RIVIERA LP PLUS

PRIMAX INTERNATIONAL August 2020 – Data may vary without notice Page • 1
The general radiographic system RIVIERA LP PLUS, designed to provide maximum flexibility and clinical
productivity in general trauma, chest, urology, abdomen and emergency applications, allowing both AP,
lateral and oblique projections with or without a table on wheels.

U-Arm design maintains constant alignment between the X-ray beam and the image detector,
regardless the tilt positions and the image receptor angle. Its extraordinary flexibility makes the system
ideal for all patients in standing, sitting or laying position, including those who are disabled or physically
restricted. The total integrated system combines modern technology of the High Frequency X-Ray
Generator with a fully motorized in all its movements, Variable SID, Anti-crushing system and two
speeds for fine adjustment all designed to give a user-friendly operation system and easy positioning
of the patient.

Total Height of the Column 224 cm


Maximum Height (Under-Table Position) 265 cm
Maximum Height (of positioner) 277.5 cm
Maximum Distance from Focal Spot of X Ray Tube to Floor 169 cm
Motorized Movements
◦ Vertical Travel 126.5 cm
◦ Source-Image Distance (SID Variable) From 100 to 180 cm
Rotation Movements
∙ Rotation of Swivel Arm +120°/-30°
∙ Rotation of Receptor ±45°
∙ Rotation of Tube-Collimator Assembly (Manual) ±180°

Column Fixing Floor-Wall


Brakes Electromagnetic brakes (in all movements)
System & Movements Counterbalanced in all movements

PRIMAX INTERNATIONAL August 2020 – Data may vary without notice Page • 2
MOVEMENTS and CONTROLS

Movement can be operated via different controls:


◦ touch screen of the Control Panel and peripheral buttons
◦ keyboard on the detector bucky
◦ Remote Control.

All the following movements can be possible:


◦ Source-Image Distance (SID).
◦ Vertical Travel.
◦ Swivel Arm Rotation and detector angulation.
◦ Automatic pre-positioning for Thorax Position and Under-table Position.

POSITIONING SCREEN PROGRAM SCREEN GENERATOR PARAMETERS


for auto-positioning SCREEN

*AUTO-POSITIONING: Position control with 10 programs for the most common exposure positions: 1. Select
the program nº with the Up and Down buttons. 2. Press “Select”, the screen changes to Positioning screen
(with the picture of the selected Programmed Position). 3. Press and hold the “Move” button from the Remote
Control for auto-positioning.
** MODIFYING PROGRAMS: From this screen, inserting a password, an already saved Program can be modified
for operator convenience. The SID, Arm Angle, Arm Height and Detector Angle of a program can be customized
and saved for further use.

PRIMAX INTERNATIONAL August 2020 – Data may vary without notice Page • 3
OTHER FUNCTIONS:
SID: Two buttons to increase or decrease the SID, Motorized movement.
Release them to lock in position. Detents at 100 cm and 180 cm are
indicated on Display. When one of these positions is reached, the
movement is blocked.

The SID position is continuously indicated in the Touch Screen Console.


When the Tube Assembly has reached a limit, a message appears on the
Screen.

VERTICAL MOVEMENT: Two buttons to control the vertical motorized


movements of Swivel Arm movements. Release them to lock in position.
The Swivel Arm height, from the Center of the Detector axis to floor, is
continuously indicated in the Touch Screen Console. When the U- Arm has
reached to the highest or lowest vertical position a message appears on
the Screen.

SWIVEL ARM ROTATION: Buttons to enable clockwise or counterclockwise


motorized rotation of the Swivel Arm. Release button to lock in position.
When the Swivel Arm detects a not allowed movement (close to floor or
ceiling), the motion is Locked and buttons blinks.
The Swivel Arm rotation is continuously indicated in the Touch Screen
Console. When the Arm has reached to an Angle limit, a message appears
on the Screen.

UNDERTABLE POSITION: Press and hold this button to place automatically the Swivel Arm
in Under-table position. If this button is released while the arm is in motion it stops
movement, press the button again to allow the Swivel Arm continue motion to Under-table
position.

THORAX POSITION: Press and hold this button to place automatically the Swivel Arm in
Thorax position. If this button is released while the arm is in motion it stops movement,
press “thorax position” button again to allow the Swivel Arm continue motion to Thorax
position.

