Professional Documents
Culture Documents
Component Locations
Error Codes
Functional Checks
Preventive Maintenance
Theory
Troubleshooting
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Alignment and Calibration
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Alignment and Calibration
Gantry Tilt
Note This procedure requires a screwdriver, a level 6. Loosen the tilt sensor mounting screws.
and a protractor.
Note Moving the Tilt Sensor more than 0.6° will cause
Tilt Check the Stop/Report Loop to open.
1. Tilt the gantry to the 0° position. Check whether the 7. Rotate the tilt sensor (with upper mounting screw as
gantry is exactly vertical with a level. If it is not, pivot) until the GPC display reads 00.0 ± 0.2°.
perform the adjustment.
8. Tighten the sensor mounting screws, taking care not
2. Tilt the gantry +30° and -30°. If either end position to let it move.
cannot be reached without hitting any mechanical
9. Switch the rotation tilt control switch (SERVICE RTC)
end stop, perform the adjustment procedure.
and the brake switch (BRAKE SWITCH) to the
Tilt Sensor Adjustment downward position on the MCU/RTC Board.
1. Remove the gantry left upper and lower covers. 10. Establish communication between the controllers by
2. Remove the cover from the MCU/RTC Board. turning the gantry off, and then on from the CT Box.
3. Boot up the system to the Patient Data screen. 11. Replace all covers and check the gantry tilt function
range with a protractor.
4. Switch the rotation tilt control switch (SERVICE RTC)
and the brake switch (BRAKE SWITCH) to the
upward position on the MCU/RTC Board.
5. Tilt the gantry with S5 (TILT POS) and S6 (TILT
NEG) on the MCU/RTC Board until it is vertical.
Check with level by placing level on flat horizontal Switches on D301 and Tilt Sensor
surface of gantry frame, usually near the top. in Left Gantry Pedestal
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Alignment and Calibration
Image Quality
Image Quality Tests 4. Look for possible rings or other artifacts according to
the table. Enter their CT values in the appropriate
Note Close all Gantry covers before performing tests.
places on the table on the next page.
Check that the temperature regulation of the
Gantry and the DMS functions properly. Wait for 5. Check the image flatness in row 2 in the table on the
the temperature to stabilize within ±2°C. next page by using:
Note The gantry cooling, with heat exchangers and a • 1,000±200 mm² area ROI in the center
dehumidifier, keeps the internal gantry tempera-
• sliding 700±100 mm² area ROI to be moved
ture at 26°C ± 4°C (78.8°F ± 7.2°F), and the rela-
tive air humidity between 20%-30%. around, but at least 1 cm from the bone
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Alignment and Calibration
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Alignment and Calibration
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Alignment and Calibration
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Alignment and Calibration
Phantom Calibrations
1. Under Service, select Show Temperature to 11. Click Body Scan. Follow the instructions.
note the DMS temperature. Check if the temperature
Note The gantry rotates for 2 minutes between calibra-
is stable within ±1° for the last ten minutes.
tion modes for temperature stabilization.
Note The gantry cooling, with heat exchangers and a
12. Backup the calibrations on the EOD disk.
dehumidifier, keeps the internal gantry tempera-
ture at 26°C ± 4°C (78.8°F ± 7.2°F), and the rela- 13. Verify the Head Calibration and Body Calibration
tive air humidity between 20%-30%. against the table below:
2. Adjust the Gantry tilt to zero.
Phantom Head Range Phantom Body Range
3. Install Head Phantom. Use laser marker to position
Water 0±4 Water 0±4
it initially on the special white groove.
Polyethylene -75±10 Aculon 120kV 100±15
4. Reset the couch position to zero at this location.
Aculon 120kV 100±15 Aculon 140kV 100±15
5. Move the couch fully down, approximately 3-5 mm
from the covers. Lexan 116±15 Teflon 120kV 890±50
Acrylic Plexi 140±15 Teflon 140kV 870±50
6. Under Service, select Phantom Calibration
and Calibrate. Teflon 120kV 1016±50
7. Select Head, 90 and 120 kV, all Slice Width modes 14. Under Misc, select Air Calibration, and click
with Air Calibration. on [CALIBRATE].
8. Click Head Scan. Follow the instructions. a. Select All, then press [SCAN].
9. Repeat for the Body Phantom. b. Click on Start Automatic Scans.
10. Select Body, All voltages, all Slice Width modes with Note This group of air cals takes 30-40 minutes.
Air Calibration.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Alignment and Calibration
Tube Change
Note Changing the X-Ray Tube will take about 8 hours 3. Remove bottom front gantry cover by taking off 6
for an inexperienced FE. nuts.
Note This procedure requires metric wrench sizes 27,
30, 32, and 36 mm, a pair of channel lock pliers
and normal service tools.
Remove Old Tube
1. If necessary, move couch down out of the way of
gantry front door. Open the gantry front cover.
Remove system power.
2. Locate tube change kit (in system crate).
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Alignment and Calibration
Connecting
Cables
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Alignment and Calibration
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Alignment and Calibration
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Alignment and Calibration
Tube Port
Cover
Oil Bottle
15. Lower old tube onto tube box top. Secure using the
bolts and spacers.
Note If you need a visual aid, look at how the tube is
mounted in the cradle that you ordered. 18. Install the oil bottle on the tube being replaced.
16. Roll the tube out from the gantry. 19. Remove tube port cover from new tube, and put it on
the tube to be returned.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Alignment and Calibration
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Alignment and Calibration
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Alignment and Calibration
2. Select Standard and High resolutions, slice width of Resolution Axis 1 Axis 2 Spacing
4x5, all kV, and all rotation times, then press [SCAN].
Std, High, 337.375 337.875 0.50
3. Click on Start Automatic Scans. Ultra-High
Note This group of air cals takes about 5 minutes. Ultra-fast 337.75 337.75 0.0
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Alignment and Calibration
3. Create a study by clicking on the index card icon. Fill Slice Thickness: 4×5 Matrix: 512 × 512
in Last Name, First Name, ID and Sex. Scan Time: 0.75 sec 90 kV, 300 mAs
4. Select Head Std QA. 12. Under Service, Service Parameters window,
5. On the Protocol Parameters window, input make these changes:
these values in the Scan and Filter tabs: Phantom Calib.:Yes Image Balance: Yes
FOV: 250 mm Filter: EB 13. Press [MANUAL] on the CT Box to scan.
Slice Thickness: 4×5 Matrix: 512 × 512 14. After the scan, access the Protocol Parameters
Scan Time: 0.75 sec 90 kV, 250 mAs window again, and take scans at 120 kV and 140 kV.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Alignment and Calibration
16. Measure the CT number, average all slices. To Head Mode Body Mode
average the CT number take the ROI CT value from Default Default
(Slice B, (Slice B,
Factor Factor
each of the 4 head scans and add them up, then SW 5mm) SW 5mm)
divide by 4. The CT number should be 0 for water
Low 90 kV 0.980 Low 90 kV 1.000
and 100 for the plastic. If it is not, proceed to next
step. Med 120 kV 1.050 Med 120 kV 1.090
High 140 kV 1.133
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Alignment and Calibration
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Alignment and Calibration
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Alignment and Calibration
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Alignment and Calibration
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Alignment and Calibration
Head 8 3.1-3.7 , --
Head 5 4.0-4.8 , ,
Head 2.5 5.6-6.8 , ,
Head 1 8.8-10.7 , ,
Head 0.5 13.1-16.5 , --
Body 5 8.6-10.4
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
COMPONENT LOCATIONS
System Overview Couch
Body Phantom Head Phantom
Gantry Front QA Phantom
C-Box Couch Service Switches
DMS Detector and GI Cards Couch Power Supply
E-Box Horizontal Release Footplate
HV Tank PTH Board
Tube Collimator PTV Controller
X-Ray Tube Rear Crash Plate Switches
Gantry Right Pedestal Remove Bellows
Remove Side Rail
Gantry Rear
Undertable Components
AC Power Brush Blocks
Tabletop and Table Support Motor/Encoders
DC Power Brush Blocks
Tabletop and Table Support Pots
Control Signals and CAN Bus
Vertical Movement Encoder
DAS RCOM Transmitter and Receiver
Noise Cancellation
CT Box
Rear Cone PDC
Gantry Left Pedestal D450 DCC Power PCB Cap Bank
D451
MCU/RTC Board
PDC 24V Power Supply
Scan Drive Frequency Controller
Table Power Switch S1 Power Safety Switch S2
Tilt Sensor
Computer Rack Power Conditioner
Computer Rack Back WCS
SGI O2 Computer Plumbing Section
Image Reconstruction System (IRS) WCC Board
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
System Overview
Gantry
Couch
Monitor
CT Box
Keyboard
PDC
(Power Computer
Distribution Rack
Cabinet)
WCS (Water
Cooling
System)
Power
Conditioner
(optional)
Wall Box
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
Gantry
Temp
Switch
Humidity and
Temp
Sensors
Scan Drive
Locking Pin
X-Ray Tube
Heat HV Tank
Exchanger
Remove Tube
Tube Cooler
screws Collimator
and Oil Patient
and lift Expansion Laser Reference
cover Bellows Lights Detector
C-Box XRS (E-Box)
S301 Service Heat
Switch Exchanger
P-Plane
Collimator Gantry Left
Pedestal
UHR
Collimator Gantry Right
Pedestal
DMS Detector
and GI Cards Gantry Rear
Overview
Heat
Exchanger System
Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
X-Ray Tube
Temp
Switch Temp and
Humidity
Sensor
Dynamic
Focal Spot
Oil Fill Connector
Port Rotor Drive
Connector
Temperature
Sensor
Connector
Gantry
Overview
Oil Line
Disconnects
System
Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
Tube Collimator
Reference A-Plane
Detector Collimator
Pre-Amp
Reference
Reference Detector
Detector Reference
Detector
Pre-Amp
A-Plane
Collimator
Gantry
Overview
System
Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
HV Tank
HV
Tank
Spark Gap
Devices
Temp
Sensor
Gantry
Overview
System
Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
C-Box
DPU2 Power
Supply
DPU1 Power
Supply
DLP Delphi
Control
ZCO
ZControl
ACU
Acquisition
Control Unit
Blank (DOM Gantry
Option) Overview
System
Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
GI (Gated
Integrator)
Cards
Detector
Modules
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
Dehumidifier
Remove
GPC (Gantry GPC Cover
Control Panel)
Main Power
Cables
To remove
covers: Triple Valve
1. Lift off (Cooling Water
top cover. Regulation) Gantry Tilt
2. Lift off Motor
bottom cover. Cooling Water Tilt
Hoses from Adjustment Gantry
WCS (Water Spindle Front
Cooling 560 VDC
System) Gantry Rear
X-Ray Overview
Power
Cables from Gantry Left
PDC (Power Pedestal
Distribution System
Cabinet) Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
Gantry Rear
Noise
Cancellation
DAS RCOM
Transmitter
and
Receiver
Rear Cone
Support
Rear Cone
Supports
1. Remove Control
screws. Signals and
DC Power CAN Bus
2. Pull cover Brush Blocks
out from (560 VDC) AC Power
bottom Brush
catches. Blocks
Blower (230 VAC)
3. Lift cover Cover Gantry
off, set
aside. Rotation
Dynamic Gantry Right
S309 Rear Braking Pedestal
Cover Safety
WARNING: Gantry Left
Interlock
Live Pedestal
Switch
voltages
present Gantry
behind back Front
panel. System
Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
DAS RCOM
Receiver
LIBA
Board Gantry Rear
Overview
DAS RCOM
Transmitter
System
Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
Noise Cancellation
Remove
Cover
Gantry Rear
Overview
System
Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
X371
Neutral
X372
Switched
230 VAC L1
X373
Neutral
WARNING:
Live voltages
present.
X374
Permanent
230 VAC L2
Gantry Rear
Overview
System
Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
X381
Ground
X383
UDC Positive
+280 VDC X382
UDC
Negative
WARNING: -280 VDC
Live voltages
present.
Gantry Rear
Overview
System
Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
X341
X343
X345
X347 Gantry Rear
Overview
System
Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
Rear Cone
Microphone
Microphone
Board
Connection
for Audio
Functions Speaker
Connection
for Power
Gantry Rear
Overview
System
Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
Tilt Sensor
Exposure
Counter
(Scan Seconds)
To remove
covers: Gantry Tilt
MCU/RTC Board
1. Lift off Frequency
top cover. Converter
2. Lift off Gantry DC
bottom cover. Supply
Gantry Rear
Overview
Circuit
Breakers Gantry
Front
Gantry Right
Filter Pedestal
System
Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
Scan Drive
Frequency
Controller
Gantry Left
Pedestal
System
Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
Tilt Sensor
Tilt
Sensor
Gantry Left
Pedestal
System
Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
MCU/RTC Board
Service Rotate
RTC Switch and Tilt
Switches
Brake
Switch
MCU
(Master
Control
Unit) RTC
(Rotation
and Control
Board)
Stop/Report
Switches
Click here Gantry Left
for Stop/ Pedestal
Report Loop
System
Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
PDC
UPS1 (optional)
D450 DCC
Power PCB Cap
Bank
Pre-Transformer
Cap Bank
(US Systems)
Pre-Charge
Autotransformer
Relay K3
T1
Cap Bank Main
Contactor K2
Charge Relay X-Ray Room
Lamp and Door
Switch Interlocks
SLIO Board
PDC Breaker
on D451
Breaker Test
D451
Switch
Table Power
PDC 24V Power
Switch S1
Supply
WCS Power
24V Power
Taps
LED
Transformer
T1
Generator Line
Fuses 80A
System Line
Fuses 40A
Transformer System
Taps Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
D450 DCC
Power PCB
Cap Bank
Cap Bank
Pre-Charge
Relay K3
Cap Bank Main
Contactor K2
Charge Relay
SLIO Board
on D451
D451
PDC
T1 Overview
System
Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
Table Power
Switch S1
Power Safety
Switch S2
PDC
Overview
System
Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
PDC
Overview
System
Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
WCS
WCC Board
F9
T1
Laptop Serial
Connector
F8
F3
K3
Electronics
Note: This is a Water to Layout Diagram
Water WCS cabinet. It uses
an external supply of cold
water to regulate gentry
Plumbing
water temperature. Section
Another option (not pictured)
is the Air to Water WCS
cabinet. It uses a built-in
compressor and Freon to
cool the water. It also
requires large air vents for
cooling the condensing
system. System
Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
Plumbing Section
Central Water
Supply Inlet
Central Water
Supply Outlet
Drain Valve
Stop Valve
Gantry Water
Supply Outlet
WCS
Overview
System
Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
WCC Board
Laptop Serial
Connector
Connect here to
run the Water
Cooling Control
Status Program
(run on laptop)
WCS
Overview
System
Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
Power Conditioner
Open
Top Cover
Open
Front
Door
Note:
The Power Conditioner System
is optional equipment. Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
Couch
Click here to see Click here to see the Slide away Base
how to position Tabletop and Table Cover to access
and fill the QA Support Motor/Encoders, Couch Service
Phantom, and the Tabletop and Table Switches and
how to position Support Pots, Circuit Breakers
the Body and the PTH Board. (See Couch System
Phantom. Power Supply.) Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
Couch
Overview
System
Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
Couch
Overview
System
Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
PTH Board
PTH
MCB
NOTE:
Remove Tabletop
and Table Support to
access PTH board.
