.

P • Record the measured value per the enclosed "Measurement Certifi-


cate".(AX00-100.814.01 / Measurement Certificate)

Function Check 0

Fluoroscopy
• Select the tube unit.
• At the generator console, reset the cGy/cm2 display with the "0 set button”.
• Insert 2.1 mm Cu.
• Completely open the collimator.
• Set the max. SID.
X • Trigger fluoro for 30 sec.
• The value of the area dose product (cGy/cm2) displayed on the console or monitor must
increase continuously.
• Remove the 2.1 Cu from the beam path.
Exposure
• Select the tube unit.
• At the generator console, reset the cGy/cm2 display with the "0 set button”.
• Set 70 kV and 10 mAs.
X • Release an exposure.
• Read and make a note of the area dose product (cGy/cm2) displayed on the console.
• Set 70 kV and 5 mAs.
X • Release an exposure.
¹ The value read on the console must be half of the above value.
• Set 70 kV and 20 mAs.
X • Release an exposure.
¹ The value read on the console must be double the above value.

Siemens AG AXD3-350.815.04.03.02 Page 73 of 82 AXIOM Iconos R100


Medical Solutions 05.07 AX SCM W II
74 Startup of Other Options

Data Printer 9.1

Two printer models can be connected. EPSON LQ 570 and Star DP8340S.
For the Star DP8340S printer model, the DIP switches must be set as follows:

Fig. 35:
Polydoros SX:
• In "Configure:Site structure:Component selection", integrate the priner connection in
the system list with "Data printer com2" .
• Select the connected printer under "printer type" (e.g. Star DP8340S).
• Select when the data are to be printed in "print selection" (the print routine must be trig-
gered using the RESET button for the following check).
• Switch on the data printer and insert paper.
X • Trigger fluoro for 5 sec.
• Make a note of the kV, mA, sec. values displayed on the generator console.
• Activate the printer with the Reset button.
• Compare the values on the label with the values noted from the generator console.
X • Release an exposure.
• Compare the values on the label with the values displayed on the generator console.
• Make sure that both labels are printed in the vertical direction and are in the middle.
• For further descriptions, see the documents included in the data printer shipment.
Polydoros LX:

NOTE The data printer is connected directly to the external KermaX disi-
play and does NOT have to be configured in the Site Structure.
Only dose values are printed out.

• Switch on the data printer and insert paper.


X • Trigger fluoro for 5 sec.
• Activate the printer with the Reset button.
• Compare the values on the label to the KermaX display.
X • Release an exposure.
• Compare the values on the label to the KermaX display.
• Make sure that both labels are printed in the vertical direction and are in the middle.
• For further descriptions, see the documents included in the data printer shipment.

AXIOM Iconos R100 AXD3-350.815.04.03.02 Page 74 of 82 Siemens AG


05.07 AX SCM W II Medical Solutions
Customer-specific Settings 75
10-
Customer-specific Settings
10Customer-specific Settings

Changing the Tilt End Position (+88 to + 90° Position) 0

• If a BWS is configured, the + 90° end position should not be changed.


• If the factory setting is changed at the customer's request (e.g.), perform the adjust-
ment of the end positions again.

NOTE If tomo is configured, the unit stop position may not be set < 88°,
otherwise tomo will be blocked with room heights of < 2.85 m or it
will not be possible to lower the tabletop.

Correction:
• In the XCS SSW,
- click on < Components >
- < ICONOS >
.
• In the ICONOS SSW,
- click on < Adjustment >
- < Calibrations >
- < Table tilt >
- < limitation of + 90° position >
- < Execute >
.
Additional steps are menu-prompted.

NOTE Following completing of all unit adjustments, there is a return to


the XCS service application program with "Cancel".

• Check the function of the tilt end position:


- Move the unit into the 70° position.
- Pull in the tabletop to ± 0.
- Move in the + 90° position direction at max. speed.
¹ The unit must stop at the previously set angle.

Only at the customer's request: Programming the Grid Movement 0

(recommendation for DLR exposures due to the moire pattern)


• <Select Configure/Site Structure>.
• Select Iconos in the “selected” box and click on ”Option”.