LOW / HIGH SPEED: Movements are performed by default at High Speed.


Press this button (lights up) to perform movements at Low Speed for
accurate positioning.
By pressing, it Lights up in the Touch Screen Console.

SAFETY LOCK: This button lights up when a safety device is activated and
when a Thorax or under-table position is reached. The Safety Lock stops

PRIMAX INTERNATIONAL August 2020 – Data may vary without notice Page • 4
any motion of the Swivel Arm. The Operator must press it before
performing any other motion. Also, this button resets all recoverable errors.
MOVE TO AUTO-POSITION: Press and hold this button once a Program nº is selected in the
Program Screen of the Touch Screen Console or from the Remote Control. The Arm moves
automatically to the selected position. Once the Program position is reached, a message
appears on the Screen.

MOTORIZED IMAGE RECEPTOR ROTATION: This motorized movement


allows a maximum of +45° Clockwise movement and a minimum of -45°
Counterclockwise movement. It includes detents at 0° and when the
Receptor is in parallel with respect to the floor.
The Detector Rotation is continuously indicated in the Touch Screen
Console and when the Detector has reached to an Angle limit, a message
appears on the Screen.

ROTATION DISPLAY: This Display shows the angle performed by the Armand the angle
performed by the Detector. It also displays some error messages.

Automatic movement to UNDER-TABLE & THORAX POSITION

When the Swivel Arm arrives to Thorax Position, the When the Swivel Arm arrives to Under-table
movement stops and the “Safety Lock” button lights Position, the movement stops and the “Safety
up. Motorized movement. Lock” button lights up. Motorized movement.

PRIMAX INTERNATIONAL August 2020 – Data may vary without notice Page • 5
VERTICAL movement - swivel arm - receptor rotation - SID adjustment

◦ Swivel Arm speed: High and low speed. ◦ SID Adjustment Speed (SID): 8.7 cm/s.
◦ Distance Focal Spot to Floor: ◦ Detents: 100 and 180 cm.
Min 42.5 cm, Max 169 cm. ◦ Variable SID:
◦ Vertical Travel: 126.5 cm. Min: 100 cm Max: 180 cm.

◦ Detents on Rotation movements at: 0° and 90°.


◦ Movement Rotation of the receptor:
◦ Rotation of Swivel Arm: +120°/-30°
+45° and -45° cm.
◦ Detent at: 0° and parallel to floor.

PRIMAX INTERNATIONAL August 2020 – Data may vary without notice Page • 6
SAFETY LIGHT BEAM
Double photocell light system, which guarantee full
safety with the patient during the execution of the
technique:
◦ Once the photocell light, at the beginning of the
"U" arch of the Swivel arm, detects the patient
on the Table, Low Speed is automatically
activated. Also lowering the Arm below the
Table-Top and rotation will not be allowed.
◦ At the end of the "U" arch of the Swivel arm, the
second photocell disables all movements when
the photocell light is interrupted (except for Arm
up movement).
The Touch Screen shows a message when the Proximity Sensor is activated.

EMERGENCY STOP FOR MOVEMENTS


◦ An Emergency Stop, located at the Receptor, disables all movements of the
Swivel Arm in case of emergency (except for the Receptor Rotation).
◦ In the Room Electrical Cabinet, there is an Emergency-OFF Switch that allows
to cut the power Supply of the Unit.
◦ Parachute system, the system has a sensor that detects the tension of the
steel cable that supports the Swivel "U" Arm, locking the system whenever it
detects a loss of tension of the cable, stopping the movement of the
equipment even with a broken cable.

DETECTOR BUMPER & ANTICRUSHING SYSTEM


◦ Anti-collision pressure sensor during the movement of the system, along the Swivel arm.
◦ Bumper with pressure sensor located on the side of the receptor.
◦ Both safety devices allow to detect any pressure during the up-down or rotation movements of
the system, stopping the unit and reversing the movement to release the pressure made.

PRIMAX INTERNATIONAL August 2020 – Data may vary without notice Page • 7
HIGH FREQUENCY GENERATOR 50 kW / 150 kV
This High Frequency X-ray Generator is designed
for conventional or digital radiography and
pulsed or continuous fluoroscopy (optional) and
is controlled by multiple microprocessors
providing an increased image / exposure
consistency, efficient operation and extended
Tube life.