Couch
Overview
System
Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
Rear Crash
Plate Switches
Couch
Overview
System
Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
After removing
screws, pull out
cover to
remove
Couch
Overview
System
Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
Remove Bellows
To detach Bellows
from the bottom:
1. Raise couch fully up.
2. Pull off Bellows.
Note:
If greater access is
required, remove
Bellows from the top.
To remove Bellows
from the top:
1. Remove Side Rail.
2. Remove bolts.
3. Pull off Bellows Rail.
4. Pull off Bellows.
Couch
Overview
System
Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
Undertable Components
Spindle
Switch
Vertical
Movement
Encoder
Couch Service
Switches
PTV Controller
Vertical
Frequency
Vertical Relay
Frequency
Converters and
Power Monitor
Couch Power
Supply
Couch Circuit
Breakers Couch
Overview
System
Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
PTV Controller
PTV
Controller is
under this
cover
Couch
Overview
System
Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
Service
Switch
Up
Down
Couch
Overview
System
Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
Couch Power
Supply
Vertical
Movement
Encoder
Couch
Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
Remove
two screws
to access
Footplate
Switches
Detail View
Couch
Overview
System
Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
QA Phantom
Couch
Overview
System
Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
Body Phantom
Head Phantom
Couch
Overview
System
Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
CT Box
Couch Feed Couch Lift
Controls and Controls and
Position Position Display
Display
Gantry Tilt
E-Stop Controls and
Button Position Display
(part of Stop/ These controls do
Report Loop) not work, as these
functions are
Audio Intercom
controlled in
Controls software
Gantry Power
Switch
X-Ray On
Indicator
Scan
Controls
System
Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
Computer Rack
EOD Drive
Power
Switch
Remove Cover
SGI O2
Computer
Power
Switch
NOTE: Image
When you turn this switch Reconstruction
on, it powers up only the System (IRS)
computer rack. IRS Power
However when you turn it Switch
off, it powers down the
entire CT System.
Computer
Rack Back
System
Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
Remove Back
Cover
Computer
Rack Front
Signal Distribution Board System
Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
SGI O2 Computer
CD-ROM Power
Drive (used Supply
for loading
IRIX Software)
Power
Switch and Hard Drive
Power
Indicator
Light
Slot for
Backup
Motherboard Hard Drive
and RAM
Audio
Network Board
Connection
Jack
Computer
Rack
System
Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
System
Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Component Locations
E-Box
D115 Anode
Remove Power
Covers
D470
Filament
Power
Inverter 1
and 2
D400 Main
Control
Board Gantry
Overview
Buffer Caps
1 and 2 System
Overview
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Error Codes
ERROR CODES
Error Codes Inside the Error Log Viewer
Click here to launch the HTML error code guide. Once in the error log viewer, you can:
• Doubleclick the error you wish to view to display a
detailed message, if available.
X-Ray System Errors
• Filter the error types by Date, Host, Application,
Click here to see a list of X-Ray System Errors, and
troubleshooting tips for resolving them. Component, Code, Level, and Severity.
• Click on the FILE menu to access these options:
– Save As - Saves all of the messages in the /usr/
Accessing the Error Log Viewer diamond.V2.0/bin/ subdirectory under the user-
1. From the Applications screen, Click on Service selected name.
and select Error Log Viewer.
– Print - Possible future error log print capability
2. On the pop up menu, select Today to view current
day's error log. – Email - Possible future error log email capability
3. Select the type of error to view, either Error or – Log Out - Shuts down the Error Log Viewer, logs
Critical. out of the applications screen to the Log In
screen.
– Exit - Exits from the Error Log Viewer and keep
the Applications screen active.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Functional Checks and Procedures
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Functional Checks and Procedures
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Functional Checks and Procedures
Autotransformer Protection
1. Switch the F99 breaker to the OFF position and then 5. If the LEDs status is OK, press and hold the S100
to the ON position. test key for at least 2 seconds. The main breaker
F99 must trip OFF to the middle position.
2. Verify that F1 and F9 breakers in the PDC are
switched ON. S100
3. Switch the power ON using the facility ON/OFF K100
switch. K101
4. Verify the status of the K101 and K100 LEDs as
follows:
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Functional Checks and Procedures
Console Power
• Check on the back of the SGI O2 Computer for
power at the plug.
• Check W8 at the Power distribution unit in the console
for 230VAC.
• Check X254 connector for 230VAC at the PDC.
• Check fuse F5 in PDC.
• Check integrity of K3 to be sure it pulled in to allow
power to fuse F5.
• Check UPS1 (optional) power.
• Check fuse F4 in PDC.
• Check phase A (L1) at F1 fuse.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Functional Checks and Procedures
CT Box
To verify the CT Box is functioning properly:
1. Click on service.
2. Select procmon.
3. Select CT box.
4. Push each button on the CT box and watch for a
response to be generated, for both pushing and
releasing each button.
Note The top portion (gantry functions) of the CT box
can not be tested with this utility.
Note The top portion of the CT Box gets power from
the GPC in the Gantry Right Pedestal. The bot-
tom of the CT Box gets power from the CIB in the
Image Reconstruction System (IRS).
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Functional Checks and Procedures
Couch
Checks Downward movement is enabled by K2 when the
following four conditions are met:
If there is a problem with couch movement:
1. Down_Enable is set by the PTV Controller
1. Check fuses F701 - F706 on the left foot stand
behind bottom cover of the table. 2. No vertical block is detected by the Rear Crash
Plate Switches
2. Check Rear Crash Plate Switches to be sure they
are not activated. 3. Motor temp is ok
3. Check F15, F16 and F8 in PDC. 4. Connection between nut and scissor is ok (S704)
4. Check K4 in PDC. Upward movement is enabled by K3 when the following
two conditions are met:
5. Verify phase 2 (L2) in the PDC is OK.
1. Up_enable is set by the PTV Controller
6. Verify Horizontal Release Footplate E703 is not
activated or stuck. 2. Motor temp is ok
7. Verify S704 connector switch on worm gear in table
base is not deactivated.
8. Check motor temp; motors could be overheating.
Refer to Couch Power logic diagram.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Functional Checks and Procedures
Couch, continued
Manual Move 4. Turn on the vertical drive (green switch) on the
You can move the couch manually, without using the bottom of the couch right side.
computer.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Functional Checks and Procedures
Daily QA Scan
1. Warm up the tube. Run a Short TC (Daily Tube
Conditioning).
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Functional Checks and Procedures
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Functional Checks and Procedures
DAS Power
Make these checks to verify DAS power: – 15v_n: -15 VDC (DPU1) fused - check DPU1 fuse
• Check PDC F9 (L2-phase 2). – 15v_p: +15 VDC (DPU1) fused - check DPU1 fuse
• Verify UPS2 is installed and working correctly – 24v_p: +24 VDC (DPU1)
(optional).
– 5v_p: +5.4 VDC (DPU1)
• Verify K9 in PDC is pulled in to allow power through.
– 5v_p1: +5.4 VDC (DPU1) fused - check DPU1 fuse
• Verify K10 in PDC is pulled in.
• If power is switched on, check AC Power Brush
• Check Gantry Rear cover safety switch. This disables Blocks X371 for 230 VAC. Also check F306 in Gantry
K10. Left Pedestal pss and F8 in PDC.
• Check AC Power Brush Blocks X373 for 230 VAC. • If DPU2 power keeps dropping out, or won’t come on
at all, check DMS temperature. Power_disable_2
• At C-Box, check:
switches the output voltage of the DPU2 on and off,
C-Box DPU2 power is 24 hour power and on after depending on the DMS temperature.
wall box is turned on:
• On C-Box ZCON D514, check F1 and F2.
– P24: +24 VDC (DPU2)
ZCON monitors all voltages of DPU1 and DPU2
– N13: -13 VDC (DPU2) (except 5v_p) and generates one bit of volt_ok which
goes into a mux on the zcon. MCB D2 on the C-Box
– P8: +8 VDC (DPU2)
ZCON reports voltage errors via the stop report loop
– P5: +5.5 VDC (DPU2) and can bus.
C-Box DPU1 power is switched after K4 (system • Check fuses F1 through F3 on C-Box Delphi (DLP).
turned on):
• Check FEE (front end electronics).
– 48v_p: +38 VDC (DPU1)
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Functional Checks and Procedures
• Store Raw: Yes 4. Run the Fee Bias test in the Service menu by
selecting the appropriate raw data from the IRS
• Diagnostic level: 1 archive.
• X-ray flag: No Note The software automatically identifies the type of
2. Perform two scans with the scan parameters from raw data, with or without x-ray.
the Data Table below; one with 4x5 mm thickness
and one with 4x1 mm thickness.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Functional Checks and Procedures
If the results are rejected, causes can be: DMS Detector If there is a problem with the anti-scatter grid, light
and GI Cards, Anti-scatter, or cable cleaning needed. A leakage may be found at end detectors (the first or the
Bias problem is usually a GI card problem. Noise is last in the module). To check this:
usually solved by cleaning the module cable, or it could • Check the Prep data. Leaking light appears as dark
be light leakage. Light leakage always means replacing lines along the detector.
the module.
• The STD (Standard Deviation) noise in a module with
Note Test reference detector at slice width 4x5 only. a light leakage problem is very high, and is outside the
limits in the Data Table below.
Caution Use extreme care when handling GI Card and
Detectors. They are very static sensitive. • Perform a scan without rotation while the GI card is
disconnected from its module. Click on Misc,
Check the GI card for biases first, then check for light Service Parameters, and change these settings:
leakage, then do the cleaning.
• Diagnostic level: 2
The procedure below can be used as is for modules not
in the center position (19-24). • Rotation: No
It is possible to swap a modules in any position in the If this effect exists, it will not appear with a
disconnected GI card.
DMS except for modules in the center position (19-24).
Do not swap them even for diagnostic purposes.
If a problem is found in any test:
Continued on next page
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Functional Checks and Procedures
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Functional Checks and Procedures
• Ignore rejects by the program if the limit lines are L200 L1500 1.2 0.01 2.5
M180 M600
misplaced by the program. H160 H330
• The criteria for signal rejects is defined in the Data W/ X-ray Reference Detector Signal Range (RAWRef-BiasRef)
Table below.
Minimum Value Maximum Value
Default parameters:
6,000 Low 11,000 Low
• Voltage: 120kV W/ X-ray All Detectors (RAW-Bias) data tolerance
• mAs: 170 Allowed deviation in % ±30
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Functional Checks and Procedures
Filming
1. Click on Processing and Viewer.
2. Select Local or EOD, and select a study to be
filmed, then click OK.
3. Select File, and then select:
• Film Images to film all images on the screen.
• Master Film to select images to film.
• Print to begin filming.
• Queue Manager to view pending filming jobs.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Functional Checks and Procedures
Gantry Fans
Verify the following fans are working:
• Two on Heat Exchangers of the X-Ray Tube.
• One behind each of the 3 radiators in the Gantry.
• Two on the Data Measurement System (C-Box).
• Four in the high voltage electronics box (E-Box).
• Two on the HV Tank high voltage transformer unit.
• One on the C-Box.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Functional Checks and Procedures
Gantry LEDs
1. Turn the switch in the upper right corner of the
Computer Rack ON. C-Box Components
ACU Z_CON Delphi DPU1 DPU2
2. Verify that the Image Reconstruction System (IRS)
power switch is ON.
3. Press the operating button on the front panel of the
SGI O2 Computer, and verify that the LED turns
green after a few seconds.
4. Turn ON the Gantry from the CT Box.
5. Disable rotation by setting S301 in the Gantry to
position 0.
6. Verify the following in the Gantry:
• All four LEDs on DPU2 are green.
• Six LEDs on DPU1 are green.
• LEDs on DELPHI, +5V, +24V, +15V and -15V,
are green.
• LEDs on Z_CON, +5V and +24V, are green.
• LED on ACU, +5V, is green
• LED on R_COM, +5V, is green.
Note LEDs shown in standby state.
R_COM
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Functional Checks and Procedures
Gantry Power
To verify gantry power:
1. Check F301 (power to rotate converter E301) in
Gantry Left Pedestal on PSS.
2. Check 3 phase power at Z301 on Gantry Left
Pedestal.
3. Check 3 phase power at X257 at PDC.
4. Check temperature switch B305, Gantry front top.
5. Check Gantry Rear cover safety switch.
6. Check for the activation of K4 in the PDC. K4 is
activated by K2, which is activated by the power on
switch.
7. Check F8 in the PDC.
8. Check F1 in the PDC.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Functional Checks and Procedures
Gantry Rotation
Use the following procedure to check rotation:
1. Enable rotation by setting S301 in the Gantry to
position 1.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Functional Checks and Procedures
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Functional Checks and Procedures
Gantry/Table Movement
From the Gantry Control Panel From the CT Box
1. On the computer, doubleclick on the MX8000 icon to 1. Ensure the system is powered up.
log in to the MX8000 software.
2. Verify the LEDs on the CT Box are ON.
2. When the log in process is finished, verify that the
Note The LEDs on the CT box are ON only when the
movements are enabled. The LEDs on the Gantry
corresponding movement is enabled. If the LED
control panel should be ON.
is OFF, the movement is disabled.
Note The LEDs in the gantry control panel are ON only
3. Check all movements from the CT box: UP and
when the corresponding movement is enabled. If
DOWN, IN and OUT, MARKERS ON and MARKERS
the LED is OFF, the movement is disabled.
OFF, and TILT.
3. Check all movements from the gantry control panel:
UP and DOWN, IN and OUT, MARKERS ON and
MARKERS OFF, and TILT.
Note All movements should be smooth and you should
not hear any unusual noise.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Functional Checks and Procedures
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Functional Checks and Procedures
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Functional Checks and Procedures
Power
Select a topic: DELPHI
• Data Acquisition System (DAS) The following voltages are monitored:
• Gantry Rotating 48V_P +238 V Fuse monitoring DPU1
Data Acquisition System (DAS) 15V_N -15 V Fuse monitoring DPU1
Z_CON 15V_P +15 V Fuse monitoring DPU1
The following voltages are monitored: 24V_P +24 V Fuse monitoring DPU1
5V_P +5.4 V monitored on MCB, D2
N13 -13 V Fuse monitoring DPU2, E541
P8 +8 V Fuse monitoring DPU2, E541
P5 +5.5 V Fuse monitoring DPU2, E541
P24 +24 V Fuse monitoring DPU2, E541
48V_P +38 V Fuse monitoring DPU1, E540
15V_N -15 V Fuse monitoring DPU1, E540
15V_P +15 V Fuse monitoring DPU1, E540
15V_P +24 V Fuse monitoring DPU1, E540
5V_P +5.4 V monitored on MCB, D2
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Functional Checks and Procedures
Power, continued
Gantry Rotating DPU2
DPU1 The Digital Power Unit 2 (DPU2) is a multiple output
The Digital Power Unit 1 (DPU 1) is a multiple output power supply that generates the voltages shown:
power supply that generates the voltages shown:
P24 +24 VDC DMS, ZCON D514 (only monitoring)
24V_P +24 VDC ZCON D514, DELPHI D515, DMS N13 -13 VDC DMS, ZCON D514 (only monitoring)
15V_N -15 VDC DELPHI D515, DOM D516, Phi_Amp P8 +8 VDC DMS, ZCON D514 (only monitoring)
D555, ZCON D514 (only monitoring)
P5 +5.5 VDC DMS, ZCON D514 (only monitoring)
15V_P +15 VDC DELPHI D515, DOM D516, Phi_Amp
D555, ZCON D514 (only monitoring)
5V_P +5.4 VDC ZCON D514, DELPHI D515, DOM
D516, ACU D513
5V_P1 +5.4 VDC Phi_Amp. D555; LPS, D522
48V_P +38 VDC DELPHI D515, ZCON D514 (only
monitoring)
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Functional Checks and Procedures
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Functional Checks and Procedures
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Functional Checks and Procedures
Stop/Report Loop
When the stop/report loop opens, all movements are
disabled immediately.