Siemens AG AXD3-350.815.04.03.02 Page 75 of 82 AXIOM Iconos R100


Medical Solutions 05.07 AX SCM W II
76 Customer-specific Settings

• “Select Reversable Grid” available.


• Confirm all windows with “OK”.
• Switch the system off and back on.
If needed, the grid delay time can be set in the Iconos SW under <Adjustment/ Calibra-
tions/ Spotfilm device/ Grid: Delay time>. See the Online Help file.

Programmiong the Light Key on the Remote Control Console 0

Depending on the customer's request: Switching on the collimator light localizer is also
possible from the ICONOS remote control console (light key). Switching off is not possi-
ble.
• To switch on the light localizer: In the ICONOS remote control console, move the
jumper on the D41 from XP4 to XP12
or
• To switch on the room light: In the ICONOS remote control console, move the jumper
on the D41 from XP4 to XP12.

Changing the Fluoroscopy Field Limitation 0

P • If the collimator plates are adjusted larger at the customer's request, determine the
amount of over-framing and record this in Quality Certificate 1.
Workflow
• Unit in 0°position. Set an SID of 115 cm Select I.I. full format.
• Place the centering cross on the tabletop; when doing this, set the X axis in the tabletop
transverse direction.
• Switch fluoroscopy ON. Collimate a radiation field of approx. 4 x 4 cm and position the
centering cross so it is centered to the radiation field.
• Open the collimator all the way. Starting from the center of the centering cross, going in
the width and length directions, read the cm division of the four X1, X2, Y1, Y2 coordi-
nates displayed on the monitor and record the values. Perform this for all I.I. formats
and all SIDs.
• In pace of the centering cross, place a cassette with film (at least 35 x 35 cm or 14 x 14")
turned 45° on the tabletop and center it to the light field. Trigger fluoroscopy with the col-
limator open all the way for all I.I. formats until sufficient film density is achieved. Make a
test film for each SID, whereby all 4 I.I. formats are visible. Evaluation is possible
because of the different film blackening of the individual formats. Make a test film for
each SID.
• Measure the width (X) and length (Y) of the radiation fields on the test films and record
the values.
Admissible Overframing of the I.I. Formats in the Centering Cross Plane:
SID 115 cm = 2.1 cm
SID 150 cm = 2.8 cm

AXIOM Iconos R100 AXD3-350.815.04.03.02 Page 76 of 82 Siemens AG


05.07 AX SCM W II Medical Solutions
Customer-specific Settings 77

For Germany only; Acceptance Test per §16 RöV:


P • Use the values entered in the Test Certificate for Fluoroscopy Field Limitation and the
test films for the Acceptance Test Certificate.

Fig. 36:

Siemens AG AXD3-350.815.04.03.02 Page 77 of 82 AXIOM Iconos R100


Medical Solutions 05.07 AX SCM W II
78 Final Work Steps
11-
Required Action
11Final Work Steps

Cover Panels, Covers 0

• Install all covers and cover panels, particularly the lateral I.I. cover panel
(1/Fig. 37 / p. 78).

Fig. 37: Side I.I. cover panel


• Check all system functions in a test operation for correct operation.
• Then perform a visual check of the sytsem, wipe down the system and touch up any
paint scratches or chips.

AXIOM Iconos R100 AXD3-350.815.04.03.02 Page 78 of 82 Siemens AG


05.07 AX SCM W II Medical Solutions
Final Work Steps 79

Check of the Ground Wires 0

Unit
Table
Ceiling rails
Monitor rackl
Less than 0.2 Ohm Monitors
Cabinets
Transformers
Consoles
I.I.
Generator / X-ray tube unit
Power Ground wire test instrument
distributor
Collimator

Fig. 38:
• Power to the system must be switched off.
• Measuring device:
- Ground wire meter
• Measurement value:
- Max. 0.2 Ohm (observe country-specific regulations!)
Exceptions: Accessory rails on the collimator, tabletop and control console.
• Test procedure:
- Measure between all exposed conductive parts of the system and the ground wire
bar in the generator.
• Recording:
P - Enter the measurement value into the "Quality Certificate".