There are "service tools" available that facilitate


configuration, calibration and remote diagnosis.
Being able to remotely update the generator
software, collect data from error logs,
equipment counters, perform or restore backups
of calibration and / or configuration data,
download data from new X-ray tubes, consult
the software versions and / or the license
installed on the computer. Even with the presence of an operator in the room, by activating the hand
switch, the generator could be calibrated remotely and simultaneously while the curves of kV, mA,
filament current versus mA, or other internal generator signals are analysed.
All the mentioned service tools along with its advanced self-diagnosis system with indication of error
codes, simplify the troubleshooting of the equipment, allowing easy maintenance and even advance the
necessary spare parts in an installation before visiting it.

Its low ripple factor and high accuracy of the radiographic parameters (KVp, mA, exposure time), reduce
soft radiation and improve the X-ray beam homogeneity allowing an improvement in the image quality
and a reduction of the patient dose. There are also optional or configurable functionalities that reduce
the kV peak times and the kV drop times, allowing to reduce even more the patient and the soft radiation
doses.

Generator with advanced features in an Ultra-Compact Size:


◦ Remote service, update and diagnosis possibility.
◦ High frequency Constant potential.
◦ Tube protection circuitry prolongs Tube life and increases system
◦ Equipped with closed loop control of X-ray Tube current, kVp and filaments, which minimizes
potential errors and the need for readjustments.
◦ Automatic line compensation due to closed loop operation of X-ray Tube current and kVp.
◦ Heat Unit storage for the X-ray Tube, even after turning On / Off the equipment.
◦ Independent memory for storing Radiographic or Fluoroscopic operating parameters. This permits
rapid switching from one technique to another. Remote service, update and diagnosis possibility.
◦ Space charge compensation.
◦ Power cabinet and electronic control.

PRIMAX INTERNATIONAL August 2020 – Data may vary without notice Page • 8
TECHNICAL SPECIFICATIONS:

GENERATOR TYPE THREE PHASE, HIGH FREQUENCY, Low/ High Speed. 1 TUBE
INPUT LINE OPERATION 400Vac
FREQUENCY 50/60kHz
MAXIMUM POWER kW 50kW, According to IEC definition (0,1s, 100Kv)
MINIMUM POWER kW 0.4 kW (40kVp 10 mA)
kVp RANGE From 40kVp to 150KVp. In steps of 1KVp
kVp ACCURACY ± (3% +1kVp)
mAs RANGE From 0.1mAs to 630*mAs in 39 steps, Renard10 Scale
mAs ACCURACY ±(10% + 0.2mAs)
mA RANGE From 10 mA to 630mA in 19 steps, Renard10 Scale
10,12.5,16,20,25,32,40,50,63,80,100,125,160,200,250,320
,400,500,630
mA ACCURACY ± (4% +1 mA)
EXPOSURE TIME RANGE 1.0msec, Interval of 1.0msec to 10,000msec (0.001 to 10
seconds)
ACCURAY EXPOSURE TIME ± (2% + 0.1 ms)
POWER OUTPUT 630mA @ 79kVp
(@ 0,1s) 500mA @ 100kVp
400mA @ 125kVp
RIPPLE FACTOR 1%
MARGING OF ERROR - kVp &
Lower than 5% in all parameters
mA/ time
automatic compensation line ± 10%Vac
mAs in AEC 500mAs
WEIGHT & DIMENSIONS L445mm x W360mm x H564.5mm 65 Kg
NOTES (*) Under requirement, mA and exposure time stations could be
configured to three different logarithmic scales by the Field
Service Engineer:
 R´10: 63mA, 630 mA y 63 ms, 630ms, 6.3 s.
 R´10(64): 64mA, 640 mA y 64 ms, 640ms, 6.4 s.
 R´10(65): 65mA, 650 mA y 65 ms, 650ms, 6.5 s.