Connector On Cable At
Component
MCU/RTC Board Component
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Functional Checks and Procedures
System Cooling
1. Under Service, select Show Temperature.
2. Screen display includes:
• Gantry temperature
• Gantry humidity
• Inlet water temperature
• DMS Cradle temperature
• DMS temperature for left and right
• Tube Heat Units readings taken with every scan
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Functional Checks and Procedures
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Functional Checks and Procedures
System Power Up
This is the sequence of events at power up. 8. Approximately 20 seconds after gantry turns on, the
decimal points on the tilt and horizontal displays
1. Turn the wall box on.
light.
2. Ensure cooling water is on.
9. Approximately 75 seconds after the SGI box is
3. Turn on the system by rotating the main power turned on, the monitor displays the login screen.
switch on the CT Box to the ON position.
Note The IRIX operating system is now running.
Note S/N9008 and older systems have a pushbut-
10. Use the left mouse button to doubleclick the MX8000
ton main power switch instead of a rotary
icon, to start loading application software.
switch. Press the white pushbutton.
4. SCSI drives in the Image Reconstruction System
(IRS) can be heard as they spin up, and all SCSI
drives LEDs light up for several seconds.
5. The EOD drive comes on.
6. Press the ON button on the front of the SGI O2
Computer.
a. The LED indicator on the SGI is red for
approximately 1 second, then turns orange.
b. After 20 seconds, the SGI box chimes and the
LED turns green.
7. The vertical, horizontal and tilt displays on the scan
box flash 3 times.
Continued on next page
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Functional Checks and Procedures
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Functional Checks and Procedures
Rear Table
Safety
Switches
Gantry
E-stops
E-stop
on
CT box
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Functional Checks and Procedures
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Functional Checks and Procedures
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Functional Checks and Procedures
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Functional Checks and Procedures
L1 The firmware of the Water Cooling Control (WCC L16 Reset_WCC is active from the Master Control
Unit (MCU).
Board) is running.
L17 The output line is active.
L2 The service function of the WCC is active.
L18 The output line is active.
L3 The regulation drive should move towards
warmer water (open bypass valve). L19 The output line, to switch the water pump ON, is
active.
L4 The regulation drive should move towards colder
water (close bypass valve). L26 WCC supply voltage is present, after the rectifier
and the onboard fuse.
L6 An error was detected in the water cooling
system.
L14 The water pump F3 overcurrent protection
responded.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Functional Checks and Procedures
3. On the laptop, select Start, Programs, 10. You should see a display similar to this:
Accessories, Communications, WWT_P20 V002 18.09.1998
HyperTerminal. =======================================================
time Tset Tact prim i_mem perc new valv 123 FP state
4. When the program window opens, click on --------!----!---!---!----!-!-----!----!---!--!-------
Hypertrm.exe. Note If you don’t see this display, try shutting the sys-
tem down and power off at the wall box for a
5. When prompted, type a name like WCS Connection
minute. Then power up the system again to
and pick any Icon. Click OK.
Standby state.
6. In the Connect To window, select Connect
11. Type STA 1 and press [ENTER]. This tells the WCC
Using: Direct to COM1. Click OK.
program to display the WCS status once every
7. For Port Settings, use: second.
• Bits per second: 9600
• Data Bits: 8
Continued on next page
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Functional Checks and Procedures
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Functional Checks and Procedures
X-Ray System
Select a topic: Check 560VDC UDC_P and UDC_N
• Collimator Testing UDC path is as follows:
• Check 560VDC UDC_P and UDC_N 1. D450 DCC Power PCB Cap Bank X7-X10 in the
PDC via the W3 cable to the X303 connection in the
• E-Box D400 LEDs Gantry Right Pedestal.
• E606 And UDC Power Supplies Table 2. X303 connection in the right leg of the Gantry via the
Collimator Testing X315 cable to the DC Power Brush Blocks
connections X382 and X383.
1. On P-Plane collimator, turn on S1 on C-Box Delphi
(D515 board) for service. 3. Slip ring connections X382 and X383 via the W505
cable to the x-ray system.
2. Move S2 and S3 on C-Box to manually open and
close collimator to test functionality. The UDC voltage varies with line voltage. If a
transformer is used (i.e., the line voltage is other than
3. On C-Box ZCO (Zcon), turn on S1 for service.
400V), the UDC voltage must be 560 V ±10%.
4. Move S2 and S3 to manually move in and out UHR
Important! During a scan, the voltage drop of UDC should
collimator to test functionality.
never be more than 80 V below the actual
UDC voltage, or a missing phase in the PDC
could result.
Check the three F100 (80A) fuses in the PDC. If any one
of the fuses is defective, replace all three.
Note Look for the yellow LEDs on the D450 for the
presence of 560VDC.
Continued on next page
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Functional Checks and Procedures
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Functional Checks and Procedures
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Logic and Block Diagrams
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Logic and Block Diagrams
Audio Intercom
IRS HOST O2
PCI CABLE
MULTI-
SERIAL
SPEAKER
W382
D342
SDB PREAMPLIFIER
W313
PSS
5V AND 12V MICRO1 MICRO2 LS1
LOUD
SPEAKER
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Logic and Block Diagrams
Couch Power
POWER SUPPLY
(UNDER TABLE) W704
G700 U4_P F704
24V Supp_24V
F705 PTH
Table_24V
SENSE_U1
SENSE
U1_1 F701
5V VCC1
HORIZONTAL
D701
U1_1 F702
5V VCC2
U2_P F703
12V 24V
PTV
U3_P
12V 12V_fur
U5_P F706 OPTIONAL
5V HARDWARE
VERTICAL
D702
Stoprel (STOP RELAY)
W701 W6
U
FILTER VERT POWER DISTRIBUTION CABINET
Z702 V VERT
FREQUENCY PDC
N_PHS2 L_PHS2 Stop relay W MOTOR
D703 CONVERTER
M702 K4
L_PHS1 N_PHS1
F15
6A F8 F1
X301 13A 44A
GANTRY STATIONARY L2
LEFT STAND
F16
X257 2A
S1
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Logic and Block Diagrams
TABLE TOP TABLE TOP TABLE TOP TABLE TOP TABLE TOP TABLE TOP TABLE TOP TABLE TOP TABLE TOP TABLE TOP TABLE TOP
MOTOR CLUTCH BRAKES (2) ENCODER POT REF SWITCH SUPPORT SUPPORT SUPPORT SUPPORT REF SWITCH
M711 (OPTO) MOTOR BRAKE POTENTIOMETER ENCODER (OPTO)
Y713 Y711 Y712 B711 R711 U711 M710 Y710 R710 B710 U710
S711B
Topmot Topclutch Topbrake K1,K2 Tabletop_pot Toprefsw
Topsafetyrear
Topsupportmot
Topsupportbrake
Supprefsw
Topsupport_pot
Ch0,ch90
PATIENT TABLE VERTICAL - PTV
Stop
D702
HTP
HORIMOVEACT PTV MCB
VERTBLOCK D2
PTH MCB Stoprep
D2
VERTMOVEACT
FOOTSW
Stoprep
PATIENT TABLE HORIZONTAL - PTH
D701
MCU
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Logic and Block Diagrams
Couch Signal
PTH M
AND TABLE TOP = MOVES 1651mm
HORIZONTAL
DRIVE M TABLE TOP SUPPORT = MOVES 350mm
CONTROL
D701
UPPER LIMIT = 1022mm
COMM
Footsw STOP STOPREP HTP GANTRY STATIONARY
VERTICAL DRIVE =
L_
PS
L_
MOVE BETWEEN 479mm
HS
PS
(19INCHES) AND
HS
2
HTP - Horz table pulses =2
1
1022mm (40INCHES)
=
LY
30
23
VA
PP
Comm
0V
SU
C
Stoprep
AC
ER
StopPHS
W
PO
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Logic and Block Diagrams
42 MODULES
DETECTOR MODULES
42 GI CARDS
GI CARDS
DPU_2 CBOX
SERIAL DATA
CAPACITIVE
SLIP RING
DATA (FIBER OPTIC)
IRS CONTROL
ACQ MCU
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Logic and Block Diagrams
+280V
UDC_pos
L1 L1 w505 UDC_pos
GENERATOR
F100 - 80A x
T1 7
slip rings
400VAC
x
L2 XFMR Z L2
LIGHTS WHEN 8
>40V w
x 3 w315
1 9
x
L3 TOP T1 L3 1
UDC_neg 0 w506 UDC_neg
D1 X17
+24 V
+24 V
ChargeRelaysClosed
PowerRelaysclosed X1
X2 K2 +24V
ClosePowerRelays
+24V
CloseChargeRelays K3
+24V
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Logic and Block Diagrams
XRAY LAMP
TEMP SWITCH -170C (380F) CONTROL
IN XFMR T1 EXT DOOR
ANNUNCIATOR SWITCH W51 for example
Most of this circuitry is designates a cable
in the PDC
X19 X15 X16
serial MCU/
+24 V
linked RTC
K1
I/o D301
SLIO DOOR
PCB TemperatureSwitch
+24V
K4
+24V
X1 XRAY >20KV
StopStatus K5
X2
D1
140F TEMP SENSOR StopReportIn
TemperatureError StopReportOut
CommPos
X3
CommPos
CommNeg
CommNeg
W51
+24V
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Logic and Block Diagrams
Dehumidifier
MCU/RTC
D301
DEHUMIDIFIER
DehumidifierPos
24V
W367 Note:
Incoming
MCU line voltage
MCB
DehumidifierNeg from F307 in
Gantry Left
D2
Pedestal.
+
HumidityPos
+15V B302
W367 HUMIDITY SENSOR
HumidityNeg
GANTRY
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Logic and Block Diagrams
4 x 5 mm A B C D
Z - AXIS
42 detector modules
ep
de
LEFT RIGHT o rs
ct
te
de
8
QUAD DMS LAYOUT
D
00
1-
D 2
01 D 67
6 5 7-
D6
INTERCONNECT CABLES
(42 CABLES) static sensitive
42 GI cards
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Logic and Block Diagrams
RIGHT DMS
LEFT DMS TEMP SENSOR 2
TEMP SENSOR 1 1-ON
1-ON 2-ON
2-OFF 3-OFF
3-ON RCOM
P9
THERMAL SWITCH
OFF T>35C
ON T<25C
CBOX GI CARD
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Logic and Block Diagrams
Gantry Rotate
PDC L1 L2 L3
GANTRY STATIONARY
MCU/RTC PCB Gantry Front Cover
K4 Safety Switches
K1 K2 RotBrakectrl W1
+24V ROTATION SAFETY
LINE FILTER Z301 SWITCH
S301
F301
10A
STOP CIRCUIT PSS
W373
K16
+24V
K1 K2 K3
W303 X2
W362
K10
StartrotCW K9 +5V K2
L1 L2 L3
K3
ROTATE FREQUENCY
StartrotCW CONVERTER L1
RTC E301 U1
X322
MCB V1 X1 X X3
D2 X9 24VrotFC Rotate Motor (Rotor)
1 M301
W1 W318
RS485 Commrot
Comm Commrot
CONTROLLER
Opto sensors
CTL_DAT_SR_neg X6 W552
X342 AP90
Service RTC CTL_DAT_RS_pos
X343 Rot idx
S4 CTL_DAT_RS_Neg
Rot CCW X344 ACU
S3 RS means ring to
Rot CW stationary
Gantry Rotate
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Logic and Block Diagrams
Gantry Tilt
TILT MOTOR
L1 AND BRAKE
M302
POWER
X5 230vac
TILT COMMAND 3 phase
X18 X1
TILT FREQUENCY
CONVERTER
E302 TILT BRAKE
X1 TILT SENSOR
X7 (Clinometer) GPC
TILT PUSHBUTTONS
B301
X4
X3
TILT + TILT -
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Logic and Block Diagrams
NOTE: GPC
CIB slot 8
Empty slot 1
PCI Backplane
Bus 2 (slots 1-4)
Mercury system Card slot 2
RACEway Acquisitor card slot 3
RACEway cable
Interlink Card PCI bridge slot 4 IRS Drives
Daughter card
Daughter card
Mercury expansion #2 slot 6 RAID 0 HD Array
BPP Backprojector card slot 7
CIB slot 8
(128 MB striped)
SCSI ADDR: 0-3
FAN
FAN
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Logic and Block Diagrams
MCB D2
comm stoprep
VCC_in Lightmarker_on
Power supply
DELPHI
D515
Lima_VCC Lima_rtn
E582
LIMA_R_2
DPU1 E581 E584 E583
E540 LIMA_R_1 LIMA_R_4 LIMA_R_3
W583
W581
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Logic and Block Diagrams
O2
O2 COMPUTER
COMPUTER
PCI SLOT
PCI SLOT
PCI SLOT
PCI SLOT
SLOT 4
PCI SLOT
PCI SLOT
PCI SLOT
PCI SLOT
PCI - PCI
BUS 0 BUS 2 BRIDGE BUS 3
2
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Logic and Block Diagrams
P-Plane Collimator
Comm Stoprep
MCB D2
OPEN
CURRENT
PWM
LIMIT MOTOR MOTION
DRIVER sign CONTROL
SERVICE
VCC2
DELPHI
POWER SUPPLY
D515
VCC2
W406 W408
W406
Det_mot- Det_mot+ Inc(5:0)
MOTOR
DPU1 M1
E450
E1 max min
DMS
ENCODER
OPTICAL SENSORS E850
P-PLANE COLLIMATOR
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Logic and Block Diagrams
MONITOR
SYNC_ERROR
MCU
MCU
RTC
MONITOR
REC_ERROR
Aquisitor
RTC
MCU
PHS
Speed_OK
XC_Gen
Ser_RX
Com_OK
n/a n/a n/a n/a
XC_Ena
Ser_Tx
HTP
n/a
AP90
AP0
Xray
Index
0 1 2 3 4 5 6 7 0 1 2 3 4 5 6 7 GANTRY
STATIONARY
DEMULTIPLEXER CONTROL MULTIPLEXER
MCU_RTC
D301
0 1 2 3 4 5 6 7 0 1 2 3 4 5 6 7
GANTRY
Ser_RX
Speed_OK
XC_Gen
K_MCU
XC_Ena
HTP
Ser_Tx
Com_OK
Xray
Index
AP90
AP0
MONITOR
ACU
LP S
MO NI TO R
ACU
ACU
ACU
ACU D513
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Logic and Block Diagrams
SLIP RING
X371
N SWITCHED 230VAC W503
W308
PDC X257 L1
X372
POWER N
DISTRIBUTION X373
L2 W326 PERMANENT 230VAC
CABINET X374
X3 DPU1
W501
X3 DPU2
W502
C-BOX
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Logic and Block Diagrams
CTL_DAT_SR_Neg X16
X342
CANBus
StopReport
CTL_DAT_RS_Pos
X343
CTL_DAT_RS_Neg
X344
W550
MCU/RTC PCB RS means ring to DLP
X2 W364 stationary
D301
CommPos X345
X6
zco
CommNeg X346
ACU
StopReport
CANBus
SLIP Acquisition
RING Control
StopRep_SR Unit
X347
StopRep_RS
X348
+12VDC X22
XMTR
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Logic and Block Diagrams
Stop/Report Loop
PINS
ResStopRepCir Signal +24V GANTRY FRONT FAÇADE
SERVICE STOP OFF ESTOPS
(from MCU) 8
+24V
S10 W362 X322 X323 W391 W390
K4 SERVICE STOP S12 X9 J39
REPORT K15 15
X16 COUCH
W52 D702 D701 IN/OUT
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Logic and Block Diagrams
UPS2
Auto Xmfr F4 D K4
Protection 16A 44A
e
Circuit l
F200
40A a
L1 y
K7 X269
L2
K9
K8
L3
T1 F5 F15
N 10A 6AF16
L1 460VAC F17 PDC 2A S2
