Model : LUC-8M
OBELIS S.A
Bd. General Wahis, 53
1030 Brussels, Belgium
TEL : +32-2-732-59-54 FAX : +32-2-732-60-03
DOC no RD – 120831 - 01
Date 2020. 07. 22
2460 259745
Rev -4-
Rotating Anode X-ray Tube Ass’y Spec. LUC-8M
GENERAL SPECIFICATION
1. Product Information
Page 2
Rotating Anode X-ray Tube Ass’y Spec. LUC-8M
1.3 Classification
2. General Data
2.1 Electrical Data
Focus 50 ㎐ 60 ㎐
Large 43 ㎾ 47 ㎾
Small 21 ㎾ 22.5 ㎾
Page 3
Rotating Anode X-ray Tube Ass’y Spec. LUC-8M
Operating Limits
Temperature .................................................................................... 10 ∼ 40 ℃
Humidity ........................................................................................... 30 ∼ 85 %
Atmospheric Pressure .................................................................. 70 ∼ 106 kPa
Shipping and Storage Limits
Temperature ................................................................................... -20 ∼ 70 ℃
Humidity ........................................................................................... 20 ∼ 90 %
Atmospheric Pressure .................................................................. 50 ∼ 106 kPa
Page 4
Rotating Anode X-ray Tube Ass’y Spec. LUC-8M
Page 5
Rotating Anode X-ray Tube Ass’y Spec. LUC-8M
400
200
300
150
200
100
100
50
0 0
0.01 0.03 0.05 0.1 0.2 0.3 0.5 1 2 3 5 10 0.01 0.03 0.05 0.1 0.2 0.3 0.5 1 2 3 5 10
80kV 100kV
40kV 40kV
300 500
110kV
90kV
125kV
250 100kV
Tube Current (㎃)
400
110kV
200 125kV
300
150
200
100
100
50
0 0
0.01 0.03 0.05 0.1 0.2 0.3 0.5 1 2 3 5 10 0.01 0.03 0.05 0.1 0.2 0.3 0.5 1 2 3 5 10
Page 6
Rotating Anode X-ray Tube Ass’y Spec. LUC-8M
500 16
400 14
300 12
200 10
100 8
0 6
4.2 4.3 4.4 4.5 4.6 4.7 4.8 4.9 5.0 5.1
Filament Current (A)
300 14
Filament Voltage (V)
Tube Current (㎃)
250 12
200 10
150 8
100 6
50 4
0 2
4.2 4.3 4.4 4.5 4.6 4.7 4.8 4.9 5.0 5.1
Filament Current (A)
Page 7
Rotating Anode X-ray Tube Ass’y Spec. LUC-8M
6. Thermal Characteristics
6.1 Anode Heating and Cooling Curves
100
90
80
Heat Storage (kJ)
70
60
50
40
30
20
10
0 1 2 3 4 5 6 7 8 9 10
Time (minutes)
800
700
Heat Storage (kJ)
600
500
400
300
200
100
Time (minutes)
Page 8
Rotating Anode X-ray Tube Ass’y Spec. LUC-8M
Anode
Large Small
Cathode
Common
Central X-ray
Terminal Connections
9 8 7 6 5 1 2 3 4 E
Focal Spot
C M A Earth
Non-connection Terminal
Center of Focal Spot
Temperature Relay
Non-connection
(Normally Closed)
A
C : Common
M : Main Winding of the Stator
A : Auxiliary Winding of the Stator
B
[Note] Do not connect terminal No.1 and No.5 or 6
in series circuit.
A : ±1.5㎜, B : ±1.5㎜
Page 9
Rotating Anode X-ray Tube Ass’y Spec. LUC-8M
8. Symbols
8.1 Symbols marked on LUC-8M
Symbol Description
CE Marking of Conformity
2460
CSA Marking
259745
Page 10
Rotating Anode X-ray Tube Ass’y Spec. LUC-8M
Symbol Description
Page 11
Rotating Anode X-ray Tube Ass’y Spec. LUC-8M
9. Label Location
9.1 Front
Page 12
Rotating Anode X-ray Tube Ass’y Spec. LUC-8M
9. Label Location
9.2 Rear
Page 13
Rotating Anode X-ray Tube Ass’y Spec. LUC-8M
10. Cautions
1) Caution for Installation, Adjustment and Maintenance
• Qualified Persons : Only qualified persons who have technical training and
professional knowledge can handle X-ray tube assembly.
• Fragile Glass : X-ray tube is assembled with glass, therefore, it can be broken with
the mechanical vibration or pulsed shock over 19.6m/s 2 (2G).
Careful handling is required to treat or transport.
• Ground Terminal : X-ray tube assembly has ground terminal. Ground cable should be
connected.
• High Voltage : All X-ray tubes operate at voltages high enough to kill through electric
shock. Never touch the high voltage delivered plugs or terminals.
When direct access to such parts is required, the primary circuit should be disabled
and high voltage capacitors/cables discharged.
• High Voltage Plug : High voltage plug should be cleaned up and free from any physical
damages. Silicon compound application is required for high voltage stability.