PRIMAX INTERNATIONAL August 2020 – Data may vary without notice Page • 9
ON/OFF Box for digital systems with integrated console

Generators controlled through a virtual


Console on a PC should be equipped with
a PC Interface Box or ON/OFF Box.
This Box comprises the controls for
“Power ON/OFF” and Hand-switch to
make radiographic exposures.
All the Exposure controls and indicators
located on the ON/OFF BOX are
positioned as follows:

ON: Generator is turned ON by pressing


this push-button (Pilot light is ON).
OFF: Generator is turned OFF by pressing this push-button.
HANDSWITCH: The X-ray Hand-switch button has three positions:
1. Off
2. Preparation
3. X-ray Exposure

1. PREP: Press the hand-switch half-way (” Prep” position) to prepare


the selected X-ray tube for exposure. The “Ready” indicator on the
Integrated Console will light when the X-ray tube is prepared and there
are no interlock failure or system faults.
After pressing this push-button, the following functions are activated:
 Anode rotation.
 Filament current switches from stand-by to the selected mA.

2.EXP: After the “Ready” indicator is illuminated on the Control


Console, fully, press the hand-switch to start the X-ray exposure.
 If the button is released before the Generator completes the selected time or the AEC (optional)
time, the exposure will be prematurely terminated and the actual mAs and Exposure Time will
be displayed.
 The “X-ray On” indicator remains illuminated on the Integrated Control Console during the
length of exposure.

3.OFF: Hand-switch disconnected.

Simultaneously to the use of the hand-switch, the indicators showed below are displayed in the
integrated control console:

PRIMAX INTERNATIONAL August 2020 – Data may vary without notice Page • 10
READY: Indicates that the technique selected is properly set, there are no interlock failures or
system faults, the anode is rotating and the X-ray Tube is ready for exposure. The “Ready” indicator
should be shown on the Integrated Console Control.

X RAY ON: Indicates that the X-ray exposure is in progress. At the same time that radiographic
exposures are being made, an audible signal sounds. The “X-ray On” indicator should be shown on the
Control Console.
ANATOMICAL PROGRAM APR
• Anatomical Programmer (APR) integrated in the digital system, with pre-programmed anatomical
views for automatic selection. The operator may introduce modifications manually in all the original
APR techniques and store them for later use.
• It has a section (various in the touch screen console) for personalized configuration for other type of
special studies not contemplated and can be configured by the user on a friendly use with all
parameters mentioned before.

AUTOMATIC EXPOSURE CONTROL (OPTION)


• Automatic Exposure Control (AEC) produces consistent images quality with excellent contrast
regardless of the radiographic technique selected. The AEC module comprises the controls for
the selection of the Exposure Detector Fields (Ion Chamber) and AEC Reset.
• Control of the Automatic Exposure Control for two Ion Chambers
• FIELD SELECTION: Each push-button indicates its related physical location of the selected field
in the AEC Exposure Detector, and it may be selected or deselected by pressing it. Any
combination of fields can be selected and the push-buttons illuminate when active.

ION CHAMBER FOR AEC (OPTION)

• Three fields Ion Chamber


• Range of KVp, from 40 to 150 kVp
• Range of dose, 0.5 at 1000 µGy/sec.
• Range of exposition time, from 0.001 to 10 sec.
• Absorption equivalent to 0.65 mm. Al.
• Differences between areas, less than 5 %
• Power supply tension, ±11.5 Vdc a ±15.75 Vdc
• Adaptation and cables for the connection between the Generator and the Chamber
• Adaptation to control more than two Ion Chambers in the system.

PRIMAX INTERNATIONAL August 2020 – Data may vary without notice Page • 11
GRIDS
◦ Dimensions 44 x 48 cm.
◦ Number of pairs of lines: 103 lines/inch.
◦ Inter spaced Aluminum material.
◦ Aluminum covered.
◦ Grid ratio 10:1, focalized at 150 cm.

X-RAY TUBE TOSHIBA E7884X


◦ Maximum Tension, 150 kV.
◦ Focus sizes:
- Small focus 0.6 mm - Large focus 1.2 mm.
◦ Maximum power:
- Small focus 20 kW
- Large focus 50 kW
◦ Maximum Current:
- Small focus 300 mA.
- Large focus 700 mA.
◦ Anode degree target angle 12o.
◦ Anode heat capacity 300 KHU.
◦ Anode Heat Dissipation Capacity 73 KHU/min.
◦ Housing Heat capacity 1,250 KHU.
◦ Housing Heat Dissipation Capacity 15 KHU/min.
◦ Anode rotation 2,700 / 3,700 r.p.m.
◦ Anode composition Rhenium & Tungsten faced; Molybdenum Target.
◦ Anode Diameter 74 mm.
◦ Filtration equivalent 0.9 mm Al.