L2 TO 10A DOOR K10
L3 X262 S1
N
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Logic and Block Diagrams
INTERCOM
WATER addr:5 (NOT USED) LASER
MARKERS CONTROL
COOLING addr:B UHR COLLIMATOR
AIR SCAN FOCAL SPOT addr:C STOP
CONTROL COMPENSATOR DEFLECTION REPORT SYSTEM
ROTATING
addr:A TUBE MONITOR LOOP POWER
GANTRY CBOX
WCS addr:9 KEYSWITCH
CAN BUS
F/O
SCAN DATA
SIGNAL DISTRIBUTION BOARD SDB
FROM RCOM
UltraWide Scsi Bus
MOUSE MONITOR
SLOT 1 EMPTY
9GB
BPP Backprojector card
Mercury expansion #2
Mercury system card slot
KEYBOARD
PCI Bridge slot 4
Mercury expansion #1
CIB slot
SCSI
Acquisitor card slot
ADDR:0
prep/raw SGI O2
8
M/B POWER
L/R AUDIO
SCSI
9GB/18GB COMM 2
ADDR:1
LED
slot 6
slot 5
slot 7
prep/raw SCSI
ADDR:0 COMM 1
9GB
SCSI
9GB/18GB
ADDR:2
SCSI
prep/raw
IRS PCI BUS 1 HCBB ADDR:1
9GB RACEway IMAGE RECONSTRUCTION PCI
SCSI RACEway cable BRIDGE 0 EXT EOD TERMINATOR
Interlink Card SYSTEM CDROM
ADDR:3 SCSI SCSI ADDR:3
prep/raw
Daughter card
Daughter card
Daughter card
ADDR:4
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Logic and Block Diagrams
W865
SIGNAL DISTRIBUTION BOARD
SDB
W326
X250
FROM T1
F18 L1 L2 L3
3A L1 L2 L3
X262
50/60 Hz
+24VDC K1
SUPPLY
K2
F3
4A L1 L2 L3
UPS1 400VAC
TO F8
Auto Xmfr F1
F4 44A
Protection 16A F14
Circuit 20A
F200 T1
40A
L1
K6
L2
L3
N
L1 460VAC
L2 TO
L3 X262
N
Power Distribution Cabinet
PDC
X263 X263
W4
Water Cooling
COOL
System OPTIIONAL
WCS
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Logic and Block Diagrams
Chilled
Water
Input Chilled Water
Output
HEAT EXCHANGER
Pressure
Pressure Gauge Port
Gauge Port
Air
Vent
Valve
Water Temp R2
2C - 16C Temp
35.6F - 60.8F Sensor
WATER TEMP
17.5C - 19.5C
63.5F -67.1F WCS
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Logic and Block Diagrams
TUBE_SYSTEM
230VAC SWITCHED OIL PUMP
X371 AND X372 E550
PDC
K4
SLIPRING
XRAY OIL COOLER
Z301
F8 13A F306 10A
230VAC L1 L1 W308 W501 L1 W519
FROM F1 44A
FAN
NTC_neg
Pmax_pos Temp_pos
S3 S2
S1
Pmax Switch
Temp Switch
POWER Pmax_neg Temp_neg NTC_pos
SUPPLY
STATIONARY FILTER FILTER FILTER
PSS
W558
GANTRY LEFT VCC VCC
STAND An An An
GANTRY DELPHI D515 MCB D2
ROTATING
Stoprep Comm
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Logic and Block Diagrams
X-Ray System
DCC POWER D450 HV
L1 Inverter 1
TANK
POS LED
D441-1 D442-1
L2 EMC
FILTER
D440
L3
NEG LED
DELPHI
Inverter 2
D441-2 D442-2
RELAYS K2 AND K3
FILAMENT ANODE
CONTROL CONTROL
SLIO D470 D115
CAN
D1
DCC CONTROL
D451
MAIN CONTROL MCB
D400 D2
MCU/
RTC
D301
E-BOX ELECTRONIC BOX
GENERATOR ACU
D513
POWER
SUPPLY C-BOX
E606 E510
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Logic and Block Diagrams
R1 - Temp Sensor
F3 3.5A
M5
K3
Communication
20A
L3 L19
L3
Valve
L4
Valve
MCU
L2 L14 Closing TEMP X15 Communication
X3
PUMP Operating
Pump ON 1.0A SENSORS
L2 Overcurrent DCV
460VAC GND
400VAC
Not Active WCC X9
WCC LEDS
L1 X260 W4 X1 F9 1A
3 32
L1
230VAC 24VAC
N WCS
K6 1 T1 31
PDC F8 3A
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Preventive Maintenance
PREVENTIVE MAINTENANCE
Select a topic:
Safety Precautions
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Preventive Maintenance
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Preventive Maintenance
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Preventive Maintenance
Couch
Select a topic: Tabletop Rails
• Lubrication 1. Move the tabletop into the gantry as far as possible.
• Lift and Feed 2. Remove the table foot end cover plate.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Preventive Maintenance
Couch, continued
Lift and Feed
Tabletop Movement
1. Turn off power to the system.
2. Move the tabletop into the gantry.
3. Attach a spring gauge to the table foot end handle.
4. Pull the tabletop fully out of the gantry steadily, and
verify that no more than 160 N is required to move
the tabletop.
Top Support Movement
1. Turn off power to the system.
2. Move the tabletop into the gantry.
3. Attach a spring gauge to the top support.
4. Pull the tabletop fully out of the gantry steadily, and
verify that no more than 160 N is required to move
the top support.
Compression Spring
Check the compression spring for damage, cracks and
breakage.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Preventive Maintenance
Dehumidifier
Replace the filter once a year, or as needed, depending
on site conditions.
Filter
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Preventive Maintenance
Gantry Fans
Verify the following fans are working:
• Two on Heat Exchangers of the X-Ray Tube.
• One behind each of the 3 radiators in the Gantry.
• Two on the Data Measurement System (C-Box).
• Four in the high voltage electronics box (E-Box).
• Two on the HV Tank high voltage transformer unit.
• One on the C-Box.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Preventive Maintenance
Gantry Rotation
1. Turn on the gantry at the CT Box. 10. Rotate the gantry several turns.
2. Lower the couch fully. 11. Repeat steps 6-10.
3. Open the gantry front cover. 12. Remove the grease gun, and wipe away any excess
lube.
4. Disable gantry rotation with S301 service switch,
located on left side of Gantry front. 13. Enable gantry rotation with S301 service switch.
5. At the CT Box, do a System Power Down of the 14. At the CT Box, do a System Power Up of the gantry.
gantry.
15. Make a scan using a 2-second rotation speed.
DANGER Sliprings are energized. 16. At the CT Box, power down the gantry.
17. Replace gantry covers.
6. Remove Gantry Rear cover.
7. With a grease gun, apply a small amount of lube to
the nipple.
Note The main bearing lube nipple is located on the
rear of the gantry, behind the slip rings, at the 2
o'clock position.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Preventive Maintenance
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Preventive Maintenance
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Preventive Maintenance
IRS
Slot #2
Mercury System Card / Slot #7
Array Processor Backprojector
Slot #3 Slot #8
Acquisitor CAN Bus
Board CIB
Slot #4
PCI (Backplane Power
Controller) Bridge Supply
Slots #5 and #6
Mercury Expansion
Card / Array Processor
Raceway
Board RAID-
configured
Hard Drives
Fan (3 or 4 drives,
depending on
version)
Fan
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Preventive Maintenance
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Preventive Maintenance
Safety Controls
Radiation Indicators E-Stops
1. Set up any scan with x-ray, and press [MANUAL 1. Set up an axial scan.
SCAN].
2. When [MANUAL SCAN] appears, press any E-stop
2. Verify the following x-ray on indicators are lit: button.
• 1 indicator at the CT Box 3. Verify the following conditions:
• 2 indicators on the two gantry control panels • System status changes to Stop
• any additional indicators on the site • All motors disengage, preventing movement
• Scanning is locked out until the E-stop is reset
4. At the CT Box, turn the gantry off and back on again.
5. Repeat this test for each E-stop button.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Preventive Maintenance
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Preventive Maintenance
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Preventive Maintenance
UPS
1. Logout of the system, but leave the gantry on.
2. Turn off power at the wallbox.
3. The console should continue to run for 7 minutes.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Preventive Maintenance
WCS
1. Turn the system off at the CT Box.
2. Open the WCS cabinet.
3. Connect a pressure gauge (in the Plumbing
Section) to the valve and check the pressure.
Pressure should be 1.8-2.2 bar.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Software and Diagnostics
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Software and Diagnostics
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Software and Diagnostics
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Software and Diagnostics
CT Box Diagnostics
1. Select Service then Procmon.
2. Select CT box in the Pmon window
3. Test the bottom 4 buttons on the CT Box and watch
the display.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Software and Diagnostics
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Software and Diagnostics
Filming
Filming Modes
• DICOM with standard network camera configurations
• AMC4 interface, external digital output to camera
(uncommon)
Configuring DICOM Camera
1. At Startup screen, login as config. Use config95
as passsword.
2. Select network.
3. Follow prompts to add a network node.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Software and Diagnostics
• CE Failures irsmon
Like the ps24 command, the irsmon command is used
• ps24 Command
after attempting reconstruction of raw data to identify a
• irsmon Command failed CE (sharc) or associated device.
Tests Hinv
• Incremental check of each sharc (CE) COMPUTER Views the devices attached to the IRS system. Replace
ELEMENT OR SHARC any devices that don't show up.
• Data pattern is transferred between every 4 sharcs
(CN) COMPUTER NODE
• RACEway (bus) integrity between all cards
• Writing from sharc to sharc through the RACEway bus
• First CE of each CN writes to the expansion memory
• Loopback test to verify SCSI comm (system card)
• Write/Read to SCSI disk Sector 0
Continued on next page
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Software and Diagnostics
1. Select unix shell from the desktop. Mass Storage PCI: Bus 2, Slot 6, Function 0, Vendor ID 0x1134,
Device ID 0x1No driver
Bit3 PCI Bridge Card: Bus 2, Slot 4
2. Type hinv and press [ENTER]. A list of the system Bit3 PCI Bridge Card: Bus 1, Slot 4
hardware, such as the following, appears: Bit3 PCI Bridge Card: Bus 0, Slot 3
Integral SCSI controller 1: Version ADAPTEC 7880
Video: MVP unit 0 version 1.4
Integral SCSI controller 0: Version ADAPTEC 7880
with no AV Card or Camera.
Vice: TRE
FLASH PROM version 4.17
On-board serial ports: 2 Note The system hardware display above is based on
On-board EPP/ECP parallel port the following IRS configuration:
1 200 MHZ IP32 Processor
FPU: MIPS R5000 Floating Point Coprocessor Revision: 1.0
CPU: MIPS R5000 Processor Chip Revision: 2.1
Data cache size: 32 Kbytes
Instruction cache size: 32 Kbytes
Secondary unified instruction/data cache size: 1 Mbyte on
Processor 0
Main memory size: 1024 Mbytes
Iris Audio Processor: version A3 revision 0
Integral Ethernet: ec0, version 1
Optical disk: unit 3 on SCSI controller 1
CDROM: unit 4 on SCSI controller 0
Disk drive: unit 1 on SCSI controller 0
CRM graphics installed
Mass Storage PCI: Bus 3, Slot 7, Function 0, Vendor ID 0x1134, Note If a card is missing from the system hardware
Device ID 0x1No driver inventory, that card is dead.
Unknown Type PCI: Bus 3, Slot 6, Function 0, Vendor ID 0x10b5,
Device ID 0x906dNo driver
Unknown Type PCI: Bus 3, Slot 5, Function 0, Vendor ID 0x10b5,
Device ID 0x9050No driver Continued on next page
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Software and Diagnostics
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Software and Diagnostics
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Software and Diagnostics
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Software and Diagnostics
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Software and Diagnostics
MX 8000 Passwords
Login Password
root pkserv99
elserv95 (mxview)
configuration config95
service service_only
terminal mode letmein (xterm)
Service (SERVICE TERMINAL)
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Software and Diagnostics
Network Configuration
Basic Network Configuration 12. Confirm the settings are correct, then click OK.
1. Ensure the the SGI O2 Computer is connected to a 13. A dialog box will open, stating The network
10baseT connection. interface "Ethernet ec0" was
successfully configured.
2. Login as root with password pkserv99.
14. Doubleclick on Modify Network Settings. This
3. In the Toolchest, select System, System
Manager. section of the network set up configures the network
routing, host table, and DNS (Domain Name Server).
4. In the left column under System
Administration, select Network and 15. Read the first screen and then click Next.
Connectivity. 16. Select enable for Primary Network Interface and
then click Next.
5. Under Network and Connectivity, select Set
Up and Start Networking. 17. Click on Add Default Route and type the Default
Route IP address. Click Next.
6. Use the host name, IP address and subnet mask
provided by the network admistrator. 18. Select enable for IP Forwarding, then click Next.
7. Read the text in the first screen, then click Next. 19. Skip step 5, Add an entry to the hosts
8. Again, read the information on the screen. Since the file. Click Next.
MX8000 has only one ethernet interface, click Next. 20. For the Set DNS name Server entries, type
the DNS info for Name Server 1. Click Next.
9. Type the Host Name and IP Address.
21. Confirm the settings are correct, then click OK.
10. Uncheck the box labeled Use Default Netmask.
22. When complete, logout of root by clicking on the
11. Enter a subnet mask of fffffc00 or whatever is
Desktop menu and then selecting Logout.
provided to you by the facility. Click Next.
