You are on page 1of 1

Manual Pre‐set procedure for KV ‐  Summit / Americomp 325.

adjust the kVp selector switches to result in 220 VAC indicated on your AC voltmeter.

Compare the kVp meter indications for all mA Re-set kVp selectors to result in voltmeter
stations with the values given below: indication of 160 VAC.
mA kVp METER mA kVp METER
Station Needle (kVp) Station Needle (kVp)
50 115 50 81
100 114 100 80
150 110 150 76
200 103 200 69
300 87 300 53

KV Compensation Kit instructions as revised in 2012 