Other Adjustments 0

• Enter the time and date in the XCU in the menu System /Realtime Clock.

Clearing the Exposure Counter 0

• In the XCS SSW program, select "Info/Operating Time".


• Select the tube unit, select "Erase tube" and confirm the message that follows
Answer "Erase tube data?" with "Yes".

Siemens AG AXD3-350.815.04.03.02 Page 79 of 82 AXIOM Iconos R100


Medical Solutions 05.07 AX SCM W II
80 Final Work Steps

Backup (Saving Data on Diskette)

NOTE After completing startup, perform a "Backup".

• Insert the backup diskette into drive A: .


• Select"Date/ Backup to disk" in the XCS SSW program and start the backup with "OK".

Erasing the XCS Error Log 0

• In the XCS SSW program, select "Diagnostic/Error Log".


• Select "Get from unit".
• Select "Delete in unit".
- Save the error log on diskette.
At the beginning of system operation, read out the log; it may be useful for compara-
tive purposes.
• Close the error log menu with "Close".
• Exit the XCS SSW program with "System/Quit".

Completing the Certificates, Reporting and Archiving 0

NOTE After completing all adjustment work steps and check measure-
ments, the responsible technician must sign and date the certifi-
cates (System, Unit Certificates, Generator Certificate, IQAP,
Installation Certificate, Startup Certificate) and the country-spe-
cific certificates (e.g. DHHS, RöV §16) to confirm that all values
have been correctly determined and recorded.

NOTE To obtain a complete and error-free DHR (Device History Record),


it is absolutely required to return these documents (certificates)
that are generated or completed during installation and startup by
fax or as a copy to the factory!
They are archived there.

SIEMENS Shanghai
Medical Equipment Ltd.
278 Jin Ha Rd. Shanghai
PRC, 201206

AXIOM Iconos R100 AXD3-350.815.04.03.02 Page 80 of 82 Siemens AG


05.07 AX SCM W II Medical Solutions
Final Work Steps 81

Labels 11.1

• Warning labels in the national language


If warning labels are available as adhesive labels in the national language (e.g. in the
System Owners Manual or even with the accessories), the appropriate warning label
must be placed over the existing one on the corresponding component. When dong
this, make sure that the label has proper adhesion.
• Additional labels
If required according to country-specific requirements, additional labels will be shipped
in the System Owners Manual in Register 9. These labels are to be affixed, as needed,
according to the instructions in Register 9. When dong this, make sure that the labels
are clearly visible and have proper adhesion.

Siemens AG AXD3-350.815.04.03.02 Page 81 of 82 AXIOM Iconos R100


Medical Solutions 05.07 AX SCM W II
82 Changes to Previous Version
12Changes to Previous Version 12-

Chapter Section Change


General Startup Power line connection Check of the rotational field is skipped.
Checked by the direction of movement
of the patient table.
Service Software Site Structure Site Structure, Correction of Errors
Service Software Check and Adjustment of A data printer for LX cannot be config-
Basic Unit Data, Site Struc- ured in the Site Structure.
ture
Beam Geometry Checking the Collimator Added.
Position
Determining the Center of
the Spotfilm Device / Center
of Monitor
Camera Rotation
Central Beam
Center of Radiation Field /
Center of Spotfilm Device
Adjusting the Centering
Cross
Installation and Startup of Compression Cabling Laying cable and cable length added.
Compression
Startup of Other Options Data Printer Distinction between Polydoros SX and
LX
Remote Service Remote service via modem removed.
Final Work Steps Completing the Certificate Address changed.
Final Work Steps Labels New label regulations for Canada.

Training Requirements
Due to significant changes in this document / new publication of this document, user training is
required.
Due to slight changes in this document, user training is not required.

AXIOM Iconos R100 AXD3-350.815.04.03.02 Page 82 of 82 Siemens AG


05.07 AX SCM W II Medical Solutions