Continued on next page
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Software and Diagnostics
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Software and Diagnostics
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Software and Diagnostics
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Software and Diagnostics
System Cooling
1. Under Service, select Show Temperature.
2. Screen display includes:
• Gantry temperature
• Gantry humidity
• Inlet water temperature
• Cradle temperature (this is the detector cradle)
• DMS temperature for left and right
• Tube Heat Units readings taken with every scan
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
THEORY
Select a topic: Gantry Tilt
Components
CAN Bus Tilt Motor
Collimator Stop
P-Plane Collimator Power Supply
Ultra High Resolution (UHR) Collimator Faults
Computer Rack Pulses from/to the Tilt Encoder
Interfaces Gantry Rotation
Power Image Reconstruction System
Signal Distribution Board IRS Architecture
AMC-4 CAN Interface Board (CIB)
Couch Stop Report Circuit
Vertical Drive Movement Power Supply
Horizontal Drive Movement CAN Interface
Couch Range 24 SHARC (CE) Configuration
CT Box Mercury Cards
DAS Data Flow Parallel Backprojector Board
DMS PCI_CON (PCI Control) Board
Scan Data Flow hinv Command
Dehumidifier Patient Alignment Lights
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Power
Couch
Data Acquisition System (DAS)
Gantry Rotating
Gantry Stationary
Power Distribution Cabinet (PDC)
Slip Ring Control Data Transfer
Stop/Report Loop
Closing the Stop Report Loop
Opening the Stop Report Loop
Water Cooling System
Components
Signals
LEDs
X-Ray Detection
X-Ray Generator
Function
Main Assemblies
Gantry Generator Power Supply (E606)
Electronic Box (E603)
High Voltage Tank (E604)
Filament Power Board (D470)
Rotating Anode Control Board (D115)
Anode Drive
X-Ray Tube Cooling Monitoring
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
CAN Bus
CAN stands for Control Area Network, which is a serial board to the MCU. The MCU would then put that
40 ma current loop network bus system inside the CT. It command on the CAN Bus, where it would be picked up
is how the CT communicates internally. by the XRS controller, which in turn would boost the
anode.
All the functions (rotation, tilt, collimation, etc) of the CT
scanner have their own computer controller, called MCB When errors are reported from the MCBs to the host
(Micro Controller Board). computer, the error code always starts with address of
the MCB reporting the error. For instance, if error E 30
Each MCB has a microprocessor, an operating system in
02 23, safety_rear_switch were found in the error
EPROM, firmware command in Flash Prom, RAM
log, this error came from the PTH Board, because the
memory, and a serial communication port.
error starts with 3.
All the MCBs are connected to each other through this
The only exceptions to this rule are the SLIO and WCC
CAN serial bus system. The following page lists the
Board. Even though they sit on the CAN bus, they
MCBs.
communicate exclusively with other MCBs (SLIO with
The Host computer is not a participant on the CAN Bus. XRS, and WCC with MCU) and are considered part of
Instead, it sends commands to the MCU/RTC Board via that other MCB's function.
a serial link, and the MCU in turn puts the commands on
MCBs can be swapped with each other as a
the CAN Bus.
troubleshooting step, but they must then be flash
Each MCB has a unique address on the CAN Bus. When programmed to work in the slot they have been placed.
the Host computer wants control an action in the gantry, (Only the OEM can flash the MCBs.)
it codes the command with the address and sends the
command out to the MCU, which in turn puts it on the
bus.
For instance, if the Host wanted to boost the anode up to
speed, it would send a serial command out the IRS CIB Continued on next page
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Collimator
Select a topic: P-Plane Positioning
• P-Plane Collimator The motion control circuit compares the MCB_DB value
with its internal actual positioning counter. According to
• Ultra High Resolution (UHR) Collimator the difference, it sets the signal sign, which connects a
P-Plane Collimator 24 V pulse width modulated voltage to the M1 motor. The
motor rotates clockwise or counterclockwise, thus
A positioning encoder (E1) on the DMS provides changing the P-Plane aperture. The motion control stops
feedback about the actual p-plane collimator position. the motor if the contents of the internal actual position
The encoder outputs the following signals to the C-Box counter is equal to the MCB_DB data.
DELPHI:
Service Switches
• CH0_POS/NEG - Increment Pulse 0°
On the C-Box:
• CH90_POS/NEG - Increment Pulse 90°
• Service - Selects service operation
• INDEX_POS/NEG - Index Pulse
• Open - Opens the p-plane collimator
All three signals are processed by the motion control
circuit. This circuit has an internal actual position counter • Close - Closes the p-plane collimator
which is incremented or decremented by the CH090 • Store
signals. The index pulse is set if the p-plane is at the
reference aperture.
Nominal P-Plane Collimator Position
The nominal P-Plane position MCB_DB is sent to the
motion control circuit by the MCBD. This value
corresponds to the selected slice thickness.
Continued on next page
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Collimator, continued
P-Plane Collimator, continued Power
LEDs The 24 VDC of the motor M1 is provided by VCC2 via
On the C-Box: C-Box DPU_1 P_24V.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Collimator, continued
Ultra High Resolution (UHR) Collimator To move the UHR collimator OUT, the current is set to
The hardware of the UHR collimator control circuit maximum after reaching the optimal limit switch OUT
position. The current is switched OFF with disable
consists of the following:
signals from the MCB. During the scan, the current is
• C-Box ZCON - Hardware position circuit motor switched to maximum; at the end it is set to low.
controller
Optical Limit Switches
• DMS - Ultra High Resolution (UHR) Collimator Define UHR collimator IN/OUT positions in UHR mode.
• M3 - DC motor for positioning IN or OUT of the UHR Service Switches
collimator
On the C-Box:
• Collimator
• Service - Select service operation
Nominal UHR Collimator Position
• UHR IN / UHR OUT- Moves the UHR collimator
The nominal position is sent to the motor controller by
MCB D2. The direction is set if the scan mode UHR. LEDs
Actual UHR Collimator Position On the C-Box:
The actual UHR collimator position is fed back by optical • UHR
limit switch IN or OUT. • HI_CUR - Displays the maximal current
UHR Collimator Positioning • UHR_LIM - When lit, UHR is in a limit switch.
To move the UHR collimator IN, set the current to
maximum after reaching the optimal limit switch IN Power Supply
position. After the current is held for one second, the The 24V for motor M3 is provided by PWR24 via C-Box
current is switched to low with a low current signal from DPU_1 P_24V. The PWR24 signal is controlled by MPE
the MCB. signals.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Computer Rack
The Computer Rack consists of two major computing Power
and processing systems:
The ON/OFF switch turns OFF the Computer Rack and
• Host SGI O2 Computer Gantry systems. The ON switch supplies power only to
the Computer Rack. The SGI O2 Computer must be
• Image Reconstruction System (IRS)
switched ON separately.
Interfaces The Power Distribution Unit (PDU) receives 230 VAC
• AC power input from PDC voltage from the Power Distribution Cabinet (PDC) and
distributes the voltage to all the components of the
• UPS Interface
Operator's Console.
• Optical fiber link with the DMS Detector and GI
Cards Signal Distribution Board
The Signal Distribution Board (SDB) is the interface
• Serial communication and control lines to/from MCU/ between the Computer Rack and the Gantry. The SDB
RTC Board
includes all the connectors which distribute the signals to
• CAN bus communication with the Gantry the system. The board is passive; it does not include any
electronic circuits. This board is located on the back of
• Audio signal lines with the Gantry
the computer workstation.
• Stop/Report Loop
AMC-4
• Network interface The Laser Multi-imager Interface provides the
• Teleservice communication connectivity between the O2 and the Laser Multi-imager.
The AMC is connected to the O2 through the parallel and
Serial Ports. This connection is for a direct connect only
and not for DICOM.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Couch
Select a topic: The spindle is held in position by a spindle brake. A tow-
rope potentiometer with signal conditioner records the
• Vertical Drive Movement
absolute couch height. The proximity release switch
• Horizontal Drive Movement defines the reference position for adjustment. Another
switch reports if an obstacle underneath the couch
• Couch Range
prevents the scissors from lowering the couch.
Vertical Drive Movement The E702 frequency converter (FC) takes single phase
Vertical motion is controlled by the D702 PTV Controller 230 VAC and converts it to three phase VAC to drive the
board (mounted at couch bottom rear). The D702 carries M702 lift motor. It requires a 230 VAC supply for drive
a D333 MCB board. The D702 controls the frequency power, a 12 VDC supply for the electronics, and
controller and motor (M702), brakes (Y702), and positioning control data. The 230 VAC is switched on and
receives vertical position from the Vertical Movement off via the D703 solid state relay which is only on when
Encoder (B702) and switches (S704, S701-703, U702). there are no errors present and the CT is reset. The 12
The D702 board also monitors the 24 VDC power VDC comes from the G700 power supply through the
coming from the G700 Couch Power Supply for the D702 board; it is not fused. The control data comes from
frequency converter and D703 relay. the D702 board as a serial data stream.
The vertical movement is achieved by rotating the ball The boundaries of couch movement are determined by
bearing screw or spindle to open or close the couch the mechanical limits of the couch. They are a function of
scissors. The spindle is driven by a 3-phase AC motor. the gantry angle, couch height and horizontal position.
Its supply is generated by a frequency converter. The These aspects define the Collision Check for which the
vertical motor is controlled by the PTV_MCB using the scanner automatically tests.
PTV frequency converter. This converter is physically the
Vertical range is 48 cm to 102 cm
same as the tilt one. There is a thermal switch to protect
the motor against overheating. Lifting speed is 10 or 20 mm/sec.
Continued on next page
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Couch, continued
Horizontal Drive Movement It also stops the Patient Table Vertical (PT_Ver D702)
The PTH Board (PT_Hor D701), with its on-board MCB controller from performing invalid vertical movements.
The PTH MCB D2 module is connected to the CAN bus
D2 controller, controls horizontal movement of the couch.
for communication, and is linked into the stop report loop
For communication, safety, and control, the board is
linked to the PTV Controller (PT_Ver D702) board, and for safety supervision.
(from there) to the rest of the unit. M711 Tabletop Motor
These drive systems handle horizontal movement: The M711 tabletop motor moves the tabletop with the
• Tabletop Drive help of a toothed belt and wheels. This 24 VDC motor
has a built-in gear of 1:36 and is supplied with pulse
• Top Support Drive width modulation voltage regulated by a motor controller.
Tabletop Drive Tabletop Motor Controller
The Tabletop Drive moves the table top (with the The tabletop motor controller is programmed by the PTH
patient) horizontally, in and out of the Gantry funnel for function controller to include acceleration, speed and the
positioning at a distance of 1651 mm (65 in.). This desired table position. From these parameters, the motor
section describes its modules and functionality. controller calculates a trapezoid speed profile for
A 24 VDC motor connected by a clutch to a toothed belt, movement regulation.
moves the tabletop. Two brakes hold the tabletop in The PWM and Sign Signals
position.
The PWM and Sign signals from the PTH Board are fed
Patient Table Horizontal (PTH) MCB D2 Module into the tabletop driver for power amplification. As a
The PTH MCB D2 (PT_Hor D701) board contains the on- prerequisite, the Stop/Report Loop (Stoppos and
board PTH MCB D2 module, with the PTH firmware Stopneg) must be closed to supply Table_24V to the
acting as function controller. This module controls driver.
horizontal movement of the tabletop and the top support. Continued on next page
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Couch, continued
Horizontal Drive Movement, continued U711 Proximity Switch
Tabletop Drive, continued The U711 proximity switch becomes active when the
B711 Increment Encoder tabletop approaches the end of its range at the back. The
U711 switch signal marks the reference tabletop position
The B711 increment encoder that is connected with and is evaluated by the PLD logic and the function
another toothed belt to the tabletop motor axis registers controller. U711 is used during adjustment and to set the
tabletop movement. The B711 is a high precision module zero position of the table.
that generates 10,000 pulses per revolution, which
correspond to 500 pulses per 1 mm (0.039 in.) of Tabletop R711 Potentiometer
tabletop movement. The tabletop Potentiometer roughly determines the
PLD Logic Module position of the tabletop after startup. The analog signal is
converted to digital as channel 1 of the D701 on-board A/
PLD logic module on the PTH Board D701 processes D-converter. The digital value is then reported to the
and evaluates the K1 and K2 signals. It then feeds high function controller. The tabletop positions and their
precision incremental pulses back into the tabletop motor corresponding potentiometer values are recorded during
controller for exact position regulation and uses a divider table adjustment.
to generate Horizontal Table Pulses (HTP) from the
incremental pulses. The HTPs are transmitted to the Y713 Clutch
PTH function controller, which sends them as differential The Y713 clutch separates the tabletop motor from the
HTP Pos and Neg signals via the PT_Ver D702 board mechanical motor drive assembly. The function controller
and the Gantry, to the DMS. applies the clutch with relay K6 and a 24 VDC supply.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Couch, continued
Horizontal Drive Movement, continued
Tabletop Drive, continued
E703 Foot Switch Pedal
The E703 Horizontal Release Footplate on the cable
duct between the Gantry and the couch, allows faster
patient positioning. This tabletop release is only enabled
when the top support is in the inner (not ICT) or outer
(ICT) position. The Foot Switch is not a safety feature. If
it is pressed, the function controller interrupts horizontal
table movement, and releases the tabletop, brakes and
clutch. In addition, the CAN bus stops Gantry tilt
movement. However, the brake for the top support stays
engaged.
Mechanical Tabletop Brakes (Y711 and Y712)
The mechanical tabletop brakes (Y711 top picture,
viewed from left, and Y712 bottom picture, viewed from
top) stop tabletop movement and hold the tabletop in
position. One brake is on the motor side of the tabletop
and the other is on the opposite side, at the Gantry end
of the toothed belt drive.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Couch, continued
Horizontal Drive Movement, continued
Top Support Drive
The Top Support Drive moves the top support (together
with the tabletop) horizontally to ICT, or normal position.
A 24 VDC motor that drives a toothed belt moves the top
support, and a brake holds it in position. A
potentiometer records the absolute position and an
increment encoder issues count pulses according to the
relative movement. A proximity reference switch defines
the reference position for adjustment of the top support.
The rear safety switches inform the PT_Hor D701 PTH
Board when an obstacle at the rear inhibits further
movement.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Couch, continued
Couch Range
Couch
Gantry Tilt Couch Feed-In
Height
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
CT Box
The CT box has two seperate inputs:
• Top half interfaces with GPC in the Gantry Right
Pedestal
• Bottom half interfaces with CIB in the Image
Reconstruction System (IRS)
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
RCVR
XMTR
W55 W590
IRS Acquisitor PCB
FIBER OPTIC FIBER OPTIC DMS
J115 RCOM
CABLE CABLE
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Dehumidifier
The gantry cooling, with heat exchangers and a The filter on the dehumidifier, in the Gantry Right
dehumidifier, keeps the internal gantry temperature at Pedestal, should be cleaned every PM.
26°C (78.8°F) ±4°C (±7.2°F), and the relative air
humidity between 20%-30%. The nominal temperature- Filter
tolerance and humidity values are adjustable by the table
T0C.
Phase L1, fused with F307 in the Gantry Left Pedestal,
is used to power the dehumidifier.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Gantry Tilt
Components In case of error, the FC sends the ErrorTilt message to
the MCB. For tilting the Gantry, the RTC_MCB has to set
The tilt circuit includes the following:
up the FC through the RS-485 controller with its values
• MCU/RTC Board D301 (CommTilt).