HIGH TENSION CABLES


• Two 9 m Ultra Flexible High Tension Cables 75 KVp with covers

PRIMAX INTERNATIONAL August 2020 – Data may vary without notice Page • 12
MANUAL COLLIMATOR
The X-ray field is defined by six pairs of shutters, four of which are
lead-lined. The six pairs of shutters move perpendicularly within
the X-ray field. Two pairs of brass shutters are located near the
focus, two are located near the entrance window and two are
located near the exit window of the X-ray beam from the
collimator.
The latter shutters serve to accurately define the X-ray field
edges.
Manual Operation, Shutter movements are manual, controlled by
two knobs on the collimator front panel.
The control board provides power and constant current to the high intensity LED light.
The Collimator is supplied with a plastic flange for collimator rotation +/- 90o

Front panel for manual selection of:


• Opening shutters.
• Filter selection (optional).
• Centered light activation.

Technical Features
• Continuous Film Coverage from Max: 48 x 48 cm at 100 cm (40")
• Maximum Radiation Leakage: 150 kVp - 4 mA.
• High luminosity provided by a White LED simulating the X-ray field. Controlled by a Light timer.
• Light Field Indicator at 100cm SID,160 lux.
• Minimum Inherent Filtration 2 mm aluminum equivalent.
• Plastic Flange for collimator rotation; 20 mm thickness.
• LASER line aligning the collimator to the Potter Bucky. The line falls on a perpendicular cross-line.
• Light source ON time for the light field is factory set at 30 seconds. It is adjustable from 30 to 45
seconds via a trimmer on the timer board.

PRIMAX INTERNATIONAL August 2020 – Data may vary without notice Page • 13
INFRARED REMOTE CONTROL

CONTROL / FUNCTIONALITIES
VERTICAL MOVEMENT:
Motorized Vertical
Safety Lock
Movement of Swivel
Arm.
IMAGE RECEPTOR
SWIVEL ARM
ROTATION: Motorized
ROTATION: Motorized
Rotation Movement of
Rotation Movement of
Image Receptor.
Swivel Arm.

SID: Motorized SID LOW SPEED: Low Speed.


Adjustment.

UNDERTABLE
POSITION: Automatic Collimator Control Button
Undertable Positioning.

THORAX POSITION:
Pre-programmed Functions
Automatic Thorax
(Not Available)
Positioning.

MOVE TO AUTO-
Tube Rotation (Not
POSITION: Automatic
Available)
Positioning Button.

PRIMAX INTERNATIONAL August 2020 – Data may vary without notice Page • 14
TABLE ON WHEELS WITH FIXED TABLE TOP (STRETCHER)

Mobile patient table, fixed height and laminated top. Locking casters on back wheels to make sure it
stays in place. The large wheels enable easy maneuverability for accurate positioning. Resistant and
light-weighted design. Thanks to its dimensions, it is easy to move from one room to another.

Maximum Patient Weight 200 Kg


Dimensions (LxWxH) 200x65x70.5cm
X-ray Transparency Area (LxW) 188x52.8cm
X-ray Absorption Factor < 1.2 mm equival. Al.
Table-Top Composition Laminated
Locking Casters 2 Units

PRIMAX INTERNATIONAL August 2020 – Data may vary without notice Page • 15
DIMENSIONS

PRIMAX INTERNATIONAL August 2020 – Data may vary without notice Page • 16
PROFESSIONAL ACQUISITION SOFTWARE FOR X-RAY IMAGES