• Rotating Anode
- Test rotation of the anode with the large focus filament at the preheat condition.
At this time, do not supply the tube voltage.
- When the hand switch “READY” button is pressed, the filament will become brighter
and the anode target will start rotating. The anode for normal speed will reach the
predetermined number of revolution in 0.4 to 1 sec., and it will continue rotation for
some time after the stator drive input is cut off. The anode for high speed shall be
automatically braked to reduce speed and continue coasting after the drive input to
the stator is cut off.
- Ensure that the anode target rotates in the proper direction. X-ray tube should always
rotate counter-clockwise, viewing from the cathode side.
• Operation Atmosphere : X-ray tube assembly is not allowed to use in the atmosphere
of flammable or corrosive gas.
• Protective Cover : X-ray tube assembly is not allowed to use without the protective cover
• Handling : Appropriate jigs or tools are required for tube installation to avoid physical
damages.
• Returning Tube : X-ray tube assembly should be repacked with the original material
when it is returned back for quality examination in our factory. Be careful to put the tube
upside cathode. If the packaging is not proper, the tube may not be correctly examined.
Page 14
Rotating Anode X-ray Tube Ass’y Spec. LUC-8M
• Safety Plate : The Safety Plate is mounted on anode housing wall. When external
cover is removed, it is visibly recognized as labeled with label 8 in Caution Labels page.
Followings are important notice of the Safety Plate:
- Function : When housing inner pressure reaches to the working pressure of the
Safety Plate, the Safety Plate collapses and vaporized oil and hot oil flow through
the collapsed Safety Plate. The external cover can protect the initial impact of oil flow,
however, it can not protect oil drop or oil leakage.
- Caution : Never touch the Safety Plate or remove the label and the screws.
Otherwise, it could cause collapse of Safety Plate.
Please note the Safety Plate is not the device to protect X-ray tube from destruction
when over-rated power input but just to reduce the impact of housing rupture.
• Unexpected Malfunction : X-ray tube assembly may have the risk to be unexpectedly
malfunctioning due to life termination or failure. If the serious problems caused by
the above risk are expected, you are requested to have a contingency plan to avoid
such a case.
• New Application : If you use the product with new application not to be mentioned
in this specification or with different type of X-ray generator, please contact to us for
confirming its availability.
Page 15
Rotating Anode X-ray Tube Ass’y Spec. LUC-8M
3) Periodical Inspection
To maintain long-time stable operation and service life, perform the periodical inspection
at every regular intervals (at least once a year) in accordance with the items described
on the following page.
For any inspection items, if necessary, instruct LUCEM Co.. Ltd. personnel to carry out
the periodical always operation the silicone compound may become dry, the coated
state changed or the withstand voltage performance degraded.
4) Others
• X-ray tube ass’y should have the X-ray limiting equipment mounted on the X-ray port
to protect unnecessary radiation.
• X-ray tube ass’y is not allowed is not allowed to use in the atmosphere of flammable
or corrosive gas.
• If the input power exceeds X-ray tube ass`y specification, it cause the over temperature
of anode, insert tube glass shatter, housing sealing parts rupture, human injury
including burns due to hot oil escape and fire accident due to flaming anode target.
The X-ray generator should have a protective function which manages input power
to be within X-ray tube ass’y specification.
• X-ray tube ass’y is covered with lead inside tube housing to protect unnecessary
radiation. The lead is harmful for human health.
Dispose the scrapped products according to the requirement of local regulation.
Page 16
Rotating Anode X-ray Tube Ass’y Spec. LUC-8M
• X-ray tube ass’y has insulation oil contained for high voltage stability.
The insulation oil is poisonous for human health.
If it is exposed to the non-restricted area, it should be disposed as following to the
local regulation.
• According to the operating conditions, the filament characteristics might be changed
to lead to the over rate exposure to X-ray tube ass’y.
To prevent the X-ray tube ass’y from being damaged, please adjust the tube current
regularly.
• Please don’t touch on X-ray tube ass’y just after operation due to high temperature.
• X-ray tube assembly is not allowed to lift more than 1 meter during transportation.
• X-ray tube assembly is not allowed to stack more than 2 stories.
• If any malfunction of the X-ray tube ass’y is noticed, please contact to us immediately.
• Avoid fire and explosion hazard. An explosion hazard may result from the use of
flammable anesthetics, skin cleansing agents and disinfectants.
• When cleaning the device, pay special attention so that cleaning liquid or any liquid may
not soak into the inside of device. If any liquid soak into the inside of device, do not
connect power line, do not use device, and contact service center.
• This equipment produces ionizing radiation. Observe proper safety practices during
operation.
• The requirement to wear personal protective equipment such as lead aprons,
dosimetry badges etc is governed by individual states and may vary from state to state.
• Other safety information follows the product-mounting system information.
• The tube assembly contains a high-vacuum product that can implode on incorrect
handling, such as by impact, striking, dropping down.
• Use of the X-ray tube assembly is forbidden if signs of mechanical, electrical or
radiation-related defects are present.
• “WARNING: To avoid the risk of electric shock, this equipment must only be connected
to a supply mains with protective earth.”
Page 17