• E302 Tilt Frequency Converter, Gantry Left Pedestal Additionally, the FC has to be started and the motor
brake has to be released. A spindle converts the motor
• M302 Tilt Motor, Gantry Right Pedestal
rotation into a linear vertical movement which tilts the
• Y2 Motor Brake, Gantry Right Pedestal Gantry forward or backward. The actual tilt angle is
• B301 Tilt Sensor, Gantry Left Pedestal constantly watched from the RTC_MCB by evaluating
the output of the tilt sensor inside the Gantry.
• Power Supply Stationary, Gantry Left Pedestal
If the desired tilt position is reached, the RTC_MCB
Tilt Motor stops the FC and engages the tilt brake.
The M302 is a three-phase asynchronous motor with a In case of emergency, when a stop is activated, the
built-in mechanical brake to hold the Gantry in a defined Gantry tilt is immediately stopped by disconnecting the
inclination. A temperature switch, inside the motor single-phase supply to the FC with Solid State relay
housing indicates overheating. D324 and activation of the motor brake.
The motor is driven by a frequency converter (FC), which Dir1Tilt enables the converter with K11, MCU/RTC
converts the single-phase 230 VAC input into the Board, (15 V from E302 are looped back).
required three-phase 230 VAC motor supply. The
generated frequency and voltage determines the tilt Dir2Tilt defines the tilt direction with K12, MCU/RTC
speed. The single-phase supply for the FC passes the Board, (15 V from E302 are sometimes looped back).
solid state relay D324 in the Gantry Left Pedestal. The BrakeTilt releases the brake by connecting 24 V to the
converter is set up from the RTC function controller by brake assembly, MCU/RTC Board.
the RS-485 controller with its values (CommTilt). Continued on next page
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
The CIB is the interface of the Host to the Gantry via the • PBB Card
CAN bus. • CIB Card
A read buffer on the board stores up to 1,000 CAN • Multi Serial Card
telegrams for the ICS.
• PCILK Card
The CAN controller is an active card. An interrupt is
asserted to the PCI bus, when a telegram is stored in the
Read Buffer.
Continued on next page
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Power
Select a topic:
• Couch
• Data Acquisition System (DAS)
• Gantry Rotating
• Gantry Stationary
• Power Distribution Cabinet (PDC)
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Power, continued
Couch three AC supply phases (U, V, and W) for the spindle
During regular operation, the Power Distribution System motor from a single input phase.
(PDS) is secured by fuse F8 and phase two is set by • PT_PWR - The G700 Patient Table DC power supply
relay K4, which is operated by the control signal voltage, (see Undertable Components) supplies the low
GAN_TI, 2. The two supply lines for the couch are fused voltages of 5 V, 12 V and 24 V that are required for
in the PDS with fuses F15 and F16 inside the PDC. horizontal and vertical movement on boards D701
and D702, and for the clutch, switches, sensors,
Components
encoders, and brakes.
• S1 - For service purposes, the Table Power Switch
S1 in the PDC can power up the Patient Table. This • PT_Ver - The Patient Table Vertical (PT_Ver) board
switch bypasses relay K4 and allows the Patient Table (see Undertable Components) gets one 5 V (U1_2)
operation, even when the unit is switched OFF. and two 12 V supplies (U2_P, U3_P). They all pass
line filters on the board. Both 12 V supplies are
• Gantry Stationary - The Patient Handling System connected to a cascade to raise the total voltage to 24
(PHS) receives two lines (L_PHS1, L_PHS2) with 230 V.
VAC, supplied from phase two of the PDS. Both
power lines pass through the Power Supply • PT_Hor - The Patient Table Horizontal (PT_Hor) PTH
Stationary (PSS) of the Gantry Stationary and run to Board gets one 5 V (U1_1) and two 24 V supplies
the PHS through the cable duct on the floor. (U4_P) from the power supply, through fuses. On the
board, all supply voltages pass line filters to reduce
• Stop Relay - The 230 VAC line passes line filter Z702 line disturbances. PT_Hor also supplies the sense
and then solid state stop relay D703, where it can be signal to the power supply for voltage regulation.
interrupted by the system stop circuit in emergency
cases.
• Frequency Converter - The E702 frequency
converter (see Undertable Components) generates
Continued on next page
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Power, continued
Couch, continued • 24V_input - The 24V_input signal is a continuous
control supply for internal switches and current loops.
Power Supply Signals
Both 12 V supplies are connected to a cascade to
• VCC 2 - 5 V supply for the Patient Table Vertical raise them to a total voltage of 24 V.
(PT_Ver) board (see Undertable Components) is
generated from a common source in PT_PWR • 24V - The 24 V line is a switched supply for the
(connected via fuse F702). brakes. It is enabled by the stop and service logic,
with relay K6. Both 12 V supplies are connected to a
• VCC 1 - 5 V supply for the Patient Table Horizontal cascade to raise them to a total voltage of 24 V.
(PT_Hor) PTH Board is generated from a common
source in PT_PWR (connected via fuse F701). It also • Supp24VnotOK - The 24 V supplies for the top
generates the sense signal for the voltage regulation. support motor are monitored by the PTH function
controller. If the voltage drops below 18 V, the
• LPHS1 - In the couch, one power phase (LPHS1) Supp24VnotOk signal is issued as a warning.
(see Undertable Components), with the frequency
converter, drives M702 spindle motor of vertical drive. • Table24VnotOk - The 24 V supplies for the table top
motor are monitored by the PTH function controller. If
• LPHS2 - The second power line (LPHS2) (see the voltage drops below 18 V, the Supp24VnotOk
Undertable Components) supplies the G700 Patient signal is issued as a warning.
Table DC power supply (PT_PWR), which generates
all low voltage used on horizontal and vertical boards. • 24VnotOk - The 24 V signals are monitored by the
PTV function controller. If the voltage drops below 18
• Table_24V - This 24 V supply is used for the V, the signal 24VnotOk is issued as a warning.
Tabletop Drive (connected via fuse F705 (see
Undertable Components)). • 12V_FUR - A 12 V line that supplies the frequency
converter with its internal operating voltage 12V_fur.
• Supp_24V - This 24 V supply is used for the Top
Support Drive (connected via fuse F704 (see
Undertable Components)). Continued on next page
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Power, continued
Data Acquisition System (DAS) Voltage OK (Volt_OK)
Z_CON The following voltages are monitored:
There are fuses for two voltages: P_24V and VCC.
N13 -13 V Fuse monitoring DPU2, E541
All voltages from DPU1 and DPU2, in C-Box, except
P8 +8 V Fuse monitoring DPU2, E541
5V_P, are monitored on MCB D2 and generate one bit of
Volt_OK. 5V_P is connected to MCB D2. P5 +5.5 V Fuse monitoring DPU2, E541
P24 +24 V Fuse monitoring DPU2, E541
48V_P +38 V Fuse monitoring DPU1, E540
15V_N -15 V Fuse monitoring DPU1, E540
15V_P +15 V Fuse monitoring DPU1, E540
15V_P +24 V Fuse monitoring DPU1, E540
5V_P +5.4 V monitored on MCB, D2
Pwr24
The power supply for the driver and stepping motors is
24 VDC that is supplied by the Digital Power Unit 1
(DPU1). The 24 V are controlled by the MCB processor
with the signal MPE (Motor Power Enable).
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Power, continued
Data Acquisition System (DAS), continued Power Supply Tube Collimator Box
DELPHI The components listed here receive power from DPU1.
DELPHI Phi_Amp
There are onboard fuses on DELPHI D515 for three The Phi Amplifier (Phi_Amp) Board is divided into two
voltages: VCC3, P_24V and VCC. parts:
VCC2 • The Z-Pre-amplifier part is supplied with +5 V, -15 V
and +15 V from the DPU1 power supply.
VCC2 is derived via a filter from P_24V.
• The Phi-pre-amplifier part is supplied with +5 V, -15 V
PownotOK and +15 V from the DPU1 power supply.
The error Power Not OK (PownotOK) is generated if the Limit Switches
voltage monitoring on DELPHI D515 shows an error. An
error message is issued and the deflection is disabled. The Optical Limit Switches inside the Tube Collimator
Box are supplied with 5 V from the DPU1 power supply.
The following voltages are monitored:
Encoder
48V_P +238 V Fuse monitoring DPU1
The Optical Encoders inside the Tube Collimator Box are
15V_N -15 V Fuse monitoring DPU1 supplied with 5 V from the DPU1 power supply.
15V_P +15 V Fuse monitoring DPU1
24V_P +24 V Fuse monitoring DPU1
5V_P +5.4 V monitored on MCB, D2
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Power, continued
Gantry Rotating The Digital Power Unit 2 (DPU2) is a multiple output
power supply that generates the voltages shown:
Power Supply
The Gantry Rotating has two AC supplies that are P24 +24 VDC DMS, ZCON D514 (only monitoring)
connected through the Power Slipring Assembly E420.
N13 -13 VDC DMS, ZCON D514 (only monitoring)
DPU1 and DPU2 P8 +8 VDC DMS, ZCON D514 (only monitoring)
The Digital Power Unit 1 (DPU 1) is a multiple output P5 +5.5 VDC DMS, ZCON D514 (only monitoring)
power supply that generates the voltages shown:
The mains supply for DPU2 is permanently on, even if
24V_P +24 VDC ZCON D514, DELPHI D515, DMS the system is in off condition. The DMS is able to switch
15V_N -15 VDC DELPHI D515, DOM D516, Phi_Amp the DPU2 on and off during this time.
D555, ZCON D514 (only monitoring)
• P24 A 24 VDC voltage from Digital Power Unit 2
15V_P +15 VDC DELPHI D515, DOM D516, Phi_Amp (DPU2) connected to the C-Box backplane and the
D555, ZCON D514 (only monitoring)
DMS.
5V_P +5.4 VDC ZCON D514, DELPHI D515, DOM
D516, ACU D513 • L1_DPU1 The 230 V phase supply for Digital Power
Unit 1 (DPU1), the generator and the oil pump of the
5V_P1 +5.4 VDC Phi_Amp. D555; LPS, D522 tube cooling. The power is connected through the
48V_P +38 VDC DELPHI D515, ZCON D514 (only sliprings from the Gantry Stationary
monitoring)
• N13 A minus 13 VDC voltage from Digital Power Unit
2 (DPU2) connected to the C-Box backplane and the
DMS.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Power, continued
Gantry Rotating, continued • 15V_P A 15 VDC voltage from Digital Power Unit 1
(DPU1) connected to the C-Box backplane.
DPU1 and DPU2, continued
• 24V_P A 24 VDC voltage from Digital Power Unit 1
• P8 An 8 VDC voltage from Digital Power Unit 2
(DPU1) connected to the C-Box backplane and the
(DPU2) connected to the C-Box backplane and the
DMS.
DMS.
• 5V_P A 5.4 VDC voltage from Digital Power Unit 1
• P5 A 5.5 VDC voltage from Digital Power Unit 2
(DPU1) connected to the C-Box backplane.
(DPU2) connected to the C-Box backplane and the
DMS. • 5V_P1 A 5.4 VDC voltage from Digital Power Unit 1
(DPU1) connected to the C-Box backplane. The
• Power_disable_2 Switches the output voltage of
difference to 5V_P is an additional fuse in the DPU1
DPU2 on and off. This is done depending on the DMS
to protect boards without an internal fuse.
temperature, when the system is in off condition.
• 15C_N A minus 15 VDC voltage from Digital Power
Unit 1 (DPU1) connected to the C-Box.
• 48V_P A 38 VDC voltage from Digital Power Unit 1
(DPU1) connected to the C-Box backplane.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Power, continued
Gantry Stationary The B305 Temperature Sensor is included in the K4
The function of the PSS, Gantry Left Pedestal: circuit. As a result, the Gantry is powered off if its air
temperature exceeds 43°C (109.4°F).
• The PSS is the input terminal and internal distribution
In the PDS, one phase (L2) is split and passes through
unit for power in the Gantry. The PSS receives three-
the Gantry to supply the PHS. Additionally, L2 is also
phases of 230V, neutral and protective earth from the
Power Distribution System (PDS) in the Power connected as the permanent Power Supply Mains2
through the sliprings to the rotating part of the Gantry.
Distribution Cabinet (PDC), and supplies all stationary
Here L2 is used for the detector heating. In the PSS,
components with power.
Mains2 is protected by F309, Gantry Left Pedestal,
• The PSS receives the UDC-voltage from the Direct passes the S309 xxx switch and is filtered from Z305,
Current Circuit (D450 DCC Power PCB Cap Bank) in Gantry Left Pedestal.
the PDC and connects line voltage and UDC-voltage
Note Mains2 is derived from L2 in the PDS without
through the sliprings to the rotating part of the Gantry.
passing K4. In this way, the 230 V supply voltage
• The 230 V required to supply the Patient Handling (AC Power Brush Blocks) for the heating detec-
System (PHS) is passed through the PSS. tor is immediately available, if the external three-
phase power connection for the PDS is switched
From the Power Distribution Cabinet (PDC), three
phases of 230 V, and neutral and protective earth are on.
connected to the Power Supply Stationary (PSS). In the A second 230 V supply for the rotating part of the Gantry,
PDS, the phases are switched with K4 relay called Mains1, is derived from L3, filtered from Z301,
corresponding to the following states: Z303 and protected from F306. This 230 V supply is
• off K4 open controlled from K4 only in when the system is on.
• system-on K4 closed
• computer-on K4 open Continued on next page
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Power, continued
Gantry Stationary, continued Single-Phase Frequency Converter to drive the tilt motor,
Three phases of 230 V, controlled by K4, are connected Gantry Right Pedestal.
to the Z301 filter in the PSS from the PDS. Phase L1, The input voltage is protected with F302, filtered by Z304
fused by F307, is used to power dehumidifier, Gantry and can be interrupted with the D324 Solid State Relay
Right Pedestal. (all in Gantry Left Pedestal) whenever the signal Stop
Phase L2 supplies the two fans, M304 and M305 for the Tilt is set from the Rotate Tilt Controller (RTC).
Gantry cooling. The third fan, M306, is driven from phase To drive the rotate motor, L1, L2, and L3 are connected
L3. These fans are permanently running in the system- to the E301 Three-Phase Frequency Converter. All three
on state. phases are protected with F301 and filtered from Z301.
Controlled from K4 in the Power Distribution Cabinet In the stop state, rotation has to be finished as soon as
(PDC), phase L1 is connected to the Z301 filter, Gantry possible. This is done by disconnecting the output of the
Left Pedestal, in the Power Supply Stationary (PSS) frequency converter from the rotate motor and
and through F303 to the E303 (both in Gantry Left connecting the braking resistors to the motor.
Pedestal) Low Voltage Power Pack. This is a multiple
power supply, that generates 5 V, 12 V and 24 V. The
Gantry Panel Controller (GPC) is supplied with 12 V and
24 V. The D344 Audio Board receives 5 V and 12 V.
Three phases of 230 V are connected to the Z301 filter in
the Power Supply Stationary (PSS) and are controlled
from K4 in the Power Distribution Cabinet (PDC). Phase
L1, fused with F303, drives the E303 low voltage power
pack. This is a multiple power supply that generates 5 V,
12 V and 24 V; all being connected to the D301 MCU/
RTC Board. Also, phase L1 is connected to the E302 Continued on next page
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Power, continued
Power Distribution Cabinet (PDC) Results after the above have occurred:
Power Up • K2 establishes its own selfhold so that K2 stays
The PDC is switched from the keyswitch in the CT Box energized if the key switch is released.
and the switch in the Computer Rack. • The potential-free contact to power on a camera
The CT Box is switched to GANTRY ON, relay K2 in the (CAM) for the digital documentation on film is closed.