DROC is a professional software dedicated to the acquisition of radiographic images from Digital
Detectors (DR) that also integrates all the controls of the X-ray generator, turning out an intuitive and
agile working system.
The professional DROC image processor is adjusted to the user's needs, granting complete control of
the capture functions in each exposition and performing the entire work cycle from the same
application, to get instant diagnostic images, and distribute them to the workstations, PACS or
Printers.
BASIC FEATURES ADVANCED FEATURES
◦ Tool for controlling the elimination or
◦ X-ray generator control with integration of
reduction of artifacts.
all exposure parameters including dose
◦ Tools for controlling the image optimization.
radiation value received by the patient
◦ Tools for reducing image noise.
during the last exposure.
◦ Tools for grid Removal.
◦ Integration with the Health Centre's
◦ Specific algorithms for image processing of
RIS/PACS systems for worklists managing
each study
and export images to external servers.
◦ Images processing that allows making ◦ Statistics management of studies carried out
measurements, rotations, annotations, including the operator & patient data, type of
horizontal and vertical inversion, grayscale exposition, exposition parameters, radiation
inversion, zoom, magnification, electronic doses, possible image rejections and their
collimation, printing, editing. causes, etc
◦ Full screen images viewing. ◦ Histograms viewing.
◦ Dose Radiation control applied to patient ◦ DVD/CD & USB recorder, recording studios
with over-dose alert. with self-running DICOM.

SOFTWARE TECHNICAL FEATURES


Selectable language user interface.
Capture of worklist information by DICOM Worklist.
Use of DICOM Procedure to transfer all patient information directly to the (HIS/RIS) system.
Easy emergency patient registration.
Positioning guide fully integrated in the system.
Specific image processing algorithms and automatic image optimization.
Instant and fast image processing acquisition.
Advanced technology for image noise reduction.
Optimized technology for calibration and dose minimization.
Configurable system for automatic deletion of images older than a programmable value, only
when they have already been stored in the PACS

PRIMAX INTERNATIONAL August 2020 – Data may vary without notice Page • 17
BASIC WORKFLOW
Registration and data entry
◦ Remote Registration: Patient data is sent from the center's Worklist Server (RIS) and the operator
simply selects the patient's name. If the hospital server sends the code of each procedure to be
performed, the program selection is done automatically, avoiding this step for the operator.
◦ Local registration: The technician can manually enter all patient data and select studies relevant to
the examination.
◦ Edit existing patient information: search patient by name, ID, access number, barcode, etc.

Double click to start the exam

◦ Default programmed with exposure techniques that can be edited for each view. Allows the
creation of new anatomical programs.

◦ X-Ray generator exposure techniques are differentiated by patient size: small, medium and large.
◦ Automatic Exposure Control.
◦ Preview in 3-5 seconds after exposure.
◦ Complete processing in 6-9 seconds after exposure.

PRIMAX INTERNATIONAL August 2020 – Data may vary without notice Page • 18
◦ Displays the Dose Rate (EI)as defined in IEC 62494-1.
◦ If the image is valid the operator will select the Accept option and the software will automatically
load the new exposure data for that patient. If the operator decides to repeat the exposure, he has
the option of rejecting the previous exposure and the system will ask him to enter the cause of
rejection (to be selected from a preset or free entry list). This information is saved for the
subsequent production of usage and dosimetry reports for the equipment.

Data output
◦ The process of sending studies to PACS, supports auto-routing and sending images to multiple
destinations at the same time.
◦ The patient dose is included in the DICOM header of each image.
◦ Dosimetry information for the RIS is provided via the DICOM MPPS (Modality Performed
Procedure Step).
◦ Printer (supports auto-routing and sending images to multiple destinations at the same time).
◦ Additional image storage via DICOM CD/DVD Recorder.
◦ In the event of loss of connectivity with the PACS, the equipment will allow new studies to
continue and when this connection is re-established, the sending of all pending images to the
PACS will be automatically activated.

PRIMAX INTERNATIONAL August 2020 – Data may vary without notice Page • 19
Image processing
All the anatomical programs have pre-
programmed image processing values
according to the requirements of the
Health Centre, however, the software
allows the operator to manipulate the
acquired images in an easy and intuitive
way, thanks to visual tools.

BASIC Image Processing


◦ Brightness and contrast control.
◦ Selection of processing curves.
◦ Scrolling through the image.
◦ Image rotation.
◦ Horizontal/Vertical Flip.
◦ Adding and deleting annotations to the image.
◦ Zoom and Automatic Size Control.
◦ Image size magnification.
◦ Display at Full Size.
◦ Electronic collimation.
◦ Trimming.
◦ Image inversion to negative.
◦ Length and angle measurements.

The software employs the advanced


SYMPHONY image processing system. This
process window is not normally used on a
day-to-day basis as it is mainly used after
the creation or modification of anatomical
programs. The advanced operators proposed
by the Health Centre have access to services
mentioned.

