FEATURES
SYSTEM CONFIGURATION
Standard Configuration
Unit Model Name Remarks
C-arm Main Body KMC-950 C-arm Unit
Standard monitor cart
Digital Imaging System
19” B/W 2 monitors
Manuals
Sterilized covers (1set of 3pieces for the I.I,
X-ray tube and C-arm main body)
Foot switch
Laser guide
Option
Unit
D.S.A (Digital Subtraction Angiography)
D.A.P (Dose Area Product)
Virtual Collimator
Thermal Printer – Sony UP897MD
Thermal Printer – Sony UP990AD
Cassette Holder (For Radiography mode)
Grid
Hand switch
INSTALLATION CONDITIONS
System Power Requirements
- Line voltage 100-120 VAC 50/60 Hz : for UL
220-230 VAC 50/60 Hz
- Line phase Single Phase
- Input power 3600VA @120 VAC: for UL
12500VA @ 230 VAC
- Line code 26.3feet
- Line plug 250VAC, 63A LEGRAND NO.LV-63A-58700
Grounding Requirements
Grounding must be provided in compliance with all applicable legal requirements for medically used electrical
equipment.
Ambient Conditions
Operation)
- Ambient Temperature 10°C~40°C
- Relative Humidity 30~75%
- Atmospheric pressure 70~106kPa
Transportation & Storage)
- Ambient Temperature -40°C~70°C
- Relative Humidity 10~85%
- Atmospheric pressure 50~106kPa
COMPLIANCE
EN 60601-1: 1990 + A1 + A2 + A12 +A13
EN 60601-1-3: 1994
EN 60601-2-7: 1998
EN 60601-2-28: 1993
EN 60601-1-2: 2001
KMC-950
SPECIFICATION
www.gemss-medical.com