PDC is activated and full power is connected to all • The LED in the Computer Rack lights up.
components via:
• If the system previously was in COMPUTER ON state
• K3 for the Computer Rack power (K1 is energized), then K1 drops and all previous
• K3 to the reserve and service sockets and second power connections are taken from K2.
monitor supply • The DMS power supply (C-Box) through K9 is active
• K4 to all Gantry components, except DMS power in all system states.
supply The system is now ready to scan.
• K5 to DCC
• K6 to WCS including a potential-free contact to power
ON, e.g., an air conditioning unit (COOL)
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Power, continued
Power Distribution Cabinet (PDC), continued Direct Current Circuit (DCC)
Service Switch S1 The D450 DCC Power PCB Cap Bank generates a DC
The Table Power Switch S1, in SERVICE position, voltage of 560 V to supply the inverters in the HV
generator. The DCC converts the filtered three-phase AC
connects L2 to the couch bypassing K4. This allows
voltage of 400 V received through F100 and the T1
table movements, with the Couch Service Switches,
even in the SYSTEM OFF state. The second contact of transformer (in the PDC) into a DC voltage of 560 V
(UDC_pos, UDC_neg), using a six-pulse rectifier to
S1 interrupts the control line for the K4 and K5 relays, so
charge a capacitor bank (both on D450).
that powering ON of the Gantry and the Generator, using
the GANTRY ON switch in the CT Box, is not possible. Power Supply
This switch is to be used during the tube replacement The D451 gets 24 V from the permanent PDC 24V
procedure to move the table up and down to pull the tube Power Supply (24V_keyboard), in the PDC, to generate
out. all required DC voltages. The D451 supplies the SLIO
Service Switch S2 with +5 V (Vcc).
The Power Safety Switch S2, in SERVICE position, Serial Linked Input Output (SLIO) Board
interrupts the control line for the K4 and K5 relays, so The Serial Linked Input Output (SLIO) board D1 is on the
that powering ON of the Gantry patient table and the D451 and serves as connection to the CAN bus. The
generator, using the GANTRY ON button in the CT Box, SLIO board is part of the CAN bus, and is only used to
is not possible, or will be switched off (same function as receive commands from the generator
second contact of S1). (ClosePowerRelays, CloseChargeRelays) or transmits
This switch can be used during service activities at the data (PowerRelayClosed, ChargeRelaysClosed) or issue
Gantry, because it inhibits the power up of the whole errors (TemperatureSwitch, TemperatureError;
system with the GANTRY ON switch in the CT Box. As a StopStatus) to the generator.
result, service can be performed safely on those
components. Continued on next page
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Power, continued
Power Distribution Cabinet (PDC), continued • Ana_gnd - Analog ground
Signals • StopStatus - Informs the generator about the state of
• ChargeRelaysClosed - Feedback to the generator the Stop/Report Loop
indicating that the charge relay is closed • TemperatureError - Generated if the temperature on
• PowerRelaysClosed - Feedback to the generator D451 exceeds 60°C (140°F). The SLIO transmits this
indicating that the power relay is closed error to the generator, responding with a fatal error.
• TemperatureSwitch - This switch is set if the • Door - This signal represents the condition of the
temperature switch, mounted on the optional T1 door-switch and is used in the MCU/RTC Board D301
transformer (in the PDC), was opened with to interrupt high tension if the door is opened while
temperatures above 170°C (338°F). As a result, relay scanning.
K5 (in the PDC) drops and the Stop-Report loop is • Xray - This signal is issued from the Generator with
interrupted. high tension above 20 kV and used to drive the X-ray
• ClosedPowerRelays - Request from the generator, warning lamp.
sent after CloseChargeRelays, to provide full energy
for the main inverter. This signal is permanently active
after power ON.
• CloseChargeRelays - Request from the Generator,
sent after power-on for a fixed duration, to close the
charge relay for a smooth charging of the empty
capacitor bank.
• P_24V - Permanent PDC 24V Power Supply Used to
energize PDC relays when the On/Off switches are
set on the CT Box Gantry Power and Computer Rack.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Stop/Report Loop
Closing the Stop Report Loop 6. After the start of the CAN communication, the MCU
sets ResStopCir for 100 ms to enable the Gantry and
Each MCB closes its Stop Report Contact as soon as the
Patient Table movements.
system is powered on if there is no malfunction detected.
Whenever a RESET is performed, the system closes the 7. K2 (MCU/RTC Board) is activated and its self-
whole Stop Report Loop in the following sequence: holding contact is closed. K2 remains energized if
ResStopCir is cleared.
1. The SGI O2 Computer (ICS, O2, Super) sends the
RESET command via the CAN bus to the MCU/RTC 8. Through the now closed K1 and K2 contacts, 24 V is
Board, and from there to all MCBs. used to enable rotation, tilt and PHS movements by:
2. The MCU sets Stop Hold to energize K3 and CLOSE • Activating the D703 (Undertable Components)
the Safety Contact which in turn provides GND to the solid state relay in the couch and connecting line
Stop Report Loop and to the Stop Circuit. power to the Couch Up/Down Circuit.
3. ResStopRepCir is set for 100 ms and looped through • Activating the D324 (Gantry Left Pedestal) solid
all participants. The function controllers have closed state relay in the Power Supply Stationary (PSS)
the relay contacts in the INIT phase already. which connects line power to the tilt frequency
4. K1 (MCU/RTC Board) is energized and its self- converter.
holding contact is closed. K1 remains energized if • Activating the K16 relay on D301 (which
ResStopRepCir is cleared. connects the output of the rotation frequency
converter (Gantry Left Pedestal) to the rotate
5. Now there is a current flow in the Stop Report Loop.
The Stop Report Loop is closed. With the raising motor, instead of the braking resistors (Gantry
Rear)) thus making rotation possible.
edge of the current, all Function Controllers get a
hardware reset. The MX8000 is in the STANDBY state and all functions
are now available.
Continued on next page
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
L1 The firmware of the Water Cooling Control (WCC L16 Reset_WCC is active from the Master Control
Unit (MCU/RTC Board).
Board) is running.
L17 The output line is active.
L2 The service function of the WCC is active.
L18 The output line is active.
L3 The regulation drive should move towards
warmer water (open bypass valve). L19 The output line, to switch the water pump ON, is
active.
L4 The regulation drive should move towards colder
water (close bypass valve). L26 WCC supply voltage is present, after the rectifier
and the onboard fuse.
L6 An error was detected in the water cooling
system.
L14 The water pump F3 overcurrent protection
responded.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
X-Ray Detection
There are four quality grades of Detector. A, B, C, and D. GI / Module Table
A is the highest, and is at the center of the DMS. D is
lowest quality, and is positioned on the outermost edges
of the DMS.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
X-Ray Generator
Function • Electronic Box (E603) - The electronic box contains
the following:
The generator’s function is to:
– EMC filter D440 shields the slip rings from the high
• Supply the high voltage for the X-ray tube kV_neg
frequency noise produced in the main inverters.
and kV_pos at one of the following levels:
– Buffers 1 and 2 serve as energy buffers for the high
– 90 kV
voltage.
– 120 kV
– Main Inverters (1 and 2) convert the 560 VDC input
– 140 kV signals UDC_pos, UDC_neg from Direct Current
Circuit (D450 DCC Power PCB Cap Bank) to a
• Generate and regulate the filament current, which
frequency regulated AC voltage (f> 20kHz) that is
indirectly regulates the tube current, and, thusly, the
dose during exposure. connected to the high voltage transformer.
• Accelerate, decelerate and monitor the anode of the – Resonance Circuits 1 and 2, each consisting of two
capacitors and one inductor, in parallel, build the
X-ray tube.
oscillating circuit for the two main inverters.
Main Assemblies – FIL Control board D470 generates the tube
The X-ray system (Gantry generator) includes the filament current.
following main assemblies:
– RA Control board D115 drives the rotating anode.
• Generator Power Supply (E606) - Receives L, N and
PE from the Power Supply Stationary (PSS) which – Main Control Board, the main controller of the
pass through the slip rings. It also generates the whole generator.
following voltages and distributes them to the
generator components: +15 V, -15 V, +5 V.
Continued on next page
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Theory
Gantry Rotation
The MCU/RTC Board is the Rotating and Tilt Controller. The rotate motor is made up of 100 permanent magnets
It is located in the left pedestal on the same main board mounted on the rotating part and 100 field coils mounted
(D301) with the MCU. It communicates with the in the stationary part. To rotate the gantry, the RTC
frequency converters for rotation and tilt. board sends commands to the rotate frequency
converter, which in turn drives the field coils with three
The RTC controls rotation by detecting the revolution
phase voltage. As the coils are energized, they create
speed and tube position by light barrier pulses from the
polarity which reacts with the permanent magnets
rotating AP slot ring and controlling the Rotate
causing gantry rotation.
Frequency Converter, Gantry Left Pedestal.
To stop rotation the RTC uses dampening resistors
The LUST converter receives three phase line voltage
(Gantry Rear) that are activated by relays K9 in RTC
from the gantry PSS which got it from the PDC. The
and K1-3 in the E322 Rotate Brake Assembly. These
converter does not regulate the speed. It generates
resistors are also engaged when the S301 safety switch
output according to its programmed parameters. When
is activated.
a fixed azimuth is required such as for TOPO, the
converter supplies a constant voltage to hold the gantry. The gantry can only be permanently stopped from
The motor is held by the magnetic attraction. This rotating by engaging the rotation safety bolt mounted in
requires that the gantry is balanced. the upper right corner of the Gantry.
There are 1440 Angular Pulses (AP0) from the LPS The GPC MCB (Gantry Right Pedestal) monitors table
D302 and one Index AP for each rotation. position, gantry tilt, gantry door switches, power levels,
temperature, humidity and lasers and will alert the MCU
MCB whenever an unsafe condition arises so that it can
stop gantry rotation.
The MCU_MCB will send a STOP signal to the Rotate
Converter and its PSS power relays whenever a STOP
condition occurs.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
TROUBLESHOOTING
Select a topic: Data Storage and Management
Can’t Communicate With Spooler
Safety Precautions Change the System Host
Core Dump File
Audio and Autovoice Hard Drive Fault
Patient Can’t Hear Tech IRS Raw Data Disk Reformat
Tech Can’t Hear Patient Detector Collimator
Patient Can’t Hear Autovoice P-Plane Collimator
CAN Bus Ultra High Resolution (UHR) Collimator
Error Log Troubleshooting Emergency Stop Switch Cannot Be Reset
MCU Troubleshooting MCU Manual Reset
Minimum Configuration Troubleshooting Filming
Computer Ping the Camera
Maintenance Tests Connecting a Dryview 8700
Shutdown Problem or Mouse Lockup Gantry Rotate
Hard Power Cycle
Gantry Tilt
Computer - SGI O2
Image Quality
Copy the O2 9 GB Drive to an 18 GB Drive
CT Numbers Out Of Calibration (Level)
Couch Ring Program
CT Box Troubleshooting Chart
DAS Data Flow RF Fiberoptic IRS
Scan Without Gantry Rotation Testing Recon
CE Check
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
CAN Bus
1. Study the error log for CAN Bus errors. Some CAN
Bus errors have extra data that can be evaluated to
find the problem. Refer to Error Log
Troubleshooting.
2. If step 1 is not conclusive then you will need to
troubleshoot using the MCU CAN Bus test points
and function controller LEDs. Refer to MCU
Troubleshooting.
3. If step 1 & 2 are not conclusive you may need to
troubleshoot by going to a minimum configuration.
Refer to Minimum Configuration Troubleshooting
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
c. Turn the CT on (Sys On). f. If it DOES NOT then the problem is in the GPC
CAN bus.
d. Wait until the scanning software has fully loaded.
e. The error log should then have the one of the
timeout errors 11 - 4E whose parameter bytes
should show that only function controllers XRS,
ZCO, DLP, DOM, ACU, GPC, PTH, & PTV are
not responding.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
a. End the software and turn the CT off. a. End the software and turn the CT off.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
Computer
Note If you see an initialization error (Master Server Hard Power Cycle
failed - “MCU Reset Failed”), then it's a CAN Bus
1. Log out from software and doubleclick shutdown.
problem or Image Reconstruction System
(IRS) problem. View error log. 2. Turn off the switch on front of console control rack.
Maintenance Tests 3. Turn off wall box breaker.
1. Reboot the system. 4. Wait 30 seconds.
2. When the Starting up the System screen 5. Re-energize wall breaker.
appears, quickly click on Stop for Maintenance.
6. Turn on the switch on front of console control rack.
3. Run the maintenance tests.
7. Turn on power on SGI O2 Computer.
• Any failures indicate a problem with the O2 computer.
8. Turn on gantry, via the CT Box (EVA on).
See IRS Raw Data Disk Reformat.
9. Click ok on warning notice and wait several
• If there are no failures, install a spare backed-up hard
minutes for boot up.
drive and retry.
Shutdown Problem or Mouse Lockup
Rightclick on background and select Restart or Quit.
Or, press the power button to do a hard power off.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
Computer - SGI O2
The Silicon Graphics computer has diagnostics for SCSI press <Enter to continue>
devices connected directly to the computer (i.e. the
6. Press [ENTER] to allow the diagnostics to complete.
EOD) and its internal components.
7. The screen will then be a grayscale and will change
Note It does not test any of the Image Reconstruction
colors while it is checking the monitor color levels.
System components.
8. At the ide>>> prompt, Type exit and press
Note This takes approximately 30 minutes to run.
[ENTER]. The system will return to the O2 task list.
1. Restart the computer.
9. Click on Start System. The system should then
2. After a few seconds, the screen will display the boot to the normal login screen.
message Starting Up The System.
Copy the O2 9 GB Drive to an 18 GB Drive
3. Click on Stop for maintenance button. The O2 See Backup Hard Drive (xfscopy).
computer task selection screen is displayed.
4. Click on Run Diagnostics.
5. All of the diagnostics should pass with the exception
of the EOD drive. The diagnostics will display the
following error:
no valid QOUT_PTR_ARRAY
(The error may be repeated for several lines)
Device 3 failed DMA test
Error Code = IP32.08.0c.01.04,
w1=00000001, w2=00000008
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
Couch
Table
Table
won't raise
won't feed
or lower
Table will
Check that gantry is not Check transfer plate
raise but
and feed table all the way Check collision sw itch
not lower
Table still Check that table crush Check Logbook then refer to
won't raise sw itches are not stuck EPST Error Code section.
W ant or lower m aladjusted Attem pt PT_Horz cal.
power to Note: OEM must do
table only PTH calibration.
Unless Service m ode is desired,
check that PDS S1 is not O N.
Check that S703 is not O N.
Activate
Check table breakers F701-F705 .
S1 in PDC
Check PDC breakers F8, F15, F16.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
CT Box
The CT Box LEDs and lights illuminate for several
seconds on initial power up. You can use this to verify the
LEDs and lights are functioning.