PRIMAX INTERNATIONAL August 2020 – Data may vary without notice Page • 20
Advanced image processing
◦ Specific algorithms for each anatomical area.
◦ Independent adjustment of response curves and image optimization parameters.
◦ Advanced histogram adjustment.
◦ Ability to distinguish between different structures and tissues of the body to provide the image
with as much information as possible.
◦ Allows images with extended dynamic range which show from soft tissues to bone trabecular
structures.
◦ Algorithms for the elimination of artifacts.
◦ High process speed thanks to the optimization of algorithms and the hardware power.

Post-processing tools
Once the image has been processed and accepted, at the operator's choice, we pass at the "Overview
screen” where we find a set of tools for image post-processing.
◦ Full size viewing of the processed image.
◦ Mesurements of:
∙ Lengths
∙ Angles
∙ Areas
◦ Displaying one or more screens at the same time.

Studies and patient dose


The software incorporates a powerful system reporting tool with on-screen
display of information or export in Excel format for easy later analysis from any
other computer as long as you have a dose meter in the system.
The software allows you to make reports by searching for the following
parameters:
◦ By date.
◦ By patient name.
◦ For accepted or rejected images.
The exported data includes the following information for every X-Ray studies:

PRIMAX INTERNATIONAL August 2020 – Data may vary without notice Page • 21
◦ Number of images taken.
◦ Type of exposure and radiographic technique used.
◦ Number of images accepted and rejected and the reason for the images rejection.
◦ Dosimetry report by image and by X-Ray study.

Specific programme for paediatrics:


The system incorporates the Paediatric Projection Program specifically for paediatric patients, with
patient specific programs:
◦ Neonate.
◦ Children under 5
◦ Children over 5

Image Quality Control tools


The software incorporates a powerful image quality control tool that allows the acquisition of images
in "No Process" (RAW) format from the same user station. RAW images have DICOM extension so
they can be opened from external image control applications. In addition, image verification can be
carried out from the application itself, as a complete module is available with the possibility of
selecting ROIs and tools for measuring the average pixel and noise values.

Functionalities of the acquisition station according to DICOM 3.0


◦ Modality Worklist SCU.
◦ Storage SCU/SCP.
◦ Modality Performed Procedure Step (MPPS).
◦ Basic Greyscale Print SCU.
◦ Storage Commitment SCU.
◦ Verification SCU/SCP.
◦ Greyscale Standard Display Function (GSDF).
◦ The dosimetry values for each and total exposures of the studies are recorded in the DICOM
headers to be sent at PACS and MPPS servers.
◦ Radiation Dose Structured Report (RDSR)

PRIMAX INTERNATIONAL August 2020 – Data may vary without notice Page • 22
TOSHIBA FIXED FLAT PANEL FDX4343R
IMAGE FORMAT
● Caesium Iodide (Csl) with Amorphous Silicon (a-Si) Photodiode
● Active Area: 430(H)×439(V)mm (16.9 x 17.3 inch)
● Pixel Matrix: 3008(H)×3072(V)
● Pixel Pitch: 143 μ m
● Cycle Time: Shot to Shot 6sec.

PERFORMANCE
● Limiting Resolution: 3.5 lp/mm typ.
● MTF (2.0 lp/mm, 70 kVp, 1x1): 30 % Typ.
● DQE (0), Quantum - Limited: > 70 %
● A/D Conversion: 14 bit

RATINGS
● Energy Range: 40 - 150 kVp
● Maximum Entrance Dose (low Gain): 4 mR / frame

INTERFACE
● Data Output: 16 bit Digital Output Ethernet (1000BASE-T)
● Command Control: Ethernet (1000BASE-T)
● X-ray Synchronization Control: External
● Power Input: DC 24V 2A (from Power Supply Unit (Option))

IMAGE ACQUISITION MODE TABLE


● Mode: 3072 lines Full Scan mode
● Frame rate (Frame/s): 1 (approx)
● Binning/non-Binning: Non-binning
● X-ray period (ms): Standard, 500 (Variable between 50 to 500). Optional, 2,300, 3,200 or 4,000.

COMPONENTS AND CHARACTERISTICS


Flat Panel Sensor Unit:
● Sensor Protection Plate: Carbon Fiber Plate
● Cooling: Natural Air Cooling
● Input: DC 24V (from AC/DC Power Supply)
● Power Consumption: Maximum 20W
● Overall Dimensions: 512 x 495 x 43 mm (W(H) x D (V) x H)
● Weight:17 kg
Power Supply Unit:
● Input: AC 100 to 240V, 50/60Hz, Output: DC 24V 1.3A 60W
● Overall Dimensions: 126 x 200 x 60mm (W(H) x D(V) x H)
● Weight: 1 kg (approx.)

PRIMAX INTERNATIONAL August 2020 – Data may vary without notice Page • 23
WORKSTATION LENOVO
State-of-the-art technology. Provides maximum reliability and powerful performance in small
dimensions.

TECHNICAL SPECIFICATIONS- i5-8500 Processor BASIC


Platform C246, 250W 92% HDMI Port Onboard HDMI Port
Intel Core i5-8500
Slim DVD Rambo 9.0mm No
Processor Processor (9MB Cache, Ultra-Slim ODD
OS
up to 4.10 GHz)
8GB DDR4 2666MHz
Memory Selection Audio Card Integrated Audio
UDIMM
Chassis Intrusion Switch
First RAID Array RAID 1 (2HDDs min) SMA
Tower
2TB Hard Drive, Additional Intel 1Gbps Ethernet I210-T1
HDD
7200RPM, 3.5", SATA3 Ethernet Server Adapter
HDD Boot Drive Yes Dust Filter Dust Filter Tower
Second Serial Port Cable
Flex Bay 5.25" Flex Module Second Serial Port
Tower

Graphic Card Integrated


Operating System Windows 10 LTSB 2016

Keyboard For English language

Dimensions 165x328x376 mm Weight 10.6Kg

PRIMAX INTERNATIONAL August 2020 – Data may vary without notice Page • 24
EV2456W- EIZO MONITOR

Ultra-Thin Bezels on All Four Sides


1 mm ultra-slim bezels on the sides and top
with black borders of 5.2mm. the bezels are
flush with the screen surface, making a
completely flat screen perfect for multi-
monitor configurations.

New stand design


The latest stand design with multiple
adjustment options. It also saves space on the
desk and a base 43 mm less in diameter than
previous stands.

Manage multiple connected monitors


Its developed software allows to easily manage monitor
settings such as power, color, brightness and gamma by
using your mouse instead of monitor control buttons. It
also allows you to manage multiple connected monitors
from the one place

Monitor Configuration for Administrators


The settings of one monitor can be saved as
an xml file and applied to other monitors via
USB or office network..

PRIMAX INTERNATIONAL August 2020 – Data may vary without notice Page • 25
TECHNICAL SPECIFICATIONS
Panel Size 24.1 inch (61.1cm), ISP color screen, LED blacklight
Viewable Image Size (H x V) 518,4 x 324 mm
Pixel Pitch 0,270 x 0,270 mm
Contraste 1000: 1
Viewing Angles 178° / 178°
Brightness 350cd /m2
Display colors 16.77 milliones
Native Resolutión 1.920 x 1.200 (16:10 ratio)
D-Sub mini 15 pin × 1, DVI-D 24 pin × 1 (with HDCP),
Video input terminals
HDMI × 1 (with HDCP), DisplayPort × 1 (with HDCP)
• DVI: 31 - 76 kHz, 59 - 61 Hz
Digital Scanning Frequency • DisplayPort: 31 - 76 kHz, 59 - 61 Hz
• HDMI: 15 - 76 kHz, 49 - 51 Hz / 59 - 61 Hz
• 1 Port for monitor control
USB • 2 Port USB hub
• 3 Standard ports
Speakers • 1.0 W +1.0 W
Audio Input/Output 3.5 mm stereo / 3.5 mm headphone
Dimensions & weight (WxHxD) with stand 531× 531.4-510.1 x 190 mm 5,7Kg
Dimensions & weight (WxHxD) without 531 x 341.9 x 45.5 mm 3.7Kg
stand
Tilt 35° Up, 5° Down
Swivel 344°
Pivot 90° (both clockwise and counter clockwise)
Maximun & Typical power consumption Max.44 W /Standard.11 W

PRIMAX INTERNATIONAL August 2020 – Data may vary without notice Page • 26