The CT box has two seperate inputs:
• Top half interfaces with GPC in the Gantry Right
Pedestal
• Bottom half interfaces with CIB in the Image
Reconstruction System (IRS)
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
2. Use password letmein if needed. Note This takes less than 20 minutes.
3. Select Format SCSI. Note This clears up several system performance prob-
lems, including when the system startup bar
takes a long time to complete during boot up.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
Detector Collimator
P-Plane Collimator 7. If the motor does not rotate freely, connect a
multimeter to the motor and measure the current.
Initialization errors include:
8. If the current does not exceed 130ma, replace the
• ERR 90 00 41 dlc_current error
Delphi, otherwise, replace the motor.
• ERR 90 00 36 dlc_deviation
LEDs:
• ERR 90 02 3F dlc_move_timeout
• IDX Index pulse 1 per revolution
• ERR 90 00 42 dlc_no_move
• LIM Optical limit switch is active
To resolve the errors:
MOTOR DRIVER: (on the Delphi D515)
1. Inspect the collimator. Check for dirt or parts
24V pulse width modulated
collected on the collimator, and check for dried
grease. Clean the collimator with alcohol.
P-Plane
2. Inspect the mechanism for defects or malfunctions. Motor
3. Perform an MCU Reset, check for proper function. Encoder
4. If an errors occurs, remove the collimator motor from
its place, but leave the electrical connection to it.
5. Set the service switch on the C-Box Delphi to ON.
Try to operate the motor by pushing and releasing
the button the Delphi.
6. If the motor rotates, the problem exists on the DMS
Detector and GI Cards.
Continued on next page
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
UHR
Collimator
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
Filming
Ping the Camera Connecting a Dryview 8700
For a filming problem, try pinging the camera. If the ip address, AE Title, etc., are configured, but you
still cannot film, ensure that Kodak service is setting up
1. Rightclick on background and select New Window
to receive an AE Title of proFILM.
Root.
This is usually missed, and unless you are careful, it
2. Use password pkserv99.
looks like our AE Title to the camera is proDISK.
3. Type cd /etc and press [ENTER].
The setup for Kodak should be port 104, AE Title of
4. Type more hosts and press [ENTER]. proFILM, and the ip address.
5. Find IP address of camera.
6. Type ping x.x.x.x and press [ENTER].
7. Press [CTRL][C] to stop ping command.
8. If no ping:
• check configuration of camera and network, be
sure IP addresses match
• check for a hub or router between console and
camera
• check cables
9. Type exit and press [ENTER] to leave the window.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
Gantry Rotate
During Is the
Reset, does the Rotation Safety
NO H1 LED TS Chart
gantry move/ Switch S301
shimmy? set to 1? 1 flash CPU Error
2 flashes DC Undervoltage or EPROM
3 flashes Overcurrent
YES
4 flashes Overvoltage
Rotate the gantry by 5 flashes Motor Overload
hand or in service YES
(D301 S1 + S3)
6 flashes Inverter Overload
7 flashes Motor Temperature
8 flashes Inverter Temperature
Is the
Do the
yellow H1 LED Check three phase
D301 AP0 and
YES on the rotation NO power at the K4 relay,
AP90 LEDs light
freq converter or the Z1 filter
up?
flashing?
NO YES
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
Gantry Tilt
Cannot tilt Check Breakers
from console Does Tilt
FC have No in Gantry Left
Vac? Pedestal: F302,
Can Tilt Is scanner Toggle ICT F313, D324
from Gantry Yes in ICT Yes button on
Keypads? mode? console Yes
Reprogram
No Frequency
Converter
No
Yes
Replace
- Ck Crash Plate Frequency
Can Tilt
- Reflash RTC MCB (OEM) Converter
from RTC in
Yes - Swap MCB with another MCB
Left
Pedestal? - Replace GPC Can tilt Replace Tilt
- Replace RTC from No Motor Spindle
console? Assembly
No
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
Image Quality
Select a topic: Symptom Possible Cause
• CT Numbers Out Of Calibration (Level) Artifact: sticks in image • Something interferring with
• Troubleshooting Chart beam path
Artifact: sticks in image • Detector
• Ring Program
Artifact: streaks only in scout • Detector
CT Numbers Out Of Calibration (Level) image but not axial
1. Perform Phantom Calibrations. Bad channels • Detector
2. Perform reference detector check. • Motherboard
3. Slice A or D values shifted (one end slice). Bad image • D501 int card
• Loose data cables
4. Focal spot shifted.
• Dirt, tape, solder on
5. Perform XRT-Z-alignment Calibration. collimator port
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
• 200 475
• 13 (8) 30 (11)
Note The program has an accuracy of ±1 detector. Use
it as a first step for channel identification. Con-
tinue with step-by step reconstruction below for
more precise identification. Continued on next page
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
IRS
CE Check
1. Logout.
Caution When sliding out the Image Reconstruction
System (IRS) card rack, be sure the fiberoptic 2. At login screen, doubleclick Service.
cable connected in back has slack. Damage to
the cable and/or connector could result. 3. On the Console window, you can see the CEs
starting. At the end, it says All CEs initialized
Testing Recon and running. Or, if any are bad, it will say which
1. Make or call up a scan with known good data (and are bad.
raw data) on the IRS.
2. Click on Processing and Offline Recon.
3. Click on IRS, highlight the scan to be reconstructed.
4. Click on OK.
5. On Add Recons screen, click OK.
6. Wait about 45 seconds for recon.
If no recon happens:
• Do an IRS Raw Data Disk Reformat.
• Check CE2 - CE25 for initialization.
• Check IRS power.
• Verify you can connect to IRS by leftclicking on it in
the Study 1 - Archives Manager window.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
No No
No
No
Check connection
Replace laser
Replace GPC
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
Power
System Power • Check K4 in PDC. If thought to be faulty, move S1 to
ON. If there is motion, K4 is bad.
If the system wont power up, check UPS in the PDC.
On the back of the Computer Rack: DAS Power
• In gantry, check the DPU2 LEDs in C-Box.
• GPC light shows power at GPC in Gantry Right
Pedestal. • Check L2 power in Gantry Left Pedestal.
• CIB light shows power at the CIB board in Image • Check AC Power Brush Blocks X374 at for 230 VAC
Reconstruction System (IRS). turned on, after K9 and K10 in PDC.
Gantry Power (Stationary and Rotating) • Check F9 in PDC.
• Check L1, L2, L3 on Gantry Left Pedestal going into • UPS2 (optional equipment) could be defective.
Z301. Refer to System Power Distribution diagram.
• Check PDC 24V Power Supply in PDC.
• Check power at X257 on PDC.
Power Distribution
• Check K4 in the PDC.
• Check F200 in PDC.
• Check F8 in the PDC.
• Check F1 in PDC.
Couch Power • Check wall box.
• Check X701 connector, Undertable Components.
• Check Teal Power Conditioner (optional equipment).
• Verify all 5 green LEDs on Couch Power Supply are
lit.
• Check power supply output.
• Check F15, F16 in PDC. Continued on next page
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
Power, continued
Power Monitoring WCS Power
• Check 2 green LEDS on G1 power supply in PDC. • Check X263 in PDC for 3-phase power.
• Check X111 incoming power in PDC. • Check K6 in PDC.
• Check D451 green LED v24 in PDC. • Check F14 in PDC.
• Check fuses F1 and F2 under top cover of Teal Power
Conditioner (optional equipment).
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
3. On the laptop, select Start, Programs, 10. You should see a display similar to this:
Accessories, Communications, WWT_P20 V002 18.09.1998
HyperTerminal. =======================================================
time Tset Tact prim i_mem perc new valv 123 FP state
4. When the program window opens, click on --------!----!---!---!----!-!-----!----!---!--!-------
Hypertrm.exe. Note If you don’t see this display, try shutting the sys-
tem down and power off at the wall box for a
5. When prompted, type a name like WCS Connection
minute. Then power up the system again to
and pick any Icon. Click OK.
Standby state.
6. In the Connect To window, select Connect
11. Type STA 1 and press [ENTER]. This tells the WCC
Using: Direct to COM1. Click OK.
program to display the WCS status once every
7. For Port Settings, use: second.
• Bits per second: 9600
• Data Bits: 8
Continued on next page
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
X-Ray Detection
Select a topic: Make sure you order the correct grade when replacing
detectors.
• General
• Anode Stator Resistance Measurements
• Check 560VDC UDC_P and UDC_N
• Initialization Errors Before Scan
• GI Card Fuses Opening
• mA Errors
• Ring Artifacts
• Streaking Artifacts
Detector Detector Detector Detector
General Grade A Grade B Grade C Grade D
There are four quality grades of Detector: A, B, C, and
To remove a detector, use only a 5mm allen wrench /
D. A is the highest, and is at the center of the DMS. D is
T- wrench. Do not use the “balled” end of the allen
lowest quality, and is positioned on the outermost edges
of the DMS. wrench. It can break when removing the normally tight
screws.
For troubleshooting, the GI cards are identical and can
Note Use anti-static wipes to combat static buildup in
be swapped. The detectors can be also be swapped.
the cables between the detectors and the GI
However, do not leave a lower grade card in a higher
grade slot. cards.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
X-Ray Generation
Select a topic: Tube Arcs
• Tube Arcs In the case of a tube arc:
• Tube Filament Resistance 1. Run XRT Filament Calibration (ma cals).
• E-Box D400 Test Points 2. Run Daily Tube Conditioning.
• Anode Stator Resistance Measurements
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
UTact
Actual high tension voltage
Test area: X210 Scale: 1V=20KV
Continued on next page
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
HV_drop
High active signal which indicates either one side or both
sides have had a high voltage arc
Test area: X270 Scale: DIGITAL
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
Test area: X200 Scale: 1V=50A Test area: X260 Scale: DIGITAL
Itact_100ma
Actual tube current
Test area: X230 Scale: 1V=100ma
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
E50 00 10 (INIT) Reload XRS FW (Note: Only Philips can do this). MCB (D-400)
Verify that you are using the correct version.
E50 00 11 (INIT) D400, MCB
E50 00 12 (INIT) D400, E-Box
E50 00 13 (INIT) Perform Single Pulse test D400, E-Box, HV Tank
(Note: Only Picker can do this).
E50 00 14 (INIT) Perform Single Pulse test D400, E-Box, HV Tank
(Note: Only Picker can do this).
E50 00 15 (INIT) Check cable D400-D432 D400, HV Tank
E50 00 16 (INIT) If parameter 2=03, check cable D400-D115 D400, FIL/RAC, X-Ray Tube
E50 00 17 (INIT) D451 (PDC), D400
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
E50 00 26 Check the voltage on the generator PS (E606) attached to the GENERATOR PS-E606, D400, E-Box
E-Box. See E606 And UDC Power Supplies Table, below.
E50 00 27 Check the voltage on the generator PS (E606) attached to the GENERATOR PS-E606, D400, E-Box
E-Box. See E606 And UDC Power Supplies Table, below.
E50 01 30 (First Check x-ray tube filament for resistance. FIL/RAC
Reset) See Tube Filament Resistance.
E50 01 32 If the x-ray tube anode bearing is very noisy, replace the x-ray FIL/RAC, X-Ray Tube
tube
E50 01 34 If parameter 1=02, the supplied DC voltage from the stationary F100 Fuses in PDC, 560VDC PS in PDC
part is too low. Check 560VDC F100-Fuses (80A in PDC unit)
E50 00 36 (INIT) E-Box
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
NO
PERFORM DOES
LONG TUBE YES ARCING
CONDITIONING OCCUR WITH ERROR
FROM HOST E 50 00 2F ?
NO
YES
GOTO
FLOWCHART
2
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Troubleshooting
MEASURE
REPLACE FILL/FPC ANODE
STARTER
RESISTANCE
REPLACE
FIL/RAC
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 How to Use this EPST
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 How to Use this EPST
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 How to Use this EPST
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 How to Use this EPST
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 How to Use this EPST
Printing
To print, perform the following actions.
1. On the Acrobat toolbar, select File, Print.
2. Select whether to print all pages, only the current
page, or a range of pages. (Default is Print All
Pages.)
3. Enable the "Shrink to Fit" option when printing the
11" x 17" block diagrams on 8.5" x 11" paper.
4. Click on OK to begin printing.
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 How to Use this EPST
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Self-Paced Training Modules
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Reference Materials
REFERENCE MATERIALS
Please select a document:
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Reference Materials
Certified Components
Component Model Number Serial Number
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Reference Materials
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Reference Materials
Tuc tube collimator box MCU Master Control Unit: coordinates gantry control
functions (e.g. state synchronization) and reports
Wcs water cooling system results to ICS
Wcc water chiller controller PHS Patient Handling System: horizontal and vertical move
of the patient table and tilt control of the gantry
Xrs x-ray system
PTH Patient Table Horizontal: control of the horizontal move
Zcon z-axis control of the patient table
ACU Acquisition Control Unit: coordinates scan control of PTV Patient Table Vertical: control of the vertical move of
rotating part the patient table
COC Collimator Control, control of tube collimator ROT Rotation Control
DLP Detector collimator, Light-marker and Phi control RTC Rotation and Tilt Control
DOM Dose Modulation (optional) XRS X-Ray Rotating System, Rotating part of the X-Ray
GPC Gantry Panel Control generator
ICS Image Control System, control computer of the whole ZCO Z-Control
CT scanner with interface to the acquisition system
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Reference Materials
Safety Precautions
Select a topic:
• Mechanical Hazards
• Electrical Hazards
• Body Protectors
• Radiation Protection
• Bloodborne Pathogens
• Working with System Components
• Torquing Screw Connections
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Reference Materials
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Reference Materials
• The Computer Rack and the Operating Console – UDC_pos and UDC_neg on right Gantry pedestal
receive power from an Uninterruptible Power Supply – UDC_pos and UDC_neg on Slip Ring Gantry rear
(UPS) located in the Power Distribution Cabinet
(PDC). This means that parts inside these units – UDC_pos and UDC_neg on generator electronic
continue to receive line voltages, even after the box
console or the various circuit breakers are switched
off. Disconnect the Rack and Console power cables
before accessing these units.
Continued on next page
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Reference Materials
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Reference Materials
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Reference Materials
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Reference Materials
in-lb. ft-lbs Nm in-lb. ft-lbs Nm 280 23.3 31.6 390 32.5 44.1
10 0.8 1.1 120 10.0 13.6 290 24.2 32.8 400 33.3 45.2
12 1.0 1.4 130 10.8 14.7 300 25.0 33.9 410 34.2 46.3
20 1.7 2.3 140 11.7 15.8 310 25.8 35.0 420 35.0 47.5
30 2.5 3.4 150 12.5 16.9 320 26.7 36.2 430 35.8 48.6
40 3.3 4.5 160 13.3 18.1 330 27.5 37.3 440 36.7 49.7
50 4.2 5.6 170 14.2 19.2 340 28.3 38.4 450 37.5 50.8
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training
GE Medical Systems Philips MX8000 Feedback to Author
FEEDBACK TO AUTHOR
If you have comments or suggestions, please email them
to:
EPST.feedback@med.ge.com
Alignment & Component Error Functional Logic & Block Preventive Software & Theory Trouble- Self-Paced Notes
Calibration Locations Codes Checks Diagrams Maintenance Diagnostics shooting Training